INVESTIGATION OF IMPACT HAMMER CALIBRATIONS

Similar documents
MODEL-BASED ANALYSIS OF THE DYNAMIC BEHAVIOUR OF A 250 KN SHOCK FORCE CALIBRATION DEVICE

Analysis of shock force measurements for the model based dynamic calibration

Investigations for the model based dynamic calibration of force transducers by using shock excitation

Modelling of a dynamic torque calibration device and determination of model parameters

Characterization of Force Transducers for Dynamic Measurements

THE NEW 1.1 MN m TORQUE STANDARD MACHINE OF THE PTB BRAUNSCHWEIG/GERMANY

Model parameter identification from measurement data as a prerequisite for dynamic torque calibration Measurement results and validation

APPLICATION OF A SCANNING VIBROMETER FOR THE PERIODIC CALIBRATION OF FORCE TRANSDUCERS

FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).

Lecture 19. Measurement of Solid-Mechanical Quantities (Chapter 8) Measuring Strain Measuring Displacement Measuring Linear Velocity

Fundamentals of Low Intensity Shock Calibration

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

Lecture 20. Measuring Pressure and Temperature (Chapter 9) Measuring Pressure Measuring Temperature MECH 373. Instrumentation and Measurements

Non-linear Modal Behaviour in Cantilever Beam Structures

Chapter 2 Lab Exercises for a Course on Mechanical Vibrations

PROGRESS IN DEVELOPMENT OF CALIBRATION SYSTEMS FOR ANGULAR VIBRATION PICKUPS

Review of modal testing

COMPLETION AND MEASUREMENT UNCERTAINTY BUDGET OF THE MULTI-COMPONENT MEASURING DEVISE FOR FORCE UP TO 1 MN AND TORQUE UP TO 2 KN M

Analysis of the Temperature Influence on a Shift of Natural Frequencies of Washing Machine Pulley

DEVELOPMENT OF DROP WEIGHT IMPACT TEST MACHINE

Force/Torque measuring facility for friction coefficient and multicomponent

Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them

Measurement Techniques for Engineers. Motion and Vibration Measurement

PLEASURE VESSEL VIBRATION AND NOISE FINITE ELEMENT ANALYSIS

UNCERTAINTY IN TORQUE CALIBRATION USING VERTICAL TORQUE AXIS ARRANGEMENT AND SYMMETRICAL TWO FORCE MEASUREMENT

OPTI 521, Optomechanical Design, Technical Paper Reviews, Dr. Jim Burge, 2011

T e c h n i c a l L e x i c o n

C. points X and Y only. D. points O, X and Y only. (Total 1 mark)

CHAPTER 4 DESIGN AND ANALYSIS OF CANTILEVER BEAM ELECTROSTATIC ACTUATORS

Dynamic Behaviour of the Rubber Isolator Under Heavy Static Loads in Aerospace Systems

EXPERIMENTAL EVALUATION OF THE MODAL DAMPING OF AUTOMOTIVE COMPONENTS WITH DIFFERENT BOUNDARY CONDITIONS

Coupling System for Ultra Precision Machining

Estimation of Rotational FRFs via Cancellation Methods

Committee Draft No. 99 To be combined with T-150 as a method B. Determination of Natural Frequency and Flexural Modulus by Experimental Modal Analysis

UNCERTAINTY EVALUATION OF SINUSOIDAL FORCE MEASUREMENT

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Name: Fall 2014 CLOSED BOOK

Transactions on the Built Environment vol 22, 1996 WIT Press, ISSN

(Refer Slide Time: 1: 19)

Investigation of Piezoelectric Force Measuring Devices in Force Calibration and Force. Standard Machines

Francisco Paulo Lépore Neto. Marcelo Braga dos Santos. Introduction 1. Nomenclature. Experimental Apparatus and Formulation

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

VMS-GeoMil. Background

Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

Applied Modal Analysis of Wind Turbine Blades

T20WN. Data Sheet. Torque transducers. Special features. Installation example with bellows couplings. B en

Structural Dynamic Behavior of a High-Speed Milling Machine

MULTI-STAGE SUBORBITAL LAUNCHER MODAL AND DYNAMIC TEST PROGRAM

D : SOLID MECHANICS. Q. 1 Q. 9 carry one mark each. Q.1 Find the force (in kn) in the member BH of the truss shown.

