MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

Similar documents
Mass determination of 1 kg silicon spheres for Avogadro project

EVALUATION OF ADSORPTION EFFECT BY USING SILICON SPHERE AND SILICON SURFACE ARTEFACTS

Status of the redefinition of the mole

Air Density Determination Using 1 kg Buoyancy Mass Comparison(III)

Traceable Mass Determination and Uncertainty Calculation

Implementation of MRA NMIs, BIPM,RMO-TCs Improvement of Global Measurement Standards NMIs, BIPM

Estimation of Uncertainty of Air Buoyancy Correction for Establishment of Primary Mass Standards

AVOGADRO CONSTANT DETERMINATION FOR A REDEFINED KILOGRAM. Horst Bettin, Michael Borys, Ingo Busch, Arnold Nicolaus, Peter Becker

Flowchart for the extraordinary calibrations prior to the redefinition of the kilogram. (BIPM proposal)

The Avogadro constant: Counting atoms in a single-crystal 28 Si sphere

EFFECT OF COVARIANCES IN A DENSITY MEASUREMENT OF SILICON SAMPLES BY HYDROSTATIC WEIGHING

Maintaining and disseminating the kilogram following its redefinition

On the possible future revision of the SI

WATT BALANCES AND THE FUTURE OF THE KILOGRAM

On the redefinition of the kilogram. Dr Philippe RICHARD

MASS AND VOLUME COMPARISONS AT MIKES

Proposed change to the definition of the kilogram: Consequences for legal metrology

Mise en pratique for the definition of the kilogram in the SI

Report on CCM Pilot Study CCM.R-kg-P1. Comparison of future realizations of the kilogram. Final Report

The Kilogram Redefined:

Water Vapour Effects in Mass Measurement

Kilogram: new definition, selected mises-en-pratique and direct link with SMD activities

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Measurement of the 30 Si mole fraction in the new Avogadro silicon material by Neutron Activation and high resolution spectrometry

CCM short note on the dissemination process after the proposed redefinition of the kilogram

Alternative set of defining constants for redefinition of four SI units

Ingvar Lindgren Symposium: New Trends in Atomic Physics II 2 3 June 2006 The Quantum SI: A possible new International System of Units

The General Conference on Weights and Measures (CGPM), at its 24th meeting,

In-vacuum realisation of the new definition of the kilogram. Stuart Davidson, National Physical Laboratory, UK JRP Coordinator - NewKILO

A Collaboration Between Ten National Measurement Institutes To Measure The Mass Of Two One Kilogram Mass Standards

Alternative set of defining constants for redefinition of four SI units

Preparation Primary Gas Reference Material (PGRM) by using NIMT Facilities

A bilateral comparison of a 1 kg platinum standard

Revising the International System (SI) of Units of Measurement. Dr Barry IngIis President, International Committee for Weights and Measures

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

THE WATT BALANCE ROUTE TOWARDS A NEW DEFINITION OF THE KILOGRAM. A. Eichenberger, H. Baumann, and B. Jeckelmann.

Guide to Weighing in Vacuum

A Three-Ratio Scheme for the Measurement of Isotopic Ratios of Silicon

EUROMET Project 702 EUROMET.M.D-K4

Ozone cross-sections (Hartley band) Progress towards new accurate values

Modernizing the SI implications of recent progress with the fundamental constants

The watt balance route towards a new definition of the kilogram

Measurement standards and the general problem of reference points in

THE NEW WATT BALANCE EXPERIMENT AT METAS

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002

SIM SIM.M.D-K3. provide. SIM.M.M-K5 for. A set of. Institute LACOMET LATU INTI. Country Costa Rica Uruguay Argentina Chile México Canada Brazil

ASIA PACIFIC METROLOGY PROGRAME

Redefinition of the kilogram, ampere, kelvin and mole: a proposed approach to implementing CIPM recommendation 1 (CI-2005)

