CR10XTCR THERMOCOUPLE REFERENCE

Similar documents
CR10XTCR THERMOCOUPLE REFERENCE

CR10TCR THERMOCOUPLE REFERENCE

INSTRUCTION MANUAL. CR10XTCR Thermocouple Reference. Revision: 1/03. Copyright (c) Campbell Scientific, Inc.

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

3WHB10K 3-WIRE HALF BRIDGE TERMINAL INPUT MODULE INSTRUCTION MANUAL

IRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL

3WHB10K 3-Wire Half Bridge Terminal Input Module

MODEL HMP45C TEMPERATURE AND RELATIVE HUMIDITY PROBE INSTRUCTION MANUAL

INSTRUCTION MANUAL. Model 207 Temperature and Relative Humidity Probe. Revision: 2/94. Copyright (c) Campbell Scientific, Inc.

CM3 PYRANOMETER INSTRUCTION MANUAL

SP-LITE SILICON PYRANOMETER INSTRUCTION MANUAL

WARRANTY AND ASSISTANCE

Model 108 Temperature Probe Revision: 4/07

CVD20 20:1 Voltage Divider Revision: 5/12

MET ONE RAIN GAGE MODELS 380, 385, 380M, 385M

INSTRUCTION MANUAL. SP Lite2 Pyranometer. August Copyright 2017 Campbell Scientific (Canada) Corp.

INSTRUCTION MANUAL. CS547 Conductivity and Temperature Probe and A547 Interface. Revision: 9/00. Copyright (c) Campbell Scientific, Inc.

Model ASPTC Aspirated Shield with Fine Wire Thermocouple

3WHB10K 3-Wire Half Bridge Terminal Input Module

INSTRUCTION MANUAL. LI200X-LC Pyranometer for MetData1 4/97. Copyright (c) Campbell Scientific, Inc.

Eppley PIR Precision Infrared Radiometer

INSTRUCTION MANUAL. CS500 Temperature and Relative Humidity Probe. Revision: 7/04. Copyright (c) Campbell Scientific, Inc.

227 Delmhorst Cylindrical Soil Moisture Block Revision: 5/13

INSTRUCTION MANUAL. CSI Model TB4 Rain Gage. Revision: 8/04. Copyright (c) Campbell Scientific, Inc.

WXT510 Weather Transmitter Revision: 4/07

IRTS-P Precision Infrared Temperature Sensor

NOVALYNX CORPORATION MODEL 110-WS-16BP BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL

Small Gage Pressure Sensor

INSTRUCTION MANUAL. CS100 Barometric Pressure Sensor 3/05. Copyright (c) Campbell Scientific, Inc.

Small Absolute Pressure Sensor

Model HFP01 Soil Heat Flux Plate Revision: 7/12

Small, Gauge Pressure Sensor

OEM Silicon Pressure Die

User s Manual: MODEL 904-T CLINOMETER PAK

SM98A Harsh Media Backside Absolute Pressure Series

HFM-6 and HFM-7 OPERATOR S MANUAL

IntraSense TM Series 1-French Wire-Connected Pressure Sensor SMI-1A AAUU

January price list. general information

LUDLUM MODEL 42-5 NEUTRON BALL CART. August 2015 Serial Number PR and Succeeding Serial Numbers

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

Low Pressure Sensor Amplified Analog Output SM6295-BCM-S

CAUTION: BEFORE YOU BEGIN, ALWAYS MAKE SURE THE WEAPON IS UNLOADED AND SAFE TO HANDLE.

Program Notes Use 2:1 voltage divider on Differential Input channel 1 to get full output range from C7

INSTRUCTION MANUAL FOR MM10 Relay

TRUMPET REFERENCE Vacuum Tube Phonostage

Field Pro Operation and Installation Manual

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

INSTRUCTION MANUAL. Model HFP01SC Self-Calibrating Soil Heat Flux Plate Revision: 3/14. Copyright Campbell Scientific, Inc.

HFM-6, HFM-7, & HFM-8 OPERATOR S MANUAL

VIBRATION METER. Instruction Manual.

IntraSense TM Series 1-French Wire-Connected Pressure Sensor Standard IntraSense TM Families

WARRANTY AND ASSISTANCE

INSTRUCTION MANUAL. CS616 and CS625 Water Content Reflectometers Revision: 2/14. Copyright Campbell Scientific, Inc.

