Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Similar documents
Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Analysis Repeatability of Trace and Major Elements in a Water Sample

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Analysis of domestic sludge using the Agilent 4200 MP-AES

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

ICP-OES Application Note Number 35

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Trace Analyses in Metal Matrices Using the ELAN DRC II

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Samples compliant with the WATER

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Fergus Keenan Shona McSheehy Julian Wills

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

Lecture 7: Atomic Spectroscopy

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Prof. Dr. Biljana Škrbić, Jelena Živančev

PRINCIPLE OF ICP- AES

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

OES - Optical Emission Spectrometer 2000

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

Agilent 8800 Triple Quadrupole ICP-MS: Understanding oxygen reaction mode in ICP-MS/MS

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS.

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Environmental Series - Trace contaminant analysis in brine using a Thermo Scientific icap 6000 Series Duo ICP

Application Note. Abstract. Authors

Partial Energy Level Diagrams

Agilent ICP-MS Journal September 2009 Issue 39

Transcription:

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System Application Note Foods Author Glenn Woods ICP-MS Specialist Agilent Technologies UK Ltd. Lakeside Business Park Cheadle Royal Cheshire SK8 3GR United Kingdom Abstract A simple methodology for the measurement of trace and ultratrace metals in edible oils is described. A selection of seed, olive, and nut oils were measured with a spike recovery being performed on a rapeseed oil recoveries of between 94% (Ag) and 107% (Sn) were achieved. Each sample was analyzed directly using an Agilent 7500cx ICP-MS with Octopole Reaction System (ORS) technology following a simple 3x dilution in kerosene. The ORS is used to remove plasma- and matrix-based interferences within a single acquisition, greatly improving the detection limit performance for the interfered-with elements, such as Cr, Fe, and Mg. The detection limits (DLs) and background equivalent concentrations (BECs) were in the µg/kg (ppb) to ng/kg (ppt) range for all elements. An 8½-hour stability run on a spiked (~1 µg/kg) rapeseed oil was performed and precision was found to be approximately % RSD for all elements except Na (4%), which was partially influenced by contamination.

Introduction Edible oils form an essential part of our daily diet. They are used in their native form for high-temperature cooking (for example, frying); as an ingredient for texture, flavor, and nutritional improvers; and as flavor carriers for other ingredients. Edible oils also have very wide usage as a processed foodstuff, for example, in their hydrogenated form as nondairy spreads and food additives. In their processed form, they are also gaining increased usage as a renewable biofuel [1]. Although the overall quality of the product is generally defined by its culinary benefits, the inorganic content of these oils has a very important role in terms of food safety and general product longevity. The mechanisms whereby inorganic constituents are incorporated into the oil include the natural uptake and preconcentration of the element by the plant, contamination during harvesting/processing (for example, from foreign bodies during harvesting or wear metals in the press), and addition or adulteration at some point. As well as the obvious importance of monitoring these elements for safety or nutritional reasons, it has been reported [] that some trace elements can act as autoxidation accelerators, altering the flavor and quality of the product over time. Some traditional approaches to elemental analysis of these materials include inductively coupled plasma optical emission spectroscopy (ICP-OES), atomic absorption spectrometry (FAAS/GFAAS), and ion chromatography (IC). These techniques each suffer severe interferences from the organic matrix; for this reason it is usual to digest the sample or perform an aqueous extraction to eliminate the organic matrix. Whilst digestion/extraction procedures are generally well developed, they are time consuming, often require additional equipment, and can introduce potential errors (for example, analyte loss through incomplete extraction or elemental volatility and contamination from reagents/vessels). In addition, many of these techniques do not offer sufficient sensitivity or detection limits for the measurement of important toxic elements. This procedure describes a simple dilution approach for sample preparation and analysis by ICP-mass spectrometry (ICP-MS). The plasma- and matrix-based interferences are efficiently attenuated using an Octopole Reaction System (ORS) operating in collision and reaction modes. Experimental Standards and internal standards were prepared by weight/weight dilution from ~1,000 mg/kg metallo-organic oils (Spex Certiprep, Metuchen, New Jersey, USA and Conostan, Conoco-Phillips, Bartlesville, OK, USA) and diluted in kerosene (Purum, Fluka Sigma-Aldrich, St. Louis, USA). Internal standards (Li, In, and Bi) were added to all samples and standards prior to analysis to compensate for viscosity differences and to allow nonmatrix-matched standards to be used; samples were prepared by simple 3x dilution in kerosene using weight/weight preparation. Three seed oils (rape, sunflower, and corn), four olive oils (Spanish, Greek, Italian, and Sicilian), and three nut oils (hazel, almond, and groundnut) were investigated. An Agilent 7500cx ICP-MS fitted with an ORS for removal of polyatomic interferences was used for this study. The instrument was operated using hydrogen, helium, and no-gas cell modes. All modes were acquired sequentially during a single visit to the sample vial. Instrumental conditions are given in Table 1. The ORS operating in He collision mode removes polyatomic interferences using a process known as kinetic energy discrimination (KED). The larger interfering species experience more collisions with He atoms and lose energy as they pass through the cell. An energy differential is applied to prevent the lower energy polyatomic ions from entering the mass filter. Optional reaction mode using H gas was used to remove the intense plasma- and matrix-based species such as 14 N and 1 C 16 O on 8 Si, 38 Ar 40 Ar on 78 Se, 40 Ar on 40 Ca, and 40 Ar 1 C on 5 Cr. The interference is neutralized or converted to another species by reaction. For interference-free analytes, the cell can be operated in either He collision or no-gas mode, that is, with no gas added to the ORS cell. Oxygen was added to the plasma (as a 50% O blend in Ar) by an additional mass flow controller to prevent carbon from condensing on the interface and ion lenses; this was used in conjunction with the organic solvent introduction kit and 1.5 mm id injector torch. Note that, while a 50% O in argon mix was used for this application, a 0% O in argon mix is the preferred option, as this provides more precise control of the O flow rate, which may be critical in certain applications, such as lowlevel S and P analysis. Table 1. 7500cx ICP-MS Operating Conditions Forward power Plasma gas Auxiliary gas Sampling depth Carrier gas Oxygen (50% O in Ar) ORS Cell Gas Flow Rates Helium (He) Hydrogen (H ) 1550W 15 L/min 1 L/min 8 mm 0.9 L/min 0.1 L/min 5.7 ml/min 6.5 ml/min

