EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Similar documents
EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Analyst Software. Automatic Optimization Tutorial

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Version 1.0 October 2007

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Application Note. Abstract. Authors. Environmental

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada

Overview. Introduction

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Determination of urea in ultrapure water by IC-MS/MS

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

APPLICATIONS TN Increased Capabilities for the Analysis of Hormones in Drinking and Waste Water Using Solid Phase Extraction (SPE) and LC/MS/MS

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis

4000 Q TRAP LC/MS/MS System. Advanced Linear Ion Trap technology at the highest level of sensitivity 4000 QTRAP. LC/MS/MS System

The Power of LC MALDI: Identification of Proteins by LC MALDI MS/MS Using the Applied Biosystems 4700 Proteomics Analyzer with TOF/TOF Optics

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

A S ENSIT IV E M E T HO D FO R T H E D E T E RM INAT IO N O F ENDO C RIN E- D IS RU P T ING COM P OUNDS IN RIV E R WAT E R BY L C / MS/MS

Benzodiazepines and Barbiturates

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Supporting Information

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

isobaric Barbiturates in Serum using LDTD- MS/MS combined with differential mobility

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Figure 1. The supported liquid extraction process using the ISOLUTE SLE plate (single well shown)

Reagents. Affinity Tag (Biotin) Acid Cleavage Site. Figure 1. Cleavable ICAT Reagent Structure.

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

16 Malachite green 17 Leucomalachite green

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Supporting Information. Sweetened Swimming Pools and Hot Tubs

Determination of trace anions in concentrated hydrofluoric acid

A New Polymer-based SPE Sorbent to Reduce Matrix Effects in Bioanalytical LC-MS/MS

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

+ butanal + 2-picolineborane. + acetic acid. + butanal + 2-picolineborane. + acetic acid

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

Determination of Pharmaceuticals in Environmental Samples

APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEY WORDS. Simultaneous high-throughput screening of melamine (MEL) and cyanuric acid (CYA).

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

Definitive EtG/EtS LC-MS/MS Analysis:

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Transcription:

Application Note EPA Method 535 EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS API 3200 LC/MS/MS System Overview Described here is the analysis of Chloroacetanilide and Acetamide degradates in drinking water using Liquid Chromatography Tandem Mass Spectrometry, LC/MS/MS. This analysis follows U.S. Environmental Protection Agencies Method 535 guidelines for sample preparation and analysis. The method used two Multiple Reaction Monitoring (MRM) transitions per analyte for both quantitation and confirmation. Monitoring a second MRM transition for each target compound adds an additional order of confidence when looking at dirty matrices, therefore, the possibility of false positive detection is reduced. Detection limits in drinking water were determined to be 0.002 to 0.004µg/L using established guidelines. Introduction The EPA has established Method 535 for the analysis of ethanesulfonic acid (ESA) and oxanilic acid (OA) degradates of chloroacetanilide herbicides in drinking water and surface water. Chloroacetanilide herbicides are extensively used for weed control on crops throughout the US. In this method degradates of Alachlor, Acetochlor and Metolachlor are monitored and a Lowest Concentration Minimum Reporting Level (LCMRL) of 0.012 to 0.014µg/L was validated for all compounds. Chromatography has also been set up to include the degradates of Dimethenamid, Flufenacet, and Propachlor in future work. The surrogate and internal standard used for this method were Dimethachlor ESA and Butachlor ESA respectively. Despite using a specific MS/MS scan, Alachlor ESA and Acetochlor ESA are structural isomers (Figure 1) that have similar product ions of m/z 80 and 121. For this reason, consistent chromatographic resolution is necessary and was achieved using a shallow stepping gradient on a Restek Ultra C 18 column (Figure 2).

Results showed consistent performance for both standards and samples over several days of work. From the product ion spectra in Figure 1, there are several unique product ions for both Alachlor ESA, m/z 158, 160, 176, and Acetochlor ESA, m/z 144, 146, 162. These product ions are not as sensitive as m/z 80 and 121, therefore they were not used. As an alternative to the method described below, using the unique product ions for both Alachlor ESA and Acetochlor ESA would eliminate the need for baseline chromatographic separation. It is important to note that using product ions other than m/z 80 and 121 will result in loss of sensitivity. Both matrix effects, such as ion enhancement and ion suppression, due to high total organic carbon (TOC) were a real concern during method development. Surrogate recoveries, matrix spikes, MDLs, and internal standard recoveries were monitored but did not indicate any interference using the method below with drinking water. Experimental The method uses an Applied Biosystems API 3200 LC/MS/MS system equipped with Turbo V source and Electrospray Ionization probe. All compounds were detected using negative ionization in Multiple Reaction Monitoring (MRM) mode using two MRM transitions for each target compound and surrogate. The following mass spectrometer conditions were used: Curtain Gas interface: 25psi, IS voltage: -4500V, Gas1: 50psi. Gas2: 50psi, Ion source temperature: 500 C, Collision gas: Medium, Interface heater: On, Vertical probe position: 2, Horizontal probe position: 5, Dwell time: 50ms. Figure 1. Product Ion spectra of structural isomers Alachlor ESA (top) and Acetochlor ESA (bottom). Figure 2. Reproducible chromatography was achieved using a gradient on a Restek Ultra C 18 3µm 100x2.1mm column. A 50ng/mL initial calibration point is shown. Sufficient baseline separation was achieved for structural isomers Alachlor ESA and Acetochlor ESA with a consistent resolution factor of 3.5 or greater.