TRACEABILITY AND CAPABILITY CONTROL OF MASS MEASUREMENT EQUIPMENT AND DRIFT STATISTICAL ANALYSIS OF NATIONAL MASS STANDARDS IN LATVIA

13th Comparison between the Swedish national kilogram and SP principal standards for one kilogram

EMRP JRP SIB-05. NewKILO. Developing a practical means of disseminating the redefined kilogram (June 2012 May 2015) Stuart Davidson, NPL, UK

2. CURRENT DEFINITION OF THE TEMPERATURE UNIT

Fundamentals of Mass Determination

On the proposed re definition of the SI

Explicit-unit formulations of proposed redefinitions of SI base units

The Avogadro Constant for the Definition and Realization of the Mole

THE USE OF X-RAY PHOTOELECTRON SPECTROSCOPY TO ESTIMATE THE STABILITY OF PRIMARY MASS STANDARDS

Final Report on CIPM key comparison of multiples and submultiples of the kilogram (CCM.M-K2)

FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).

THE PLANCK-BALANCE A SELF-CALIBRATING PRECISION BALANCE FOR INDUSTRIAL APPLICATIONS

Open Access Repository eprint

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

arxiv:physics/ v1 [physics.class-ph] 28 Jul 2005 On a natural definition of the kilogram and the ampere

SI base units. SI : Système International d'unités (International System of Units)

Fundamental Constants and Units

Mass and Derived Quantities

BUREAU INTERNATIONAL DES POIDS ET MESURES

Research on Air density Measurement for measuring weights

The International Association for the Properties of Water and Steam

BUREAU INTERNATIONAL DES POIDS ET MESURES

The origins and history of the International System of Units, SI. Terry Quinn

Determination of Ar molar mass for the Boltzmann constant. Inseok Yang, Jin Bok Lee, Jin Seog Kim Korea Research Institute of Standards and Science

Comparison Euramet 898

Guest Forum. ph Measurement using the Harned Cell. The Screening Committee Lecture for the First Dr. Masao Horiba s Award

Introduction. (hqp:// units/new- si/)

INTERNATIONAL COMPARISON OF WATER TRIPLE POINT CELLS LEADING TO A MORE PRECISE DEFINITION OF THE KELVIN

Automated volume measurement for weihts using acoustic volumeter

Automatic Comparison of Weights and Mass Standards

DRAFT B, Final Report on CIPM key comparison of 1 kg standards in stainless steel (CCM.M-K1)

Le système SI et son évolution; le rôle de la seconde dans le future SI Terry Quinn

FUNDAMENTAL PROBLEMS IN METROLOGY

The new SI another proposal on the future revision of the International System of Units

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

Teaching and learning basic chemical concepts of the atomic

Report on the. CCM Workshop on the mise en pratique of the new definition of the kilogram. BIPM, Sèvres, November 2012

Final Report August 2010

How do you really know what the temperature is? Michael de Podesta

Cosmological Ouroboros. Sheldon Lee Glashow (Nobel prize in Physics 1979)

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

How to define the base units of the revised SI from seven constants with fixed numerical values

High Precision Mass Measurement in Automation Tatjana Ivanova 1, Jānis Rudzītis 2

Chapter 10. Thermal Physics. Thermodynamic Quantities: Volume V and Mass Density ρ Pressure P Temperature T: Zeroth Law of Thermodynamics

CCQM-K45: Sn in tomato paste key comparison

Preparation of nitrous oxide (N 2 O) in air standard being traceable to SI by gravimetric method

Goodbye, Mr. kilogram!