Results and Discussion Table displays the detection limit (DL) and background equivalent concentration (BEC) for the kerosene standard blank. The BEC is an indication of the calibration offset expressed as a concentration (due to elemental contamination in the kerosene matrix). The DLs and BECs were in the µg/kg (ppb) or ng/kg (ppt) range for all elements measured. Table. Detection Limits (3s) and Background Equivalent Concentrations for the Standard Blank Solution (Data reported as µg/kg [ppb].) Element Mass Mode DL BEC Be 9 No gas 0.0153 0.034 B 10 No gas 0.344 4.63 Na 3 No gas.99 18.8 Mg 4 H 1.37 8.9 Si 8 H 5.17 53.3 P 31 He 10 5 K 39 H 0.649 1.34 Ca 40 H 0.568.95 Ti 47 He 0.15 0.0396 V 51 He 0.0198 0.0435 Cr 5 H 0.0935 0.077 Mn 55 He 0.049 0.057 Fe 56 He 0.0447 0.19 Co 59 He 0.0113 0.045 Ni 60 He 0.05 0.0367 Cu 63 He 0.055 0.681 Cu 65 He 0.073 0.675 Zn 66 He 0.0393 0.0685 As 75 He 0.0896 0.19 Sr 88 No gas 0.0335 0.11 Mo 95 No gas 0.5 0.411 Ag 107 No gas 0.0374 0.083 Cd 111 No gas 0.11 0.138 Sn 118 No gas 0.0173 0.0901 Sb 11 No gas 0.061 0.087 Ba 137 No gas 0.047 0.145 W 18 No gas 0.0111 0.031 Hg 01 No gas 0.0147 0.107 Pb 08 No gas 0.0074 0.0595 In order to test the cell's performance in gas mode, several interfered elements were also measured with no gas in the cell. The effect of the 40 Ar 1 C interference can be seen in Figures 1a and 1b ( 5 Cr determined in H reaction mode and no-gas mode): the BEC has been reduced from over 80 ppb to less than 0.08 ppb. Figures a and b (Mn determined in He collision mode and no-gas mode) demonstrate the removal of the 38 Ar 17 O (interference due to oxygen addition) and 40 Ar 15 N (from air entrainment and residual nitrogen in the matrix) both minor polyatomics but nevertheless present at mass 55. The BEC is reduced from > 1. ppb without cell gas to ~0.05 ppb in gas mode. 6 5 4 3 1 0 5 Cr (H mode) R = 0.9999 0 10 0 30 40 50 60 70 80 90 100 Figure 1a. 1600 1400 100 1000 800 600 400 00 Cr (H reaction mode). Note excellent curve fit and minimal calibration offset (BEC = 0.077 ppb), indicating efficient interference (ArC) removal. 5 Cr (no gas mode) R = 0.3708 0 0 10 0 30 40 50 60 70 80 90 100 Figure 1b. 3.5 3.0.5.0 1.5 1.0 0.5 Cr (no-gas mode). Note very poor curve fit due to extremely high calibration offset (approximate BEC 83 ppb) due to the intensity of the ArC interference. 55 Mn (He mode) R = 1.0000 0.0 0 1 3 4 5 6 7 8 9 Figure a. Mn (He collision mode). Low BEC (0.053 ppb) indicates efficient interference (ArO, ArN) removal. 10 9 8 7 6 5 4 3 1 0 Figure b. 55 Mn (no gas mode) R = 0.9997 0 1 3 4 5 6 7 8 9 Mn (no-gas mode). Small but significant offset most likely due to 38 Ar 17 O and 40 Ar 15 N interferences equivalent to 1.3 ppb BEC. 3

