Routine MS Detection for USP Chromatographic Methods

Similar documents
The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

UPLC Intact MASS Analysis Application Kit

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Successfully Scaling and Transferring HPLC and UPLC Methods

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Basic Principles for Purification Using Supercritical Fluid Chromatography

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Peptide Isolation Using the Prep 150 LC System

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS)

Dilution(*) Chromatography

Fast methods for the determination of ibuprofen in drug products

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Developing Analytical Chromatographic Methods for Pharmaceutical Stability Investigations

LC Technical Information

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

[application note] ACQUITY UPLC/SQD ANALYSIS OF POLYMER ADDITIVES. Peter J. Lee, and Alice J. Di Gioia, Waters Corporation, Milford, MA, U.S.A.

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

[ a ppl ic at ion no t e ]

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

HPLC Praktikum Skript

UPLC Method Development and Validation

Multi-Channel SFC System for Fast Chiral Method Development and Optimization

Varian Galaxie Chromatography Data System for Preparative HPLC

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

It s hot! ZIC -chilic

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

PosterREPRINT AN INTERACTIVE PHYSICOCHEMICAL PROPERTY PROFILING SOFTWARE FOR EARLY CANDIDATE ANALYSIS IN DRUG DISCOVERY INTRODUCTION

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

with diode array detection

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Selecting Detectors for Compounds with No Optical Absorbance

ACQUITY UPLC Peptide BEH C 18, 130Å and 300Å Columns

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

BEST PRACTICES IN THE ANALYSIS OF RAPIFLUOR-MS LABELED GLYCANS USING THE ACQUITY QDa DETECTOR (PERFORMANCE MODEL)

Chromatography. Gas Chromatography

APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEY WORDS. Simultaneous high-throughput screening of melamine (MEL) and cyanuric acid (CYA).

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

ImprovIng ADmE ScrEEnIng productivity In Drug DIScovEry

IN QUALITATIVE ANALYSIS,

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

[ Care and Use Manual ]

Determination of trace anions in concentrated hydrofluoric acid

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Modernizing the USP Monograph for Acetaminophen

ultimate efficiency unleashed [ Cortecs 1.6 µm columns ]

Expediting Achiral SFC Method Development Using a Multi-Channel SFC System with MS Detection

High Performance Liquid Chromatography

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

ANALYSIS OF BASIC COMPOUNDS

Transferring Yesterdays Methods to Tomorrows Technology: The Evolution from HPLC to UPLC TM

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

for Acclaim Mixed-Mode HILIC-1 Column

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TOF LC/MS Combined with Advanced MassHunter Data Processing Software

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Transcription:

Daniel S. Root, Thomas E. Wheat, and Patricia McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS This approach enables mass spectral analysis of peaks directly from unmodified USP HPLC chromatographic methods. INTRODUCTION The USP analytical methods monographs are widely used in the development and quality control of drug substances and finished drug products. These methods are typically robust and are suitable for the identification and assay of active pharmaceutical ingredients (APIs) and their related substances and impurities. There are circumstances, however, where alternative analytical methods are necessary. By far, the more common circumstance revolves around the needs of validation. In order to satisfy the selectivity and specificity criteria common to any methods validation, peak identification and purity are required. In this case, alternative methods are routinely employed to complement the original optical detection methods. Less common, but no less important, is the circumstance of the unexpected peak appearing in a routine analysis. Undoubtedly, this kind of serious event will necessitate a formal investigation within which early steps will focus on the identification of the offending peak. Again, alternative analytical methods must be engaged to achieve this goal. WATERS SOLUTIONS ACQUITY UPLC System with 2D-LC Technology ACQUITY QDa Detector MassLynx 4.1 Software with Waters Pump Control KEY WORDS ACQUITY QDa, mass spectrometry, USP, heart-cutting, 2D Mass spectral detection (MS) is the most predominant form of alternative detection. The richness of mass spectral data makes it an obvious first choice. Unfortunately, the majority of USP methods use mobile phases made up of nonvolatile buffers, limiting detection to UV or other optical methods. To utilize MS, time-demanding fraction collection, evaporation, and reconstitution then become part of the process. To overcome this problem, numerous protocols have been proposed to convert these chromatographic methods to completely volatile mobile phases, making the methods compatible with mass spectral analysis. This redevelopment can be complicated and costly. Furthermore, it can prove difficult to demonstrate that the original chromatographic selectivity has been perfectly preserved. The ACQUITY UPLC System with 2D-LC Technology and integrated ACQUITY QDa Mass Detector offers a solution to this challenge. With this system, a peak of interest can be transferred from an unmodified USP HPLC method, trapped, the mobile phase replaced with a volatile MS-friendly mobile phase, and the peak eluted to a mass spectral detector. This application note presents a systematic protocol for using MS detection directly with unmodified USP methods. The USP assay for Irbesartan is used as an example to illustrate the peak transfer protocol. 1