An Agilent 1200 HPLC system was used consisting of a binary pump, autosampler with thermal unit, and column oven. Chromatographic separation was achieved on a Restek Ultra C 18 3µm 100x2.1mm column using mobile phases, A: 5mM ammonium acetate, B: methanol with the gradient in Table 1. A 25µL injection volume was used. Sample preparation was performed using Solid Phase Extraction (SPE). Restek Carbora 90, 6.0mL tube size, SPE cartridges were used. Cartridges were conditioned with 20mL of 10mM ammonium acetate/methanol and then rinsed with 30mL reagent water. Cartridges were not allowed to go dry at any time during the sample loading process. After conditioning, 250mL of sample was prepared by adding 25-30mg ammonium chloride and spiked with 5µL of a 12µg/mL surrogate standard, mixed, and then loaded using a vacuum manifold at a flow rate of 10-15mL/min. After loading, each cartridge was rinsed with 5mL reagent water and then allowed to dry using nitrogen. Cartridge elution was performed using 15mL of 10mM ammonium acetate/methanol at gravity flow. The extracts were concentrated to dryness using a gentle stream of nitrogen in a heated water bath, 60-70 C. Finally, 1mL of 5mM ammonium acetate/reagent water and 10µL of a 5µg/mL internal standard solution were added and transferred to an autosampler vial. Six calibration points (0.5, 1.0, 10.0, 50.0, 100.0, and 125.0ng/mL) prepared in 5mM ammonium acetate/reagent water, were used for the initial calibration curve. A linear fit was used with 1/x weighting and a correlation coefficient, r, of 0.995 or greater was achieved. All stock and primary dilution standards were prepared in methanol and stored at 4 C. Table 1: LC GRADIENT ON A RESTEK ULTRA C 18 COLUMN Time (min) Flow Rate (µl/min) A (%) B (%) 0.00 250 80 20 4.00 250 70 30 10.0 250 70 30 15.0 250 50 50 17.0 250 15 85 18.0 250 15 85 18.1 250 80 20 28.0 250 80 20 Table 2: Detection in Multiple Reaction Monitoring (MRM) Analyte Quantifier Quantifier MRM Ratio Range MRM MRM (±20%) Alachlor OA 264 / 160 264 / 158 0.24 0.36 Acetochlor OA 264 / 146 264 / 144 0.23 0.35 Alachlor ESA 314 / 80 314 / 121 0.28 0.42 Metolachlor OA 278 / 206 278 / 174 0.10 0.15 Acetochlor ESA 314 / 80 314 / 121 0.33 0.50 Metolachlor ESA 328 / 80 328 / 121 0.32 0.48 Dimethachlor ESA (Surrogate) Results and Discussion Method development was performed using the automated Quantitative Optimization feature of the Analyst Software. Each target compound, surrogate, and internal standard was infused into the mass spectrometer 300 / 80 300 / 121 0.39 0.59 Butachlor (IS) 356 / 80 - - Two MRM transitions were monitored for all target analytes and the surrogate. Only 1 MRM was used to monitor the internal standard (IS). at a low flow rate of 10µL/min. Quantitative Optimization identified the precursor ion, the most sensitive product ions for each compound, and optimized all compound dependant parameters automatically. Results are shown in Table 2.

MRM area ratios were used for confirming detection. Each ratio, displayed in Table 2, was calculated by determining the MRM ratio of each calibration standard and then taking the average of all standards. A ±20% range was then applied to each unknown sample. The Analyst Reporter automatically flagged any unknown sample with a calculated MRM ratio outside the established 20% range. Once method development was completed an Initial Demonstration of Capability (IDC) was performed. First an initial demonstration of low system background was run by preparing a Laboratory Reagent Blank (LRB). For each analyte, detection in the prepared LRB needed to be < 1/3 of the MRL detection (Figure 3). To validate the proposed MRL, seven replicate LRBs were spiked at a concentration of 0.013µg/L and processed through the sample preparation procedure above. All chromatographic peaks for both quantifier and qualifier MRM transitions required a signal to noise ratio of at least 3:1. Using the proposed procedure for calculating an LCMRL in EPA Method 535 a calculated detection limit of 0.004µg/L or less was determined for all analytes (Table 3). After the MRL was confirmed an Initial Demonstration of Precision on Accuracy was performed. Four replicate LRBs were fortified at a concentration of 0.2µg/L. The Percent Relative Standard Deviation (%RSD) for all analytes was 20% and the average recovery was within ±30% of the true value. Therefore the method satisfied the precision and accuracy requirements (Table 4). Figure 3. Extracted Ion Chromatograms (XIC) for the Laboratory Reagent Blank (LRB) for all target analytes are displayed in the left column and the XIC of the proposed Method Reporting Limit (MRL) are in the right column.

