Advanced Topic: Tips and Tricks of HPLC Separations and Troubleshooting. Rita Steed LC Advanced Topics Seminar Series August 25, 2010.

Similar documents
Tips and Tricks of HPLC System Troubleshooting

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think!

HPLC Column Troubleshooting: Is It Really The Column? LC Columns and Consumables. Robyn C. Woolley Online Application Engineer July 24, 2007.

Choosing Columns and Conditions for the Best Peak Shape

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

HPLC Column Troubleshooting

William E. Barber, Ph.D. Applications Chemist October

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

HPLC Method Development with Eclipse Plus: Standard Practices and New Columns. Agilent Technologies

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

High Resolution Fast LC

High Performance Liquid Chromatography

High Performance Liquid Chromatography

LC and LC/MS Column Selection Flow Chart

New To HPLC Avoiding Beginner Pitfalls. Rita Steed LC Columns Application Engineer April 24, 2013

HPLC Method Development: Standard Practices and New Columns

penta-hilic UHPLC COLUMNS

Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput

InertSustainSwift C8

penta-hilic UHPLC COLUMNS

Reversed Phase Solvents

HPLC Method Development: Standard Practices and New Columns. Agilent Technologies February 2007

[ CARE AND USE MANUAL ]

Fast Separation of Vastly Different Compounds by Isocratic HPLC

Too Polar for Reversed Phase What Do You Do?

Basic principles of HPLC

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Validated HPLC Methods

Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies

Acclaim Mixed-Mode WCX-1

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

State-of-the-art C18 HPLC Columns

COMPARISON GUIDE. to C18 Reversed Phase HPLC Columns. Comparison Data on 60 Commonly Used C18 Phases. Stationary Phase Specifications

Comparison of different aqueous mobile phase HPLC techniques

HPLC Separation Robustness and Ruggedness Assessing the Effects of Experimental Variables During Method Development

for Acclaim Mixed-Mode HILIC-1 Column

C18 Column. Care & Use Sheet

Shodex TM ODP2 HP series columns

HPLC Separation Fundamentals. Ed Kim Application Engineer Agilent Technologies, Inc January 14, 2009


Reversed Phase Chromatography: Mobile phase Considerations

Solid Phase Extraction Method Development Tips and Tricks

Does one-size-fit-all?

(HILIC) Bill Champion Agilent Technologies, Inc opt 3/opt3/opt 2 HILIC - Agilent Restricted

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

ProntoSIL C18-EPS Reversed-Phase HPLC Columns

Packings for HPLC. Packings for HPLC

A C18 Silica Column With Exceptional Temperature and ph Stability

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Chemistry Instrumental Analysis Lecture 31. Chem 4631

LC Technical Information

for Acclaim Mixed-Mode WCX-1

ACQUITY UPLC HSS Columns

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Hypersil BDS Columns TG 01-05

Kromasil. for your analytical HPLC. Kromasil. The way to peak performance in liquid chromatography

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

High Performance Liquid Chromatography

[ Care and Use Manual ]

"Theory and Practice of High Speed Chromatography for Bioanalysis" Stuart Coleman March 20, 2007

New Product Update. HPLC Columns

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

ProPac WCX-10 Columns

XSEL EC T C HA RGED SU RFAC E HYBRID (C SH) COLUMNS

Alltech Alltima HP Introduction

[ Care and Use Manual ]

Tips and tricks to increase HPLC column lifetime. Dr. Frank Michel

The new Agilent 1290 Infinity Quaternary LC

FAQ's. 1 - Which column would be the most appropriate for my application?