E X P E R IM E N TA L Method conditions System: ACQUITY UPLC System with 2D-LC Technology Alpha pump: ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM) Beta pump: ACQUITY UPLC I-Class Binary Solvent Manager (BSM) Dilution pump: ACQUITY Isocratic Solvent Manager (ISM) Sample manager: ACQUITY UPLC H-Class Sample Manager Flow-Through-Needle (SM-FTN) Column manager (CM-A) configured for two active pre-heaters, 2D column tubing kit (p/n 25764) and 2D valve kit (p/n 25952) UV detection: ACQUITY UPLC Photodiode Array (PDA) Detector 22 nm Mass detection: ACQUITY QDa Detector Software: MassLynx 4.1 SCN 888 with Waters pump control Chromatographic Parameters USP method HPLC column: XSelect HSS T3 4.6 x 25 mm 5 μm (L1) Mobile phase: Buffer solution and acetonitrile (6:4) Buffer solution:.55 Phosphoric acid in water adjusted to ph 3.2 with triethylamine Flow profile: Isocratic Flow rate: 1. ml/min Injection: 1 µl LC-MS method UPLC column: BEH C 18, 2.1 x 5 mm, 1.7µm Mobile phases: Water, acetonitrile Dilution pump:.1 formic acid in water,.2 ml/min Flow profile: Gradient to 95 Acetonitrile, 3 minutes,.5 ml/min Mass detector: ACQUITY QDa Ionization mode: ESI+, scan Acquisition range: Full scan 1 5 m/z SIR functions: Irbesartan m/z 429.2 Irbesartan related compound A (RC-A) m/z 447 2

Sample Irbesartan and Irbesartan related compound A (RC-A) were purchased from USP (Rockville, MD 2852). Irbesartan sample and standard solutions were prepared as per the USP monograph for Irbesartan. 1 The peak transfer protocol When designing a multi-dimensional protocol, it is useful to begin with a simple bulleted list describing the main tasks and steps of the process. A progression then follows translating the protocol into the actual plumbing and valve events, and finally, transcribing these events into a form that can be entered into the method editor and gradient table of an instrument method. For this peak transfer, the process can be initially described by the following protocol. The standard USP method is run with UV detection. In fact, the hallmark of this approach is that the USP HPLC method can be run without any changes or modifications When the peak begins to elute from the HPLC column, entered in the method as a timed event, the following steps occur: flow from the PDA is diverted to the UPLC Column flow from the Beta pump is delivered to the HPLC column flow from the Dilution pump mixes with the Beta pump mobile phase diluting the organic composition of the flow entering the UPLC column (at-column dilution) so that the compound does not continue to migrate along the UPLC column flow from the UPLC column is sent to waste The USP mobile phase from the Alpha pump elutes from the HPLC column directly to waste The peak is now trapped at the head of the UPLC column The MS-compatible mobile phase from the Beta pump, which is free of any non-volatile additives, flushes the UPLC column, replacing the USP mobile phase Flow through the UPLC column remains diverted to waste until the flushing is complete With the original USP mobile phase flushed out by the additive-free Beta pump flow, the flow from the UPLC column is diverted to the ACQUITY QDa Detector The compound of interest, still retained on the UPLC column, is then eluted to the ACQUITY QDa Detector with an increasing gradient of acetonitrile The ACQUITY QDa Detector then monitors the eluate using a single ion channel for expected compounds and full scan mode to collect full mass spectral information for use in potential unknown identification 3

To execute these steps in the above protocol, a specific configuration of fluidic connections was made using the two, 2-position, 6-port valves of the ACQUITY UPLC Column Manager. Figures 1 through 4 show the column manager plumbing and describe the key events associated with the steps in the peak transfer process. To complete the protocol all of the described tasks, plumbing, and pump and valve events must be transcribed into the format of an instrument method. An example of a final gradient table with pump parameters and valve events can be seen in Table 1. Individual module parameters have been consolidated into a single table. The time settings marked with an asterisk are the start and stop times for the peak transfer window. In all methods of this type, as the settings for the start and end of the peak transfer window change, so do all subsequent timed events in the method. Dilution Pump PDA detector Mixer UPLC column Alpha pump Beta pump HPLC column ACQUITY QDa Detector Figure 1. Step 1. The flow from the Alpha pump (USP mobile phase) is directed to the HPLC column. The sample is injected and the USP method is run with UV detection (red flow path). The Beta pump flows through the equilibrated UPLC column (blue path), then out to waste. Dilution pump PDA detector Alpha pump Beta pump Mixer UPLC column HPLC column 2 4 1 5 ACQUITY QDa Detector Figure 2. Step 2. When the desired peak begins to elute, a timed event, the left valve turns to position two. The Alpha pump flow is diverted to waste and the flow from the Beta pump is directed through the HPLC column eluting the peak of interest onto the UPLC column (Blue flow path). At the same time, the flow from the dilution pump is combined with that from the HPLC column at a low-volume mixer. The reduced organic composition of the diluted mobile phase facilitates the trapping of the peak of interest on the UPLC column. 4

