OXEA - Online Elemental Analyzer

Similar documents
Multi-element process analyzer

TECHNICAL DESCRIPTION SPECTRAFLOW ONLINE ANALYZER for CROSSBELT APPLICATION

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

TECHNICAL DESCRIPTION SPECTRAFLOW ON LINE ANALYZER for BELT CONVEYOR APPLICATION

NSD GRADEL FUSION NEUTRON GENERATORS

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer

Downhole XRF-Logging:

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

Standardless Analysis by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited

X-ray fluorescence analysis - calibration of the X-ray energy detector

Nuclear Lifetimes. = (Eq. 1) (Eq. 2)

S8 TIGER Series 2 for ASTM D 6443

An Optimized Gamma-ray Densitometry Tool for Oil Products Determination

Radioactivity. Lecture 6 Detectors and Instrumentation

Complete and balance these equations to show the reactions during electrolysis. Na Na (2)

EM SPECTRUM, WAVELENGTH, FREQUENCY, AND ENERGY WORKSHEET

UV3000. Accurate, precise, and portable ambient gas point analyzer

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR

XUV 773: X-Ray Fluorescence Analysis of Gemstones

2100TR Liquid Scintillation Counter

Honors Chemistry: Chapter 4- Problem Set (with some 6)

DAY LABORATORY EXERCISE: SPECTROSCOPY

On-line XRF analysis of multiple elements in liquid process streams C-QUAND

Chem 1A Chapter 5 and 21 Practice Test Grosser ( )

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

PYCKO SCIENTIFIC LIMITED

GAMMA RAY SPECTROSCOPY

X-ray spectroscopy: Experimental studies of Moseley s law (K-line x-ray fluorescence) and x-ray material s composition determination

Agilent G2350A Atomic Emission Detector (AED)

CHAPTER 4 NUCLEONIC GAUGES. B.D. Sowerby

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219

Sample Copyright. Academic Group ATOMIC STRUCTURE 1. Topics covered in this chapter:

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?

Elemental analysis by X-ray f luorescence. High performance EDXRF elemental analyzer

He-3 Neutron Detectors

Q1. The chart shows the processes involved in the manufacture of nitric acid from ammonia.

Performance of high pressure Xe/TMA in GEMs for neutron and X-ray detection

The Fundamental Ideas in Chemistry

TIIVISTELMÄRAPORTTI (SUMMARY REPORT)

Biophysics Collaborative Access Team Basic Techniques for EXAFS Revision date 6/25/94 G.Bunker. Optimizing X-ray Filters

Budget Uncertainty and Minimum Detectable Concentrations for the INAA Laboratory of the ETRR-2

INSTRUMENTAL NEUTRON ACTIVATION ANALYSIS (INAA)

Information on the test material EDS-TM002 and the BAM software package EDX Spectrometer Test for determination of the spectrometer performance

THE COMPTON EFFECT Last Revised: January 5, 2007

C2.6 Quantitative Chemistry Foundation

POWER + FLEXIBILITY IN ELEMENTAL ANALYSIS

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Channa Asela

Introduction to Blackbody Sources

Duncan. Electrons, Energy, & the Electromagnetic Spectrum Notes Simplified, 2-D Bohr Model: Figure 1. Figure 2. Figure 3

In 1807 Davy did an electrolysis experiment to produce potassium. Davy first tried to electrolyse a solid potassium salt to produce potassium

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

IdentiFINDER Digital Hand Held Spectrometer & Dose Rate Meter for Portable Applications

X-Ray Fluorescence and Natural History

novaa 800 D Atomic Absorption Spectrometer

Efficiency and Attenuation in CdTe Detectors

Partial Energy Level Diagrams

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

INORGANIC ELEMENT FUNCTIONAL GROUP DATABASE ON PULVERIZED COAL SURFACE BASED ON XPS METHOD

Overview of X-Ray Fluorescence Analysis

Atomization. In Flame Emission

ICSE Board Class IX Chemistry Paper 5 Solution

X-ray Fluorescence Imaging Following Synchrotron Beam Excitation

Absorption of X-rays

X-ray Absorption Spectroscopy

Gamma and X-Ray Detection

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

print first name print last name print student id grade

Choose words from the list to complete the sentences below. In an atom, the particles with a negative charge are called...