Mass standard for the french watt balance experiment - pure iridium -

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Transcription:

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT Michael Borys, Michael Gläser, Michael Mecke Physikalisch-Technische Bundesanstalt (PTB), Germany ABSTRACT Spheres made of a silicon single crystal with a mass of about 1 kg are used as density standards and in the context of a project for the determination of the Avogadro constant. The status of the mass determination of such silicon spheres at the PTB is presented. Special facilities and procedures used for the mass determination in air are described. Results for a single sphere and the mass difference between two spheres are given with a discussion of the measurement uncertainty and mass stability. 1. INTRODUCTION The Avogardo constant N A represents an equal number of elementary entities, e. g. atoms or molecules, in one mole of a substance, e. g. in 12 g of the carbon isotope 12 C or 28 g of the silicon isotope 28 Si. Microscopic and macroscopic quantities are linked by the Avogadro constant. Thus the unified atomic mass unit is defined by [1] 3 10 kg 1u = (1) N A and e. g. for a silicon single crystal, the Avogadro constant can be derived from mole volume M V n N A = =, (2) atomic volume m V 0 where M is the molar mass of silicon, V the volume and m the mass of a silicon body, V 0 the volume and n the number of atoms of the unit cell. Based on Eq. (2), a value N A = 6,0221353 10 23 mol -1 was recently derived from measurements of the lattice parameter, volume, mass and molar mass of a silicon single crystal with a relative standard uncertainty (k = 1) of 3,4 10-7 [2]. This result was derived from a single silicon crystal sphere named AVO#1. The AVO#1 sphere is one of three spheres, which were fabricated at CSIRO (National Metrology Laboratory, Australia) from the silicon crystal WASO 04. Details of the measured crystal properties and parameters of the sphere AVO#1 are given in [2]. Between June 2000 and September 2004 several measurements for the determination of the mass of the sphere AVO#1 and the mass difference between the spheres AVO#1 and AVO#2 in air were performed at the PTB. The procedures, experimental set-up and the results are described and discussed in the following. 1

2. EXPERIMENTAL SET-UP AND PROCEDURES The 1 kg mass comparator used for the mass determination of 1 kg silicon spheres is based on the hydrostatic principle, i. e. instead of the usual counterweight the weight force is compensated by hydrostatic buoyancy in water (Fig. 1). This mass comparator has been developed at the PTB and was operational in 1983. Several improvements were carried out until the end of 1997. The hydrostatic comparator is operated in a pressure-resistant chamber and has an exchange mechanism with six positions. Over a cardan joint the balance pan hangs on a rod. The rod is connected with the floating system. An electromagnetic force compensation system ensures constant submergence depth during mass comparisons. With this comparator a pooled standard deviation of about 2 µg can be achieved. A detailed description of this comparator is given in [3]. Pressure-resistant chamber Float system Internal water container Electromagnetic force compensation system Exchange mechanism for six mass standards Fig. 1: PTB s liquid balance based on the hydrostatic weighing principle The hydrostatic mass comparator is located in an air-conditioned room with a temperature stabilisation better 0,05 K within 12 h. Over the same period the temperature changes inside the pressure-resistant chamber are smaller than 0,01 K. According to the recommendations given for the handling of silicon spheres by the CSIRO- NML a resting of the highly polished silicon surface on any hard surface, especially metal surface, has to be avoided. Therefore a special support ring with three conducting Teflon pins and a mass of about 10 g was constructed and used for the weighing process (Fig. 2). The uncertainty contribution of this support ring is included in the uncertainty of the auxiliary weights in the uncertainty budget (Tables 2 and 3). For the measurements described in the following the spheres were cleaned with pure ethanol. The period between cleaning and weighing varied between 40 days and about 12 hours. It is known that after cleaning it takes several days for the surface of a stainless steel weight to be covered again with a water layer in equilibrium with the ambient humidity [6]. An increase of the mass by water adsorption, however, could not be observed. It is assumed that the sorption process was either already completed at the time of the weighing or significantly smaller than the measurement uncertainty (Table 1). 2

