Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Similar documents
Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Study of Residual Solvents in Various Matrices by Static Headspace

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

USP<467> residual solvents

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Headspace Technology for GC and GC/MS: Features, benefits & applications

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Fast methods for the determination of ibuprofen in drug products

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Introduction to Fourier Transform Infrared Spectroscopy

Analysis of BTEX in Natural Water with SPME

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

HILIC Method Development in a Few Simple Steps

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

STANDARD OPERATING PROCEDURES

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

Monitoring Protein PEGylation with Ion Exchange Chromatography

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

of mass spectrometry

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

A Capillary Gas Chromatographic Procedure for the Analysis of Nine Common Residual Solvents in Water-Insoluble Bulk Pharmaceuticals

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Fast, Effective XPS Point Analysis of Metal Components

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Application. Gas Chromatography February Introduction

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

ApplicationNOTE ULTRA TRACE ANALYSIS OF DIOXINS AND FURANS IN HUMAN ADIPOSE TISSUE USING SFE-LC EXTRACTION/CLEANUP AND THE WATERS AUTOSPEC ULTIMA NT

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

GAS CHROMATOGRAPHY (GC)

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

Complete Materials Deformulation Using TGA-IR

Advances in Sample Preparation for Accelerated Solvent Extraction

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Introduction to Fourier Transform Infrared Spectroscopy

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Transcription:

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling Andrea Caruso and Massimo Santoro, Thermo Fisher Scientific, Milan, Italy Application Note 1316 Key Words USP Method <467>, residual solvents, excipients, headspace analysis, valve and loop sampling Goal The purpose of this application note is to describe the analysis of residual solvents in pharmaceutical products in accordance with USP Method <467>. Introduction US Pharmacopeia (USP) method <467> delineates the procedure to identify and quantify any organic volatile solvent present in pharmaceuticals and excipients as a result of the production process. The method details the chromatographic system and analytical conditions to be used and addresses the acceptable limits for each solvent. This application note describes the analysis of residual solvents in accordance with USP method <467> using the Thermo Scientific TRACE 131 GC, the new Thermo Scientific TriPlus 3 Headspace autosampler, and the Thermo Scientific Dionex Chromeleon chromatography data system (CDS) software. The residual solvents tested in this work, are chemical residues present at trace levels in pharmaceutical substances and their products. These solvents may be the result of the manufacturing process or from packaging and storage. Pharmaceutical products should contain low levels of residual solvents as determined by safety data. The USP 467 method sets the concentration limit for each solvent and describes the analysis of 53 solvents grouped according to their health hazards: Class 1 solvents (known to cause unacceptable toxicities or to have environmental effects) will be avoided in the manufacturing process. Class 2 solvents (associated with less severe toxicity) will be limited. Class 3 solvents (less toxic) will be used where practical. GC headspace (HS) analysis is recommended for identification and quantification of Classes 1 and 2 solvents. Experimental Three analytical procedures are used for identification and quantification of the residual solvents: Procedure A screens for and confirms the solvents presence using a G43 (volatiles) column. Procedure B confirms the solvents identity using a G16 (wax) column. Procedure C, which also uses a G43 column, is required to quantify the amount of residual solvents present. The USP method suggests the use of a GC and a valve-and-loop headspace sampling system. Accordingly, a TRACE 131 GC and the TriPlus 3 Headspace autosampler are used here. The data are collected and processed with Chromeleon 7.1 CDS. A Thermo Scientific TraceGOLD TG 624 3 m.32 mm 1.8 µm (P/N 2685-339) column is used for Procedures A and C. A TraceGOLD TG WaxMS 3 m.32 mm 5 µm (P/N 2688-143) column is used for Procedure B.

