The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Similar documents
Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Author. Abstract. Introduction

Application Note. Authors. Abstract. Introduction. Environmental

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Analyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments. Application. Introduction. Authors. Abstract.

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

MassHunter Software Overview

GC Analysis of Polybrominated Flame Retardants Application

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Agilent 240 Ion Trap GC/MS

MassHunter TOF/QTOF Users Meeting

Mass Spectrometry. Electron Ionization and Chemical Ionization

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

Agilent G2350A Atomic Emission Detector (AED)

Signal, Noise, and Detection Limits in Mass Spectrometry

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Identifying Disinfection Byproducts in Treated Water

Positive and Nondestructive Identification of Acrylic-Based Coatings

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

The Analysis of Organophosphate Pesticides by LC/MS Application

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

CEE 772: Instrumental Methods in Environmental Analysis

ISO INTERNATIONAL STANDARD

Making Sense of Differences in LCMS Data: Integrated Tools

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

WADA Technical Document TD2003IDCR

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Application Note. Abstract. Authors. Environmental

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

LC-MS Based Metabolomics

Mass Spectrometry (MS)

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Analysis of domestic sludge using the Agilent 4200 MP-AES

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Introduction. Chapter 1. Learning Objectives

Using the ACD/MS Manager Software with Agilent 1100 Series LC/MS Systems. Application Note

Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Agilent G3212 GC-APCI Source

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Agilent 5977A Series GC/MSD System

Cerno Application Note Extending the Limits of Mass Spectrometry

GC-CI-MS analysis of TMS derivatives

brochure brief 5973Network Mass Selective Detector General Description and Configurations

Determination of Polybrominated Diphenylethers (PBDE) in Sediment and Sewage Sludge Application

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Modernizing the USP Monograph for Acetaminophen

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Agilent MassHunter Quantitative Data Analysis

SIFT-MS. technology overview

Multi-residue analysis of pesticides by GC-HRMS

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Characterization and Classification of Heroin from Illicit Drug Seizures Using the Agilent 7200 GC/Q-TOF

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Analyst Software. Automatic Optimization Tutorial

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

4000 MS Users Guide Internal Ionization

TIME OF FLIGHT GC/MSTOF

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Transcription:

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note Authors Chris Sandy Agilent Technologies Wokingham, Berkshire, UK Jean-Francois Garnier Agilent Technologies Massy, Essonne, France Harry Prest Agilent Technologies, Inc. Santa Clara, CA USA Abstract The full power of chemical ionization (CI) mass spectrometry (MS) is now conveniently available through the complete integration of CI on the 5975 inert MSD (mass selective detector). CI has been fully automated by providing auto-tuning with electronic control of the reagent gases. All modes of MS operation in a single sequence, completely free of operator intervention, are now possible. Analytes can be sensitively determined in samples by implementing alternating positive- or negative-ci (PCI or NCI) modes. On the 5975 inert MSD, CI is as easy as EI (electron impact ionization). Introduction CI lends selectivity and specificity to the ionization process and is a powerful tool in MS. As an alternative approach to EI ionization, CI can provide enhanced analyte detection in complicated matrices or assistance in confirmation of compound identity. In most cases, based on the particulars of the compounds of interest, a choice between PCI and NCI modes can be made and optimized. This optimization requires selecting an appropriate reagent gas and exploring the various parameters such as source temperature, reagent pressure or flow, etc., to obtain the desired performance. This has been a manual process in the past but with the new integrated electronic control of the CI gases available in the 5975 inert MSD, many of these tasks can be automated. In some cases the analytical situation is not clear-cut and samples contain analytes that require acquisition in PCI and NCI modes. Because the parameters that optimize these two modes of performance are often very different, manual implementation of the appropriate methods was the rule. Again, integration of CI control makes it possible to alternate modes without compromising the important method parameters. The new CI control provides Automated EI ionization tuning of the CI source allowing EI spectra to be acquired in series with CI Completely electronic control of the CI gas flow Electronic selection of one of two possible CI reagent gases Method optimization by survey of reagent flow Most importantly, this allows the user to perform EI, PCI, and NCI acquisitions within a sequence by alternating between methods. This technical note provides examples of some important features in the enhanced CI control of the 5975 inert MSD.