Nuclear techniques in mineral exploration, extraction, and processing

K 40 activity and Detector Efficiency

Procedure for determining radionuclides in foodstuffs at elevated levels of contamination by gamma spectrometry

POSITION SENSITIVE DETECTION OF CONCEALED SUBSTANCES EMPLOYING PULSED SLOW NEUTRONS

CHEMISTRY. SCIENCE Paper 2. (Two hours) You will not be allowed to write during the first 15 minutes.

Full file at

An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity

A.P. Chemistry Practice Test - Ch. 7, Atomic Structure and Periodicity

Atomic and nuclear physics

CHEMISTRY. SCIENCE Paper 2

EDS User School. Principles of Electron Beam Microanalysis

Unit 3. Atoms and molecules

ß + RADIATION. Choose your application of. chromatography controlled. chemical purity TLC PET GABI. PET GABI 2x2 Nal (Tl) 0 1V output.

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

CHEMICAL REACTIONS & EQUATIONS

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

AS 101: Day Lab #2 Summer Spectroscopy

Secondary Ion Mass Spectroscopy (SIMS)

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

INTERNATIONAL JOURNAL OF SCIENTIFIC & TECHNOLOGY RESEARCH VOLUME 3, ISSUE 2, FEBRUARY 2014 ISSN

C2.6 Quantitative Chemistry Foundation

Atomic and nuclear physics

Mercury Stack Monitor SM-4

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

Lab NUC. Determination of Half-Life with a Geiger-Müller Counter

Transcription:

02 25 08

OXEA - Online Elemental Analyzer OXEA (Online X-ray Elemental Analyzer) is based on the X-ray fluorescence technology (XRF) which is well known in the laboratory field. With the aid of a patented procedure, this method was made online capable. The instrument is available in different versions and is used successfully on all kinds of materials, e.g. coal, ore and cement, and also in the food industry. The OXEA instruments include a very powerful software package to detect and exactly analyze X-ray fluorescence spectra. The measured concentrations are depicted graphically, saved to a database and are available to the user on request as data telegram. Optionally, the XRF analyzer may be combined with the microwave moisture meter PMD 2450. The data of both instruments are evaluated by the software. As a result of the demands on the measurement geometry, the measurement is preferably carried out on a representative bypass, where the partial flow taken is supplied to the measurement device on a small conveyor belt. An installation above the main conveyor belt using a special sledge is also possible. Figure 1: OXEA 3000 on a sledge 2

The Principle of Measurement Irradiated matter is ionized by X-ray radiation. If sufficient excitation energy is available, the inner electrons will be extracted from the atomic compound. The created gaps in the inner electron shell are filled by electrons from the outer shells. The energy being released in the course of this process will be emitted as photons, i.e. as X-ray radiation. The energy of this radiation is characteristic for the element which the respective photon has created. The detectors used convert the X-ray radiation into electrical pulses, with the pulse height being proportional to the energy of the photons. Energy-dispersive X-ray spectroscopy means that the pulse height of each photon is now determined and assigned as an event to one channel each. This device is called multi-channel analyzer. If one reads out after an adequate time of measurement the data stored in the multi-channel analyzer and plots the number of events versus the channel number, then one gets the spectrum. This spectrum can be divided into background and characteristic lines. The intensity of the line is a measure of the concentration of the respective element. When determining the concentration, however, one has to take into account that characteristic photons may be absorbed by other elements on the way from the emitting element to the detector. The line of the analyzed element will thus be reduced and the lines of the low energy element will be depicted inflated. This reciprocal action is called matrix effect. The matrix effect can be compensated for by means of mathematical methods. The respective correction factors can be determined for a specific product during calibration of the instrument. If one knows the measurement geometry and the required data of the X-ray tube and the detector exactly, one may also calculate these correction factors on the basis of so-called fundamental parameters which can be taken from tables. OXEA utilizes the energy-dispersive method of X-ray fluorescence using high resolution detectors and thus allows product-specific matrix compensation as well as the use of fundamental parameters. 3

