Study of Nanoparticles in a Few Rivers in North East Italy using SP-ICP-MS

Similar documents
A Comparison of Microsecond vs. Millisecond Dwell Times on Particle Number Concentration Measurements by Single Particle ICP-MS

Accurate Sizing and Precise Counting of 10 nm Gold Nanoparticles using the Enhanced Sensitivity of the NexION 2000 ICP-MS

Analysis of Cerium Dioxide Nanoparticles in Soil Using Single Particle ICP-MS

Single particle analysis using the Agilent 7700x ICP-MS

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

ICP-MS-based characterization of inorganic nanoparticles sample

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

TITLE NANOPARTICLE SUSPENSIONS COUNTING AND SIZING NANOPARTICLES VIA SINGLE PARTICLE INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY

Detection of engineered cerium oxide nanoparticles in soils

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

Characterization of nanoparticles in aqueous samples by ICP-MS

SP-ICP-MS Analysis of Size and Number Concentration in Mixtures of Monometallic and Bimetallic (Core-shell) Nanoparticles

DISCLAIMER: This method:

Analysis of NIST Gold Nanoparticles Reference Materials Using the NexION 300 ICP-MS in Single Particle Mode

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Analysis of domestic sludge using the Agilent 4200 MP-AES

CEINT/NIST PROTOCOL REPORTING GUIDELINES FOR THE PREPARATION OF AQUEOUS NANOPARTICLE DISPERSIONS FROM DRY MATERIALS. Ver. 2.0

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

Development and validation of the test guideline on agglomeration behavior of nanomaterials in different aquatic media.

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

ICP-OES Application Note Number 35

Rapid Detection of Americium-241 in Food by Inductively-Coupled Plasma Mass Spectrometry

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Assessing the Fate of Nanoparticles During Coagulation Using SP-ICP-MS. ICP - Mass Spectrometry APPLICATION. Introduction

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Determination of Nanoparticle Sizeand Number Concentration using the npquant Evaluation Module for Qtegra ISDS. The world leader in serving science

Basic Digestion Principles

Observation of size-independent effects in nanoparticle retention behavior during asymmetric-flow field-flow fractionation

NICKEL-63/59 IN WATER

Precision and accuracy of protein size determination using the ActiPix TDA200 Nano-Sizing System

Application of Selenium Speciation Analysis to Elucidate Limitations with Accepted Total Selenium Methods

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

TECHNETIUM-99 IN SOIL

Screening Methods for Metal-Containing Nanoparticles in Water APM 32

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Sample Analysis Design PART II

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

Unilever s approach to assuring the safety of novel nanomaterials - supporting the risk assessment science

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

TECHNETIUM-99 IN WATER

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Trace Analyses in Metal Matrices Using the ELAN DRC II

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

Investigation of Silver Nanoparticle Sorption and Dissolution in Quartz Columns

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Samples compliant with the WATER

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Structural effects on catalytic activity of carbon-supported magnetite. nanocomposites in heterogeneous Fenton-like reactions

Nanocrystalline Cellulose:

In a typical routine, the pristine CNT (purchased from Bill Nanotechnology, Inc.) were

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Supporting Information

enable measurement. This method separates these isotopes effectively.

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Introduction to Nanoparticle Tracking Analysis (NTA) Measurement Principle of ZetaView

Protocol. Indranil Chowdhury Chemical and Environmental Engineering, University of California, Riverside

Analytical Methods for Nanomaterials in Food

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Food safety and nanotechnology: Results from LENA Project. Richard Winterhalter

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

This installment of Atomic Perspectives discusses. Single-Particle ICP-MS: A Key Analytical Technique for Characterizing Nanoparticles

URANIUM IN SOIL. Analytical Procedure (2 GRAM SAMPLE) 1. SCOPE

Lab 4 Major Anions In Atmospheric Aerosol Particles

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

Accurate Determination of TiO 2 Nanoparticles in Complex Matrices using the Agilent 8900 ICP-QQQ

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Imaging gold nanoparticles in mouse liver by laser ablation inductively coupled plasma mass spectrometry

