Beers Law Instructor s Guide David T. Harvey

Similar documents
An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13)

Designing Acid-Base Titrations: Instructor s Guide David T. Harvey

Beer's Law and Data Analysis *

Chem 310 rd. 3 Homework Set Answers

Investigating Food Dyes in Sports Beverages. Sample

DETERMINATION OF AN EQUILIBRIUM CONSTANT

Spectroscopy II Introduction: Spectrophotometry and Fluorometry

EXPERIMENT #3 A Beer's Law Study

Bio 120 Lab 5: Quantitative Analysis

This activity has been password protected to prevent modification. In order to request an unprotected version of this activity, contact

you-try-it-04.xlsx Step-by-Step Guide ver. 9/4/2009

Experiment 7: Adsorption Spectroscopy I, Determination of Iron with 1,10 Phenanthroline

Concentrations that absorb. Measuring percentage transmittance of solutions at different concentrations

Introduction. The amount of radiation absorbed may be measured in a number of ways: Transmittance, T = P / P 0 % Transmittance, %T = 100 T

Beer's- Lambert Law and Standard Curves. BCH 312 [Practical]

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

SPECTROPHOTOMETRY/BEER S LAW LECTURE HONORS CHEMISTRY NAME

Lab Investigation 4 - How could you make more of this dye?

Chemistry 213. A KINETIC STUDY: REACTION OF CRYSTAL VIOLET WITH NaOH LEARNING OBJECTIVES

Spectrometric Determination of the Acid Dissociation Constant of an Acid-base Indicator

DETERMINATION OF K c FOR AN EQUILIBRIUM SYSTEM

Spectrophotometry. Introduction

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

Experiment 6: Determination of the Equilibrium Constant for Iron Thiocyanate Complex

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Experiment 2: The Beer-Lambert Law for Thiocyanatoiron (III)

Chem 321 Lecture 18 - Spectrophotometry 10/31/13

MOLEBIO LAB #4: Using a Spectrophotometer

Determining the Concentration of a Solution: Beer s Law

ANALYSIS OF ZINC IN HAIR USING FLAME ATOMIC ABSORPTION SPECTROSCOPY

MORE LIGHTS, COLOR, ABSORPTION!

Determining the Concentration of a Solution: Beer s Law

Experiment 11 Beer s Law

Experiment 11 Beer s Law

CHEMISTRY SEMESTER ONE

Instructor s Advance Preparation

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Spectroscopy Primer. for ultraviolet and visible absorbance spectroscopy. by Stephanie Myers Summer 2015

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Basic Instrumentation. Learning Objectives:

Experiment 13I THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Lab 5: Calculating an equilibrium constant

Determining the Conductivity of Standard Solutions

Introduction to Spectroscopy: Analysis of Copper Ore

Experiment 7A ANALYSIS OF BRASS

Student Exploration: Bohr Model of Hydrogen

UV-Visible Absorption Analysis of a Two Component System

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample

Practical 1P4 Energy Levels and Band Gaps

Practical 1P4 Energy Levels and Band Gaps

The Spectrophotometer and Atomic Spectra of Hydrogen Physics 246

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

UV/VIS SPECTROMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

1901 Application of Spectrophotometry

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

Experiment 12: SPECTROSCOPY: EMISSION & ABSORPTION

WISE Regression/Correlation Interactive Lab. Introduction to the WISE Correlation/Regression Applet

Introduction to Spectroscopy: Analysis of Copper Ore

EXPERIMENT 14. ACID DISSOCIATION CONSTANT OF METHYL RED 1

Math 2311 Written Homework 6 (Sections )

CHEM 254 EXPERIMENT 9. Chemical Equilibrium-Colorimetric determination of equilibrium constant of a weak acid

Clinical Chemistry (CHE221) Professor Hicks Week 1. Statistics Made Slightly Less Boring and Introduction to Spectrophotometry. Accuracy vs Precision

Determination of an Equilibrium Constant

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Exp 03 - Reaction Rate

Spectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages )

MOLECULAR AND ATOMIC SPECTROSCOPY

Spectrophotometry Materials

Spectrometer User s Guide

Regression on Faithful with Section 9.3 content

Lab #12: Determination of a Chemical Equilibrium Constant

R O Y G B V. Spin States. Outer Shell Electrons. Molecular Rotations. Inner Shell Electrons. Molecular Vibrations. Nuclear Transitions

