Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Similar documents
Determination of Adsorbable Organic Halogen in Wastewater

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Inline sample preparation for the determination of anions in sodium hydroxide

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of adsorbable organic halogen in wastewater using a combustion ion chromatography system

Determination of trace anions in concentrated hydrofluoric acid

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Dionex IonPac AS28-Fast-4µm column

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determination of urea in ultrapure water by IC-MS/MS

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Analysis of various clays with Metrohm Combustion IC

The determination of trace anions in concentrated phosphoric acid

Expanded capabilities in ammonium and amine detection

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

columns IonPac AS17-C Anion Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Determination of trace anions in concentrated glycolic acid

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of trace anions in organic solvents using matrix elimination and preconcentration

The Determination of Trace Anions in Concentrated Phosphoric Acid

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Determination of trace anions in high-nitrate matrices by ion chromatography

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Thermo Scientific. Anion-Exchange Column

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Determination of Trace Anions in Organic Solvents

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of Methylamine in Drug Products

Determination of Trace Anions in Concentrated Hydrofluoric Acid

Overcoming Challenging Matrices in Ion Chromatography

columns IonPac AS18 Anion-Exchange Columns

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

anion-exchange column

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

columns IonPac AS22 Anion-Exchange Column

Determination of Trace Anions in Concentrated Glycolic Acid

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Fast methods for the determination of ibuprofen in drug products

Elemental Analysis: CHNS/O characterization of polymers and plastics

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography

Thermo Scientific Dionex eluent suppressors for ion chromatography

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur using pyrohydrolysis

Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

Technical Report. Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Waters. Ion Chromatography

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Combustion Ion Chromatography. Fast and reliable determination of halogens and sulfur

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Determinations of Heat Stable Amine Salts and Amines in Amine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Utility of the Charge Detector in Ion Chromatography Applications

columns IonPac SCS 1 Silica Cation Separator

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

METHOD 9076 DETERMINATION OF INORGANIC ANIONS BY ION CHROMATOGRAPHY PRODUCTS BY OXIDATIVE COMBUSTION AND MICROCOULOMETRY

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

Elemental Analysis: NCS characterization of paper

Transcription:

APPLICATION UPDATE 72588 Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent Authors Manali Aggrawal and Jeffrey Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Combustion Ion Chromatography (CIC), Dionex IonPac AS23 column, suppressed conductivity detection, polymer, Dionex Aquion IC Goal To develop a CIC method for the determination of Cl, Br, and S in all forms of plastic materials Introduction Polyethylene (PE) is a thermo-plastic created from the polymerization of ethylene, a process that produces long straight chains of hydrocarbon monomers. By adjusting the polymerization process, different kinds of polyethylene with varying degrees of branching in their molecular structures can be made. PE compounds are classified by their molecular weight, density, and branching into major categories; the most common are linear lowdensity polyethylene (LLDPE), low-density polyethylene (LDPE), high-density polyethylene (HDPE), and ultrahigh molecular weight polyethylene (UHMWPE). Different polymer materials are used for different purposes. Polyethylene is widely used in packaging (plastic bags, plastic films, containers including bottles, etc.). Plastic materials consist not only of plastic polymers, as many additives are used to improve different properties of the plastic.

Some additives prevent degradation of the polymer during processing (typical for polyvinyl chloride (PVC)), while others improve resistance to fire or prevent degradation by environmental factors (e.g., UV, temperature, humidity, microorganisms). Halogen and sulfur-containing compounds are often added as plasticizers, flame retardants, and heat stabilizers. Due to the widespread use of plastic/polymers and their subsequent impact on the environment, it is important to know the content of the halogens and sulfur when polyethylene materials are disposed or recycled. Many countries are adapting to regulate these substances. 2 Plastic samples can be very difficult to analyze as sample preparation is required to extract analytes or remove interfering matrices, and these sample preparation techniques can be labor intensive. Combustion ion chromatography (CIC) has been demonstrated for the determination of halogens and sulfur in a variety of difficult samples including coal. 3 In Application Note 72349, a Thermo Scientific Dionex IonPac AS-HC-4µm anion-exchange column was combined with a Thermo Scientific Dionex Integrion HPIC system, which used an electrolytically generated potassium hydroxide (KOH) eluent and suppressed conductivity detection, to determine halogens and sulfur in plastic materials. 4 This application update reports an alternative method using carbonate/bicarbonate eluent with a Thermo Scientific Dionex Aquion IC system. The Dionex Aquion IC system is a simple, inexpensive, and compact platform with straightforward operation for basic ion analysis. The Thermo Scientific Dionex IonPac AS23 carbonate eluent anion-exchange column is used for this application. Here we demonstrate that a Mitsubishi AQF-200H combustion system combined with Dionex Aquion IC system can be used for automated qualitative and quantitative analysis of halogens and sulfur in plastic samples. Four plastic samples were analyzed for their chlorine (Cl), bromine (Br), and sulfur (S) content. Method accuracy was evaluated using a polymer certified reference material (ERM EC680K) for Cl, Br, and S and spike recovery experiments for each analyte. Experimental Equipment Thermo Scientific Dionex Aquion IC system including: Dual Piston Pump Column Oven Temperature Control Thermo Scientific Chromeleon Chromatography Data System (CDS) software, version 7.2 SR5 Mitsubishi AQF-200H system, comprised of: Electric Furnace, HF-20 Gas Absorption Unit, GA-20 External Solution Selector, ES-20 Automatic Boat Controller, ABC-20 2

