Novel Electrodeionization Devices: Applications in Inorganic Analysis

Similar documents
Determination of trace anions in concentrated hydrofluoric acid

Expanded capabilities in ammonium and amine detection

Determination of urea in ultrapure water by IC-MS/MS

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Trace Cations in Power Plant Waters Containing Morpholine

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Trace ion chromatography

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Determination of Adsorbable Organic Halogen in Wastewater

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

columns IonPac AS17-C Anion Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Sample Preparation for Ion Chromatography

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm)

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

columns IonPac AS18 Anion-Exchange Columns

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

IonPac Trace Cation Concentrator

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Dionex IonPac AS28-Fast-4µm column

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

High-Performance Chromatography Superior isocratic separation of

columns IonPac SCS 1 Silica Cation Separator

Ion Chromatography (IC)

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

Determination of Methylamine in Drug Products

The Determination of Trace Anions in Concentrated Phosphoric Acid

Overcoming Challenging Matrices in Ion Chromatography

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Determination of Trace Anions in Organic Solvents

columns IonPac AS22 Anion-Exchange Column

Suppression in anion chromatography. More sensitive analysis of anions and organic acids

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Thermo Scientific. Anion-Exchange Column

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of Trace Anions in Concentrated Glycolic Acid

anion-exchange column

Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0

The determination of trace anions in concentrated phosphoric acid

Determination of Trace Anions in Concentrated Hydrofluoric Acid

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Theory of Ion Chromatography

Determination of trace anions in concentrated glycolic acid

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Ion Analysis of Hydraulic Fluids in Fracking Process

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Determination of trace anions in organic solvents using matrix elimination and preconcentration

suppressors Eluent Suppressors for Ion Chromatography

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

Inline sample preparation for the determination of anions in sodium hydroxide

ION CHROMATOGRAPHY SYSTEM S 150

Thermo Scientific Dionex eluent suppressors for ion chromatography

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Determination of Common Inorganic Anions in Environmental Waters Using a Hydroxide Selective Column

"Analysis of Inorganic Anions in Potomac Water, Sediment and Floc by Ion Chromatography" FINAL REPORT. by James E. Girard, Ph.D.

Determination of Cr(VI) in Water, Waste Water, and Solid Waste Extracts

ISO INTERNATIONAL STANDARD

Ion chromatography. Ion chromatography

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis

Air Monitoring. Semi-continuous determination of ambient air quality

Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography

Electrodeionization (EDI)

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

Sulfur speciation in mining leachates by ion exchange with direct UV detection, using a perchlorate mobile phase.

881 Compact IC pro / 882 Compact IC plus for ion chromatography. Intelligent. Reliable. Precise.

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

Automation of the radiochemical procedures for the sequential separation of radionuclides

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

PRODUCT MANUAL. for the. IONPAC CG14 GUARD COLUMN (2 x 50 mm, P/N ) (4 x 50 mm, P/N )

2011_ST20_Taylor.pdf. SSPT Document not peer-reviewed by CORESTA

IonPac SCS 1 IonPac SCG 1

Determination of trace anions in high-nitrate matrices by ion chromatography

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

Data Pack. Ion Chromatography Methods. June 2009

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

California Environmental Protection Agency Air Resources Board

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Transcription:

