Flat Fields and Flux Calibrations for the COS FUV Channel at Lifetime Position 4

Similar documents
Verification of COS/FUV Bright Object Aperture (BOA) Operations at Lifetime Position 3

COS FUV Dispersion Solution Verification at the New Lifetime Position

SMOV Absolute Flux Calibration of the COS FUV Modes

Updates to the COS/NUV Dispersion Solution Zero-points

COS/FUV Spatial and Spectral Resolution at the new Lifetime Position

Characterizing the COS OSM1 Drift in the Dispersion Direction

COS Cycle 17: 20 programs 149 external orbits 446 internals FUV Detector Sensitivity Monitor 36 0 Oct. 3, 2010

Cycle 20 COS Calibration Plan. Jerry Kriss & COS/STIS Team 9/13/2012

Summary of COS Cycle 23 Calibration Plan

Cycle 24 COS Calibration Plan. Paule Sonnentrucker Gisella De Rosa & COS Team 9/12/2016

Wavelength Calibration Accuracy for the STIS CCD and MAMA Modes

Extraction of Point Source Spectra from STIS Long Slit Data

COS FUV Focus Determination for the G140L Grating

Absolute Flux Calibration for STIS First-Order, Low-Resolution Modes

Improvements to the COS FUV G130M and G160M Wavelength Solutions at Lifetime Position 1

Cycle 21 COS Calibration Plan. Julia Roman-Duval Justin Ely & COS/STIS Team 9/10/2013

Calibration of ACS Prism Slitless Spectroscopy Modes

Impacts of focus on aspects of STIS UV Spectroscopy

Updated flux calibration and fringe modelling for the ACS/WFC G800L grism

Predicted Countrates for the UV WFC3 Calibration Subsystem using Deuterium Lamps

Delivery of a new ACS SBC throughput curve for Synphot

Fringe Correction for STIS Near-IR Long-Slit Spectra using Contemporaneous Tungsten Flat Fields

New On-Orbit Sensitivity Calibration for All STIS Echelle Modes

The TV3 ground calibrations of the WFC3 NIR grisms

Cosmic Origins Spectrograph Instrument Mini-Handbook for Cycle 12

The in-orbit wavelength calibration of the WFC G800L grism

Cosmic Origins Spectrograph Instrument Handbook for Cycle 23

The WFC3 IR Blobs Monitoring

Dark Rate of the STIS NUV Detector

Improved Photometry for G750L

SBC FLATS: PRISM P-FLATS and IMAGING L-FLATS

Cosmic Origins Spectrograph Instrument Handbook for Cycle 21

NICMOS Status and Plans

Summary of the COS Cycle 18 Calibration Program

Reduction procedure of long-slit optical spectra. Astrophysical observatory of Asiago

COS Monthly Status Review. August 19, Ball

Introduction of near-infrared (NIR) spectroscopy. Ken-ichi Tadaki (NAOJ)

ACS CCDs UV and narrowband filters red leak check

EXPOSURE TIME ESTIMATION

Study of the evolution of the ACS/WFC sensitivity loss

Cosmic Origins Spectrograph Instrument Handbook for Cycle 19

WFC3 IR Blobs, IR Sky Flats and the measured IR background levels

arxiv: v1 [astro-ph.im] 1 Jun 2016

NEON Archive School 2006

Calibration Goals and Plans

SOLSTICE I and II: Ultraviolet Variability. Marty Snow, Bill McClintock, Tom Woods, and Gary Rottman Laboratory for Atmospheric and Space Physics

A Fast Algorithm for Cosmic Rays Removal from Single Images

NIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.

The HST/STIS Next Generation Spectral Library

Grism Sensitivities and Apparent Non-Linearity

Lab 4 Radial Velocity Determination of Membership in Open Clusters

Optical/IR Observational Astronomy Spectroscopy. David Buckley, SALT

UVIT IN ORBIT CALIBRATIONS AND CALIBRATION TOOLS. Annapurni Subramaniam IIA (On behalf of the UVIT team)

Exploring Data. Keck LRIS spectra. Handbook of CCD Astronomy by Steve Howell Chap. 4, parts of 6

WFC3/UVIS Photometric Transformations

arxiv:astro-ph/ v1 12 Nov 2003

A Python Script for Aligning the STIS Echelle Blaze Function

Trace and Wavelength Calibrations of the UVIS G280 +1/-1 Grism Orders

Observer Anomaly(?): Recent Jitter and PSF Variations

arxiv: v1 [astro-ph.im] 13 Jun 2012

The Accuracy of WFPC2 Photometric Zeropoints

Cross-Talk in the ACS WFC Detectors. I: Description of the Effect

Astronomical Techniques

Investigating the Efficiency of the Beijing Faint Object Spectrograph and Camera (BFOSC) of the Xinglong 2.16-m Reflector

