Standard Practices for Air Speed Calibration Testing

Similar documents
Review of Anemometer Calibration Standards

Anemometer Calibration Requirements for Wind Energy Applications

REPORT OF SPECIAL TEST

A Procedure for Classification of Cup-Anemometers

3.2 Wind direction / wind velocity

Memorandum. From: Erik Hale, Meteorology Group Assistant Manager Date: January 8, 2010 Re: NRG #40 Transfer Function Recommendation.

Anemometry Anemometer Calibration Exercise

POWER QUALITY MEASUREMENT PROCEDURE. Version 4 October Power-Quality-Oct-2009-Version-4.doc Page 1 / 12

and methods. Manometers. Pressure-based measurement of velocity magnitude and direction. Anemometers, thermal probes. Temperature measurements.

An Essential Requirement in CV Based Industrial Appliances.

Turbines and turbine sets Measurement of emitted airborne noise Engineering/survey method

(ADVANCED) FLOW MEASUREMENTS Dr. János VAD, associate professor, Dept. Fluid Mechanics, BME

Lab Section Date. ME4751 Air Flow Rate Measurement

Windcube TM Pulsed lidar wind profiler Overview of more than 2 years of field experience J.P.Cariou, R. Parmentier, M. Boquet, L.

NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

Measurement method for the proficiency testing program

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

EUROVENT 8/12 SOUND TEST METHOD FOR DUCTED FAN COIL UNITS. FCP 2008 testing procedures FCP acoustical test method pag. 1

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Applied Fluid Mechanics

ISO INTERNATIONAL STANDARD. Meteorology Wind measurements Part 1: Wind tunnel test methods for rotating anemometer performance

Applied Fluid Mechanics

ISO 3741 INTERNATIONAL STANDARD

INTERNATIONAL STANDARD

Welcome to the world of wind energy. Wind Potential. Dr. D. V. Kanellopoulos OPWP Renewable Energy Training Program December 2016 Muscat, Oman

ECMA Measurement of airborne noise emitted and structure-borne vibration induced by small air-moving devices Part 1: Airborne noise measurement

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

TEST METHOD FOR STILL- AND FORCED-AIR JUNCTION-TO- AMBIENT THERMAL RESISTANCE MEASUREMENTS OF INTEGRATED CIRCUIT PACKAGES

ALASTAIR MCLACHLAN Applications Cameron

Part 2: Methods for special reverberation test rooms

On the correlation of the acoustic signal of microphones mounted on a flat plate to the turbulence of an impinging jet

INTERNATIONAL STANDARD

Measurement and Industrial Instrumentation

Characterisation and classification of RISØ P2546 cup anemometer

Sonic Anemometer Classification

1 Wind Turbine Acoustics. Wind turbines generate sound by both mechanical and aerodynamic

ISO INTERNATIONAL STANDARD. Ships and marine technology Marine wind vane and anemometers

Noise Testing and Prediction Methods for Multi-Point Ground Flares

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Document No 10/6298/001/GLA/O/R/001 Issue : B2. Kingspan Renewables Ltd. Kingspan KW6 Wind Turbine Noise Performance Test

Fluid Dynamic Simulations of Wind Turbines. John Abraham, Brian Plourde, Greg Mowry University of St. Thomas

Module 3: Velocity Measurement Lecture 16: Validation of PIV with HWA. The Lecture Contains: Hotwire Anemometry. Uncertainity

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Advisory Circular. U.S. Department of Transportation Federal Aviation Administration

Uncertainties in Acoustics

SONIC THERMOMETRY TODAY. W.R. Dagle and H.A. Zimmerman. Applied Technologies, Inc., Longmont, CO 80501

Technical Requirements. Version No. 1.0 Effective Date 10/1/2008

ISO 5136 INTERNATIONAL STANDARD. Acoustics Determination of sound power radiated into a duct by fans and other air-moving devices In-duct method

Expedited Filing Draft August 22, 2017

Designation: NWI DIF Horizontal Hail impact Standard

Method of Measuring Machinery Sound Within an Equipment Space

ISO INTERNATIONAL STANDARD. Acoustics Acoustic insulation for pipes, valves and flanges

IECRE OPERATIONAL DOCUMENT

DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER

An Introduction to the Differences Between the Two Most Recognized Force Standards

Site assessment for a type certification icing class Fraunhofer IWES Kassel, Germany

Theory and Design for Mechanical Measurements

Test Report EN (2008)

Standard Test Method for Field Measurement of Sound Power Level by the Two- Surface Method 1

Comparison of Noise Test Codes when Applied to air Compressors

Symmetry of Turbulent Characteristics Inside Urban Intersection

Solar Flat Plate Thermal Collector

NEMA MW (R2014) Dynamic Coefficient of Friction of Film-Insulated Magnet Wire

Omega Chemicals Emissions Testing Report

Automated data logging instrumentation system for wind speed and direction measurements

Wind Turbine Generator System Acoustic Noise Test Report. Whisper H40 Wind Turbine

ISO INTERNATIONAL STANDARD

FLOW MEASUREMENT INC 331 Industrial process measurement 2018

Effects of screens set characteristics on the flow field in a wind tunnel.