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

A Study of Beer s Law Prelab

Basics of UV-Visible Spectroscopy *

Determining the Concentration of a Solution: Beer s Law

Background The power radiated by a black body of temperature T, is given by the Stefan-Boltzmann Law

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Chapter 18. Fundamentals of Spectrophotometry. Properties of Light

EXPERIMENT 6 INTRODUCTION TO SPECTROSCOPY

Reflection = EM strikes a boundary between two media differing in η and bounces back

CHEMICAL KINETICS E + 2B 2C + D (1)

CEE 772: Instrumental Methods in Environmental Analysis

Experiment 4 Radiation in the Visible Spectrum

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Chemistry 112 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium

Absorption photometry

Kinetics of Crystal Violet Bleaching

CHEM Lab 7: Determination of an Equilibrium Constant using Spectroscopy

Experimental Procedure Overview

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Determining the Rate Law and Activation Energy for the Methyl Blue Reaction:

SPECTROPHOTOMETERS. Visible and UV-Visible

Beer s Law 29 January 2019

BETA-RAY SPECTROMETER

MEASUREMENT: PART II

AIM To verify Beer - Lambert s law and to determine the dissociation constant (Ka) of methyl red, Spectrophotometrically.

09/05/40 MOLECULAR ABSORPTION METHODS

Transcription:

Beers Law Instructor s Guide David T. Harvey Introduction This learning module provides an introduction to Beer s law that is designed for an introductory course in analytical chemistry. The module consists of seven investigations: Investigation 1: Recalling Beer s Law Investigation 2: Absorbance Spectra Investigation 3: Calibration Curves Investigation 4: Fundamental Limitation to Beer s Law Investigation 5: Instrumental Limitation to Beer s Law: Polychromatic Radiation Investigation 6: Instrumental Limitation to Beer s Law: Stray Light Investigation 7: Chemical Limitation to Beer s Law The learning module is programmed in R (www.r-project.org) using the Shiny package, which allows for interactive features. Each investigation includes a brief introduction, an explanation of the controls sliders, radio buttons, and check boxes available to the user and the type of plots, tables, and statistical summaries produced by the underlying code. Each investigation also includes one or more questions to answer. The purpose of this document is to provide instructors with additional background on the program s features and data sets, to provide representative examples of the results students might generate, and to provide suggestions of possible ways to make use of the module s investigations. Some Background Details on the Learning Module s Data The data set included with this learning module consists of two objects: a dataframe with 351 rows and 20 columns, and a numerical vector with 14 values. As shown here, each row of the data frame provides information for one wavelength in the range 400 nm to 750 nm, and each of the 20 columns provides values for a different variable: Column 1: wavelengths Column 2: values for a 5.000 mm standard solution of the analyte (noise free) Column 3: transmittance values for a 5.000 mm standard solution of the analyte (noise free) Column 4: power spectrum for the reference (noise free) Column 5: power spectrum for the sample (noise free) Column 6: ɛb values for the analyte (noise free) Column 7: values for a blank (with 0.2%T noise added) Columns 8-20: values for a series of standards with concentrations of 0.100 mm, 0.200 mm, 0.400 mm, 0.500 mm, 0.600 mm, 0.800 mm, 1.000 mm, 2.000 mm, 4.000 mm, 5.000 mm, 6.000 mm, 8.000 mm, and 10.000 mm (with 0.2%T noise added) The numerical vector contains the concentrations of analyte in each of the standards, as noted above. For those wishing to explore the data, files are available at the application s Github site. The file BL.RData contains both the dataframe and the vector and can be loaded directly in R. The file rawdata.csv contains the dataframe only and be opened in Excel or read into R. 1

