Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa

Similar documents
Final Report on Key Comparison EURAMET.M.P-K13 in the Range 50 MPa to 500 MPa of Hydraulic Gauge Pressure

High pressure comparisons between seven European National Laboratories

High pressure comparison among seven European national laboratories

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

EURAMET.M.P-S9 / EURAMET 1170, LOOP2 Comparison in the negative gauge pressure range -950 to 0 hpa FINAL REPORT

Interferometric determination of thermal expansion coefficient of piston/cylinder unit preliminary investigation

Final report. T. Priruenrom 1), W. Sabuga 2) and T. Konczak 2) March 2013

National Physical Laboratory New Delhi

Final report Comparison Identifier: APMP.SIM.M.P-K1c Other designation: NIST-NPLI/Pressure/2

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

An Examination of the Uncertainty in Pressure of Industrial Dead-Weight Testers Used For Pressure Calibrations in Different Environments

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

A Method for Verifying Traceability in Effective Area for High Pressure Oil Piston-Cylinders

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

EURAMET.M.P-S9 / EURAMET 1170, LOOP1 Comparison in the negative gauge pressure range -950 to 0 hpa FINAL REPORT

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Physikalisch- Technische Bundesanstalt

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

PTB S 16.5 MN HYDRAULIC AMPLIFICATION MACHINE AFTER MODERNIZATION

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge

Certification of a High Capacity Force Machine for Testing of Load Cells According to OIML R60

EURAMET Project no Inter-comparison of a 1000 L proving tank

THE PURPOSE AND IMPLEMENTATION OF A MULTI-PISTON PRESSURE CALIBRATION CHAIN

COOMET.M.V-S2 (587/RU-a/12) COOMET SUPPLEMENTARY COMPARISON IN THE FIELD OF MEASUREMENTS OF LIQUIDS KINEMATIC VISCOSITY

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

Euramet project 1187 Comparison of Instrument Current Transformers up to 10 ka. Technical protocol (March 2012)

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

Comparison protocol EURAMET project

Portuguese Institute for Quality. Contents

Low gauge pressure comparisons between MIKES, Metrosert and FORCE Technology

Calculation of effective area A 0 for six piston cylinder assemblies of pressure balances. Results of the EUROMET Project 740

Ajchara Charoensook, Chaiwat Jassadajin. National Institute of Metrology Thailand. Henry Chen, Brian Ricketts and Leigh Johnson

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

INVESTIGATION OF THE CALIBRATION CAPABILITIES OF A 1 KN M AND 10 KN M REFERENCE CALIBRATION MACHINE

MASS AND VOLUME COMPARISONS AT MIKES

Euramet project 1075 Bilateral comparison

MODEL FOR UNCERTAINTY ESTIMATION IN COMPARISON CALIBRATION OF THERMOCOUPLES

Centre for Metrology and Accreditation, MIKES

ANNEXE 8. Technical protocols for the interlaboratory comparisons

Increased Requirements for Higher Nominal Forces, Necessities and Possibilities to Measure Them

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

A capacitive probe for measuring the clearance between the piston and the cylinder of a gas piston gauge

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

Physikalisch-Technische Bundesanstalt

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Measurement & Uncertainty - Basic concept and its application

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Comparison of V and 10 V DC Voltage References

Appendix B1. Reports of SMU

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

Final Report on Key Comparison APMP.M.P-K13 in Hydraulic Gauge Pressure from 50 MPa to 500 MPa

EUROMET Project 702 EUROMET.M.D-K4

BIPM/CIPM key comparison CCM.FF-K Final Report for Volume of Liquids at 20 L and 100 ml - Piloted by Centro Nacional de Metrología (CENAM)

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

CERTIFICATE OF ACCREDITATION

Final Report on the Torque Key Comparison CCM.T-K2 Measurand Torque: 0 kn m, 10 kn m, 20 kn m Dirk Röske 1), Koji Ogushi 2)

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Document No: TR 12 Issue No: 1

Bilateral comparison on micro-cmm artefacts. between PTB and METAS. Final report

TOLERANCES AND UNCERTAINTIES IN PERFORMANCE DATA OF REFRIGERANT COMPRESSORS JANUARY 2017

INVESTIGATION OF TRANSFER STANDARDS IN THE HIGHEST RANGE UP TO 50 MN WITHIN EMRP PROJECT SIB 63 Falk Tegtmeier 1, Michael Wagner 2, Rolf Kumme 3

Traceable Mass Determination and Uncertainty Calculation

Report on EURAMET.M.M-S9

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

Available online at ScienceDirect. Andrej Godina*, Bojan Acko

Measurement & Uncertainty - Concept and its application

Final Report on Key Comparison APMP.M.P-K7 in Hydraulic Gauge Pressure from 10 MPa to 100 MPa

R SANAS Page 1 of 7

Can there be a metrology for psychometricians?