Determination of Zinc, Cadmium, Lead and Copper in Effluent Sample

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

RARE EARTH FLUORIDE MICROPRECIPITATION

Analytical Method Determination of airborne metal concentrations by ICP-MS Sampling using Solu-Sert filter capsules

Dr. Christoph Johann Wyatt Technology Europe GmbH Copyright Wyatt Technology Europe GmbH All Rights reserved 1

Formation of iron oxide nanoparticles for the photooxidation of water: Alteration of finite size effects from ferrihydrite to hematite

Low Level Phosphorus: A Method Comparison Study

Fate, Transport, and Toxicity of Nanoparticles in the Environment

Supplementary information for:

ICP-MS-based characterization of inorganic nanoparticles sample preparation and off-line fractionation strategies

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Appendix II- Bioanalytical Method Development and Validation

Automation of the radiochemical procedures for the sequential separation of radionuclides

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Nanoparticles in food and non food Recent methods and measurements

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Environmental Consequences of Nanotechnologies

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Transcription:

Research Article imedpub Journals www.imedpub.com Nano Research & Applications DOI: 10.21767/2471-9838.100030 Study of Nanoparticles in a Few Rivers in North East Italy using SP-ICP-MS Elisa Piccoli *, Denis Mazzilis, Emanuele Gava and Stefano De Martin Laboratory of Inorganic Micropollutants, Regional Environmental Protection Agency of Friuli Venezia Giulia (ARPA-FVG), Italy * Corresponding author: Dr. Elisa Piccoli, Laboratory of Inorganic Micropollutants, Regional Environmental Protection Agency of Friuli Venezia Giulia (ARPA-FVG) 42 Colugna Avenue, 33100 Udine, Italy, Tel: +3904321918351; E-mail: elisa.piccoli@arpa.fvg.it Rec date: January 27, ; Acc date: February 09, ; Pub date: February 12, Citation: Piccoli E, Mazzilis D, Gava E, De Martin S () Study of Nanoparticles in a Few Rivers in North East Italy using SP-ICP-MS. Nano Res Appl Vol. 4: No. 1: 1. Abstract The use of nanoparticles (NP) has increased in these years. They are used in many fields, for example industrial and commercial sectors. The large employment may compromise the human and the ecosystem health, because the fate/damages/consequences of NP in the environment are still not known. SP-ICP-MS is a fast method and it can be used for a screening of environmental samples, it combines high sensitivity with fast analytical speed. This is a water surface study, few rivers in the North East of Italy are analysed. 3 litres of surface water are filtered with a vacuum pump with membrane by 0.22 µm pore. The filter is put into a test tube with 30 ml of ultrapure water and then sonicated for 20 minutes. The analyses are performed on a SP-ICP-MS NexION 350D Perkin Elmer using Syngistix Software. Transport efficiency and particle calibration are performed with standard gold nanoparticles of 30 nm, 60 nm and 90 nm. The detection limit (DL) is calculated from the analysis of 5 replicates of ultrapure water. The recovery test is made on water surface both spiked and unspiked and it appears to be between 61,8% and 97%. Considering the high dilution of NP in an environmental complex matrix, this study evaluates the efficiency of the concentration via filtration for SP-ICP-MS analysis of nanoparticles of ZnO, CeO 2, TiO 2, Ag. The results show that the analysis in SP-ICP-MS combined with the filtration enhance the detection sensitivity and the characterization of nanoparticles. Keywords: SP-ICP-MS; Nanoparticles; Environmental pollution; River; Filtration Introduction This preliminary study puts the attention on propagation of nanoparticles (NP) into few rivers in Friuli Venezia Giulia (FVG)- Italy. The use of NP has been increasing in recent years. The scientific and engineering communities are driven heavily by government investments on nanotechnology applications. NP are used in industrial and commercial sectors (e.g. foods, pesticides, personal care products, pharmaceutical, and medicine). The large use may compromise the humans and the ecosystem health (EFSA 2016 2017) [1]. In order to understand the fate of NP in the environment, it is necessary to study all the characteristics that permit their correct identification. Six characteristics contribute to determine nanomaterials: size, shape, surface charge, surface porosity, composition and structure [2]. As suggested in EPA APM 32 [3], detecting, quantifying and characterizing the NP in environmental samples with traditional technique is difficult because they are present in a very low concentration and there are a lot interference coming from background natural colloid. The most used techniques to determine the NP and to achieve a reliable and quantitative determination of NP were hydrodynamic chromatography (HDC) and field flow fractionation (FFF) in combination with detectors such as multiple angle light scattering (MALS) and inductive coupled mass (ICP-MS). These methods require a complex preparation of the samples with the aim of removing interfering matrix and meet a low detection limit. To determine the real particle size number distribution, several methods are available: nanoparticle tracking analysis (NTA), differential mobility analyzers (DMA) and single particle inductive coupled mass spectrometry (SP-ICP-MS). The use of FFF-ICP-MS permits to determine the size of NP. This method has more sensitivity than ICP-MS however the NP interactions with the porous membrane wall of the separation system can lead to peak tailing and even irreversible adsorption. In addition, the time of migration of NP is very high and it cannot permit to use this as routine method. Using an analytical approach as HDC permits to lower the resolution than FFF-ICP-MS, but the separation of the NP is more rapid. HDC-ICP-MS provide elemental information to size distribution, therefore the application for natural environmental media is limited. Imagines techniques as scanning electron microscopy (SEM) and transmission electron microscopy (TEM) are used for the characterizing of shape, structure, particle size and particle size distribution, agglomeration of NP; in this case the quantification is precluded. SP-ICP-MS is by far the common tool used for NP analysis because of its trace element capability and extremely low detection limits is ideally suited to characterization of NP. The information obtained using this technique are concentration, particle size and particle size distribution, agglomeration and Copyright imedpub This article is available from: http://nanotechnology.imedpub.com/ 1

