Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Similar documents
Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

The new Agilent 1290 Infinity Quaternary LC

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Fast methods for the determination of ibuprofen in drug products

Determination of Alkaloids in Goldenseal Using Agilent Bond Elut Plexa Solid Phase Extraction Sorbent for Cleanup and HPLC-DAD Analysis

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Modernizing the USP Monograph for Acetaminophen

Application Note. Authors. Abstract. Food Testing & Agriculture. Stephen E. Reichenbach University of Nebraska Lincoln, NE, USA

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

with diode array detection

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

Quantification of Furocoumarins

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Transferring HPLC methods from the HP 1090 Series to the Agilent 1100 Series HPLC system

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Agilent LC & LC-MS Cannabis Testing

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Figure 1. Structures for Vitamin B2 and Vitamin B1.

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Rapid quantification of Chinese medicine Zuo Jin Pill using rapid resolution liquid chromatography

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

The Agilent 1290 Infinity II Multi Method Solution

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of PEGylated Proteins with Agilent AdvanceBio SEC Columns

Appendix II- Bioanalytical Method Development and Validation

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

High Resolution Fast LC

Application. Introduction. Authors. Abstract. Hydrocarbon Processing

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Routine MS Detection for USP Chromatographic Methods

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Volume 6, Issue 2, January February 2011; Article-015

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Peptide Isolation Using the Prep 150 LC System

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Application Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Transcription:

Agilent Application Solution Analysis of acaricides in honey Application Note Food Authors mau 12 1 A.G. Huesgen Agilent Technologies, Inc. Waldbronn, Germany 8 6 4 2 2 4 6 8 1 12 14 16 18 Abstract Varroosis is a bee disease caused by the mite Varroa Jacobsoni, which endangers beekeeping all over the world. In order to prevent economic losses, beekeepers treat their colonies with acaricides. Using these acaricides inside beehives implies a risk of direct pollution of honey and other hive products, therefore maximum residue levels (MRLs) on honey have been fixed in many countries to protect consumers. In this Application Note, an HPLC method was developed and validated for four acaricides,,,, and using the Agilent 126 Infinity LC system. In addition, an UHPLC method was developed using the Agilent 129 Infinity LC system, which saves time and solvent consumption, and results in better sensitivity.

Introduction Acaricides are widely used to keep the possibility of infection by Varroa Jacobsoni under control.,,, and are frequently used. These acaricides belong to four different chemical families, see Figure 1. N (MW 293.) Amidine H3 C Br H CH 3 P S N (MW 362.77) rganophosphorus N CH 3 CH 3 CH 3 CI The use of these pesticides implies the risk of contaation of consumer products like honey and other beehive products. Therefore, many countries have set maximum residue limits (MRL) to reduce the health risk for consumers. Two norms were defined in ECC Regulation 2377/9 for and :.1 and.2 mg/kg respectively. 1,2 To detere acaricides in honey several LC methods have been used. 3-6 This method was developed based on these publications. Experimental Instrument and software An Agilent 126 Infinity Binary LC system consisting of the following modules was used: Agilent 126 Infinity Binary Pump (G1312B) Agilent 126 Infinity Autosampler and sample thermostat (G1367E, G 133B) Agilent 126 Infinity Thermostatted Column Compartment (G1316A) Chromatographic parameters Agilent 126 Infinity Diode Array Detector (G4212B) with 1-mm path length flow cell The UHPLC analysis was developed and performed using an Agilent 129 Infinity LC system consisting of the following modules: Agilent 129 Infinity Binary Pump (G422A) Agilent 129 Infinity Autosampler and sample thermostat (G4226A and G133B) Agilent 129 Infinity Thermostatted Column Compartment (G1316C) Agilent 129 Infinity Diode Array Detector (G4212A) with 1-mm Max-Light flow cell Columns: Agilent ZRBAX Eclipse Plus C18, 4.6 15 mm, 5 µm (p/n 95993-92) Agilent ZRBAX RRHD Eclipse Plus C18, 2.1 5 mm, 1.8 µm (p/n 959757-92) Software: Agilent ChemStation B.4.2 Parameter Conventional method Fast UHPLC method Br (MW 428.1) Benzilate Column: Agilent ZRBAX Eclipse Plus C18, 4.6 x 15 mm, 5 µm Agilent ZRBAX RRHD Eclipse Plus C18, 2.1 5 mm, 1.8 µm Mobile phase: water (A), acetonitrile (B) water (A), acetonitrile (B) Gradient: 4 to 9% B in 15 4 to 9% ACN in 1.16 H At 2 9% B At 2.1 4 % B At 25 4 % B At 1.557 9% B At 1.56 4% B At 1.93 4% B Flow rate: 1 ml/.9 ml/ H (MW 394.4) Isoflavonoid Figure 1 Chemical structure of acaricides. Column temp: 3 C 3 C DAD: Injection volume:.5 sec, 1 Hz, 29 nm (rotenone), 29 nm (), 23 nm (), 29 nm (), Identification through UV spectra 2 µl with 6 sec for exterior needle wash with methanol, sample compartment was cooled to 1 C and kept dark.13 sec, 4 Hz, 29 nm (rotenone), 29 nm (), 23 nm (), 29 nm (), Identification through UV spectra 1.5 µl with 6 sec for exterior needle wash with methanol, sample compartment was cooled to 1 C and kept dark 2

