Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Similar documents
HPLC Column Troubleshooting: Is It Really The Column? LC Columns and Consumables. Robyn C. Woolley Online Application Engineer July 24, 2007.

Advanced Topic: Tips and Tricks of HPLC Separations and Troubleshooting. Rita Steed LC Advanced Topics Seminar Series August 25, 2010.

Tips and Tricks of HPLC System Troubleshooting

Tips and Tricks of Faster LC Analysis Without Capital Investment It may be easier than you think!

William E. Barber, Ph.D. Applications Chemist October

HPLC Column Troubleshooting

Choosing Columns and Conditions for the Best Peak Shape

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

ProntoSIL C18-EPS Reversed-Phase HPLC Columns

Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput

Hypersil BDS Columns TG 01-05


penta-hilic UHPLC COLUMNS

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

LC and LC/MS Column Selection Flow Chart

COMPARISON GUIDE. to C18 Reversed Phase HPLC Columns. Comparison Data on 60 Commonly Used C18 Phases. Stationary Phase Specifications

C18 Column. Care & Use Sheet

ACQUITY UPLC HSS Columns

Validated HPLC Methods

HPLC Method Development with Eclipse Plus: Standard Practices and New Columns. Agilent Technologies

New To HPLC Avoiding Beginner Pitfalls. Rita Steed LC Columns Application Engineer April 24, 2013

penta-hilic UHPLC COLUMNS

[ CARE AND USE MANUAL ]

Alltech Alltima HP Introduction

for Acclaim Mixed-Mode HILIC-1 Column

Reversed Phase Chromatography: Mobile phase Considerations

FAQ's. 1 - Which column would be the most appropriate for my application?

State-of-the-art C18 HPLC Columns

High Performance Liquid Chromatography

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Acclaim Mixed-Mode WCX-1

High Resolution Fast LC

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Basic principles of HPLC

LC Technical Information

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

P R O D U C T B U L L E T I N. Fused-Core particle technology for hyper-fast and super-rugged HPLC columns

HPLC Method Development: Standard Practices and New Columns

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Comparison of different aqueous mobile phase HPLC techniques

Reversed Phase Solvents

HPLC Separation Fundamentals. Ed Kim Application Engineer Agilent Technologies, Inc January 14, 2009

Bulletin 826B. How to identify, isolate, and correct the most common HPLC problems

Do you need trouble-free speed? The NEW Chromolith HR

Packings for HPLC. Packings for HPLC

Fast Separation of Vastly Different Compounds by Isocratic HPLC

InertSustainSwift C8

Inertsil ODS-EP Technical Information

A C18 Silica Column With Exceptional Temperature and ph Stability

Too Polar for Reversed Phase What Do You Do?

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Do you need trouble-free speed? The NEW Chromolith HR

Comparison Guide to C18 Reversed Phase HPLC Columns

A Pre-Clinical Pharmacokinetic-Pharmacodynamic Modelling and Biodistribution Studies of Donepezil Hydrochloride by Validated HPLC Method

Minimizing Solvent Impact on Purification of Nitrogencontaining

Solid Phase Extraction Method Development Tips and Tricks

InertSustainSwift C8

New Product Update. HPLC Columns

[ Care and Use Manual ]

ProPac WCX-10 Columns

Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

Shodex TM ODP2 HP series columns

SPE Introduction. general chromatography. solid phase extraction. How to Choose an SPE Product. Solid-Phase Processing Methods

Bringing uhplc Performance to the Separation of Peptides

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

HPLC Separation Robustness and Ruggedness Assessing the Effects of Experimental Variables During Method Development

zorbax Packed by Phenomenex Material Characteristics High efficiency Guaranteed quality Phenomenex columns packed with Zorbax

Place a collection tube beneath cartridge. 3 ml methylene chloride/isopropyl alcohol/ hydrochloric acid (60/40/1) Collect the eluant

Optical Isomer Separation Columns and Packing Materials

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

ACQUITY UPLC Peptide BEH C 18, 130Å and 300Å Columns

[ Care and Use Manual ]

(HILIC) Bill Champion Agilent Technologies, Inc opt 3/opt3/opt 2 HILIC - Agilent Restricted

HPLC Column Classification

for Acclaim Mixed-Mode WCX-1

COLUMN GUIDE - Usage and maintenance of HPLC columns

8. Methods in Developing Mobile Phase Condition for C18 Column

HPLC Method Development: Standard Practices and New Columns. Agilent Technologies February 2007

Kromasil. for your analytical HPLC. Kromasil. The way to peak performance in liquid chromatography

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

High Performance Liquid Chromatography

Technical Guide Thermo Scientific Syncronis HPLC Columns

[ CARE AND USE MANUAL ] XSELECT HSS XP 2.5 µm COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. GETTING STARTED III. COLUMN USE

