Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Similar documents
Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analytical Trap Comparison for USEPA Method 8260C

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Methanol Extraction of high level soil samples by USEPA Method 8260C

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Volatile Organic Compounds in Every Day Food

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Maximizing Sample Throughput In Purge And Trap Analysis

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Optimizing. Abstract: is standardd. procedures. altered to

Analysis of BTEX in Natural Water with SPME

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Study of Residual Solvents in Various Matrices by Static Headspace

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Solid Phase Micro Extraction of Flavor Compounds in Beer

Optimizing Standard Preparation

Headspace Technology for GC and GC/MS: Features, benefits & applications

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Solid Phase Micro Extraction of Flavor Compounds in Beer

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Determination of Volatile Organic Compounds in Air

Volatile Hydrocarbons (VH) in Waters by GC/FID

STANDARD OPERATING PROCEDURES

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Enhanced Method Detection Limits for irm-gc/ms of Volatile Organic Compounds Using Purge and Trap Extraction

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

HYDROCARBONS, AROMATIC 1501

METHOD 8015C NONHALOGENATED ORGANICS BY GAS CHROMATOGRAPHY

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

METHOD 8015D NONHALOGENATED ORGANICS USING GC/FID

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

Maximizing Production While Minimizing Costs

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

Blood Alcohol Determination using Static Headspace Analysis with Optimized Sample Throughput

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

GAS CHROMATOGRAPHY (GC)

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Application. Gas Chromatography February Introduction

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Author. Abstract. Introduction

Ed George and Anaïs Viven Varian, Inc.

Certified BTEX in Unleaded Gas Composite

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Thermal Desorption Technical Support

Atlantic RBCA. Guidelines for Laboratories. Tier I and Tier II Petroleum Hydrocarbon Methods

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Application Note 032

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

Transcription:

Abstract Purge and Trap (P&T) is a concentration technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible for trapping the analytes purged from the sample and then releasing them upon heating for further analysis by the GC and GC/MS. This study will evaluate a new analytical trap that is designed by Teledyne Tekmar. The packing material in this new trap improves the recovery and performance of Volatile Hydrocarbon Fractions found in Gasoline Range Organics (GRO) when prepared using Purge & Trap and Headspace sample concentration techniques used in tandem with GC. Introduction Purge and Trap (P&T) concentration is a typical technique used for the analysis of Volatile Organic Compounds (VOCs). The major component of any P&T system is the analytical trap. This trap is responsible for retaining the VOCs during purge and then releasing the analytes upon desorbing. The requirements for an analytical trap are as follows: 1. At low temperatures, it must retain the desired analytes while allowing oxygen and water to pass through unimpeded. 2. It must release the analytes quickly and efficiently upon heating 3. Must not contribute any volatiles of interest to the system. 4. Should have a reasonable price and lifetime. The latest advancement in analytical traps by Teledyne Tekmar is the #11-VPH-Trap. This trap provides the features of Tekmar s #9 analytical trap while modifying the sorbent bed so a lesser amount of methanol is retained. The #11-VPH-Trap is ideal for running the Wisconsin Gasoline Range Organics (GRO) method (e.g. methyl-tert-butyl-ether, toluene, benzene, ethylbenzene, xylenes, 1, 2, 4- trimethylbenzene, 1, 3, 5-trimethylbenzene and naphthalene) and BTEX methods. To evaluate the new #11-VPH-Trap, the analytical trap examined three Tekmar products the Atomx an automated VOC sample prep system (U-Shape), Velocity XPT Purge and Trap (straight trap) with an AQUATek 100 autosampler, and a HT3 headspace analyzer (straight trap).

