DRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s)

Similar documents
DRI OXYCODONE APPLICATION BECKMAN COULTER DxC 700 AU

CEDIA CYCLOSPORINE PLUS (LOW RANGE) APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

CEDIA COCAINE OFT ASSAY APPLICATION (EU) BECKMAN COULTER AU680

CEDIA OPIATE OFT ASSAY APPLICATION (UK) - Beckman Coulter AU680

CEDIA COCAINE OFT ASSAY APPLICATION (UK) BECKMAN COULTER AU680

Activity. This procedure is to be used in conjunction with the package insert for PLAC Test for Lp-PLA 2

CEDIA Cyclosporine PLUS High Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System, VITROS 5,1 FS and 4600 Chemistry Systems

CEDIA Cyclosporine PLUS Low Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems

Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400

CEDIA Tacrolimus Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems

THC PRINCIPLE REF ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

COCM Cocaine Metabolite

For In Vitro Diagnostic Use. Rx Only. ANNUAL REVIEW Reviewed by Date Reviewed by Date

Psychoactive Drugs Urine LC-MS/MS Analysis Kit User Manual

For In Vitro Diagnostic Use

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010

Gentian Canine CRP Immunoassay Application Note for Siemens ADVIA * 1800/1650

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Hydrogen Peroxide Colorimetric Detection Kit

Human Creatinine Serum Detection Kit

CALC PRINCIPLE REF & REF A28937 & A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

BÜHLMANN fcal turbo. Reagent Kit. Calprotectin turbidimetric assay for professional use B-KCAL-RSET-U. Intended use

Instructions for use. Glycine ELISA BA E-2100

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Monensin ELISA Kit. Catalog Number KA assays Version: 11. Intended for research use only.

For In Vitro Diagnostic Use. Rx Only. ANNUAL REVIEW Reviewed by Date Reviewed by Date

PRINCIPLE. Kit Reorder # Kit Reorder # Kit Reorder # A17652 Kit Reorder # A ANNUAL REVIEW Reviewed by: Date.

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Instructions for use. Chromogranin A ELISA TM E-9000

The NGAL Test Reagent Kit

PMSG (Pregnant Mare Serum Gonadotropin) ELISA

Micro Volume QuEChERS kit

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

1. Intended use. 2. Introduction. 3. Assay description. 4. Limitations. Instruction For Use Gyrolab Generic PK Kit

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Date. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

CHGA (Human) ELISA Kit

101Bio.com. Total Aflatoxin ELISA Kit. better than big brands and save. Cat. #: T21131 (1 x 96-well plate)

Cystatin C ELISA. For the quantitative determination of cystatin C in human biological samples.

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit

Beta-2 M Beta-2-Microglobulin

CTGF (Connective Tissue Growth Factor) ELISA

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Testosterone ELISA, EIA

Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit

MyBioSource.com. This package insert must be read in its entirety before using this product.

IgM (Canine) ELISA. For the quantitative determination of IgM in canine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

ab MDA Assay Kit (competitive ELISA)

IgD ELISA. For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information

Creatinine Colorimetric Detection Kit

IgD ELISA. For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information

IgG (Canine) ELISA. Please see Appendix A for Reference Serum information. For Research Use Only. Not For Use In Diagnostic Procedures.

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Mouse IgE ELISA. For the quantitative determination of IgE in mouse serum or plasma. For Research use Only. Not For Use In Diagnostic Procedures.

IgG (Rabbit) ELISA Kit

IgA ELISA. For the quantitative determination of IgA in human serum and plasma. Please see Appendix A for Reference Serum Information

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

Human hepatitis B virus surface antigen(hbsag) ELISA Kit

Histamine Research ELISA

Rat Beta-2 Microglobulin ELISA

Mouse KIM-1 ELISA. For the quantitative determination of Kidney Injury Molecule in mouse serum, plasma, or urine.

Human papillomavirus,hpv ELISA Kit

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit

CERTIFICATE OF ANALYSIS Nitric Oxide Assay Kit

Aflatoxin M1 (AFM1) ELISA Kit

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG

Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit

Canine brain natriuretic peptide,bnp ELISA Kit

Zearalenone ELISA Kit

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #:

Rabbit IgA ELISA Kit

This immunoassay kit allows for the in vitro quantitative determination of Aflatoxin M1 concentrations in milk, milk power.

C3 (Mouse) ELISA Kit. Catalog Number KA assays Version: 15. Intended for research use only.

Rat Ferritin ELISA. For the determination of ferritin in serum and plasma of rats. For Research Use Only. Not For Use In Diagnostic Procedures.

