Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Similar documents
Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Standard Operating Procedure for the Analysis of Fresh Water Samples for Orthophosphorus CCAL 34B.2

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

ALLOWAY METHOD OUTLINE

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Chloride by Flow Injection Analysis (FIA)

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

Hexavalent Chromium by Flow Injection Analysis (FIA)

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Organic Carbon, Total (Low Level) (UV Promoted, Persulfate Oxidation) CAS # Total Organic Carbon TOC C

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Method for Nitrite determination on Low-Range Samples

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Standard Operating Procedure for the Digestion and Analysis of Fresh Water Samples for Total Phosphorus and Total Dissolved Phosphorus CCAL 35B.

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

Test Method: CPSC-CH-E

Standard Operating Procedure VUSP M

LOW-RANGE TOTAL CHLORINE

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA)

METHOD DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY

Spectrophotometric Determination of Ferrocyanide in Effluents

California Environmental Protection Agency Air Resources Board

United States EPA Method 415.3

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Carbonate content. SCAN-N 32:98 Revised White, green and black liquors and burnt lime sludge

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

STANDARD OPERATING PROCEDURES

QAM-I-116 Preparation of Labware

Standard Operating Procedure for: ph using Oakton ph 5+ Handheld ph Meter. Missouri State University. and

Standard Operating Procedures. Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC)

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

Standard Operating Procedure

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments

Sulfotepp impurities in Chlorpyrifos EC formulations

Facilities Management

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

Fusion UV/Persulfate TOC Analyzer

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Standard Operating Procedure VUSP N

GUIDELINES FOR THE SAFE USE OF PYROPHORIC LIQUID REAGENTS

STANDARD OPERATING PROCEDURES

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Bromide by Flow Injection Analysis (FIA)

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

Absorption Spectra of Cyanine Dyes

Approximate Volatile Acids by Titration

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Zinc Metal Determination Perkin Elmer Atomic Absorption Spectrometer AAnalyst Procedures

SYRACUSE UNIVERSITY RADIATION PROTECTION PROGRAM APPLICATION FOR USE OF RADIOACTIVE MATERIALS

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

This protocol is a modified version of the DuBois assay (DuBois et al. 1956) for the quantification of total carbohydrate content of a sample.

Laboratory Certification Workshop

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

SECTION D.2 AMMONIA NITROGEN

Method for estimation of iodine in urine

Standard Operating Procedure for: Total Phosphorus (3010R02 Total P.doc) Missouri State University. and

APPLICATION FOR AUTHORIZATION

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

DRAFT EAST AFRICAN STANDARD

Technical Procedure for General Laboratory Equipment

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

Micro Volume QuEChERS kit

425 Jordan Road SOP. No. 10 Rev. No. 1.3 Troy, NY Date: 03/14/2012 Page 1 of 7

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

Fusion UV/Persulfate TOC Analyzer

Safety in the Chemistry Laboratory

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

Nitrogen, Total Inorganic

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Manganese III Digestion Method * (with chloride removal)

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Responsibilities: Effective Date: November Revision Date: February 8, VP, Facilities and Construction Management. Issuing Authority:

USGS Troy WSC Laboratory Inductively Coupled Plasma- NH4Cl Soil Extracts SOP 425 Jordan Road Rev. No. 2.0 Troy, NY Date: 03/16/2012 Page 1 of 7

Analysis of domestic sludge using the Agilent 4200 MP-AES

3. Chemical Hygiene Plan: Laboratory Standard Operating Procedures. A. Laboratory Specific Information and Signatures

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Method 630. The Determination of Dithiocarbamate Pesticides in Municipal and Industrial Wastewater

CHEMISTRY DEPARTMENT

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Examining the Role of TOC Analyzers in the Pharmaceutical Laboratory

Transcription:

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis, Oregon Prepared by Kathryn Motter

Page 2 of 10 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 Table of Contents 1.0 Scope and Application... 3 2.0 Summary of Method... 3 3.0 Definitions... 3 4.0 Interferences... 4 5.0 Safety... 4 6.0 Equipment and Supplies... 4 7.0 Reagents and Standards... 5 7.1 Preparation of Reagents... 5 7.2 Preparation of Standards... 5 8.0 Sample Handling and Storage... 6 9.0 Quality Control... 6 10.0 Calibration and Standardization... 7 11.0 Procedure... 7 11.1 Shimadzu TOC-VCSH Instrument Operating Parameters... 7 11.2 Procedure... 8 11.3 System Notes... 8 12.0 Data Analysis and Calculations... 8 13.0 Method Performance... 8 14.0 Pollution Prevention... 9 15.0 Waste Management... 9 16.0 References... 9 17.0 Tables, Diagrams, Flowcharts, and Validation Data... 10 18.0 Document Revision History... 10

