ACS after SM4: RELATIVE GAIN VALUES AMONG THE FOUR WFC AMPLIFIERS

Similar documents
FLAT FIELDS FROM THE MOONLIT EARTH

Earth Flats. 1. Introduction. Instrument Science Report ACS R. C. Bohlin, J. Mack, G. Hartig, & M. Sirianni October 25, 2005

SBC FLATS: PRISM P-FLATS and IMAGING L-FLATS

WFC3 TV2 Testing: UVIS-2 Dark Frames and Rates

Study of the evolution of the ACS/WFC sensitivity loss

The in-orbit wavelength calibration of the WFC G800L grism

Updated flux calibration and fringe modelling for the ACS/WFC G800L grism

Persistence in the WFC3 IR Detector: Spatial Variations

WFC3 Calibration Using Galactic Clusters

Cross-Talk in the ACS WFC Detectors. I: Description of the Effect

SPACE TELESCOPE SCIENCE INSTITUTE. ACS Polarization Calibration: Introduction and Progress Report

Here Be Dragons: Characterization of ACS/WFC Scattered Light Anomalies

NICMOS Status and Plans

NIRSpec Performance Report NPR / ESA-JWST-RP Authors: T. Böker, S. Birkmann & P. Ferruit Date of Issue: 20.5.

First On-orbit Measurements of the WFC3-IR Count-rate Non-Linearity

WFC3/UVIS and IR Multi-Wavelength Geometric Distortion

11041,11042,11043: ACS CCD Daily Monitor

WFC3/IR Persistence as Measured in Cycle 17 using Tungsten Lamp Exposures

The Impact of x-cte in the WFC3/UVIS detector on Astrometry

ACS CCDs UV and narrowband filters red leak check

WFC3 TV2 Testing: IR Channel Read Noise

Absolute Flux Calibration for STIS First-Order, Low-Resolution Modes

WFC3/UVIS Photometric Transformations

Calibration of ACS Prism Slitless Spectroscopy Modes

WFC3 IR Blobs, IR Sky Flats and the measured IR background levels

SBC L-Flat Corrections and Time-Dependent Sensitivity

Analysis of wavelength shifts reported for X-shooter spectra

An Algorithm for Correcting CTE Loss in Spectrophotometry of Point Sources with the STIS CCD

Accuracy of VIMOS night-time and daytime wavelength calibrations

Time Dependence of ACS WFC CTE Corrections for Photometry and Future Predictions

Breathing, Position Drift, and PSF Variations on the UVIS Detector

Improved Photometry for G750L

High Signal-to-Noise, Differential NICMOS Spectrophotometry

TIPS-JIM Meeting 13 October 2005, 10am, Auditorium

The TV3 ground calibrations of the WFC3 NIR grisms

Data Reduction - Optical / NIR Imaging. Chian-Chou Chen Ph319

Grism Sensitivities and Apparent Non-Linearity

Exploring Data. Keck LRIS spectra. Handbook of CCD Astronomy by Steve Howell Chap. 4, parts of 6

WFC3 TV3 Testing: Orbital Cycling Effects on IR Images

Astrometric Performance of STIS CCD CTI Corrections on Omega Cen Images

1 The Preliminary Processing

HST Temporal Optical Behavior: Models and Measurements with ACS

Focus-diverse, empirical PSF models for the ACS/WFC

On-orbit Calibration of ACS CTE Corrections for Photometry

Report on the new EFOSC2 VPH grisms

NICMOS Cycles 13 and 14 Calibration Plans

Technical Note 103. Response Enhancement for the Loral CCDs. using Ultra-Violet Flooding. A. P. Oates (RGO)

COS FUV Focus Determination for the G140L Grating

Commissioning of the Hanle Autoguider

WFPC2 Clocks-ON Close Out

Dark Rate of the STIS NUV Detector

NICMOS Focus Field Variations (FFV) and Focus Centering

ACS PSF variations with temperatures

Evaluation of a temperaturebased

Delivery of a new ACS SBC throughput curve for Synphot

Wavelength Calibration Accuracy for the STIS CCD and MAMA Modes

WFC3 Cycle 18 Calibration Program

Calibration Goals and Plans

Kepler photometric accuracy with degraded attitude control

RADIATION DAMAGE IN HST DETECTORS

Reduction procedure of long-slit optical spectra. Astrophysical observatory of Asiago