2 Samples and Standard Preparation Standard stock solutions for Class 1, Class 2 mix A, and Class 2 mix B residual solvents are prepared as described by the method: Class 1: transfer 1 ml of USP Class 1 mixture RS (Sigma-Aldrich 4131-U) to a 1 ml flask previously filled with 9 ml dimethyl sulfoxide, dilute to volume with water and mix. Transfer 1 ml of this solution to a 1 ml flask previously filled with 5 ml water, dilute to volume with water, and mix. Transfer 1 ml of this solution to a 1 ml flask previously filled with 5 ml water, dilute to volume with water, and mix. Class 2 mix A: Transfer 1 ml of USP Class 2 mix A RS (Sigma-Aldrich 4132-U) to a 1 ml flask, dilute to volume with water, and mix. Class 2 mix B: Transfer 1 ml of USP Class 2 mix B RS (Sigma Aldrich 4133-U) to a 1 ml flask, dilute to volume with water, and mix. Prepare standard solutions for Class 1, Class 2 mix A, and Class 2 mix B residual solvents directly into the headspace (HS) vial as described by the method: Class 1: Transfer 1 ml of Class 1 standard stock solution to a 2 ml HS vial; add 5 ml of water, cap the vial, and mix. Class 2 mix A: Transfer 1 ml of Class 2 mix A standard stock solution to a 2 ml HS vial; add 5 ml of water, cap the vial, and mix. Class 2 mix B: Transfer 5 ml of Class 2 mix B standard stock solution to a 2 ml HS vial; add 1 ml of water, cap the vial, and mix. Prepare the test stock solution and the test solution as follows: Test stock solution: Transfer 25 mg of the article under test to a 25mL flask, dissolve, bring to volume with water, and mix. Test solution: Transfer 5 ml of Test Stock Solution to a 2 ml HS vial; add 1 ml of water, cap the vial, and mix. Analysis Procedure A The chromatographic system is equipped with an instant-connect split/splitless (SSL) injector, an instantconnect FID detector, and a 3m.32 mm 1.8 µm TG 624 column. Helium is used as carrier gas with a constant flow of 2.2 ml/min and a split ratio of 1:1. The instantconnect injector and instant-connect detector are kept at temperatures of 14 C and 25 C respectively. TriPlus 3 HS parameters are configured as follows: Equilibration time of the sample is 45 min with vial shaking set to high Oven temperature set to 8 C Manifold temperature set to 9 C Transfer line temperature set to 15 C Pressurization mode is set to Pressure, and the pressure is set to 1 bar and the pressure equilibration time to min Loop filling mode is set to Pressure, and the pressure is set to.5 bar with an equilibration time of min Loop size is 1 ml Injection mode is Standard, and injection time is set to.5 min Sample line is purged after injection for 1 min at 8 ml/min Vial venting is set to On The GC oven parameters are configured as follows: Hold at 4 C for 2 minutes Raise at 1 C/min until oven temperature reaches 24 C Maintain oven temperature at 24 C for 2 min According to USP <467>, the signal-to-noise ratio of 1,1,1 trichloroethane in Class 1 standard solution must be no less than 5, and the resolution between acetonitrile and methylene chloride in Class 2 mix A standard solution must be no less than 1. Figure 1 shows the signal-to-noise ratio of 1,1,1-trichloroethane in Class 1 standard solution obtained in this analysis, and figure 2 shows the resolution between acetonitrile and methylene chloride in Class 2 mix A standard solution. Figure 1. Signal-to-noise ratio (S/N) of 1,1,1-trichloroethane in Class 1 standard solution analyzed according to Procedure A. S/N calculated at 73.6 Figure 2. Resolution (R) between acetonitrile and methylene chloride in Class 2 mix A standard solution analyzed according to Procedure A. R = 1.12