COMPLETION AND MEASUREMENT UNCERTAINTY BUDGET OF THE MULTI-COMPONENT MEASURING DEVISE FOR FORCE UP TO 1 MN AND TORQUE UP TO 2 KN M

PROTOCOL OF MASS AND VOLUME COMPARISONS BETWEEN SIM NMIs

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Technical protocol for the CCL-K11 key comparison of optical frequency/wavelength standards.

CMM Uncertainty Budget

First International Involute Gear Comparison

Automated volume measurement for weihts using acoustic volumeter

Study on a thermal diffusivity standard for the laser flash method measurements 1

The dimensional calibration of piston-cylinder units to be used for pressure metrology and the re-determination of the Boltzmann constant

Final Report on the EURAMET.PR-K1.a-2009 Comparison of Spectral Irradiance 250 nm nm

A bilateral comparison of a 1 kg platinum standard

Physikalisch Technische Bundesanstalt

Force Key Comparison EUROMET.M.F-K2. (50 kn and 100 kn) EURAMET Project No 518. Final Report. 14 July Pilot: NPL, United Kingdom

Uncertainty restrictions of transfer click torque wrenches

ENCAPSULATION AND UNCERTAINTIES OF STRAIN-GAUGE SENSORS FOR STRESS-MONITORING OF CONSTRUCTIONS

EURAMET Project 858: Hydrostatic weighing exchange of experiences. Project outline

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

Key Words: pressure standard, controlled-clearance piston gauge, effective area, uncertainty

g/cm³ DAkkS Calibration Laboratory for Temperature and Density High-precision calibration for temperature and density

Final report Results and evaluation of key comparison CCM.P-K12.1 for very low helium flow rates (leak rates)

Multi-Capacity Load Cell Concept

Transcription:

EURAMET - THE EUROPEAN ASSOCIATION OF NATIONAL METROLOGY INSTITUTES Comparison Identifier: EURAMET 1414 (Registered in KCDB as EURAMET.M.P-S18) Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa L. Grgec Bermanec 1, D. Pantić 2, B. Ramač 2 1 2 HMI/FSB-LPM Croatian Metrology Institute, Faculty of Mechanical Engineering and Naval Architecture- University of Zagreb, Laboratory for Process Measurement, Croatia Directorate of Measures and Precious Metals (DMDM) of the Republic of Serbia Laboratory for mass, pressure and force, Serbia 1

Abstract Bilateral comparison was organized between the Laboratory for Process Measurement of the Croatian Metrology Institute (HMI/FSB-LPM) and the Pressure Laboratory of the Directorate of Measures and Precious Metals of the Republic of Serbia (DMDM). Laboratory for Process Measurement of HMI acted as the pilot laboratory. The aim of the comparison was to evaluate the degree of equivalence in the determination of effective area and elastic distortion coefficient, considering respective uncertainties of the two laboratories. Measurements were done on the pressure balance in gauge mode, with oil as transmitting medium, in the gauge pressure range 10-80 MPa. The results of the comparison successfully demonstrated that the hydraulic gauge pressure standards are equivalent within their claimed uncertainties. 2

Contents: 1. Introduction... 4 2. Participating Laboratories... 4 3. Description of the Laboratory standards and traceability... 4 3.1 LPM Laboratory standard... 4 3.2. DMDM Laboratory standard... 6 4. Transfer standard... 7 5. Circulation of the transfer standard... 8 6. Calibration methods and uncertainty evaluation... 9 6.1. DMDM calibration method... 9 6.2. LPM calibration method... 11 7. Results... 13 8. Conclusions and discussion... 14 9. References... 15 3