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

PYRIPROXYFEN TECHNICAL

ACE Ultra Inert Base Deactivated HPLC Columns

HPLC Praktikum Skript

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Characterization of Carbon SPE for the Extraction of Polar Analytes

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Dilution(*) Chromatography

[ CARE AND USE MANUAL ]

Open Column Chromatography, GC, TLC, and HPLC

Liquid Chromatography

Transcription:

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 0, 000 HPLC Column Troubleshooting What Every HPLC User Should Know :00 a.m. EST Telephone Number: 86-650-06 Chair Person: Tim Spaeder

HPLC Column Troubleshooting What Every HPLC User Should Know Slide 4

HPLC Components Pump Injector/Autosampler Column Detector Data System/Integrator All of these components can have problems and require troubleshooting. Slide 5

Categories of Column Problems A. Pressure B. Peak shape C. Retention Slide 6

Pressure Issues Column Observations High pressure Potential Problems Plugged frit Column contamination Plugged packing Slide 7

Determining the Cause and Correcting High Back Pressure Check pressure with/without column - many pressure problems are due to blockages in the system or at the guard If Column pressure is high: Wash column Eliminate column contamination and plugged packing high molecular weight/adsorbed compounds precipitate from sample or buffer Back flush column Clear plugged frit Change frit Clear plugged frit Slide 8

Column Cleaning Flush with stronger solvents than your mobile phase. Reversed-Phase Solvent Choices in Order of Increasing Strength Use at least 5 ml of each solvent for analytical columns Mobile phase without buffer salts 00% Methanol 00% Acetonitrile 75% Acetonitrile:5% Isopropanol 00% Isopropanol 00% Methylene Chloride* 00% Hexane* * When using either Hexane or Methylene Chloride the column must be flushed with Isopropanol before returning to your reversed-phase mobile phase. Slide 9

Column Cleaning Normal Phase Solvent Choices in Order of Increasing Strength Use at least 50 ml of each solvent 50% Methanol : 50% Chloroform 00% Ethyl Acetate Slide 0

How to Change a Frit Column Inlet Frit Column Body Compression Ferrule Wear gloves Do not allow bed to dry Do not touch the column - body heat will extrude packing Do not overtighten Female End Fitting Male End Fitting Slide

Preventing Back Pressure Problems Use column protection - Guard columns - In-line filters Sample Preparation Appropriate column flushing Filter buffered mobile phases Slide

Preventing Back Pressure Problems: In-Line Devices Mobile Phase From Pump Pre-Column Injector Guard Filter Column Analytical Column Filter and Guard Column Act on Sample Pre-Column Acts on Mobile Phase To Detector Slide 3

Preventing Back Pressure Problems: Sample Preparation Solvent/Chemical Environment Particulate/Aggregate Removal Filter samples Centrifugation Solid Phase Extraction (S.P.E.) Cartridges or Plates Disks or Membranes Slide 4

Peak Shape Issues Split peaks Peak tailing Broad peaks Poor efficiency Many peak shape issues are also combinations - i.e. broad and tailing or tailing with increased retention Slide 6

Split Peaks Can be caused by: Column contamination Partially plugged frit Column void 0.0 Injection solvent effects.5 5.0 7.5 0.0 Slide 7

Split Peaks Column Contamination Column: StableBond SB-C8, 4.6 x 50 mm, 5 µm Mobile Phase: 60% 5 mm Na HPO 4, ph 3.0 : 40% MeOH Flow Rate:.0 ml/min Temperature: 35 C Detection: UV 54 nm Sample: Filtered OTC Cold Medication:. Pseudoephedrine. APAP 3. Unknown 4. Chlorpheniramine Injection Injection 30 Injection After Column Wash with 00% ACN 4 4 3 4 3 3 0 5 0 5 0 5 0 5 0 5 0 5 Slide 8

Split Peaks Injection Solvent Effects Column: StableBond SB-C8, 4.6 x 50 mm, 5 µm Mobile Phase: 8% H O : 8% ACN Injection Volume: 30 µl Sample:. Caffeine. Salicylamide A. Injection Solvent 00% Acetonitrile B. Injection Solvent Mobile Phase 0 0 0 0 Slide 9

Determining the Cause of Split Peaks. Complex sample matrix or many samples analyzed - likely column contamination or partially plugged frit. Mobile phase ph 7 - likely column void due to silica dissolution (unless specialty column used) 3. Injection solvent stronger than mobile phase - likely split and broad peaks, dependent on sample volume Slide 0

Peak Tailing, Broadening and Loss of Efficiency Can be caused by: Column secondary interactions Column void Column contamination Column aging Column loading Extra-column effects Slide

