Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Similar documents
Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

Organochlorine Pesticides by GCxGC-ECD

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Rtx -CLPesticides Columns Reduce Analysis Time; Reduce System Maintenance

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Introduction to Gas Chromatography

U.S. EPA Method 8270 for multicomponent analyte determination

The Polyarc System Product Brochure. A Breakthrough Innovation Revolutionizing GC/FID Technology

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

Evaluating CLP and EPA Methods for Pesticides in Water Using Agilent J&W DB-CLP1/DB-CLP2 GC Columns

Activity in the FID Detection Port: A Big Problem if Underestimated

Author. Abstract. Introduction

Understanding Gas Chromatography

Yokogawa Analytical Systems Inc., Nakacho, Musashino-shi, Tokyo , Japan.

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Application Note. Abstract. Authors. Introduction

SUPELCO. Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns

Application. Gas Chromatography February Introduction

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Introduction to Capillary GC

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

Eliminating the Need for Standards and Calibration in GC/FID Analysis of Hydrocarbons and Oxygenates

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Introduction to Capillary GC

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

2401 Gas (liquid) Chromatography

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Chemistry Instrumental Analysis Lecture 27. Chem 4631

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

GC Analysis of Polybrominated Flame Retardants Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Trajan SGE GC Columns

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

STANDARD OPERATING PROCEDURES

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Fast, Accurate Environmental Testing

Study of Residual Solvents in Various Matrices by Static Headspace

Fast and Accurate Analysis of PBDEs in a Single Run, Including BDE-209

Fast and Accurate Analysis of PBDEs in a Single Run, Including

The Analysis of Organophosphate Pesticides by LC/MS Application

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Residual solvents analysis using an Agilent Intuvo 9000 GC system

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Chapter 31 Gas Chromatography. Carrier Gas System

Sulfotepp impurities in Chlorpyrifos EC formulations

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Application Note. Abstract. Authors. Introduction. Food Safety

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Volatile organic compounds (VOCs):

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

METHOD 8082A. POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY

Application. Gas Chromatography March 1998

A Method for Simultaneous Analysis of 174 Pesticides in Grape Using GC-MS/MS

Secrets of GC Column Dimensions

Analysis of BTEX in Natural Water with SPME

Evaluation of Capillary Columns for General Performance Parameters

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Solid Phase Micro Extraction of Flavor Compounds in Beer

Application of gas chromatography atmospheric pressure chemical ionization mass spectrometry for analysis of contaminants in environmental samples

METHOD DETERMINATION OF CHLORINATED PESTICIDES, HERBICIDES, AND ORGANOHALIDES BY LIQUID-SOLID EXTRACTION AND ELECTRON CAPTURE GAS CHROMATOGRAPHY

Transcription:

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor Application Note Pesticides Authors Charlie Spanjers and Paul Dauenhauer University of Minnesota, Twin Cities Amundson Hall 421 Washington Ave SE Minneapolis, MN 55455 Andrew Jones Activated Research Company 7561 Corporate Way Eden Prairie, MN 55344 Andrew.Jones@activatedresearch.com Abstract The quantification of pesticides in food using gas chromatography (GC) is traditionally a time-consuming process that involves the painstaking calibration of detector response for each analyte. The conversion of analytes to methane before their detection with a flame ionization detector (FID) results in a response that is proportional to the number of carbon atoms in the analytes and thereby eliminates the need for detector calibration. In this note, we use this approach and show the application of the Polyarc reactor to the quantification of dilute pesticides. The response factors of compounds in a commercial 22-component organochlorine pesticide mixture (200 µg/ml of each component) were 1.00 ± 0.09 with an average deviation from unity of 4%. This is compared to an FID-only analysis with response factors of 0.83 ± 0.10 with an average deviation from unity of 17%. Additional testing of the Polyarc reactor for six single-analyte pesticide solutions prepared from pure components provided response factors of 1.00 ± 0.04 with a mean deviation of 2%. Because all compounds have a response factor of one when using the Polyarc reactor, calibration to determine response factor is no longer required. We also show that direct-connect splitless liners prevent discrimination of analytes in the GC injector port. Introduction Quantification of pesticides and other molecules by GC/FID is often a time-consuming process because the response factors for each analyte must first be determined. Response factors (RF) are typically defined as: area 1 mol C1 RF (mol % C) =, area 2 mol C2 where (1) and (2) are the analyte and internal standard, respectively, area is the integrated GC/FID peak area (i.e., the integrated detector response), and mol C are the injected moles of carbon of the component (the concentration of the component in terms of carbon content in the sample could also be used). Response factors in GC/FID analyses are dependent on the chemical structure of the molecule. The presence of heteroatoms, such as O, N, P, S, and Cl, change the response of given analyte in an FID detector. The Polyarc reactor eliminates the differences in FID response per carbon atom by converting all organic compounds to methane before their detection in the FID: Since all carbon-containing compounds are converted to methane, the response of the FID is equivalent for all molecules on a per-carbon atom basis. 1

