Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

Similar documents
Chloramine (Mono) and Nitrogen, Free Ammonia

Chlorine, Free and Total, High Range

Chlorine, Free and Total, High Range

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows

Method to 0.50 mg/l NH 3 N Powder Pillows

CHLORINE, TOTAL (0 to 4.00 mg/l)

Chlorine, Free and Total, Low Range

Nitrogen, Total Inorganic

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Quaternary Ammonium Compounds

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Peracetic Acid (PAA) and Hydrogen Peroxide (H 2 O 2 )

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

DPD Test N Tube Method *

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

Trihalomethane Formation Potential (THMFP)

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Oxygen Demand, Chemical

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

Oxygen Demand, Chemical

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

DR/4000 PROCEDURE. HARDNESS, Total

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

PAR Method Method to 2.0 mg/l Pb TNTplus 850

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

DR/4000 PROCEDURE. IRON, Total

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

NITROGEN, AMMONIA, High Range, Test N Tube

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

DR/4000 PROCEDURE LEAD. 2. The display will show: 3. Fill a 250-mL graduated cylinder to the 250-mL mark with sample.

NITROGEN, TOTAL, HR, Test N Tube

Mercaptoacetic Acid Method*

NITROGEN, TOTAL, Test N Tube * Method 10071

TPH (Total Petroleum Hydrocarbons)

NITRITE, Low Range (0 to mg/l NO 2- -N)

SILICA, Low Range (0 to 1.60 mg/l)

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Understanding Chlorine Measurement. Rebecca Luedee Environmental Sales

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

ALLOWAY METHOD OUTLINE

Manganese III Digestion Method * (with chloride removal)

Parameter Method Range # of Tests Code Page. Acidity (as % Oleic acid) titration % acidity 6 HI

Polychlorinated Biphenyls (PCB) in Soil

Reagents for Chemical Test Kits

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific

p-dimethylaminobenzaldehyde Method*

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

The content of this operation manual is protected by copyright laws. No part of this operation manual may be reproduced in any form.

Analysis of Free Ammonia in Chloramination Applications Using Lab Method and the APA6000

GENERAL INSTRUCTIONS GENERAL PREPARATION

METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation)

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

Photometer AL250. Scroll Memory (SM) N.I.S.T Traceability. Highlights

RAPID KJELDAHL BENCHNOTES

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

SMART Spectro. L Mott TEST INSTRUCTIONS. v MN. Spectro pho to m eter

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

TCEQ Directed Assistance Module (DAM) No. 5: Understanding and Controlling the Chloramination Process

The test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.

SILICA, Low Range SiLR

AP Chemistry Summer Assignment

Ground Water & UST Samples: Containers, Preservation and Hold Times Table

SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33

Nitrogen total LR VARIO TT mg/l N b) Persulphate Digestion

Transcription:

Chlorine Free, Indophenol, 10241 Indophenol 1 1 Patent pending. DOC316.53.01256 Method 10241 (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows Scope and Application: For determining residual free chlorine levels in the presence of manganese, chloramines and other oxidants which interfere with DPD colorimetric, DPD titrimetric, and amperometric methods for free chlorine. For use in potable water, chlorinated drinking water, swimming pool water and treated wastewater effluent. Test preparation How to use instrument-specific information The Instrument-specific information table shows requirements that can vary between instruments. To use this table, select an instrument, then read across to find the corresponding information required to perform this test. Table 1 Instrument-specific information Instrument Sample cell Cell orientation Adapter Program DR 5000 4864302 Orientation key toward user A23618 66 DR 3800 5940506 1-cm (flat) path aligned with arrow on adapter LZV585 (B) 66 DR 2800 5940506 1-cm (flat) path aligned with arrow on adapter LZV585 (B) 66 DR 2700 5940506 1-cm (flat) path aligned with arrow on adapter LZV585 (B) 66 DR/2500 5940506 1-cm (flat) path aligned with slots in cell holder 66 DR/2400 5940506 1-cm (flat) path aligned with light path. Refer to the user manual for more information. 5946600 66 DR/850 & DR/890 4864302 Slots in the 2 clock position in cell holder 110 Pocket Colorimeter II Monochloramine 4864302 Diamond mark faces keypad Chlorine Before starting the test: Put the protective cover or lid over the cell compartment during measurements. The sample and reagent from one analysis can contaminate other analyses and interfere with the test results. Be sure to rinse the cells and caps several times with deionized water or with the sample water to be tested before each test. Do not switch the sample cell caps between the Blank and Sample during the analysis. Tap sample cells lightly on a hard surface or slowly invert the cells to remove air bubbles from the cell walls. Be sure to keep the cap on the sample cells when not used to avoid ammonia contamination. Page 1 of 8