Teflon pins Support ring Ring of exchange mechanism Fig. 2: Support ring for silicon spheres with conducting Teflon pins The main source of uncertainty for the mass determination of a silicon sphere in air is the uncertainty of the buoyancy correction. Between a platinum-iridium prototype and a silicon sphere of about 1 kg a volume difference of about 380 cm 3 has to be considered. If the air density is determined according to the air density equation, internationally recommended by the Comité International des Poids et Mesures (CIPM), the relative standard uncertainty of the air density u(ρ)/ρ is in the order of magnitude of 1 10-4 [4, 5] which leads to an uncertainty contribution of about 50 µg. This contribution can be significantly reduced if two buoyancy artefacts with a similar mass and surface, but different volumes, are used for a gravimetric determination of the air density [7, 9]. With these buoyancy artefacts the relative standard uncertainty for the air density determination can be reduced to less than 5 10-5, which leads to a standard uncertainty of the buoyancy correction of less than 25 µg. 3. RESULTS AND DISCUSSION 3.1 Mass determination of silicon sphere AVO#1 The mass of the sphere AVO#1 was determined between June 2000 and March 2001 and in a second period between October 2003 and September 2004. Figure 3 shows the sphere AVO#1 together with the platinum-iridium prototype No. 70 and buoyancy artefacts on the six-position load mechanism of the hydrostatic mass comparator. The history of the mass determination of the sphere AVO#1 is given in Table 1. The traceability was realised by the prototypes No. 55 and No. 70 and the 1 kg stainless steel standards No. 6 and No. F. In the period between 2000 and 2004 all prototypes were calibrated annually, in January, against the German national prototype No. 52, which was calibrated at the Bureau International des Poids et Mesures (BIPM) in 1996. Since that time it had been corrected for a long-term drift known from previous calibrations at the BIPM and by comparison in June 2002 with the prototype No. 70, which was calibrated at the BIPM in April 2002, and in July 2003 with the prototype No. 55, which was calibrated at the BIPM in April 2003. The 1 kg stainless steel standard No. 6 was calibrated in October 2000 against the prototype No. 70, which was calibrated against the prototype No. 52 in January 2000 and the 1 kg stainless steel standard No. F was calibrated in March 2001 against the 1 kg stainless steel primary standard No. G, which was calibrated against the prototype No. 52 in January 2001. For each set of measurements given in Table 1 the measurement uncertainty was estimated. For example in Tables 2 and 3, the uncertainty budget for the last measurement in September 2004 is given for both, the mass determination of the sphere AVO#1 with air density determination for buoyancy correction according to the CIPM formula 1981/91 and with buoyancy artefacts. Figure 4 shows a summary of the results. The systematic difference between the values for the air density determination according to the CIPM formula and with buoyancy artefacts is 3

remarkable. A systematic difference between the two methods is already known from previous measurements [5, 9]. In a summary of air density measurements performed by different laboratories with buoyancy artefacts, Picard et al. reported recently on a mean relative air density difference of approximately 6,4 10-5 between the gravimetric method and the CIPM formula 1981/91 [10]. These results were confirmed by a redetermination of the argon content of air by Park et al. [11]. Together with the most recent CODATA recommendation for the molar gas constant [12], these new values for the air composition would increase the air density of the CIPM formula by 1 ρ/ρ = 7,6 10-5, i. e. the curve for the CIPM formula in Fig. 4 would be shifted up to 35 µg depending on the volume difference between the silicon sphere and the reference weight. Fig. 3: Silicon sphere AVO#1 together with platinum-iridium prototype No. 70 and buoyancy artefacts on the six position load mechanism of the hydrostatic mass comparator Table 1: History of mass determination of silicon sphere AVO#1 at the PTB (Index CIPM: Air density determination according to the CIPM formula 1981/91, Index BA: Gravimetric air density determination with buoyancy artefacts) Beginning End m CIPM / g U CIPM / g (k = 2) m BA /g U BA / g (k = 2) t / C h / % p / hpa Ref. 22.06.2000 11.07.2000 1000,914779 8,5 10-05 1000,914876 4,3 10-05 21,24 21,35 45 60 1013,8 1016,2 Pt 55 26.09.2000 30.09.2000 1000,914790 8,5 10-05 1000,914868 4,3 10-05 20,26 20,29 51 59 1010,7 1012,9 Pt 55 10.01.2001 17.01.2001 1000,914832 8,6 10-05 1000,914861 3,8 10-05 19,96 20,11 48 62 1002,9 1007,7 St 6 19.01.2001 27.01.2001 1000,914815 8,6 10-05 1000,914842 3,8 10-05 19,97 19,99 48 62 1006,5 1007,6 St 6 27.03.2001 30.03.2001 1000,914837 8,6 10-05 1000,914872 4,3 10-05 19,51 19,61 49 57 1009,9 1010,5 St F 07.10.2003 17.10.2003 1000,914808 8,5 10-05 1000,914915 5,0 10-05 19,88 19,93 52 56 1006,3 1007,1 Pt 55 27.04.2004 07.05.2004 1000,914834 8,5 10-05 1000,914874 4,8 10-05 19,96 19,99 41 46 1006,5 1008,2 Pt 70 02.09.2004 11.09.2004 1000,914869 8,5 10-05 1000,914968 4,8 10-05 19,84 20,02 44 52 1006,7 1008,6 Pt 70 14.09.2004 25.09.2004 1000,914826 8,5 10-05 1000,914933 4,8 10-05 19,95 20,00 43 53 1007,3 1007,9 Pt 70 1 This value is valid only, if for the CIPM formula the more exact value of the molar mass of dry air is obtained by an additional measurement of the concentration of carbon dioxide. If the concentration of carbon dioxide is not considered, a shift of about ρ/ρ = 6,5 10-5 would be obtained. 4

