Performance of Semi-micro and Single Drop Analysis by Using. Flash Xenon Lamp Spectrophotometer

Similar documents
Effective DNA Absorption Range of SpectroArt 200 by Using. Semi-micro and Micro Cuvettes

Protein assay of SpectroArt 200

Using Dolphin-1D microtiter assay to establish protein

Spectrophotometric quantification of nucleic acids using the nanovette microliter cell accessory with a DU 730 spectrophotometer

BioDrop DUO dsdna Application Note

Nano Spectrophotometer LNS-A10

The Plus for your lab!

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Lecture 5. More on UV-visible Spectrophotometry: Beer s Law and Measuring Protein Concentration

Adenoviral Reference Material Working Group Bid Submission Form Participation in Assignment of Particle Concentration RFP 8.0

Instrumental Technique: Cuvette. Md Rabiul Islam

TL6 Ultra Micro-volume Spectrophotometer

MOLEBIO LAB #4: Using a Spectrophotometer

Degree of labeling (DOL)

Bradford Reagent, 5x

EXPERIMENT #3 A Beer's Law Study

Protein Quantitation using a UV-visible Spectrophotometer

Determining the Concentration of a Solution: Beer s Law

Spectrophotometry Materials

Quantitative Determination of Proteins

IRDye 800CW Protein Labeling Kit Low MW

Spectroscopy. Page 1 of 8 L.Pillay (2012)

FLS980 Series Reference Guide

Low-volume, High Throughput Workflow for Analysis of Nucleic Acid Samples for Biobanking

Supporting Information s for

Application Note TABLE OF CONTENTS

Basics of UV-Visible Spectroscopy *

Lab 3: Protein Determination and Enzyme Assay of Crab and Onion Samples

Optical density measurements automatically corrected to a 1-cm pathlength with PathCheck Technology

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

Protein assay. Absorbance Fluorescence Emission Colorimetric detection BIO/MDT 325. Absorbance

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

Quantification Strategies Using the High Sensitivity Protein 250 Assay for the Agilent 2100 Bioanalyzer. Technical Note. Abstract

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

INSTRUCTION MANUAL. VWR myspec spectrophotometer

Interaction of Gold Nanoparticle with Proteins

Spectrophotometer for Life Science. BioSpec-nano C101-E112D

Microplate Spectrophotometer

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Protease Inhibitor Cocktail A (1 tablet / 7 10 ml, Roche Cat# ) Protease inhibitor Cocktail B (0.5ml per 250ml, Calbiochem Cat# )

PowerWaveX Select and KC4 : A Multifunctional System for Today s Laboratory Environment

GE Healthcare Life Sciences. Spectrophotometry. Handbook

Nucleic Acid Quantitation in Microplates

MOHAWK COLLEGE OF APPLIED ARTS AND TECHNOLOGY CHEMICAL AND ENVIRONMENTAL TECHNOLOGY DEPARTMENT. Lab Report ROOM NO: FE E309

High-throughput Isolation of Genomic DNA From Buccal Swab on the Eppendorf epmotion 5075 VAC

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Eppendorf UVette. Instructions for use

QuickZyme Total Protein Assay (to be used with acid hydrolyzates)

An ultra-sensitive, versatile, and affordable single tube luminometer for Life Science Research.

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Applications Note 228 September 2010

Microdetermination of phosphorus using the SPECTRAmax PLUS microplate spectrophotometer: choice of microplate, cuvette or test tube assay formats

Chirascan 6-Cell Peltier Cell Holder: Rapid Optimisation of Buffer Conditions for Stabilising Protein Therapeutics

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Spectrophotometry. Introduction

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Experiment 13I THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Module 3. Establishing Standard Curve (Bradford Assay)

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

Atomic Absorption Spectrometer ZEEnit P series

A Spectrophotometric Analysis of Calcium in Cereal

Measurement of L-Malic Acid in Wines Using the SpectraMax Plus 384 Microplate Reader

NOTE: The color of the actual product may differ from the color pictured in this catalog due to printing limitation.