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

1.7 m Fortis UHPLC Columns

InertSustainSwift C8

High Performance Liquid Chromatography

Comparison Guide to C18 Reversed Phase HPLC Columns

8. Methods in Developing Mobile Phase Condition for C18 Column

Ensure reproducible Polymeric SPE workflow for high-throughput bioanalysis applications

ACE Ultra Inert Base Deactivated HPLC Columns

[ CARE AND USE MANUAL ]

A Look at Column Choices. Ed Kim Application Engineer Agilent Technologies, Inc February 12, 2009

for Acclaim Trinity TM P1

High Pressure/Performance Liquid Chromatography (HPLC)

Dilution(*) Chromatography

Phenyl-Hexyl. UHPLC Columns. Alternate, complementary selectivity to C18 and C8 bonded phases

Method Development Kits

State - of - the -art C18 HPLC Columns

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Zirconia: the Ideal Substrate for Ion-Exchange LC and LC-MS

Acclaim Mixed-Mode WAX-1 Columns

Hints for Strong Ion Exchange Resins

Efficiently analyze your most challenging compounds. Deliver high-quality results faster. Agilent ZORBAX Column Selection Guide for HPLC

Use of the SFC-MS Splitter Kit G

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis - HPLC A.136. Primesep 5 µm columns B.136

ACE SuperC18. ACE UHPLC and HPLC Columns. Ultra-Inert UHPLC and HPLC Columns with Extended ph Stability

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

Do you need trouble-free speed? The NEW Chromolith HR

Transcription:

Advanced Topic: Tips and Tricks of HPLC Separations and Troubleshooting Rita Steed LC Advanced Topics Seminar Series August 25, 200 Page

Troubleshooting Steps You Have Recognized There is a Problem! How Do You Fix It? st Did System Suitability or Sample Fail? 2 nd Review Method for Compliance - Is The Procedure Being Followed Properly? - Are Instrument Settings Correct? 3 rd Ask More Questions! - When Did the System Last Function Properly? - Has Anything Been Changed? 4 th Review ALL parameters! - The Obvious Is Not Always the Cause - Was There More Than One Change? Page 2

HPLC System Components Pump Injector/Autosampler Column Detector Data System/Integrator Problems Can Be Related to All Components in the System Page 3

Column and System Problems A. Pressure B. Peak shape C. Retention Page 4

Pressure Issues Column Observations High Pressure Low Pressure Potential Problems - Plugged frit - Column contamination - Plugged packing - Leak - Flow Incorrect Page 5

Determining the Cause and Correcting High Back Pressure Check pressure with/without column - many pressure problems are due to blockages in the system or guard col. Remove Column - Pressure Still High? Remove Guard Pressure Still High? If Column pressure is high: Back flush column Clear dirty frit surface Wash column Eliminate column contamination and plugged packing High molecular weight/adsorbed compounds Precipitate from sample or buffer Change frit Clear plugged frit PREVENT THIS! Page 6

Column Cleaning Flush with stronger solvents than your mobile phase Reversed Phase Solvent Choices In Order of Increasing Strength Use at least 25 ml of each solvent for analytical columns This Is time consuming Often performed offline Mobile phase without buffer salts 00% Methanol 00% Acetonitrile 75% Acetonitrile:25% IPA 00% Isopropanol 00% Methylene Chloride* 00% Hexane* Must reverse to Re-equilibrate Tip: When using either Hexane or Methylene Chloride; The column must be flushed with Isopropanol before returning to your reverse phase mobile phase. Page 7

Changing a Frit May Not Be a Good Idea May not be possible with new generation columns May damage high performance columns Column Inlet Frit Column Body Compression Ferrule Wear gloves Do not allow bed to dry Do not touch the column - body heat will extrude packing Do not overtighten Female End Fitting Male End Fitting Tip: Prevention is a Much Better Idea! Page 8

The Trick Prevention Techniques - A Better Choice! Use column protection - In-line filters - Guard columns Filter samples Filter buffered mobile phases Easy Sample clean-up (i.e. SPE) Appropriate column flushing Not as Easy Page 9

Inexpensive Filters Prevent Column Frit Plugging Regenerated Cellulose (RC) Recommended Universal hydrophilic membrane; Compatible..with most solvents - aqueous and organic High purity, extremely low extractables and..binding More Uniform Surface Different than Other Cellulose Filters!! In-line Filters Easy to use and replace Frits available in 0.2, 0.5 and 2.0µ porosity Much less expensive than a column Easier and faster to replace than a column frit Page 0

What Are Common Peak Shape Issues?. Split peaks 2. Peak tailing 3. Broad peaks Many peak shape issues are also combinations - i.e., broad and tailing..or tailing with increased retention Symptoms do not necessarily affect all peaks in the chromatogram Each of these problems can have multiple causes Page