Experimental An Agilent 7890A GC equipped with a split/splitless inlet and a Polyarc reactor (ARC PA-RRC-A02) was used for the analysis. Air (zero grade, Praxair) and H 2 (99.999%, Praxair) were supplied to the FID and to the ARC manual flow control module (PA-CAS-A07). Helium (99.999%, Praxair) was used as the carrier gas. The system was configured with the column connected from the split/splitless inlet to the Polyarc /FID and to the FID-only. An Agilent G1544-80700 direct connect liner was used, and the inlet was operated in splitless mode. A commercial pesticide sample (Supelco 8081 pesticide mix, 22 components, 200 μg ml -1 in hexane:toluene 50:50) was used without further modification. GC conditions Front inlet Split/Splitless Inlet temperature 300 ºC Inlet flow 2.5 sccm He Septum purge flow 3 sccm (switched) Oven 40 C (3 min), 10 C/min to 130 C (2 min), 3 C/min to 280 C Column DB-5, 30 m, 320 μm, 0.25 μm film thickness Syringe 10 µl Injection volume 0.1 µl splitless FID conditions Temperature 300 C H2 1.5 sccm Air 350 sccm Makeup 0 sccm (He) Sampling rate 50 Hz Polyarc reactor conditions Setpoint 293 C H2 35 sccm Air 2.5 sccm Results and Discussion The chromatograms for a 22-component organochlorine pesticide test mixture analyzed with and without the Polyarc reactor are shown in Figure 1; molecular structures for the compounds are shown in Figure 2. The full widths at half maximum (FWHM) of the peaks are, on average, 8% greater when using the Polyarc reactor in conjunction with the FID. The increased broadening has a negligible influence on the chromatographic separation and all compounds are resolved. The peak positions and shapes are also very similar with and without the Polyarc reactor. The peak areas (i.e., a surrogate for sensitivity) obtained with the Polyarc reactor are 13% to 43% larger (average of 22%) than those obtained with the FIDonly. The elution order was determined with mass spectrometry on a separate GC/MS system. Hexane was used as the internal standard for the analysis. The response factors for the 22 analytes are shown in Figure 3 and Table 1 in order of elution. The response factors for the Polyarc reactor are 1.00 ± 0.09 with a mean deviation from unity of 4%. The response factors for the FID-only system have a lower average response of 0.83 ± 0.10 and an increased deviation from unity with an average deviation of 17.1%. The error bars in Figure 3 represent the error obtained from the propagation of the expanded uncertainty in analyte concentration (95% confidence intervals). Within these confidence intervals, the Polyarc reactor provides equal response factors (RF = 1) for every organic compound, because all compounds are completely converted to methane before detection in the FID. Thus, calibrations to determine response factor are not necessary when using the Polyarc reactor and even the concentration (in carbon amount) of unknown peaks could be determined using RF = 1. Table 1 shows the measured concentrations of the analytes in the mixture using RF = 1 for all components. The average error in concentration is less than 5%, which is within the uncertainty of the measurement. The minimum detectable limit (MDL) for the Polyarc reactor is estimated to be about 100 ng/ml (76 ppb) based on the peak areas obtained in this study and the smallest peak area distinguishable from noise. 2

Figure 1. Comparison between the chromatograms obtained for a 22-component organochlorine pesticide test mixture using the Polyarc /FID and FID-only. Figure 2. Molecular structures of select compounds analyzed in this study. 3

Figure 3. Response factors obtained for a 22-component organochlorine pesticide mixture (200 µg/ml in hexane:toluene 1:1) using the Polyarc reactor (top) and FID-only (bottom). 4