Table 3 displays the quantitative results obtained for the samples, including spike recovery data for the rapeseed sample. Recoveries for all elements ranged from a low of 93.7% for silver to a high of 107.1% for tin; generally, the recoveries for all elements were excellent, considering that the standards were not matched exactly for viscosity. Overall, the nut oils had higher elemental levels than the other oils studied, which is probably an indication of the different biological pathways of the plants rather than any specific contamination during production. Instrumental stability was tested over a period of 8.5 hours using a spiked (~1 µg/kg) rapeseed oil. Figure 3 displays the normalized stability plot. Precision was found to be around % for most elements (Na RSD was higher at 4%, most likely due to some airborne contamination occurring during the run), which clearly demonstrates the applicability of the technique to routine analysis of this organic matrix. Table 3. Quantitative Results for the Oil Materials (Corrected for Dilution) (All results presented as µg/kg. spike recoveries are presented for the rapeseed material.) Seed oil Olive oil Nut oil Rapeseed spike recovery Sunflower Rape Corn Spanish Greek Italian Sicilian Groundnut Hazel Almond Spk amt Measured % Rec Be 9 No gas 0.0675 1. 0.4549 0.4394 0.115 0.0591 0.0747 0.0994 0.0089 0.0734 16.3 17.84 10.1 B 10 No gas 15.36 1.13 N/D N/D N/D N/D.89 N/D 41.53 3.53 1.4 14.33 106.5 Na 3 No gas 131.1 168. 488. 199. 441.6 13.6 304.9 160 73.9 938 Spike too low Mg 4 H N/D N/D 74.58 N/D N/D 43.43 N/D 98.9 1740 8133 1.4 1.07 97.3 Si 8 H 45.9 N/D 43.98 1.486 356.7 404.7 481.7 51.9 37.6 19.8 1.4 1.78 103.1 P 31 He 38.4 910 545.4 164.8 51. 31.3 115.4 9196 11410 60860 Spike too low K 39 He 53.5 155.03 139.5 100.4 144.1 186.6 08.3 1031 415 15540 Spike too low Ca 43 He 16.49 10.43 5.54 7.88 9.14 5.31 8.57 6.46 68.7 1785 1.4.81 99.9 Ti 47 He 0.6974 0.4051 0.9066 0.3111 1.1 0.168 1.681 0.9585 80.1 3.568 1.4 1.6 98.5 V 51 He 0.116 0.0383 0.068 N/D N/D N/D N/D 0.0034 1.07 0.1468 1.4 13.01 104.6 Cr 5 H 1.671 1.95 1.515 1.043.918.043 3.048 1.06 5.09.681 1.4 13.69 100.0 Mn 55 He 0.7513.905.5 1.513 1.069 0.8085 0.3445 9.736 341.5 170.9 1.4 15.67 10.9 Fe 56 He 16.81 0.03 13.3 18.79 1.9 14.36 1.04 6.47 83.09 358. 1.4 3.64 101.7 Co 59 He 0.46 0.8684 13.89 4.869 3.185 1.355 0.5555 1.75 0.0679 1.47 14.8 15.38 98.1 Ni 60 He 0.403.841.83 0.398 0.3685 0.13 0.1193 0.9994 1.19 0.6101 1.4 16.01 106. Cu 63 He 5.60.03 5.351 1.56 10.3 6.393 4.11.137.538 5.084 1.4 34.6 98.6 Zn 66 He 1.9.164 4.415.867 3.934 4.035 0.5509 1.73 5.54 119.9 1.4 15.06 104.0 As 75 He 0.73 0.434 4.467.35 0.6854 0.1878 0.3433 5.95 0.6574 1.4 17.4 18.4 103.3 Se 78 H 1.187 0.8585 3.090 0.584 1.8861 N/D 3.414 0.58.305 1.836 1.4 13.54 10.7 Sr 88 No gas 0.4056 0.688.44 1.584 0.539 0.6583 0.14 1.417 10.54 119.9 15.4 16.57 103.1 Ag 107 No gas 0.43 0.373 0.934 0.3644 0.0758 0.177 0.09 0.048 N/D 0.616 1.4 11.95 93.7 Cd 111 No gas 0.6787.054 0.6848 1.50 0.3816 0.396 0.795 N/D N/D N/D 1.4 14.58 101.0 Sn 118 No gas N/D N/D 0.949 N/D N/D 0.1151 N/D N/D 77.1 0.331 1.4 13.8 107.1 Ba 137 No gas 0.509 0.466 1.11 0.5088 0.098 0.1067 N/D 3.019 9.43 76.7 1.4 13.54 105.5 W 18 No gas N/D N/D 4.366.314 0.3513 N/D N/D 0.3609 N/D 0.165 1.4 1.93 104.3 Hg 01 No gas 0.5693 0.465 0.579 0.3813 0.6913 0.5076 0.774 0.344 0.3807 0.3391 Not in spike mix Pb 08 No gas 0.11 0.608 0.5475 0.45 0.1735 0.1308 0.3311 0.6437 0.4816 1.61 1.4 13.46 103.6 4