Table 2: Uncertainty budget for the mass determination of the sphere AVO#1 with air density determination for buoyancy correction according to the CIPM formula 1981/91 [4, 8] (values for September 2004) Parameter Standard uncertainty Sensitivity coefficient Uncertainty contribution X i u(x i ) c i u i (m) / g CIPM formula 7,7 10-05 kg / m 3 3,8 10-04 m 3 3,0 10-05 Pressure 3 Pa 4,5 10-06 g / Pa 1,4 10-05 Temperature 1,0 10-02 K 1,7 10-03 g / K 1,7 10-05 Humidity 5,0 10-03 4,1 10-03 g 2,1 10-05 Mole fraction CO 2 1,0 10-05 1,8 10-01 g 1,8 10-06 Mass of reference/auxiliary weights 8,3 10-06 g 1,0 8,3 10-06 Volume of AVO#1 3,6 10-05 cm 3 1,2 10-03 g / cm 3 4,3 10-08 Volume of reference/auxiliary 1,1 10-03 cm 3 1,2 10-03 g / cm 3 1,3 10-06 weights 2 Weighing difference 5,8 10-07 g 1,0 5,8 10-07 u(m) 4,3 10-05 Table 3: Uncertainty budget for the mass determination of the sphere AVO#1 with air density determination for buoyancy correction based on buoyancy artefacts (values for September 2004) Parameter Standard uncertainty Sensitivity coefficient Uncertainty contribution X i u(x i ) c i u i (m) / g BA mass difference in vacuum 2,4 10-06 g 4,7 1,1 10-05 BA weighing difference in air 4,1 10-06 g 4,7 1,9 10-05 BA sorption correction 2,4 10-07 g 4,7 1,1 10-06 BA volume difference 5,6 10-04 cm 3 5,8 10-03 g / cm 3 3,2 10-06 Mass of reference/auxiliary weights 8,3 10-06 g 1,0 8,3 10-06 Volume of AVO#1 3,6 10-05 cm 3 1,2 10-03 g / cm3 4,3 10-08 Volume of reference/auxiliary weights 2 1,1 10-03 cm 3 1,2 10-03 g / cm 3 1,3 10-06 Weighing difference 5,8 10-07 g 1,0 5,8 10-07 u(m) 2,4 10-05 In the period between June 2000 and March 2001 the PTB buoyancy artefacts No. A3 were used for the gravimetric determination of the air density. From several previous measurements, the mass difference of these artefacts in vacuum has been proven to be stable, i. e. within a drift of less than 2 µg per year. In October 2003 it was not possible to use these artefacts because they were used in the context of a EUROMET project at other laboratories [13]. Therefore the PTB buoyancy artefacts No. A5 were used for the succeeding measurements. As also seen in Fig. 4, the air density determined by these artefacts seems to be higher than the air density determined by the A3 artefacts. The mean difference between the curve for the CIPM formula and the curve for the A3 artefacts is 58 µg, and 88 µg for the curve for the A5 artefacts. Even though both values are within the given uncertainties a systematic influence seems to be obvious, which was the reason for several comparison measurements performed between the A5 and the A3 artefacts in the year 2004. As a result a systematic difference of about 7,9 10-5 was confirmed. This value corresponds to a difference between 29 µg and 36 µg in Fig. 4. The reason for this systematic shift of the A5 artefacts is still unclear and the subject of current investigations. 2 Since the uncertainty of the volume of the reference and auxiliary weights is already part of the uncertainty of the mass of these weights, this contribution has to be considered in the calculation of the combined uncertainty with a negative sign (cf. [3], p. 271). 5