composition. The shape of NP is assumed spherical. Traditional ICP-MS detects metal ions based on their mass to charge ratio. The analysis in single particles requires a different approach that measures dissolved elements with ICP-MS. Two parameters are very important: the dwell time (reading time) and the settling time (overhead and processing time). In a traditional ICP-MS multiple intensity are integrated with a long dwell time (0.03-0.1 s) to produce an overall metal concentration for the sample. In SP-ICP-MS each intensity reading is combined with a shorter dwell time (100 µs or less) and plotted individually. The principal assumption behind SP- ICP-MS is that each pulse represents a single particle event which depends on short dwell time, constant flow rate and a sufficient low particle concentration. The better situation for analysis of nanoparticles is when the settling time is zero because every nanoparticle is detected. This parameter also permits a fast data acquisition with two benefits: to ensure an accurate particle counting and to eliminate the changes that particles miss or that only partial ion clouds from particles detect, as suggested in Chady [4]. Another important parameter in SP-ICP-MS is the transport efficiency. It is defined as the ratio of the amount of analyte entering the plasma and the amount of analyte aspirate. During each new analytical session, the transport efficiency must be determined for a good result [5]. This study puts the attention on the presence of NP in a few rivers in FVG and it describes the suitability of SP-ICP-MS for routine analysis, including adequate and simple sample preparation methods. This method researches nanoparticle of TiO 2, CeO 2, ZnO and Ag. To determine nanoparticles in the environment a very sensitive analytical technique is needed because the concentration of nanoparticles is very low in surface waters. Ten rivers surface water samples, downstream of productive settlements, are collected and analysed. Methods Reagents, standards and materials Ionic Au calibration standards 1.25, 2.5, 5, 10 µg/l are used as ionic calibrations and made from 1000 mg/l. It is purchased from Ultra Scientific. Three sizes of Au (30-60-90 nm) with nanoparts carboxylic acid, purchased from Perkin Elmer, are used for the particles calibration and the transport efficiency determination. Nanopowder of TiO 2 <100 nm particle size (BET), CeO 2 <50 nm particle size (BET) ZnO <50 nm particle size (BET) and Ag dispersion 40 nm particle size 0.02 mg/ml in aqueous buffer contains sodium citrate as stabilizer. All the nanoparticles standards are purchased from Sigma- Aldrich. Samples are filtered with a decontaminated vacuum pump with cellulose acetate membrane filters with a 0.22 µm pore size purchased from Millipore. HNO 3 suprapure (Panreac), H 2 SO 4 hiperpure (Panreac) and H 2 O 2 suprapure 30% (Merck) are needed for the digestion of the samples. Instruments Single Particle Inductively Coupled Plasma-mass Spectrometry (SP-ICP-MS) Perkin Elmer NexION 350D with Syngistix Nano Application Software ModuleVers.1.0 is used for this study. For the analysis of the NP dissolved concentration, Milestone Ultrawave and DigiPrep digestors (both with dedicated and decontaminated vessels) are used for the mineralization of the samples; to avoid the aggregation of the NP before the analysis, it s necessary the use of a sonicator. Experimental protocol Ten rivers (downstream of productive settlements) of different nature in FVG have been chosen for this study. Each sample is put into a polypropylene bottle and stored in the dark in a fridge at 4 C [6]. The samples of three rivers of different typologies have been selected: one river affected by industrial activities (A), one river exposed to an urban pollution (B) and one river in estuarine context (C). For the evaluation of the total metals content, river water is treated with two different mineralization methods. To determine Ag and Zn total, the digestion is made according to UNI EN ISO 15587-2 [7], the mineralization is performed with an addition of 6.25 ml of nitric acid suprapure to 25 ml of sample water. The sample is mineralized in an open digestion system (DigiPrep) at 95 C for at least 3 hours. To extract Ti and Ce total from surface water a different treatment is required, as suggested in Poscher et al. [8]. 10 ml of H ₂ SO 4 are added to 5 ml of water, and the solution is mineralized via microwave, according to UNI EN 16173 [9]. After the digestion, few drops of hydrogen peroxide are added cautiously (to enhance the extraction of Ti and Ce) until the apparition of an orange-red colour. All extracts are brought to a volume of 50 ml with deionized water and analysed with ICP-MS NexION 350 D (Table 1). Table 1 Range of concentrations obtained from samples analysis. Constituents Minimum Maximum Ti 3,7 µg/l 11.2 µg/l Ce 0.2 µg/l 0.4 µg/l Zn 0.12 µg/l 7.9 µg/l Ag 0.64 µg/l 0.80 µg/l Table 2 DLs estimated accoring to ISO/TS 19590: 2017. Size detection limit (nm) CeO 2 TiO 2 Ag ZnO 23.7 49.6 24.2 31.4 To evaluate the efficiency of the filtration in the NP research, three litres of water are filtered with a vacuum pump with cellulose acetate membrane filters with a 0.22 µm pore size to capture nanoparticles [10]. The filter is put into a vessel 2 This article is available from: http://nanotechnology.imedpub.com/