THE OBSERVATION AND ANALYSIS OF STELLAR PHOTOSPHERES

PACS Spectroscopy performance and calibration PACS Spectroscopy performance and calibration

Report on the new EFOSC2 VPH grisms

Impressions: First Light Images from UVIT in Orbit

XMM-Newton Optical-UV Monitor: introduction and calibration status OM instrument and calibration

Persistence in the WFC3 IR Detector: Spatial Variations

The Effective Spectral Resolution of the WFC and HRC Grism

First On-orbit Measurements of the WFC3-IR Count-rate Non-Linearity

Physics 476LW Advanced Physics Laboratory Atomic Spectroscopy

UNIVERSITY COLLEGE LONDON. Measuring a Stellar Spectrum

Ultra-narrow-band tunable laserline notch filter

Astronomical frequency comb for calibration of low and medium resolution spectrographs

Astronomy. Optics and Telescopes

WFC3/IR Persistence as Measured in Cycle 17 using Tungsten Lamp Exposures

An Algorithm for Correcting CTE Loss in Spectrophotometry of Point Sources with the STIS CCD

Here Be Dragons: Characterization of ACS/WFC Scattered Light Anomalies

AS750 Observational Astronomy

In this lab you will measure and quantify emission spectra from several different visible light sources.

arxiv:astro-ph/ v3 7 Apr 2005 Space Telescope Science Institute 2, 3700 San Martin Drive, Baltimore, MD 21218, U.S.A.

Studies of diffuse UV radiation

Performance of the NICMOS ETC Against Archived Data

Optical/NIR Spectroscopy A3130. John Wilson Univ of Virginia

Chopping and Nodding for Mid-Infrared Astronomy Kevin Volk December 2007

To determine the wavelengths of light emitted by a mercury vapour lamp by using a diffraction grating.

FUV Grating Performance for the Cosmic Origins Spectrograph

Detection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics

Southern African Large Telescope

Astrometric Performance of STIS CCD CTI Corrections on Omega Cen Images

SBC L-Flat Corrections and Time-Dependent Sensitivity

AST 301, Lecture 2. James Lattimer. Department of Physics & Astronomy 449 ESS Bldg. Stony Brook University. January 29, 2019

TECHNICAL REPORT. Doc #: Date: Rev: JWST-STScI , SM-12 August 31, Authors: Karl Gordon, Ralph Bohlin. Phone:

Future HST Observations of Europa and its Plumes: Findings and Recommendations

Astro 500 A500/L-15 1

The x Factor: Determining the Strength of Activity in Low-Mass Dwarfs

Optimal resolutions for optical and NIR spectroscopy S. Villanueva Jr.* a, D.L. DePoy a, J. L. Marshall a

You, too, can make useful and beautiful astronomical images at Mees: Lesson 3

Transcription:

Instrument Science Report COS 2018-20(v1) Flat Fields and Flux Calibrations for the COS FUV Channel at Lifetime Position 4 William J. Fischer 1, Marc Rafelski 1, and Gisella de Rosa 1 1 Space Telescope Science Institute, Baltimore, MD 25 September 2018 ABSTRACT As part of the calibration of the fourth lifetime position (LP4) of the Cosmic Origins Spectrograph (COS) Far-Ultraviolet (FUV) detector, we derived low-order flat fields (L flats) and sensitivities. To do this, spectra of the star WD 0308 565 were obtained with the G130M, G160M, and G140L gratings, and spectra of the star GD 71 were obtained with the G160M grating, all through the Primary Science Aperture (PSA). Observations were executed with all cenwaves and all FP-POS with the exception of G130M/1055 and G130M/1096, which remained at LP2. The L flats and sensitivities at LP4 were derived with the same method as at LP3. The L flats at LP4 agree with those at LP3 to within 2%, depending on grating and detector position. The sensitivities at LP4 are as expected after accounting for the previously characterized time-dependent sensitivity of the FUV detector. The relative and absolute flux calibration accuracies at LP4 are within the specifications of 2% and 5%. Contents Introduction (page 2) Observations (page 3) Generation of L Flats and Sensitivity Curves (page 4) Comparison of LP4 Flats and Sensitivities to Those at LP3 (page 7) Operated by the Association of Universities for Research in Astronomy, Inc., for the National Aeronautics and Space Administration.