Model 3024 Albedometer. User s Manual 1165 NATIONAL DRIVE SACRAMENTO, CALIFORNIA WWW. ALLWEATHERINC. COM

VIBRATION ANALYSIS AND REPAIR PROCESS FOR THE VENTILATION SYSTEM FOR SMOKE DRAIN IN THE THERMAL POWER PLANT

Standard Test Method for Coefficient of Linear Thermal Expansion of Plastics Between 30 C and 30 C with a Vitreous Silica Dilatometer 1

SENSORS CLIMATRONICS CORPORATION (631) MANUAL TACMET II WEATHER STATION, EMI 2.0 INSTALLATION 1.0 INTRODUCTION

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Compressible Gas Flow

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

Investigation of fan vibration

A new lidar for water vapor and temperature measurements in the Atmospheric Boundary Layer

Standard Test Method for Thermal Conductivity of Refractories 1

Dennis: SODAR-Based Wind Resource Assessment

Addendum to ARI Standard , Air Terminals September 2002

ME332 FLUID MECHANICS LABORATORY (PART I)

ELECTRICAL AND THERMAL DESIGN OF UMBILICAL CABLE

The Effect of Endplates on Rectangular Jets of Different Aspect Ratios

ME411 Engineering Measurement & Instrumentation. Winter 2017 Lecture 11

WEATHER STATION FOR SOLAR FARM MONITORING

Measurement of fluid flow in closed conduits

Jet Aircraft Propulsion Prof. Bhaskar Roy Prof A M Pradeep Department of Aerospace Engineering Indian Institute of Technology, Bombay

RIVERBANK ACOUSTICAL LABORATORIES

LP PYRA Installation and Mounting of the Pyranometer for the Measurement of Global Radiation:

CALIFORNIA POLYTECHNIC STATE UNIVERSITY WIND RESOURCE ASSESSMENT. A thesis. presented to. the faculty of California Polytechnic State University,

ISO 376 Calibration Uncertainty C. Ferrero

Integrated Circuits Thermal Test Method Environment Conditions - Natural Convection (Still Air)

Canal Velocity Indexing at Colorado River Indian Tribes (CRIT) Irrigation Project in Parker, Arizona using the SonTek Argonaut SL

Technical Explanation for Axial Fans

OIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Turbine Meter TRZ 03 PRODUCT INFORMATION. Reliable Measurement of Gas

TECHNICAL NOTES NATIONAL ADVISORY COMMITTEE FOR AERONAUTICS NO. 341

Centralized Forecasting Registration and Communication Requirements for Distribution Connected Variable Generators. IESO Training

Transcription:

Standard Practices for Air Speed Calibration Testing Rachael V. Coquilla Bryza Wind Lab, Fairfield, California Air speed calibration is a test process where the output from a wind measuring instrument or anemometer is related to a reference wind speed. In some cases such as wind sensors deployed for measuring atmospheric winds, a transfer function generated from a calibration table is requested and provided in the test report. Currently, there are three published standards that present test protocols for rotating anemometers: ASTM D 5096 02, ISO 17713 1, and IEC 61400 12 1. For sonic anemometers, two published standards are referenced: ASTM D 6011 96 and ISO 16622. Rotating and sonic anemometers are regularly used for weather measurements. For smaller air speed instruments such Pitot tube systems and hot wire anemometers, a standard calibration practice is provided in ASTM D 3796 09. A common requirement from all of these standards is that anemometer calibration is to be conducted in a uniform flow, low turbulence wind tunnel, in which steady state wind conditions would isolate the performance of the wind sensor from certain factors such as vibrations and offangle winds. Procedures in IEC 61400 12 1 and ASTM D 3796 09 also specify that the wind tunnel reference speeds is to be measured using a Pitot tube system. On the other hand, ASTM D 5096 02, ISO 17713 1, ASTM D 6011 96, and ISO 16622 does not stipulate a particular measuring system for the reference wind speed, which allows the possibility of incorporating other types of wind measurement systems. At the NIST Air Speed Calibration Facility, the reference wind speed is measured using an LDA (Laser Doppler Anemometer). Although standards have offered general practices, several methods in air speed calibration testing have also been deployed. This presentation provides a review of current air speed calibration standards and a guide for manufacturers and end users in defining test protocols with respect to the application for the wind measuring instrument. A. Introduction Wind measuring instruments, velocity meters, or anemometers come in various types. One of the oldest types is a Pitot tube system, in which a measurement of the change in pressure in a moving fluid is related to the local velocity. There are also hot wire or hot film anemometers which relate the change in temperature or current across a wire to wind speed. Some of the latest models, such as sound emitting systems, are designed based on the Doppler effect, where the change in frequency of a waveform is related to the local velocity of the fluid moving across the waveform. The most common type are rotating anemometers with either cups or propellers. Rotating anemometers are designed so that its rotation rate is related to the incoming wind speed. Typical methods of generating an output signal from a rotating anemometer can be done through a reed switch or through the pole interaction of a magnet to a coil. Currently, there are six published standards that define the procedures for calibrating wind measuring instruments: ASTM D 5096 02, ISO 17713 1, and IEC 61400 12 1 for rotating anemometers, ASTM D 6011 96 and ISO 16622 for sonic anemometers, and ASTM D 3796 09 for S type Pitot tubes and, as commonly referenced, for standard Pitot tubes and hot wire anemometers. ASTM D 5096 02 and ISO 17713 1 are generally referenced for rotating anemometers applied in general meteorology applications and are referred in procedures for industries such as weather and nuclear. In the wind energy industry, IEC 61400 12 1 is referred for the calibration of rotating anemometers used is wind turbine power 2012 Measurement Science Conference Page 1 of 11