Peak Splitting Caused By Disrupted Sample Path Flow path disrupted by void Sample allowed to follow different..paths through column Poorly packed bed settles in use High ph dissolves silica Normal Double Peaks Tip: Similar Effect Can be Caused by Partially Plugged Frit Page 2

Split Peaks from Column Contamination Column: StableBond SB-C8, 4.6 x 50 mm, 5 mm Mobile Phase: 60% 25 mm Na 2 HPO 4, ph 3.0 : 40% MeOH Flow Rate:.0 ml/min Temperature: 35 C Detection: UV 254 nm Sample: Filtered OTC Cold Medication:. Pseudoephedrine 2. APAP 3. Unknown 4. Chlorpheniramine Injection Injection 30 Injection After Column Wash with 00% ACN 2 2 2 4 4 3 4 3 3 0 5 0 5 0 5 0 5 0 5 0 5 Tip: Column washing eliminates the peak splitting, which resulted from a contaminant on the column How could this be prevented? (Guard Column, SPE clean up of samples, Periodic column wash) Page 3

Split Peaks from Injection Solvent Effects Column: StableBond SB-C8, 4.6 x 50 mm, 5 mm Mobile Phase: 82% H 2 O : 8% ACN Injection Volume: 30 ml Sample:. Caffeine 2. Salicylamide A. Injection Solvent 00% Acetonitrile 2 2 B. Injection Solvent Mobile Phase 0 0 0 0 Tip: Injecting in a solvent stronger than the mobile phase can cause peak.shape problems such as peak splitting or broadening Trick: Keep Organic Concentration in Sample Solvent < Mobile Phase Page 4 Group/Presentation Title Agilent Restricted August 25, 200Month ##,

Peak Tailing, Broadening and Loss of Efficiency May be caused by: Column secondary interactions Column contamination Column aging Column loading Extra-column effects Page 5

Peak Shape: Tailing Peaks Symmetry >.2 Causes Some Peaks Tail Secondary - Retention Effects. Residual Silanol Interactions. Small Peak Eluting on Tail of Larger Peak. Normal Tailing Normal Tailing All Peaks Tail Extra-Column Effects. Build up of Contamination on Column Inlet. Heavy Metals. Bad Column. Page 6

Peak Tailing Identifying Column Secondary Interactions Column: Alkyl-C8, 4.6 x 50 mm, 5mm Mobile Phase: 85% 25 mm Na 2 HPO 4 ph 7.0 : 5% ACN Flow Rate:.0 ml/min Temperature: 35 C Sample:. Phenylpropanolamine 2. Ephedrine 3. Amphetamine 4. Methamphetamine 5. Phenteramine 3 No TEA USP T f (5%)..29 2..9 3..63 4. 2.35 5..57 2 3 4 5 2 4 5 0 mm TEA USP T F (5%)..9 2..8 3..20 4..26 5..4 0.0 2.5 5.0 TIme (min) 0.0 2.5 5.0 Tip: Mobile phase modifier (TEA) competes with sample for surface ion exchange sites at mid-range ph values Page 7

Peak Tailing Low ph Minimizes Secondary Interactions for Amines Column: Alkyl-C8, 4.6 x 50 mm, 5mm Mobile Phase: 85% 25 mm Na 2 HPO 4 : 5% ACN Flow Rate:.0 ml/min Temperature: 35 C Sample:. Phenylpropanolamine 2. Ephedrine 3. Amphetamine 4. Methamphetamine 5. Phenteramine ph 3.0 USP T f (5%) 4..33 2 3 4 5 2 3 4 5 ph 7.0 USP T f (5%) 4. 2.35 0.0 2.5 5.0 0.0 2.5 5.0 Tip: Reducing mobile phase ph reduces silanol interaction and peak tailing. Page 8