A brief description of how this data was created is in order. The power spectrum for the reference is taken from the emission profile of a tungsten-halogen visible source; the original data is from Thor labs for their model SLS201 light source and gives power as a function of wavelength from 400 nm to 2600 nm in 10 nm intervals. This data was trimmed to the range 400 nm to 750 nm and additional values were added at 1 nm intervals using a linear interpolation over each 10 nm interval. The resulting power spectrum was assigned as the reference and is stored in the data frame s fourth column. The values for the standard solution of 5.000 mm analyte were simulated in R using a function written for that purpose; the resulting spectrum is the sum of three Gaussian curves centered at 450 nm, 550 nm, and 650 nm, with respective standard deviations of 30 nm, 30 nm, and 20 nm, and respective peak heights of 25, 10, and 50. The resulting spectrum is shown here in blue (with values stored in the data frame s second column) and has a maximum at 649 nm; the dashed red lines show the three Gaussian functions that make up the spectrum, although these are not relevant to the remainder of the learning module. 0.0 0.4 0.8 400 450 500 550 600 650 700 750 wavelength This spectrum was used to generate the ɛb values in the data frame s sixth column by dividing by the standard s concentration of 0.00500 M, and the transmittance values in the data frame s third column were calculated as T = 10 A. The sample s power spectrum was calculated as P samp = T P ref and stored in the data frame s fifth column. Finally, the spectrum for the blank and for each standard was created using the following equation A = ɛbc (corr) where the standard values for ɛb are in the data frame s sixth column and the concentrations of the standards are in their vector, and where the correction factor (corr) corr = c(1.000, 0.9987, 0.9974, 0.9948, 0.9934, 0.9922, 0.9896, 0.9870, 0.9624, 0.9271, 0.9040, 0.8835, 0.8316, 0.7733) is designed to make obvious a fundamental deviation from Beer s law when the analyte s concentration is greater than 2.000 mm. Noise was added to each value by first converting it to a transmittance, then adding a value drawn from a random normal distribution centered on zero with a standard deviation of 0.002 (0.2%T), and, finally, converting the transmittance back to an value. The resulting spectra for the blank and for all 13 standards are shown here 2

0.0 0.5 1.0 1.5 400 450 500 550 600 650 700 750 wavelength Investigation 1 This investigation asks students to remember that Beer s law relates a sample s to the concentration of the absorbing species, the distance light travels through the sample, and the absorbing species molar absorptivity. The intent here simply is to review how these four values relate to each other. Question 1. For the first set of cuvettes, students should recognize that the values are identical when viewed from the side because ɛ, b, and C are constant, but that when viewed from the top, the values increase proportionally with the increase in the solution s height. For the second set of cuvettes, students should recognize that the product of b and C is the same for all four cuvettes when viewed from the top, a result that is consistent with a constant value for ɛ. Question 2. Students should recall that molar absorptivity provides a measure of the relative probability that the analyte will absorb a photon of light of a particular wavelength, and that it has general units of cm -1 M -1, or, more generally, length -1 concentration -1. Question 3. The molar absorptivity is (4X) -1 cm -1 M -1. Investigation 2 This investigation presents students with the power spectra for the reference and for a standard solution that is 5.000 mm in analyte, the data for which are available as a.csv file using the Download button. Question 1. Students should recall that A = logt and that logt = ɛbc. Question 2. To create the transmittance spectrum and the spectrum, students should use the equations T = P sam /P ref and A = logt, which yields the results shown here 3

transmittance 0.0 0.4 0.8 400 450 500 550 600 650 700 750 wavelength (nm) 0.0 0.4 0.8 1.2 400 450 500 550 600 650 700 750 wavelength (nm) Students can check their work by using the radio buttons to display the transmittance or the spectrum and to view the first six values for each. Investigation 3 In this investigation, students are presented with the spectra for the blank and for a set of standards with concentrations of 0.200 mm, 0.400 mm, 0.600 mm, 0.800 mm, and 1.000 mm. The slider, which increments in steps of 1 nm, allows students to adjust the analytical wavelength and to retrieve the data for the standards at this wavelength. After plotting the calibration data and completing a regression analysis, students can evaluate their results by checking the box for Show Model? and examining the resulting regression model. Question 1. The calibration curve and the regression model when using the default wavelength of 450 nm are shown here 4