of 50 ml then 30 ml of deionized water are added. Before determining and quantifying nanoparticles, the samples and standards are sonicated for 20 minutes. Particle size detection limit is performed according to ISO/TS 19590 [11]. The size detection limit is determined by three times the standard deviations (added to the mean size detected above the background) (Table 2). For accurate results, the sample flow rate should be determined (as in Syngistix Nano Application module) [12]. Ten millilitres of DW are weighted and aspirated through the nebulizer with a pump set 20 rpm for three minutes and the vial is weighted again. This operation is made for 5 times, then the flow rate is calculated on 5 readings average. The transport efficiency is calculated during the calibration. This operation is made every day and the average value of transport efficiency resulted between 4.85% and 7.96% (Table 3). Dissolved Au standards (1.25-2.5-5.0 and 10.0 µg/l) are prepared in 1% HCl acid. Au NP (30 60 and 90 nm) are used to determine transport efficiency and particle calibration (Figure 1) (Table 4). Table 3 Example of parameters used in an analysis. Mode Gas Channel Gas Flow (ml/min) RPq AFT (V) Sample Flow Rate (ml/min) Dwell Time (μs) Scan Time (s) TE (%) STD None 0 0,5-0,246 25 30 6,25 Even though nanoparticles have got a diameter between 1 and 100 nm, they form agglomerates and aggregates. For this reason, they don t pass through the pores of the filter. Table 4 Calibration parameters. Variables Dissolved Particle Slope 7.782E+09 3.535E+10 Figure 1 Example of calibration in dissolved and particle. Intercept 0.0578-8.1727 R2 0.99938 0.98719 Table 5 comparison between filtered and unfiltered spiked samples. Analyte Data collected % recovery Sample Mean Size No. of Part. Conc. Diss. Conc. Mean Size No. of Part. Conc. Diss. Conc. (nm) Peaks (parts/ml) (µg/l) (nm) Peaks (parts/ml) (µg/l) Ag 106.905 Ce 139.905 Ti 46.9518 Zn 65.926 filtered 57,93 2780,00 462216,31 0,692 unfiltered 59,98 7126,00 1184803,39 1,144 filtered 64,07 2806,00 466539,20 0,868 unfiltered 72,14 6758,00 1123617,92 1,554 filtered 92,29 4736,00 787430,38 0,953 unfiltered 113,23 9510,00 1581178,82 2,649 filtered 65,63 1764,00 293291,21 2,633 unfiltered 88,31 2708,00 450245,24 8,024 96,60 39,00 39,00 60,40 88,80 41,50 41,50 55,90 81,50 49,80 49,80 36,00 74,30 65,10 65,10 32,80 Table 6 Evaluation of the % recovery of total analytes. Variables Ti Ce Ti Ce Dilution 200 X 200 X 500 X 500 X Theorical value (µg/l) 12000,00 16300,00 12000,00 16300,00 Experimental value (µg/l) 11216,31 11421,27 9949,49 15095,54 % of recovery 93,47 70,07 82,91 92,61 Copyright imedpub 3