performance testing. This standard defines the procedures for generating the power curve of a wind turbine. A power curve gives the predicted wind turbine power production estimate for a particular wind speed. Since wind power is proportional to the cube of the wind speed, errors in the wind speed measurement translates to a much larger deviation in the predicted power output of a wind turbine. This predicted power output is essentially one of the key parameters for estimating the annual energy production from a wind power plant, which is vital in securing power purchase agreements. Thus, wind speed measurement is critical in the performance evaluation of a wind turbine. Wind measurement from sonic anemometers is determined from the change in the propagation of sound waves between a sound transmitter and a receiver due to the magnitude of the incoming flow of atmospheric wind. Since these types of sensors are dependent on the speed of sound, they are essentially most sensitive to the local atmospheric density and to the angle of the incoming wind. As defined in ASTM D 6011 96 and ISO 16622, base calibration for sonic anemometers is performed in a zero wind chamber, which involves the measurement of the acoustic pathlength and transient times. To fully complete the performance test of a sonic sensor, it is also necessary to evaluate the sensor through various temperature and pressure environments in the zero wind chamber and to measure the response of the sensor when subjected to various angles of incoming flow simulated in a wind tunnel. In general, the common requirement from all six standards is that calibration of wind measuring instruments is performed in a controlled, uniform flow, low turbulence wind tunnel or duct. Such conditions are essential in isolating the response of a wind sensor to the direct incoming, horizontal wind flow, which is the primary measurement collected in most applications. The following sections discuss particular details in the calibration methods of each of the standards. B. ASTM D 5096 02 Cup and Propeller Anemometer Test Procedures A guideline for performance testing of cup or propeller anemometers used for general meteorology applications is defined in ASTM D 5096 02, originally published in 1990. The purpose of this standard is not only to provide a method of calibration that produces a transfer function but also to determine certain performance characteristics (i.e., starting threshold, distance constant, and off axis response) for cup or propeller anemometers using a wind tunnel facility. The following table lists the wind tunnel requirements specified by ASTM D 5096 02. Table 1: Wind tunnel requirements for ASTM D 5096 02 cup and propeller anemometer testing. Wind Tunnel Characteristic Blockage Wind Speed Capability Minimum Requirement Anemometer front area is less than 5% of the test section cross section area Must be capable of reaching speeds up to at least 50% of the application range and must maintain speed within +/ 0.2 m/s Flow Uniformity Flow profile in the test section must be constant to within 1% Turbulence Air Density Uniformity Wind Speed Reading Must be less than 1% in the test section Density profile in test section must be less than 3% difference Maintain a relative accuracy of 0.1 m/s to its traceable source Air density is a function of the local conditions. Thus, during tests the ambient pressure and temperature in the test section are to be monitored and reported. All instruments used to measure the wind speed and the local conditions must be traceable to NIST (National Institute for Standards and Testing) or to 2012 Measurement Science Conference Page 2 of 11