1. Introduction At the EURAMET TC-M meeting in 2015, it was agreed to carry out a bilateral supplementary comparison (SC) between LPM and DMDM of their hydraulic gauge pressure standards in the range up to 80 MPa. The comparison in that range was motivated by the fact that, in both National Metrology Institutes (NMI s), their pressure balances provide traceability and that DMDM does not have CMC entries published in the BIPM database. DMDM provided a transfer standard (TS) for this comparison. A pressure balance, with nominal accuracy of 0.015%, was circulated between the two laboratories. The measurand was the effective area of the pistoncylinder, in gauge mode. This SC is identified as EURAMET.M.P-S18 in the BIPM key comparison database. 2. Participating Laboratories Details about participating laboratories and responsible persons are given in Table 1. Table 1: Details about participants Laboratory Address Person responsible Croatian Metrology Institute (HMI), Faculty of Mechanical Engineering and Naval Architecture (FSB) - University of Zagreb Laboratory for Process Measurement (LPM) Pilot Laboratory Directorate of Measures and Precious Metals (DMDM) of the Republic of Serbia Laboratory for mass, pressure and force Participating laboratory Ivana Lučića 5 Zagreb Croatia Mike Alasa 14, 11000 Beograd Serbia Prof. dr. sc. Lovorka Grgec Bermanec Tel : 01 6168 488 e-mail : lovorka.grgec@fsb.hr Dragan Pantić, dipl.ing. Tel : +381-11-20-24-417 e-mail : pantic@dmdm.rs 3. Description of the Laboratory standards and traceability 3.1 LPM Laboratory standard LPM has been developing temperature, pressure and humidity standards for more than 40 years. More than 15 years' traceability of pressure measurements is achieved through the German national pressure laboratory PTB (Physikalisch-Technische Bundesanstalt). To maintain compliance with MRA, LPM participated in several international comparisons (through DKD Deutscher Kalibrierdienst, now DAkkS, Deutsche Akkreditierungsstelle and EURAMET) with other NMIs to support calibration measurement capabilities (CMCs). First, cross-float was validated with PTB measurements for accreditation purposes in 2002. LPM successfully participated in two key comparisons through EURAMET namely: EURAMET.M.P.-K8 (up to 0.2 4

MPa) [10] and EURAMET.M.P-K1c, (up to 7 MPa) as well as in EURAMET project 1125 whose purpose was to compare approaches to effective area and uncertainty estimation [1]. CMC entries of LPM were published in BIPM database in 2008. LPM standard used in this SC was a commercially manufactured oil pressure balance with details described in Table 2 and Figure 1. Zero-pressure effective area,a 0, and pressure distortion coefficient () were determined by cross-floating method in PTB in 2006, 2011 and 2017. Table 2: Details of the LPM laboratory standard Manufacturer / Type Measurement range Pressurements / P7000 Series 3 to 140 MPa PCU S/N X 0467 Nominal area of the piston 4 10-6 m 2 Piston-cylinder unit thermal coefficient Tungsten carbide + Steel 16.6 10-6 C -1 Total mass 60 kg Weights set S/N 59689 Pressure reference level Working liquid used Bottom of the piston at crossfloating position sebacate Figure 1: LPM piston-cylinder unit X 0467 5

3.2. DMDM Laboratory standard The properties of the DMDM pressure standard and measurement conditions are presented in Table 3. All uncertainties in the table are standard ones. DMDM pressure balance and pistoncylinder unit are shown in Figure 2. Laboratory standard of DMDM is traceable to Swedish National Laboratory (SP, nowadays RISE). A 0 and were determined by cross-floating method in SP in 2015. Table 3: Details of the DMDM laboratory standard Manufacturer Desgranges et Huot 5303 S CP PCU S/N 2496 Measurement range 2 to 80 MPa Material of piston Tungsten carbide Material of cylinder Tungsten carbide Operation mode Free-deformation Zero-pressure effective area (A 0 ) at reference temperature in mm 2 4.90287 Relative uncertainty of A 0 in 10-6 19 Pressure distortion coefficient () in Pa -1 3.06 10-13 Uncertainty of in MPa -1 0.91 10-13 Relative uncertainty of mass pieces in 10-6 0.75 Linear thermal expansion coefficient of piston (α p ) in ºC -1 4.5 10-6 Linear thermal expansion coefficient of cylinder (α c ) in ºC -1 4.5 10-6 Local gravity (g) in m/s 2 9.80582 Relative uncertainty of g in 10-6 25 Height difference between pressure balances in mm -45 Uncertainty of h in mm 1 Weights set model 40 kg Weights set S/N 1927 Pressure reference level Bottom of the piston at cross-floating position Direction of the rotation CCW Working liquid Di(2)-ethyl-hexyl-sebacate (DHS) Rotation speed 30 rpm 6