Page 3 of 10 Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1 1.0 Scope and Application 1.1 This method details the determination of Dissolved Inorganic Carbon in fresh waters by acidification and sparging followed by infrared analysis. The practical range of determination for this method as reported by the manufacturer is 0 25000 ug/l. The laboratory determined Method Detection Limit is 50 ug/l. 2.0 Summary of Method 2.1 Samples are acidified with phosphoric acid and sparged with zero air to convert inorganic carbon to volatilized carbon dioxide. The CO 2 formed is transported to the detector in a carrier gas stream and measured directly by an infrared detector. The measured CO 2 consists of carbon derived from carbonates, hydrogen carbonates and dissolved carbon dioxide in the sample. 3.0 Definitions 3.1 DI water: Water that has been through a deionization system to produce water similar to ASTM Type I reagent with 16.7 Mohms resistivity (ASTM) (Reference 16.3). 3.2 Method Detection Limit (MDL): The minimum concentration of an analyte that can be measured and reported with 99% confidence, based on a one-sided 99% confidence interval (t-value at a significance level of 0.01 and n-1 degrees of freedom) from at least seven repeated measurements of a low concentration standard measured within an analysis run. MDL = ts Where,

Page 4 of 10 t = Student s t value at a significance level of 0.01 and n-1 degrees of freedom s = standard deviation of at least seven repeated measurements of a low level standard 4.0 Interferences 4.1 Atmospheric contact must be limited to minimize carbon dioxide outgassing or uptake. 4.2 Contact with organic material may contaminate a sample. Care must be taken in sample handling and storage to minimize exposure. 4.3 This procedure is applicable only to homogeneous samples which can be reproducibly injected by microliter syringe into the instrument. 4.4 Purgeable organic carbon that may be volatized during the sparging process is considered interference in the analysis. 5.0 Safety 5.1 The toxicity or carcinogenicity of each reagent has not been precisely determined; however, each chemical should be regarded as a potential health hazard. Exposure to these chemicals should be reduced to the lowest possible level. Cautions are included for known extremely hazardous materials. 5.2 The following chemicals have the potential to be highly toxic or hazardous. For detailed explanations, consult the MSDS. 5.2.1 Phosphoric Acid 5.2.2 Hydrochloric Acid 5.2.3 Sodium Carbonate 6.0 Equipment and Supplies Note: Brand names, suppliers and part numbers are for illustrative purposes only. No endorsement is implied. Equivalent performance may be achieved using apparatus and materials other than those specified here, but demonstration of equivalent performance that meets the requirements of this method is the responsibility of the laboratory. 6.1 Shimadzu TOC-VCSH Analyzer with IC Reaction Vessel

Page 5 of 10 6.1.1 Shimadzu ASI-V Autosampler 6.1.2 Instrument Controller 6.1.3 Data Collection Software 6.1.4 CO 2 Absorber (soda lime) 6.2 Ultra Zero Air compressed air and regulator 6.3 40 ml borosilicate vials and septum caps 6.4 Laboratory glassware and pipettes 6.5 Balance 6.6 Safety glasses 6.7 Nitrile gloves 6.8 Lab coat or apron 6.9 Laboratory exhaust fume hood 6.10 60 ml Leur Lock Syringes 6.11 Syringe Filters (0.45 um pore size, nylon filter, 25 mm) 7.0 Reagents and Standards 7.1 Preparation of Reagents 7.1.1 25% Phosphoric Acid (weight %) Dilute 50 ml of concentrated phosphoric acid (85 %) to a final volume of 250 ml with DI water. A final concentration accuracy of ± 2% is acceptable. 7.1.2 ph 2 Hydrochloric Acid (system rinse) Dilute 5 ml of 2N HCl to 1 L with DI water. 7.1.3 2N Hydrochloric Acid Add 25 ml of concentrated HCl to 125 ml of DI water. 7.2 Preparation of Standards 7.2.1 Calibration Standards: Standards are prepared by dilution of a single element standard purchased from a vendor that provides traceability to NIST

Page 6 of 10 standards. Working standards for calibration are prepared at concentrations stated below. Standard # mg C/L 1 0.20 2 0.50 3 1.00 4 2.00 5 5.00 6 7.50 7 10.00 8 15.00 9 20.00 7.2.2 Second Source Check Standard: Standard is prepared by dilution of a single element standard purchased from a vendor that provides traceability to NIST standards, this from a source other than that of the calibration standards. Check standard concentration should be mid-calibration range for high and low ranges; currently 1.00 and 10.00 mg C/L. 8.0 Sample Handling and Storage 8.1 Unfiltered samples are delivered to the lab in sealed syringes and held refrigerated until time of analysis. Samples are analyzed within 48 hours of delivery to the lab. Immediately prior to analysis, samples are filtered through syringe filters directly into clean analysis vials, minimizing contact with the atmosphere and mixing of air into the sample. Samples are filled to the brim with reverse meniscus and capped immediately with septum caps. 9.0 Quality Control 9.1 Preparation of stock standards is recorded on worksheets and documented by weight of standard added to a given flask before dilution to volume with DI water. All records of certification are kept on file at the CCAL Laboratory. 9.2 Blank: DI water run before and after the calibration, and after every check standard to monitor drift and carry-over. 9.3 Second Source Check Standard: Run every ten samples to monitor drift and analysis efficiency.