any other internationally recognized standards body. The following is a list of the critical steps and requirements from the calibration procedure or transfer function test in ASTM D 5096 02. 1) Install the anemometer at an angle of attack maintained within 0.5. 2) Set the wind tunnel flow at the desired speed. Ensure that the wind tunnel has stabilized to each test speed before collecting data. The time to stabilize generally depends on the start up response and the distance constant of the anemometer undergoing test. 3) Acquire wind tunnel speed and anemometer rotation rate along with test section environmental conditions. Collect data for 30 to 100 seconds at each test speed and at a sampling rate of at least 100 Hz. 4) Calibration is performed for a speed range starting at two times the anemometer threshold, U o, up to 0.5 times the maximum application speed, U max. Incremental test speeds are then defined according to the following table of ascending and descending speeds, giving a total number of 20 test speeds. Table 2: ASTM D 5096 02 Calibration Test Speed Protocol Ascending Speeds 2 times U o 3 times U o 4 times U o 5 times U o 6 times U o 0.1 times U max 0.2 times U max 0.3 times U max 0.4 times U max 0.5 times U max Descending Speeds 0.5 times U max 0.4 times U max 0.3 times U max 0.2 times U max 0.1 times U max 6 times U o 5 times U o 4 times U o 3 times U o 2 times U o 5) Conduct a least squares regression analysis using the collected wind speed and anemometer rotation data set to determine and report the slope and offset of the calibration transfer function. 6) Calculate and report the wind speed residuals by first determining a predicted wind speed value for a range of anemometer rotation rates using the determined linear transfer function and then subtracting this predicted value from the measured wind speed. 7) Extrapolation of data beyond the range of actual measurement is not recommended Thus, if the maximum test speed is 50% of U max, then the transfer function only applies to that particular speed range. Along with procedures to calibrate an anemometer, ASTM D 5096 02 also specifies methods of determining three types of anemometer performance characteristics: 1) starting threshold, 2) distance constant and 3) off axis response. Starting threshold is the wind speed at which the anemometer starts to rotate. Distance constant, directly analogous to time constant, is the distance that air passes an anemometer as the anemometer reaches a certain equilibrium speed. In order to have an anemometer that is most sensitive or responsive to the incoming wind, the instrument is typically designed with the lowest distance constant. The off axis evaluation studies the angular response of an anemometer, in 2012 Measurement Science Conference Page 3 of 11

which the procedure subjects the anemometer to various vertical angle rotations to determine its sensitivity to upslope and downslope winds as evident in complex terrain. C. ISO 17713 1 Cup and Propeller Anemometer Test Procedures ISO 17713 1, released in May 2007, is an international standard for the performance testing of rotating anemometers. This standard was developed from ASTM D 5096 02 with updates to certain procedures. Requirements for the wind tunnel facility are similar. For anemometer calibration testing, a method of determining wind speed measurement resolution was added, adding a guideline in the data acquisition time interval for each test speed. ISO 17713 1 requires that a measurement resolution of 0.1 m/s or better be maintained. When using an anemometer pulse count technique, measurement resolution is calculated based on the wind distance passage per output pulse divided by the data acquisition time interval. ISO 17713 1 also redefined the test speed range recommended in ASTM D 5096 02, so that it would primarily cover the full range of intended use for the anemometer but also address the working range of the wind tunnel test facility. This test speed protocol also includes 5 points in the non linear range of the anemometer (i.e., near threshold, U o ). In the ASTM standard, calibration was performed up to 50% of the application speed. ISO 17713 1 requires that at least five additional test speeds evenly spaced up to the application speed be included and that linear regression or transfer function analysis should not take into account test data acquired near the anemometer threshold speed. Similar to the recommendation from ASTM, ISO 17713 1 suggests not using the resulting calibration transfer function to extrapolate beyond the calibration test range since this would result to increased measurement uncertainties. A typical test speed protocol offered by ISO 17713 1 is as follows: Table 3: ISO 17713 1 Calibration Test Speed Protocol 2 times U o 3 times U o 4 times U o 5 times U o 6 times U o 0.5 times U max 0.4 times U max 0.3 times U max 0.2 times U max 0.1 times U max 0.6 times U max 0.7 times U max 0.8 times U max 0.9 times U max U max Procedures for determining rotating anemometer starting threshold, distance constant, and off axis response are also defined in ISO 17713 1 and are similar to that defined in ASTM D 5096 02. D. IEC 61400 12 1 Cup Anemometer Test Procedures A key publication referenced in the wind energy industry is IEC 61400 12 1, ed. 1, released in 2005. This particular standard was first introduced to this standard by MEASNET (the International Measuring Network of Wind Energy Institutes). The purpose of this standard is to provide uniform methodology that will ensure consistency, accuracy, and reproducibility in the measurement and analysis of power performance by wind turbines. In IEC 61400 12 1, cup anemometers are the preferred sensors for wind turbine power performance testing. Cup anemometers are known to provide the lowest uncertainty in the measurement of atmospheric winds and have proven to be the most durable open air instruments for wind turbine testing applications. Similar to ASTM D 5096 02 and ISO 17713 1, the IEC 61400 12 1 specifies for cup anemometer calibration to be conducted in a uniform flow, low turbulence wind tunnel facility. The following table lists the IEC performance requirements for a wind tunnel facility. 2012 Measurement Science Conference Page 4 of 11