Scientific Validation of Flats and Sensitivities (page 7) Conclusions (page 12) Change History (page 12) References (page 12) 1. Introduction The COS FUV detector is susceptible to gain sag, a reduction in the ability of the detector to convert incoming photons to electrons that becomes more severe with continued use. To counteract this effect, the location where FUV spectra fall on the detector (the lifetime position or LP) is periodically moved to an unused and therefore pristine region. On July 23, 2012, spectra were moved to LP2, offset by +3.5 from LP1 in the cross-dispersion direction. On February 9, 2015, spectra were moved to LP3, offset by 2.5 from LP1. Operations of the COS FUV detector at LP4, located at 5.0 from LP1, started on October 2, 2017 for all cenwaves except 1055 and 1096 of the G130M grating, which remained at LP2. At each new lifetime position, the resolution changes, worsening with increased angular separation from LP1 (Fox et al. 2018, ISR 2018-07), the flat field may change due to the new location of spectra on the detector, and the throughputs are remeasured to verify their consistency with the established requirements. To maintain the targeted absolute and relative flux calibration accuracies for COS (< 5% and < 2%, respectively) and meet other specifications, each change in lifetime position requires a new calibration effort. In this ISR, which is part of a series describing the LP4 calibration program, we report on the generation of new flat fields and sensitivity curves for LP4. The COS flat field is conventionally separated into an L flat, which contains variations in the detector response that have low spatial frequency, and a P flat, which accounts for pixel-to-pixel variations. The L flats are available as reference files and are incorporated in the standard COS calibration routine. The P flats are not applied in routine calibration. Instead, the influence of pixel-to-pixel variations can be reduced with the standard COS observing strategy, where spectra are obtained at multiple positions (FP-POS) spaced along the dispersion direction of the detector and then co-added. The maximum signal-to-noise ratio attainable with this strategy depends on the grating and segment and varies from 30 to 50 (Table 5.6 of the COS Instrument Handbook). If higher signal-to-noise ratios are desired, P flats may be useful; they are available upon request from the STScI help desk. At LP1 and LP2, L flats were constructed by iteratively decomposing net-count spectra into a position-dependent flat field and a wavelength-dependent spectrum. The flat field was interactively fit with a low-order polynomial or spline function, and sensitivity curves were then calculated from spectra that had been flat-fielded with the final Instrument Science Report COS 2018-20(v1) Page 2

Table 1. Spectra Used for Flat Creation and Flux Calibration Program 14910 Star Grating Cenwaves Segment WD 0308 565 G130M 1222, 1291, 1300, 1309, 1318, 1327 BOTH WD 0308 565 G160M 1577, 1589, 1600, 1611, 1623 BOTH WD 0308 565 G140L 1105 FUVA WD 0308 565 G140L 1280 BOTH GD 71 G160M 1577, 1589, 1600, 1611, 1623 FUVA L flats (Massa et al. 2014, ISR 2013-09). This approach required input from the user in the form of manual selection of nodes for the fitting and manual creation of a function to fit the 1180 Å feature (Section 3.1). A different procedure was used at LP3 (Debes et al. 2016, ISR 2016-15). In the revised procedure, there is limited input from the user, with a fixed number of equally separated nodes for each spectrum, although the assumptions are retained that, for a given detector setting, the flat field depends only on position and the sensitivity curve depends only on wavelength. Debes et al. showed that their more easily reproducible approach gave similar results to the original one. At LP4 we continue to use the procedure developed for LP3. In this report we describe the construction of L flats and sensitivity curves for LP4 based on data that were obtained there in April 2017, before the move to LP4 for general observing programs. In Section 2 we list the calibration observations used in the program, and in Section 3 we review the methodology for constructing the flats and sensitivity curves and plot them. In Section 4 we compare the results to those obtained at LP3, in Section 5 we discuss the scientific validation of the files, and in Section 6 we give our conclusions. 2. Observations The spectra used to derive the flats and sensitivities were obtained through the Primary Science Aperture (PSA) at LP4 in program 14910, Fourth COS FUV Lifetime Position: Cross-Dispersion Profiles, Flux, and Flat-Field Calibration (PI M. Rafelski), which was executed in April 2017. Table 1 lists the observed cenwaves. The only cenwaves that were not observed are those that remained at LP2 (G130M/1055 and 1096). Spectra of the hot subdwarf WD 0308 565 were obtained with all modes, and, due to the diminishing flux of that target at long wavelengths, spectra of the white dwarf GD 71 were obtained for segment A of all G160M cenwaves. All modes were observed at all four FP-POS. Instrument Science Report COS 2018-20(v1) Page 3