Modal Analysis Technique for Anisotropic Composite Laminates

Structural Dynamics Model Calibration and Validation of a Rectangular Steel Plate Structure

INVESTIGATION OF TRANSFER STANDARDS IN THE HIGHEST RANGE UP TO 50 MN WITHIN EMRP PROJECT SIB 63 Falk Tegtmeier 1, Michael Wagner 2, Rolf Kumme 3

Experimental Modal Analysis of a Flat Plate Subjected To Vibration

XIX IMEKO World Congress Fundamental and Applied Metrology September 6 11, 2009, Lisbon, Portugal

Francisco Paulo Lépore Neto. Marcelo Braga dos Santos. Introduction 1. Nomenclature. Experimental Apparatus and Formulation

Chapter 7 Vibration Measurement and Applications

Physics 2001/2051 The Compound Pendulum Experiment 4 and Helical Springs

HOMOGENEOUS ELECTRORHEOLOGICAL FLUIDS APPLIED TO VIBRATION CONTROL

COUPLED USE OF FEA AND EMA FOR THE INVESTIGATION OF DYNAMIC BEHAVIOUR OF AN INJECTION PUMP

DEVELOPMENT OF MEASUREMENT STANDARD FOR DYNAMIC PRESSURE AT MIKES

Model tests and FE-modelling of dynamic soil-structure interaction

Investigations to determine the dynamic stiffness of elastic insulating materials

DROP-WEIGHT SYSTEM FOR DYNAMIC PRESSURE CALIBRATION

DEVELOPMENT OF A NOVEL ACTIVE ISOLATION CONCEPT 1

Integration of measured receptance into a time domain simulation of a Multi Body Model using SIMPACK

NV-TECH-Design: Scalable Automatic Modal Hammer (SAM) for structural dynamics testing

Identification of damage in a beam structure by using mode shape curvature squares

CHAPTER 6 FRICTION AND WEAR ANALYSIS FOR BUSHING

EXPERIMENTAL MODAL ANALYSIS OF A SCALED CAR BODY FOR METRO VEHICLES

Experimental and Numerical Modal Analysis of a Compressor Mounting Bracket

440. Simulation and implementation of a piezoelectric sensor for harmonic in-situ strain monitoring

FINAL REPORT OF THE BILATERAL COMPARISON OF THE CALIBRATIONS OF STANDARD WEIGHTS BETWEEN CENAM-MEXICO AND INEN-ECUADOR SIM.M.M-S4 (SIM.7.

Dynamic characterization of multi-component sensors for force and moment

Estimation of Unsteady Loading for Sting Mounted Wind Tunnel Models

C7047. PART A Answer all questions, each carries 5 marks.

Compensation-based non-destructive automatic quality control using acoustic methods

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

RESILIENT INFRASTRUCTURE June 1 4, 2016

Strain Measurement. Prof. Yu Qiao. Department of Structural Engineering, UCSD. Strain Measurement

Characterizing, Simulating, and Eliminating Vibration Induced Counts in Measurement While Drilling Gamma Ray Detectors. authors

Introduction to Continuous Systems. Continuous Systems. Strings, Torsional Rods and Beams.

ASSESMENT OF THE EFFECT OF BOUNDARY CONDITIONS ON CYLINDRICAL SHELL MODAL RESPONSES

MV Module 5 Solution. Module 5

The use of transmissibility properties to estimate FRFs on modified structures

Lecture #2: Split Hopkinson Bar Systems

Grandstand Terraces. Experimental and Computational Modal Analysis. John N Karadelis

Engineering Science OUTCOME 2 - TUTORIAL 3 FREE VIBRATIONS

A Repeated Dynamic Impact Analysis for 7x7 Spacer Grids by using ABAQUS/ Standard and Explicit

EVALUATION OF STATIC AND DYNAMIC PARASSITIC COMPONENTS ON THE INRIM 1 MN PRIMARY FORCE STANDARD MACHINE BY MEANS THE 500 kn SIX-COMPONENT DYNAMOMETER

Random Eigenvalue Problems in Structural Dynamics: An Experimental Investigation

Mechanical Shock Testing for LIGA Materials Characterization

STATICS Chapter 1 Introductory Concepts

NONLINEAR CHARACTERISTICS OF THE PILE-SOIL SYSTEM UNDER VERTICAL VIBRATION

CONTRIBUTION TO THE IDENTIFICATION OF THE DYNAMIC BEHAVIOUR OF FLOATING HARBOUR SYSTEMS USING FREQUENCY DOMAIN DECOMPOSITION

10 Measurement of Acceleration, Vibration and Shock Transducers

Varuvan Vadivelan. Institute of Technology LAB MANUAL. : 2013 : B.E. MECHANICAL ENGINEERING : III Year / V Semester. Regulation Branch Year & Semester

