The Role of the Carbon Cartridge in the One-Pass System

Similar documents
Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

DUAL CARTRIDGE SYSTEM FOR THE EXTRACTION OF ACIDS, BASES, AND NEUTRALS IN WATER

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

Semivolatile Organic Reference Materials

Appendix 2 Data Distributions for Contaminants in Water and Sediment Samples in Response to the Deepwater Horizon Oil Spill, 2010

2. September 2015 Sub-Concrete Slab Soil Investigation

Certificate of Accreditation

Amec Foster Wheeler Environment & Infrastructure, Inc BIG SHANTY ROAD, NW, SUITE 100 KENNESAW, GEORGIA (770)

POTENTIAL VAPOR INTRUSION ENVIRONMENTAL

Non-Potable Water A2LA Certificate

Drying Solvent Extracts using a Membrane

Certificate of Accreditation

October 26, Dear Mr. White:

Certificate of Accreditation

Certificate of Accreditation

Methods and Analytes Requiring Laboratory Certification Attachment 3

317 Elm Street Milford, NH (603) Fax (603)

Table 1 Soil and Groundwater Sampling Summary Table

STANDARD OPERATING PROCEDURES SOP: 1821 PAGE: 1 of 33 REV: 0.0 DATE: 11/25/94

Table of Contents ~ 1 ~

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Certificate of Accreditation

STANDARD OPERATING PROCEDURES

March 6, Mr. John Meyer DRI Coordinator Tampa Bay Regional Planning Council 4000 Gateway Center Blvd. Pinellas Park, FL 33782

Semivolatile Organic Compounds By Gas Chromatography / Mass Spectrometry (GC/MS)

STATE OF ILLINOIS ENVIRONMENTAL PROTECTION AGENCY NELAP - RECOGNIZED ENVIRONMENTAL LABORATORY ACCREDITATION. is hereby granted to

METHOD 8410 GAS CHROMATOGRAPHY/FOURIER TRANSFORM INFRARED (GC/FT-IR) SPECTROMETRY FOR SEMIVOLATILE ORGANICS: CAPILLARY COLUMN

What Does the EPA Method Round Robin Data Really Tell Us?

SUBSURFACE (PHASE II) INVESTIGATION REPORT

Page 1 of 7 ADEQ Water Div NPDES 9/18/2017 8:52 AM

ANALYTICAL REPORT. Attn: Steve Johnson

Presentation Title. Isotope Dilution: Where it doesn t work Where we should expand its use. August 10, David I. Thal Principal Chemist

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

UCT ENVIRO-CLEAN HL DVB

Appendix C. Validation Standard Operating Procedures

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

MULTIRESIDUE ANALYSIS OF PRIORITY POLLUTANTS IN SURFACE WATERS USING EXACT MASS GC-TOF MS

PHASE II ENVIRONMENTAL SITE ASSESSMENT RICKENBACKER MARINA PARCELS 3301 RICKENBACKER CAUSEWAY MIAMI-DADE COUNTY, FLORIDA

APPENDIX D Laboratory Analytical Reports

Toxics Release Inventory (TRI) Basis of OSHA Carcinogens

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

METHOD 8270D SEMIVOLATILE ORGANIC COMPOUNDS BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS)

Pursuant to Chapter 4 of the Code of Ethics, ETAD members may not manufacture or sell dyes referred to in Annexes A and B. Annex A

Analysis of Organic Pollutants by Micro Scale Liquid-Liquid Extraction. and On-column Large Volume Injection Gas Chromatography. Mark S.

LABORATORY DATA CONSULTANTS, INC Loker Ave. West, Suite 220, Carlsbad, CA Bus: Fax:

ANALYTICAL REPORT. Attn: Steve Johnson

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

~WJuliJ LABORATORY DATA CONSULTANTS, INC.

Spakes, Rita. Rec d Digitally S1-R3 S W M D BS>FILE <RDS. TestAmerica Denver THE LEADER IN ENVIRONMENTAL TESTING AFIN: PMT#: DOC ID#:

The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane

Reviewing Material Safety Data Sheets to Verify Significant Drinking Water Threats

Sample Prep Solutions for Environmental Contaminants

TECHNICAL MEMORANDUM. SUBJECT: Round 1 Sediment Investigation and Plan for Round 2 Investigation Sparrows Point Phase I Area

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

A Review of Solid Phase Sorbents Over Time As Applied to Environmental Applications. Page 1 Biotage

Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood Using GC/MS

Determination of 24 PAHs in Drinking Water

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Selection of a Capillary

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

TETRA TECH, INC. December 8, Lynn Nakashima Berkeley Regional Office 700 Heinz Avenue, Suite 200C Berkeley, California 94710