3. Generation of L Flats and Sensitivity Curves Our approach assumes that the flat fields reflect the detector response and are thus a function of the detector coordinates only, while the sensitivity curves describe the remaining throughput of the telescope plus instrument and are thus a function only of wavelength. Flats are generated for each grating and segment, while sensitivity curves are generated for each grating, segment, and cenwave. We used the IDL procedure derive cos flats.pro (G. Becker) to generate the flat fields and sensitivity curves with the method outlined in Debes et al. (2016). The net counts are assumed to be the product of the flux from the source, the flat field, and the sensitivity, or N(x, λ) = S(λ)F (x)r(λ), where N(x, λ) are the net counts at detector position x and wavelength λ, S(λ) is the flux, F (x) is the flat field, and R(λ) is the response (sensitivity) function. We observed HST standards, where S λ is well known, and we used multiple cenwave and FP-POS settings to sample each position on the detector at multiple wavelengths. This breaks the degeneracy between x and λ, allowing the flat and the sensitivity to be constrained independently. To fit the L flats, we used a cubic spline with 20 nodes spaced evenly over the range in x covered by well calibrated data. For the sensitivity curves, we used piecewise quadratic interpolation. Generally there were 15 nodes over the range in λ covered by all four FP-POS, except for G140L, where 30 points were used, and for segment B of G130M, where 20 points were used to give flexibility in fitting the 1180 Å feature (Section 3.1). The fitting is performed on net counts spectra that were calibrated using a flat field that contains only grid wire shadows and impostors. The shadows are due to grid wires, 17 on segment A and 22 on segment B, that improve the quantum efficiency of the detector and induce a drop in response of about 20% over a width of about 50 pixels (Ely et al. 2011, ISR 2011-03). The impostor grid wire shadows appear only on segment B and are an artifact introduced by the geometric distortion correction (Section 5.8.1 of the COS Instrument Handbook). Each FUV grating (G130M, G160M, G140L) and segment (A or B) has its own flat field of 16384 columns by 1024 rows. (Relatively little data are obtained on segment B with G140L, so this flat is a copy of the one for G130M/B.) The L flats derived here depend on column (the dispersion direction), and each is replicated over 1024 rows (the cross-dispersion direction). Grid-wire shadows and impostor shadows, which are the only features included in the flats that vary with row, are then added back into the flats. The sensitivity curves are provided in a FITS table with one entry for each combination of cenwave, segment, and aperture. They are corrected for the time-dependent sensitivity (TDS) of each grating as measured in FUV TDS calibration programs (De Rosa 2018, ISR 2018-09). Figures 1 and 2 show the one-dimensional flats and sensitivity curves derived with these procedures. Sensitivities for the Bright Object Aperture Instrument Science Report COS 2018-20(v1) Page 4

Figure 1. L flats for LP4. In the delivered reference files, the L flats shown here, which depend only on detector column, are replicated for each row of the detector. The signatures of the gridwire shadows and impostors are then included. Because only the extreme red end of G140L/B is well calibrated, the flat for G130M/B is used for that segment. (BOA), not shown, were obtained by multiplying those derived for the PSA by the wavelength-dependent transmission curve of the BOA (Figure 8 of Massa et al. 2010, ISR 2010-02). Instrument Science Report COS 2018-20(v1) Page 5

Figure 2. Sensitivity curves at LP4 for each cenwave. Curves for segments A and B are shown in the same panel. The curves for G140L are also shown with a logarithmic vertical axis to highlight their behavior at short wavelengths. Instrument Science Report COS 2018-20(v1) Page 6