ACCELEROMETER BASED IN SITU RUNOUT MEASUREMENT OF ROTORS

Chapter 23: Principles of Passive Vibration Control: Design of absorber

Preliminary Examination - Dynamics

Transcription:

IMEKO 23 rd TC3, 13 th TC5 and 4 th TC22 International Conference 30 May to 1 June, 2017, Helsinki, Finland INVESTIGATION OF IMPACT HAMMER CALIBRATIONS M. Kobusch 1, L. Klaus 1, and L. Muñiz Mendoza 2 1 Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116 Braunschweig, Germany, michael.kobusch@ptb.de 2 Centro Nacional de Metrología (CENAM), km 4.5 Carretera a Los Cués, Municipio El Marques, Querétaro, C.P. 76246, México Abstract: This paper presents recent theoretical and experimental investigations of calibration measurements of impact hammers using different dynamic calibration set-ups with primary traceability methods. In the scope of a scientific collaboration between the German national metrology institute PTB and the Mexican CENAM, measurements of selected hammers were conducted at both institutes. The analysis of the measurement signals in the time and frequency domain is presented and discussed. Keywords: Modal hammer, dynamic calibration, shock force, sensitivity. 1. INTRODUCTION Impact hammers (or modal hammers) are a widely used handheld tool in science and industry for the modal testing (or experimental modal analysis) of mechanical components and interconnected mechanical structures. The instrumented hammers feature an integrated force transducer to measure the dynamic force stimulus that excites the mechanical structure under test. The main purpose of these experimental tests is the determination of the test structure s impulse responses, mode shapes, resonances, and damping by means of additional acceleration sensors which pick up the shock response at specific points of interest [1]. The appropriate calibration of impact hammers has gained importance in order to satisfy measurement accuracy demands and the needs of the quality management employed. Considering the related mechanical transducers for acceleration and force, it is well known that these transducers show a frequency-dependent response which basically depends on their mechanical design and the mechanical coupling to the environment. Consequently, the establishment of traceable calibrations of impact hammers has become a topic of recent research activities at the national metrology institute (NMI) level. In the scope of a scientific collaboration between the German national metrology institute PTB and the Mexican CENAM, methods for the primary calibration of impact hammers have been investigated at both institutes. This paper presents dynamic calibration measurements of a selected impact hammer specimen conducted at the experimental calibration set-ups of both institutes. A hammer stroke generates an approximately half-sine-shaped force pulse of short duration. The dynamic output signal of the impact hammer is then compared to a measured reference force, which is the inertia force of a known reaction mass body at which the hammer impacts. The experimental challenges in measurement and data analysis, considering the time and the frequency domain, are introduced and discussed in the following. 2. MEASUREMENT PRINCIPLE The current research activities focus on dynamic calibrations following primary methods for the traceable measurement of the hammer force pulse. The reference force is determined as the product of the mass and acceleration of the reaction mass body, with the mass value obtained by weighing, and the time-dependent acceleration determined by an acceleration measurement. Figure 1 illustrates the principle of the dynamic impact hammer calibration. Figure 1. Model of the impact hammer calibration. The hammer strikes the reaction mass m, and its timedependent course of acceleration a(t) is picked up at the opposing body face by an attached acceleration sensor (marked in green) or by a contactless beam of a laser vibrometer (red). Both dynamic signals, the hammer s force

