Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

Similar documents
Elemental Analysis: NCS characterization of paper

Elemental Analysis: CHNS/O characterization of polymers and plastics

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

SKALAR. your partner in chemistry automation. Primacs SN Protein-Nitrogen Analyzer

SKALAR. your partner in chemistry automation. Primacs SNC Total Carbon and Total Nitrogen Analyzer

ECS8020 ORGANIC ELEMENTAL ANALYZER CHNS-O Analyzer

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Defining quality standards for the analysis of solid samples

Compact Orifice FlowMeter

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Determination of Carbon d13c isotopic composition from DIC samples

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Finnigan TC/EA. High Temperature Conversion Elemental Analyzer

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Speed and flexibility

δ 13 C of DIC and carbonate samples: comparison of traditional mass spectrometry methods with an infrared spectrometry method

Automated and accurate component detection using reference mass spectra

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

ISO INTERNATIONAL STANDARD. Surface chemical analysis X-ray photoelectron spectroscopy Repeatability and constancy of intensity scale

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Kjeltec 2300 Analyzer Unit

Expanded capabilities in ammonium and amine detection

INTERNATIONAL STANDARD

Fast methods for the determination of ibuprofen in drug products

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

of mass spectrometry

costech instruments ECS 4010 Elemental Combustion System CHNS-O

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

The art of elemental analysis

ISO 1871 INTERNATIONAL STANDARD. Food and feed products General guidelines for the determination of nitrogen by the Kjeldahl method

ISO INTERNATIONAL STANDARD

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

AOAC Official Method 2016.xx. Determination of Total Sulfur in Fertilizers by High Temperature Combustion

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

The key to your laboratory s success

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Letter to non European Union customers

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

UDK. Distillation Units Series. A Full Range of Solutions for Kjeldahl Distillation LABSOLUTIONS

Fast and effective. quality monitoring for breweries. Thermo Scientific Gallery Plus Beermaster

Introduction to Fourier Transform Infrared Spectroscopy

Thermo Scientific TPDRO 1100 Advanced Catalyst Characterization

Pyrolysis process for the treatment of food waste. Barbora Grycová VSB - Technical University of Ostrava Czech Republic

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Analysis of domestic sludge using the Agilent 4200 MP-AES

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Application Note No. 116 / 2013 Nitrogen determination in sodium nitrate

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

ACRYSOL TM RM-8WE Rheology Modifier

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

ISO INTERNATIONAL STANDARD. Water quality Determination of trace elements using atomic absorption spectrometry with graphite furnace

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

EA-IRMS: Fast and Precise Isotope Analysis of Liquids on a Delta V Isotope Ratio MS with a High Temperature Conversion Elemental Analyzer

TPDRO 1100 Advanced Catalyst Characterization. For Temperature Programmed Desorption, Reduction, Oxidation and Chemisorption

Introduction to Fourier Transform Infrared Spectroscopy

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Quantification of total nitrogen according to the Dumas method (Musts and Wines) (Resolution Oeno 13/2002)

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Ideal and powerful solution

Monitoring Protein PEGylation with Ion Exchange Chromatography

INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD

MetWorks Metabolite Identification Software

DMA-80L TECHNICAL SPECIFICATIONS

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

HYPHENATED TECHNOLOGY GUIDE

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Agilent G2350A Atomic Emission Detector (AED)

SpeedDigester K-436, K-439 / KjelFlex K-360 Nitrogen Determination in Nitrate-Free Fertilizer according to the micro-kjeldahl Method 074/2011

ISO INTERNATIONAL STANDARD. Plastics Determination of thermal conductivity and thermal diffusivity Part 3: Temperature wave analysis method

ACRYSOL TM RM-8W Rheology Modifier

Transcription:

APPLICATION NOTE 42255 Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases Authors Dr. Liliana Krotz, Dr. Francesco Leone, and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy Keywords Argon, Combustion, Food Quality, Infant Food, Labeling, Nitrogen, Protein Goal To demonstrate the performance of the Thermo Scientific FlashSmart Elemental Analyzer for food quality and labeling purposes, while showing compliance to international standards requirements. Introduction The nutritional composition of infant food plays a very important role in food industries for research and quality control purposes. Food market globalization requires accurate control of product characteristics to protect commercial value, to safeguard consumer health, and to maintain manufacturer reputation. The selection of infant food products available in the market is very wide, and consumers decisions depend on a set of information that includes quality valuation. New regulations regarding all processed food and most raw foods include a series of tests aimed at determining food contents and their contribution to a healthy diet for infants and children. Official regulations establish protein content and labeling requirements which enable consumers to make quality comparisons.

