Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Similar documents
Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

Determination of Pharmaceuticals in Environmental Samples

Benzoylecgonine in Urine by SAMHSA GC/MS

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

HYPHENATED TECHNOLOGY GUIDE

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

HYPHENATED TECHNOLOGY GUIDE

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Routine MS Detection for USP Chromatographic Methods

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Extraction of Aflatoxins and Ochratoxin from Dried Chili Using ISOLUTE. Myco Prior to LC-MS/MS Analysis

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Screening of pesticides in water using SPE on line

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Opiates in Urine by SAMHSA GC/MS

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Computer-assisted analysis of complex natural product extracts

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

The new Water Screening PCDL

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

LC/Q-TOF Workflows for Comprehensive Micropollutant Analysis

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

TANDEM MASS SPECTROSCOPY

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Agilent Standard Operating Procedure

A S ENSIT IV E M E T HO D FO R T H E D E T E RM INAT IO N O F ENDO C RIN E- D IS RU P T ING COM P OUNDS IN RIV E R WAT E R BY L C / MS/MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

Quantitative analysis of mitragynine in human urine by high performance liquid chromatography-tandem mass spectrometry

Quantification of Rice Aroma, 2-Acetyl-1-Pyrroline (2-AP), Using TurboMatrix Headspace Trap Coupled with GC/NPD and GC/MS

Accurate Mass Analysis of Hydraulic Fracturing Waters

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Separation and Characterization of Indian and Australian Sandalwood Oils

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

SUSPECT AND NON-TARGET SCREENING OF ORGANIC MICROPOLLUTANTS IN WASTEWATER THROUGH THE DEVELOPMENT OF A LC-HRMS BASED WORKFLOW

Agilent G3212 GC-APCI Source

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer

MRMS axelerate rapidly detected micropollutants and plant response metabolites in poplar leaves

A New Polymer-based SPE Sorbent to Reduce Matrix Effects in Bioanalytical LC-MS/MS

Determination of urea in ultrapure water by IC-MS/MS

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

U.S. EPA Method 8270 for multicomponent analyte determination

Application Note. Edgar Naegele. Abstract

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Transcription:

application Note Liquid Chromatography/ Mass Spectrometry Author Sharanya Reddy PerkinElmer, Inc. Shelton, CT USA Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF Introduction Identifying the presence of emerging pollutants in surface water samples is a growing area of concern in the environmental field. 1,2 Many of these pollutants are introduced into the surface waters anthropogenically through municipal waste water. Among the emerging pollutants, pharmaceuticals and personal care products (PPCPs) have been detected at parts per million and parts per trillion concentrations in surface waters. The presence of PPCPs suggests inefficient removal of these compounds by current sewage treatment processes. We present a study of PPCPs in river water samples from the northeastern United States using UHPLC-TOF-MS for both targeted and non-targeted analytes. Unlike a triple quadrupole, which is operated in multiple reaction monitoring mode for screening only predefined targeted analytes, the time-of-flight (TOF) mass spectrometer provides full spectrum accurate mass data that can be used to analyze and identify an unlimited number of compounds, without prior knowledge of target analytes or when reference standards are not available. In this study we show how high mass accuracy information provided by the PerkinElmer AxION 2 TOF along with the priopriatory AxION EC ID software can be used to identify unknown analytes in surface river waters.

