Thermal Analysis Excellence

Similar documents
Automated multi-vapor gravimetric sorption analyzer for advanced research applications

DYNAMIC VAPOR SORPTION ANALYZER FULLY AUTOMATED COMPACT BENCH-TOP DVS ANALYZER, FOR FAST AND ACCURATE SORPTION MEASUREMENTS.

DVS Intrinsic Compact and Economical Dynamic Vapor Sorption System

DVS INTRINSIC. The easy-to-use solution to complex water sorption challenges.

Thermal Analysis Premium

Surface Measurement Systems

Moisture Sorption Analysis of Pharmaceuticals

Moisture Sorption Analysis of Pharmaceuticals

TA INSTRUMENTS. Vapor Sorption Analysis

Understanding Importance of Water Sorption Isotherm Shape, Hysteresis, and Models on Pharmaceutical Materials

Technical brochure DuraLac H TABLETING DIRECT COMPRESSION ANHYDROUS LACTOSE

Pharmaceutical Characterisation. Dr. Lidia Tajber and Dr. Krzysztof Paluch School of Pharmacy and Pharmaceutical Sciences, Trinity College Dublin

High Pressure DSC Differential Scanning Calorimeter

Characterization of Solid State Drugs by Calorimetry

Complete Moisture Analysis

ASAP: Accelerated Stability Assessment Program. Improved Protocol and Data Analysis for Accelerated Shelf-Life Estimation of Solid.

Crystallization behavior of freeze-dried lactose and lactitol as detected from the loss of sorbed water

Val Joly (16-20 June 2014)

Preformulation & Solid State Chemistry Services

TOPEM the new advanced multi-frequency TMDSC technique

Apparent Melting: A New Approach to Detecting Drug-Excipient Incompatibility

Gravimetric Analysers for the Characterisation of the Sorption Properties of Materials.

11/23/ (4) Harmonization Stage 6: <1241> Water Solid Interactions in Pharmaceutical Systems (USP34 NF29 2S) BRIEFING


Complete Moisture Analysis

WATER ACTIVITY MEASUREMENT WATER ACTIVITY

Application Note. Fundamentals of Moisture Sorption Isotherms. Hysteresis

dissolved into methanol (20 ml) to form a solution. 2-methylimidazole (263 mg) was dissolved in

Hiden Isochema. Gravimetric Gas & Vapor Sorption Analyzers. Hiden Isochema IGA Series. Advancing Sorption Analysis

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

FORCE TENSIOMETER K100 THE HIGH-END SOLUTION FOR MEASURING SURFACE AND INTERFACIAL TENSION

Particle Characterization of Pharmaceutical Products by Dynamic Image Analysis

Measuring the flow properties of powders. FT4 Powder Rheometer. freemantechnology

C80. Calvet Calorimeter From ambient to 300 C. A trademark of KEP Technologies group

International Journal of Advanced Chemical Science and Applications (IJACSA)

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer

The Application of the Accelerated Stability Assessment Program (ASAP) to Quality by Design (QbD) for Drug Product Stability

Particle Size and Shape Analysis in PHARMACEUTICAL INDUSTRY.

Sorption-Desorption Behavior and Characterization of Cyclophosphamide

LUMiSizer. Dispersion Analyser. The All-in-One. Dispersion Analyser. Stability Demixing Consolidation Particle Sizing

News & Trends for Thermal Analysis

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed

PTG-NIR Powder Characterisation System (not yet released for selling to end users)

Supplementary Information

BIO & PHARMA ANALYTICAL TECHNIQUES. Chapter 5 Particle Size Analysis

EXPERIMENT 6 Empirical Formula of a Compound

CHAPTER 7 SUMMARY AND CONCLUSIONS. constitutes the objective of many a research project in the field of pharmaceutical

SARA Pharm Solutions

Adsorption Processes. Ali Ahmadpour Chemical Eng. Dept. Ferdowsi University of Mashhad

Karl Fischer Titration Of Food : Towards Complete Automation

HYGROTHERMAL CHARACTERISTICS OF PUMICE AGGREGATE CONCRETE USED FOR MASONRY WALL BLOCKS

Automated Stability Testing Delivering Rapid Product Development

The Vacuum Sorption Solution

Mercury Stack Monitor SM-4

Method development in thermal analysis. Part 1:

Alcolyzer Plus. Beer Analyzing System. ::: Unique Density & Concentration Meters

CHEM*3440. Thermal Methods. Thermogravimetry. Instrumental Components. Chemical Instrumentation. Thermal Analysis. Topic 14

DSC Methods to Quantify Physical Aging and Mobility in Amorphous Systems: Assessing Molecular Mobility