One of the tests used in the production process is the determination of protein content through nitrogen analysis. It is very important to have an accurate and precise technique for periodically performing this test, and an automated technique that provides fast analysis, excellent reproducibility, and avoids the use of toxic chemicals is ideal. An alternative to the classical Kjeldahl method, based on the Dumas (combustion) method, has been developed and is approved by different organizations. CO 2 and H 2 O traps, a GC column, and were finally detected by a thermal conductivity detector (Figure 2). A complete report is automatically generated by the Thermo Scientific EagerSmart Data Handling Software and displayed at the end of the analysis. Using the nitrogen data obtained and a protein factor, the software automatically calculates the protein content. The Thermo Scientific FlashSmart Elemental Analyzer (Figure 1), based on the dynamic combustion method (Dumas method), provides rapid and automated nitrogen determination without use of hazardous chemicals and offers advantages in precision over traditional methods. The FlashSmart Elemental Analyzer allows runs at both high and low nitrogen levels with no need to change configurations and without matrix effects. Figure 2. FlashSmart Elemental Analyzer nitrogen configuration. Table 1 shows the analytical conditons using helium and argon as carrier gases. Table 1. Analytical conditions Analytical Conditions Parameter Helium Carrier Gas Argon Carrier Gas Combustion Reactor Temperature 950 C 950 C Figure 1. Thermo Scientific FlashSmart Elemental Analyzer. However, due to a possible worldwide shortage and increase in the cost of helium, the FlashSmart Analyzer can work with an alternative gas, argon which is readily available. Methods The Elemental Analyzer operates according to the dynamic flash combustion (modified Dumas method) of the sample. Powdered samples were weighed in tin containers and liquid samples were adsorbed on the inert material Chromosorb, then weighed in tin containers. Both were introduced into the combustion reactor with oxygen via the Thermo Scientific MAS Plus Autosampler. After combustion, the produced gases were carried by a helium or argon flow to a second reactor filled with copper. Then they were swept through Reduction Reactor Temperature 840 C 840 C Oven Temperature 50 C 50 C Carrier Gas Flow 140 ml/min 60 ml/min Reference Gas Flow 100 ml/min 60 ml/min Total Run Time 400 sec 600 sec Sample Delay 10 sec 10 sec Oxygen Flow 300 ml/min 300 ml/min OxyTune Category GC Column @ 30 sec: aspartic acid, Cat. A: samples 1, 3, 4, 5, 6, 7, 8, 9, 10 and 11 Cat. C: urea/water solution and liquid samples 12 and 13 Category E (d = 1.4) for sample 2 0.5 m, installed outside the oven @ 30 sec: aspartic acid, @ 30 sec: all powder samples Cat. C: urea/water solution and liquid samples 12 and 13 1 m, installed in the oven Note: The EagerSmart Data Handling Software provides a new option AGO (Argon Gas Option) that allows the user to manage the flow of argon gas during the run.

Results To evaluate the performance of the system using helium and argon as carrier gases, several infant food samples (powders and liquids) with different nitrogen concentrations were chosen. Table 2 shows the sample information and Table 3 shows the standards used for calibration and the weight of each sample. K factor was used as the calibration method in all cases. Table 2. Sample information. Ref. No. Matrix Type Consuming Age 1 Milk based powder, sucrose free 1 3 years 2 Milk based powder From birth onward 3 Milk based powder 1 3 years 4 Milk based powder 1 3 years 5 Milk based powder 3 years onward 6 Milk based powder 1 8 years 7 Milk based powder 6 12 months 8 Milk based powder, sucrose free 0 6 months 9 Milk based powder 1 10 years 10 Milk based powder From birth onward 11 Milk based powder, sucrose free From birth onward 12 Milk based powder From birth onward 13 Milk based powder From birth onward (premature babies) Table 3. Standard and sample weight. Helium Carrier Gas Argon Carrier Gas Standard Sample Weight (mg) Standard Sample Weight (mg) Aspartic acid (10.52 N%) 50-100 mg Urea/water solution (0.46 N%) 160-200 mg 1 200 220 1 110 115 2 210 220 2 110 115 3 220 230 3 110 115 4 200 220 4 110 115 5 170 180 5 110 115 6 200 220 Aspartic acid (10.52 N%) 50-70 mg 6 110 115 7 190 200 7 110 115 8 200 210 8 110 115 9 200 220 9 110 115 10 200 240 10 110 115 11 200 220 11 110 115 12 180 220 Urea/water solution (0.46 N%) 12 220 270 13 180 220 260-270 mg 13 220 270