Peak Tailing High ph Eliminates Secondary Interactions for Amines Column: ZORBAX Extend-C8, 4.6 x 50 mm, 5 m m Mobile Phase: See Below, Flow Rate:.0 ml/min, Temperature: RT, Detection: UV 254 nm Sample:. Maleate 2. Scopolamine 3. Pseudoephedrine 4. Doxylamine 5. Chlorpheniramine 6. Triprolidine 7. Diphenhydramine ph 7 30% 20 mm Na 2 HPO 4 70% MeOH ph 30% 20 mm TEA 70% MeOH 2,3 4 7 4 7 3 t 5 R = 8.5 t R =.4 5 6 2 6 0 5 0 5 0 Peak shape and retention of this sample of basic compounds improves at high ph where column has high IEX activity. Why? Page 9

Peak Tailing - Column Contamination Tip: Quick test to determine if column is dirty or damaged Trick: Reverse column and run sample If improved; Possible cleaning will help No improvement; Column damaged and needs to be replaced QC test forward direction QC test reverse direction QC test after cleaning 00% IPA, 35 C Plates T f. 7629 2.08 2. 2043.64 3. 3727.69 4 3355.32 2 3 Plates T f. 7906.43 2. 2443.2 3. 7999.9 4 7098.25 2 3 Plates T f. 7448.06 2. 2237.2 3. 5366. 4 9067.7 2 3 4 4 4 0.0 2.5 5.0 0.0 2.5 5.0 0.0 2.5 5.0 Column: StableBond SB-C8, 4.6 x 250 mm, 5mm Mobile Phase: 20% H 2 O : 80% MeOH Flow Rate:.0 ml/min Temperature: R.T. Detection: UV 254 nm Sample:. Uracil 2. Phenol 3. 4-Chloronitrobenzene 4. Toluene Page 20

Peak Tailing/Broadening Sample Load Effects Columns: 4.6 x 50 mm, 5mm Mobile Phase: 40% 25 mm Na 2 HPO 4 ph 7.0 : 60% ACN Flow Rate:.5 ml/min Temperature: 40 C Sample:. Desipramine 2. Nortriptyline 3. Doxepin 4. Imipramine 5. Amitriptyline 6. Trimipramine Tailing Eclpse XDB-C8 USP T f (5%) A. High Load x0 C. Broadening Competitive C8 Plates A B..60.70 2. 2.00.90 3..56.56 4. 2.3.70 5. 2.5.86 6..25.25 B. 0 5 0 Low Load D. 0 5 0 C D. 850 594 2. 85 7842 3. 2776 623 4. 2539 8359 5. 2735 0022 6. 589 0725 0 5 T im e ( m in ) 0 5 T im e (m in ) Tip: Evaluate Both Volume and Mass Loading Page 2 Group/Presentation Title Agilent Restricted August 25, 200Month ##,

Unknown Phantom Peaks Column: Extend-C8, 4.6 x 50 mm, 5 mm Mobile Phase: 40% 0 mm TEA, ph : 60% MeOH Flow Rate:.0 ml/min Temperature: R.T. Detection: UV 254 Sample:. Maleate 2. Pseudoephedrine 3. Chlorpheniramine Sample : Chlorpheniramine maleate Peak : maleate 2 Sample 2 : Chlorpheniramine maleate and Pseudoephedrine Peak : maleate Peak 2: pseudoephedrine Peak 3: chlorpheniramine (from st injection) Plates. 5922 2. 9879 3. 779 3 Phantom peak from first injection 0 5 0 0 5 0 5 Tip: The extremely low plates for moderately retained peaks are an indication of a very late eluting peak from a preceding run. Page 22

Extra-Column Dispersion Increasing Extra-Column Volume Use short, small internal diameter tubing between the injector and column and between the column and detector. Make certain all tubing connections are made with matched fittings. Use a low-volume detector cell. Inject small sample volumes. Page 23

Peak Broadening Extra-Column Volume Column: StableBond SB-C8, 4.6 x 30 mm, 3.5 mm Mobile Phase: 85% H 2 O with 0.% TFA : 5% ACN Flow Rate:.0 ml/min Temperature: 35 C Sample:. Phenylalanine 2. 5-benzyl-3,6-dioxo-2-piperazine acetic acid 3. Asp-phe 4. Aspartame 0 ml extra-column volume 50 ml extra-column volume (tubing) 2 3 4 3 2 4 0.0 0.5.0.5 2.0 0.0 0.5.0.5 2.0 Page 24

