AFM Imaging In Liquids. W. Travis Johnson PhD Agilent Technologies Nanomeasurements Division

Similar documents
Imaging Nucleic Acids with the AFM. W Travis Johnson PhD Agilent Technologies Nanomeasurements Division

Basic Laboratory. Materials Science and Engineering. Atomic Force Microscopy (AFM)

Lecture 4 Scanning Probe Microscopy (SPM)

Atomic and molecular interactions. Scanning probe microscopy.

Scanning Force Microscopy

Lecture 12: Biomaterials Characterization in Aqueous Environments

Module 26: Atomic Force Microscopy. Lecture 40: Atomic Force Microscopy 3: Additional Modes of AFM

Contents. What is AFM? History Basic principles and devices Operating modes Application areas Advantages and disadvantages

Scanning Probe Microscopy. Amanda MacMillan, Emmy Gebremichael, & John Shamblin Chem 243: Instrumental Analysis Dr. Robert Corn March 10, 2010

Scanning Force Microscopy II

Atomic Force Microscopy imaging and beyond

Intermittent-Contact Mode Force Microscopy & Electrostatic Force Microscopy (EFM)

Outline Scanning Probe Microscope (SPM)

General concept and defining characteristics of AFM. Dina Kudasheva Advisor: Prof. Mary K. Cowman

STM: Scanning Tunneling Microscope

Agilent Technologies. Scanning Microwave Microscopy (SMM)

Scanning Probe Microscopy. L. J. Heyderman

Interfacial forces and friction on the nanometer scale: A tutorial

Point mass approximation. Rigid beam mechanics. spring constant k N effective mass m e. Simple Harmonic Motion.. m e z = - k N z

Techniken der Oberflächenphysik (Techniques of Surface Physics)

Instrumentation and Operation

Supplementary Figure 1 a) Scheme of microfluidic device fabrication by photo and soft lithography,

Visualization of Nanoscale Components Using Low Cost AFMs Part 2. Dr. Salahuddin Qazi

AFM for Measuring Surface Topography and Forces

NIS: what can it be used for?

INTRODUCTION TO SCA\ \I\G TUNNELING MICROSCOPY

Scanning Tunneling Microscopy

AFM: Atomic Force Microscopy II

Oxford Scholarship Online

INDIAN INSTITUTE OF TECHNOLOGY ROORKEE NPTEL NPTEL ONLINE CERTIFICATION COURSE. Biomedical Nanotechnology. Lec-05 Characterisation of Nanoparticles

Imaging Methods: Scanning Force Microscopy (SFM / AFM)

Contents. Preface XI Symbols and Abbreviations XIII. 1 Introduction 1

Nitride HFETs applications: Conductance DLTS

Water. Hydrogen Bonding. Polar and Nonpolar Molecules. Water 8/25/2016 H 2 0 :

SUPPLEMENTARY INFORMATION

Lecture Note October 1, 2009 Nanostructure characterization techniques

VEDA - Virtual Environment for Dynamic Atomic Force Microscopy

Measurements of interaction forces in (biological) model systems

Introduction to Scanning Probe Microscopy Zhe Fei

Chapters 11 and 12: Intermolecular Forces of Liquids and Solids

Today s SPM in Nanotechnology

The living world has a hierarchy of organizational levels - from molecules to ecosystems

Electonegativity, Polar Bonds, and Polar Molecules

Magnetic Force Microscopy

PHY 481/581. Some classical/quantum physics for the nanometer length scale.

Fundamentals of Atomic Force Microscopy Part 2: Dynamic AFM Methods

L8: The Mechanics of Adhesion used by the Gecko

3.052 Nanomechanics of Materials and Biomaterials Thursday 02/08/06 Prof. C. Ortiz, MIT-DMSE I LECTURE 2 : THE FORCE TRANSDUCER

Keysight Technologies Carbon Nanotube Tips for MAC Mode AFM Measurements in Liquids. Application Note

Introduction to Nanomechanics: Magnetic resonance imaging with nanomechanics

Understanding the properties and behavior of groups of interacting atoms more than simple molecules

Polarity Notes. What is polarity?

The four forces of nature. Intermolecular forces, surface forces & the Atomic Force Microscope (AFM) Force- and potential curves

Scanning Probe Microscopy (SPM)

The Fluid-Coupled Motion of Micro and Nanoscale Cantilevers

And Manipulation by Scanning Probe Microscope

Development of AFM Platform for Directly Measuring the Nanoparticle-Nanoparticle and Nanoparticle-Cell Interactions

Physical Chemistry - Problem Drill 01: Chemistry and Physics Review

Simple Harmonic Motion and Damping

CHEMISTRY Matter and Change. Chapter 12: States of Matter

Nanostructure. Materials Growth Characterization Fabrication. More see Waser, chapter 2

Chapter 10. Dipole Moments. Intermolecular Forces (IMF) Polar Bonds and Polar Molecules. Polar or Nonpolar Molecules?