WFPC2 Dark Current vs. Time

Internal Flat Field Monitoring II. Stability of the Lamps, Flat Fields, and Gain Ratios

Atmospheric Extinction

IRS-TR 04001: Photoresponse and Read Noise Trends

Guidelines. H.1 Introduction

The Effective Spectral Resolution of the WFC and HRC Grism

FPA#64 and new substrate removed FPAs for WFC3-IR: a trade-off study

estec NIRSpec Technical Note NTN Spectral intensity of the MSA glow sources Abstract:

WFPC2 Cycle 7 Calibration Plan

The Accuracy of WFPC2 Photometric Zeropoints

COS Cycle 17: 20 programs 149 external orbits 446 internals FUV Detector Sensitivity Monitor 36 0 Oct. 3, 2010

Monitoring The HRC-S UV Rate: Observations of Vega

Ground and On-Orbit Characterization and Calibration of the Geosynchronous Imaging Fourier Transform Spectrometer (GIFTS)

A Python Script for Aligning the STIS Echelle Blaze Function

You, too, can make useful and beautiful astronomical images at Mees: Lesson 3

Spitzer Space Telescope

Impacts of focus on aspects of STIS UV Spectroscopy

Lab 4: Differential Photometry of an Extrasolar Planetary Transit

SMOV Absolute Flux Calibration of the COS FUV Modes

TECHNICAL REPORT. Doc #: Date: Rev: JWST-STScI , SM-12 August 31, Authors: Karl Gordon, Ralph Bohlin. Phone:

Status of the EPIC calibration

Requirements for the Star Tracker Parallel Science Programme

REDUCED LIGHT CURVES FROM CAMPAIGN 1 OF THE K2 MISSION

The WFC3 IR Blobs Monitoring

Trace and Wavelength Calibrations of the UVIS G280 +1/-1 Grism Orders

Point-Source CCD Photometry with STIS: Correcting for CTE loss

BetaDrizzle: A Redesign of the MultiDrizzle Package

Data Processing in DES

The SDSS Data. Processing the Data

Keck Adaptive Optics Note 1069

Predicted Countrates for the UV WFC3 Calibration Subsystem using Deuterium Lamps

Determining the Orbital Period of the Cataclysmic Variable CSS1204 Using Data from the Vatican Advanced Technology Telescope

Planet scintillation as a probe for ground-layer turbulence

CCD OPERATION. The BBD was an analog delay line, made up of capacitors such that an analog signal was moving along one step at each clock cycle.

Week 7. Observing happens the week after midterm.

Photometric Techniques II Data analysis, errors, completeness

NIRSpec Performance Report NPR / ESA-JWST-RP Author(s): Date: October 11, Stephan Birkmann, Catarina Alves de Oliveira

Characterisation & Use of Array Spectrometers

Attempts to Mitigate Trapping Effects in Scanned Grism Observations of Exoplanet Transits with WFC3/IR

Transcription:

Instrument Science Report ACS 2009-03 ACS after SM4: RELATIVE GAIN VALUES AMONG THE FOUR WFC AMPLIFIERS R. C. Bohlin, A. Maybhate, & J. Mack 2009 October 8 ABSTRACT For the default setting of gain=2, the individual gain values of the four WFC amplifiers are determined from internal flat field observations. The average absolute gain remains unchanged, but matching the flat fields at the boundaries of the four quadrants provides a more accurate determination of the relative gains among the four separate amplifiers. 1. Introduction Following the HST Servicing Mission in 2009 May (SM4), new electronics are now available for controlling ACS. The available gains (G) with the new electronics are 0.5, 1.0, 1.4, and 2.0 (e-/dn); but only G=2 is being used for scientific observations in cycle 17. In cycle 18, only G=2 will be supported. CCDs can approach signal detection at the quantum noise limit; however, CCDs are basically analog detection devices. The analog signal charge is read, amplified, and then digitized with an A-to-D converter, which for ACS is 16 bits (0-65535). Because of read noise, distinguishing one electron of signal is rarely useful, so that the signal can be compressed with little loss of information before the digitization step. For example at G=2, one Analog Data Unit (ADU) of digitized signal corresponds to 2 electrons of detected charge. Therefore, the ACS/WFC CCD detector with its full well of maximum charge of ~85000 electrons can be digitized without saturating the A-to-D converter at G=2. A nearly full well of 80000 electrons will be digitized for the telemetry stream at a value of 40000 ADU, which is well below the A- to-d converter saturation limit of 65535. With the new electronics package installed during SM4, the new absolute gain values for the four ACS WFC detector quadrants (A, B, C, D) have been measured during the SM4 ACS Optimization Campaign by P. L. Lim (private comm.). The accuracy of this photon-transfer method that analyzes the noise properties of the signal is only ~5%. 1