For the Atomx, a calibration curve by summation, Replicate Lab Control Spikes (RLCS), Duplicate Lab Control Spikes (DLCS), Limit of Detection (LOD) and Limit of Quantitation (LOQ) were determined for water, high-level soil and low-level soil samples. 1 For the Velocity XPT a calibration by summation, RLCS, DLCS, LOQ and LOD were evaluated for water samples only. 1 The HT3 headspace analyzer generated individual calibration curves for all ten compounds listed above. Atomx Automated VOC Sample Prep System Instrument Parameters The Atomx, equipped with a #11-VPH-Trap, and an Agilent 6890 GC/FID were utilized for this study. Tables 1-4 show the GC/FID and P&T conditions for water, soil, and methanol extraction applications. GC Parameters GC: Agilent 6890 Detector FID Column Restek RTX-VMS 20m x 0.18mmID x 1um Oven Program: Inlet: 220 C Column Flow 0.9mL/min Gas: Helium Split: 80:1 Pressure: 21.542 psi Inlet: Split/Split less 40 C for 4 min; 16 C/min to 100 C for 0 min; 30 C /min to 200 C for 4 min, 15.083 min runtime FID 250 C, Hydrogen Flow 35.0mL/min, Air Flow 400.0mL/min, Constant Column and Makeup Flow 35mL/min Table 1: GC/FID Parameters

Atomx Water Parameters Variable Value Variable Value Valve oven Temp 140 C Dry Purge Flow 100mL/min Transfer Line Temp 140 C Dry Purge Temp 20 C Sample Mount Temp 90 C Methanol Needle Rinse Off Water Heater Temp 90 C Methanol Needle Rinse Volume 3.0mL Sample Vial Temp 20 C Water Needle Rinse Volume 7.0mL Sample Equilibrate Time 0.00 min Sweep Needle Time 0.50min Soil Valve Temp 50 C Desorb Preheat Time 245 C Standby Flow 10mL/min GC Start Signal Start of Desorb Purge Ready Temp 40 C Desorb Time 2.00 min Condensate Ready Temp 45 C Drain Flow 300mL/min Presweep Time 0.25 min Desorb Temp 250 C Prime Sample Fill Volume 3.0mL Methanol Glass Rinse Off Sample Volume 5.0mL Number of Methanol Glass Rinses 1 Sweep Sample Time 0.25 min Methanol Glass Rinse Volume 3.0mL Sweep Sample Flow 100mL/min Number of Bake Rinses 1 Sparge Vessel Heater On Water Bake Rinse Volume 7.0mL Sparge Vessel Temp 40 C Bake Rinse Sweep Time 0.25 min Prepurge Time 0.00 min Bake Rinse Sweep Flow 100mL/min Prepurge Flow 0mL/min Bake Rinse Drain Time 0.40 min Purge Time 11.00 min Bake Time 2.00 min Purge Flow 40mL/min Bake Flow 200mL/min Purge Temp 20 C Bake Temp 280 C Condensate Purge Temp 20 C Condensate Bake Temp 200 C Dry Purge Time 2.00 min Table 2: Atomx Water Parameters (Parameters highlighted in yellow were not used.)

Atomx Soil Parameters Variable Value Variable Value Valve oven Temp 140 C Purge Time 11.00 min Transfer Line Temp 140 C Purge Flow 40mL/min Sample Mount Temp 90 C Purge Temp 20 C Water Heater Temp 90 C Condensate Purge Temp 20 C Sample Vial Temp 40 C Dry Purge Time 2.00 min Prepurge Time 0.00 min Dry Purge Flow 100mL/min Prepurge Flow 0mL/min Dry Purge Temp 20 C Preheat Mix Speed Slow Methanol Needle Rinse Off Sample Preheat Time 0.00 min Methanol Needle Rinse Volume 3.0mL Soil Valve Temp 100 C Water Needle Rinse Volume 7.0mL Standby Flow 10mL/min Sweep Needle Time 0.25 min Purge Ready Temp 40 C Desorbs Preheat Time 245 C Condensate Ready Temp 45 C GC Start Signal Start of Desorbs Presweep Time 0.25 min Desorbs Time 2.00 min Water Volume 10mL Drain Flow 300mL/min Sweep Water Time 0.25 min Desorbs Temp 250 C sweep Water Flow 100mL/min Bake Time 2.00 min Sparge Vessel Heater Off Bake Flow 400mL/min Sparge Vessel Temp 20 C Bake Temp 270 C Purge Mix Speed Fast Condensate Bake Temp 200 C Table 3: Atomx Soil Parameters (Parameters highlighted in yellow were not used.)