Consumables Thermo Scientific Dionex AS23 Eluent Concentrate; Sodium Carbonate/Bicarbonate Concentrate (00 ); P/N 0646 Dionex IonPac AS23 Analytical, (2 250 mm); P/N 06445 Dionex IonPac AG23 Guard, (2 50 mm); P/N 06443 Thermo Scientific Dionex AERS 500 Anion Electrolytically Regenerated Suppressor (2 mm); P/N 08254 Instrument setup and installation Configuring the modules in the Chromeleon CDS software To configure the IC system:. Start the Chromeleon Instrument Controller program. 2. Select the link, Configure Instruments, to open the Chromeleon Instrument Configuration Manager. 3. Right-click on the local computer name. 4. Select Add an Instrument. 5. Enter an appropriate name (for example: AquionCIC). 6. Add the following modules to this instrument configuration: Aquion IC system and Remote Inject, as detailed below. Figure. Creating a configuration. Adding Dionex Aquion IC system module. Select Add a Module, Aquion IC System (Figure ). A window with multiple tabs will automatically open up. 2. Select the Model Serial No. on the General tab. (Figure 2) The Mitsubishi AQF-200H system communicates with the Dionex Aquion IC system using TTL signals. Figure 2. Selecting model serial number. 3

3. Open the TTL inputs and make the selections shown in Figure 3. Figure 5. Configuring the Remote Inject Module. Figure 3. Dionex Aquion TTL inputs. Physical TTL connections With a pair of twisted wires, make the connections in Table (Figures 6 and 7). Adding Remote Inject Module Select Add a Module, Generic, and Remote Inject (Figure 4). The default device name is InjectValve and needs to be renamed (for example: InjectValve_CIC). Select Remote Start on Inject Port (Figure 5). Table. TTL connections. Dionex Aquion IC System Mitsubishi AQF-200H System Red Wire TTL- IN Black Wire TTL GND 2 Figure 4. Adding the Remote Inject Module to the instrument. 4

Direct injection using the injection valve in the Mitsubishi GA-20 module This is the most common combustion IC setup. In this setup, the sample loop is installed on the Mitsubishi GA-20 injection valve. When the sample is ready for IC analysis, the Mitsubishi GA-20 module loads the sample loop. The schematic is shown in Figure 8. This setup is recommended for isocratic applications with pressures up to 3000 psi. Please see Technical Note 722 for more details. 5 Figure 8. Direct injection using the injection valve in the Mitsubishi GA-20 module. Figure 6. TTL connections at the rear panel of the Mitsubishi AQF-200H system. Reagents and standards Deionized (DI) water, Type I reagent grade, 8 MΩ cm resistance or better Thermo Scientific Dionex Combined Seven Anion Standard I, 50 ml (P/N 056933) Thermo Scientific Dionex Chloride Standard (000 mg/l), 00 ml (P/N 03759) Thermo Scientific Dionex Sulfate Standard (000 mg/l), 00 ml (P/N 03760) Potassium bromide, Crystal BAKER ANALYZED ACS Reagent (P/N 2998-0) European Reference Material ERM - EC680K Polyethylene (Trace Elements Low Level), ~00 g, ARMI/LGC standards Figure 7. TTL connections at the rear panel of the Dionex Aquion IC system. 5