If the peak response of any peak of the article under test other than 1,1,1-trichloroethane in the test solution is greater than or equal to a corresponding peak in the standard solution or if the peak corresponding to 1,1,1-trichloroethane is greater than or equal to 15 times the peak response corresponding to 1,1,1-trichloroethane in the Class 1 standard solution, the analyst must proceed to Procedure B to verify the identity of the peak on a different stationary phase, otherwise the article meets the method s requirements. Procedure B In Procedure B of USP <467>, the chromatographic system is equipped with an instant-connect split/splitless (SSL) injector, an instant-connect FID detector, and a TraceGOLD TG WaxMS 3 m.32 mm 5 µm column. The carrier gas used is helium with a constant flow of 2.2 ml/min and a split ratio of 1:1. The instant-connect injector and detectors are kept at temperatures of 14 C and 25 C respectively. The oven ramp is configured as follows: Hold at 5 C for 2 minutes Raise at 6 C per minute until temperatures reaches 165 C Maintain oven at 165 C for 2 minutes. This procedure is applied to confirm the peak identification of Procedure A, so a different stationary phase is used to analyze those solvents. The signal-to-noise ratio of benzene (in Class 1 standard solution) must be no less than 5, and the resolution between acetonitrile and cis-dichloroethene (in Class 2 mix A standard solution) must be no less than 1. Figures 3 and 4 show the signalto-noise ratio of benzene in Class 1 standard solution and the resolution between acetonitrile and cis-dichloroethene in Class 2 mix A standard solution, respectively. If the peak response of any peak found in the test solution is greater than or equal to a corresponding peak in the standard solution, the analyst must proceed to Procedure C to quantify the peak; otherwise the article meets the method s requirements. Procedure C For each peak identified and verified with Procedures A and B, prepare a standard stock solution according to the following instructions: Transfer an accurate volume of an individual standard corresponding to each residual solvent identified to a suitable container. Dilute with water to obtain a solution having a final concentration of 1/2 of the concentration limit value as stated by the USP method. Prepare a standard solution by transferring 1 ml of Class 1 standard stock solution to a 2 ml headspace vial, adding 5 ml of water, capping the vial, and mixing. Prepare a spiked test solution by transferring 5 ml of test stock solution to an appropriate headspace vial, adding 1 ml of the standard stock solution, capping the vial, and mixing. Figure 3. The signal-to-noise ratio (S/N) of benzene in Class 1 standard solution analyzed according to Procedure B. S/N calculated at 73.7 Figure 4. Resolution (R) between acetonitrile and cis-dichloroethene in Class 2, mix A standard solution, analyzed according to procedure B. R = 3.59 Perform the analysis following the instructions for Procedure A with the same instrument configuration on the standard solution, the test solution, and the spiked test solution. Calculate the amount in ppm for each residual solvent found using the formula: 5(C/W) [ru/(rst-ru)] C is the concentration in µg per ml of the reference standard in the standard stock solution; W is the weight in grams of the article under test used to prepare the test stock solution; ru and rst are the peak responses of each residual solvent obtained from the test solution and the spiked test solution, respectively. Linearity and repeatability tests To assess system linearity and the overall system performance, calibration curves are run for the solvents analyzed, preparing the calibration points as follows: For Class 1 solvents, prepare 6 calibration points: 2, 1,.5,,, and.5 fold the concentration of Standard stock solution, step 2 for each solvent. For Class 2 mix A solvents prepare 5 calibration points:,,.5,.2, and.1 fold the concentration of Standard stock solution for each solvent. For Class 2 mix B solvents prepare 5 calibration points:,,.5,.2, and.1 fold the concentration limit of Standard stock solution for each solvent. 3

4 Results and Discussion Figures 5 9 show examples of the calibration curves of the residual solvents benzene, dicholoroethane, THF, o-xylene, and Tetraline. All the calibration curves obtained for each standard show outstanding linearity..3 y = 31x -.16 R 2 =.9997 Benzene 7 y =.463x - 24 R 2 =.9996 o-xylene 5.15 6 5 4 3 2.5 1.5 1 1.5 2 Figure 5. Calibration curve of benzene 2.5 2 4 6 8 1 12 14 Figure 8. Calibration curve of o-xylene 5 5.5 y =.398x +.2 R² =.99946 Dichloroethane 1 2 3 4 5 6 Figure 6. Calibration curve of dichloroethane 2 1.8 1.6 1.4 1.2 1.8.6.4 y = 1.7173x +.1777 R 2 =.99913 Tetraline.4.6.8 1 1.2 Figure 9. Calibration curve of tetraline 5 y =.313X -.8 R 2 =.99998 THF 5 1 2 3 4 5 6 7 8 Figure 7. Calibration curve of THF