Reagents and materials All chemicals and solvents used were HPLC grade, and highly purified water from a Milli Q water purification system was used. Acetonitrile gradient grade was purchased from Merck (Darmstadt, Germany). All standards were ordered from Sigma Aldrich, Germany Preparation of standards To perform the validation tests, a dilution series was set up, see Table 2. For the stock solution and all other dilutions, Acetonitrile was used as solvent. The stock solution contents are shown in Table 1. Compound 18.65 18.25 28.55 18.9 Weighted sample (mg/1 ml) Table 1 Concentration of stock solution. The stock solution was stored at 4 C in the refrigerator and was stable for at least two months. Dilutions 2 to 4 are typical concentrations obtained after sample preparation of honey samples. Precautions is very light sensitive and should be kept in the dark. The light of the autosampler should be switched off. should not be heated above 4 C due to decomposition. Sample preparation Sample preparation was done in the following way: 1 g honey + 1 ml Acetonitrile Mechanical stirring for 3 utes 8 ml of the supernatant liquid are transferred into 25 ml rotary vessel. Evaporate to dryness at 3 C with vacuum rotary evaporator from Büchi. Reconstitute in 5 µl acetonitrile. Transfer to a LC vial and inject 2 µl. Spiking procedure: 1 g spiked with 2 µl of dilution 2 (~.1 mg/kg) or dilution 3 (~.5 mg/kg) Mechanical stirring for 1 utes Procedure The following steps were taken to develop and validate the method on the Agilent 126 Infinity LC system: Method development: Standards were injected to elute all peaks in a reasonable time (~ 2 utes) with an Agilent 126 Infinity LC system using standard-bore 4.6 mm id columns. Method validation: Area and RT precision, LD/LQ, linearity (relevant range), recovery rates, and accuracy (from spiked matrix), robustness (column temperature, flow, gradient steepness, wavelength, injection volume) were evaluated. Sample preparation: ne relevant matrix was chosen. Unspiked honey and two spiked honeys were prepared and analyzed. Analysis: Injection of spiked real-life sample with quantification and identification through UV spectra was performed. Having developed and validated the conventional method, the analysis was transferred to the Agilent 129 Infinity LC system to develop an UHPLC method. Method transfer to UHPLC: An UHPLC method was developed increasing speed and sensitivity using the Agilent 129 Infinity LC system and short sub-2-µm columns. Proof of UHPLC method performance: Precision of area and RT, LD/LQ was tested. Compound 1st 1:2 of stock solution () 2nd 3rd 4th 5th 6th 7th 8th 186.5 93.25 46.625 23.3125 11.65625 5.828125 2.91463 1.45731 182.5 91.25 45.625 22.8125 11.4625 5.73125 2.851563 1.425781 285.5 142.75 71.375 35.6875 17.84375 8.921875 4.46938 2.23469 189 94.5 47.25 23.625 11.8125 5.9625 2.953125 1.476563 Table 2 Dilution series. 3