Table 2A. Combustion parameters. AQF-200H Sample size: ~20 35 mg (2 3 pellets) Sample boat: Quartz Pyrolysis tube: Quartz tube / Quartz wool Absorption solution: H 2 O 2 aqueous solution (30 ppm) Mode: Constant volume HF-20 Furnace inlet temp.: 900 C Furnace outlet temp.: 000 C Argon flow (Carrier): 200 ml/min Oxygen flow (Combustion agent: 400 ml/min GA-20 Absorption tube: 0 ml Sample loop: 00 µl Final absorption solution volume: 0 ml Absorption solution volume*: 3.5 ml Water supply scale: 3 Argon flow for humidification: 00 ml/min GA-20 washing parameters Water injection time: 20 s Drain time: 2 s Washing times: 3 Gas line washing time: 0.5 s Gas line washing interval: 3 s Gas line washing times: 3 Washing time of sample absorption line: 5 s Syringe washing times: 3 GA-20 gas line collection parameters Collection time: 0.5 s Collection interval: 3.0 s Times: 3 Table 2B. Combustion parameters (continued). GA-20 injection parameters Washing time for injection start: 0 s Sample purge time: 5 s Sample absorption time: 0 s ABC-20/ASC-250L Argon time: 0 s Oxygen time: 600 s Position (mm) Wait time (s) Speed (mm/s) 30 90 20 60 90 0.2 End 90 20 Cool 60 40 Home 20 40 Table 3. IC parameters. IC parameters System: Columns: Dionex Aquion IC Dionex IonPac AS23 Analytical, 2 250 mm Dionex IonPac AG23 Guard, 2 50 mm Eluent: 4.5 mm Na 2 CO 3 / 0.8 mm NaHCO 3 Flow rate: 0.36 ml/min Column temp.: 30 C Detection: Suppressed conductivity with Thermo Scientific Dionex ERS 500 Electrolytically Regenerated Suppressor, recycle mode Suppresser current: 9 ma System backpressure: ~2500 psi Background conductance: ~8.5 9.5 µs Noise: ~8.4 0 ns/min peak-to-peak Run time: 8 min *This is the starting absorption volume. Final absorption volume is adjusted to a constant about 0 ml and is determined accurately according to the manual. 6 6

Preparation of solutions and reagents 000 mg/l stock chloride solution Dionex Chloride Standard (000 mg/l), 00 ml (P/N 03759) 000 mg/l stock sulfate solution Dionex Sulfate Standard (000 mg/l), 00 ml (P/N 03760) 000 mg/l stock bromide solution Weigh 0.49 g of dry potassium bromide into a 25 ml polypropylene bottle and tare the balance. Add 00 g of DI water to make a 000 mg/l bromide stock solution. Cap the bottle and shake to completely dissolve the solid material. The standard is stable for one week when stored at 4 C. Working standard solutions Deliver the appropriate volume of the 000 mg/l stock solution into a 25 ml polypropylene bottle and bring to volume (by weight) with DI water. For this application, calibration standards were prepared at 0., 0.2, 0.4, 0.5, 0.75, and.25 mg/l. Aliquots were stored at 40 C and thawed prior to use. Results and discussion By using a CIC system consisting of an automatic quick furnace AQF-200H and Dionex Aquion Ion Chromatography (IC) system, the concentrations of halogens and sulfur were determined in plastic materials. Combustion and IC parameters used for this application are listed in Tables 2 and 3. About 20 to 35 mg of plastic sample are weighed in a quartz sample boat and introduced into system, where the sample is thermally decomposed in argon (Ar) atmosphere, then combusted in oxygen atmosphere. Halogens in the sample are converted to hydrogen halide, and halogen gas and sulfur to sulfur oxide. These components are collected into absorbing solution and converted to halide ion and sulfate ion. For sulfur-containing samples, hydrogen peroxide is added to the absorbing solution to oxidize hydrogen sulfite, resulting from the incomplete oxidation of sulfur, which will be quantified as sulfate in aqueous solution. The resulting solution is analyzed by injecting into an IC system. The Dionex IonPac AS23 column can be used with isocratic carbonate/bicarbonate eluents and suppressed conductivity detection. Due to the higher background conductivity of the carbonate eluent, the carbonate peak does not interfere with the analysis of most of the common anions. However, achieving very low detection limits using a carbonate/bicarbonate eluent system can prove difficult. Figure 9 shows the chromatogram of the combusted plastic certified reference material (ERM-EC680K) containing the anions chloride, bromide, and sulfate separated with a Dionex IonPac AS23 column. The run time is 8 min, slightly longer than that with hydroxide eluent on the Dionex Integrion HPIC system. 4 Depending upon the need, faster run times can be achieved with other carbonate eluent anion-exchange columns such as the Dionex IonPac AS22 fast column. 7 0 - µs/cm 3 Cl SO 4 2-8 0 2 4 6 8 0 2 4 6 8 Br Figure 9. Chromatogram of ERM EC680K. ERM EC680k Boat Blank 7