Table 1. Molecular information, peak areas, response factors, and concentrations for the analytes in the 22-component organochlorine pesticide mixture tested with the Polyarc reactor. # Measured Analyte Peak Area Analyte MW (g/mol) Carbon Number RF a Expanded Uncertainty Concentration (µg/ml) b Expanded Uncertainty (µg/ml) 1 Tetrachloro-m-xylene 1583 244.0 8 199 1.00 0.07 200 2 alpha-hexachlorohexane 1038 290.8 6 200 1.04 0.08 207 3 gamma-hexachlorohexane 1033 290.8 6 200 1.03 0.08 206 4 beta-hexachlorohexane 1029 290.8 6 192 1.07 0.07 206 5 delta-hexachlorohexane 1025 290.8 6 200 1.02 0.06 205 6 Heptachlor 1216 373.3 10 200 0.94 0.06 187 7 Aldrin 1571 364.9 12 199 0.99 0.06 197 8 Heptachloroepoxide Isomer B 1228 389.3 10 200 0.99 0.05 197 9 gamma-chlordane 1181 409.8 10 200 1.00 0.06 199 10 Endosulfan I (alpha) 1068 406.9 9 196 1.02 0.06 199 11 alpha-chlordane 1214 409.8 10 200 1.03 0.06 205 12 Dieldrin 1477 380.9 12 200 0.97 0.06 193 13 4-4 -DDE 2233 318.0 14 197 1.06 0.06 209 14 Endrin 1390 380.9 12 200 0.91 0.06 182 15 Endosulfan II 1110 406.9 9 200 1.04 0.07 207 16 4-4 -DDD 1994 320.0 14 195 0.96 0.06 188 17 Endrin Aldehyde 1639 380.9 12 197 1.09 0.08 214 18 Endosulfan Sulfate 997 406.9 9 199 0.93 0.15 186 19 4,4 -DDT 1953 354.5 14 192 1.06 0.10 204 20 Endrin Ketone 1536 380.9 12 199 1.01 0.09 201 21 Methoxychlor 2109 345.7 16 200 0.94 0.09 188 22 Decachlorobiphenyl 1169 498.7 12 199 1.01 0.16 200 a Response factors determined using hexane as the internal standard. b Concentrations reported assuming RF = 1 for all analytes. Additional testing of the Polyarc reactor with different molecules and higher concentrations was performed on six solutions of pesticides dissolved in methanol (Figure 4). The analyte concentrations ranged from 1-4 mg/ml. Methanol was used as both the solvent and the internal standard for this analysis. The response factors shown in Figure 4 are 1.00 ± 0.04 with a mean deviation from unity of 0.02. This data provides further confirmation that the Polyarc reactor leads to calibration-free analyses as the response factor is equivalent for all molecules. 5

deactivated direct-connect splitless liner (e.g., the Agilent G1544-80700 direct connect liner, or the Restek Uniliner ). The direct-connect liners make a sealed connection with the column, preventing preferential vaporization and contact of analytes with metal components within the GC inlet. Furthermore, the inlet liners are deactivated and free of quartz wool to prevent adsorption and subsequent hold-up of the analytes. The data shown above for the 22- component pesticide test mixture was collected with a direct-connect liner in splitless mode. Splitless injections, in general, and direct-connect liners, in particular, require smaller injection volumes to avoid overloading the inlet and column. Peak broadening and fronting can be minimized with the appropriate injection volumes and inlet temperature, although in some cases poor peak shapes are unavoidable. Figure 4. Response factors obtained for six pesticides analyzed with the Polyarc reactor Selecting the correct inlet liner When performing quantitative GC analyses it is important that inlet conditions are optimized to prevent discrimination of analytes in the GC system. Sources of analyte discrimination include preferential vaporization in the inlet and hold-up. Hold-up is defined as the retention of small amounts of analyte within the inlet, column, or other location of a GC. If analyte discrimination is occurring within an instrument, it will result in inaccuracies, independent of the analysis and detection methods. For this reason, it is very important that discrimination be eliminated using an appropriate instrument setup. For analytes that are prone to discrimination such as high molecular weight compounds or sticky compounds with reactive functional groups, it is best to use a As an example of poor inlet conditions, the GC was initially equipped with a liner designed primarily for split injections (Agilent part no. 5190-2295) and used in splitless injection mode. The response factors obtained with the Polyarc (Figure 5) deviate from unity, demonstrating that absolute injection of all components is not occurring. This is in stark comparison to the results that were obtained using a direct-connect liner (Figure 3) where RF = 1 for all compounds. The low response factor for methoxychlor, for example, indicates that this compound is either retained within the GC inlet or escaping the system through the septum purge or split vent. A response factor greater than one indicates that the compound has a higher injection efficiency than the internal standard, tetrachloro-m-xylene. The discrimination present with the split liner and not the direct-connect liner may be the result of differences in the design of the liners, or adsorption sites present on the split liner from the quartz wool or residual molecules. These results stress the importance of selecting a proper inlet liner when analyzing high molecular weight and/or highly reactive analytes. 6

Figure 5. Response factors obtained for a 22-component organochlorine pesticide mixture (200 µg/ml in hexane:toluene 1:1) with a poor absolute injection of analytes onto the column. Conclusions The Polyarc reactor saves time and makes analysis easier by eliminating the need for calibration in GC/FID analyses of all organic molecules. The ability of the Polyarc reactor to perform analysis without calibration is unparalleled in the field of analytical chemistry. Quantification of rare molecules for which standards are prohibitively expensive, or do not exist, is made possible with the Polyarc reactor because response factors are unity for all compounds. An additional benefit of the Polyarc reactor is its ability to detect and remedy inlet discrimination issues through the analysis of standard component mixtures; if response factors deviate from unity, there may be a problem with the GC (e.g., system leak, discrimination, compound degradation, etc.). Contact Us For more information or to purchase a Polyarc system, please contact us at 612-787-2721. Please visit our website for details and additional technical literature, www.activatedresearch.com. Activated Research Company shall not be liable for errors contained herein, or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. 2016 Activated Research Company, LLC Printed in the USA December 14, 2015 PA-APP-1570 7