Chem 11 Unit 4 POLARITY, MOLECULE SHAPE, and BEHAVIOUR

Supplementary Figures

Lorentz Contact Resonance for viscoelastic measurements of polymer blends

Kinetic Molecular Theory of Gases used to account for Ideal Gas Behavior when gases approach high temperatures and low pressures

IMFA s. intermolecular forces of attraction Chez Chem, LLC All rights reserved.

Liquids and Solids The Condensed States of Matter

MS482 Materials Characterization ( 재료분석 ) Lecture Note 11: Scanning Probe Microscopy. Byungha Shin Dept. of MSE, KAIST

MEGASONIC CLEANING OF WAFERS IN ELECTROLYTE SOLUTIONS: POSSIBLE ROLE OF ELECTRO-ACOUSTIC AND CAVITATION EFFECTS. The University of Arizona, Tucson

Scanning Tunneling Microscopy

Introduction to Scanning Tunneling Microscopy

Chapter 2 Correlation Force Spectroscopy

Supporting Information. for. Angew. Chem. Int. Ed. Z WILEY-VCH Verlag GmbH & Co. KGaA

What makes water so special?

Atoms, Molecules, and Life

Atomic Force Microscopy (AFM) Part I

Adventures in nanoscale mechanics. Peter M. Hoffmann Department of Physics & Astronomy Wayne State University

Santosh Devasia Mechanical Eng. Dept., UW

Chapter 11. Intermolecular Forces and Liquids & Solids

Improving the accuracy of Atomic Force Microscope based nanomechanical measurements. Bede Pittenger Bruker Nano Surfaces, Santa Barbara, CA, USA

CNPEM Laboratório de Ciência de Superfícies

CHAPTER 11: Intermolecular Forces, Liquids, and Solids. Are there any IDEAL GASES? The van der Waals equation corrects for deviations from ideality

Fabrication of ordered array at a nanoscopic level: context

The Powerful Diversity of the AFM Probe

Electrochemical Scanning Tunneling Microscopy

Lorentz Contact Resonance for viscoelastic measurements of polymer blends

16 years ago TODAY (9/11) at 8:46, the first tower was hit at 9:03, the second tower was hit. Lecture 2 (9/11/17)

Liquids & Solids: Section 12.3

THE CHEMISTRY OF LIFE

Program Operacyjny Kapitał Ludzki SCANNING PROBE TECHNIQUES - INTRODUCTION

What determines whether a substance will be a solid, liquid, or gas? Thursday, April 24, 14

Lecture Presentation. Chapter 11. Liquids and Intermolecular Forces. John D. Bookstaver St. Charles Community College Cottleville, MO

bio-molecular studies Physical methods in Semmelweis University Osváth Szabolcs

CHEMISTRY - TRO 4E CH.11 - LIQUIDS, SOLIDS & INTERMOLECULAR FORCES

SUPPLEMENTARY NOTES Supplementary Note 1: Fabrication of Scanning Thermal Microscopy Probes

Chapter 3. Water and the Fitness of the Environment

Supplementary Information for. Effect of Ag nanoparticle concentration on the electrical and

Chapter 10. Liquids and Solids

How Does a Microcantilever Work?

Transcription:

AFM Imaging In Liquids W. Travis Johnson PhD Agilent Technologies Nanomeasurements Division

Imaging Techniques: Scales Proteins 10 nm Bacteria 1μm Red Blood Cell 5μm Human Hair 75μm Si Atom Spacing 0.4nm DNA 2nm Virus 50nm Cell 30μm Ant 5mm 10-10 10-9 10-8 10-7 10-6 10-5 10-4 10-3 10-2 10-1 1 Angstrom Nanometer TEM AFM Micron Millimeter Meter Near Field Optical Optical Microscope SEM

A C B D AFM Basic Configuration Z Y X

Liquid Cell and Perfusion (flow through) Cell

Intermolecular/Atomic Interactions Important in AFM Ionic Interactions (charge-charge) Si AFM probes (SiO - ) SiN AFM probes (SiO - and NH 3+ ) VDW interactions ( electrostatic between uncharged molecs) Induced Dipole (aka Dispersion or London Forces; temp., induced & fluctuating dipoles) Dipole-Dipole (perm dipoles) VDW repulsion (VDW radius) Atoms near end of probe vs atoms further away Hydrophobic (nonpolar; decay exp with distance) Surface tension (meniscus forces, capillary; in air) Can Modify Ionic, VDW, Hydrophobic Properties Perform Chemistry on AFM probe (coat with hydrophobic or charged silanes)