Calibration There are two methods for calibration. Method uses direct injection of liquid standards through the External Solution Selector ES-20. In this method, calibration standards containing the components of interest are injected and measured directly through IC. Using the calibration curve of each component, the concentration of the component in the sample is calculated using the formula below: Concentration (mg/l) = Cc * A * s * 000 Cc is component concentration (mg/l), A is absorption solution at constant volume (ml), and s is sample weight in mg. Absorption solution at constant volume is determined following instructions in the Mitsubishi Automatic Combustion Unit Model AQF-200H manual. 6 The dilution factor is approximately 0,000 when a 0 ml absorption tube is used. The weight of sample (S) and dilution factor (A*000) are then entered into the data sequence pane of the Chromeleon CDS software console window, which automatically calculates the concentration of the component in the sample. Method 2 uses injection of liquid standards through the Liquid Sample Changer Module ASC-250L. In this method, standard solution prepared with organic solvent is injected into the combustion unit where it is first combusted and then measured by IC. Using the calibration curve of each component, the concentration of each component in the sample is calculated: Concentration (mg/l) = Cc * s Because both the standards and samples are combusted in the same way, the dilution correction factor is not needed. Only the weight of each sample is entered into the data sequence pane of the Chromeleon CDS software. Sample analysis Four plastic materials listed in Table 4 were analyzed for their Cl, Br, and S content. Figure 0 shows chromatograms of plastic samples (, 2, 3, and 4) containing various anions. Besides chloride and sulfate, fluoride, nitrite, and nitrate are also observed in these samples. The concentrations of anions (Cl, Br, and S) were calculated using the calibration method as discussed above in calibration section. Bromide was not found in any of the four samples analyzed. This could be that because of bromide s toxicity it is no longer used as a plastics additive. Sample #2 (transfer pipette) was found to contain very little or almost no chlorine and sulfur. In the samples, sulfur was found to be in low concentration except for in Sample #3 (plastic wrap) where it was almost 60 mg/l. Plastic wrap contains mostly PVC material and PVC has high amounts of sulfur-containing organotin compounds used as heat stabilizers. Table 4. List of plastic materials used in this study. - -3 µs/cm 2 - Sample # Sample Polymer Type Wash bottle LDPE 2 Transfer pipette LDPE 3 Plastic wrap Mostly PVC 4 Bubble wrap LDPE or HDPE Cl NO 2 2 SO 4 F Sample 4 Sample 3 Sample 2 Sample 0 2 4 6 8 0 2 4 6 8 Figure 0. Chromatograms of polymer samples,2, 3, and 4. 8

Precision and accuracy The precision and accuracy for Cl, Br, and S were evaluated using a certified reference material of lowdensity polyethylene (EC680k), by European Reference Materials (ERM, Belgium). Table 5 lists the amount of Cl, Br, and S along with their certified values. The percent relative standard deviation (RSD) is less than 2.0 % for all three anions indicating good precision. The recovery is in the range 98 00%. Table 5. Amount of chlorine, bromine, and sulfur in ERM EC680K. Sample Chlorine Bromine Sulfur 00 94.7 74.0 2 02 96. 77. 3 02 95.5 74.8 4 00 95.0 75.8 5 00 95.0 77. 6 02 93.8 76.4 7 0 94.7 77.3 8 00 95.4 74.5 Average 0 95.0 75.9 Certified value 02 96 76 % Recovery 98.7 99.0 99.8 Figures, 2, 3, and 4 are the chromatograms of the spiking experiment for samples, 2, 3, and 4, respectively. Table 6 contains spike recovery results for chloride and sulfate in the four plastic samples. The percent recovered ranged from a low of 87.8% to a high of 28% for chloride and 88.8% to 96.4% for sulfate. µs/cm 2 Cl 2 SO Sample 4 Br spiked with ERM 0 Sample - -9 0 2 4 6 8 0 2 4 6 8 Figure. Chromatograms of unspiked sample, and sample spiked with ERM pellet. 3 0 Cl SO 4 2 Br Sample 2 spiked with ERM Sample 2 Recovery For the recovery experiment, the plastic sample was spiked with CRM pellets and then the percentage recovery was calculated using the equations below: C spiked sample - C unspiked sample Recovery % = * 00 C added µs/cm -8 0 2 4 6 8 0 2 4 6 8 Figure 2. Chromatograms of unspiked sample 2, and sample 2 spiked with ERM pellet. C spiked sample = Cc *A * * 000 Weight of plastic sample C unspiked sample = Cc *A * * 000 Weight of plastic sample 2 - Cl SO 4 2 Br Sample 3 spiked with ERM Sample 3 C added = Cc *A * * 000 Weight of CRM pellet µs/cm where Cc is the component concentration obtained from the chromatogram, A is the absorption solution at constant volume (ml) determined following the instructions in the Mitsubishi Automatic Combustion Unit Model AQF-200H manual. 6-9 0 2 4 6 8 0 2 4 6 8 Figure 3. Chromatograms of unspiked sample 3, and sample 3 spiked with ERM pellet. 9