3.1 Removal of the 1180 Å Feature Wood s anomalies (Wood 1902; Maystre 2012) appear near wavelengths λ = d(sin α ± 1)/n, where d is the groove separation of the grating, n is a non-zero integer, and α is the incidence angle. This is the wavelength for which a scattered wave propagates tangentially to the grating surface. For the COS G130M grating, λ d, and it is therefore subject to such an anomaly, which manifests as a dip in sensitivity on Segment B, near 1180 Å. Since the wavelength of a Wood s anomaly depends on the incidence angle, the wavelength of the dip varies mildly with cenwave as the optics select mechanism (OSM1) is rotated to different positions. Figure 3 shows how we fit and remove the 1180 Å feature by including extra nodes in the fitting function. The fitting function then automatically traces the feature, and it is removed when we divide by the fit. No assumptions are made about the central wavelength, depth, or shape of the sensitivity depression. 4. Comparison of LP4 Flats and Sensitivities to Those at LP3 In Figure 4 we compare the L flats derived for LP3 by Debes et al. (2016) to those we derived for LP4. The L flats differ by up to 2% over most of each segment, with variations of up to 8% at the edges of segment A. These are likely due to subtle differences in the response of the detector at 2.5 from LP3. The COS FUV gratings and detector segments decline in sensitivity with time. This time-dependent sensitivity (TDS) is characterized and reported elsewhere (e.g., De Rosa 2017, ISR 2017-10). A direct comparison of sensitivities between LP3 and LP4 is not possible due to the difficulty in disentangling the TDS from other effects, but there is no indication of a significant change in sensitivity at LP4 beyond that expected from TDS. 5. Scientific Validation of Flats and Sensitivities In this section we demonstrate that the L flats and sensitivity curves can be applied to spectra of the net counts to yield accurately flux-calibrated spectra. In Figures 5 and 6, we show the effect of the flat correction. Uncorrected counts, in the left column of the figure, disagree at a given wavelength when sampled at different positions on the detector with different cenwaves. After the flat fields are applied, on the right column of the figure, each cenwave gives the same number of counts at each wavelength. This shows that the spectra have been corrected for elements of the detector response that vary slowly with position. The flat correction leaves cenwave 1222 mildly out of agreement with the other G130M cenwaves on segment A; the effect is more significant on segment B. This was Instrument Science Report COS 2018-20(v1) Page 7

Figure 3. Removal of the 1180 Å feature from G130M/B spectra. Purple arrows point to the affected wavelengths. The extreme cenwaves are plotted to show how the wavelength of the feature increases slightly with increasing cenwave, from 1170 Å for cenwave 1222 to 1186 Å for cenwave 1327. The top plot of each pair shows the net counts, the spline nodes, and the sensitivity function. (For clarity, the sensitivity function is scaled to align with the spectrum.) The bottom plot of each pair shows how spectra calibrated with a sensitivity function that includes the 1180 Å feature align well with the model, including in the region around the feature. also the case at LP3. We have assumed that the conversion from counts to flux takes place in two steps: a flat correction that depends only on grating, not cenwave, and a Instrument Science Report COS 2018-20(v1) Page 8

Figure 4. Comparison of L flats at LP3 and LP4 (gray and salmon curves). The blue curves show their ratios. (The units on the vertical axis apply to the flats; their ratios are dimensionless.) Except at the edges of the segments, the L flats from LP3 and LP4 generally agree to within 2%. flux correction that depends on the cenwave. For the extreme OSM1 rotation of cenwave 1222, the assumption that the flat field is independent of cenwave appears to be mildly violated. The residual disagreement of 1222 with other cenwaves is removed in the flux calibration, as demonstrated in the following paragraphs. Instrument Science Report COS 2018-20(v1) Page 9

Figure 5. Spectra of WD 0308 565 that demonstrate the effectiveness of the flat correction for segment A. Cenwaves are color-coded the same way as in Figure 2. The flat correction is successful because, after it is applied (right column), spectra from different cenwaves agree at each wavelength. Emission features at 1302 Å (top row) and 1216 Å (bottom row) are due to O I and Ly α airglow, respectively. For G130M, cenwave 1222 (pink) is not completely brought into agreement with the others. This is rectified in the flux calibration step; see the text for discussion. Instrument Science Report COS 2018-20(v1) Page 10

Figure 6. Spectra of WD 0308 565 that demonstrate the effectiveness of the flat correction for segment B. Cenwaves are color-coded the same way as in Figure 2. The flat correction is successful because, after it is applied (right column), spectra from different cenwaves agree at each wavelength. Emission features at 1216 Å and 1302 Å (top row) are due to Ly α and O I airglow, respectively. For G130M, cenwave 1222 (pink) is not completely brought into agreement with the others, and the 1180 Å feature persists. These issues are rectified in the flux calibration step; see the text for discussion. G140L has just one cenwave that uses segment B, so no comparison is possible. In Figures 7, 8, and 9, we show the effect of the flux calibration. Spectra calibrated with the L flats and sensitivities derived for LP4 are binned by 0.6 Å for the M modes and 4.8 Å for the L modes. The spectra agree with the models in that the residuals average to about zero and have little bin-to-bin scatter. Over all modes, the mean difference between the calibrated observed flux and model flux is at most 0.5%. The standard deviations in the observed-to-model flux ratios are small. For the M modes other than 1222/B, they are less than 1.5%. At the blue end of 1222/B, the spread is larger; it is 1.1% at wavelengths greater than 1100 Å but climbs to 2.9% at shorter wavelengths. For the L modes, the standard deviations of the residuals are less Instrument Science Report COS 2018-20(v1) Page 11