System Connections Can Cause Peak Broadening The system has been optimized and : All tubing lengths are minimum Smallest diameter tubing used Proper flow cell column Symptom: Still seems to have too much extra column volume What is wrong? Have you made the connections properly? Page 25

Column Connectors Used in HPLC Swagelok Waters 0.090 in. 0.30 in. Parker 0.090 in. Rheodyne 0.70 in. Valco 0.080 in. Uptight 0.090 in. Troubleshooting LC Fittings, Part II. J. W. Dolan and P. Upchurch, LC/GC Magazine 6:788 (988) Page 26

What Happens If Connections Are Poorly Made? Wrong too long Ferrule cannot seat properly Wrong too short X Mixing Chamber If Dimension X is too long, leaks will occur X If Dimension X is too short, a dead-volume, or mixing chamber, will occur Page 27

Stainless Steel and Polymer Fittings Which type is used and when? Stainless Steel (SS) fittings are the best choice for reliable high pressure sealing Agilent uses Swagelok type fittings with front and back ferrules which give best sealing performance throughout all our LC systems PEEK (<400b bar System Pressure) fittings are ideal where: Connections are changed frequently, i.e. connecting columns Pressure is less critical PolyKetone Easy, hand tighten column connection 600 bar Pressure Rating PN: 5042-8957 (0/pk) Fits to SS Tubing Page 28

Changes in Retention Can Be Chemical or Physical May be caused by: Column aging Column contamination Insufficient equilibration Poor column/mobile phase combination Change in mobile phase Change in flow rate Different gradient delay volumes Page 29

Column Aging/Equilibration Causes Retention/Selectivity Changes Column - Initial Column - Next Day Column - After Cleaning with % H 3 PO 4 /Equilibration 2 2 0 3 5 9 2 5 0 3 5 9 2 5 The primary analyte was sensitive to mobile phase aging/ conditioning of the column The peak shape was a secondary issue (metal chelating compound) resolved by de-activating the active metal contamination Page 30 Group/Presentation Title Agilent Restricted August 25, 200Month ##,

: : : : : : : : : Metal Sensitive Compounds Can Chelate Hint: Look for lone pair of electrons on :O: or N which can form 5- or 6-membered ring with metal H C O H C O M +2 OH + M +2 O H Salicylaldehyde 6-membered ring complex N : M +2 OH C C O N OH M +2 8-hydroxyquinoline 5-membered ring complex a-benzoinoxomine 5-membered ring complex Page 3

Acid Wash Can Improve Peak Shape Before Acid Wash After Acid Wash 50 00 mls % H 3 PO 4 HO OH OH HO OH. 2.. 2. OH OH OH 2 Columns: ZORBAX SB-Phenyl 4.6 x 50 mm Mobile Phase: 75% 25 mm ammonium phosphate buffer 25% ACN Flow Rate:.0 ml/min. Temperature: Sample Size: RT 5 ml 2 Tf: 3.7 Tf:.2 A % H 3 PO 4 solution is used on SB columns, 0.5 % can be used on endcapped columns. Page 32

Example: Change in Retention/Selectivity Unintended Mobile Phase Variation Tip: The Source of the Problem is Often Not the Obvious Change Column Column 2 Column 2 - Fresh mobile phase 0 4 6 Time(min) 0 2 3 4 5 6 7 0 4 6 Time(min) I have experimented with our mobile phase, opening new bottles of all mobile phase components. When I use all fresh ingredients, the problem ceases to exist, and I have narrowed the problem to either a bad bottle of TEA or phosphoric acid. Our problem has been solved. Page 33 Group/Presentation Title Agilent Restricted August 25, 200Month ##,

Tip: Dwell Volume Differences Between Instruments Can Cause Changes in Retention and Resolution V D = 0.43 ml Column: ZORBAX Rapid Resolution Eclipse XDB-C8 4.6 x 75 mm, 3.5 µm 0 0 20 30 40 V D = 2.0 ml 0 0 20 30 40 Mobile Phase: Gradient, 0-00 %B in 52.5 min. Flow Rate: Temperature:25 C Injection: Detection: Sample: A: 5/95 methanol/ 25 mm phosphate ph 2.50 B: 80/20 methanol/25 mm phosphate ph 2.50 0.5 ml/min 5 ml 250 nm Mixture of antibiotics and antidepressants Upper trace simulates actual run data entered into DryLab 3.0 software Lower trace is simulated chromatogram for larger V D Page 34