Table 4: Wind tunnel requirements for IEC 61400 12 1 cup anemometer calibration testing. Wind Tunnel Characteristic Description Minimum Requirement Blockage Ratio Flow Uniformity Horizontal Wind Gradient Turbulence Intensity Ratio of the anemometer plus mount frontal area to the total wind tunnel test section area Percent difference in wind speed within the test section volume Dynamic pressure differential at the area covered by the rotating cup anemometer Ratio between the wind speed standard deviation to the mean speed Not to exceed 0.1 for open test sections, 0.05 for closed sections Less than 0.2% in the longitudinal, transversal, and vertical directions Must be less than 0.2% Must be less than 2% Unlike the previously discussed standards, IEC 61400 12 1 specifies that cup anemometers are to be specifically calibrated to a Pitot static tube system. Wind speed from a Pitot static tube system is calculated using the differential pressure measured at the inlet of the Pitot static tube and also using the measurements of the local conditions inside the wind tunnel test section (i.e. ambient pressure, temperature, humidity). IEC 61400 12 1 also requires that equipment used to measure the differential pressure from the Pitot static tube, the ambient pressure, temperature, and relative humidity must be traceable to a national standards authority such as NIST (National Institute for Standards and Testing). The data acquisition system designed to collect such measurements must also be of minimal interference and noise in that the calculated velocity from the Pitot tube system maintains a resolution of at least 0.02 m/s. The following is a list of required conditions, defined in IEC 61400 12 1, to be included in a cup anemometer calibration procedure. 1) The test anemometer shall be installed on a mounting stand similar to that used in field installation and shall be positioned to the incoming flow at a maximum declination of 1. 2) Prior to every calibration round, the integrity of the experimental set up shall be verified by means of comparative calibration of a reference anemometer. 3) Prior to calibration, the anemometer shall be run in for about 5 minutes to avoid the effect that large temperature variations may have on the mechanical friction of the anemometer bearings. 4) Calibration will be conducted for a wind speed range of 4 m/s to 16 m/s at a 1 m/s interval, according to the following sequence of test speeds: 4, 6, 8, 10, 12, 14, 16, 15, 13, 11, 9, 7, and 5 m/s. This rising and falling speed sequence is to identify any hysteresis during the test. 5) Before test data acquisition, the wind tunnel must be given adequate time to stabilize to the set wind speed. According to the IEC, wind tunnel stability can be assumed if two successive 30 sec means are within 0.05 m/s of each other. 6) For test data collection at the desired speed, the sampling frequency shall be at least 1 Hz at a sampling interval of 30 seconds, which could be increased for low resolution anemometers. 7) Anemometer calibration shall be reported with a thorough assessment of calibration uncertainty (i.e., wind tunnel speed), according to ISO guidelines. To verify the integrity of the wind tunnel test facility, IEC 61400 12 1 recommends that the facility designate a reference anemometer that would undergo at least five calibration tests at various times and different environmental conditions. The integrity of the test process is verified by ensuring that such 2012 Measurement Science Conference Page 5 of 11