0.00 0.10 0.20 0.0 0.2 0.4 0.6 0.8 1.0 ## ## Call: ## lm(formula = abs ~ conc) ## ## Residuals: ## 1 2 3 4 5 6 ## -9.842e-04 1.363e-03-4.008e-04 6.801e-04-6.901e-04 3.164e-05 ## ## Coefficients: ## Estimate Std. Error t value Pr(> t ) ## (Intercept) 0.0003692 0.0007172 0.515 0.634 ## conc 0.0690523 0.0011844 58.302 5.18e-07 *** ## --- ## Signif. codes: 0 '***' 0.001 '**' 0.01 '*' 0.05 '.' 0.1 ' ' 1 ## ## Residual standard error: 0.0009909 on 4 degrees of freedom ## Multiple R-squared: 0.9988, Adjusted R-squared: 0.9985 ## F-statistic: 3399 on 1 and 4 DF, p-value: 5.183e-07 Question 2. Students should note that the intercept s value essentially is constant for all wavelengths with all values clustering near zero; they should note, as well, that the slope varies with wavelength and is steepest where the analyte has the strongest. Both observations, of course, are consistent with Beer s law, which predicts that the slope of the calibration curve is ɛb, and thus increases as the molar absorptivity increases, and that there is no in the absence of analyte. Question 3. The optimum analytical wavelength is 649 nm as this is where the analyte has its maximum ; any value within ±2 nm or ±3 nm is reasonable. Investigation 4 This investigation introduces students to the fundamental limitation of Beer s law. Using the two available sliders, students select the analytical wavelength and adjust the range of concentrations to include in the calibration curve. Highlighting in the calibration plot shows which standards are included in the calibration model. 5

Question 1. Students will need to consider what it means for a particle to be an independent absorber. A reasonable assumption is that a particle of analyte is independent if its behavior is not affected by the presence of other particles, and that this is more likely for lower concentrations of analyte. Question 2. When using 649 nm as the analytical wavelength, and when including all 14 standards, the calibration curve 0.0 0.5 1.0 1.5 0 2 4 6 8 10 is a poor fit to the data, showing evidence, in particular, of distinct curvature. When limiting the standards to those with concentrations less than or equal to 2.000 mm analyte, the calibration curve provides a good fit to the data. 0.0 0.5 1.0 1.5 0 2 4 6 8 10 Students may choose to limit the calibration curve to a maximum concentration of 1.000 mm or extend it to a maximum concentration of 4.000 mm; although these are not unreasonable choices, for standards not included in the calibration curve, consistent negative deviations from the calibration equation are evident only for concentrations of analyte greater than 2.000 mm. Question 3. Students should recall that to interpolate is to predict an analyte s concentration given an that falls within the values used to construct the calibration curve, and that to extrapolate is to predict an analyte s concentration given an that is greater than that for the most concentrated standard. The calibration curve from Question 1 should convince students that interpolation is not possible if the mathematical model a straight-line in this case does not correctly match the data; they should recognize from the calibration curve from Question 2 that extrapolating beyond the calibration data 6

will lead to a determinate error and that this error becomes larger for greater extrapolations. Investigation 5 In this investigation, students examine the effect on Beer s law of a non-monochromatic source, using sliders to select the analytical wavelength and the offset for the interfering wavelength. Question 1. Students should recall from Investigation 2 that transmittance is the ratio of P sam to P ref, where each term is a sum of contributions at each wavelength. Students should recognize that T = P samλ1 + P samλ2 P refλ1 + P refλ2 and, perhaps with additional guidance, they should be able to derive the following equation that relates the combined transmittance of light at both wavelengths to the source s power at each wavelength and the analyte s at each wavelength; note: it may help to nudge students in this direction. T = P samλ1 + P samλ2 P refλ1 + P refλ2 = T ( λ1p refλ1 + T λ2 P ) refλ2 A ɛbc = P λ1 refλ1 + ( A ɛbc) P λ2 refλ2 P refλ1 + P refλ2 P refλ1 + P refλ2 With this equation, it is easy to see that a negative deviation from Beer s law occurs when the the analyte absorbs less strongly at the interfering wavelength than at the analytical wavelength. Question 2. For an analytical wavelength of 649 nm, increasing the offset from ±5 nm to ±50 nm results in an increasingly more negative deviation from Beer s law. Students should note that the deviation from Beer s law is negligible for an offset of ±5 nm because there is little difference in at these two wavelengths. They also should note that the deviation is greater for an offset of +50 nm than an offset of -50 nm, as shown here 0.0 0.1 0.2 0.3 0.4 analytical wavelength only w/ 50 nm offset w/ +50 nm offset 0.0 0.5 1.0 1.5 2.0 because the peak is not symmetrical around the wavelength of maximum such that the of 0.371 at 699 nm is much smaller than the of 0.448 at 599 nm. Question 3. Students should discover two things as they explore this question. First, as the offset increases from its default value of ±0 nm, the negative offset yields a negative deviation from Beer s law and the positive offset yields a positive deviation from Beer s law; students should be able to explain this using the same logic as in Question 2. Second, for a positive offset of +45 nm and a positive offset of +50 nm, there is 7