µs/cm 4 - -3 Cl SO 4 2-7 0 2 4 6 8 0 2 4 6 8 Figure 4. Chromatograms of unspiked sample 4, and sample 4 spiked with ERM pellet. Br Sample 4 spiked with ERM Sample 4 Conclusion The results demonstrate that the Dionex Aquion IC system combined with the Mitsubishi AQF-200H system can be used for both qualitative and quantitative analysis of Cl, Br, and S in plastic samples. The method was validated using certified reference material ERM EC680K with recoveries within 98 to 00%. This automated method is simple, easy to use, saves time, and produces fewer environmental contaminants than other sample preparation procedures, such as acid digestions or back extractions from organic solvents. This analysis will allow polymer manufacturers to monitor halogen/ sulfur concentration levels to ensure compliance with government regulations. Table 6. Spike recovery for samples, 2, 3, and 4. Unspiked Sample # Added Spiked Recovery (%) Chlorine 20.6 98.6 35 6 Sulfur 0 82.2 76.6 93.2 Sample #2 Chlorine 3.09 95.8 06 07 Sulfur 2.30 79.5 75.5 92. Sample #3 Chlorine 90.4 97.5 76 87.8 Sulfur 62.0 76.9 30 88.8 Sample #4 Chlorine 55.3 98. 8 28 Sulfur 30.8 79.9 08 96.4 References. Deanin, R.D. Additives in plastics. Environ Health Perspect. 975,, 35 39. 2. DIN EN 6232-3-2: Determination of certain substances in electrotechnical products - Part 3-2: Screening of the halogens (fluorine, bromine and chlorine) in polymer and electronics by Combustion - Ion Chromatography (C-IC), 206. [Online] http://www.beuth.de/de/norm-entwurf/din-en-6232-3-2/25640093 3. Thermo Fisher Scientific Application Note 45: Determination of Halogens in Coal Using Combustion Ion Chromatography, 206 Sunnyvale, CA. [Online] https://tools.thermofisher.com/content/sfs/brochures/an-45-ic-halogens-coal- AN7844-EN.pdf 4.Thermo Fisher Scientific Application Note 72349: Determination of Chlorine, Bromine, and Sulfur in Polyethylene Materials using Combustion Ion Chromatography, 207. Sunnyvale, CA. [Online] https://appslab.thermofisher.com/app/3830/ an72349-determination-chlorine-bromine-sulfur-polyethylene-materials-using-combustion-ion-chromatography 5. Thermo Fisher Scientific Technical Note 722: Combustion Ion Chromatography with a Dionex Integrion HPIC System, 207. Sunnyvale, CA. [Online] https://tools. thermofisher.com/content/sfs/brochures/tn-722-ic-integrion-tn722-en.pdf 6. Mitsubishi Chemical Analytech. Operation Manual for NSX-200 Series Automatic Combustion Unit Model AQF-200H, Instruction Manual of Absorption Unit GA-20 section. 7. Thermo Fisher Scientific Dionex IonPac AS22 Fast IC Columns. [Online] https://www.thermofisher.com/order/catalog/product/079936 8. Thermo Fisher Scientific Manufacturing and Processing Learning Center - Polymers and Plastics Information. https://www.thermofisher.com/us/en/home/industrial/ manufacturing-processing/manufacturing-processing-learning-center/polymers-plastics-information.html Find out more at thermofisher.com/combustionic 208 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise stated. Mitsubishi is a trademark of Mitsubishi Corporation. BAKER ANALYZED is a registered trademark of Avantor Performance Materials, LLC. ERM is a registered trademark of European Community International Governmental Organization Erpn Cmnty TM Ofc. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AU72588-EN 08S