1000,91505 1000,91500 1000,91495 CIPM formula 1981/91 Buoyancy artefacts 1000,91490 Mass / g 1000,91485 1000,91480 1000,91475 1000,91470 1000,91465 Cleaning Cleaning Cleaning Cleaning Cleaning 15.3.00 1.10.00 19.4.01 5.11.01 24.5.02 10.12.02 28.6.03 14.1.04 1.8.04 17.2.05 Date Fig. 4: Mass of silicon sphere AVO#1 determined with air density determination for buoyancy correction according to the CIPM formula 1981/91 and with buoyancy artefacts (uncertainty bars indicate expanded uncertainties U (k = 2)) In January 2001 and September 2004 the mass of the sphere AVO#1 was measured immediately before and after the cleaning procedure. In both cases, a considerable mass reduction between 20 µg and 40 µg could be observed. This corresponds to a mass reduction per surface area between 0,07 µg/cm 2 and 0,15 µg/cm 2. 3.2 Measurement of mass differences between the spheres AVO#1 and AVO#2 The mass difference between the spheres AVO#1 and AVO#2 was determined in October 2003 and April/May 2004. Table 4 gives a summary of these measurements and Table 5 an example of the uncertainty budget. Since for the measurement of the mass difference between two spheres no buoyancy correction is necessary, the uncertainties are significantly smaller than for the mass determination of one sphere. The difference of 14 µg between the two values is, however, out of the sum of the two expanded uncertainties (10,8 µg). It is assumed that the different history of the spheres in the half year between the measurements and a not perfect cleaning process are two reasons for this result. Table 4: Results for the determination of the mass difference between the silicon spheres AVO#1 and AVO#2 Beginning End m 1-2 / g U m / g (k = 2) T / C h / % p / hpa 07.10.2003 17.10.2003-0,013478 5,0 10-6 19,88-19,93 52-56 1006,3-1007,1 27.04.2004 07.05.2004-0,013492 5,8 10-6 19,96-19,99 41-46 1006,5-1008,2 6