Atomx Methanol Extraction Parameters Variable Value Variable Value Valve Oven Temp 140 C Dry Purge Flow 100mL/min Transfer Line Temp 140 C Dry Purge Temp 20 C Sample Mount Temp 90 C Methanol Needle Rinse On Soil Valve Temp 100 C Methanol Needle Rinse Volume 2.0mL Standby Flow 10mL/min Water Needle Rinse Volume 7.0mL Purge Ready Temp 40 C Sweep Needle Time 0.25min Condensate Ready Temp 45 C Desorb Preheat Temp 245 C Presweep Time 0.25min GC Start Signal Star of Desorb Methanol Volume 7mL Desorb Time 2.00min Sparge Vessel Heater On Drain Flow 300mL/min Sparge Vessel Temp 40 C Desorb Temp 250 C Prepurge Time 0.00min Methanol Glass Rinse On Prepurge Flow 0mL/min Number of Methanol Glass Rinses 1 Sample Mix Speed Fast Methanol Glass Rinse Volume 3.0mL Sample Mix Time 4.00min Number of Water Bake Rinses 1 Sample Settle Time 2.00min Water Bake Rinse Volume 7.0mL Sample Sweep Time 0.25min Bake Rinse Sweep Time 0.25min Sample Sweep Flow 100mL/min Bake Rinse Sweep Flow 100mL/min Purge Time 11.00min Bake Rinse Drain Time 0.40min Purge Flow 40mL/min Bake Time 2.00min Purge Temp 20 C Bake Flow 200mL/min Condensate Purge Temp 20 C Bake Temp 280 C Dry Purge Time 2.00min Condensate Bake Temp 200 C Calibration / Results Table 4: Atomx Methanol Extraction Parameters A 50ppm working calibration standard was prepared in methanol using a 1000ppm Wisconsin GRO/PRVO standard. Calibration standards were then serially diluted with de-ionized water over a range of 5-200ppb. A 25ppm surrogate standard of fluorobenzene was prepared in methanol and transferred to one of the three standard addition vessels on the Atomx. The Atomx delivered the surrogate in 5µL aliquots to the sample to give a final concentration of 25ppb.

Using Agilent Chemstation software, the summation of the peak areas of each standard from methyltert-butyl-ether (MtBE) to naphthalene were used to generate the water and low-level soil calibration curves seen in Figures 1-2. The area of the surrogate fluorobenzene was subtracted from the total area to give the correct area requird for the Wisconsin GRO calibration.the calibrations passed method criteria with a correlation coefficient (r 2 ) of 0.9978 for water and r 2 of 0.9990 for low level soil. Using the #11-VPH-Trap it will allow for the resolution of MtBE down to 5ppb allowing for a calibration curve to be run from 5 to 200ppb. 100000000 80000000 Response 60000000 40000000 20000000 0 #11-VPH-Trap Water Calibration y = 446398x + 2E+06 R² = 0.9978 0 25 50 75 100 125 150 175 200 225 Concentration ppb Figure 1: Wisconsin GRO water calibration on a range from 5-200ppb using #11-VPH-Trap on Teledyne Tekmar s Atomx 80000000 #11-VPH-Trap Low level Soil Calibration Response 60000000 40000000 20000000 0 y = 330331x + 3E+06 R² = 0.999 0 50 100 150 200 250 Concentration (ppb) Figure 2: Wisconsin GRO low level soil calibration on a range from 5-200ppb using #11-VPH-Trap on Teledyne Tekmar s Atomx