SPE Introduction. general chromatography. solid phase extraction. How to Choose an SPE Product. Solid-Phase Processing Methods

SPE Basics: Sorbents, Sorption, and Multiclass Analytes. Martha J.M. Wells, Ph.D. EnviroChem Services

The Suite for Environmental GC Analysis

RE: 2005 Annual Report: AR (Boyd Point Facility)

Selection of a Capillary GC Column

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

Polycyclic Aromatic Hydrocarbons in Water by GC/MS PBM

Appendix J NPRI Substances

Hypercrosslinked Gel-type Polystyrene Resin. Applications Guide

San Francisco Bay Ambient Water Monitoring Interim Report

Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas

Laboratory Report Number(s) Lone Tree Road (Deep) 10/5/15, 10/19/15, 6/13/16. Address Sample Date(s) Analyses Requested

Application of GC/HRToF for the identification and comparison of tobacco smoke constituents

A Single-Method Approach for the Analysis of Volatile and Semivolatile Organic Compounds in Air Using Thermal Desorption Coupled with GC MS

STANDARD OPERATING PROCEDURES

Guidance on the Major Source Determination for Certain Hazardous Air Pollutants

N_HW1 N_HW1. 1. What is the purpose of the H 2 O in this sequence?

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

CERTIFICATE OF ANALYSIS

Appendix 1: Polycyclic Aromatic Compounds: Nomenclature and Analysis

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes

EL608. Deodorants [EL /1/ ]

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

General considerations Dye manufacturers achieve an effective impurity profile by good manufacturing practices such as: - implementation of a suitable

Hints for Strong Ion Exchange Resins

Chromatography. Gas Chromatography

APPLICATIONS TN Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and HPLC

HICHROM. Chromatography Columns and Supplies. LC COLUMNS Vydac. Catalogue 9. Hichrom Limited

EPA Region III BTAG Freshwater Sediment Screening Benchmarks 8/2006. Bioaccumulative o

Method 8270C PAH. FD- Field Duplicate Samples

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

TABLE OF CONTENTS. LIST OF FIGURES 1. Site Location Map 2. Assessment Map. LIST OF APPENDICES I. Laboratory Analytical Report

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

BRCC CHM 102 Class Notes Chapter 13 Page 1 of 6

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Hypersil BDS Columns TG 01-05

Transcription:

The Role of the Carbon Cartridge in the One-Pass System William Jones and Zoe Grosser, Horizon Technology, Inc. Key Words Solid phase extraction, spe, one-pass system, carbon cartridges Introduction Horizon Technology is a leader in sample preparation using solid phase extraction (SPE) for semi/nonvolatile extraction. One of the recent introductions is a system to extract a full range of acid/base neutral compounds from various types of water samples using a One-Pass system. In this case, rather than extract water at ph 2, adjust the ph of the water to 12 and re-extract, the water is passed once through a multi-media SPE disk and then a carbon cartridge before being delivered to waste. The elegance of this approach is the elimination of the basification step, which requires additional labor adjusting the ph and time passing the sample through the disk a second time after the first elution of acids and neutrals. The additional benefit of omitting the basification step eliminates the formation of insoluble metal hydroxides that can restrict the flow of sample through the disk or add significant time if a liquid-liquid extraction is used. Figure 1. Schematic of the One-Pass system loading of sample (1) and elution of the disk (1,2) and cartridge (3)