Trick: Measure and Correct for Dwell Volume (V D ) If V D > V D2 Compensate for longer V D by adding an isocratic hold to V D2, such that... Hold + V D2 = V D If V D < V D2 Delay injection, such that V D2 - delay = V D 0004P.PPT Page 35

Mobile Phase ph and Buffers Why Are These So Important in HPLC? ph affects ionization Silica surface of column Sample components of interest Buffers Resist changes in ph and maintain retention Improve peak shape for ionizable compounds Affects Column Life Low ph strips bonded phase High ph dissolves silica Page 36

Minimize Change in Retention/Selectivity Lot-to-Lot Evaluate: All causes of column-to-column change* Method ruggedness (buffers/ionic strength) ph sensitivity (sample/column interactions) *All causes of column-to-column change should be considered first, especially when only one column from a lot has been tested. Page 37

Lot-to-Lot Selectivity Change Related to ph Choice ph 4.5 - Lot ph 3.0 - Lot 2-Base 2 3 4-Base 4 3 0 2 4 6 8 0 2 4 6 8 0 2 4 6 8 0 2 4 6 8 ph 4.5 - Lot 2 ph 3.0 - Lot 2 2 2-Base 3 4-Base 4 3 0 2 4 6 8 0 2 4 6 8 0 2 4 6 8 0 2 4 6 8 ph 4.5 shows selectivity change from lot-to-lot for basic compounds ph 3.0 shows no selectivity change from lot-to-lot Page 38 Group/Presentation Title Agilent Restricted August 25, 200Month ##,

Why Worry About ph? ph, pka and Weak Acids RCOOH RCOO - + H + COOH COO _ K a = [RCOO - ][H + ] [RCOOH] + H + K a = 6.4 x 0-5 pk a = 4.2 At ph 4.2 the sample exists as benzoic acid and the benzoate ion in a ratio of :. Peak shape can be poor At ph 5.2 9% of the sample exists as the benzoate ion. RP retention decreases. At ph 3.2 9% of the sample exists as benzoic acid. RP retention increases. Page 39

Effect of ph on Peak Shape at or Near the Sample pka Column: ZORBAX SB-C8 4.6 x 50 mm, 5 mm Mobile Phase: 40% 5 mm KH 2 PO 4 : 60% ACN Flow Rate:.0 ml/min Temperature: RT ph 4.4 ph 3.0 CH 3 CHCOOH CH 2 CH(CH 3 ) 2 Ibuprofen pk a = 4.4 0 2 3 4 5 6 7 8 9 0 0 2 3 4 5 6 7 8 9 0 Inconsistent and tailing peaks may occur when operating close to an analyte s pka and should be avoided. Page 40

Why Worry About ph? ph, pka and Weak Bases R 3 NH + R 3 N + H + K a = [R 3 N][H + ] [R 3 NH + ] CHOCH CH N 2 2 + H CH 3 CH 3 CHOCH CH N 2 2 + CH 3 CH 3 H + K a = x 0-9 pk a = 9 At ph 9 the sample exists as protonated and unprotonated diphenhydramine in a ratio of :. Peak shape can be poor. At ph 0 9% of the sample exists as unprotonated diphenhydramine. At ph 8 9% of the sample exists as protonated diphenhydramine. Page 4

retention time (min) Change in Retention with ph for Ionizable Compounds is Compound Dependent 2 0 8 6 Acetylsalicylic acid Pyridine Codeine Procainamide Amphetamine Caffeine pka 3.5 pka 5.2 pka 8.0 pka 9.2 pka 9.9 pka 4 4 2 Mobile Phase: 45% MeOH: 55% 20 mm Phosphate Buffer 0 ph 2.5 ph 6.5 ph 8 ph.5 ph Page 42

Importance of ph and Buffers A Practical Example Why the Sample Dictates Use What Happens When Buffer Used Effectively What Happens When Buffer Ignored or Used Improperly Page 43