Peak Tailing Column Secondary Interactions Column: Alkyl-C8, 4.6 x 50 mm, 5µm Mobile Phase: 85% 5 mm Na HPO 4 ph 7.0 : 5% ACN Flow Rate:.0 ml/min Temperature: 35 C Sample:. Phenylpropanolamine. Ephedrine 3. Amphetamine 4. Methamphetamine 5. Phenteramine No TEA USP TF (5%)..9..9 3..63 4..35 5..57 3 4 5 3 4 5 0 mm TEA USP TF (5%)..9..8 3..0 4..6 5..4 0.0.5 5.0 TIme (min) 0.0.5 5.0 Peak tailing eliminated with mobile phase modifier (TEA) at ph 7 Slide

Peak Tailing Column Secondary Interactions Column: Alkyl-C8, 4.6 x 50 mm, 5µm Mobile Phase: 85% 5 mm Na HPO 4 : 5% ACN Flow Rate:.0 ml/min Temperature: 35 C Sample:. Phenylpropanolamine. Ephedrine 3. Amphetamine 4. Methamphetamine 5. Phenteramine ph 3.0 USP TF (5%) 3 ph 7.0 USP TF (5%) 4..33 4 5 3 4..35 4 5 0.0.5 5.0 0.0.5 5.0 Reducing the mobile phase ph reduces interactions with silanols that cause peak tailing. Slide 3

Peak Tailing Column Contamination Column: StableBond SB-C8, 4.6 x 50 mm, 5µm Mobile Phase: 0% H O : 80% MeOH Flow Rate:.0 ml/min Temperature: R.T. Detection: UV 54 nm Sample:. Uracil. Phenol 3. 4-Chloronitrobenzene 4. Toluene Plates TF. 769.08. 043.64 3. 377.69 4 3355.3 3 Plates TF. 7906.43. 443. 3. 7999.9 4 7098.5 3 Plates TF. 7448.06. 37. 3. 5366. 4 9067.7 3 4 4 4 0.0.5 5.0 0.0.5 5.0 0.0.5 5.0 QC test forward direction QC test reverse direction QC test after cleaning 00% IPA, 35 C Slide 4

Peak Tailing/Broadening Sample Load Effects Columns: 4.6 x 50 mm, 5µm Mobile Phase: 40% 5 mm Na HPO 4 ph 7.0 : 60% ACN Flow Rate:.5 ml/min Temperature: 40 C Sample:. Desipramine. Nortriptyline 3. Doxepin 4. Imipramine 5. Amitriptyline 6. Trimipramine Tailing Eclipse XDB-C8 USP TF (5%) A. High Load x0 C. Broadening Competitive C8 Plates A B C D..60.70..00.90 3..56.56 4..3.70 5..5.86 6..5.5 0 5 0 0 5 0. 850 594. 85 784 3. 776 63 4. 539 8359 5. 735 00 6. 589 075 B. Low Load D. 0 5 0 0 5 0 Slide 5

Peak Broadening, Splitting Column Void Mobile Phase: 50%ACN: 50% Water : 0.% TEA (~ ph ) Initial After 30 injections Multiple peak shape changes can be caused by the same column problem. In this case a void resulted from silica dissolved at high ph. Slide 6

Broad Peaks Unknown Phantom Peaks Column: Extend-C8, 4.6 x 50 mm, 5 µm Mobile Phase: 40% 0 mm TEA, ph : 60% MeOH Flow Rate:.0 ml/min Temperature: R.T. Detection: UV 54 Sample:. Maleate. Pseudoephedrine 3. Chlorpheniramine Sample : Chlorpheniramine maleate Sample : Chlorpheniramine maleate and Pseudoephedrine Plates. 59. 9879 3. 779 3 Phantom peak from first injection 0 5 0 0 5 0 5 The extremely low plates are an indication of an extremely late eluting peak from the preceding run. Slide 7

Peak Tailing Injector Seal Failure Column: Bonus-RP, 4.6 x 75 mm, 3.5 µm Mobile Phase: 30% H O : 70% MeOH Flow Rate:.0 ml/min Temperature: R.T. Detection: UV 54 nm Sample:. Uracil. Phenol 3. N,N-Dimethylaniline Plates USP TF (5%). 35.7. 349.48 3. 543.5 3 3 Plates USP TF (5%). 3670.45. 0457.09 3. 0085.00 0.0 0.5.0.5 Before 0.0 0.5.0.5.0 After replacing rotor seal and isolation seal Overdue instrument maintenance can cause peak shape problems. Slide 8