The Wisconsin GRO method requires for Quality Control (QC) checks to demonstrate the capabilty to generate acceptable accuracy and precision. 1 For the water/soil, each matrix must run a RLCS, DLCS, LOD and LOQ. Each of these test listed above must fall between required limits that are set forth in the Wisconsion GRO method. Results of these test can be found in Table 5-6. Table 5 shows the capabilty of the Atomx to generate acceptable accuracy and precision for both water and soil matrices using the #11-VPH-Trap. Table 6 shows that the LOQs for both matrices are below the required limits set by the method. Water Quality Control Theoretical (ppb) Experimental (ppb) %Recovery Experimental %RSD RLCS 100 +/-20% 95.73* 95.73 2.47 DLCS 100+/-20% 89.0 89.0 Low Level Soil Quality Control RLCS 100 +/-20% 105.08* 105.08 1.59 DLCS 100+/-20% 97.48 97.48 High Soil Quality Control (Methanol Extracted) Theoretical (ppm) Experimental (ppm) %Recovery Experimental %RSD RLCS 10+/-120-75% 8.75* 87.5 3.96 DLCS 10+/-120-75% 8.61 86.1 Table 5: Replicate Laboratory Control Spike (RLCS) and Duplicate Laboratory Control Spike (DLCS) Results for water and soil. (*n= 5 replicates) Water/Soil Quality Control Matrix LOD* (ppb) LOQ*(ppb) Water 1.56 4.96 Low Level Soil 1.854 5.89 High level Soil 1.16 5.14 Table 6: Results for for the Lower Limit of Detection (LOD) and Limit of Quantitation LOQ for both water and soils. (*n=7 replicates) All the experimental values fall with in the acceptable range set forth by the Wisconsin GRO Method. The method states the LOQ for soil should be less then 10ppm and for groundwater it should be 0.1ppm or less. 1

Velocity XPT (straight trap) with an AQUATek 100 Autosampler Instrument Parameters The second part of the #11-VPH-Trap study was to validate the trap on a system that requires a straight analytical trap. A Velocity XPT concentrator with an AQUATek 100 autosampler was used in tandem with a GC/FID. The Agilent 6890 GC/FID parameters can be found in Table 1 while the Velocity XPT and AQUATek 100 instrument parameters can be found in Table 7. Velocity P&T and AQUATek 100 Parameters Variable Value Variable Value Valve Oven Temp 150 C Desorb Time 2.00 min Transfer Line Temp 150 C Desorb Temp 250 C Purge Ready Temp 35 C Drain Flow 300 ml/min Dry Flow Standby Temp 40 C Bake Time 2.00 min Standby Flow 5 ml/min Bake Temp 280 C Sparge Vessel Heater On Dry Flow Bake Temp 270 C Pre-Purge Time 0.00 min Bake Flow 200 ml/min Pre-Purge Flow 40 ml/min Pressurize Time 0.35 min Preheat Time 1.00 min Sample Transfer Time 0.35 min Sample Temp 40 C Rinse Loop Time 0.30 min Purge Temp 20 C Sweep Needle Time 0.30 min Purge Flow 40 ml/min Bake Rinse On Dry Purge Time 2.00 min Bake Rinse Cycle 1 Dry purge Temp 20 C Bake Rinse Drain Time 0.35 min Dry Purge Flow 200 ml/min Presweep Time 0.25 min GC Start Start of Desorb Water Temp 90 C Desorb Preheat Temp 245 C Calibration / Results Table 7: Velocity XPT and AQUATek 100 parameters. (AQUATek 100 parameters are listed in blue) A 50ppm working calibration standard was prepared in methanol using a 1000ppm Wisconsin GRO/PRVO standard from Restek. Calibration standards were then serially diluted with de-ionized water over a range of 5-200ppb. A 25ppm surrogate of fluorobenzene was prepared in methanol and transferred to one of the two standard addition vessels on the AQUATek100. The autosampler delivered the surrogate in 5µL aliquots to the sample for a final concentration of 25ppb.