Figure 2: DMDM pressure balance and piston-cylinder unit 4. Transfer standard The transfer (TS) standard that was circulated between the two laboratories is a pressure balance manufactured by Desgranges et Huot, consisting of a base, a piston-cylinder unit (PCU) and a set of weights. The transfer standard was provided by DMDM. Table 4 summarizes the basic characteristics of the transfer standard. 4.5 10-6 4.5 10-6 Table 4: Technical characteristic of transfer standard Manufacturer Desgranges et Huot Pressure balance model 5304 N CP Pressure balance S/N 3617 PCU model HUILE PCU S/N 8755 Nominal area of the piston 4.9 10-6 m 2 Linear thermal expansion coefficient of piston (α p ) in ºC -1 Linear thermal expansion coefficient of cylinder (α c ) in ºC -1 Weights set model 50 kg Weights set S/N 4159 Pressure reference level Bottom of the piston at cross-floating position Working liquid Di(2)-ethyl-hexyl-sebacate (DHS) Direction of the rotation CCW Rotation speed 30 rpm Piston circumference 0.007917 m 7

The DMDM provided calibration certificates for the following masses: discs (mass set), piston and mass carrying bell. The mass set consists of 8 mass pieces listed in the Table 5. The material density of the mass pieces is m = (7920 140) kg/m 3. Table 5: Masses of transfer standard Nominal value Marking Conventional mass in g 4 kg 4159 4 kg 4000.000 5 kg 4159 1 5kg 5000.008 5 kg 4159 2 5kg 5000.018 5 kg 4159 3 5kg 5000.013 5 kg 4159 4 5kg 5000.003 5 kg 4159 5 5kg 4999.998 5 kg 4159 6 5kg 5000.003 5 kg 4159 7 5kg 4999.998 Conventional mass for piston and mass carrying bell are 200.0254 g and 800.0065 g, respectively. 5. Circulation of the transfer standard The measurements with the transfer standard were performed in accordance with the schedule given below. The time allocated for each laboratory for the measurements was three weeks. One week was also allocated for transportation of the artifact. The equipment to be circulated was separated in two packages, under the same ATA-carnet, and there were no problems during the comparison. Table 6: Measurement schedule Country NMI Period Responsible Croatia FSB-LPM December 2015 Serbia DMDM January 2016 Lovorka Grgec Bermanec Pantić Dragan and Boris Ramač 8

6. Calibration methods and uncertainty evaluation According to the Technical protocol [2] it was required that each laboratory finds the effective area of TS at the following nominal pressures: 100 bar, using weights 4 kg, piston and weight carrier 200 bar, using weights 4 kg, 1, piston and weight carrier 300 bar, using weights 4 kg, 1, 2, piston and weight carrier 400 bar, using weights 4 kg, 1, 2, 3, piston and weight carrier 500 bar, using weights 4 kg, 1, 2, 3, 4, piston and weight carrier 600 bar, using weights 4 kg, 1, 2, 3, 4, 5, piston and weight carrier 700 bar, using weights 4 kg, 1, 2, 3, 4, 5, 6, piston and weight carrier 800 bar, using weights 4 kg, 1, 2, 3, 4, 5, 6, 7, piston and weight carrier Calibration methods used in both laboratories are based on [3] with specific details described in the following sections. 6.1. DMDM calibration method During all measurements both the laboratory and the transfer standard were operated with their piston-cylinder assemblies rotating counter clockwise (CCW) and at almost constant rotation rate by means of their motor-drivers. PCU temperature was measured by a platinum resistance thermometer (PRT) built into the laboratory standard mounting post, traced to the international temperature scale (ITS-90) at DMDM Temperature Laboratory. Its standard measurement uncertainty was 0.05 C. DMDM Mass Laboratory calibrated mass of the weights of the laboratory and the transfer standard with a relative standard uncertainty of 0.75 10-6. Position and fall rates of the laboratory and transfer standard pistons were measured by Float Position Sensors connected to Ruska 2456 Piston Gauge Monitor. Fall rates data acquisition was done with a WinPrompt software and a specific DMDM software program that perform processing of image taken by a simple camera placed in front of window for the piston position display. Each fall rate measurement has been made at each pressure at temperature around 20 C, laboratory and transfer standard were cross-floated using the fall rates of both pistons as an equilibrium criterion. To reach the equilibrium trim masses were applied to both pressure balances. DMDM calibration setup is shown in Figure 3. 9