Comparison with other Measurement Methods The measurement accuracy of the OXEA instrument can be compared with those of PGNAAinstruments which are much more costly and complex and which require high energy neutron radiation for excitation. The table below shows the variables or elements that are relevant for coal. As you can see, only a few elements at the beginning of the periodic system cannot be measured with the XRF system OXEA. All other elements are detected with similar accuracy by both methods. To determine the calorific value, an online water content determination is required for both methods. Measurement PGNAA XRF Standard Analysis Moisture Ash Sulphur Calorific value (Btu/lb.) lb. SO 2 /MBtu Elemental Analysis Hydrogen Carbon Nitrogen Sodium Aluminium Silicium Sulphur Chlorine Oversensitive! Potassium Calcium Titan Iron Instruments working according to the dual-energy method can only determine the ash content. This principle of measurement is based on the absorption of Am- and Cs-rays by ash-producing chemical elements. Since a general absorption is being measured, the measurement is significantly dependent on the variations in ash composition. These variations can only be compensated for partially, even if a third energy source is used. In any case, a XRF measurement device is to be preferred for determination of the ash content, because it offers a lot more opportunities and higher measurement accuracies. 4

Application Examples 1 Coal 1.1 Sulphur Content The end of the pipe method, i.e. the detection of the SO 2 -content of flue gases, does not allow any optimum utilization of the flue gas desulphurization, since the sulphur content of the fuel which is already stored in the bunker cannot be influenced any more and provisions to reduce SO 2 can be taken only with a significant delay. Through continuous monitoring of the coal mixture with OXEA in front of the bunkers, these provisions can be carried out in due time and variations of the sulphur content in flue gas can be corrected much better. The OXEA measuring system can be installed behind the main sampling device which takes standard samples for quality assurance. These may also be used for calibration of the online instrument. Further harmful substances of coal, such as chlorine or arsenic can be determined, and thus controlled, at the same time. Below you see a sulphur content measurement in coal. Sulphur in coal - online measurement (RMSE = 0.14 wt.-%, correlation = 0.97) XRF Value, wt.-% 2,5 2,0 1,5 1,0 0,5 0,0 0,4 0,9 1,4 1,9 2,4 Lab value, wt.-% wt.-% 2,5 2,0 1,5 1,0 0,5 0,0 Sulphur in coal - online measurement (RMSE = 0.14 wt.-%, correlation = 0.97) Lab 2 20 25 31 36 45 50 55 60 68 76 Sample RFA 5

1.2 Ash Content and Elemental Compensation of Coal Figure 2: User interface of the OXEA Analyse software OXEA covers the entire ash composition. Typically, the processed coal is checked online. An empirical value for the attainable accuracy is 0.3 0.7%. In some cases it may be a better approach to check the tailings in order to get an impression of the losses. Figure 2 shows a typical tailings spectrum. The ash content is determined using special algorithms. Optionally, one can also monitor elements online that are of particular interest, e.g. chlorine. Figure 3 shows the calibration of the tailings measurement carried out with OXEA Labor. The calibration curve is in addition shown as trend line. 6

Figure 3: OXEA Labor software with ash calibration of mountains 2. Recyclable Material and Harmful Substance Content in Mineral Raw Materials OXEA can be used to determine chemical elements with an atomic number >10. Thus, OXEA is ideally suited to check the recyclable material content of mineral raw materials in processing plants. Importent elements which helps to assess quality may also be determined. It is important to have information on harmful substances such as silicium in magnesite, mercury in iron ore or lead in lime. 7