Dilution pump PDA detector Mixer UPLC column Alpha pump Beta pump HPLC column ACQUITY QDa Detector Figure 3. Step 3. With the peak transfer to the UPLC column complete, the left valve returns to position 1 and the HPLC column is now regenerated. The Beta pump flushes the UPLC column to waste with 1 water retaining the analyte. Dilution Pump PDA detector Mixer UPLC column Alpha pump Beta pump HPLC column ACQUITY QDa Detector Figure 4. Step 4. With the mobile phase flush of the UPLC column complete, the right valve turns to position two and the peak of interest is eluted to the ACQUITY QDa Detector with a gradient of increasing acetonitrile. QSM (Alpha pump) A: Buffer B: Acetonitrile BSM (Beta pump) A: Water B: Acetonitrile Table 1. An example gradient table with pump parameters and valve events for the peak transfer protocol. The asterisks denote the start and stop time of the peak transfer window. ISM (Dilution pump).1 Formic in water Flow A B Curve Flow A B Curve Flow A. 1. 6 4.2 6 4. 1 1 12.9*.1 6 4 11.2 6 4 6. 14.4*.1 6 4 6.2 6 4 6. 1 1 14.41.1 6 4 6. 1. 1 1 18.4.1 6 4 6.5 1 6. 1 1 18.41.1 6 4 6.5 1 6. 1 2 21.4 1. 6 4 6.5 5 95 6. 1 2 25. 1. 6 4 6.2 6 4 1. 1 1 Left valve CM-A Right valve 5

At-column dilution (ACD) At-column dilution is a technique used to facilitate the injection of large volumes of sample diluent containing high amounts of organic solvent onto reversed-phase columns without compromising the chromatography. Typically, when a large volume of highly organic diluent is injected onto a reversed-phase column, poor analyte retention and chromatographic peak distortions are observed. The magnitude of the distortion will be dependent on the relative organic composition of the sample diluent and the hydrophobicity of the analytes, and can vary from a shift in retention to peak splitting and smearing. During a heart-cut, or peak transfer, a specific volume of the mobile phase, containing the peak of interest, is transferred to a second-dimension column. If that cut contains a high percent of organic, the same poor retention and chromatographic distortions mentioned above can be observed on the second-dimension column. By adding an additional aqueous dilution flow to the injected or cut volume prior to the second-dimension column, the organic composition can be significantly reduced, resulting in improved retention and eliminating possible chromatographic peak distortions. This is the basis of Waters at-column dilution [US Patent (Waters Corporation) 6,79,361 B2]. A sample diagram of at-column dilution is shown in Figure 5. When using ACD with a peak transfer, the flow from the dilution pump is set to 1 aqueous and mixes with the eluate from the first dimension column prior to reaching the head of the seconddimension column. The final organic composition of the mobile phase as it flows into the second-dimension column is determined by the proportion of the two flow rates. In this application, the dilution pump flow matched that of the Beta pump during peak transfer. This lowered the organic composition of the mobile phase from 4 acetonitrile in the USP mobile phase to 2 passing through the second-dimension column. Previous method development experiments (data not shown) confirmed this to be an appropriate composition for adequate retention of Irbesartan and its related compounds. 4 organic =.2 ml /min From HPLC column From Dilution pump Low volume mixer or T 1 aqueous =.2 ml/min 2 organic =.4 ml/min To HPLC column Figure 5. Principle of Waters at-column dilution applied to this peak transfer. AU 2.2e-1 2.e-1 1.8e-1 1.6e-1 1.4e-1 1.2e-1 1.e-1 8.e-2 6.e-2 4.e-2 2.e-2. Unidentified peak Irbesartan 5. 7.5 1. 12.5 15. 17.5 2. 22.5 Figure 6. UV chromatogram of the USP assay for Irbesartan. RESULTS AND DISCUSSION Peak transfer The USP assay for Irbesartan was run, with UV detection, as a single-dimension method on the ACQUITY UPLC System with 2D-LC Technology. The UV chromatogram obtained with this method can be seen in Figure 6. The expected Irbesartan API peak (retention time 15.3 min) is present, but a small, unidentified peak (12.7 min) is also observed. This separation exemplifies the two circumstances mentioned previously. The Irbesartan API peak will be transferred for the purposes of confirmation as would be required in the context of a validation. Being an unexpected peak, the unknown at 12.7 minutes will be transferred and mass spectral data collected to begin characterization. 6