Figure 3: DMDM calibration setup The air density was calculated from measured temperature, pressure and humidity by following equation ρ a = 0.34848 p 0.009024 h e0.0612 t 273.15+t (1) This equation has a relative uncertainty of 2 10 4 in the range 900 hpa < p < 1100 hpa, 10 C < t < 30 C and hr < 80 %. Atmospheric pressure was measured with resolution of 1 Pa and standard uncertainty of 3 Pa. Laboratory temperature was measured with resolution of 0.01 ºC and standard uncertainty of 0.05 ºC. Relative air humidity was measured with resolution of 0.01 % and standard uncertainty of 0.5 %. All measurements were performed with Vaisala PTU-301 traceable to FSB-LPM for atmospheric pressure and DMD for temperature and relative air humidity. The following density () in dependence on pressure (p) and temperature (t) as well as the surface tension () of DHS were used in the calculations: ρ(p,t) 912.7 0.752(p/MP a) 1.65 10 4 o 3 1 7.8 10 t/ C 20 1 0.01 kg/m =31.2(10.05)mN/m 3 2 6 p/mpa * 1.5 10 p/mpa 3 (2) 10

Uncertainty evaluation was based on ISO GUM and EURAMET procedure. 6.2. LPM calibration method Transfer standard (pressure balance under test) was hydrostatically balanced against the LPM standard of known effective cross-sectional area. The basis of the comparison was determination of the loads at which each balance would individually reach equilibrium of the system at precisely the same pressure. State of equilibrium was identified by rate-of-fall technique, which means that load of each assembly is carefully adjusted by means of trimming weights until both pistons were falling with their natural rates [4]. Fig. 4 shows a common system configuration for a single medium hydraulic cross-float. M M t E E h t T T R S IV IV IV PV Figure 4: LPM calibration setup The pressure balance under test, T, is connected to a standard pressure balance, E, and a pressure generator, S. The loads, M, on both pistons are adjusted until they are hydrostatically balanced at the required pressure. Difference in altitude, h, between the reference levels of the E and T balances as well as the temperatures of the environment, tested and standard piston-cylinder unit are measured. Mathematical model used to calculate pressure balance effective area at zero pressure is known as the P method which requires all loading forces acting on the system to be evaluated and summed, including those due to internal fluid-head and buoyancy effects. For each calibration point the effective pressure, p r, measured at the reference level of the E is calculated, using the known characteristic of the standard piston/cylinder assembly according to equation (3). 11

p r = [ [m i (1 a i )] V ( f a )] g + c mi A 0 (1+λ p) [1+( p + c ) (t t r )] (3) Where: p r is the pressure generated by the reference pressure standard at its reference level, m i are the individual mass values of the weights applied on the piston, including all floating elements, ρ mi are the densities of the weights, ρ a is the density of air, V additional volume due to a free volume, a conical end or a step on the piston, ρ f is the density of the fluid, g is the local gravity, is the surface tension of the liquid, c is the circumference of the piston or its extension at the level where it emerges from the oil, A 0 effective area at null pressure, λ pressure distortion coefficient, p is an approximate value of the measured pressure p r, α p is the linear thermal expansion coefficient of the piston, α c is the linear thermal expansion coefficient of the cylinder, t is the measured temperature of the piston-cylinder assembly during its use, is the reference temperature of the piston-cylinder assembly. t r For each calibration point A p was calculated as: A p = [m i (1 a i )] g cos + c mi (p r +( f a )g h) [1+( p + c ) (t t r )] (4) Where A p h is effective area at the reference temperature t r, of an pressure balance calibrated in gauge mode by a reference pressure balance, is difference between the altitude of the balance reference level and the altitude of the point where the pressure has to be measured (or reference level of the pressure balance under calibration), is the angle of deviation of the piston axis from verticality. Following the ISO [5] and EURAMET [3] recommendations and [6], uncertainties u(a p ) for each calibration point were calculated. 12