Page 7 of 10 9.4 Method Detection Limit (MDL): Established for each analyte. Based on a one-sided 99% confidence interval (t-value) from at least seven repeated measurements of a low concentration standard. The t-distribution value is multiplied by the standard deviation of the population (n-1) to obtain the MDL. 9.5 Analytical Duplicate: Separate analysis from the same sample aliquot. One per ten samples if additional syringe is provided. 9.6 Standard recoveries are tracked over time to monitor overall performance. 9.7 When sample volume allows, samples with concentration greater than the highest calibration standard are diluted and reanalyzed. 10.0 Calibration and Standardization 10.1 Balances: calibrated yearly by external vendor. 10.2 Pipette delivery checked by weight to within 1% of theoretical weight of aliquot volume. 10.3 The instrument is calibrated every analysis run. The r 2 value of each calibration curve should be greater than 0.999. 10.4 Check standard recovery must be within 10% of theoretical value. 11.0 Procedure 11.1 Shimadzu TOC-VCSH Instrument Operating Parameters Compressed Air Pressure: Supply pressure @ tank regulator 600 kpa Carrier Gas Pressure 200 kpa Carrier Gas Flow Rate 150 ml/min Sample Volume 500 ul Acid Addition 25 % by wt H 3 PO 4 Analysis IC Calibration Method Linear Regression Catalyst NA Sparge Time 4 minutes Washes 2 Combustion Temperature NA

Page 8 of 10 11.2 Procedure 11.2.1 Stabilize system with ph ~2 HCl (sets of 10 repeated measurements). Monitor peak profile for consistency. 11.2.2 Follow manufacturer s instructions for analysis. Refer to instrument manual. 11.2.3 After use, sample vials and septum are rinsed well with DI water and soaked overnight in 0.5 N HCl, rinsed well again and soaked overnight in DI water and air dried. The vials and septum caps are stored in sealed containers until use. 11.3 System Notes 11.3.1 Yearly preventative maintenance as suggested by the manufacturer is required for optimum performance. 11.3.2 Periodic regeneration of the IC solution with phosphoric acid. Regeneration is performed prior to each analysis set if the instrument has not been turned off. 12.0 Data Analysis and Calculations 12.1 A standard curve is prepared in Excel by plotting standard response against known concentration. Sample response is compared to the standard curve to determine concentration. 12.2 All raw data is copied into a template for calibration and sample concentration determination. Carbon in DI water that is used to prepare calibration standards must be corrected for in determination of sample concentration. The calibration curve is shifted to pass through zero and the sample concentration is determined from this corrected curve. 12.2 When sample volume allows, samples with concentration greater than the highest calibration standard are diluted and reanalyzed. 13.0 Method Performance 13.1 This method was validated through inter-laboratory studies. The CCAL Water Analysis Laboratory participates in the National Water Research Institute s (NWRI) Environment Canada Proficiency Testing (PT) Program.

Page 9 of 10 14.0 Pollution Prevention 14.1 The chemicals used in this method pose little threat to the environment when properly managed. 14.2 All standards and reagents should be prepared in volumes consistent with laboratory use to minimize the volume of disposable waste. 14.3 For further information on pollution prevention consult Less is better: Laboratory Chemical Management for Waste Reduction, available from the American Chemical Society s Department of Government Relations and Science Policy, 1155 16 th Street NW, Washington D.C. 20036, (202) 872-4477. 15.0 Waste Management 15.1 It is the laboratory s responsibility to comply with all federal, state and local regulations governing waste management, and to protect the environment by minimizing and controlling all releases from fume hoods and bench operations. Compliance with all sewage discharge permits and regulations is required. 15.2 For further information on waste management, consult The Waste Management Manual for Laboratory Personnel, and Less is Better: Laboratory Chemical Management for Waste Reduction, both available from the American Chemical Society s Department of Government Relations and Science Policy, 1155 16 th Street NW, Washington DC, 20036. 16.0 References 16.1 USEPA. 2004. National Wadeable Streams Assessment: Water Chemistry Laboratory Manual. EPA841-B-04-008. U.S. Environmental Protection Agency, Office of Water and Office of Research and Development, Washington DC. 16.2 Code of Federal Regulations. Protection of Environment. Section 40, Appendix B to Part 136. Definition and procedure for the determination of the method detection limit. Revision 1.11. Revised July 1, 1990. Office of the Federal Register, National Archives and Records.

Page 10 of 10 16.3 ASTM. American Society for Testing and Materials. Standard Specifications for Reagent Water. D1193-77 (Reapproved 1983). Annual Book of ASTM Standards, Vol. 11.01. ASTM: Philadelphia, PA, 1991. 16.4 Shimadzu TOC-V CPH/CPN & TOC-Control V Software User s Manual. Shimadzu Corporation; Analytical & Measuring Instruments Division; Kyoto, Japan. 17.0 Tables, Diagrams, Flowcharts, and Validation Data 17.1 NA 18.0 Document Revision History Original Document: February 2010 Version: 21A.0 Edit Date: New Version: 20A.3 Section 6.2: add Ultra Zero Air requirement Section 7.2.1: add additional high concentration calibration standards Section 7.2.2: change to use of prepared standards and add high check standard Section 13: add NWRI participation General editing