IgG (Bovine) ELISA Kit

Mouse Retinol Binding Protein 4 ELISA

KIM-1 ELISA. For the quantitative determination of Kidney Injury Molecule in various biological samples.

Paraquat ELISA Kit. Catalog Number KA assays Version: 17. Intended for research use only.

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Mouse Cholecystokinin (CCK) ELISA Kit

HUMAN MYELOPEROXIDASE

IgE (Rat) ELISA Kit. Catalog Number KA assays Version: 05. Intended for research use only.

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Hepatitis B virus IgM ELISA Kit

The kit is a sandwich enzyme immunoassay for in vitro quantitative measurement of GFP in biological agents. Reagents Quantity Reagents Quantity

Canine Erythropoietin,EPO ELISA Kit

IgE (Dog) ELISA Kit. Catalog Number KA assays Version: 05. Intended for research use only.

Cortisol (Horse) ELISA Kit

Instructions for use 5-HIAA ELISA BA E-1900

IgA (Guinea Pig) ELISA Kit

IgA ELISA. For the quantitative determination of IgA in human serum and plasma. For Research Use Only. Not for Use in Diagnostic Procedures.

For the quantitative determination of IgE in equine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Human Lipocalin-2/NGAL ELISA

Mouse Creatinine Urinary Detection Kit

Procine GHb ELISA Kit

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Transcription:

DRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s) Beckman Coulter Reorder Number A39935 Homogeneous enzyme immunoassay for the qualitative or semi-quantitative determination of ecstasy drugs in human urine. For In Vitro Diagnostic Use Only Intended Use The information provided in this application sheet is intended as a supplement to the package insert. Refer to the package insert for information on intended use, reagent storage, reagent preparation, specimen collection, specimen storage, quality control, and additional performance data. Ordering Information Materials available from your local Beckman Coulter Representative: Item DRI Ecstasy Assay Kit (100 ml) Beckman Coulter Reorder Number A39935 DRI Negative Calibrator DRI Ecstasy 250 ng/ml Calibrator DRI Ecstasy 500 ng/ml Calibrator DRI Ecstasy 750 ng/ml Calibrator DRI Ecstasy 1000 ng/ml Calibrator MGC Select DAU Control Set A44121 A39936 A39937 A39938 A40566 A40567 User-Defined Reagent Cartridge (pkg. of 12) 442835 For Technical Support please contact your local Beckman Coulter Representative. DRI is a registered trademark of Thermo Fisher Scientific Corporation. March 2008 Page 1 of 11

Reagent Storage Refer to the package insert for information on reagent storage. NOTE: It is not recommended to leave the reagent on-board for more than 60 days. Procedure for Analyzer Refer to the operator s manuals for information on analyzer operation. Dispense adequate amounts of Antibody/Substrate Reagent A and Enzyme Conjugate Reagent E into appropriate compartments of a User Defined Cartridge (PN 442835) as shown in the table below: User Defined Cartridge DRI Ecstasy Assay Kit Compartment A Compartment B Antibody/Substrate Reagent A 48 ml Enzyme Conjugate Reagent E 18 ml Calibrator Information Refer to the package insert for information on calibration. To use qualitative mode monitoring, calibration shall be carried out every 14 days or as indicated by control recovery. To use semi-quantitative mode monitoring, calibration shall be carried out every 14 days or as indicated by control recovery. March 2008 Page 2 of 11

Application Parameter Parameters DRI Ecstasy Assay chemistry parameter for qualitative mode on Synchron System(s). INSTRUMENT PARAMETERS: SYNCHRON CX UniCel DxC Synchron LX Chemistry Name Ecstasy NA a Test Name XTCX XTCX Reaction Type: Rate 1 Rate 1 Units ng/ml ma/min Decimal Precision/Precision: X.X X.X Reaction Direction: Positive Positive Calculation Factor: 0 1000 Math Model: Linear DAT Cal. Time Limit: 336 336 Number of Calibrators: 2 3 #1 0.0 0.0 #2 500.0 500.0 #3 1000.0 #4 #5 #6 Primary Wavelength: 340 nm 340 nm Secondary Wavelength: 650 nm 650 nm Sample Volume: 15 μl 10 μl REAGENTS: Primary Inject (first)/first Inject Compartment/Component: A A Volume/Dispense Volume: 200 μl 125 μl Add Time/Inject Time: NA a NA a Primary Inject (first)/second Inject Compartment/Component: None None Volume/Dispense Volume: 0 μl NA a Add Time/Inject Time: NA a -180 a&b Second Inject/Third Inject Compartment/Component: B B Volume/Dispense Volume: 75 μl 75 μl Add Time/Inject Time: 368 sec 276 sec REAGENTS: Blank Start Read: 237 sec 292 sec Continued on next page March 2008 Page 3 of 11