calibration tests are repeatable to less than 0.5% at 10 m/s. A second recommendation is for the facility to participate in round robin testing with other calibration facilities using a designated reference anemometer. The average calibration of the reference anemometer should agree with those of other facilities to within 1% over the range of 4 to 16 m/s. IEC 61400 12 1 also defines the test procedures required to determine cup anemometer classification, which provides a rating for the anemometer based on its performance characteristics in a particular environment. A classification number is used to help wind turbine evaluators determine a suitable anemometer for power performance testing. Classification involves sensitivity tests designed to evaluate the response of an instrument when subjected to various conditions such as atmospheric flow complexity, turbulence, and density changes. Through the standard, cup anemometers are classified based on the combined results from three types of sensitivity tests: 1) angular response, 2) dynamic effects due to rotor torque acceleration and deceleration, and 3) bearing friction torque for a range of environmental temperature conditions. Cup anemometer classification is also calculated primarily for two types of terrain, Class A and Class B, the latter being the more complex. A number preceding the class letter is the classification number calculated from the prescribed sensitivity tests. More specifically, this number identifies the relative error in the anemometer s indicated wind speed. For turbine power performance evaluation, IEC 61400 12 1 recommends that an anemometer of Class 1.7A or Class 2.5B or lower shall be employed depending on terrain conditions at the turbine test site. For conditions not specified under Class A or B terrain, an anemometer of Class 1.7S or lower is suggested. E. ASTM D 6011 96, Sonic Anemometer Test Procedures ASTM D 6011 96 is a published standard for sonic anemometer performance testing procedures. Sonic anemometers evaluated under this standard will obtain the calibration transfer characteristics necessary for general meteorological applications. One of the most difficult challenges is that sonic anemometers are essentially three dimensional sensors in which the horizontal wind speed reading is greatly affected by vertical or off axis flow. Thus, performance testing proves to be much more complicated than that for cup or propeller anemometers. This ASTM standard provides a method for assessing the performance of sonic anemometers that use inverse time solutions to measure wind velocity components and the speed of sound. Wind tunnel requirements for sonic anemometer performance evaluation are similar to those defined in ASTM D 5096 02 for cup or propeller anemometer testing. An added condition is that wind tunnel speeds from 1.0 to 10 m/s are to be maintained at +/ 0.1 m/s or better. In order to understand the full performance characteristics of a newly designed sonic anemometer, ASTM D 6011 96 recommends measurements of the: 1) acoustic pathlength, 2) system delay, 3) system delay mismatch, 4) thermal stability range, 5) velocity resolution, 6) shadow correction, 7) velocity calibration range, and 8) the acceptance angle. The first five performance tests are conducted in a zero wind chamber, in which the base calibration transfer characteristics of the instrument are obtained. Shadow correction, velocity calibration range, and the acceptance angle criteria are evaluated in a controlled wind tunnel facility to verify the overall response of the sonic anemometer. Due to the three dimensional sensitivity of sonic anemometers, shadow correction must be performed in a wind tunnel facility. This test process defines the sonic anemometer response or deviation at multiple orientations around the horizontal and vertical plane with respect to the incoming wind. For this testing, it is required that an apparatus that would rotate the anemometer around the horizontal and vertical plane be maintained with an angular alignment resolution of 0.5. Shadow correction creates a look up table or correction function, which may be used to adjust the sonic anemometer wind speed output according to its sensitivity to the particular off angle position from the 2012 Measurement Science Conference Page 6 of 11

incoming flow. The following steps summarize the procedures in ASTM D 6011 96 to conduct sonic performance tests in a wind tunnel facility: Step 1: With the sensor positioned vertically at zero angle of attack, select a low wind speed (about 2 m/s or lower) to test for orientations +/ 60 deg around its vertical axis at 10 deg increments. Repeat this procedure for a mid range speed (about 5 to 6 m/s) and then for a high wind speed (about 10 m/s or greater). Step 2: Repeat Step 1 for angles of attack at 5 deg increments starting with the sensor tilted 15 deg into the wind to 15 deg tilted away from the wind. F. ISO 16622 Sonic Anemometer Test Procedures An international standard for conducting performance testing on sonic anemometers is ISO 16622, which was released in September 2002. The goal of this standard is similar to that of ASTM D 6011 96 in that it provides the procedures that capture the three dimensional response of a sonic anemometer. ISO 16622 defines four test methods: 1) Zero wind chamber test, which determines the zero wind offset for the instrument. 2) Wind tunnel test, which characterizes the deviation of the wind speed due to the angle of the incoming flow (i.e. shadow correction testing) 3) Pressure chamber test, which defines the operational air density range for the instrument. 4) Field test, which evaluates the sensor under possible adverse environmental conditions difficult to simulate in a laboratory. Requirements for a wind tunnel facility are similar to that in ASTM D 6011 96. However, ISO 16622 requires that the wind tunnel be capable of producing wind speeds that cover the full application range of the sonic anemometer to be calibrated. Each steady setting of the wind speed must be maintained to within +/ 0.2 m/s, preferably +/ 0.1 m/s. For angle testing, the rotating fixture must have a 1 angular resolutions and +/ 0.5 repeatability. The wind tunnel test protocol in ISO 16622 is defined as follows: Step 1: With the sensor positioned vertically at zero angle of attack, select a low wind speed (about 10% of the maximum test speed) to test for orientation +/ 360 deg around its vertical axis at 5 deg increments. Repeat step for 18%, 32%, 56%, and 100% of the maximum test speed. Step 2: With the sensor positioned vertically at zero angle of attack and rotated at the worst case orientation (where the sensor reading is most disturbed by support structures), perform tests at 1%, 1.7%, 2.8%, 4.6%, 7.7%, 13%, 21%, 36%, 60%, and 100% of the maximum test speed. If such a wind speed range is not feasible for the wind tunnel lab, perform tests at 2%, 3%, 5%, 7%, 11%, 18%, 27%, 42%, 65%, and 100% of the maximum test speed. Repeat step for the best case orientation (where the sensor is least disturbed by support structures). Step 3: Repeat Steps 1 and 2 with the sensor tilted 15 deg into the wind and then tilted 15 deg away from the wind. G. ASTM D 3796 09 Standard Calibration for S type Pitot Tubes ASTM D 3796 09, first introduced in 1979, defines the procedures for the calibration type S Pitot tubes, which are used for flow measurements in ducts. Although this standard is defined for a specific instrument, it is commonly referred for the calibration of small flow measuring instruments such as standard Pitot tubes and hot wire anemometers. Similar to the previous standards discussed, ASTM D 2012 Measurement Science Conference Page 7 of 11