An exact measure of the gain for any particular mode is of little import, because any error is compensated exactly in the calibration of the absolute flux sensitivity. However, to avoid different sensitivity calibrations for different modes, the exact value of the gain relative to a fiducial gain value is important and should be measured precisely. There are two types of relative gain a) the ratio of the signals for the same source in different nominal gain settings and b) the relative response among the four WFC quadrants at the same gain setting. Following a precise measurement for b), the ratios a) should be the same for all four ABCD quadrants. For ACS, Gilliland (2004) measured the type a) gain ratio by comparing the same star fields taken at the different gains. Using this same procedure for calibrating the post-sm4 electronics, the exact new G=2 value relative to the same original pre-sm4, fiducial G=1 setting should be measured by comparing stellar signals in the same field as previously used. The fractional adjustment from the star field analysis will apply to the 2.036 average and to the individual ABCD gains quoted in line 3 of Table 1. The analysis of the 47 Tuc star field by Mack, Desjardins, & Bohlin (2009 in prep) is used to define the type a) relative gains and produce the final gain values listed in the last line of Table 1. For the type b) relative gain that is the topic of this ISR, internal flat field observation define the relative gain among the ABCD quadrants. The relative gain between pairs of amplifiers, i.e. A/B on chip 1 and C/D on chip 2, can be derived from internal flat field images to ~0.1% precision by matching the signal across the two boundaries. The flat fields also provide a more precise determination of the gain ratio between the two WFC chips. However, the chip1/chip2 sensitivity ratio is wavelength dependent. The chip1/chip2 relative gains are reconciled using the internal flats at F625W following the method of Bohlin, Hartig, and Sparks (2002). 2. ANALYSIS Table 1 shows the average gain and the gain ratios for pre-failure in line 1 and after SM4 in lines 2-4. The average gain value of 2.036 for the G=2 setting in Table 1 is unchanged by our adjustments from line 2 to line 3 for the individual A, B, C, and D amplifiers. Table 2 lists the observations analyzed and discussed below. As normally displayed and as in Figures 1-2, the positions of A to D are in order like you read a book with A at the top left and D at the bottom right. The relative gains are measured along eight pixel wide strips at adjacent edges of the amplifier pairs on each chip and at the adjacent edges, where the two chips are butted together. The IDL program gainrat.pro compares the sum of three F625W internal lamp images obtained on day 190 (July 9) of 2009 to a similar set obtained on 2007 January 11. While our relative A/B and C/D gains measured from the relative response at the adjacent edges of the AB and CD amp pairs are accurate and repeatable to 0.1%, relative values between the chips depends on the assumption of equal sensitivities at F625W. The sensitivity from chip 1 to chip 2 along the chip gap is expected to be similar, because the two chips were cut from the same thinned silicon wafer and are positioned in the same orientation as on the original wafer. With the Table 1 gains derived to make chip 1 match chip 2 for F625W, the 2

average chip1/chip2 response varies with wavelength, i.e. filter, by <0.7%, except for F850LP, where chip 1 is 1.2% more sensitive than chip 2. However, this same sensitivity variation with filter repeats in the flat field reference files themselves, so that the flux calibrations with the new electronics should remain the same for both chips. 3. UNCERTAINTIES While the gain values derived here attempt to make the four quadrants within a raw flat field image match, the more important purpose is to make the SM4 flat fields match the pre-failure flats obtained at the same CCD set point temperature of -81C. For the F435W, F625W, and F814W internal flats obtained on 2009 July 9-10, the match is better than ~0.2% for any systematic differences in 8x8=64 bins of 512x512 pixels, except for the lower left corner of F435W where that bin is 0.3% low. However, the F850LP internal flat shows changes of up to 0.4% over an extended region, as shown in Figure 1, while Figure 2 shows the expected uniform ratio for F814W. The pattern in Figure 1 is strongly reminiscent of Fig. 2 of Bohlin et al. (2005), where the cause is attributed to the known shutter light leak of the WFC. The light that leaks around the shutter has a fixed equivalent exposure time that depends on how long the lamp is on with the shutter closed. Thus, shorter exposures have a larger fractional contamination from extra spurious signal in the blacker region of our Figure 1. The three 2007 images comprising the denominator of Figure 1 have an exposure time of 2s, while the 2009 numerator images have a 4.2s duration. Thus, the denominator has a larger fractional contamination, making the ratio less than one in the contaminated region. The exposure times of the other three filters with 2009 data are all identical to the exposure times of the comparison data obtained in 2007. With a maximum difference of ~1% between the values in line 2 vs. line 3, our repartitioned gains in Table 1 agree with the absolute gain measurements from the SM4 Optimization Campaign. 3