CI Instrument Interface Enhancements CI Reagent Control Figure 1 shows the new CI control panel that replaces the knob and buttons of the 5973N massflow controller (MFC). The gas flow setting is still as the percent of the maximum 5 ml/min flow of the MFC. This panel is accessible from the TUNE View and from the Instrument Control View (under CI Controls or by selecting Edit Mass Parameters). Control of the gas flow is provided as well as selection between the A and B channels. The Shutoff turns all gases off. Note that changes to the CI gas flow and other tune parameters are saved with a TUNE file and this file must be associated with a method to be executed. Automated CI tuning In both modes of CI operation, the CI TUNE WIZARD assists the user in configuring the reagent or buffer gas, selecting reagent ions and their ratios, and the tuning ions of the CI tuning gas itself, perfluorodimethyltrioxadodecane (PFDTD). As previously, PCI tuning with PFDTD is available only in methane, and in NCI, both methane and ammonia use the PFDTD tuning compound. PCI tuning Previously the user manually adjusted the methane CI gas flow to obtain the proper reagent ion ratios for PCI tuning. This is now automated in both methane and ammonia PCI tunes. After flow adjustment, a PCI Autotune can be executed from the pull down menu. Additionally, a survey can be conducted of the reagent ion ratios as a function of reagent flow. This allows the user to select various reagent ion ratios to meet specific PCI criteria. NCI tuning Methane and ammonia NCI tuning are fully automated, and the 5975 inert MSD software sets default buffer gas flows for tuning for both gases as an initial tune. Application Examples Figure 1. The new CI control panel. EI tuning of CI source A menu command automatically executes tuning of the CI source for EI operation, Figure 2. This tuning procedure, like all CI tuning, requires that methane be connected to the instrument at channel A. In developing CI methods for a known list of compounds, the recommended procedure is quite straightforward. Develop a suitable GC method in EI, then switch to CI and acquire a standard in the various CI modes. Figure 3 shows an example for the sequence table for such a survey. When a particular mode provides the best analysis, optimize that mode. The criteria for what is the best may be most favorable spectra, or greatest response, or it may include issues concerning the sample matrix, such as minimizing mass spectral interferences. Figure 2. EI tuning of CI source. 2

Figure 3. Example of CI survey sequence. Figure 4 shows the reconstructed total ion chromatogram (RTIC) for one of several phthalates acquired in the two CI modes with two reagent gases. As expected, the spectra in each mode are all different. For these compounds, PCI with ammonia reagent gas provides the greatest response in full scan and the most distinctive spectrum. PCI NH 3 391 M + NCI NH 3 277 221 M _ 390 149 221 PCI CH 4 NCI CH 4 167 M _ 390 113 279 M + 391 148 Figure 4. RTIC in PCI and NCI with methane and ammonia CI gases (middle panel) of a dioctyl phthalate. Note the differing spectra and total responses indicated by the arrows. 3

In Figure 4, the temperature of the source had a profound effect on response and especially chromatographic peak shape. This is most pronounced in NCI mode. Figure 5 shows an example of optimizing source temperature for the higher molecular weight brominated diphenyl ethers (BDE). Typically NCI favors lower source temperatures but in this case the loss in sensitivity is offset by improved chromatography achieved by the better volatilization of the analytes at higher source temperatures. This data was produced by a sequence where the four methods were the same except for the NCI tune files which had different source temperatures. Scan 75 to 975 u; ECNI CH 4 at varying ion source temperatures Deca-BDE 150 C 200 C Octa-BDE 150 C 200 C Nona-BDE 150 C 200 C 300 C 250 C 250 C 300 C 250 C 300 C Figure 5. Effect of optimizing NCI source temperature. CI reagent flow is especially important. In a continuation of the polybrominated diphenyl ethers example, Figure 6 shows the dramatic influence of buffer gas flow into the source. In PCI mode with ammonia reagent gas, this is especially important as ammonia source pressure has a profound effect on the ratio of protonated species ([M+H] + ) to the adduct species ([M+NH 4 ] + ) of the analyte molecule ([M]). 4

Scan 75 to 975 u; ECNI CH 4 at varying ion source pressures for same PBDEs in Figure 5 2.0 ml CH 4 /min 1.0 ml CH 4 /min Figure 6. Influence of source pressure on NCI response. Figure 7 presents such data for the benzophenone PCI checkout compound. Using the reagent ion ratios allows the user to optimize the ratio of the target and qualifying ion for their analysis and to reproduce the results in future tunes. A convenient way to gain this information is through the FLOW SURVEY menu item in CI tune. This feature steps the reagent flow over a user selectable range and presents the results for the response and ratio of responses for two reagent gas ions. It is knowledge of this ratio that is key to optimizing and reproducing PCI performance. [M+NH 4 ] + [M+H] + 0.05 0.09 0.32 0.53 0.75 Figure 7. PCI with ammonia reagent gas. Influence of source pressure on benzophenone spectrum as a function of source species seen in the ratio of m/z 35 : m/z 18. 5

www.agilent.com/chem A convenient way of conducting a series of flow experiments is by employing the command setflow # in a sequence, where # represents the setting of the reagent MFC. See Figure 8. This allows a single CI method to be used with different reagent gas flows in a sequence. Similarly, the ability to shutoff the CI gas at the end of a sequence is given by the sequence commands CIGASOFF and PAUSE, Figure 8. Conclusions The full power of CI/MS is conveniently available for the analyst through the complete integration of CI on the 5975 inert MSD. CI has been fully automated by providing auto-tuning with electronic control of the reagent gases. Now all modes of mass spectrometric operation in a single sequence that is completely free of operator intervention is possible. This allows analytes to be sensitively determined in samples by implementing alternating PCI or NCI modes. On the 5975 inert MSD, CI is as easy as EI. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. Figure 8. Commands for changing and shutting off CI gas flow in a sequence. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2005 Printed in the USA December 7, 2005 5989-4347EN