Qualitative and Quantitative Analysis of Pesticides in Horse Feed Matrix Using Orbitrap MS

Similar documents
Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Key Words Q Exactive Focus, Orbitrap, veterinary drugs, small molecule HRAM quantitation, small molecule HRAM screening, UHPLC, vdia

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

DETERMINATION OF PESTICIDES IN FOOD USING UPLC WITH POLARITY SWITCHING TANDEM QUADRUPOLE LC/MS/MS

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Automated and accurate component detection using reference mass spectra

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Quan/Qual Analyses. Unmatched Confidence for. Thermo Scientific Q Exactive Orbitrap LC-MS/MS System. Identify Quantify Confirm

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Introduction to Fourier Transform Infrared Spectroscopy

of mass spectrometry

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

Multi-residue analysis of pesticides by GC-HRMS

Pesticides in Cannabis

Fast, Effective XPS Point Analysis of Metal Components

Simplified Workflow and Improved Confidence in Screening Methods on Time of Flight Mass Spectrometry

for XPS surface analysis

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

LC-MS/MS multiresidue method used for pesticides analysis in fruits and vegetables

Multiresidue Analysis of 100 Pesticides in Food Samples by LC/Triple Quadrupole Mass Spectrometry Application

MetWorks Metabolite Identification Software

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Application News. No. C154. Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Safeguarding Food from Pesticides by UHPLC After Extraction with the QuEChERS Method

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Introduction to Fourier Transform Infrared Spectroscopy

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Overview. Introduction

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

An Evaluation of Various High-Resolution, Accurate-Mass Scan Modes for In Vitro Drug Discovery Screening

HILIC Method Development in a Few Simple Steps

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

Improving Experimental Efficiency for Drug Metabolite Profiling and Identification

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Triumphs and Challenges of High-Resolution Mass Spectrometry in Comprehensive Residue Screens

Advanced Data Acquisition and Data Processing Workflows to Identify, Quantify and Confirm Pesticide Residues

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Higher Confidence in Identification with QTRAP LC/MS/MS Systems when Screening and Quantifying Pesticides in Fruit and Vegetable Samples

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

IN QUALITATIVE ANALYSIS,

High-Field Orbitrap Creating new possibilities

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Targeted protein quantification

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

Trading Rules in Targeted/Unknown Multiresidue Pesticides Analysis Using UHPLC-Orbitrap MS and Data Mining Technologies

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

Fast and Accurate Identification of Pesticides by Direct Analysis in Real Time (DART) Ionization with Orbitrap Mass Spectrometry

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Determination of Pesticide Multiresidues in Apple, Pear and Grape using modified QuEChERS and analysis by LC-QTOF MS

Direct Analysis using Paper-Spray Mass Spectrometry: Method Development for the Rapid Screening of Drugs of Abuse for Forensic Toxicology

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

Rapid Pesticide Screening and Identification Using the High Resolution All Ions MS/MS Technique

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

dynamic MRM for pesticide analysis

Pure Chromatography. Food Safety Applications. Abstract. Introduction

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzcloud Library

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Transcription:

Qualitative and Quantitative Analysis of esticides in Horse Feed Matrix Using rbitrap M laf cheibner and Maciej Bromirski, Thermo Fisher cientific, Bremen, Germany Application Note 573 Key Words Exactive lus, UHLC, high resolution, accurate mass, high throughput, rbitrap, Data-Dependent All-Ion Fragmentation, ExactFinder Goal To test the ability of a high-resolution, accurate-mass benchtop rbitrap mass spectrometer to achieve high sensitivity and selectivity when analyzing modern, very-short-gradient UHLC separations of complex samples. Introduction roductivity of a liquid chromatograph-mass spectrometer (LC-M) system is measured in samples per day. To achieve higher productivity, modern ultra-high-performance LC-M (UHLC-M) methods use very short gradients. Chromatographic peak widths are often below 5 seconds at the base. A high-resolution, accurate-mass (HR/AM) mass spectrometer operating in full-scan mode must be able to provide a sufficient number of scans ( 10) across the chromatographic peak without compromising sensitivity and selectivity. As reported earlier, a resolving power in excess of 50,000 (FWHM at m/z 200) combined with a mass extraction window of 5 ppm is necessary to ensure selectivity comparable to established M/M techniques. 1 The Thermo cientific Exactive lus rbitrap mass spectrometer (Figure 1) is the second generation of the Exactive product family. It features two major changes over the first generation instrument. First, in the ion optics the tube-lens / skimmer assembly has been replaced by an -Lens (Figure 2) that provides significantly higher ion transmission, increasing the instrument s sensitivity. econd, the rbitrap mass analyzer and related electronics have been improved, 2 resulting in higher scan speed and resolution, as well as improved polarity switching. As a result, the range of resolving power is from 17,500 to 140,000 at m/z 200, with a maximum scan rate of 12 Hz. Figure 1. Exactive lus mass spectrometer with Accela 1250 UHLC Figure 2. Exactive lus ion optics and mass analyzer components In this research, the Exactive lus instrument was used to analyze extracts of horse feed spiked with common pesticides.

2 Experimental ample reparation QuEChER extracts of horse feed were spiked with 85 common pesticides (Table 1) at levels of 10 and 100 ppb, and diluted 1:1 with acetonitrile. ix calibration standards with the 85 pesticides in acetonitrile were mixed 1:1 with horse feed matrix that, through previous analysis, was proven to be free of pesticides. The final calibration levels were 5, 10, 25, 50, 100, and 150 ppb (5 150 µg/kg). Table 1. esticides spiked into QuEChER extracts esticide Chemical Formula esticide Chemical Formula Acephate C 4 N Indoxacarb C 22 ClF 3 7 Acetamiprid C 10 Cl Iprovalicarb C 18 H 28 Aldicarb Isofenphos-methyl N Aldicarb-sulfone Isofenphos-oxon C 15 H 24 N 5 Azinphos-ethyl C 12 2 Isoprothiolane C 12 2 Azinphos-methyl C 10 2 Isoproturon C 12 Azoxystrobin C 22 5 Bromacil C 9 H 13 Br Linuron C 9 Mepanipyrim H 13 Bromuconazole C 13 Br Carbaryl C 12 N Carbendazim C 9 Metconazole Methiocarb Methiocarb-sulfone C 17 Cl C 11 N C 11 N Carbofuran C 12 N Carbofuran-3-hydroxy C 12 N Methoxyfenozide C 22 H 28 Metobromuron C 9 Br Chlorfluazuron C 20 Cl 3 F 5 Monocrotophos N 5 Clofentezine H 8 Cymiazole C 12 Cymoxanil Cyproconazole C 15 Cl Cyromazine C 6 N 6 Demeton--methyl-sulfone C 6 5 2 Napropamide C 17 N Nitenpyram C 11 Cl methoate C 5 N xamyl H 13 encycuron C 19 Cl henmedipham C 16 Dichlorvos C 4 H 7 irimicarb C 11 Diethofencarb N Difenoconazole C 19 Diflubenzuron ClF 2 Dimethoate C 5 N 2 Disulfoton C 8 H 19 3 Disulfoton-sulfone C 8 H 19 3 Diuron C 9 H 1 0 Ethiofencarb C 11 N Fenamiphos C 13 N Fenazaquin C 20 Fenhexamid N Fenobucarb C 12 N rochloraz C 15 Cl 3 ropamocarb C 9 H 20 ropoxur C 11 N rosulfocarb N rosulfuron C 15 F 3 N 5 ymetrozine C 10 N 5 yraclostrobin C 19 Cl yridaphenthion pinosyn-a C 41 H 65 N 10 pinosyn-d C 42 H 67 N 10 piroxamine C 18 H 35 N Tebufenozide C 22 H 28 Fenoxycarb C 1 7H 19 N Tebufenpyrad C 18 H 24 Cl Fenthion C 10 2 Flucycloxuron C 25 H 20 ClF 2 Flufenoxuron C 21 ClF 6 Formetanate C 11 Furathiocarb C 18 H 26 5 Hexaflumuron C 16 H 8 F 6 Hexythiazox C 17 Cl Imazalil Teflubenzuron H 6 F 4 Tetraconazole C 13 F 4 Thiabendazole C 10 H 7 Thiacloprid C 10 Cl Thiodicarb C 10 3 Trichlorfon C 4 H 8 Cl 3 Trifloxystrobin C 20 H 19 F 3 Triflumuron C 15 ClF 3 Imidacloprid C 9 ClN 5