P a ge 2 A schematic of the system is shown in Figure 1. The solubility range of compounds for a wide suite of semivolatile analytes, such as seen in US EPA method 625 or 8270 is amazingly disparate and includes compounds from infinitely soluble methyl methanesulfonate to benzo(k)fluoranthene @ 8.00x10-7 g/l. 1,2 Liquid-liquid extraction (LLE) cannot be expected to efficiently extract very water soluble compounds into a non-polar solvent like methylene chloride. Recovery of highly water soluble compounds is very dependent on the matrix, such as the ionic strength of the sample matrix. In SPE there is no single sorbent that can handle this range of solubility or charged state of the analytes, so a multimodal media disk (Hydrophilic/Lipophilic groups along with cation exchange sites on the resin) is used for the majority of compounds. Compounds that are unretained on the disk are captured by the Max Detect Carbon Cartridge. These compounds typically have water solubility of ³ 10g/L or may be anionic in nature, e.g. N-Nitroso compounds like N- Nitrosodimethylamine (NDMA) and short chain alkyl-sulfonates. The One-Pass disk is resin-based offering several benefits over silica-based sorbents like C18. The sorbent requires no activation with methanol and is stable with samples at extreme acidic and basic conditions. The resin has several times more reversed-phase retention capacity over silica as the resin backbone is the sorbent whereas the bonding capacity of C18 is only found on the surface. Silica C18 loses C18 through hydrolysis at ph <2 and the silica begins to dissolves at a ph above 7. The Max Detect Carbon cartridge has been designed to be complementary to the One-Pass disk as it can capture analytes passing through as fast as 100 ml/min. The cartridge contains activated carbon, not non-porous graphitized carbon, making it especially suitable for this purpose. In addition the cartridge captures compounds through a backflush process to minimize the solvent used in elution. In some cases the carbon cartridge is not needed in the One-Pass system when the compounds retained on the carbon cartridge are not of interest. Table 1 shows a table of selected compounds showing the distribution in their elution between the acid fraction, ion exchange fraction and the carbon fraction. The utility of the carbon cartridge can also be seen in Figures 2 and 3. Figure 2 shows a full suite of compounds extracting using liquid-liquid (blue) compared to the One-Pass process (red). The compounds are named in Appendix 1, but include a full suite of semivolatile compounds. It is easily seen in the compound by compound comparison that the SPE One-Pass procedure works very well and in many cases gives a more accurate recovery. Figure 3 shows the same set of compounds extracted with the One-Pass system and the compound bar graph colored to indicate which fraction contained the analyte of interest. You can readily see that each component contributes to the compound recovery and makes the system effective.

P a ge 3 Table 1. Selected Compound Elution Percentage in Acid/Neutral, Cation Exchange, or Carbon Carbridge Fraction Analyte Number ACID/Neutral CATION EXCHANGE CARBON Target Compounds CAS NO # N-Nitrosodimethylamine (NDMA) 62-75-9 1 0 0 100 Pyridine 110-86-1 2 0 100 0 2-Picoline (2-Methylpyridine) 109-06-8 3 0 100 0 N-Nitrosomethylethylamine 10595-95- 4 6 5 0 95 Methyl methanesulfonate 66-27-3 5 2 5 93 N-Nitrosodiethylamine 55-18-5 7 20 0 80 Ethyl methanesulfonate 62-50-0 8 8 0 92 Phenol-d6 (surr) 13127-88- 9 3 78 0 22 Phenol 108-95-2 10 80 0 20 Aniline 62-53-3 11 0 100 0 Benzyl alcohol 100-51-6 18 50 0 50 N-Nitroso-pyrrolidine 930-55-2 21 7 0 93 N-Nitroso-morpholine 59-89-2 22 16 0 84 o-toluidine 95-53-4 27 0 100 0 N-Nitrosopiperidine 100-75-4 30 42 0 58 4-Chloroaniline 106-47-8 40 2 98 0 3-Nitroaniline 99-09-2 61 40 60 0 2-Naphthylamine 91-59-8 69 0 100 0 1-Naphthylamine 134-32-7 70 0 100 0 5-Nitro-o-toluidine 99-55-8 75 72 28 0 Azobenzene 103-33-3 78 97 3 0 4-Aminobiphenyl 92-67-1 86 3 97 0 Methapyrilene 91-80-5 93 0 100 0 Benzidine 92-87-5 95 0 100 0 p-dimethylaminoazobenzene 60-11-7 98 2 98 0 3,3'-Dimethylbenzidine 119-93-7 99 0 100 0 3,3'-Dichlorobenzidine 91-94-1 102 0 100 0

P a ge 4 Figure 2. Full suite of analytes from US EPA Method 8270 recovered from spiked reagent using Liquid-Liquid extraction (blue) compared to the One-Pass system (red) Figure 3. Full suite of analytes from US EPA Method 8270 recovered from spiked reagent water using a One-Pass system and drying and evaporation using the DryVap system. The bar color indicates the fraction in which the compound was eluted.

P a ge 5 Conclusion The One-Pass system using the One-Pass disk and carbon cartridge is an effective way to extract a large suite of analytes with different characteristics with good recovery. The carbon cartridge can be used to capture additional analytes that are more polar and have large water solubility. If these compounds are not of interest, the One-Pass system can be used without the carbon cartridge. In addition, the ability to avoid formation of emulsions is an important step in making the extraction more predictable and avoiding time spent to break an emulsion. References 1. US EPA Method 8270E, can be obtained from https://www.epa.gov/sites/production/files/2017-04/documents/method_8260d_update_vi_final_03-13-2017_0.pdf 2. US EPA Method 625.1, can be obtained from https://nepis.epa.gov/exe/zypdf.cgi?p100lvhc.pdf?dockey=p100lvhc.pdf. 16 Northwestern Drive. Salem, NH 03079. Tel: 603-893-3663. www.horizontechinc.com T N0191711_01