7. Results The participating laboratory provided, three weeks after the completion of its measurements, results of the calibration, in the form of an EXCEL sheet, prepared for this purpose by the pilot laboratory. The results to be provided are the values of the effective area A p at each calibration point, as well as the values of A 0 at 20 C and pressure distortion coefficient (λ), all with their respective expanded uncertainties. En values were calculated: E = n /M - M / LPM U +U lab 2 2 lab LPM (5) Where M LPM is measurement result of LPM, M lab is measurement result assigned by DMDM, U LPM is uncertainty of M LPM. U lab is the uncertainty of M lab. Results are regarded as equivalent when the resultant E n for any measurement is between +1 and -1. Table 7: Summary results of the effective area obtained by LPM and DMDM: Combined Combined Nominal Average of A p Average of A uncertainty p uncertainty pressure A p (LPM) A U(LPM) p (DMDM) U(DMDM) MPa mm 2 mm 2 mm 2 mm 2 En 10 4.90240 0.0004 4.90264 0.00043 0.41 20 4.90251 0.0003 4.90256 0.00039 0.10 30 4.90247 0.0003 4.90251 0.00038 0.08 40 4.90245 0.0003 4.90245 0.00037 0.00 50 4.90233 0.0003 4.90237 0.00037 0.08 60 4.90227 0.0003 4.90229 0.00037 0.04 70 4.90214 0.0003 4.90220 0.00037 0.13 80 4.90209 0.0003 4.90212 0.00037 0.06 13

Average effective area (mm 2 ) Final report on EURAMET.M.P-S18 4,9032 4,9030 4,9028 y(dmdm) = -7.35714E-07x + 4.90272E+00 4,9026 4,9024 4,9022 4,9020 DMDM LPM Linear (DMDM) Linear (LPM) 4,9018 4,9016 y(lpm) = -5.642857E-07x + 4.902598E+00 0 100 200 300 400 500 600 700 800 900 Nominal pressure (bar) Figure 5: Average values of effective areas at each nominal pressure, LPM and DMDM To determine zero-pressure effective area A 0 of TS and its distortion coefficient (λ), linear mode based on average reported areas is used for each laboratory: A p = A 0 (1 + λ p) (6) Table 8: Results for zero-pressure effective area and distortion coefficient Value LPM Combined uncertainty Value DMDM Combined uncertainty En A 0 (mm 2 ) 4,902598 0,0003 4,90272 0,0004 0,24 (MPa -1 ) -1.13E-06 3.70E-07-1.51E-06 4.80E-07 0.63 8. Conclusions and discussion The aim of proposed comparison was to evaluate the degree of equivalence in the determination of effective area and elastic distortion coefficient, considering respective uncertainties of the two laboratories. Measurements were done on the pressure balance in gauge mode, with oil as transmitting medium, in the gauge pressure range from 10 up to 80 MPa. With the uncertainty claimed by each laboratory the En values for effective area determination are in the range from 0.01 to 0.41. The E n value for A 0 is 0,24 while it is 0,63 for 14

distortion coefficient. The results of the comparison successfully demonstrated that the hydraulic gauge pressure standards are equivalent within their claimed uncertainties. 9. References [1] P. Otal, I. Morgado, D. Steindl, N. Medina Martín, A. Lefkopoulos, A. Altintas, L. Grgec Bermanec, S. Bursic, J. Setina, M. Bergoglio, I. Spohr, C. Wüthrich, M. Rantanen, C. Vámossy1, W. Sabuga, I. Koçaş; Evaluation of crossfloat measurements with pressure balances Results of EURAMET project 1125. // PTB Mitteilungen. 3 (2011) [2] Technical Protocol for EURAMET bilateral comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa [3] Calibration of Pressure Balances, EURAMET cg-3 Version 1.0, European Association of National Metrology Institutes, 2011. [4] L. Grgec Bermanec, D. Zvizdic, V. Simunovic, Development of Method for Determination of Pressure Balance Piston fall Rate. // ACTA IMEKO The e-journal of the International Measurement Confederation (IMEKO). Vol 3 (2014). [5] BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML, Guide to the expression of uncertainty in measurement (GUM), Geneva, 1995. [6] Zvizdić D.; Grgec Bermanec L.; Bonnier G.; Renaot E.; Veliki T.: Uncertainty Budget of Pressure Balance Effective Area Determined by Comparison Method, Proceedings of XVII IMEKO World Congress, Topic 16 - Pressure and Vacuum Measurement, pp 2003-2007, Dubrovnik, 2003. 15