3796 09 requires that the calibration is to be performed in a well developed, stable flow test section. However, unlike the previously discussed standards, such test section conditions may be achieved in a small duct where the controlled steady flow is generated by an upwind blower. Test facility requirements defined by ASTM D 3796 09 are as follows: 1) To ensure fully developed flow, the test section is to be located 8 equivalent diameters downstream and 2 diameters upstream from the nearest flow disturbance such as an upwind blower or the exit of the duct. 2) The flow system shall have the capacity to generate at least four different, steady state velocities from 5.08 and 25.4 m/s. 3) Reference velocity is measured using a standard Pitot tube connected to a differential pressure gage capable of measuring to within 60.13 mm of water or 1.2 Pa (60.005 inch of water). 4) A measurement position in the duct is located so that the blockage from the sensors and mount is less than 3%. H. Conclusion Wind measuring instruments are essential in various applications for different industries. This paper presented a review and summary of six published standards that define wind instrument calibration procedures: ISO 17713 1, ASTM D 5096 02, and IEC 61400 12 1 for rotating anemometers, ASTM D 6011 96 and ISO 16622 for sonic anemometers, and ASTM D 3796 09 for small units such as Pitot tubes and hot wire anemometers. A summary of the wind tunnel facility requirements and anemometer test protocols is presented in the Appendix A to C. Of these standards, IEC 61400 12 1 is the only publication specifically related to the wind energy industry, particularly for the performance testing and classification of cup anemometers used in wind turbine power performance testing. Overall, most test facilities are capable of performing anemometer test procedures from these standards; however, some may be limited in wind speed test range. Recommended sonic anemometer test procedures define the necessary steps to understand the complete response of a newly designed sonic sensor; however, such procedures prove to be quite rigorous for most test facilities, particularly, if performing calibration checks of field installed anemometers. In most cases, sonic procedures from standards are normally revised according to the application of the sensor and its manufacturer recommendations. 2012 Measurement Science Conference Page 8 of 11

Appendix A: Summary of Wind Tunnel Requirements for Wind Instrument Testing Wind Tunnel Characteristic Wind Speed Capability Flow Uniformity Horizontal Wind Gradient Turbulence Intensity Air Density Uniformity Data Acquisition Resolution Blockage Repeatability Interlaboratory Comparison Description Controlled wind speed range Percent difference in wind speed or velocity profile within the test section volume Pressure differential at the area covered by a rotating sensor Ratio between the wind speed standard deviation to its mean Percent difference in density within test section volume Method of collecting reference wind speed data Ratio of the wind sensor projected area plus mount to the test section cross sectional area Consistency of repeated tests Comparison to other test facilities Summary of Minimum Requirements from Standards IEC 61400 12 1: generate controlled speeds from 4 to 16 m/s ASTM D 5096 02, ISO 17713 1: produce speeds at least from 0 to 50% of application speed and maintained within +/ 0.2 m/s ASTM D 6011 96: generate speeds of at least 1 to 10 m/s within +/ 0.1 m/s or better ISO 16622: produce speeds over full application speed and maintained within +/ 0.2 m/s, preferably +/ 0.1 m/s ASTM D 3796 09: generate at least four different, steady state velocities from 5.08 to 25.4 m/s, differential pressure gage is capable of measuring to within 60.13 mm of water or 1.2 Pa (60.05 inch of water) IEC 61400 12 1: less than 0.2% ASTM D 5096 02, ASTM 6011 96, ISO 17713 1, and ISO 16622: uniform velocity profile to within 1% ASTM D 3796 09: to ensure fully developed flow, the test section is to be located 8 equivalent diameters downstream and 2 diameters upstream from the nearest flow disturbance IEC 61400 12 1: must be less than 0.2% IEC 61400 12 1: must be less than 2% ASTM D 5096 02, ASTM 6011 96, ISO 17713 1, and ISO 16622: must be less than 1% ASTM D 5096 02 & ISO 17713 1: less than 3% difference IEC 61400 12 1: maintain resolution of 0.02 m/s, sample frequency of at least 1 Hz for 30 sec ASTM D 5096 02, ASTM 6011 96, ISO 17713 1, and ISO 16622: maintain resolution of 0.02 m/s, sample frequency of at least 10 Hz for 30 to 100 sec IEC 61400 12 1: not to exceed 10% for open test sections and 5% for closed test sections ASTM D 5096 02, ASTM 6011 96, ISO 17713 1, and ISO 16622: less than 5%, preferably closer to 1% IEC 61400 12 1: repeated reference tests are repeatable to less than 0.5% at 10 m/s IEC 61400 12 1: round robin testing with other calibration facilities are within 1% over the range of 4 to 16 m/s 2012 Measurement Science Conference Page 9 of 11