Liquid Chromatography A Thermo cientific Accela UHLC system consisting of an Accela open autosampler in combination with an Accela 1250 UHLC pump was used. A 2 minute chromatographic gradient of water and methanol, both spiked with 0.1% formic acid, was applied resulting in a total chromatographic cycle time of 5 minutes (Figure 3). Ten microliters of each sample were injected onto a Thermo cientific Hypersil GLD F column (50 x 2.1 mm, 1.9 µm particle size) with a flow rate of 800 µl/min. This resulted in peak widths of 3 6 seconds for the analytes of interest. For improved component identification, it would have been useful to have fragmentation scans on the analytes of interest. However, continual switching between full-scan and all-ion fragmentation scan modes (F/AIF) would have required resolution to be reduced to maintain the number of scans. As an optimal solution, data-dependent AIF scans (dd-aif) were introduced into the full scans (F/dd-AIF) by means of a mass inclusion list containing the masses of the spiked components. ne AIF scan was triggered for each target compound as soon as the abundance of the target compound crossed a given intensity threshold in a full scan. This significantly reduced the number of fragmentation scans and kept the overall data rate close to what could have been achieved in full-scanonly mode. Method details are shown in Figure 5. 3 Figure 3. Chromatographic gradient Mass pectrometry Given that resolution in excess of 50,000 was needed for this application, the Exactive lus system was set to a resolving power of 70,000 at m/z 200, resulting in a scan rate of 3.7 Hz. As shown in Figure 4, this provided 13 scans across a 3.2 second peak. Figure 4. cans achieved across a narrow chromatographic peak Figure 5. Exactive lus instrument method setup Data Analysis The same data set was used for quantitative and qualitative data processing. Thermo cientific ExactFinder software version 2.0 was used to process the data. Qualitative processing included targeted screening in combination with general unknown screening. The 85 common pesticides were selected using built-in databases from ExactFinder software. These selection could be exported directly into the mass inclusion list used by the Exactive lus instrument method to trigger the dd-aif scans. No further optimization of the LC-M system was needed.

4 Results and Discussion Quantitative Analysis The six calibration standards, with spike levels ranging from 5 to 150 µg/kg, were analyzed to establish calibration curves for each of the target pesticides. The majority of pesticides eluted at between 1.3 and 3.0 minutes, so a number of target components and matrix components coeluted (Figure 6). However, the extracted ion chromatograms of most target components were free from additional peaks, demonstrating that the 5 ppm extraction window combined with the resolving power of the mass spectrometer provided very high selectivity. Linear calibration curves were achieved for nearly all target pesticides (example shown in Figure 7), confirming that the compounds could be clearly distinguished from the matrix. Figure 6. Extracted chromatograms demonstrate coelution of target and matrix compounds (only 20 traces compound shown) Figure 7. Example of quantitative results from one target compound (tetraconazole)