Novel Electrodeionization Devices: Applications in Inorganic Analysis John M. Riviello and Archava Siriraks ABSTRACT Electrodeionization (EDI) is a well established technique for the production of ultrapure water (UPW). In this paper the authors discuss novel electrodeionization devices which have been developed primarily for use in chemical analysis. These EDI devices can be incorporated directly into the analytical instruments for on-line production of UPW, or can be used off-line for automated sample dilution or sample preparation. Applications of these EDI devices in trace analysis of inorganic ions will be shown. INTRODUCTION Electrodeionization (EDI) is a well established technique for the production of ultrapure water (UPW). Conventional EDI uses mixed bed depletion chambers for the removal of contaminant ions. Recent advances in EDI cell chemistries have resulted in the improved removal of weakly ionized contaminants and improved resistance to fouling from calcium and magnesium [1]. To date, all commercial EDI systems have been developed for process and laboratory level UPW production. UPW is the most commonly used reagent in inorganic analysis. Techniques such as ion chromatography, atomic spectroscopy and inorganic mass spectrometry rely on UPW in order to achieve low detection limits. In most instances, water is taken from a bench top laboratory water purifier and used to make reagents or transferred to a container for sample processing. This non-point-of-use process can result in contamination of the water from container leaching, bacterial growth or absorption of contaminates from the air. In addition, the quality of the UPW produced from bench top water purifiers varies over time as the mixed polishing beds become expended. This variation in water quality results in inconsistent chemical blanks, which compromise detection limits. EDI offers the advantage of being a dynamic process and as a result can produce a consistent UPW product. We have used this advantage of EDI to develop point-of-use EDI devices, which are designed for integration into the analytical instrument and analysis process. In this paper we will demonstrate the use of these EDI devices in trace analysis using ion chromatography (IC). EXPERIMENTAL The Trovion CIRA electrolytic purifiers (Trovion Pte. Ltd., Singapore) were used as point-of-use EDI devices for IC analysis. Figure 1 shows a schematic of a CIRA purifier used for the analysis of anions by IC. Figure 2 shows a CIRA purifier used for the analysis of cations by IC. The CIRA devices were used with a Dionex RFIC (Reagent Free Ion Chromatography) ICS-1500 ion chromatograph (Dionex Corp, Sunnyvale, CA). For anion analysis by IC, a Dionex AS 17 separator (2 mm) column was used with an ASRS300 suppressor. Potassium hydroxide eluent was generated using the RFIC electrolytic eluent generator. For cation analysis, a Dionex CS12A separator column (2 mm) was used with a CSRS300 suppressor. Methane sulfonic acid was generated using the RFIC electrolytic eluent generator. The Dionex RFIC electrolytic eluent generation system uses a deionized water feed to electrochemically produce eluents. A Trovion CIRA 10X electrolytic purifier was used to deionize the water inline with the electrolytic eluent generation system. The CIRA 11A electrolytic purifier is a layered EDI device with two discreet depletion chambers as shown in Figure 1. The CIRA purifiers are designed to operate in a flow rate range of 0.1 to 10 ml min 1 and their compact size makes them ideal for integrating into analytical instrumentation. CIRA can be configured so that water flows through one or both depletion chambers in parallel or serial fashion. For the experiments performed below, the suppressed anion eluent (water containing trace anions) exiting the conductivity cell of the IC was passed through the anion-mixed bed depletion chamber. The product water is deionized and the anion level in the product water is below 0.1 ng kg 1. This UPW is then used to process the next 2009 by PowerPlantChemistry GmbH. All rights reserved. 32 PowerPlant Chemistry 2009, 11(5)

PPChem Figure 1: Schematic of a CIRA 11A electrolytic purifier. Figure 2: Schematic of a CIRA 11C electrolytic purifier. sample or standard. The UPW from the CIRA 11A is used to rinse the large volume sample loop, which prevents sample carry-over and lowers the blank. The cation depletion chamber of the CIRA 11A can be used to remove cations from the sample or from the suppressor waste stream. Table 1 shows some of the properties of the CIRA 11A and 11C. For cations analysis, the suppressed eluent (water containing trace cations) is passed through the cation-mixed bed depletion chamber. The product water is deionized and the cation level in the product water is below 0.1 ng kg 1. The UPW from the CIRA 11C is used to rinse the large volume sample, which prevents sample carry-over and reduces the blank. The anion depletion chamber of the CIRA 11C can be used to remove anions from the sample or the suppressor waste stream. This approach of purifying the waste stream and using the purified waste stream for calibration and sample processing is referred to as IonPrep. The diagram in Figure 3 shows this basic configuration of the IC system for ultra trace analysis with IonPrep and the CIRA purifier [2]. With the addition of a large volume autosampler, this configuration allows for fully automated calibration and analysis at part-per-trillion (ppt or ng kg 1 ) levels. Note that in the configuration shown in Figure 3, only one analytical pump is required. Sample and standards are drawn into the sample loops using a peristaltic pump. This configuration allows for fully automated operation, and the CIRA purifier results in extremely low and consistent chemical blanks. Figure 4 shows a Dionex RFIC ICS-1500 ion chromatograph configured with the Trovion IonPrep system with CIRA for ultra trace analysis. Property 11A 11C Layered chamber Anion-mixed bed Cation-mixed bed Void volume (layered chamber) 800 µl 800 µl Homogeneous chamber Cation Anion Void volume (homogeneous chamber) 200 µl 200 µl Typical voltage 25 V @ 30 ma 25 V @ 30 ma Flow rate 0.1 to 10 ml min 1 0.1 to 10 ml min 1 Table 1: Properties of CIRA electrolytic purifiers. PowerPlant Chemistry 2009, 11(5) 33