2.1 Potassium Content The common method for determination of the potassium content is the measurement of the natural Gamma radiation of the isotope K-40. However, with high concentrations this method is inaccurate. In particular for concentrate we recommend the use of OXEA. Below you see OXEA measurements in the range of 8-60% K2O. Potassium calibration in KCl salt RMSE = 0.87 wt.-%, correlation = 0.99 70 60 Meas. Value (XRF) 50 40 30 20 10 0 0 10 20 30 40 50 60 70 Lab value Potassium calibration in KCl salt RMSE = 0.87 wt.-%, correlation = 0.99 Lab XRF wt.-% 65 55 45 35 25 15 5 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Sample 8

3 Ores Figure 4: OXEA 3000 on bypass measurement belt OXEA is ideally suited to determine the recyclable material content or the concentration of accompanying minerals of ores. Below we will present the silicium content in magnesite ore as an example. Measurements were carried out online on the measurement belt, as shown in Figure 4. The only material preparation that was carried out was drying and reducing the partial flow to < 3 mm. The calibration curve is depicted as x-y-chart and as trend line. The accuracies obtained with the low energy element Si are excellent. A table shows, in addition, the accuracies obtained for this application for the major accompanying element. 9

Silicium determination in magnesite ore (calibration) (RMSE = 0.037 wt.-%, correlation = 0.984) Meas. Value, XRF 1,2 1,0 0,8 0,6 0,4 0,2 0,0 0,0 0,2 0,4 0,6 0,8 1,0 1,2 Lab value Silicium determination in magnesite ore (trend line) (RMSE = 0.1 wt.-%, correlation = 0.968) Lab RFA wt.-% 2,0 1,8 1,6 1,4 1,2 1,0 0,8 0,6 0,4 0,2 0,0 1 6 11 16 21 26 31 36 41 46 51 Sample Std. dev. Al 2 O 3 SiO 2 CaO TiO 2 Mn 2 O 3 Fe 2 O 3 Absolute, wt.-% 0.02 0.04 0.03 0.002 0.006 0.05 Relative, % 3.8 3.6 0.1 4.0 1.7 1.6 Table 1: Accuracies obtained with OXEA on magnesite ore 10

4 Elemental Analysis of Cement OXEA with its online XRF technology is suited for the determination of different elements in cement. This is illustrated below using an SO 3 -measurement as an example; the other elements are summarized in a table. The measurements were carried out under online-like conditions using the OXEA Atline, i.e. the measurements were carried out without sample preparation (pressing, melting). Only the measurement surface was smoothed using a special mechanical facility. SO 3 Measurement in cement (calibration) RMSE = 0.16 wt.-%, correlation = 98.7% Meas. value (XRF), wt.-% 6 5,5 5 4,5 4 3,5 3 2,5 2 2 2,5 3 3,5 4 4,5 5 5,5 Lab value, wt.-% SO 3 Measurement in cement (trend line) Lab XRF 6 5 wt.-% 4 3 2 1 4 7 10 13 16 19 22 25 28 31 34 37 40 43 46 49 52 Sample Std. dev. Al 2 O 3 SiO 2 SO 3 Cl K 2 O CaO Fe 2 O 3 Absolute, wt.-% 0.14 0.35 0.16 0.04 0.05 0.38 0.08 Relative, % 2.3 1.6 2.9 4.6 3.2 0.6 1.8 Table 2: Accuracies obtained on cement with OXEA Atline without sample preparation 11

5 Trace Element Analysis Even smallest amounts of various elements, for example, arsenic or mercury, can be determined online in a variety of materials. Figure 5: Section of XRF coal spectrum with arsenic line (20 ppm) Another example is the zinc measurement in the range between 0.04 0.14 %. Zn measurement in catalyst (RMSE = 0.01 wt.-%, correlation = 0.97) 0,14 Meas. Value (XRF) 0,12 0,10 0,08 0,06 0,04 0,02 0,02 0,04 0,06 0,08 0,10 0,12 0,14 Lab value 12