little apparent deviation from Beer s law; students should recognize that there is no deviation from Beer s law if the at the analytical wavelength is identical to the at the interfering wavelength. Question 4. Students should recall that a scanning monochromator s slit width determines the range of wavelengths that pass through the sample, and that a narrower slit width will make the source radiation appear less polychromatic, resulting in a smaller deviation from Beer s law; they should recognize, as well, that a narrower slit width will decrease the total power of source radiation that reaches the detector, which increases the importance of noise. Investigation 6 This investigation allows students to explore how stray light affects the linearity of a Beer s law calibration curve using sliders to select the analytical wavelength and the amount of stray light, which is given as a percentage of P ref for the analytical wavelength. Question 1. Using the same general approach as in Investigation 5, but replacing P refλ2 and P samλ2 with P stray T = P sam + P stray P ref + P stray students should recognize that the effect of stray light is to increase P sam + P stray relatively more than P ref + P stray ; thus, T increases and A decreases, resulting in a negative deviation to Beer s law. Question 2. For all wavelengths and for all amounts of stray light, students should see that the effect of stray light on is greater for higher concentrations of analyte and greater when using an analytical wavelength where the analyte has a stronger. A typical result when using an analytical wavelength of 649 nm and 50% stray light is shown here 0.0 0.1 0.2 0.3 0.4 without stray light with stray light 0.0 0.5 1.0 1.5 2.0 Using the values from the tables, in the presence of 50% stray light the 2.000 mm standard s is 55.4% of its in the absence of stray light; for the 0.200 mm standard, the in the presence of 50% stray light is 65.7% of its in the absence of stray light. Question 3. Students should have little problem identifying as a source of stray light external radiation from the room that finds its way into the spectrometer s detector compartment; students may need some nudges to recognize that scattering of source radiation within the spectrometer might allow some of the scattered radiation to reach the detector without passing through the sample. The first source of stray light is minimized by better designing the spectrophotometer so that external radiation cannot enter the spectrophotometer; the second source of stray light is minimized by improving the spectrophotometer s optics. 8

Investigation 7 The final investigation allows students to explore the effect of an analyte s acid-base chemistry on its Beer s law calibration curve. Controls are provided to select the analyte s pk a value, to select the absorbing species, to allow for buffering, and, when using buffering, to set the buffer s ph. Question 1. Students should recognize that the negative deviation when HA is the sole absorbing species indicates that the concentration of HA is not the same as the analyte s reported total concentration. With additional thought and some hints, as needed students should recognize that because HA is a weak acid, some of it has dissociated into its weak base form, A. Question 2. Students should predict that the stronger HA is as an acid, the greater the deviation from Beer s law when HA is the absorbing species; thus, they should predict that the deviation is greater when HA has a pk a of 3 and smaller when HA has a pk a of 7. Question 3. Although both sets of conditions result in a negative deviation from Beer s law, the deviation is concave when HA is the absorbing species 0.0 0.1 0.2 0.3 0.4 no acid base chemistry w/ acid base chemistry: only HA absorbs 0.0 0.5 1.0 1.5 2.0 and convex when A is the absorbing species 0.0 0.1 0.2 0.3 0.4 no acid base chemistry w/ acid base chemistry: only A absorbs 0.0 0.5 1.0 1.5 2.0 Students can explain this behavior by writing the K a expression 9

K a = [ H + ] [ A ] and noting that as the concentration of HA increases, the ph decreases, the [ H +] increases, and the ratio [ A ] [HA] becomes smaller. As a result, the percentage of analyte present as HA increases as the total concentration of analyte increases. Question 4. For an analyte with a pk a of 5 and HA as the absorbing species, students should recognize that they need to buffer the ph such that HA is the only significant species; a ph of 3 or less will work. Question 5. For an analyte with a pk a of 7 and A as the absorbing species, students should recognize that they need to buffer the ph such that A is the only significant species; a ph of 9 or more will work. Question 6. To prepare the calibration standards, students should recognize that they need to dilute the stock standard using a solution of buffer with the appropriate ph. The buffer, of course, will need sufficient buffer capacity to hold the ph constant for all concentrations of analyte. To prepare samples for analysis, a suitable amount of sample is brought into solution using the same buffer such that the analyte s concentration falls within the calibration curve; again, the buffer capacity must be sufficient to ensure that the ph of the sample matches that of the standards. [HA] 10