transducer output U H(t) and the acceleration a(t), are recorded with a sufficiently high sampling rate. The figure further illustrates the basic model to describe the dynamic behaviour of the calibration set-up. The mechanical structure is modelled by a lumped mass model, i.e. a series arrangement of spring-coupled rigid mass bodies. In accordance with the dynamic modelling of force transducers [2], the hammer s integrated force transducer is described by two parts (masses m 1, m 2) coupled by an elastic spring (stiffness k) and connected to the hammer tip and the hammer mass, respectively. The hammer signal U H(t) is assumed to be proportional to the spring compression. The magnitude and duration of the force pulses can be varied by applying differently strong hammer strokes and by using hammer tips of different hardness. Figure 2 shows typical hammer signals in the time and frequency domain which were obtained with different hammer tips. For a given impact magnitude of the applied hammer stroke, i.e. the momentum defined as the product of the hammer mass and impact velocity, soft hammer tips generate long pulses of small force amplitude, and hard tips short pulses of large amplitude. In the frequency domain, long pulses correspond to lower frequency contents, and short pulses to higher frequency contents. arrangements. Considering a pendulum suspended at both ends with one common pivot point, the mass performs a rotational motion which results in a radial distribution of acceleration at the front surface. In the case of a string geometry with two pivot points, the pendulum shows a parallelogram-type motion with an even acceleration over its front surface. This geometry however exhibits a transversal motion component. In order to guarantee a linear motion without parasitic rotational and transversal components, a linear air bearing is further investigated. This linear motion is advantageous compared to the pendulum designs, but the costs of purchase are considerably higher and will thus limit its versatility, as modifications of the set-up are barely possible. (a) Figure 3. Calibration set-ups using a pendulum (top) and an airborne (bottom) reaction mass body. (b) Figure 2. Hammer signals measured with five different hammer tips in the time (a) and frequency (b) domain. The calibration set-ups realized at PTB and CENAM use reaction mass bodies (various mass values available) which are either suspended by strings (pendulum arrangement) or guided by an air bearing to assure a defined axial motion with low friction. Figure 3 visualizes the principal motion of the hammer-excited reaction mass body for three different Regarding the repeatability in measurement, it has to be noted that an impact hammer is a handheld instrument, and the user-dependent test conditions may affect the calibration results. The variability of manually guided hammer strokes includes the applied momentum, the curved hammer motion, the location and direction of the impact force, as well as the mechanical interaction with the operator s arm. Although these influences might be negligible, the great variability is a challenge for well-defined calibration conditions. Impact hammers are commonly equipped with a piezoelectric force transducer of high stiffness, and therefore, of high bandwidth. Dynamic effects due to inertia forces of the hammer s mechanical structure are assumed to be small. However, the shock pulses might be very short and dynamic effects due to inertia forces of the particular calibration setup, i.e. from the reaction mass, the hammer tips, or the

hammer mass, should not be generally ruled out. Whereas these effects may be correctly modelled by the dynamic model sketched in Figure 1, parasitic vibrations from shockexcited modal vibrations, which could be observed at the shock force calibration devices [2], will not be described. In general, it is necessary to know and compensate for the dynamic behaviour of both signal measuring chains in the frequency range of interest. This consideration includes the measuring amplifiers and signal conditioners [3], the data acquisition electronics as well as the data processing software. Whereas dedicated charge amplifiers can be calibrated prior to the calibration measurements, an appropriate compensation is not possible in the case of force or acceleration sensors with integrated electronics, the socalled IEPE (Integrated Electronics Piezoelectric) sensors. 3. SIGNAL ANALYSIS The impact hammer calibration measurements are based on the analysis of two pulse-shaped signals, the hammer output U H(t) and the acceleration a(t) of the reaction mass m which gives the reference force F(t) = m a(t). The signal ratio U H / F then leads to the sought calibration result, the shock sensitivity S sh. Figure 4 illustrates the two principal methods of the signal analysis in the time and frequency domain. peak fitting procedure. In the case of a high-pass characteristic of the measurement instrumentation, which is always the case for a piezoelectric sensor, ground level offsets may be corrected by a trapezoidal correction if necessary [4]. For a set of calibration measurements with varying pulse heights, the shock sensitivity can be calculated as the averaged pulse peak ratio of each individual hammer pulse, or can be determined by a linear regression of the hammer and acceleration peak values (cf. Figure 9). Modal testing with impact hammer excitations usually analyses spectral measurement data. The spectral range of the force excitation is an important criterion (cf. Figure 2). In this regard, it is useful to directly specify the sensitivity of the hammer output from frequency domain data (Figure 4b). Here, the shock sensitivity can be expressed as the ratio of the spectral amplitude values at a specific frequency, e.g. at 100 Hz. 4. EXPERIMENTAL MEASUREMENTS Within the scope of a scientific collaboration between the German and the Mexican NMIs, calibration measurements of impact hammers have been conducted at both institutes to investigate the procedures and methods and to obtain a first comparison of calibration results. The different experimental dynamic calibration set-ups are shown in Figure 5. (a) (b) Figure 4. Signal analysis in the time domain (a) and frequency domain (b). Considering a signal analysis in the time domain (Figure 4a), shock calibration results are typically based on the evaluation of pulse peak ratios. To suppress affecting signal noise, the pulse peak values may be determined by a Figure 5. Impact hammer calibration set-ups applying a pendulum mass at CENAM (top), at PTB (centre), and an airborne mass body at PTB (bottom).