Table 1. Average Absolute Gain and Relative Individual Gain Values Average Amp A Amp B Amp C Amp D Comment 1.992 2.002 1.945 2.028 1.994 Gilliland (2004) 2.036 2.090 1.918 2.092 2.045 SM4 optimization campaign 2.036 2.074 1.936 2.071 2.065 This work 1.983 2.020 1.886 2.017 2.011 Mack et al.(2009 in prep) Table 2. Internal Lamp WFC Observations Analyzed ROOTNAME DATE-OBS Exp(s) PROPOSID FILTER j9v9w1obq 2007-01-10 17.6 11052 F435W j9v9w1odq 2007-01-10 17.6 11052 F435W j9v9w1ofq 2007-01-10 17.6 11052 F435W ja9aw5y0q 2009-07-09 17.6 11374 F435W ja9aw5y2q 2009-07-09 17.6 11374 F435W ja9aw5y4q 2009-07-09 17.6 11374 F435W j9v9w7snq 2007-01-11 3.0 11052 F625W j9v9w7spq 2007-01-11 3.0 11052 F625W j9v9w7srq 2007-01-11 3.0 11052 F625W ja9aw6y6q 2009-07-09 3.0 11374 F625W ja9aw6y8q 2009-07-09 3.0 11374 F625W ja9aw6yaq 2009-07-09 3.0 11374 F625W j9v9wbxlq 2007-01-12 2.9 11052 F814W j9v9wbxnq 2007-01-12 2.9 11052 F814W j9v9wbxpq 2007-01-12 2.9 11052 F814W ja9aw7duq 2009-07-10 2.9 11374 F814W ja9aw7dwq 2009-07-10 2.9 11374 F814W ja9aw7dyq 2009-07-10 2.9 11374 F814W j9v9wcxsq 2007-01-12 2.0 11052 F850LP j9v9wcxuq 2007-01-12 2.0 11052 F850LP j9v9wcxwq 2007-01-12 2.0 11052 F850LP ja9aw8xtq 2009-07-09 4.2 11374 F850LP ja9aw8xvq 2009-07-09 4.2 11374 F850LP ja9aw8xxq 2009-07-09 4.2 11374 F850LP 4

Fig. 1. The ratio for F850LP of the sum of three internal lamp images obtained on day 190 in 2009 (lpwi09190sm03f850lp) to a similar sum obtained on day 12 of 2007 (lpwi07012sm03f850lp). The new gain values from the last line of Table 1 are applied to the 2009 data. The scaling is from 0.995 to 1.005 per the grayscale reference bar at the top. The blacker region near the chip 1 and 2 boundary between amps A and C indicates a systematic difference in the flat fields before and after SM4 at the ~0.4% level. This difference is caused by the shutter light leak. 5

Fig. 2. As in Figure 1 for the ratio in F814W. The artifacts at the <0.1% level in both Figures are probably caused by slight differences in illumination of the filter because of mechanical positioning tolerances of the filter wheel, as suggested by the slight vignetting visible in the lower left and upper right corners. The offset of 0.12% between amps A and B is due to a 0.01% offset in the numerator amplified by a 0.11% offset of the denominator. 6

REFERENCES Bohlin, R. C., Hartig, G., & Sparks, Wm. 2002, Instrument Science Report, ACS 2002-03, (Baltimore:STScI) Bohlin, R. C., & Mack, J., Hartig, G., & Sirianni, M. 2005, Instrument Science Report, ACS 2005-12, (Baltimore:STScI) Gilliland, R. L. 2004, Instrument Science Report, ACS 2004-01, (Baltimore:STScI) Mack, J., Desjardins, T., & Bohlin R. 2009, Instrument Science Report, ACS 2009-xx, (Baltimore:STScI), in preparation 7