The USP method was run on the ACQUITY UPLC System with 2D-LC Technology configured for peak transfer (heart-cut). The valve event settings were adjusted to transfer a portion of the large API peak. After the USP mobile phase is flushed from the UPLC column, an organic gradient of acetonitrile was used to elute the analyte to the ACQUITY QDa Detector. The SIR results confirmed the identity of the API as Irbesartan m/z 429.2 (Figure 7). The protocol was repeated for the small, unidentified peak at 12.7 minutes. A common impurity in the formulation is Irbesartan related compound A (RC-A). A selective ion channel for this specific mass was screened. Subsequent elution of this unknown peak to the mass spectral detector confirmed the analyte as Irbesartan RC-A by SIR at m/z 447.1 (Figure 8). 31 Irbesartan SIR m/z 429.2 2. 2.25 2.5 2.75 21. 21.25 21.5 21.75 22. 22.25 22.5 TIC m/z 1-5 2. 2.25 2.5 2.75 21. 21.25 21.5 21.75 22. 22.25 22.5 Figure 7. Confirmation of the Irbesartan API peak. In each transfer, the USP UV method was run unaltered. The transfer window for the peak transfer, determined from the time settings for the valve events, was adjusted to divert either part or all of the volume of the selected peaks. For a large, concentrated peak such as the API, only a small portion is required to be transferred and eluted to the mass spectral detector for mass assignment. For smaller peaks, such as RC-A, it was desirable to transfer the entire volume of the chromatographic peak to ensure the best possible spectral data. This would be a necessity for quantification. 1 RC-A SIR m/z 447.1 18.5 19. 19.5 2. 2.5 21. TIC m/z 1-5 Repeatability To assess the repeatability of the peak transfer, five consecutive injections were performed. There was no variation in the seconddimension elution time of the transferred RC-A peak. Area repeatability for the five injections had a RSD of 4.2 (Figure 9), indicating the automated peak transfer process was highly repeatable. 18.5 19. 19.5 2. 2.5 21. Figure 8. Confirmation of the unexpected peak as Irbesartan RC-A by SIR. 1 1 2. 21. 22. 23. 24. 25. 26. 2. 21. 22. 23. 24. 25. 26. 1 1 1 2. 21. 22. 23. 24. 25. 26. 2. 21. 22. 23. 24. 25. 26. Injection Area 1 18947 817 3 17823 4 17126 5 1721 mean 17853 s 75 RSD 4.2 2. 21. 22. 23. 24. 25. 26. Figure 9. Repeatability of the automated protocol. 7

At-column dilution The use of at-column dilution in this protocol was critical to the quality of the chromatographic results. Figure 1 compares the gradient eluted RC-A peak observed with and without ACD. With the dilution pump off (no ACD), the Beta pump elutes the transferred peak volume to the UPLC column using the USP mobile phase containing 4 acetonitrile. Upon final gradient elution, the RC-A peak is asymmetric, fronting strongly. The organic composition of the undiluted mobile phase was too high for good retention on the second dimension. With the dilution pump flowing (ACD) and the flow rate matching that of the Beta pump during the peak transfer, the organic composition of the mobile phase was reduced by half to 2 acetonitrile. The lower percent of organic improved the retention of RC-A on the second-dimension column, exemplified by the symmetrical RC-A peak. It is important to note that RC-A is a moderately hydrophobic analyte. A more hydrophilic compound would have displayed more severe peak distortions (splitting, retention shifts). CONCLUSION Mass spectral detection of peaks from USP analyses can be problematic due to the high ionic strength, non-volatile mobile phases often used with these methods. When MS is needed, the choices have been fraction collection and solvent exchange prior to a reanalysis, or the development of a separate method using a volatile mobile phase. Both choices incur penalties in time and cost. The ACQUITY UPLC System with 2D-LC Technology system configured for peak transfer can directly transfer a selected peak from a USP chromatographic method to the ACQUITY QDa Mass Detector for mass spectral analysis with no fraction collection, and no redevelopment of the USP method. Configuring the system for at-column dilution will further ensure the quality and repeatability of the mass spectral data collected from the transferred peak. References 1. USP Monograph Irbesartan, USP37-NF32, 3396. The United States Pharmacopeial Convention, official from August 1, 214. Without ACD 1 19. 19.9.4 19.6 19.8 2. 2.2 2.4 2.6 With ACD 2 19. 19.9.4 19.6 19.8 2. 2.2 2.4 2.6 Figure 1. Waters at-column dilution improves the quality of the chromatographic result. Waters, ACQUITY, ACQUITY UPLC, UPLC, MassLynx, QDa, XSelect, and The Science of What s Possible are registered trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 215 Waters Corporation. Produced in the U.S.A. February 215 72536EN AG-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 2 F: 1 58 8799 www.waters.com