Qualitative Analysis Qualitative analysis was carried out as a combination of targeted analysis and general unknown screening. In a first step, targeted analysis was carried out. In a second step, all peaks not identified in the targeted search were automatically forwarded for general unknown screening. 5 The same list of analytes used for quantitative analysis (Table 1) was applied for the targeted search. Retention time, isotopic pattern match, fragment search, and library search were used as confirmation criteria for targeted search. The fragment information for the analytes of interest and the fragmentation spectra for the library search were taken from databases included with the ExactFinder software. Even at the lower end of the concentration range, most components quantified could be easily confirmed on all four stages of confirmation (see Figure 8). With its built-in reporting capabilities, the ExactFinder software version 2.0 provided a quick, easy overview of the screening results. Figure 8. Qualitative results as displayed by the ExactFinder software It quickly became clear that sufficient resolution was the key to successful full-scan quantitation and screening of complex samples like the ones analyzed in this work. As shown in Figure 9, most analyte signals were surrounded by numerous matrix signals. nly sufficient resolving power ensured proper separation of analyte and matrix signals. This applies to the monoisotopic signals used for analysis as well as for the isotopic signals used for confirmation. The peaks of interest showed a resolution of close to 60,000. It was apparent that significantly lower resolving power at these masses would have led to interference and merged signals, causing significant mass shifts. The mass shifts would have led to false negatives or would have required to widening of the extraction window. Widening the extraction window would have lowered the selectivity of the analysis and resulted in false positives.

Application Note 573 Figure 9. Isotopic pattern match of pencycuron. Green boxes mark the isotope signals surrounded by matrix signals The general unknown screening carried out on the remaining peaks offers several options for automatic identification of the found peaks: database search, elemental composition determination based on isotopic pattern matching, spectral library search, and internet search. For the samples, roughly 15,000 components were detected; all of them went through the identification process. Database and spectral library searches were carried out using built-in resources. Internet search was carried out using a selection of databases listed in the Chempider online search portal. Numerous additional contaminants could be identified, especially pesticides and a selection of aflatoxins (results not shown). Conclusion HR/AM analysis is a versatile method for residue analysis. It offers full quantitation capabilities in combination with unrestricted target and unknown screening options. Ultra-high resolution delivered by the rbitrap mass analyzer in the Exactive lus mass spectrometer provides reliability and selectivity comparable to established M/M techniques. The Exactive lus mass spectrometer is compatible with UHLC without compromising resolution or mass accuracy. References 1. Kaufmann, A., Butcher,., Maden, K., Walker,. & Widmer, M. Comprehensive comparison of liquid chromatography selectivity as provided by two types of liquid chromatography detectors (high resolution mass spectrometry and tandem mass spectrometry): Where is the crossover point?. Anal. Chim. Acta 2010 Jul 12; 673(1), 60-72. 2. cheibner,.; Damoc, E.; Denisov, E.; Hauschild, J.; Lange,.; Czemper, F.; Kholomeev, A.; Makarov, A.; Wieghaus, A.; Bromirski, M. Improved analysis of biopharmaceutical samples using an M-only rbitrap mass spectrometer. resented at the 60th AM Conference on Mass pectrometry and Allied Topics, Vancouver, Canada, May 20 24, 2012, Th28 #623. Acknowledgements We would like to thank Dr. Thorsten Bernsmann from CVUA-MEL, Muenster, Germany, for preparation of the samples used in this study. www.thermoscientific.com 2012 Thermo Fisher cientific Inc. All rights reserved. I is a trademark of the International tandards rganization. Chempider is a registered trademark of ChemZoo, Inc. All other trademarks are the property of Thermo Fisher cientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher cientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. pecifications, terms and pricing are subject to change. Not all products are available in all countries. lease consult your local sales representative for details. Thermo Fisher cientific, an Jose, CA UA is I Certified. Africa-ther +27 11 570 1840 Australia +61 3 9757 4300 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-ther +43 1 333 50 34 0 Finland/Norway/weden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CI +43 1 333 50 34 0 outh Africa +27 11 570 1840 pain +34 914 845 965 witzerland +41 61 716 77 00 UK +44 1442 233555 UA +1 800 532 4752 AN63644_E 10/12