INSTRUMENT PARAMETERS: SYNCHRON CX UniCel DxC Synchron LX End Read: 300 sec 308 sec Reaction 1 Start Read: 96 sec 336 sec End Read: 144 sec 396 sec Reaction 2 Start Read: NA a NA a End Read: NA a NA a USABLE RESULT RANGE: Lower Limit: 0.000 b 0.000 b Upper Limit: 99999.999 b 99999.999 b ERROR DETECTION LIMITS: Reagent Blank/Blank ABS Low Limit: -1.5 b -1.5 b ABS High Limit: 1.5 b 2.2 b Reaction/Reaction 1 ABS Low Limit: -1.5 b -1.5 b ABS High Limit: 1.5 b 2.2 b Reaction 2 ABS Low Limit: NA a -1.5 b ABS High Limit: NA a 2.2 b SUBSTRATE DEPLETION Initial Rate: 99.999 99.999 Delta ABS: 1.5 b 2.2 b MULTI POINT SPAN: 0.000 b 0.000 b a NA=Not applicable b Denotes default value Continued on next page March 2008 Page 4 of 11

Parameters DRI Ecstasy Assay chemistry parameter for semi-quantitative mode on the Synchron System(s). INSTRUMENT PARAMETERS: SYNCHRON CX UniCel DxC Synchron LX Chemistry Name Ecstasy NA a Test Name XTCX XTCX Reaction Type: Rate 1 Rate 1 Units ng/ml ng/ml Decimal Precision/Precision: X.X X.X Reaction Direction: Positive Positive Calculation Factor: 0 1 Math Model: 1 1 Cal. Time Limit: 336 336 Number of Calibrators: 5 5 #1 0.0 0.0 #2 250.0 250.0 #3 500.0 500.0 #4 750.0 750.0 #5 1000.0 1000.0 #6 Primary Wavelength: 340 nm 340 nm Secondary Wavelength: 650 nm 650 nm Sample Volume: 15 μl 10 μl REAGENTS: Primary Inject (first)/first Inject Compartment/Component: A A Volume/Dispense Volume: 200 μl 125 μl Add Time/Inject Time: NA a NA a Primary Inject (first)/second Inject Compartment/Component: None None Volume/Dispense Volume: NA a Add Time/Inject Time: NA a -180 a&b Second Inject/Third Inject Compartment/Component: B B Volume/Dispense Volume: 75 μl 75 μl Add Time/Inject Time: 368 sec 276 sec REAGENTS: Blank Start Read: 237 sec 292 sec Continued on next page March 2008 Page 5 of 11

INSTRUMENT PARAMETERS: SYNCHRON CX UniCel DxC Synchron LX End Read: 300 sec 308 sec Reaction 1 Start Read: 96 sec 336 sec End Read: 144 sec 396 sec Reaction 2 Start Read: NA a NA a End Read: NA a NA a USABLE RESULT RANGE: Lower Limit: 0.00 0.00 Upper Limit: 1000.0 1000.0 ERROR DETECTION LIMITS: Reagent Blank/Blank ABS Low Limit: -1.5 b -1.5 b ABS High Limit: 1.5 b 2.2 b Reaction/Reaction 1 ABS Low Limit: -1.5 b -1.5 b ABS High Limit: 1.5 b 2.2 b Reaction 2 ABS Low Limit: NA a -1.5 b ABS High Limit: NA a 2.2 b SUBSTRATE DEPLETION Initial Rate: 99.999 b 99.999 b Delta ABS: 1.5 b 2.2 b MULTI POINT SPAN: (1-2) 0.004 0.009 (2-3) 0.024 0.046 (3-4) 0.031 0.037 (4-5) 0.016 0.013 (5-1) 0.082 0.106 a NA=Not applicable b Denotes default value March 2008 Page 6 of 11