Instructor s Advance Preparation

Polyethylene Glycol (PEG), High Sensitive ELISA

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Far UV Performance of the LAMBDA 850/950 UV/Vis and UV/Vis/NIR Research Spectrophotometers

R O Y G B V. Spin States. Outer Shell Electrons. Molecular Rotations. Inner Shell Electrons. Molecular Vibrations. Nuclear Transitions

CHAPTER V ANALYTICAL METHODS ESTIMATION OF DICLOFENAC. Diclofenac (gift sample from M/s Micro Labs Ltd., Pondicherry)

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

800CW Protein Labeling Kit - Low MW

BCA Protein Quantitation Kit

Making competence visible Flexible. Reliable. Personal.

Hepatitis B virus IgM ELISA Kit

Determining the Concentration of a Solution: Beer s Law

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Colorimetry Extinction coefficient ( ) Lambda max ( max) Qualitative vs. quantitative analysis

SPECTROPHOTOMETRY AND SPECTROMETRY - CONCEPT AND APPLICATIONS

FTIR Spectrometer. Basic Theory of Infrared Spectrometer. FTIR Spectrometer. FTIR Accessories

Colorimetric Protein Assays Thermo Scientific NanoDrop Spectrophotometers

The SAXS Guide. Getting acquainted with the principles. New Edition with special contributions

Photochemistry of Nitro-Polycyclic Aromatic Hydrocarbons. August 9, 2011

Chemistry 112 SPECTROPHOTOMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

Method of Enzyme Assay

Outline of Recombinant DNA technology. Application of UV spectroscopy in recombinant DNA technology

microplates and cuvettes

JAPAN QUALITY. Automated Clinical Analyzer

Automated Clinical Analyzer

Kang, Lin-Woo, Ph.D. Professor Department of Biological Sciences Konkuk University Seoul, Korea nd Semester

IRDye 680RD Protein Labeling Kit Low MW

Eppendorf BioSpectrometer kinetic: activity measurements of enzymes immobilized on a mesoporous silica matrix

SCANNING UV VISIBLE SPECTROPHOTOMETER

Photoluminescence Spectrometer (FLS980)

Fluorescence Spectrophotometry

JAPAN QUALITY. Automated Clinical Analyzer

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

Solution reduction synthesis of amine terminated carbon quantum dots

Chemical Equilibrium: Finding a Constant, Kc

Transcription:

Application Note SpectroArt 2 8/9/27 Performance of Semi-micro and Single Drop Analysis by Using Flash Xenon Lamp Spectrophotometer Aim: To investigate the accuracy, reproducibility and linearity of Wealtec s SpectroArt 2 spectrophotometer when creating standard curves measured within 2-8 nm. Furthermore, the aim is to study the performance of a Hellma TrayCell ultra microvolume cuvette when fitted into SpectroArt 2. INTRODUCTION Spectroscopy is a widely used technique for measuring the interaction of molecules in a sample with electromagnetic radiation. SpectroArt 2 spectrophotometer utilises a xenon flash lamp as light source. Hence, it can be used for a wide range of spectrophotometric applications within both UV and visible light-wavelength areas. In 25 Hellma GmBH & Co. KG launched the TrayCel ultra microvolume cuvette which is a device for measuring minute biological or chemical samples. The Hellma TrayCell can be used to measure samples as small as.7 µl. The two different TrayCell caps available generate.2 or mm path lengths respectively which create virtual dilution factors of :5 or :, compared to the : dilution factor created by the usual mm light path cuvette. Hence, with the TrayCell, there is no need to dilute the sample which therefore can be retrieved non-affected after the measurement, if this is desired. A schematic presentation of the principle of the TrayCell is provided in fig. 2a. With SpectroArt 2 spectrophotometer, you can increase the dynamic range of your measurements by increasing the number of times the xenon flash lamp flashes for a single detector reading. A high light throughput and dynamic range is a requirement when utilising devices for measuring very small volume samples, like the Hellma TrayCell ultra microvolume cuvette. The TrayCell is fitted into the cuvette holder as any other standard cuvette, and the position of the cell is adjusted until the best light throughput and signal to noise ratio is obtained. The cuvette is not removed during the measurement. This guarantees a continuously identical position of the aperture in the light beam and no variation in comparison to the reference measurement. --