Experimental Conditions Sample collection and preparation: River samples (400 ml) were collected in 1 L amber bottles. Samples were filtered through Whatman glass fiber filters (GF/C, 1.2 μm) and stored at 4 C until analysis. Prior to analysis, the samples were extracted through C-18 solid phase extraction (SPE) cartridges. SPE extraction: Phenomenex Strata -X SPE cartridges (500 mg/6 ml) were used for extraction. The cartridges were initially conditioned with methanol (5 ml) followed by water (5 ml). The filtered river sample was loaded on the cartridges with or without a spike of standard PPCPs (100 ng each) and extracted through the cartridge at 5-6 ml/min. The cartridge was dried under vacuum for ~20 mins prior to eluting with methanol (4 ml) and acidified methanol (containing 2% formic acid, 4 ml). The eluate was dried under nitrogen to ~0.3 ml and diluted to 0.5 ml with water. Results In the present study the AxION 2 TOF MS was used to analyze nine targeted PPCPs and several unknown compounds present in the river water samples. The separation of the nine target analytes was achieved under 3 mins on column (Figure 1). The SPE extraction procedure resulted in an estimated 75% or greater recovery of the analytes (based on 100 ppb standard spiked in MilliQ water). Using the AxION 2 TOF MS in the higher sensitivity TrapPulse mode, showed excellent sub ppb instrument detection limits for most analytes (detection limits for caffeine and theobromine were 4.0 and 8.0 ppb, respectively) and linearity over a very wide concentration range (0.5 to 500 ng/ml). Liquid chromatography conditions: Pump: PerkinElmer Flexar FX-10 pump Mobile phase A: Water containing 0.1% formic acid Mobile phase B: Acetonitrile containing 0.1% formic acid Gradient conditions: 10% B to 90% B in 5 min Injection volume: 3 μl in partial fill mode Column used: PerkinElmer Brownlee SPP C-18, 2X50 mm, 2.7 μm Figure 1. Shows the separation of PPCPs in under 2.5 min. Mass spectrometry conditions: Mass spectrometer: PerkinElmer AxION 2 TOF Ionization source: PerkinElmer Ultraspray 2 (Dual ESI source) Ionization mode: positive Pulse mode: 100-800 m/z Spectral acquisition rate: 1.5 spectra/sec Capillary exit voltage: 100 V Trap mode: 100-800 m/z (D7:42, D8:63) Internal calibration: Internal calibration was done using two ions m/z 118.08625 and 622.02896 as lock mass ions. 2

Figure 2. Identification of carbamazepine, diphenylhydramine, sulfamethoxazole and fluoxetine in river sample using accurate mass and retention time matching with standards. Identification of target and non-target analytes in river water The surface river water samples were collected downstream from a sewage treatment plant and were processed and analyzed as described above. This analysis revealed the presence of carbamazepine, diphenylhydramine, sulfamethoxazole and fluoxetine in the river water samples (Figure 2). The presence of these compounds was confirmed by accurate mass and retention time matching with standards. 3

Figure 3. Shows a major unknown target analyte eluting at ~2.7 mins. a b Figure 4 a and b. Accurate mass and isotope ratio of unknown analyte was entered into AxION EC ID software to determine elemental composition. The AxION EC ID software searches against the PubChem database and by sorting for decreasing AIDs (chemical listed for most number of assays), the first possible candidate for the given elemental composition was given as N, N-Diethyl-m-toluamide (DEET), a common ingredient in insect repellents. 4

Conclusions The PerkinElmer AxION 2 TOF operated in the higher sensitivity TrapPulse mode resulted in instrument detection limits of <1 ppb for majority of the PPCPs. The surface river water sample was screened for both target and non-target PPCPs. The accurate mass and isotope ratio provided by the TOF confirmed presence of several of the target PPCPs including antidepressant drugs in the surface waters. The non-target PPCPs in the sample were identified using accurate mass information along with powerful database search tools such as the AxION EC ID software. Figure 5. Presence of DEET was confirmed by accurate mass data The river sample data was further examined for unknown non-target analytes. A major chromatographic peak eluting at ~2.7 min (Figure 3) was analyzed using AxION EC ID software. Accurate monoisotopic mass and isotope ratio information is used by AxION EC ID to search the PubChem database (or other data bases) for potential molecular formula matches and provide a ranked summary of the potential matches as well as suggestions for possible compound structures for a given elemental composition. In this example, the elemental composition, C 12H 17NO was listed with highest score (Figure 4a) and the top candidate was identified as N, N-Diethyl-m-toluamide, (DEET) based on the number of active AIDS (Assay IDs) reported (Figure 4b). DEET is a common ingredient in insect repellent and the accurate mass of the unknown compound was verified to be within 2 ppm mass error of the expected mass of DEET (Figure 5). The presence of DEET in the river sample was further confirmed by retention time matching with the standard. Using a similar approach, acetaminophen, a commonly used analgesic drug was also identified in the river water (data not shown). References 1. Daughton, C.G., Jones-Lepp, T.L., Pharmaceuticals and personal care products (PPCPs) in the environment. Scientific and Regulatory Issues, ACS Symp. Ser. 791, Oxford University Press, Washington, U.S. (2001). 2. Daughton, C.G., Environ. Impact Assess. Rev. 24 (2004) 711. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2012, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 010131A_01