Using Agilent Chemstation software, the summations of each standard from methyl-tert-butyl-ether to naphthalene were used to generate the water calibration curves seen in Figure 3. The area of the surrogate fluorobenzene was subtracted from the total area to give the correct value required for the Wisconsin GRO calibration.the calibration pass method criteria with a correlation coefficient (r 2 ) of 0.9998. 1500 #11-VPH Trap Water Calibration Response 1000 500 0 y = 6.0698x + 15.922 R² = 0.9998 0 25 50 75 100 125 150 175 200 225 Concentration ppb Figure 3: Wisconsin GRO water calibration on a range from 5-200ppb using #11-VPH-Trap on Teledyne Tekmar s Velocity XPT Purge and Trap Concentrator AQUATek 100 Autosampler. Since the AQUATek 100 is a water autosampler, only the water portion of the Wisconsin GRO method could be evaluated. RLCS, DLCS, LOD and, LOQ were preformed. Results of these tests can be found in Table 8-9. Table 7 shows the capabilty of the Velocity XPT with a AQUATek 100 to generate acceptable accuracy and precision for water samples using the #11-VPH-Trap. Table 9 shows that the LOQs for both matrices are below the required limits set by the method. Water Quality Control Theoretical (ppb) Experimental (ppb) %Recovery Experimental %RSD RLCS 100 +/-20% 98.44* 98.44 4.20 DLCS 100+/-20% 97.10 97.10 Table 8: Replicate Laboratory Control Spike (RLCS) and Duplicate Laboratory Control Spike (DLCS) Results for water. (*n= 5 replicates) Water Quality Control Matrix LOD* (ppb) LOQ*(ppb) Water 1.22 3.90 Table 9: Results for for the LOD and LOQ for both water. (*n=7 replicates)

All the experimental values fall with in the acceptable range set forth by the Wisconsin GRO Method. The method states for LOQ the groundwater should be 0.1ppm or less. 1 HT3 Automated Headspace Vial Sampler Instrument Parameters The dynamic (trap) mode of the HT3 headspace system was used to evaluate the GRO analysis. The HT3 was connected to an Agilent 6890 GC/FID. HT3 parameters are listed in Table 10. The Agilent 6890 GC/FID system parameters are listed in Table 1. Dynamic (Trap) Variable Value Variable Value Platen/Sample Temp 75 C Dry Purge Time 2.00 min Valve Oven Temp 150 C Dry Purge Flow 100mL/min Transfer Line Temp 150 C Dry Purge Temp 25 C Standby Flow Rate 50mL/min Desorb Preheat 245 C Sample Preheat Time 0.00 min Desorb Temp 250 C Standard and Sample Preparation Sweep Flow Rate 200mL/min Desorb Time 1.00 min Sweep Flow Time 5.00 min Trap Bake Temp 270 C Preheat Mixer 0.00 min Trap Bake Time 3.00 min Trap Material #11 Trap Bake Flow 200mL/min Table 10: Dynamic HT3 Headspace parameters. A 100ppb stock Internal Standard (IS) solution was prepared by adding 50 µl of a 2000ppm fluorobenzene standard to 1 L of de-ionized water. Using a Wisconsin GRO/PVOC mix a 50ppm stock standard was made. Calibration standards were then serially diluted with de-ionized water containing a 100ppb IS of fluorobenzene. The working range of the calibration ranged from 5-200ppb for Wisconsin GRO. 5 ml of the working standards were transferred to 22 ml headspace vials with a glass syringe and capped for the GRO analysis.