Table 3: Minimum reporting limit confirmation Analyte LCMRL Standard HRPIR Lower PIR Upper PIR Extract Sample On Column Deviation LOD (µg/l) LOD (µg/l) LOD (fg) Alachlor OA 0.013 0.28 1.1 72.4 142.3 0.868 0.003 86.8 Acetochlor OA 0.014 0.27 1.1 74.0 141.9 0.843 0.003 84.3 Alachlor ESA 0.013 0.18 0.7 82.1 127.9 0.569 0.002 56.9 Metolachlor OA 0.013 0.21 0.8 75.7 128.2 0.651 0.003 65.1 Acetochlor ESA 0.012 0.29 1.1 62.4 134.7 0.897 0.004 89.7 Metolachlor ESA 0.012 0.18 0.7 76.1 122.5 0.576 0.002 57.6 Finally the recoveries of the internal standard and surrogate were monitored over a period of 48 hours. Samples were analyzed consecutively over this time and the recovery and %RSD of Dimethachlor ESA, surrogate, and Butachlor ESA, internal standard, were calculated. The results, shown in Table 5, for the surrogate indicated that the sample preparation efficiency is acceptable. In addition, internal standard recoveries show that the mass spectrometer is maintaining consistent sensitivity over long analysis times. Most importantly, results of both QC analytes indicate that there is no ion suppression or enhancement taking place that may affect the results of the target analytes. TABLE 4: Initial demonstration of precision and accuracy Analyte Average Recovery (%) % RSD Alachlor OA 96.6 8.5 Acetochlor OA 97.0 8.9 Alachlor ESA 92.5 8.6 Metolachlor OA 95.0 8.5 Acetochlor ESA 94.3 8.0 Metolachlor ESA 94.8 8.9 Dimethachlor ESA (Surrogate) 100.1 9.2 Table 5: Internal standard and surrogate recoveries Analyte Spike Level (μg/l) Average % RSD Recovery (%) Dimethachlor ESA (Surrogate) 0.24 97.3 12.7 Butachlor ESA 0.20 87.4 23.4 Surrogate and internal standard recoveries were within acceptable limits of 70-130% and 50-150% respectively.

Conclusion A method following US EPA guidelines for Method 535 has been presented. This method shows superior sensitivity for ethanesulfonic acid and oxanilic acid degradates of chloroacetanilide herbicides Alachlor, Acetochlor, and Metolachlor using SPE and LC/MS/MS with an API 3200 system. A Lowest Concentration Minimum Reporting Level (LCMRL) of 0.012 to 0.014µg/L was verified with a calculated detection limit of 0.004µg/L or less. Two MRM transitions were used for both quantitation and confirmation of target analytes. Surrogate and internal standard recoveries indicate that there is no matrix interference. This method is easily transferable to any Applied Biosystems/MDS SCIEX LC/MS/MS system and is available upon request. Author Christopher Borton (Applied Biosystems, Foster City, CA) Acknowledgements The authors would like to thank Dr. Paul Winkler at GEL Analytics in Golden, Colorado for providing sample preparation materials, Dr. Ali Haghani of MWH laboratories for providing standards of all compounds for this study and Restek Corporation for supplying the HPLC column and SPE cartridges. Reference 1 U.S. Environmental Protection Agency Method 535: Measurement of Chloroacetanilide and other Acetamide Herbicide Degradates in Drinking Water by Solid Phase Extraction and Liquid Chromatography / Tandem Mass Spectrometry (LC/MS/MS) Version 1.1 (April 2005) J.A. Shoemaker, M.V. Bassett Not for use in diagnostic procedures. 2007 Applera Corporation and MDS Inc. All Rights Reserved. Applera, Applied Biosystems and AB (design) are registered trademarks of Applera Corporation or its subsidiaries in the US and/or certain other countries. API 3200, Curtain Gas and Turbo V are trademarks and Analyst is a registered trademark of Applied Biosystems/MDS SCIEX, a joint venture between Applera Corporation and MDS Inc. All other trademarks are the sole property of their respective owners. 11/2007 Publication 114AP65-01 Headquarters 850 Lincoln Centre Drive Foster City, CA 94404 USA Phone 650.638.5800 Toll Free 800.345.5224 www.appliedbiosystems.com International Sales For our office locations please call the division headquarters or refer to our Web site at www.appliedbiosystems.com/about/offices.cfm