Normalized Recovery.0 1.8 1.6 1.4 1. 1.0 0.8 0.6 0.4 0. 0.0 Na P S Ti V Cr Mn Fe Co Ni Cu Zn As Sr Mo Ag Cd Sn Sb Ba W Pb 1:00:00 AM 1:13:00 AM 1:7:00 AM 1:41:00 AM 1:55:00 AM 1:09:00 AM 1:3:00 AM 1:37:00 AM 1:50:00 AM :04:00 AM :18:00 AM :3:00 AM :46:00 AM 3:00:00 AM 3:14:00 AM 3:8:00 AM 3:41:00 AM 3:55:00 AM 4:09:00 AM 4:3:00 AM 4:37:00 AM 4:51:00 AM 5:05:00 AM 5:18:00 AM 5:3:00 AM 5:46:00 AM 6:00:00 AM 6:14:00 AM 6:8:00 AM 6:4:00 AM 6:55:00 AM 7:09:00 AM 7:3:00 AM 7:37:00 AM 7:51:00 AM 8:05:00 AM 8:19:00 AM 8:33:00 AM Time Figure 3. Normalized stability plot for a spiked (~1 µg/kg) rapeseed oil. Repeat measurements were taken over a period of 8 hours 33 minutes. Conclusions The measurement of trace and ultratrace elements in edible oils by 7500cx (ORS) ICP-MS is routine, sensitive, and selective. The ORS efficiently attenuates matrix- and plasma-based interferences to concentrations below background contamination, providing ppt-level detection performance for almost all elements. Furthermore, the simplicity of the method setup and sample preparation lends itself to a routine environment, as very little sample workup or instrument optimization is required. The result is improved productivity and a reduced likelihood of preparation errors. References 1. Agilent Technologies publication, Direct Elemental Analysis of Biodiesel by 7500cx ICP-MS with ORS, 5989-7649EN, April 17, 008. Eunok Choe and David Min, Mechanisms and Factors for Edible Oil Oxidation, Comprehensive Reviews in Food Science and Food Safety, 5, p. 169 186, 006 For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 008 Published in the USA December 9, 008 5989-9888EN