This is demonstrated with a recovery test. NP-spiked sample of river water (with the following concentration: Ag 2 µg/l, TiO 2 20 µg/l, CeO 2 40 µg/l and ZnO 200 µg/l) is divided in two quotes. The first is analysed directly and the second (3 L) is filtered. The filter is analysed in the same way of the previous experiments (Table 5). To evaluate more precisely the efficiency of recovery using the filtration, a spiked surface water (with a known concentration of TiO2 and CeO2 NP) is filtered and the total metal content on the filter is analysed (according to UNI EN ISO 17294-1:2007 and UNI EN ISO 17294-2:2016) [13,14] (Table 6). Results To characterize the presence of nanoparticles in water rivers is needed to concentrate them into a filter. This operation allows the reduction of the matrix interference and lowering of background signal as shown in Table 6. In particular there is an abatement in dissolved concentration of Zn up to 67.2%, which permit a reduction of mean size of 25.7%. Figures 2 and 3 show the difference between filtered and unfiltered sample in the river C, in the case of TiO 2 NP. Figure 2 Result of TiO 2 20 µg/l in water filtered. The remarkable reduction of the background signal permits a lower detection limit of the particle size; therefore, the filtration allows to remove the interference of dissolved Ti. As a result, the signal is cleared than the unfiltered one. These images are from the SP-ICP-MS management software. Table 7 Percentage of recovery for river A and B TiO2, CeO2 NP. Rivers Mean particle concentration of filtered samples Mean particle concentration of unfiltered samples River A (TiO2) 675151,6 864259,3 78.1 River B (TiO2) 225147,0 364500,9 61.8 River A (CeO2) 240908,4 248283,5 97 % of recovery (filtered/not filtered parts/ml ratio) Table 7 shows the results of recovery tests for rivers A B in case of TiO2 and CeO2 NP. Figure 3 Result of TiO 2 20 µg/l in water unfiltered. Discussion This research proved that SP-ICP-MS is a good technique to identify and characterize the nanoparticles in surface water. The use of SP-ICP-MS presents some advantages: the high sensitivity of this method permitted to evaluate the presence of NP in a lower concentration (in order of µg/l or ng/l); the simple treatment of the surface waters reduce the cross contamination of sample and the transformation of nanoparticles in ions; Furthermore, the filtration with a vacuum pump using a cellulose filter of 0.22 µm allows to lower the DL both concentration and size. This is possible because the filtration takes out the dissolved ions and depresses the signal of background with a good recovery of nanoparticles onto the filter; moreover, it s demonstrated the positive effect of the filtration in removing the matrix interferences, which could cause a significant false positive in the evaluation of NP. Table 8 Comparison between experimental DL and results from Lee et al. in 2014. DL particle size Lee et al. in 2014 Ag 24.2 13 CeO 2 23.7 21 TiO 2 49.6 91.8 The size detection limit of this study is compare with the data obtained from Lee et al. [15]. The results showed in Table 8. They are similar for CeO 2 and Ag. In the case of TiO 2 there is a significant reduction of DL probably due to removal of matrix interferences. Conclusion The work of near future will be to compare this study with another analytical method as SEM or TEM. The future development will put the attention on the bioaccumulation and biological effects of NP to evaluate the risk assessment perspective. 4 This article is available from: http://nanotechnology.imedpub.com/