The repeatability was tested for each residual solvent under the conditions of Procedures A and B; results are in Tables 1 and 2. Excellent repeatability for both retention time and peak area were achieved as reported in Tables 1 and 2, even at concentrations much lower than the Safety Concentration limit stated in the USP <467> method. 5 Table 1. Repeatability for residual solvents using USP <467> Procedure. Procedure A Class 1 RSD % (n = 5) Class 2 mix A RSD % (n = 5) Class 2 mix B RSD % (n = 5) 1,1-Dichloroethene.95 Methanol.79 Hexane 1.19 1,1,1-Trichloroethane 1.3 Acetonitrile.65 Nitromethane 1.15 Benzene.89 Dichloromethane 1.6 Chloroform.68 1,2-Dichloroethane.94 trans 1,2-Dichloroethene 1.53 1,2-Dimethoxyethane 1.23 cis 1,2-Dichloroethene 1.38 Trichloroethylene 1. Tetrahydrofuran.58 Pyridine 1.43 Cyclohexane 1.13 2-Hexanone.51 Methylcyclohexane 2.76 Tetraline.71 1,4-Dioxane 1.19 Toluene 1.5 Chlorobenzene 1.22 Ethylbenzene 1.34 m-xylene/p-xylene 1.14 o-xylene 1.27 Table 2. Repeatability for residual solvents using USP <467> Procedure B. Procedure B Class 1 RSD % (n = 5) Class 2 mix A RSD % (n = 5) Class 2 mix B RSD % (n = 5) 1,1-Dichloroethene.93 Methanol 1.79 Hexane 1.46 1,1,1-Trichloroethane 1.6 Acetonitrile 1.67 Nitromethane 1.48 Benzene 1.5 Dichloromethane 1.56 Chloroform.47 1,2-Dichloroethane 1.4 trans 1,2-Dichloroethene 3.39 1,2-Dimethoxyethane.85 cis 1,2-Dichloroethene 1.3 Trichloroethylene.95 Tetrahydrofuran 1.31 Pyridine.82 Cyclohexane 2.86 2-Hexanone.71 Methylcyclohexane 1.61 Tetraline.42 1,4-Dioxane 2.87 Toluene 2.9 Chlorobenzene 2.25 Ethylbenzene 2.56 m-xylene/p-xylene 1.89 o-xylene 1.63

6 Results and Discussion The chromatograms for each class mixture are shown in Figures 1 to 17. Figure 1. Class 1 standard solution analyzed according to Procedure A Figure 14. Class 2 mix A analyzed according to Procedure B Figure 11. Class 2 mix A analyzed according to Procedure A Figure 15. Class 2 mix A analyzed according to Procedure B zoomed in on dioxane Figure 12. Class 2 mix B analyzed according to Procedure A Figure 16. Class 2 mix B analyzed according to Procedure B Figure 13. Class 1 analyzed according to Procedure B Figure 17. Class 2 mix B analyzed according to Procedure B

Figure 18 shows the headspace chromatogram obtained when following Procedure A on a sample made by mixing the three solutions (Class 1, Class 2 mix A, and Class 2 mix B). Application Note 1316 Figure 18. Headspace chromatogram of a sample obtained by mixing the three solutions tested according to procedure A. (Class 1, Class 2 mix A and Class 2 mix B) Conclusion All the USP <467> method acceptance criteria were met, and the chromatograms show good peak shapes and resolution. These data demonstrate that headspace-gas chromatography is a quantitative technique that easily meets the requirements of the USP <467> method. The system configuration of the TRACE 131 GC, TriPlus 3 Headspace, and Chromeleon CDS is a solid, fully integrated, and reliable platform to perform the analysis according to the USP <467> method with full auditing capability. Operating this system configuration and method in a regulated environment is highly possible obtaining accurate, dependable results. References United States Pharmacopeia, Chemical Tests, <467> Residual Solvents, official from August 212. www.thermoscientific.com 213 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Sigma-Aldrich is a trademark of Sigma-Aldrich Co., LLC. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa-Other +27 11 57 184 Australia +61 3 9757 43 Austria +43 1 333 5 34 Belgium +32 53 73 42 41 Canada +1 8 53 8447 China +86 1 8419 3588 Denmark +45 7 23 62 6 Europe-Other +43 1 333 5 34 Finland/Norway/Sweden +46 8 556 468 France +33 1 6 92 48 Germany +49 613 48 114 India +91 22 6742 9434 Italy +39 2 95 591 Japan +81 45 453 91 Latin America +1 561 688 87 Middle East +43 1 333 5 34 Netherlands +31 76 579 55 55 New Zealand +64 9 98 67 Russia/CIS +43 1 333 5 34 South Africa +27 11 57 184 Thermo Fisher Scientific S.p.A. Milan, Italy is ISO 91:28 Certified. Spain +34 914 845 965 Switzerland +41 61 716 77 UK +44 1442 233555 USA +1 8 532 4752 AN1316_E 2/13S