NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON ELEMENTAL ASSAY U TITRATION

Similar documents
TECHNICAL WORKING GROUP ITWG GUIDELINE ON LABORATORY APPLICATIONS OF HIGH-RESOLUTION GAMMA SPECTROMETRY

TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON SECONDARY ION MASS SPECTROMETRY (SIMS)

TECHNICAL WORKING GROUP ITWG GUIDELINE ON IN-FIELD APPLICATIONS OF HIGH- RESOLUTION GAMMA SPECTROMETRY FOR ANALYSIS OF SPECIAL NUCLEAR MATERIAL

NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE THERMAL IONISATION MASS SPECTROMETRY (TIMS)

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

ABSTRACT 1. INTRODUCTION

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1

Samples compliant with the WATER

Test Method: CPSC-CH-E

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Environmental Sample Analysis Advances and Future Trends

The role of LSD spikes in safeguarding nuclear reprocessing plants R Wellum, Y Aregbe, A Verbruggen, S Richter

Chapter 4 Chemical Quantities and Aqueous Reactions

Standard Methods for the Examination of Water and Wastewater

OFFICE OF NAVAL RESEARCH. Project NR FINAL REPORT - PART H (Final Summary) ACIDIC AQUEOUS SOLUTION. John B. Hunt and Henry Taube

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Titrimetric Determination of EASTMAN Color Developing Agent, CD-3, in Process ECN-2 Developer with Sulfato Cerate

Current State of Extraction Don t Be Deceived! Sharon F. Webb, Ph.D. Director of Quality Program

7) Applications of Nuclear Radiation in Science and Technique (1) Analytical applications (Radiometric titration)

Exam 2 Chem 311 Fall 2002 You must do (1 or 2), (3 or 4), (5 or 6), (7 or 8), and one other problem!!! All problems are worth 20 points

Preparation and Certification of IRMM-075. A. Verbruggen, A. Alonso, R. Eykens, F. Kehoe, H. Kühn, S. Richter, R. Wellum, Y.

2. QUALITY CONTROL IN MASS SPECTROMETRY

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

Preparation and Certification of IRMM-081a and IRMM-086 Spike Isotopic Reference Materials

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

Evaluation of Nondestructive Assay Characterization Methods for Pipe- Over-Pack Containers

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Introduction to Pharmaceutical Chemical Analysis

Method development for analysis of single hot particles in Safeguards swipe samples

ICP/MS Multi-Element Standards

ANALYSIS CERTIFICATE OF

2. Define what is meant by an oxidizing and reducing agent. Give a good example of each.

ITWG CMX-4 Exercise - LLNL s Experience RMCC XI Workshop

INTRODUCTION TO VOLUMETRIC ANALYSIS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Partner: Judy 29 March Analysis of a Commercial Bleach

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

REFERENCE MATERIALS

Unit 4a: Solution Stoichiometry Last revised: October 19, 2011 If you are not part of the solution you are the precipitate.

Chapter 4: Verification of compendial methods

Application Note #119 Alkalinity and Chloride by Titration

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Rapid Analytical Methods for Determination of Actinides

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Chapter 2: Mass Relations in Formulas, Chemical Reactions, and Stoichiometry

Chemistry 213. Electrochemistry

Acid and Base. Properties of Umass Boston

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Certificate of Analysis

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

Chemistry 301 Test #1

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Standard Methods for the Examination of Water and Wastewater

Karl Fischer Volumetric Titration Theory and Practice

EXPERIMENT 23 Lab Report Guidelines

ITU-Nuclear Safeguards and Security Unit

NICKEL-63/59 IN WATER

ITWG - A Platform for International Cooperation in Nuclear Forensics

Chapter 16 Redox Titrations. In general: Determining the amount of reductant in a sample through titration with a strong oxidant.

Background Statement for SEMI Draft Document #5329 NEW STANDARD: GUIDE FOR TRIS(DIMETHYLAMINO) SILANE (3DMAS)

CHAPTER 4 TYPES OF CHEMICAL REACTIONS & SOLUTION STOICHIOMETRY

AP Chemistry Unit #4. Types of Chemical Reactions & Solution Stoichiometry

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Redox Titrations. -the oxidation/reduction reaction between analyte and titrant. -the equivalence point is based upon:

CHM 111 Final Fall 2012

AGE DETERMINATION OF HIGHLY ENRICHED URANIUM

11/3/09. Aqueous Solubility of Compounds. Aqueous Solubility of Ionic Compounds. Aqueous Solubility of Ionic Compounds

Chapter 4: Chemical Quantities and Aqueous Reactions

Chapter 4 Reactions in Aqueous Solution

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Zinc electrode. 1M zinc sulphate solution

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Basic Digestion Principles

CALCIUM (EGTA-Titrimetric)

Ch 4-5 Practice Problems - KEY

CHEM 107 (Spring-2004) Exam 2 (100 pts)

PURPOSE: 1. To illustrate an oxidation-reduction titration with potassium permanganate 2. To determine the percent mass of iron in an unknown.