References 1. European Food Safety Authority (EFSA) (2016) Annual report of EFSA Scientific Network of Risk Assessment of Nanotechnologies in Food and Feed. pp. 1-14. 2. Salamon AW, Courtney P, Shuttler I (2010) Nanotechnology and engineered nanomaterials: A Primer. PerkinElmer FAQ Manuals 1-19. 3. United States Environmental Protection Agency (EPA) (2011) APM 32 screening methods for metal-containing nanoparticles in water. 4. Chady SKN (2014) Single particle inductively coupled plasma mass spectrometry: Understanding how and why. Perkin Elmer White Paper. pp. 1-5. 5. Pace HE, Rogers NJ, Jarolimek C, Coleman VA, Higgins CP, et al. (2011) Determining transport efficiency for the purpose of counting and sizing nanoparticles via single particle inductively coupled plasma-mass spectrometry. Anal Chem 15: 83. 6. Donovan AR, Shi H, Adams CD, Stephan C (2015) Monitoring cerium dioxide and zinc oxide nanoparticles through drinking water treatments using single particle ICP-MS. PerkinElmer Application Note 38-45. 7. UNI EN ISO 15587-2 (2002) Water quality-digestion for the determination of selected elements in water-part 2: Nitric acid Digestion 1-16. 8. Poscher A, Luidold S, Schnideritsch H, Antrekowitsch H (2016) Extraction of lanthanides from spent polishing agent. Rare Earths Industry 209-222. 9. UNI EN 16173 (2012) Sludge, treated bioiste and soil-digestion of nitric acid soluble fractions of elements. UNI-Ente Nazionale Italiano di Unificazione database 1-11. 10. Gondikas AP, Von der Kammer F, Reed RB, Wagner S, Ranville JF, et al. (2014) Release of TiO2 nanoparticles from sunscreens into surface waters: A One-Year survey at the old danube recreational lake. Enviromental Sci Technol 5415-5422. 11. ISO/TS 19590 (2017) Nanotechnology Size distribution and concentration of inorganic nanoparticles in aqueous media via SP-ICP-MS. ISO-International Organization for Standardization database 1-19. 12. Syngistix Nano Application module rev. 1.0 (2016) Single Particle- Inductively Coupled Plasma Mass Spectrometry (SP- ICPMS). PerkinElmer instrument manual 1-26. 13. UNI EN ISO 17294-1 (2007) Water quality: Application of inductively coupled plasma mass spectrometry (ICP-MS). Part 1: General guidelines. UNI - Ente Nazionale Italiano di Unificazione database. pp. 1-33. 14. UNI EN ISO 17294-2 (2016) Water quality: application of inductively coupled plasma mass spectrometry (ICP-MS). Part 2: Determination of selected elements including uranium isotopes. UNI - Ente Nazionale Italiano di Unificazione database. pp. 1-31. 15. Lee S, Bi X, Reed RB, Ranville JF, Herckes P, et al. (2014) Nanoparticle size detection limits by single particle ICP-MS for 40 elements. Environmental Sci Technology. pp. 1-31. Copyright imedpub 5