4WFB120, 4WFB350, 4WFB1K 4 Wire Full Bridge Terminal Input Modules Revision: 5/07

Using CSIM11 ph and ORP Probes with Campbell Scientific Dataloggers

Medium Pressure Sensor Analog Output

Series FEATURES AND BENEFITS INTRODUCTION. Figure 1 Power Derating Curve. Table 1 Tolerance and TCR vs. Resistance

Franck-Hertz Experiment

Installation Manual. Models 4911A/4911. VW Rebar Strain Meters

e/m APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model SE D 5/ PASCO scientific $5.

TEMPERATURE SENSOR. Model TH-T. Roctest Limited, All rights reserved.

IR/Thermistor Pen. Instruction Manual

Thermocouple TC-K Sensor 0 C to 1200 C

Today s Weather 24 Hour Forecaster

Instruction Manual. DTP Series DC Termination Panels

Avery Dennison LW Series Retroreflective Window Sticker Film Issued: July 2013

Protective Coatings: Non-lead orange paint RAL red (optional) non-lead paint Hot dipped galvanized conforming to ASTM A-153.

FHR

INSTRUCTION MANUAL ANALOG COLORIMETER

RAYTECH USA, Inc. 90C Randall Avenue, Woodlyn, PA USA Tel Fax Instruction Manual

INSTRUCTIONS FOR ATTACHING THE NORDIC PLOW TO THE ACTION TRACKCHAIR

INSTRUCTION MANUAL. 229 Heat Dissipation Matric Water Potential Sensor Revision: 5/09. Copyright Campbell Scientific, Inc.

SkyScan EWS-PRO - Manual -

Automatic and Manual Safety Pilot Kits

Bulk Metal Foil Technology Industrial Grade Miniature Voltage Divider with TCR Tracking of 1.5 ppm/ C and Ratio Stability of 0.

TB-2012M BAROMETRIC PRESSURE SENSOR USER S MANUAL

TEMPERATURE COMPENSATION FOR MEAS PRESSURE SENSORS APPLICATION NOTE

WELDED STEEL PIPE NIPPLE PRICE SHEET

1450 First Ave. Page 1 of 10 Chippewa Falls, WI (800) or (715) Fax: (715)

WHEN BRIGHTNESS COUNTS

Exercise 1: Thermocouple Characteristics

Bulk Metal Foil Technology Power and Current Sensing Resistors with TCR of 2 ppm/ C, Tolerance to ± 0.01 % and power up to 10 W

CHROMA Colorimeter Model 252

WELDED STEEL PIPE NIPPLE PRICE SHEET

Mini Environmental Quality Meter

Instruction Sheet COOL SERIES DUCT COOL LISTED H NK O. PR D C FE - Re ove r fro e c sed rea. I Page 1 Rev A

FIELD GUIDE FOR HORUS SCOPES

M2 SERIES THERMOSTATS 0 F to 240 F, Narrow Differential, Hermetically Sealed ½

FEATURES MAXIMUM - 55 C MAXIMUM

Product Data Sheet PD-0036-B. 3M PC Boardmount Plug and Receptacle, 2 mm 95X Series

IF YOU NEED ANY FURTHER HELP WITH YOUR GALILEO TELESCOPE PLEASE CALL MON-FRI 9AM - 5PM E.S.COM

CAUTION: BEFORE INSTALLATION, CONFIRM THAT THE CROSSBOW IS UNLOADED AND IS IN SAFE OPERATING CONDITION.

BMXART0814 analog input module M340-8 inputs - temperature

Temperature Profile for 36 Marker Spheres on 3M 788 ACCR Conductor 3M Company

AN2970 Application note

3M Pak 50 Boardmount Socket, P50 series P50-XXXS-XXX-EA. 3M Pak 50 Boardmount Plug, P50 series P50-XXXP-XXX-EA

MKIII -MB USER S GUIDE 09/18/18. Rev. 3

Transcription:

CR10XTCR THERMOCOUPLE REFERENCE REVISION: 01/01/03 COPYRIGHT (c) 1995-2003 CAMPBELL SCIENTIFIC, INC.

This is a blank page.