VDW Interactions: AFM Probe - Sample Cantilever Deflection in Force (N) or Voltage (V) VDW Repulsion contact non-contact Distance (tip position) WDW radius Dipole-Dipole Attractions Adhesion greatly exaggerated

H Bonding (RO.. H, RS.. H, RN.. H) Ionic Interactions Change (salt bridges) Hydrodynamic Forces Viscous damping Viscosity & Probe Geometry Increase in Effective Mass Res Freq Reduction & Peak Broadening Compression Repulsive Force Viscosity in Confined Areas (Viscoelasticity) Liquid Lowers Q Factor AFM Probe Sensitivity & Responsiveness Q100 Air ~ Q2-3 Liquid Low Q Damage Soft Samples Q Further Decreases as Tip Surface Thermal Fluctuations (< noise) Oscillatory Forces (F liquid ) Smooth Surfaces (liquid order, packing) Rough Surfaces (liquid asymmetry) Can Cause Artifacts Size of the Liquid Molecules Liquid New Freqs Arise for Dynamic (AC) Imaging Modes Additional Forces in Liquid

Oscillatory Forces & Viscosity Force curve showing increase in viscosity of confined liquid between tip and sample. As tip approaches surface, viscosity increases. O Shea Jpn. J. Appl. Phys. 40 (2001) 4309 AFM tip surface roughness can disrupt tip-liquidsurface interface. Roughness introduces disorder, reduces packing and helps to minimize oscillatory forces. Smaller solvent molecules also increase averaging. AFM tip immersed in liquid. In high res imaging, force between tip-sample (F image ) should be small. Solvent forces (F liquid ) also interact between tip and sample.

Immobilization Chemistry is Critical for Samples in Liquid (e.g., Biological Samples) OH OH OH OEt Si O Si O Si + EtO Si NH 2 OEt APTES mica EtOH Electrostatic Attachment positively charged protein OEt EtO Si NH 2 OH O OH Si O Si O Si Negatively charged mica O O H H Glutaraldehyde EtO Si N H OH Si OEt O OH O Si O Si OH Glut-Mica O H 2 N H 2 N H 2 N NH 2 H 2 N NH 2 Proteins (lysine groups) Covalent Protein Immobilization on Mica

Contact Mode Imaging y X Dynamic in x and y Tip in contact or near contact with surface Small vertical force Probe exerts lateral forces on sample Damage to soft samples Weakly bound samples move which lowers lateral resolution X Topography (height) Deflection (error) Data Channels (typical) Topography Signal representing the voltage applied to Z piezo in order to maintain constant force between tip and sample. Deflection Servo input or error signal which shows difference between setpoint and actual deflection. Shows edges clearly

Deflection Images of Live BCE Cell LOW FORCE HIGH FORCE 200 nm 100 um 25 um

AC Mode Imaging z y X X Piezoelectric transducer shakes cantilever holder at or near resonant frequency of cantilever Dynamic in x, y, & z Intermittent contact Interaction with sample reduces oscillation amplitude. Reduction in amplitude used as feedback signal Soft surfaces stiffened by viscoelastic response Impact forces predominately vertical Large vertical force, small lateral force Higher lateral resolution than contact mode ~ V

Short Range Interactions (Attraction/Repulsion) Effect Resonance Frequency Attractive gradient is equivalent to additional spring tension on tip, which reduces resonance frequency of cantilever ΔF Attractive Repulsive Attractive ΔF Repulsive Repulsive gradient is equivalent to additional spring compression on tip, which increases resonance frequency of cantilever

Phase Imaging Drive Signal Detected Signal Before contact 500nm Contact with hard material dampens amplitude but has little effect on phase Topography image of block copolymer Contact with soft material effects phase and amplitude Phase lag related to material stiffness 500nm Phase image of block copolymer

AC AFM Data Channels (typical) Topography Derived from voltage applied to Z piezo needed to keep oscillation amplitude constant Amplitude Error signal from photo detector that shows the change in amplitude Phase Phase lag of cantilever response with respect to drive signal Topography Amplitude Phase Image courtesy of Nanotechnology Center of Tsinghua University AC Mode images of inner surface of blood vessel in buffer

AC AFM in Liquid AAC Mode A solenoid applies an AC field to piezoelectric transducer Transducer-shakes the cantilever holder Reduction of amplitude signals contact Forest of Peaks in Liquids MAC Mode Cantilever is coated with magnetic film Magnetic field drives cantilever ONLY Cleaner Oscillation in Liquids

Human Rhino Virus 2 (HRV2) 300 nm 70 nm 20 nm Protein Loops 10 nm 0 1.5 nm HRV2 Absorbed on Mica/NiCl Kienberger et al., Single Mol. 2 (2001) 99

In situ Experiment: RNA Release from HRV2 ph= 4.1 2 hours 300 nm 300 nm 900 nm Kienberger et al., J. Virol. 78 (2004) 3203

Thank You Questions?