AQA A2 CHEMISTRY TOPIC 5.3 REDOX EQUILIBRIA BOOKLET OF PAST EXAMINATION QUESTIONS

AP Chemistry Honors Unit Chemistry #4 2 Unit 3. Types of Chemical Reactions & Solution Stoichiometry

GRAVIMETRIC ANALYSIS

Experiment 5: Determining the Stoichiometry and Products of a Redox Reaction

Behind the scenes: Scientific analysis of samples from nuclear inspections in Iraq

5.2.3 Electrode Potentials and Fuel Cells Redox

CERTIFICATE OF. Single Analyte Custom Grade Solution g/ml (measured at 20 ± 4 ºC) Graviametric NIST SRM Lot Number: See Sec. 4.

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) EDTA NIST SRM 928 Lot Number: 928

Lecture 0403 Displacement Reactions, Molarity, Stoichiometric Calcs.

ANALYSIS CERTIFICATE OF 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

CHEM N-12 November In the electrolytic production of Al, what mass of Al can be deposited in 2.00 hours by a current of 1.8 A?

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

LACTIC ACID. The method is applicable to the determination of lactic acid and lactate salts (Note 2) in light or heavy steepwater.

Chapter 6. Types of Chemical Reactions and Solution Stoichiometry

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L

Joint Research Centre (JRC)

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

INTERNATIONAL STANDARD

Semester 1 Review Chemistry

Transcription:

NUCLE A R FORENSIC S INTERN ATION A L TECHNIC A L WORK ING GROUP ITWG GUIDELINE ON ELEMENTAL ASSAY U TITRATION

EXECUTIVE SUMMARY In this guide, the assay of uranium through titration refers to a methodology determining the U content in material where U is the major constituent. U assay through titration is a well-tested and well-understood method developed over the last 50 to 60 years as part of nuclear safeguards verification in different nations. Other U assay techniques commonly used across the nuclear industry include ignition gravimetry and X-ray fluorescence. These methods are described in separate ITWG guideline documents. Mass spectrometric techniques such as thermal ionization mass spectrometry (TIMS), isotope dilution mass spectrometry (IDMS), or inductively coupled plasma mass spectrometry (ICP-MS) with multi collector, can also be used for U assay, but are also discussed in different ITWG documents. All these U assay methods are considered primarily destructive assay techniques and leave the sample in a form which may or may not be useful for other analyses. The analysis time quoted for the method assumes a standard eight hour day and 5 day work weeks. Times will be different if shift work schedules are available. Uncertainties are expressed using GUM terminology expressed in relative percent. Uncertainties from ASTM/ISO methods, the 2010 International Target Values (ITV), and/or documented historical knowledge are reported. This document was designed and printed at Lawrence Livermore National Laboratory in 2017 with the permission of the Nuclear Forensics International Technical Working Group (ITWG). ITWG Guidelines are intended as consensus-driven best-practices documents. These documents are general rather than prescriptive, and they are not intended to replace any specific laboratory operating procedures.

1. INTRODUCTION Chemical titration is a standard method for the determination of the U concentration of nuclear fuel material for accountability measurements or accountability verifications. In chemical titration, the sample is made to react with an exactly measured amount of a selective reagent of known composition, leading to the completion or characteristic end point of a well-known stoichiometric reaction. Uranium titration can occur under a large variety of conditions using many different reagents; however, the most commonly used and accepted technique is based on a method developed by Davies and Gray in 1964. This method can determine precise uranium assay values with minimum interferences and without requiring separations prior to the assay of uranium materials. A system using computer controlled titrator and an automated sample system is seen below (Fig. 1). The uranium is reduced to U(IV) by excess Fe(II) in strong phosphoric-sulfamic acid, and excess Fe(II) is selectively oxidized by nitric acid in the presence of a Mo(VI) catalyst. Then the U(IV) is titrated by a dichromate solution or by cerium sulfate solution with a potentiometric titration using vanadyl as electrochemical enhancer. The overall process for the dichromate titration is: Cr 2 O 7 2- + 3U 4+ + 2H + = 2Cr 3+ + 3UO 2 2+ + H 2 O 2 = the electron change in U 2. USE FOR NUCLEAR FORENSICS There are many techniques available to apply to a bulk nuclear forensic sample depending on the precision desired, instrumentation availability, expected impurities/ interferences, access to sufficient quantities of standards or reference materials, and the amount of sample provided for assay. The Davies and Gray technique is robust and precise, making this the method of choice for U assay when sufficient material is available. 3. SAMPLE REQUIREMENTS Sample/Standard requirement: 25 mg/titration. However, preparation of samples/standards containing 250 mg of U are needed to meet the uncertainty of 0.1%, k = 2. The remainder of the sample may be used for other analyses. This method is not appropriate for U as a minor sample component. This method is applicable to the determination of uranium in uranium metal, uranium oxides and nitrides, uranium-plutonium oxides, carbides, and nitrides, and to solutions containing uranium and plutonium. The method can be used to To determine the mass fraction (C U ) of uranium in a sample, C U = WT(m c /2m a ) where: W T = the relative atomic mass of the uranium in the sample = the titer of the dichromate solution in equivalent/gram Fig. 1. Automated titrator set up in a fume hood at Atomic Weapons Establishment for uranium assay in the United Kingdom. m c = the mass of dichromate solution used to reach the endpoint in gram m a = the mass of sample aliquant used in the titration, in grams 1