Figure : SpectroArt 2 spectrophotometer a b Figure 2. Figure a.shows the principle behind the Hellma TrayCell. The sample is pipetted directly onto the Quartz SUPRASIL (QS) window of the TrayCell and the cap housing an aluminium-coated QS mirror is placed on top of the sample, creating a well defined light path as well as keeping the sample from drying out. Light from the xenon flash lamp of the spectrophotometer is deflected towards the sample window of the cell and then reflected by the mirror in the cap. The resultant transmitted light reaches the detector of the spectrophotometer. During filling and cleaning stages, the cell remains in the photometer. This guarantees a continuously identical position of the aperture in the light beam and no variation in comparison to the reference measurement. Figure b shows Hellma TrayCell microvolume cuvettes of two different heights as well as traycell caps. Images from Hellma GmBH & Co. KG. -2-

MATERIAL AND METHODS To demonstrate the efficiency and reproducibility of SpectroArt 2 spectrophotometer as well as its compatibility and with the Hellma TrayCell at different wavelengths, a number of dilution series were constructed and measured utilising the Protein -function of SpectroArt 2. For 28 nm protein measurements, bovine serum albumin (BSA) (Sigma-Aldrich, St Louis MO, U.S.A) diluted in PBS, was used. For the Bradford assay, BSA (Sigma) was diluted in PBS before the addition of Bio-Rad Protein Assay (Bio-Rad, Hercules CA, U.S.A). For DNA standard curves, bp EZ-ladder (EZ-labs, Taipei, Taiwan) diluted in ddh 2 O was used. For the Bradford assay, the Bradford Assay at 595 nm was used. For DNA samples, the wavelength was set to 26 nm using the Customized Wavelength assay and for UV protein, 28 nm was chosen. Standard curve samples were measured using a mm light path Starna 8 quartz cuvette as well as the Hellma TrayCell fitted with a mm light path cap. To increase light throughput and signal to noise ratio, a flash-setting of 8 and an average of 5 (the number of times, in average, that the detector reads a signal) was utilised when measuring with the Hellma TrayCell. When measuring with the mm light path cuvette, the default settings of flash and average were utilised. The cells were thoroughly cleaned in between runs with Hellmanex cleaning solution and plenty of ddh 2 O to remove any protein or colour residue from the cells. Lint-free cotton swabs and lint-free wipes were used when carefully cleaning the TrayCell. -3-

AU 595 AU 595 nm AU 26 AU 26 AU 28 AU 28 RESULTS,9,8,7,6,5,4,3,2, Starna 8 cuv R 2 =,99 5 5 2 25 a Hellma Tray,2,8,6 R 2 =,99,4,2 -,2 5 5 2 25 -,4 b Starna 8 cuv Hellma Tray,4,35,3,25,2 R 2 =,99,5,,5 5 5 2 25,8,6 R 2 =,99,4,2 5 5 2 25 conc DNA µ c conc DNA µ d Starna 8 cuvette B Hellma TrayCell Bradford,2,8,6,4 R 2 =,99,8,6,4 R 2 =,9956,2,2 5 5 2 25 -,2 5 5 2 25 3 e g/ml f -4-

Figure 3 a-e: Standard curves created by dilution series of BSA or DNA measured by a mm light path Starna 8 quartz cuvette (a,c,e) or by the Hellma TrayCell using the mm cap fitted into SpectroArt 2 (b,d,f) showing the R 2 -value, representing the linearity of the curve. All measurements were performed on 4 µl samples. For measurements with the Starna cuvette, default settings of flash and in average detection numbers were utilised. When measuring with the tray cell, a flash setting of 8 and average number of 5 detections in average was used. For comparative reasons, the same solutions were used to measure with both the Starna cuvette and the Hellma TrayCell. All samples were measured five times and show standard deviations of <.2. Mean values obtained when measuring with the TrayCell are multiplied by, because of the virtual dilution factor generated by the mm light path in comparison with the mm light path. Tove Sivik Application Manager Email: info@wealtec.com Wealtec BioScience Co. Ltd. 27Fl. No. 29- Sec.2, Jungjeng E. Rd., Danshuei Jen, Taipei, Taiwan 257 TEL: +886-2-889-8587 FAX: +886-2-889-8589; http://www.wealtec.com -5-