Collect the following items: Description Freechlor F Reagent Solution Quantity 5 drops Monochlor F Reagent Pillows 2 Sample cell (refer to Instrument-specific information) 2 See Consumables and replacement items for reorder information. Indophenol Method Stored Programs 66 Monochloramine LR Start 1. Select the test. Insert the correct adapter (refer to Instrument-specific information). 2. Fill 2 sample cells to the 10-mL line with sample. Mark one cell as the Blank. Mark the other cell as the Sample. 3. Add 5 drops of the Freechlor F Reagent to the Sample cell. 4. Cap and invert the cell to mix. Zero 5. Add the contents of one Monochlor F Reagent Powder Pillow to the Blank cell and to the Sample cell. Cap and shake both cells 20 seconds to dissolve. 6. Press TIMER>OK. A 5-minute reaction will start. If the sample temperature is less than 18 C (64 F), refer to Table 2 for the correct reaction time. 7. When the timer stops, invert the Blank cell to mix. Insert the cell into the cell holder. 8. ZERO the instrument. The display will show: 0.00 mg/l Cl 2 Page 2 of 8

Indophenol Method (continued) 9. Invert the Sample cell to mix. Insert the cell into the cell holder. READ the results in mg/l Cl 2. Color reaction time Use Table 2 to find the reaction time when the sample temperature is less than 18 C (64 F). The color is stable for 30 minutes after the reaction time is reached. If a temperature measurement is not available, read the sample after the color is stable. Table 2 Color development based on sample temperature Sample Temperature C F Reaction Time (minutes) 5 41 10 7 45 9 9 47 8 10 50 8 12 54 7 14 57 7 16 61 6 18 64 4 20 68 3 23 73 2.5 25 77 2 greater than 25 greater than 77 2 Page 3 of 8

Interferences Interfering substance The substances listed in Table 3 have been tested for interference and do not interfere at or below the indicated levels. Refer to Table 4 for substances that do interfere with the test. Table 3 Non-interfering substances Interference level Alanine 1 mg/l N Aluminum 10 mg/l Al 3+ Bromide 100 mg/l Br - Bromine 15 mg/l Br 2 Calcium 1000 mg/l as CaCO 3 Chloride 18,000 mg/l Cl - Chlorine Dioxide 5 mg/l ClO 2 Chromium (III) 5 mg/l Cr 3+ Copper 10 mg/l Cu Cyanide 10 mg/l CN - Dichloramine 10 mg/l as Cl 2 Fluoride 5 mg/l F - Glycine 1 mg/l N Iodine 4 mg/l I 2 Iron (II) 10 mg/l Fe 2+ Iron (III) 10 mg/l Fe 3+ Lead 10 mg/l Pb Manganese (7+) 3 mg/l MnO 4 Nitrate 100 mg/l NO 3 N Nitrite 50 mg/l NO 2 N Oxone 1 (potassium peroxomonopersulfate) 30 mg/l Phosphate 100 mg/l PO 3-4 Silica 100 mg/l SiO 2 Sulfate 2600 ppm SO 2-4 Tyrosine 1 mg/l N Urea 10 mg/l N Zinc 5 mg/l Zn 2+ 1 Oxone is a registered trademark of E.I. du Pont de Nemours & Co., Inc. Table 4 Interfering substances Interfering substance Interference level Ozone 1 > 1 mg/l O 3 Sulfide 1 > 0.5 mg/l S 2-1 This compound does not usually exist with free chlorine. Page 4 of 8