determine uranium in mixed oxides in which the uranium:plutonium ratio varies from 20:1 to 2:1 and in mixed carbides and nitrides containing uranium and plutonium in a ratio of 4:1 or greater. This method is generally free from interferences seen, however, the following elements that are not normally present in uranium materials could interfere when at mg or higher levels: Ag, V, Pt, Au, Ru, Os, I, Sn, As, Sb, Mo, and Mn. Elements that interfere but are removed by fuming a sample prior to analysis are F, Cl, and Br. 4. PRO S AND CON S OF THE TECHNIQUE Pro s: Several laboratories are known to produce accurate results with uncertainties at or below 0.10% for samples containing 20 30 mg of uranium. International Target Value uncertainties range from 0.14% to 0.28%, k = 2 depending on the material analyzed. This method is considered the gold standard for uranium assay. This method can be automated. See Figure 1 above. The manual method does not require special instrumentation and can be used when sample numbers are small and infrequent. Method can handle a broad range of materials with little to no interferences. Pure metals, alloys, oxides, salts, carbides, and nitrides do not pose any difficulty once the sample is dissolved. Method does not typically need any chemical separations, as little will interfere with the assay. Titrants are very stable and can be calibrated versus a number of available uranium assay CRM/SRMs. Con s Chemical separations could be required if unexpectedly large impurities are present. Large amount of material required for initial dissolution to achieve low uncertainties for the method. 5. FAQ Good, commercially available instruments are available in the $10,000 $20,000 range. However, the method can be set up for <$5000 using a manual weighing burette, voltmeter, analytical balance, and stir plate. A group of 4 samples can be analyzed in 1 week, less if the method is automated. Routine analysis of a control material is recommended. Best practices require true replicates of sample (unique sampling leading to each analysis). This allows one to determine if a material is homogenous or not at the sampled level as well as prevent erroneous data due to errors during weighting, dissolution, aliquoting, or other handling of the sample. This method requires chemical assay CRM/SRM. Material is left over for other analyses. 2

Care must be taken in the preparation of the materials and standards being analyzed as the method is sensitive enough to detect even small losses of material. It is recommended that approximately the same amount of sample, standard and control material be analyzed every time. This ensures similar chemical conditions apply for all analyses and helps keep the uncertainty associated with the analysis low. Calibration must be done each day that the measurement is performed. Some of the regents used may have limited shelf life and can require weekly preparation. If reagents are prepared and not used in the same week, their effectiveness is reduced and some reactions used to force or maintain the U in the desired oxidation state will not go to completion causing bias and uncertainty to increase. If using Ce(IV) titrant, care must be taken to prevent the burette tip from touching the solution. If this happens, the Ce(IV) can precipitate in the tip and clog the delivery of the titrant. 2. ASTM C1267-11 Standard Test Method for Uranium by Iron (II) Reduction in Phosphoric Acid Followed by Chromium (VI) Titration in the Presence of Vanadium. 3. ASTM C1204-02 (Reviewed 2008) Standard Test Method for Uranium in Presence of Plutonium by Iron(II) Reduction in Phosphoric Acid Followed by Chromium(VI) Titration. 4. ISO 7097-1 Nuclear fuel Technology Determination of Uranium in Solutions, Uranium Hexafluoride and Solids Part 1: Iron(II) reduction/potassium dichromate oxidation titrimetric method (First Edition). 5. ISO 7097-2 Nuclear fuel Technology Determination of Uranium in Solutions, Uranium Hexafluoride and Solids Part 2: Iron(II) reduction/cerium(iv) oxidation titrimetric method (First Edition). 6. Zendel M., Donohue D.L., Kuhn E., Deron S., Biro T., Handbook of Nuclear Chemistry, chpt. 63, section 63.4.5.2, Uranium Titration, pgs 2972-2973, 2011. DOCUMENT REVISION HISTORY If using Ce(IV) titrant, care must be taken when approaching the endpoint of the titration as the potential break is very sharp and easy to over titrate as compared to titrating with the Cr 2 O 7 titrant. Version No. Version Date Document INFL-AUTI Description of Changes Changes made by 6. REFERENCES 1. Davies, Gray W., A Rapid and Specific Titrimetric Method for Precise Determination of Uranium Using Iron(II) Sulfate as Reductant, Talanta 11, 1964, 1203. 1 April 2017 Initial Draft L. Colletti, L. Walker, L. Tandon (authors) 3