Figure 3: IC schematic using CIRA electrolytic purifier. ASRS ASRS suppressor Figure 4: IC system with IonPrep for trace ion analysis. Figure 5: IC anion blank with IonPrep. 34 PowerPlant Chemistry 2009, 11(5)

PPChem Figure 6: IC anion calibration at the ng kg 1 level using IonPrep. RESULTS Figure 5 shows a typical system blank that results when using a 20 ml sample loop. The trace anions (sub ng kg 1 ) shown in this chromatogram result from concentrating 20 ml of the conductivity cell waste which was purified by passing through the anion-mixed bed depletion chamber of the CIRA 11A. The peak at 11.8 min is carbonate, which is not analyzed in this system. Figure 6 shows an overlay of a calibration run at the 1, 5 and 10 ng kg 1 levels for the common anions. The calibration is performed automatically by using a single partper-billion anion standard, loading the small loop (10 µl) and then diluting with the CIRA water in the 20 ml loop. This results in a 2 000 fold dilution of the ng kg 1 standard, which allows for calibration at the ng kg 1 level. This novel approach is feasible as a result of the point-of-use UPW produced from CIRA [3]. The UPW from CIRA is used for dilution of the standard and rinsing of the system between samples. As a result, detection limits in the low to sub ng kg 1 level can be achieved routinely for monitoring ionic contamination in UPW. CONCLUSION Point-of-use electrolytic water purifiers, based on EDI technology, have been developed for integration into analytical instrumentation. Using the electrolytic purifiers with ion chromatography, we have demonstrated the ability to measure common ions at the low part-per-trillion level. The point-of-use electrolytic purifiers will find use in other inorganic analytical applications including automated sample pretreatment and reagent purification and preparation. REFERENCES [1] Liang, L. S, Ultrapure Water Journal 2003, 20(October), 13. [2] Product Document LPN1919-01, 2008. Dionex Corporation, Sunnyvale, CA, U.S.A. [3] US Patent Application, 2006. US2006-0231403. Courtesy of the Engineers' Society of Western Pennsylvania. Paper presented at the 69th International Conference, San Antonio, TX, U.S.A., October 26 30, 2008. PowerPlant Chemistry 2009, 11(5) 35

THE AUTHORS John M. Riviello (B.S., Chemistry, San José State University, M.S., Chemistry, University of California, Santa Cruz, both in California, U.S.A.) is a co-founder and chief executive officer of Trovion Company. He has over 25 years of experience in the analytical instrument industry, primarily in chromatography R&D. John Riviello has developed major technologies for trace metal analysis, especially in the area of interfacing ion chromatography with inorganic spectroscopy. His extensive research with electrolytic ion exchange resins and membranes led directly to the development of a new class of ion chromatography instrumentation. John Riviello has worked for Dionex Corporation and Transgenomic, Inc. in a variety of technical and management positions. He holds 17 US patents, with several patents pending, and has authored 15 publications in peer-reviewed journals. CONTACT John M. Riviello Trovion Company 2591 S. Bascom Ave, Ste. I Campbell, CA 95008 U.S.A. E-mail: jriviello@trovion.com Archava Siriraks Trovion Pte. Ltd. 14 Little Rd #05-02 Tropical Industrial Bld Singapore 536987 E-mail: asiriraks@trovion.com Archava (Dew) Siriraks (B.S., Agricultural Chemistry, De La Salle Araneta University, Manila, Philippines, Ph.D., Analytical Chemistry, American University, Washington, D.C., U.S.A.) is a co-founder, chief scientific officer and operations manager of Trovion Company. He is a recognized expert in ion chromatography and trace analysis. Archava Siriraks previously worked for Dionex Corporation as a research scientist and has extensive experience in ion chromatography applications development. For the past nine years, Archava Siriraks has owned and operated Archemica International, an analytical instrument distributor and service provider based in Bangkok, Thailand. 36 PowerPlant Chemistry 2009, 11(5)