Importance of ph and Buffers - A Practical Example Optimized Isocratic Conditions for Cardiac Drugs 2 3 4 Column: StableBond SB-C8, 4.6 x 50 mm, 5 mm Mobile Phase: 45% 25 mm NaH 2 PO 4, ph 3.0 55% MeOH Flow Rate: 2.0 ml/min. Temperature:35 C Detection: UV 254 nm Sample: Cardiac Drugs. Diltiazem 2. Dipyridamole 3. Nifedipine 5 2 4. Lidoflazine 3 4 5. Flunarizine 5 0 2 3 4 5 6 7 8 9 0 2 3 4 0 2 3 4 5 6 Page 44

I Don t Have Time to Make Buffers or Adjust ph Column: StableBond SB-C8 4.6 x 50 mm, 5 mm Mobile Phase: A: 20% H 2 O B: 80% MeOH Flow Rate:.0 ml/min. Temperature:35 C UV Detection: 254 nm Sample: Cardiac Drugs Even at very high % MeOH Most Components Strongly Retained with Poor peak Shape Due to IEX at Surface 0 5 0 5 20 25 Buffers are critical to good retention and peak shape in many separations. Page 45

What If You Work Outside the Buffer Range? 2 Unsuitable Peak Shape Columns: StableBond SB-C8 4.6 x 50 mm, 5 mm Mobile Phase: A: 30% 25 mm NaH 2 PO 4, ph 4.8 unbuffered B: 70% MeOH Flow Rate:.0 ml/min. Temperature:3 5 C UV Detection: 254 nm Sample: Cardiac Drugs. Diltiazem 2. Dipyridamole 3. Nifedipine 4. Lidoflazine 5. Flunarizine 5 3 4 0 5 0 5 20 25 Page 46

Don t Forget - Match Column to ph of Mobile Phase for Maximum Column Lifetime High ph and Room Temperature (ph RT) Mobile Phase: 50%ACN: 50% Water : 0.2% TEA (~ ph ) Initial After 30 injections Tip: Use Columns Designed for chosen ph Page 47

Detection Issues Recognize where the problem originates Is it a consequence of technique? Is it expected due to use of certain mobile phase components? Can it be corrected by adjusting detector parameters? Answers will help find a solution! Let s explore some problems and solutions Page 48

Chromatographic Results with Wrong Lamp at 24 nm Wavelength OEM Lamp Lamp from Generic Source Tip: Could also be a symptom of aging lamp Page 49

Expanded View of Chromatographic Results Generic Source Lamp at 24 nm Wavelength OEM Lamp Peak S/N = 50 Peak 2 S/N = 400 Peak 3 S/N = 300 Lamp from Generic Source Peak S/N = 5 Peak 2 S/N = 50 Peak 3 S/N = 50 Tip: Poor S/N makes it difficult to detect low level impurities Page 50

Effect of Detector Response Time The system is operating well - the settings were poorly made Slow data rates can hinder impurity detection and reduce sensitivity Response Time 0. sec st peak =.2 sec At 20 pts/sec = 24 pts/sec st peak =.2 sec At 5 pts/sec = 6 pts 0.2 sec 0.5 sec.0 sec 2.0 sec 0 0. 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9.0 Agilent 00 DAD Agilent 00 WPS with ADVR Column: Poroshell 300SB-C8 2. x 75 mm, 5 mm Mobile Phase: A: 95% H 2 O, 5% ACN with 0.% TFA B: 5% H 2 O, 5% ACN with 0.% TFA Flow Rate: 2 ml/min Temperature:70 C Detector: Piston stroke: 20 UV 25 nm Sample:. Neurotensin3. Lysozyme 2. RNaseA 4. Myoglobin Tip: Adjust the response rate of your detector for best peak detection. Page 5

Conclusions HPLC column problems are evident as High pressure (prevention better than the cure) Undesirable peak shape Changes in retention/selectivity Often these problems are not associated with the column and may be caused by instrument and chemistry issues. ph of mobile phase Instrument connections Detector settings Metal contamination Start with the correct questions Find the answers Answers will lead to solutions Page 52