473-SHX Dew Point Mirror

DVS Application Note 101

White Paper. Perform Conductivity Measurements In Compliance with USP <645>

Properties of a New Type of Plaster Containing Phase-Change Material

Agilent Cary 630 FTIR sample interfaces combining high performance and versatility

TGA Thermal Gravimetric Analysis

Determination of Solubility Parameters using Inverse Gas Chromatography

Reliable, fast and accurate WVTR measurement technology. Dr. Jörg Koch SEMPA SYSTEMS GmbH

ISEmax CAM40/CAS40. Technical Information

Progress on Moisture Swing Sorbents. Klaus S. Lackner Columbia University

Supplementary Information

The integration of NeSSI with Continuous Flow Reactors and PAT for Process Optimization

PMF2000 Series DESCRIPTION FEATURES APPLICATIONS MAXIMUM RATINGS MASS AIR FLOW SENSOR

Design, Testing and Pharmaceutical Applications of a Gas Pressure Controller Device for Solid - Gas Microcalorimetric Titration

ActiPix SDI300 Surface Dissolution Imaging System

FORCE TENSIOMETER K100 THE HIGH-END SOLUTION FOR MEASURING SURFACE AND INTERFACIAL TENSION

473 Dew Point Hygrometer

White Paper. Moisture Analyzer Routine Testing

DVS. Water Vapor Sorption Analyzer ADVENTURE

Heat stress index. Dewpoint. pressure. altitude

Pharmaceutical Polymers for Tablets and Capsules

Semi-automatic titrators PH-Burette. A great ph-meter with integrated burette. Incredible price! Infinite possibilities

DM70. Hand-held Dewpoint Meter. Truly Fast Response - Superior Long-term Stability

Study of Proton conductivity of a 2D flexible MOF and a 1D coordination polymer at higher temperature

PHARMACEUTICAL BULLETINS

Adsorption boundary curve influenced by step interval of relative humidity investigated by Dynamic Vapour Sorption equipment

Characterization of Water Adsorption and Absorption in Pharmaceuticals

Effect of temperature and humidity on vegetable grade magnesium stearate

DVS Application Note 46

High-Pressure Volumetric Analyzer

Formulation of Low Dose Medicines - Theory and Practice

Humidity Mass Transfer Analysis in Packed Powder Detergents

473 Dew Point Hygrometer

Phase Behavior of Water and Cyclodextrin Systems

Polymorphism A Critical Consideration in Pharmaceutical Development, Manufacturing, and Stability

New Horizons in Accelerated Stability Modeling-- Tablet Dissolution, Tablet Disintegration and Product Appearance

3D Thermal-Diffusion Analysis on a Moisture Loaded Epoxy Sample

Method Development. Creating the Perfect Standard Operating Procedure (SOP) 2007 HORIBA, Ltd. All rights reserved.

Labcell Water Activity Instruments

Analyzing & Testing. Photocalorimetry Photo-DSC. Method, Technique, Applications. Photo-DSC 204 F1. Leading Thermal Analysis

Transcription:

Thermal Analysis Excellence Sorption Test Systems SPS11-1µ SPSx-1µ High Load SPSx-1µ Advance SPS23-1n Multi-Sample Moisture Sorption Analysis Simple, Accurate and Reliable

Sorption Test Systems Moisture Sorption Analysis Multiple Samples with Highest Accuracy Virtually all materials interact with moisture to some extent. Understanding moisture induced changes of material properties is key to making decisions regarding processing, packaging, storage and shelf life of a product. Features and benefits of the SPS Moisture Sorption Test Systems: The only true multi sample sorpton test system up to 23 samples can be measured simultaneously High dynamic load range thanks to integrated METTLER TOLEDO balance Excellent baseline stability internal reference ensures drift compensation Outstanding humidity accuracy polymer humidity sensor enables measurements near % RH Optimized sorption kinetics due to ideal surface to volume ratio and laminar flow over the sample Extra samples for additional analyses up to 24 samples in parallel positions for DSC, XRD, microscopy, etc. Attractive options extend application range permeability kit, CCD video camera, NIR and RAMAN interface available The heart of all sorption test systems is an integrated METTLER TOLEDO balance which ensures unbeatable accuracy. 2