than 1.8% between 1100 and 2000 Å, and they rise to 5 8% above and below this range, where the absolute sensitivities are low. 6. Conclusions The method developed to derive flats and sensitivities at LP3 continues to work well at LP4. We applied this method to spectra of WD 0308 565 and GD 71 obtained in April 2017 at LP4. The resulting L flats differ from those derived for LP3 by 2% depending on detector position and grating, and the sensitivities are as expected after accounting for time dependence. The resulting flat fields bring adjacent cenwaves into excellent agreement, and the sensitivity curves generate spectra that agree with models to within the specifications of 2% relative and 5% absolute. Change History for COS ISR 2018-20 Version 1: 25 September 2018- Original Document References Debes, J. H., Becker, G., Roman-Duval, J., et al. 2016, Instrument Science Report COS 2016-15, Third COS FUV Lifetime Calibration Program: Flatfield and Flux Calibrations De Rosa, G. 2018, Instrument Science Report COS 2018-09, Cycle 24 COS/FUV Spectroscopic Sensitivity Monitor Ely, J., Massa, D., Ake, T., et al. 2011, Instrument Science Report COS 2011-03, COS FUV Gridwire Flat Field Template Fox, A., James, B., Roman-Duval, J., Rafelski, M., & Sonnentrucker, P. 2018, Instrument Science Report COS 2018-07, The Spectral Resolution of the COS FUV Channel at Lifetime Position 4 Massa, D., Ely, J., Osten, R., et al. 2014, Instrument Science Report COS 2013-09, Updated Absolute Flux Calibration of the COS FUV Modes Massa, D., Keyes, C., Penton, S., Bolin, R., & Froning, C. 2010, Instrument Science Report COS 2010-02, SMOV Absolute Flux Calibration of the COS FUV Modes Maystre, D. 2012, in Springer Ser. in Optical Sciences 167, Plasmonics, ed. S. Enoch & N. Bonod (Berlin: Springer-Verlag), 39 Wood, R. W. 1902, Philos. Mag. 4, 396 Instrument Science Report COS 2018-20(v1) Page 12

Figure 7. A demonstration of the accuracy and precision of the flux calibration for both segments with the G130M grating. Colored symbols are the COS data for WD 0308 565, binned by 0.6 Å. Colors map to cenwaves as in previous figures. The black curves are the model spectra. Absorption and airglow lines and the detector edges are excluded from the analysis. The overall agreement between the colored symbols and the black curves (< 0.5% mean difference) demonstrates good absolute flux calibration, while the small scatter in the colored symbols (standard deviations < 1.5% above 1100 Å) demonstrates good relative flux calibration. Instrument Science Report COS 2018-20(v1) Page 13

Figure 8. A demonstration of the accuracy and precision of the flux calibration for both segments with the G160M grating. Colored symbols are the COS data, binned by 0.6 Å. Colors map to cenwaves as in previous figures. The black curves are the model spectra. We used GD 71 for segment A and WD 0308 565 for segment B. Absorption and airglow lines and the detector edges are excluded from the analysis. The overall agreement between the colored symbols and the black curves (< 0.5% mean difference) demonstrates good absolute flux calibration, while the small scatter in the colored symbols (standard deviations < 1.5%) demonstrates good relative flux calibration. Instrument Science Report COS 2018-20(v1) Page 14

Figure 9. A demonstration of the accuracy and precision of the flux calibration for both segments with the G140L grating. Colored symbols are the COS data for WD 0308 565, binned by 4.8 Å. Colors map to cenwaves as in previous figures. The black curves are the model spectra. Absorption and airglow lines and the detector edges are excluded from the analysis. The overall agreement between the colored symbols and the black curves (< 0.5% mean difference) demonstrates good absolute flux calibration, while the small scatter in the colored symbols (standard deviations < 1.8% between 1100 and 2000 Å) demonstrates good relative flux calibration. Instrument Science Report COS 2018-20(v1) Page 15