At CENAM, the impact hammer calibration set-up used two differently sized pendulum masses made of stainless steel and an accelerometer (Endevco 2270) which was axially mounted via a threaded stud bolt. The total mass of these reaction masses amounted to 0.77 kg and 1.47 kg, respectively. The measurement instrumentation further consisted of a charge amplifier (Brüel & Kjær 2635) and a data acquisition and IEPE conditioning system (National Instruments PXI-4461). At PTB, impact hammer calibration measurements were conducted with a pendulum mass of 1.0 kg and an airborne mass of 8.2 kg. In both cases, the acceleration of the reaction mass was obtained by a laser vibrometer (Polytec OFV 503) and an acceleration sensor (PCB 352B10) glued at the front surface. The instrumentation further consisted of a charge amplifier (Kistler type 4011), an IEPE conditioner (PCB Piezotronics 482C series), and a data acquisition system (National Instruments PXI-5124, PXI-5922). The experimental tests and results given in the following are only to be considered as examples to adequately describe the investigated methods and procedures of an impact hammer calibration. All the measurements presented in this paper were conducted with a Kistler 9726A20000 impact hammer. The selected hammer applies an IEPE sensor of 20 kn force range, features a hammer head of 500 g, and can be equipped with five interchangeable hammer tips of different hardness. Figure 6 shows the pulse parameters measured with a Kistler 9726A20000 impact hammer using five differently stiff hammer tips for a set of 26 individual measurements each. The experimental tests showed that the pulse width slightly decreases for stronger hammer strokes, which results from a nonlinear contact stiffness of the hammer tip. A dependence on the operator was not observed. A corresponding spectral analysis of these 130 pulses is given in Figure 7, showing the averaged amplitudes of the shock sensitivity for each of the five hammer tips. At low frequencies, the sensitivities obtained with the two hardest tips (steel, plastic) obviously differ from those with the three softest ones (rubber), which clearly indicates that dynamic effects from the inertia force of the hammer tip might be seen. The directly screwed hardest tips have a mass of about 2 g, and the softer tips use an adapter which results in a much higher mass of about 37 g. Simultaneous measurements with an acceleration sensor and a laser vibrometer demonstrate that the motion should be preferably measured at the surface centre to minimize influences from modal vibrations (bending modes) and, in particular, from low-frequent oscillations of the reaction mass, which will be excited by oblique or non-centric hammer strokes. In this context, a low-frequent ringing was commonly observed for soft hammer tips, i.e. long shock pulses, and for a mass suspension (or guidance) of comparably higher lateral stiffness, e.g. at the airborne 8 kg reaction mass. Furthermore, the calibration measurements obtained with the hammer tips made of rubber demonstrate that the impacting surfaces experience some kind of sticking effect, as the rebounding impact hammer measures a small tension force when the shock pulse ends. Figure 6. Pulse parameters achieved with a Kistler 9726A20000 impact hammer for five hammer tips of different materials, 26 pulses each, measured at CENAM using a pendulum mass of 1.47 kg. Figure 7. Spectral sensitivity for five hammer tips, average of 26 pulses each, measured at CENAM. In general, short pulses obtained with the hard hammer tips significantly excite modal resonances of the reaction mass, the hammer, and the acceleration sensor. This noise has to be appropriately low-pass filtered for an analysis in the time domain. Figure 8 gives an example demonstrating the importance of this filtering and shows the two acceleration signals measured at the airborne mass body with an impact hammer excitation using the steel tip. The signals filtered by a low-pass filter of 100 khz cut-off frequency are dominated by modal oscillations. A 15 khz low pass (Bessel, 8 th order) was needed to suppress the strongest noise components in order to obtain a smooth pulse. This signal still exhibits a small ringing of about 18 khz, which belongs to the first longitudinal resonance of the mass body. A first comparison of the results of the calibration measurements conducted at PTB and CENAM is presented in Figure 9. The analysis of the shock pulse data of the impact hammer selected as an example was performed in the time domain. The diagram shows the distribution of the measured pulse peak values of the reference force and the hammer output for hammer strokes obtained with the steel tip, about 50 data points of each institute. A linear