True Multi Sample Sorption Tests Allow for High Sample Throughput Measure more samples in shorter time With the SPS sorption test systems, the sorption behavior of up to 23 samples can be determined simultaneously. Especially in test series with multiple samples or high sample volumes, accumulated measurement times of single-sample sorption test systems often are far too long. In a single sample instrument, the data to the left would have taken 11 months to generate. In the SPS, the data generation took just 3 days. Weight change in % Temperature in C Relative humidity in % 12 1 9 1 8 8 7 6 6 5 4 4 3 2 2 1 1 2 3 4 5 Time in days Accurate sample comparison In the SPS sorption test systems, all samples are measured simultaneously and under the same conditions. As the experimental conditions are identical, even the smallest differences between samples and batches can be detected. The curves on the left show the sorption behavior of various batches of the same active pharmaceutical ingredient with different drying kinetics and hydration properties. Weight change in % 27 2 13 6-1 6 12 18 24 3 36 42 Time in minutes 4 2 r.h. in% Access to complete sorption behavior When using single-sample instruments, the time available for each measurement is often limited and the individual climatic settings can only be maintained for very short periods. Important information on sample behavior, e.g. stability, may therefore be missed. The figure shows a comparison of the same compound measured in a single-sample and a SPS instrument. The SPS measurement shows small reproducible effects missed in the other measurement because equilibrium conditions were not reached. Standard DVS with suspension balance Sample material in a small pile Gas flow SPS with top loading balance Gas flow Sample material in a thin layer Optimized sorption kinetics Proper sample arrangement and climate control create optimal conditions for sorption kinetics. No diffusion delays, boundary layers or humidity gradients occur in the sample thanks to optimized surface-to-volume ratio and gas flow directly over the sample surface. 3

Choice of models For All Requirements A Suitable System The SPS Sorption Test Systems are available in four different configurations which are adapted to different measurement needs. A broad range of options including digital imaging or RAMAN-spectroscopy can be added to the system which can contribute to an even deeper understanding of sample behavior. SPS11-1µ Best suited for voluminous and bulky samples i.e. in quality control. Can be fitted with up to 11 high volume samples with a weight up to 8 g and offers a resolution of 1 µg. SPSx-1µ High Load The SPSx systems offer the best of both worlds: high precision or high load combined with best versatility regarding number of samples. Both can be operated with 11 or 23 samples. The High Load model offers the ability to run the largest samples thanks to a dual range balance with a resolution of 1 µg up to 22 g and a resolution of 1 µg up to 22 g. SPSx-1µ Advance The Advance offers better precision (±2 µg, vs ±5 µg for the High Load) with a resolution of 1 µg up to a maximum load of 22 g. SPS23-1n Offers high resolution for pharma applications and pre-formulation. Up to 23 samples can be measured simultaneously. Up to 2 g can be determined with a resolution of.1 µg. Options By adding special sensor mounts, SPS systems can be equipped with a CCD camera or a RAMAN PhAT probe, each with full software integration. A 21 CFR part 11 software option is available for operation in highly regulated environments. Other special options allow for multisample permeability or large objects testing. 4

Key 1 Humidity sensor 2 Heat exchanger 3 Autosampler 4 Tray 5 Sample 6 Balance 7 Fans 8 Heated column 9 Air flow Relative humidity in % 1 9 8 7 6 5 4 3 2 1 Lower temperature limit +5 C Operation range Temperature limit of balance calibration +4 C Upper temperature limit +6 C Superior measurement principle Sample material is placed in small dishes on a rotating sample tray. Humidified air flows over the samples in a laminar fashion. The atmosphere is thermostated by a heat exchanger with large exchange area. A solid state sensor in the back accurately measures humidity and temperature. By continuously turning and lowering the tray, the dishes are placed onto the highly sensitive balance and weighed at fixed time intervals. A large range of temperature and humidity can be achieved in a measurement. -2-1 +1 +2 +3 +4 +5 Additional samples The SPS provides additional sample positions on the inner ring of each tray. These samples are subjected to the same humidity and temperature conditions but they are not weighed. They are therefore conditioned identically to the main sample but can be taken out any time for measurements with other analytical techniques like DSC, XRD, SEM. Weight change in %.1 Empty sample pan.8.6.4.2...5 1. 1.5 2. 2.5 -.2 Time in days -.4 -.6 -.8 -.1 1 Temperature in C Relative humidity in % 9 8 7 6 5 3. 4 3 2 1 Excellent baseline stability Much more reliable results can be achieved due to excellent baseline stability because of the measurement principle of discontinuous weighing and reference measurements before and after each weighing cycle. The figure shows full range stepwise humidity increase at 5 different temperatures. The weight signal as purple dots is very stable without any drift. 5