P a ge 6 Appendix LLE One Pass SPE 1 N-Nitrosodimethylamine (NDMA) 50.5 73.3 2 Pyridine 8.4 46.6 3 Phenol 45.8 83.0 4 Aniline 26.3 65.7 5 Bis(2-chloroethyl) ether 90.3 83.8 6 2-Chlorophenol 95.8 94.7 7 1,3-Dichlorobenzene 74.3 73.9 8 1,4-Dichlorobenzene 75.8 74.0 9 Benzyl alcohol 77.5 100.5 10 1,2-Dichlorobenzene 79.8 76.5 11 2-Methylphenol 88.8 99.6 12 Bis(2-chloroisopropyl) ether 83.0 87.6 13 Acetophenone 99.8 93.3 14 3-Methylphenol & 4-Methylphenol 79.0 102.2 15 N-Nitrosodi-n-propylamine 93.3 96.3 16 Hexachloroethane 66.5 72.8 17 Nitrobenzene 97.5 89.4 18 Isophorone 99.0 95.9 19 2-Nitrophenol 103.0 90.1 20 2,4-Dimethylphenol 98.0 100.2 21 Bis(2-chloroethoxy)methane 92.8 95.2 22 Benzoic acid 46.3 40.2 23 2,4-Dichlorophenol 108.5 97.5 24 1,2,4-Trichlorobenzene 79.0 83.5 25 Naphthalene 85.5 86.4 26 4-Chloroaniline 63.0 85.7 27 Hexachlorobutadiene 68.5 74.7 28 p-chloro-m-methylphenol 110.3 98.0 29 2-Methylnaphthalene 89.8 87.6 30 1,2,4,5-Tetrachlorobenzene 90.3 87.7 31 Hexachlorocyclopentadiene 54.8 63.4 32 2,4,6-Trichlorophenol 114.3 101.4 33 2,4,5-Trichlorophenol 116.5 97.8 34 2-Chloronaphthalene 101.0 92.9 35 2-Nitroaniline 105.8 100.8 36 Dimethyl phthalate 106.3 101.1 37 Acenaphthene 98.5 94.9 38 2,6-Dinitrotoluene 110.8 101.7 39 3-Nitroaniline 85.3 79.6 40 Acenaphthylene 89.3 94.7 41 2,4-Dinitrophenol 76.3 91.5

P a ge 7 LLE One Pass SPE 42 4-Nitrophenol 58.5 98.2 43 Dibenzofuran 96.5 96.1 44 2,4-Dinitrotoluene 111.3 101.1 45 Diethyl phthalate 107.0 101.2 46 Fluorene 97.3 96.6 47 4-Chlorophenyl phenyl ether 99.0 95.7 48 4-Nitroaniline 93.8 95.1 49 4,6-Dinitro-2-methylphenol 85.0 98.4 50 NDPA/DPA 90.1 96.3 51 Azobenzene 94.3 97.4 52 4-Bromophenyl phenyl ether 101.8 98.7 53 Hexachlorobenzene 102.5 98.1 54 Pentachlorophenol 100.0 99.1 55 Phenanthrene 102.0 98.5 56 Anthracene 107.8 97.7 57 Carbazole 104.8 101.0 58 Di-n-butyl phthalate 116.3 100.5 59 Fluoranthene 108.5 98.1 60 Benzidine 2.6 63.8 61 Pyrene 101.8 97.9 62 Butyl benzyl phthalate 115.3 98.1 63 Benz(a)anthracene 111.0 97.6 64 3,3'-Dichlorobenzidine 94.6 75.4 65 Chrysene 105.0 99.1 66 Bis(2-ethylhexyl) phthalate 114.5 99.1 67 Di-n-octyl phthalate 11.5 97.0 68 Benzo(b)fluoranthene 128.3 95.5 69 Benzo(k)fluoranthene 106.5 97.6 70 Benzo(a)pyrene 100.5 96.0 71 Indeno(1,2,3-cd)pyrene 139.3 95.0 72 Dibenz(a,h)anthracene 114.8 94.9 73 Benzo(g,h,i)perylene 106.5 90.6 1 2-Fluorophenol (surr) 67.0 66.1 2 Phenol-d6 (surr) 47.4 70.3 3 Nitrobenzene-d5 (surr) 103.2 89.0 4 2-Fluorobiphenyl (surr) 104.8 92.4 5 2,4,6-Tribromophenol (surr) 131.4 101.4 6 p-terphenyl-d14 (surr) 113.6 98.3