Table 4 shows the nitrogen/protein data obtained from samples listed in Table 3 by using helium and argon as carrier gases. Each sample was analyzed in triplicate. The protein factor used to calculate the protein content through the nitrogen value was 6.38. Table 5 presents a summary of the protein averages comparing data from helium gas, argon gas, and from the Kjeldahl analysis. The protein % determined by Kjeldahl analysis was performed at least 3 4 times for every sample, and includes the acceptable range for results. The data comparison demonstrates the performance of the instrument using both gases. Table 4. Nitrogen/protein data using helium and argon as carrier gas. Sample Ref. No. 1 2 3 4 5 6 7 8 9 10 11 12 13 Helium as carrier gas Argon as carrier gas N% Protein % RSD% N% Protein % RSD% 2.33 2.79 2.80 2.79 2.44 2.42 2.43 4.95 4.97 4.94 2.51 2.49 2.09 2.09 2.07 0.199 0.201 0.195 0.449 0.461 0.450 14.90 14.89 14.92 14.26 14.22 14.26 17.78 17.88 17.80 15.54 15.47 15.51 10.25 10.26 10.28 31.59 31.72 31.54 15.97 15.96 16.00 10.78 10.79 10.81 15.83 15.91 15.81 13.32 13.35 13.21 10.78 10.80 10.77 1.27 1.28 1.24 2.87 2.94 2.87 0.11 0.16 0.32 0.21 0.16 0.30 0.13 0.14 0.31 0.54 0.13 1.54 1.42 2.31 2.30 2.14 2.16 2.13 2.75 2.77 2.74 2.42 2.41 2.40 1.57 1.57 1.59 4.87 4.86 4.85 2.52 1.67 1.64 2.46 2.44 2.06 2.04 2.07 1.66 0.202 0.204 0.196 0.475 0.477 0.461 14.76 14.68 14.93 13.66 13.78 13.57 17.54 17.65 17.50 15.45 15.41 15.34 10.03 10.00 10.12 31.06 30.99 30.91 16.06 15.97 15.84 10.68 10.53 10.49 15.67 15.81 15.59 13.15 13.04 13.18 10.56 10.59 10.55 1.29 1.30 1.25 3.03 3.04 2.94 0.85 0.76 0.43 0.34 0.60 0.23 0.71 0.90 0.72 0.55 0.18 2.08 1.86

Table 5. Summary of protein results and comparison with Kjeldahl method. Ref. No. Kjeldahl values FlashSmart EA using helium carrier gas FlashSmart EA using argon carrier gas N% Protein % Acceptable range N% Protein % N% Protein % 1 2.35 15.0 14.25 15.75 14.91 2.32 14.79 2 2.19 14.0 13.30 14.70 14.24 2.14 13.67 3 2.66 17.0 16.15 17.85 2.79 17.82 2.75 17.57 4 2.35 15.0 14.25 15.75 2.43 15.50 2.41 15.40 5 1.57 10.0 9.50 10.50 10.26 1.58 10.05 6 4.80 30.6 29.07 32.13 31.62 4.86 30.99 7 2.51 16.0 15.20 16.80 15.98 15.96 8 1.72 11.0 10.45 11.55 10.79 1.66 10.57 9 15.8 15.00 16.59 15.85 2.46 15.69 10 2.01 12.8 12.16 13.44 2.08 13.30 2.06 13.12 11 1.60 10.2 9.69 10.71 10.78 1.66 10.57 12 0.19 1.24 1.178 1.302 0.198 1.27 0.201 1.28 13 0.45 2.88 2.736 3.024 0.453 2.89 0.471 3.00 Conclusions All data shown were obtained with excellent repeatability, and no matrix effect was observed when changing the sample. Good repeatability was obtained with the FlashSmart EA Nitrogen Analyzer using argon as carrier gas, and data from argon and helium were comparable. The data is also comparable with data obtained by the Kjeldahl method. As a complete automatic system, the FlashSmart Elemental Analyzer can analyze nitrogen in a wide content range, in solid and liquid samples, and without the use of sample digestion or the toxic chemicals used in traditional methods. The Dumas combustion method has been approved and adopted by official organizations such as ASBC, AOAC, AACC, AOCS, ISO, and IDF (International Dairy Federation) Standard. The FlashSmart Elemental Analyzer meets all stringent requirements of modern laboratories such as flexibility, repeatability, sensitivity, automation, speed, and low cost per analysis. Acknowledgement Thanks to the Ministry of Health Laboratory, Haifa, Israel, for procuring the wide range of infant food samples analyzed by the FlashSmart Elemental Analyzer. Find out more at thermofisher.com/oea 2017 Thermo Fisher Scientific Inc. All rights reserved. AOAC is a trademark of The Association of Official Analytical Chemists. AOCS is a trademark of The American Oil Chemists Society. AACC is a trademark of The American Association of Cereal Chemists. ASBC is a trademark of The American Society of Brewing Chemists. IFFO is a trademark of The International Fishmeal and Fish Oil Organization. All trademarks are the property of Thermo Fisher Scientific. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. AN42255-EN 0217