WARRANTY AND ASSISTANCE The CR10XTCR THERMOCOUPLE REFERENCE is warranted by CAMPBELL SCIENTIFIC (CANADA) CORP. ( CSC ) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. CSC's obligation under this warranty is limited to repairing or replacing (at CSC's option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CSC. CSC will return such products by surface carrier prepaid. This warranty shall not apply to any CSC products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CSC is not liable for special, indirect, incidental, or consequential damages. Products may not be returned without prior authorization. To obtain a Return Merchandise Authorization (RMA), contact CAMPBELL SCIENTIFIC (CANADA) CORP., at (780) 454-2505. An RMA number will be issued in order to facilitate Repair Personnel in identifying an instrument upon arrival. Please write this number clearly on the outside of the shipping container. Include description of symptoms and all pertinent details. CAMPBELL SCIENTIFIC (CANADA) CORP. does not accept collect calls. Non-warranty products returned for repair should be accompanied by a purchase order to cover repair costs.

CR10XTCR Thermocouple Reference 1. General The Model CR10XTCR Probe contains a thermistor for measuring temperature and is used as a reference temperature for thermocouples measured with the CR10X. It is installed between the two terminal stripes in the wiring panel. After all the thermocouples are wired, the wiring panel cover is installed to minimize temperature gradients on the terminal strip. 1.1 Specifications Temperature Measurement Range: -35 to +50 C Thermistor Interchangeability Error: Typically <±0.2 C over 0 C to 60 C ±0.4 @ -35 C. Polynomial Linearization Error: <±0.5 C over -35 C to +50 C. 2. Accuracy Temperature Sensor The overall probe accuracy is a combination of the thermistor's interchangeability specification, the precision of the bridge resistors, and the polynomial error. In a "worst case" all errors add to an accuracy of ±0.4 C over the range of -24 to 48 C and ±0.9 C over the range of -38 C to 53 C. The major error component is the interchangeability specification of the thermistor, tabulated in Table 1. For the range of 0 to 50 C the interchangeability error is predominantly offset. The bridge resistors are 0.1% tolerance with a 10 ppm temperature coefficient. Polynomial errors are tabulated in Table 2 and plotted in Figure 1. TABLE 1. Thermistor Interchangeability Specification Temperature ( C) Temperature Tolerance (± C) 40 0.40 30 0.40 20 0.32 10 0.25 0 to +50 0.20 TABLE 2. Polynomial Error -40 to +56 <±1.0 C -38 to +53 <±0.5 C -24 to +48 <±0.1 C 1

CR10XTCR Thermocouple Reference FIGURE 1. Error Produced by Polynomial Fit to Published Values 3. Installation and Wiring The wiring diagram for the CR10XTCR is shown in Figure 1. The probe requires a single-ended analog input channel and an excitation channel. After the CR10XTCR and thermocouple are connected to the datalogger, install the wiring panel cover (Figure 2 and 3). Analog Ground; Clear (AG) CR10XTCR LT#343403 Switched Excitation; Black (E3) Single-Ended Input; Red (1H) FIGURE 2. Wiring Diagram 2

DEF SE DEF LOGAN, UTAH 1 2 1 7 8 4 3 4 2 5 5 6 3 6 AG G G G SW 12V CTRL SW 12V G 12V POWER IN CR10X WIRING PANEL MADE IN USA CR10XTCR Thermocouple Reference G G H L AG H L AG E3 AG G G EARTH CROUND SE 9 10 L AG H 11 12 G G H L AG H L AG H L AG E3 AG G SDM G G H L AG H L AG H L AG E3 12V 12V WIRING PANEL NO. FIGURE 3. Wiring Panel, CR10XTCR, and Wiring Panel Cover Logan, Utah Clear SW 12 SE DIFF 7 8 9 10 11 12 4 5 6 G G H L AG H L AG H L AG E3 AG G G 5V 5V G SW G Red CR10XTCR LT#343403 Black DIFF SE 1 2 3 4 5 6 1 2 3 G G H L AG H L AG H L AG E1 AG E2 G P1 G P2 G C8 C7 EARTH GROUND TYPE TYPE T THERMOCOUPLE T FIGURE 4. CR10XTCR Installed on the CR10X Wiring Panel 4. Programming Instruction 11 is used to measure the CR10XTCR. Instruction 11 provides AC excitation, makes a single ended voltage measurement, and calculates temperature with a fifth order polynomial. A multiplier of 1.0 and an offset of 0.0 yields temperature in Celsius. For Fahrenheit, use a multiplier of 1.8 and an offset of 32. CAUTION The reference temperature for thermocouples must be in degrees Celsius. 3