Table 5: Uncertainty budget for the determination of the mass difference between the spheres AVO#1 and AVO#2 (values for April/May 2004) Parameter Standard uncertainty Sensitivity coefficient Uncertainty contribution X i u(x i ) c i u i (m) / g Mass of auxiliary weights 3,2 10-06 g 1,0 3,2 10-06 Volume of auxiliary weights 2 1,2 10-03 cm 3 1,2 10-03 g/cm3 1,4 10-06 Volume of AVO#1 and AVO#2 9,8 10-05 cm 3 1,2 10-03 g/cm3 1,2 10-07 Weighing difference 7,3 10-07 g 1,0 7,3 10-07 u(m) 2,9 10-06 3. SUMMARY AND CONCLUSIONS The mass of the silicon sphere AVO#1 was determined in several measurements within the period June 2000 and September 2004 with an expanded uncertainty between 38 µg (buoyancy artefacts) and 86 µg (CIPM formula). The main source for the uncertainty is the buoyancy correction. When a pair of buoyancy artefacts was used in air and in vacuum, the uncertainty of the air density could be determined with a considerably smaller uncertainty than by use of the CIPM formula 1981/91 for the air density. Also, a systematic difference between the buoyancy correction based on a determination of the air density by the CIPM formula 1981/91 and a gravimetric determination with buoyancy artefacts was observed. One reason for this difference is probably the deviation between the air composition used for the CIPM formula 1981/91 and the recently measured values. In addition, a possible unexpected drift of the buoyancy artefacts A5 has to be considered. Within the given uncertainties no systematic drift of the mass of the sphere AVO#1 could be observed. The mass difference between two silicon spheres can be determined with a significantly smaller uncertainty than the mass of a single sphere. Between the spheres AVO#1 and AVO#2 a mass difference was measured in October 2003 and April/May 2004 with an expanded uncertainty of 5 µg and 6 µg, respectively. Although a relative standard uncertainty of about 2 10-8 could be obtained for the mass determination of a single sphere, for an investigation of the mass drift of silicon spheres a further reduction of the measurement uncertainty and a unified handling and cleaning procedure is considered to be indispensable. A reduction of the measurement uncertainty by 50% seems to be realistic within the next two or three years. REFERENCES [1] Mohr, P. J., Taylor, B. N., CODATA Recommended Values of the Fundamental Physical Constants: 1998, J. Phys. Chem. Ref. Data, 28, 1999,pp. 1713-1852 [2] Becker, P., Bettin, H., Danzebrink, H.-U., Gläser, M., Kuetgens, U., Nicolaus, A., Schiel, D., De Bièvres, P., Valkiers, S., Taylor, P., Determination of the Avogadro constant via the silicon route, Metrologia, 40, 2003, pp. 271-287 [3] Kochsiek, M., Gläser, M. (eds.), Comprehensive mass metrology, Berlin, WILEY-VCH Verlag Berlin, 2000, pp. 475-476 [4] Giacomo, P., Equation for the determination of the density of moist air (1981), Metrologia, 18, 1982,pp. 33-40 [5] Picard, A., Fang, H., Three methods of determining the density of moist air during mass comparisons, Metrologia, 39, 2002, pp. 31-40 7

[6] Schwartz, R., Precision determination of adsorption layers on stainless steel mass standards by mass comparison and ellipsometry, Part I: Adsorption isotherms in air, Metrologia, 31, 1994, pp. 117-128 [7] Gläser, M., Schwartz, R., Mecke, M., Experimental determination of air density using a 1 kg mass comparator in vacuum, Metrologia, 28, 1991, pp. 45-50 [8] Davis, R. S., Equation for the determination of the density of moist air (1981/91), Metrologia, 29, 1992, pp. 67-70 [9] Picard, A., Fang, Mass comparisons using air buoyancy artefacts, Metrologia, 41, 2004, pp. 330-332 [10] Picard, A., Fang, H., Gläser, M., Discrepancies in air density determination between thermodynamic formula and a gravimetric method: Evidence for a new value of the mole fraction of argon in air, Metrologia, 2004, (accepted for publication) [11] Park, S. Y., Kim, J. S., lee, J. B., Esler, M. B., Davis, R. S., Wielgosz, R. I., A redetermination of the argon content of air for buoyancy corrections in mass standard comparisons, Metrologia, 2004 (accepted for publication) [12] CODATA, http://physics.nist.gov/cuu/constants/ [13] EUROMET Mass & derived quantities project 144, Measurement of air density using specially designed masses Addresses of the Authors: Michael Borys, Michael Gläser, Michael Mecke, Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig, Germany, E-mail: Michael.Borys@PTB.de, Internet: http://www.ptb.de/en/org/1/11/111 8