Peak Tailing Extra-Column Volume Column: StableBond SB-C8, 4.6 x 30 mm, 3.5 µm Mobile Phase: 85% H O with 0.% TFA : 5% ACN Flow Rate:.0 ml/min Temperature: 35 C Sample:. Phenylalanine. 5-benzyl-3,6-dioxo--piperazine acetic acid 3. Asp-phe 4. Aspartame 0 µl extra-column volume 50 µl extra-column volume (tubing) 3 4 3 4 0.0 0.5.0.5.0 0.0 0.5.0.5.0 Slide 9

Determining the Cause of Peak Tailing Evaluate mobile phase effects - alter mobile phase ph and additives to eliminate secondary interactions Evaluate column choice - try column with high purity silica or different bonding technology Reduce sample load Eliminate extra-column effects Flush column and check for aging/void Slide 30

Retention Issues Retention time changes (t r ) Capacity factor (retention) changes (k ) Selectivity changes (α) Slide 3

Changes in Retention Same Column, Over Time May be caused by: Column aging Column contamination Insufficient equilibration Poor column/mobile phase combination Change in mobile phase Change in flow rate Other instrument issues Slide 33

Mobile Phase Change Causes Change in Retention 60% MeOH: 40% 0.%TFA Fresh TFA Added to Mobile Phase 0 30 0 0 30 Volatile TFA evaporated/degassed from mobile phase. Replacing it solved problem. Slide 34

Column Aging/Equilibration Causes Retention/Selectivity Changes Column - Initial Column - Next Day Column - After Cleaning with % H 3 PO 4 0 3 5 9 5 0 3 5 9 5 0 3 5 9 5 The primary analyte was sensitive to mobile phase aging of the column. The peak shape was a secondary issue resolved by flushing the column. Retention and peak shape were as expected after cleaning. Slide 35

Determining the Cause of Retention Changes Same Column. Determine k, α, and t r for suspect peaks. Wash column 3. Test new column - note lot number 4. Review column equilibration procedures 5. Make up fresh mobile phase and test 6. Check instrument performance Slide 36

Change in Retention/Selectivity Column-to-Column Different column histories (aging) Insufficient/inconsistent equilibration Poor column/mobile phase combination Change in mobile phase Change in flow rate Other instrument issues Slight changes in column bed volume (t r only) Slide 37

Example Change in Retention/Selectivity Column-to-Column Mobile Phase Variation Column Column Column - Fresh mobile phase 4 6 3 4 5 6 7 4 6 I have experimented with our mobile phase, opening new bottles of all mobile phase components. When I use all fresh ingredients, the problem ceases to exist, and I have narrowed the problem to either a bad bottle of TEA or phosphoric acid. Our problem has been solved. Slide 38

Determining the Cause of Retention Changes Column-to-Column. Determine k, α, and t r for suspect peaks. Test new column - note lot number 3. Determine column history of all columns 4. Review column equilibration procedures 5. Make up fresh mobile phase and test 6. Check instrument performance Slide 39

Minimize Change in Retention/Selectivity Lot-to-Lot Evaluate: All causes of column-to-column change* Method ruggedness (buffers/ionic strength) ph sensitivity (sample/column interactions) *All causes of column-to-column change should be considered first, especially when only one column from a lot has been tested. Slide 40

Lot-to-Lot Selectivity Change ph -Base ph 4.5 - Lot -Base ph 3.0 - Lot 3 4-Base 4-Base 3 0 4 6 8 0 4 6 8 0 4 6 8 0 4 6 8 -Base -Base ph 4.5 - Lot ph 3.0 - Lot 3 4-Base 4-Base 3 0 4 6 8 0 4 6 8 0 4 6 8 0 4 6 8 ph 4.5 shows selectivity change from lot-to-lot for basic compounds ph 3.0 shows no selectivity change from lot-to-lot, indicating silanol sensitivity at ph 4.5 Slide 4

Experimental Conditions for Classifying Column Selectivity Changes Bonded-Phase Test Silanol Activity Test Mobile Phase: 85% CH 3 OH 5% H O Mobile Phase: 60% CH 3 OH 40% H O Chromatographic probes. Naphthalene. Anthracene Chromatographic probes. Dimethylaniline. Toluene 0 4 6 8 0 0 4 6 8 0 α value changes of >0% suggest changes in bonded-phase or silica Slide 4

Evaluate Retention Changes Lot-to-Lot Eliminate causes of column-to-column selectivity change Re-evaluate method ruggedness - modify method Determine ph sensitivity - modify method Classify selectivity changes Contact manufacturer for assistance Slide 43

Conclusions HPLC column problems are evident as: High pressure Undesirable peak shape Changes in retention/selectivity Often these problems are not associated with the column and may be caused by instrument and experimental condition issues. Slide 44