Results and Data Interpretation Refer to the package insert for information on data interpretation. Typical Precision A properly operating Synchron system should exhibit precision values less than or comparable to the following: Within-run Precision (Qualitative) Rate ma/min Mean SD %CV Mean SD %CV Mean SD %CV Negative Control (375 ng/ml) (500 ng/ml) Positive Control (625 ng/ml) 83 0.76 0.91 115.6 1.50 1.29 118.3 1.77 1.49 102 0.83 0.81 146.2 1.95 1.33 149.6 1.91 1.28 123 0.73 0.60 172.9 1.42 0.82 176.9 2.16 1.22 Total Run Precision (Qualitative) Rate ma/min Mean SD %CV Mean SD %CV Mean SD %CV Negative Control (375 ng/ml) (500 ng/ml) Positive Control (625 ng/ml) 83 1.00 1.20 116 2.45 2.12 118 3.96 3.34 102 1.19 1.16 146 3.36 2.20 150 3.92 2.62 123 1.08 0.89 173 3.08 1.78 177 4.25 2.40 Within-run Precision (Semi-quantitative) Conc. ng/ml Mean SD %CV Mean SD %CV Mean SD %CV Negative Control (375 ng/ml) (500 ng/ml) Positive Control (625 ng/ml) 342.4 4.83 1.33 361.7 7.64 2.11 368.6 6.68 1.81 498.2 5.04 1.01 502.6 8.01 1.59 502.3 10.17 2.03 633.5 6.67 1.05 634.8 13.07 2.06 629.7 11.49 1.82 Continued to next page March 2008 Page 7 of 11

Total Precision (Semi-quantitative) Conc. ng/ml Mean SD %CV Mean SD %CV Mean SD %CV Negative Control (375 ng/ml) (500 ng/ml) Positive Control (625 ng/ml) 342.4 8.86 2.45 361.7 9.24 2.55 368.6 13.24 3.59 498.2 7.09 1.42 502.6 9.80 1.95 502.3 10.61 2.11 633.5 7.76 1.22 634.8 15.25 2.40 629.7 15.32 2.43 Continued on next page March 2008 Page 8 of 11

Cross Reactivity Ecstasy compounds and ecstasy structurally related compounds were tested for cross reactivity. The table below presents representative results for Synchron instruments. Ecstasy Compounds That Tested Positive Cross-Reactant Tested Conc. ng/ml 3,4-Methylenedioxymethamphetamine (MDMA) 500 3,4-Methylenedioxyamphetamine (MDA) 900 3,4-Methylenedioxyethylamphetamine (MDEA) 400 N-Methylbenzodioxazolylbutananmine (MBDB) 2000 Benzodioxazolylbutananmine (BDB) 1500 p-methoxyamphetamine (PMA) 4700 p-methoxymethamphetamine (PMMA) 1700 Ecstasy Structurally Related Compounds That Tested Negative Instruments Synchron CX UniCel DxC Synchron LX Cross-Reactant Concentration ng/ml ng/ml 3,4-Methylenedioxymethamphetamine (MDMA) 500 500 d-amphetamine 600,000 600,000 l-amphetamine 10,000 30,000 d,l-amphetamine 40,000 75,000 l-ephedrine 800,000 800,000 d-methamphetamine 200,000 300,000 l-methamphetamine 10,000 10,000 d,l-methamphetamine 20,000 30,000 Phentermine 50,000 100,000 d,l-phenylpropanolamine 800,000 800,000 d-pseudoephedrine 1,000,000 1,000,000 Continued on next page March 2008 Page 9 of 11

Method Comparison Refer to the Accuracy Section of the Product Insert for c Comparison A total of 138 clinical samples were tested in the Qualitative and Semiquantitative modes on the Synchron and compared to the results by. The data were analyzed and presented in the table below. Qualitative + - + - + - - 7 c 81-7 c 81-7 c 81 : : Semi-quantitative: + - + - + - - 7 c 81-7 c 81-7 c 81 : : c Of the 138 total clinical samples confirmed by (Gas Chromatography / Mass Spectrometry) seven were between 500 ng/ml and 634 ng/ml (+25% of cutoff concentration). Due to limitations in sample volume, not all Ecstasy derivatives and/or metabolites were tested for. March 2008 Page 10 of 11

ADDITIONAL INFORMATION This guideline has undergone limited technical evaluation and is intended to provide guidance only for the use of this reagent on the UniCel DxC, SYNCHRON LX and CX Clinical Systems. The performance characteristics have not been established. You should perform additional testing before reporting diagnostic results. Information on sample preparation, expected values, quality control, as well as warnings and precautions related to the use of this reagent may be obtained from the package insert. Instrument operating instructions are contained in the SYNCHRON LX Operations Manual, SYNCHRON CX Operating Instructions, or UniCel DxC Systems Instructions For Use (IFU) Manual. Since Beckman Coulter does not manufacture the reagent or perform quality control or other tests on individual lots, Beckman Coulter cannot be responsible for the quality of the data obtained which is caused by performance of the reagent, any variation between lots of reagent, or protocol changes by the manufacturer. SHIPPING DAMAGE If damaged product is received, notify your Beckman Coulter Clinical Support Center. March 2008 Page 11 of 11