Calibration / Results The peak areas of the Wisconsin GRO standards from 5 to 200ppb were used to calculate the Response Factors (RF) by both external and internal standard calculations. The method requires %RSD less than 20% and correlation coefficients (r 2 ) greater than 0.99, these values are shown in Table 11. 5 ml Sample Data Calibration Data %RSD of the Compound Correlation Coefficient of the Compound Response Factor Compound #11-VPH-Trap #11-VPH-Trap MtBE 8.59 0.9999 Benzene 3.92 0.9996 Toluene 2.32 0.9993 Ethylbenzene 2.50 0.9995 m-, p-xylene 1.02 0.9995 o-xylene 2.39 0.9998 1,3,5-Trimethylbenzene 4.60 0.9999 1,2,4-Trimethylbenzene 4.00 0.9999 Naphthalene 5.82 0.9998 Table 11: Calibration data showing the %RSD and correlation coefficient for the calibration curve (5-200ppb)

Figure 4: Chromatogram showing the separation 100ppb Wisconsin GRO standard from Methanol (MeOH) on the Tekmar HT3 Using the #11-VPH-Trap enabled resolution of MtBE from the methanol peak seen in Figure 4. For all ten compounds listed in Table 11 all the %RSD were below 10%, while the correlation coefficient were all greater than the requirements of 0.99 or better. Trap Comparison Throughout this study, the #11-VPH-Trap was compared to other traps for differences in performing this method. The first difference is the ability of the #11-VPH-Trap to resolve MtBE down to 5ppb while the M-Trap can only resolve MtBE down to 10ppb. Figure 5 shows an overlay comparing chromatograms of a 5ppb Wisconsin GRO water sample on the #11-VPH-Trap and the M-Trap.

Figure 5: 5ppb chromatogram showing an overlay of Wisconsin GRO standard on the #11-VPH-Trap (Blue) and M-Trap (Red). Using the #11-VPH-Trap will allow the analyst to generate a calibration curve for all compounds over a range of 5-200ppb for the Atomx automated VOC sample prep system, the Velocity XPT concentrator, and the HT3 automated headspace vial sampler. The second difference uncovered in the trap comparison was each trap s methanol retention. If the trap retains too much methanol, the associated solvent peak could affect, or in some cases completely mask the MtBE due to the similar retention times. Figure 6 shows an overlay of the solvent peak of three traps: the #11-VPH-Trap, The #9 Trap, and the M-Trap.

Figure 6: An overlay of the solvent peaks for all three traps. As Figure 6 demonstrates the VOCARB 3000 Trap retains the most amount of methanol due to the fact that it is typically used in full VOC list which require stronger sorbent materials in order to trap lighter volatiles such as Freon for example. The #11-VPH-Trap and M-Trap are designed for the analysis of Wisconsin GRO and BTEX compounds, which do not contain many polar compounds. Since these two traps do not require full analyte list the sorbent beds can be modified so that each trap retains a smaller amounts of methanol. The methanol does not interfere with the first analyte of interest (MtBE), making the #11-VPH-Trap an idea analytical trap for GRO analysis. The third and final difference while noticed between the #11-VPH-Trap and the M-Trap was the bake back pressure. The #11-VPH-Trap behaved similarly to the the Tekmar #9 trap, the M-Trap showed evidence of increasing bake back pressure. This increasing pressure can lead to early degradation of the trap resulting in more frequent replacement of the analytical trap.

Conclusion The #11-VPH-Trap is the latest advancement in analytical traps by Teledyne Tekmar. This analytical trap provides the features of the Tekmar #9 analytical trap, but also reduces the sorbent bed so that lesser amount of methanol is retained. The trap keeps its ability to trap the polar and non-polar compounds found in gasoline range organics (ie: Wisconsin GRO and BTEX methods). This study validates the use of the #11-VPH-Trap in three of Tekmar products the Atomx, Velocity XPT with an AQUATek 100 autosampler, and the HT3 headspace analyzer. By using the #11-VPH-Trap the solvent peak is greatly reduced and does not interfere with the first analyte, MtBE. The #11-VPH-Trap also has a reduced bake pressure than the M-Trap which translates to a longer life cycle of the trap and reduces overall operation costs. References 1. Modified GRO Method for Determine Gasoline Range Organics Wisconsin DNR September 1995 PUBL-SW-140