Results and Discussion Separation and detection A conventional method was developed using an Agilent ZRBAX Eclipse Plus C-18, 4.6 15 mm, 5 µm particles column. A gradient from 4 to 9% organic was used and a flow rate of 1 ml/. Figure 2 demonstrates the excellent separation. nly water and acetonitrile were used to achieve the separation the four compounds. Buffers or further modifiers were not needed. Two wavelengths were needed to measure all compounds at their absorbance maximum. For,, and, a wavelength of 29 nm was used for calibration. For, a wavelength of 23 nm was used. The characteristic spectra (Figure 3) of the four compounds were used to create a UV spectral library. This library, along with the retention times, helped identify the compounds in the honey sample. Method validation Limit of detection (LD) and Limit of quantitation (LQ) The limit of detection and quantitation was evaluated using dilution 8. In Figure 5, the chromatogram for this concentration is shown. mau 5 4 3 2 1 Compound 1 12 14 16 18 Figure 2 Analysis of the four Acaricides on Agilent ZRBAX Eclipse Plus C18, 4.6 x 15 mm, 5 µm column. 1:2 dilution of stock solution () 186.5 182.5 285.5 189 The analyte concentration that provides a signal to noise ratio (S/N) of > 3 was considered as LD and compound concentration with 1 * LD was considered as LQ, which corresponds to dilution 8. verall the LD is <.2 and LQ (1*LD) is < 1.8. Precision of retention times and areas Retention time RSD values for all four acaricides across the linearity levels were calculated. The precision of retention times for six consecutive runs was typically <.5% RSD. The precision of the complete sequence was <.16 % RSD, over 7 runs within 29 hours. The Precision of Areas was tested over the complete dilution series. The results are combined in Figure 4. Even though the peak height of dilution 8 is only ~.1 to.2 mau, the RSD value of < 6.5%, is very good. All other RSD values were < 2% over the complete series and < 1% RSD from dilution 5 on. Figure 3 Spectra of acaricides. RSD(%) 8 6 4 2 Precision of area 8 dil 7 dil 6 dil 5 dil 4 dil 3 dil 2 dil 1 dil RSD area Figure 4 Precision of areas over the complete dilution series, over 6 consecutive runs for each concentration. 4

Linearity To test the linearity 1 to 8 were used. Each linearity solution was injected six times and the area was used to construct the linearity curve. The linearity of all four compounds was >.9999 for the coefficient of correlation over the complete dilution series. mau 5 4 3 2 1 Compound 8th dilution. 1.45731 1.425781 2.23469 1.476563 Carry over To test the carry over behavior dilution 1 was injected followed by the injection of pure solvent. No carry over was observed for the concentration range used. Robustness test Five critical parameters were changed and data collected in 1 replicate injections. Values from the last six replicates were used for the analysis. Allowed deviation for retention time and area was set to ± 3.% and Figure 5 Analysis of dilution 8. 9 1 11 12 13 14 15 16 ± 5% respectively. Robustness of the method was tested using dilution 1. The results showed that retention time shifts > 3% have to be expected for gradient slope differences of ± 1%, see Table 3. Regarding robustness related to compound amounts, it is critical to change the injection volume and the flow rate. Robustness results indicate that the method is reliable for normal usage and to a great extent the performance remains unaffected by deliberate change in parameters. However, some parameters are critical and must be carefully controlled. Parameters changed Flow ± 2% Standard: 1 ml/ TCC ± 5% Standard: 3 C Inj ± 5% Standard: 2 µl Gradient slope ± 1% Wavelength ± 3 nm 29 nm 23 nm 254 nm Changes High: 1.4 ml/ Low:.96 ml/ High: 31.5 C Low: 28.5 C 2.23 1.87 1.59 1.47 3 + 1.99 1.94 1.69 1.57 3.3.2.18.27 3 +.79 +.76 +.67 +.64 3 High: 21 µl +.26 +.2 +.19.14 3 Low: 19 µl.32.24.23.19 3 55% in 15 + 1% 45% in 15 1% DAD 293 nm 233 nm 257 nm DAD 287 nm 227 nm 251 nm 3.91 4.75 5.3 5.64 3 + 4.66 + 5.68 + 6.41 + 6.83 3.2.1.1 3.2.2 3 ± % Deviation limits Table 3 Robustness tests related to retention times. 5