CR10XTCR Thermocouple Reference EXAMPLE 1. Sample CR10X Instructions 01: Temp 107 Probe (P11) 01: 1 Rep 02: 1 IN Chan 03: 3 Excite all reps w/exchan 3 04: 15* Loc [:Ref_Temp] 05: 1 Mult 06: 0 Offset * Proper entries will vary with program and input location. 5. Instruction 11 Details Understanding the details in this section are not necessary for general operation of the CR10XTCR with Campbell Scientific's dataloggers. Instruction 11 outputs a precise 2 VAC excitation and measures the voltage drop due to the sensor resistance (Figure 5). The thermistor resistance changes with temperature. Instruction 11 calculates the ratio of voltage measured to excitation voltage (Vs/Vx) which is related to resistance, as shown below: Vs/Vx = 1000/(Rs+249000+1000) where Rs is the resistance of the thermistor. See the measurement section of the datalogger manual for more information on bridge measurements. Instruction 11 then calculates temperature using a fifth order polynomial equation correlating Vs/Vx with temperature. The polynomial coefficients are given in Table 3. The polynomial input is (Vs/Vx)*800. Resistance and datalogger output at several temperatures are shown in Table 4. Black Red Clear FIGURE 5. CR10XTCR Reference Temperature Schematics 4

CR10XTCR Thermocouple Reference TABLE 3. Temperature, Resistance, and Datalogger Output Temperature C Resistance OHMS Output C -40.00 4067212-39.18-38.00 3543286-37.55-36.00 3092416-35.83-34.00 2703671-34.02-32.00 2367900-32.13-30.00 2077394-30.18-28.00 1825568-28.19-26.00 1606911-26.15-24.00 1416745-24.11-22.00 1251079-22.05-20.00 1106485-20.00-18.00 980100-17.97-16.00 869458-15.95-14.00 772463-13.96-12.00 687276-11.97-10.00 612366-10.00-8.00 546376-8.02-6.00 488178-6.05-4.00 436773-4.06-2.00 391294-2.07 0.00 351017-0.06 2.00 315288 1.96 4.00 283558 3.99 6.00 255337 6.02 8.00 230210 8.04 10.00 207807 10.06 12.00 187803 12.07 14.00 169924 14.06 16.00 153923 16.05 18.00 139588 18.02 20.00 126729 19.99 22.00 115179 21.97 24.00 104796 23.95 26.00 95449 25.94 28.00 87026 27.93 30.00 79428 29.95 32.00 72567 31.97 34.00 66365 33.99 36.00 60752 36.02 38.00 55668 38.05 40.00 51058 40.07 42.00 46873 42.07 44.00 43071 44.05 46.00 39613 46.00 48.00 36465 47.91 50.00 33598 49.77 52.00 30983 51.59 54.00 28595 53.35 56.00 26413 55.05 58.00 24419 56.70 60.00 22593 58.28 5

CR10XTCR Thermocouple Reference TABLE 4. Polynomial Coefficients Coefficient Value C0-53.4601 C1 90.807 C2-83.257 C3 52.283 C4-16.723 C5 2.211 6. Electrically Noisy Environments AC power lines can be the source of electrical noise. If the datalogger is in an electronically noisy environment, the CR10XTCR temperature measurement should be measured with the AC half bridge (Instruction 5) with the 60 Hz rejection integration option on the CR10X (see Section 13 of the datalogger manual for more information on noise). Instruction 11's fast integration will not reject 60 Hz noise. Example 2. Sample CR10X Instructions Using AC Half Bridge 01: AC Half Bridge (P5) 01: 1 Rep 02: 22 7.5 mv 60 Hz rejection Range 03: 1 IN Chan 04: 3 Excite all reps w/exchan 3 05: 2000 mv Excitation 06: 15* Loc [:Ref_Temp] 07: 800 Mult 08: 0 Offset 02: Polynomial (P55) 01: 1 Rep 02: 15* X Loc Air Temp 03: 15* F(X) Loc [:Ref_Temp] 04: -53.46 C0 05: 90.807 C1 06: -83.257 C2 07: 52.283 C3 08: -16.723 C4 09: 2.211 C5 * Proper entries will vary with program and input location assignments. 6