Applications Broad Application Range As a multitude of different materials can be investigated using the SPS systems, it is ideal both for research and development and quality control. Application areas include food, pharmaceuticals, polymers, chemical compounds, textiles, cosmetics and construction materials. Examples of effects and materials that can be investigated with SPS systems Effects Moisture uptake and release Formation of hydrates Deliquescence Amorphous content Stability and caking Moisture-induced glass transition Sorption and desorption kinetics Permeability and moisture diffusion Materials Food powders Processed food Biscuits Grains, seeds APIs Excipients Tablets Polymers Formation of drug hydrates The two figures show the water vapor sorption isotherms of two hydrating drugs. Drug A is a classical stoichiometric monohydrate. Drug B only changes at very high humidity of 9% r.h. into a hydrated form. Desorption runs over several well observable dehydration steps. The distinct hysteresis between sorption- and desorption curve is a characteristic for stoichiometric hydrates. 6

Isotherms of excipients The three figures show the water vapor sorption isotherms of different excipients. Microcrystalline cellulose Avicel follows the classical sorption behavior of polysaccarides with distinct hysteresis between sorption- and desorption curve. Saccharose exhibits liquefaction above a critical humidity of 8% r.h. Lactose monohydrate adsorbs only.13% water over the entire humidity range. Nevertheless a well distinguished isotherm can be determined. Amorphous content in spray dried lactose With increasing relative humidity amorphous lactose adsorbs water and at certain water content the lactose becomes mobile enough to start crystallizing. This begins at 5% r.h.. Since crystalline lactose adsorbs much less moisture, a release of excess water is observed. Because lactose crystallizes to the monohydrate form, drying of the crystallized sample will cause a weight difference before and after the whole treatment. This weight difference is the additional water molecule that accompanies each lactose molecule in the α-lactose monohydrate crystal lattice. From this the amount of amorphous lactose that was present in the original sample can be calculated. Weight Change in %.8.7.6.5.4.3.2.1 2% r.h. Wrapping film A Wrapping film B Aluminum foil Temperature Relative humidity 6% r.h. 8% r.h. 1% r.h. 5 1 15 2 25 3 35 4 time in hours 1 9 8 7 6 5 4 3 2 1 Temperature in C/r.h. in % Permeation measurement Using the permeation kit water vapor permeability can be measured. The experiment shows dynamic sorption curves of two different wrapping films which were measured with mole sieve located in the permeation pans. Aluminum foil which is water tight was measured as control. From the slope the diffusion in grams per square meter and day can be calculated according to EN ISO 7783-1: Part 1: Dish method for free films. Aluminum foil adsorbs a small amount of surface water only. The diffusion through the two films is dependent on the relative humidity. Film A has slightly higher diffusion coefficient than film B. 7

SPS Specifications SPS11-1µ SPSx-1µ High Load SPSx-1µ Advance No. of samples 11 11 or 23 11 or 23 23 SPS23-1n Sample volume 12 ml 12 or.5 ml 12 or.5 ml.5 ml Sample surface 2 cm 2 2 or 1.3 cm 2 2 or 1.3 cm 2 2 or 1.3 cm 2 Additional samples 12 12 or 24 12 or 24 24 Balance Weighing range.3... 8 g.3... 22 (22) g.1... 22 g.5... 2 g Dynamic range full range full range full range full range Resolution 1 µg 1 (1) µg 1 µg.1 µg Repeatability ±2 µg RMS ±5 (2) µg RMS ±2 µg RMS ±1.5 µg RMS Balance model WXS25 WXS26 WXS26 XP2U Temperature Temperature control Peltier element Temperature range 5 6 C 5 6 C 5 6 C 5 4 C Temperature accuracy over time ±.1 C, uniformity ±.5 C Humidity Humidity control Humidity range Polymer type sensor... 98% r.h. Humidity accuracy ±.6% r.h. (... 1% r.h.) at 23 C ±5 C Long term stability Dimensions W x D x H / Weight Power supply Gas supply Operating pressure better than 1% r.h. per year 488 x 63 x 437 mm (124 mm with open lid) / 68 kg (without keyboard and display) 1 23 VAC / 5... 6 Hz Compressed air / nitrogen, dry and oil free 22... 145 psi Options Software validation package according to 21 CFR part 11 Permeability testing kit Camera system RAMAN kit NIR kit Large objects tray www.mt.com/na-dvs For more information Mettler-Toledo GmbH, Analytical CH-863 Schwerzenbach, Schweiz Tel. +41 44 86 77 11 Fax +41 44 86 72 6 Quality certificate. Development, production and testing according to ISO 91. Environmental management system according to ISO 141. Subject to technical changes 1/217 Mettler-Toledo GmbH, 3394383 Marketing MatChar / MarCom Analytical European conformity. The CE conformity mark provides you with the assurance that our products comply with the EU directives.