Sample collection, preservation and storage Testing applications Analyze samples for chlorine immediately after collection. Free chlorine is a strong oxidizing agent and it is unstable in natural waters. It reacts rapidly with various inorganic compounds and more slowly oxidizes organic compounds. Many factors, including reactant concentrations, sunlight, ph, temperature and salinity influence the decomposition of free chlorine in water. Avoid plastic containers because they can have a large chlorine demand. Pretreat glass sample containers to remove any chlorine demand by soaking in a dilute bleach solution (1 ml commercial bleach to 1 liter of deionized water) for at least 1 hour. Rinse thoroughly with deionized or distilled water. If sample containers are rinsed thoroughly with deionized or distilled water after use, only occasional pre-treatment is necessary. A common error in testing for chlorine is not obtaining a representative sample. If sampling from a tap, let the water flow for at least 5 minutes to ensure a representative sample. Let the container overflow with the sample several times, then cap the sample containers so there is no headspace (air) above the sample. If sampling with a sample cell, rinse the cell several times with the sample, then carefully fill to the 10-mL mark. Perform the chlorine analysis immediately. Finished chlorinated drinking waters and distributions systems Finished waters contain free chlorine and various levels of organic chloramines and inorganic contaminants. It is generally assumed that the free chlorine has reacted with all easily oxidizable species present and the remaining free chlorine is likely present in a steady-state equilibrium. Replicate analyses for free chlorine on this type of water should give equivalent results. It is especially important when testing water where free chlorine residual levels are low to observe all precautions that refer to sample cell cleanliness, water temperature and sampling techniques. At breakpoint These waters may contain a mixture of free chlorine, chloramines and nuisance residuals depending on water temperature, mixing efficiencies, sampling location and distance beyond the theoretical breakpoint. It is important to note that the water can be in a state of "dynamic equilibrium" and the chemical speciation can change rapidly, especially if one is at or near breakpoint. The chemical speciation can change dynamically in both the Blank cell and the Sample cell. The test must be conducted immediately on these types of samples. Test results may be difficult to replicate on duplicate samples depending on the dynamics of the water. Test results are best used to identify free chlorine trends and to monitor changes due to different mixing efficiencies, sampling locations, temperature changes, increased chlorine feed rates, etc. In chloramination kinetic studies These waters will contain a mixture of free chlorine and chloramines depending on water temperature, mixing efficiencies, sampling locations, feed rates for chlorine and ammonia and contact time. It is important to note that the water is in a state of "dynamic equilibrium" and the chemical speciation can change rapidly depending on water conditions. The chemical speciation can change dynamically in both the Blank cell and the Sample cell. The test must be conducted immediately on these types of samples. Test results may be difficult to replicate on duplicate samples depending on the dynamics of the water. Test results are best used to identify free chlorine trends and to monitor changes based on changes in mixing efficiencies, sampling locations, water temperature changes, increased chlorine feed rates, etc. With other oxidants such as Oxone, permanganate, chlorine dioxide, bromine and iodine It is assumed that the free chlorine residual has stabilized in the presence of the other oxidants. Replicate analyses for free chlorine on this type of water should give equivalent results. The levels of alternate oxidants that can be present without interference have been tested only in laboratory bench studies (refer to Table 3). Field data for free chlorine in the presence of these oxidants is not available. Page 5 of 8