Appendix B: Summary of Wind Tunnel Test Protocols for Rotating Anemometers Standard IEC 61400 12 1 ASTM D 5096 02 ISO 17713 1 Test Speed Protocol for Rotating Anemometers 4 to 16 m/s at 1 m/s increasing & decreasing increments (i.e. 4, 6, 8, 10, 12, 14, 16, 15, 13, 11, 9, 7, and 5 m/s) Test speeds start at 2 times the anemometer threshold, U o, up to 0.5 times the maximum application speed, U max, in the following ascending to descending speeds: Ascending Speeds Descending Speeds 2 times U o 0.5 times U max 3 times U o 0.4 times U max 4 times U o 0.3 times U max 5 times U o 0.2 times U max 6 times U o 0.1 times U max 0.1 times U max 6 times U o 0.2 times U max 5 times U o 0.3 times U max 4 times U o 0.4 times U max 3 times U o 0.5 times U max 2 times U o Test speeds cover the range of intended use for the anemometer and the working range of the wind tunnel, which includes 5 points in the non linear range of the anemometer (near threshold, U o ). Typical test protocols are as follows: 2 times U o 0.1 times U max 0.6 times U max 3 times U o 0.2 times U max 0.7 times U max 4 times U o 0.3 times U max 0.8 times U max 5 times U o 0.4 times U max 0.9 times U max 6 times U o 0.5 times U max U max Appendix C: Summary of Wind Tunnel Test Protocols for Sonic Anemometers Standard ASTM D 6011 96 ISO 16622 Wind Tunnel Test Protocol for Sonic Anemometers Step 1: With the sensor positioned vertically at zero angle of attack, select a low wind speed (~ 2 m/s or lower) to test for orientations +/ 60 deg around its vertical axis at 10 deg increments. Repeat this procedure for a mid range speed (~ 5 to 6 m/s) and then for a high wind speed (~ 10 m/s or greater). Step 2: Repeat Step 1 for angles of attack at 5 deg increments starting with the sensor tilted 15 deg into the wind to 15 deg tilted away from the wind. Step 1: With the sensor positioned vertically at zero angle of attack, select a low wind speed (about 10% of maximum) to test for orientation +/ 360 deg around its vertical axis at 5 deg increments. Repeat for 18%, 32%, 56%, and 100% of the maximum test speed. Step 2: With the sensor positioned vertically at zero angle of attack and rotated at the worst case orientation (where the sensor reading is most disturbed by support structures), perform tests at 1%, 1.7%, 2.8%, 4.6%, 7.7%, 13%, 21%, 36%, 60%, and 100% of the maximum test speed. If such a wind speed range is not feasible for the wind tunnel lab, perform tests at 2%, 3%, 5%, 7%, 11%, 18%, 27%, 42%, 65%, and 100% of the maximum test speed. Repeat step for the best case orientation (where the sensor is least disturbed by support structures). Step 3: Repeat Steps 1 and 2 with the sensor tilted 15 deg into the wind and then 15 deg away from the wind. 2012 Measurement Science Conference Page 10 of 11

References (1) ASTM D 3796 09 (2009). Standard Practice for Calibration of Type S Pitot Tubes. American Society for Testing and Materials (ASTM). (2) ASTM D 5096 02 (2007). Standard test method for determining the performance of a cup anemometer or propeller anemometer. American Society for Testing and Materials (ASTM). (3) ASTM D 6011 96 (2008). Standard test method for determining the performance of a sonic anemometer/thermometer. American Society for Testing and Materials (ASTM). (4) IEC 61400 12 1 1 st ed (2005). Wind turbines Part 12 1: Power performance measurements of electricity producing wind turbines. International Electrotechnical Commission (IEC). (5) ISO 16622 (2002). Meteorology Sonic anemometers/thermometers Acceptance test methods for mean wind measurements. International Organization for Standardization (ISO). (6) ISO 17713 1 (2007). Meteorology Wind Measurements Part 1: Wind tunnel test methods for rotating anemometer performance. International Organization for Standardization (ISO). 2012 Measurement Science Conference Page 11 of 11