regression finally gives the averaged pulse peak ratio, i.e. the shock sensitivity to be determined. The calibration results of 0.222 mv/n (PTB) and 0.219 mv/n (CENAM) deviate by only 1 % and agree with the specified nominal sensitivity of 1 mv/lbf (0.225 mv/n). Figure 8. Filtering of the noisy acceleration signals using different low-pass (LP) filters, impact hammer with steel tip, measured at PTB s airborne reaction mass of 8 kg. 5. DISCUSSION AND CONCLUSIONS This paper presents investigations of impact hammer calibration measurements conducted within the scope of a scientific collaboration between PTB and CENAM. It was found that the different set-ups may give consistent results. The calibration results obtained at both institutes could be compared for the first time. For the impact hammer tested as an example, the measured shock sensitivities deviate in the order of 1 %. However, some parasitic influences from the mechanical set-up or the experimental test conditions could be observed, which may affect the calibration results. Hammer tips of different mass and stiffness gave slightly different shock sensitivities, probably due to dynamic effects within the hammer head. In addition, considerably strong disturbances resulted from modal vibrations of the shock-excited reaction mass body and the accelerometer. In particular for short pulses of high frequency content, the appropriate filtering or minimization of parasitic modal vibrations is very important for the pulse analysis in the time domain. The resonances of the reaction mass body on which the impact hammer strikes should be as high as possible, in order to achieve a broad calibration bandwidth. For this purpose, the modal vibrations of a cylindrical reaction mass were investigated by a finite element (FE) modal analysis. Considering a steel body of 1 kg mass and 60 mm diameter, its two lowest longitudinal modes are found at about 45 khz (Figure 10). These modes represent the axisymmetric in-phase (left) and out-of-phase vibration of the cylinder end faces, respectively. Definitely excited by the hammer strokes, the longitudinal vibration modes will set the limits to the useable bandwidth of the acceleration measurement that provides the reference force signal. In contrast, parasitic influences from other modal vibrations, e.g. non-axisymmetric bending modes, which may even appear at lower frequencies, are suspected to be much smaller and may be neglected if the acceleration is sensed at the body axis. To find the best geometrical shape of the cylindrical body, the influence of the aspect ratio (ratio of length to diameter) is shown in Figure 11. The diagram visualizes the modal frequencies of the first two longitudinal modes as well as of the lowest resonance, which is either plate bending, torsion or rod bending. The highest modal frequencies can be expected for an aspect ratio of about 0.8. Figure 9. Comparison of the pulse peak values measured at PTB and CENAM: Kistler 9726A20000 impact hammer with steel tip; the gradient determined by a linear regression represents the shock sensitivity of the hammer. Figure 10. FE calculation of the first two longitudinal modes at 44.2 khz (left) and 46.2 khz of a 1 kg cylinder made of steel with a diameter of 60 mm.

Figure 11. FE calculation of the first modal frequencies of a 1 kg cylinder made of steel in dependency of its aspect ratio length to diameter. The previous FE calculations describe the resonances of a cylinder made of steel. The influence of the material can be estimated from the analytical solutions of the longitudinal vibrations of an elastic rod. Considering a cylindrical body of constant mass and aspect ratio, its resonance frequencies f e depend on the material s density and the modulus of elasticity E the Poisson s ratio is assumed to be fixed, and it holds the proportionality f e E 1/2-1/6. To maximize the resonance frequencies of a given mass, the modulus of elasticity should be as high as possible. Compared with the material stainless steel (density about 7.9 g/cm 3 ), the resonances of aluminium (2.7 g/cm 3 ) and brass (8.4 g/cm 3 ) are estimated to be 30 % lower, tungsten alloy (18 g/cm 3 ) is almost 20 % higher, and the non-oxide ceramics silicon nitride (3.2 g/cm 3 ) is even 70 % higher. Steel seems to be a good choice for a metallic material, however, the exotic materials tungsten alloy and silicon nitride may be considered for special needs. In future, the research will continue to investigate and optimize the presented experimental set-ups for the dynamic calibration of impact hammers as well as the mathematical procedures for data analysis and uncertainty estimation. This work will have to consider the calibration methods based on the time domain and the frequency domain analysis as well. 6. REFERENCES [1] D.J. Ewins, Modal testing; theory, practice and application, 2 nd ed., Research Studies Press, 2000. [2] M. Kobusch, S. Eichstädt, L. Klaus and T. Bruns, Investigations for the model-based dynamic calibration of force transducers by using shock excitation, ACTA IMEKO, 4 (2), pp. 45 51, 2015. DOI: 10.21014/acta_imeko.v4i2.214 [3] L. Klaus, T. Bruns, H. Volkers, Calibration of bridge-, charge- and voltage amplifiers for dynamic measurement applications, Metrologia, vol. 52 (1), pp. 72-81, 2015. DOI: 10.1088/0026-1394/52/1/72 [4] International Standard ISO 16063-13:2001 Primary shock calibration using laser interferometry.