Accuracy check Important Note: This procedure is only valid for stabilized or equilibrated free chlorine samples. Standard additions method (sample spike) Required for accuracy check: Chlorine Standard Solution, 2-mL PourRite Ampule, 25 30 mg/l PourRite Ampule Breaker Pipet, TenSette, 0.1 1.0 ml Pipet tips Method performance 1. After reading test results, leave the sample cell (unspiked sample) in the instrument. 2. Select standard additions from the instrument menu (if available). Refer to the instrument user manual for specific instructions. Refer to step 9 if a standard additions menu is not available. 3. Enter the average chlorine concentration from the label or certificate that is enclosed with the standard solution ampules. 4. A summary of the standard additions procedure will be displayed. Press OK to accept the default values for standard concentration, sample volume and spike volumes. After the values are accepted, the unspiked sample reading will appear in the top row. 5. Open one standard solution ampule. 6. Use the TenSette Pipet to prepare 3 spiked samples: add 0.1 ml, 0.2 ml, and 0.3 ml of standard to three 10-mL portions of fresh sample. 7. Follow the Indophenol Method test procedure for each of the spiked samples, starting with the 0.1 ml sample spike. Measure each of the spiked samples in the instrument. 8. Select GRAPH to view the results. Select IDEAL LINE (or best-fit) to compare the standard addition results to the theoretical 100% recovery. 9. If a standard additions menu is not available, calculate the percent recovery: a. Put the spiked sample in the cell holder and press READ. b. Calculate the concentration of chlorine that was added to the sample: 0.1 ml concentration of chlorine standard (mg/l Cl 2 ) mg/l chlorine added = ----------------------------------------------------------------------------------------------------------------------------------------------- 10.1 ml c. Find the expected result of the spiked sample from the sum of the unspiked sample result plus the concentration of chlorine that was added (step b). d. Calculate the percent recovery from the actual (step a) and the expected (step c) results. Note: If results are not within acceptable limits (± 10%), be sure that the sample volumes and sample spikes are measured accurately. The sample volumes and sample spikes that are used should agree with the selections in the standard additions menu. If all procedures are followed correctly but the standard additions results are not within acceptable limits, the sample may contain an interference. In a single laboratory with a standard solution of 3.51 mg/l chlorine (as Cl 2 ) and a single lot of reagent with a single instrument (DR 5000), a single operator obtained a standard deviation of ± 0.04 mg/l Cl 2. Page 6 of 8

Summary of method An ammonia solution at a ph of 8.3 is added to a sample containing free chlorine. The free chlorine is immediately converted into monochloramine (NH 2 Cl). The monochloramine is then determined by the indophenol method using Monochlor F Reagent. In this method, the monochloramine reacts with a substituted phenol in the presence of a cyanoferrate catalyst to form an intermediate monoimine compound. The intermediate compound couples with excess substituted phenol to form a green-colored indophenol compound, which is proportional to the amount of free chlorine present in the sample. A sample blank containing Monochlor F Reagent compensates for background color from the reagent and sample. Test results are measured with a spectrophotometer at 655 nm or with a colorimeter at 610 nm. Consumables and replacement items Required reagents Description Quantity/Test Unit Catalog number Freechlor F Reagent Solution 5 drops 50-mL SCDB 2964926 Monochlor F Reagent Pillows 2 100/pkg 2802299 Recommended standards Description Unit Catalog number Chlorine Standard Solution, 2-mL PourRite Ampule, 25 30 mg/l 20/pkg 2630020 Chlorine Standard Solution, 2-mL PourRite Ampule, 50 75 mg/l 20/pkg 1426820 Chlorine Standard Solution, 10-mL Voluette Ampule, 50 75 mg/l 16/pkg 1426810 Optional reagents and apparatus Description Unit Catalog number Ampule Breaker, 2-mL PourRite Ampules each 2484600 Ampule Breaker Kit, 10-mL Volulette Ampules each 2196800 Pipet, TenSette, 0.1 to 1.0 ml each 1970001 Pipet Tips, for TenSette Pipet 1970001 50/pkg 2185696 Pipet Tips, for TenSette Pipet 1970001 1000/pkg 2185628 Clippers (shears) each 2369400 Thermometer, non-mercury, 10 to 225 C each 2635700 Wipers, disposable, 28 x 37 cm 188/pkg 2932800 Page 7 of 8

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company, 2009. All rights reserved. Printed in the U.S.A. Updated July 2009, Edition 1