Parameters changed Flow ± 2% Standard: 1 ml/ TCC ± 5% Standard: 3 C Inj ± 5% Standard: 2 µl Gradient slope ± 1% Wavelength ± 3 nm 29 nm 23 nm 254 nm Changes High: 1.4 ml/ Low:.96 ml/ 3.13 3.33 2 3.87 5 + 5 + 4.4 + 6 + 5.16 5 High: 31.5 C + 3.13 + 2.78 + 3 + 1.94 5 Low: 28.5 C + 2.5 + 2.2 + 2.5.65 5 High: 21 µl + 6.88 + 6.67 + 7 + 5.81 5 Low: 19 µl + 3.75 + 3.9 + 3.5 + 4.52 5 55% in 15 + 1% 45% in 15 1% DAD 293 nm 233 nm 257 nm DAD 287 nm 227 nm 251 nm + 1.25 +.56 + 2 +.65 5 + 1.25 + 1.1 ± 2.5 + 5.16 5 + 2.5 2.2 4.5 + 3.23 5 6.88.56 + 3.23 5 ± % Deviation limits Table 4 Robustness test related to compound amounts. Recovery rates The honey used was a fully liquid honey which contained none of the tested acaricides. The honey was spiked with the four compounds to obtain a resulting concentration of.1 and.5 mg/ kg of the acaricides. In Figure 6, the chromatogram of honey spiked with ~.5 mg/kg is shown after sample preparation. The recovery rate was for honey spiked with: 5 ng/g,.5 mg/kg = recovery 67 88% 1 ng/g,.1 mg/kg = recovery 84 15% shows the lowest recovery rate, this may be due to decomposition. In addition, the UV spectral library was used to identify the compounds using spectral match. Method transfer to UHPLC method A UHPLC method was established for the separation of acaricides using the Agilent method translator. This tool enables one to easily convert methods from either binary or quaternary pump systems to optimized methods for the Agilent 129 Infinity LC system. The analysis time was decreased to 2 utes using a short sub-2-µm column, see Figure 7. The benefits of the fast analysis are: Time savings versus conventional of 92% Solvent savings versus conventional of 92.8% In Figure 7, all peaks are baseline separated and show sufficient resolution and good peak shape. Partial validation of ultra fast method The following parameter were tested Limit of detection and quantitation using dilution 8 RSD of retention times and areas for dilution 3, 4, and 5 The RSD for retention times over 38 runs was <.12%, which is comparable to results obtained for the conventional method. The RSD of areas was typically <.9% for amounts < 6 to < 1 ng injected amount. Even though only 1.5 µl were injected, the LD was.2.6 ng injected amount and the LQ (1*LD) <.6 injected amount. This is overall 3 times better than measured for the conventional method. 6

mau 8 6 4 2 6 8 1 12 14 16 18 Figure 6 Analysis of pesticides in honey in a concentration range of.5 mg/kg and identification through UV spectral comparison. Conclusion A conventional method for the analysis of,,,, and in honey was developed and validated using the Agilent 126 Infinity Binary LC system. The method is robust and suitable for the quantitation of concentrations of acharicides <.5 mg/kg in honey. The extraction method is based on the extraction of the compounds with Acetonitrile and is fast and simple to perform. Recovery rates between 67 and 88% are typically achieved for the.5 mg/kg range. The LD is typically <.2 ng injected amount. Faster results with significant decrease in solvent consumption and time can be achieved by applying an UHPLC method using the Agilent 129 Infinity LC system. In addition, lower limits of detection can be achieved. mau 12 1 8 6 4 2 2 4 6 8 1 12 14 16 18 Figure 7 Comparison of conventional method versus ultra fast method. 7

References 1. Committee for veterinary medicinal products, (Bees), http://www. ema.europa.eu/docs/en_gb/document_library/maximum_residue_ Limits_-_Report/29/11/ WC51419.pdf 6. Sabine Adamczyk, Regina L azaro, Consuelo P erez-arquillu e, Antonio Herrer, Deteration of synthetic acaricides residues in beeswax by high-performance liquid chromatography with photodiode array detector, Analytica Chimica Acta 581 (27) 95 11 2. Residues of veterinary drugs in honey and possible approaches to derive MRLs for this commodity ftp://ftp.fao. org/docrep/fao/11/i659e/i659e1. pdf 3. Anne-Claire Martel, Sarah Zeggane, Deteration of acaricides in honey by high-performance liquid chromatography with photodiode array detection, Journal of Chromatography A, 954 (22) 173 18 4. E. Korta, A. Bakkali, L.A. Berrueta, B. Gallo*, F. Vicente, Study of semiautomated solid-phase extraction for the deteration of acaricide residues in honey by liquid chromatography, Journal of Chromatography A, 93 (21) 21 29 5. Jae Jeong Ryoo, Se Hoon Kim, Young Han Jeong, Hwan-Suk Do, Jeong Eun Ryu,Hae Young Kwon, Ji Yoon Jeong, Hye Jin Park, Soon Ho Lee, Moo Ki Hong, and Jongki Hong, Simultaneous Deteration of,,, Cymiazole and 2,4-Dimethylaniline in Korean Honey Samples by High-Performance Liquid Chromatography, Bull. Korean Chem. Soc. 28, Vol. 29, No. 5, 143ff www.agilent.com/chem/lc Agilent Technologies, Inc., 211 Published in USA, September 1, 211 Publication Number 599-8717EN