6 Layer Thickness Determination Layer thicknesses which contain a specific chemical element in a constant concentration can be determined very reliably with OXEA. For example, the charts diagrams below show the measurement of a silicium layer on paper. In combination with an optional traversing frame, the sensor can cover the layer thickness over the entire width of the paper track. Calibration of silicium coating on a sheet of paper (RMSE = wt.- 0.16 %, correlation 0.96) 2,4 2,2 Meas. value (XRF) 2,0 1,8 1,6 1,4 1,2 1,0 0,8 0,7 1,2 1,7 2,2 2,7 Lab value Calibration of silicium coating on a sheet of paper (RMSE = wt.- 0.16 %, correlation 0.96) Lab XRF 3,0 2,5 2,0 wt.- % 1,5 1,0 0,5 0,0 1 2 3 4 5 6 7 8 9 11 12 13 14 15 Sample 13

7 Application in the Food Industry OXEA can be employed with success for quality control in the food industry. The diagram below shows that the Zink content can be determined very accurately in the ppm range. Other elements such as K, Ca, Fe can also be determined online without any problem. Zn measurement in food (RMSE = 3 ppm, correlation = 0.997) 120,0 Meas. value (XRF), [ppm] 100,0 80,0 60,0 40,0 20,0 0,0 0,0 20,0 40,0 60,0 80,0 100,0 120,0 Lab value, [ppm] Zn measurement in food (RMSE = 3 ppm, correlation = 0.997) 120,0 100,0 80,0 Lab RFA [ppm] 60,0 40,0 20,0 0,0 1 5 6 7 8 10 12 13 15 16 18 Sample 14

8 OXEA Atline Sometimes a measurement of individual samples which have been taken at different operating points is preferred to a fixed online measurement. For this application we have developed OXEA-Atline which works according to the same principles as the online device. In this case, the sensor is provided with a substructure which contains a rotary plate on which the sample can be moved. The maximum particle size should be < 10 mm, as for the online measurement. The sample quantity may be in the range between 100 to 1000 cm 3. With the aid of a special evaluation routine the software can partially compensate for changes in particle size distribution, which normally have a very strong influence on the intensity of the spectrum. The measuring instrument may be connected to the in-house network. Thus, the quality data of a quality management system become accessible. The quality data are logged in the program database. The built-in software modules allow the user to create daily or shift-related quality reports. Figure 6: OXEA Atline 15

Technical Data for OXEA Online Devices Sensor Unit / Atline Dimensions (L x W x H) Weight Protection type High voltage supply Dimensions Weight Protection type 15.7 x 13.4 x 8.7 / 16.1 x 19.7 x 9.0 Inches 400 x 340 x 220 / 408 x 500 x 228 mm 68,4 lbs / 99,2 lbs 31 kg / 45 kg 29.9 x 29.9 x 8.3 Inches 760 x 760 x 210 mm 110,3 lbs 50kg IP 66 IP 66 Avg. Particle Size < 1 < 25 mm Max. particle size < 2 < 50 mm Operating temperature -22 F to 122 F -30 C to 50 C Power supply 100-240V VAC nominal, 16 A X-ray tube Model: IT XR S1-25 25 kv 1,6 ma X-ray detector Peltier-cooled silicium drift detector Type: IT XR D2-5/155-8.0Be 155 ev FWHM resolution @5.9 kev, 55 Fe Moisture meter Precision microwave detector PMD 2450 Transmission measurement (attenuation and phase) Computer Intel Celeron processor, 2000 MHz 160 GB hard drive or better or better Monitor 17 flat screen or comparable device Software OXEA evaluation software Microsoft Windows XP Pro Microsoft Office (Option) LapLink Modem Network interface card 56 Kbps standard or network interface card Requires a free phone line or internet connection for service Analysis period Short approximate value 1-2 minutes Licensing Single licence Radiation outside the Below the detection limit analyzer Accuracies under regular conditions Typical RMSD for coal: 5% of measuring range but not < ± 0.4-0.7% ash The achievable accuracy will be determined by measuring your samples. ± 0.3-0.6% water content ± 0.05-0.15% sulphur ± 100-150 calorific value (Btu/lb.) Performance test In accordance with ISO/DIS 15239 Service (optional) Service contract Check / Calibration Interfaces PCS system DeviceNet, Modbus, DUST 3964R, Profibus Optional 16