Dionex IonPac AS28-Fast-4µm column

Similar documents
columns IonPac AS17-C Anion Exchange Column

Thermo Scientific. Anion-Exchange Column

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

columns IonPac AS18 Anion-Exchange Columns

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

anion-exchange column

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

columns IonPac AS22 Anion-Exchange Column

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Determination of trace anions in concentrated hydrofluoric acid

Thermo Scientific Dionex IonPac AS11 and AS11-HC Anion-Exchange Column

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

columns I o n P a c AS11 and AS11-H C Anion-E x c h a n g e C o l u m n s

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Concentrator and Trap Columns. Column: IonPac AG15, AS15, 2 mm. Eluent Source: EG40 Concentrator. Temperature: 30 C Column: IonPac AC15 (2 50 mm)

for IonPac AG11-HC IonPac AS11-HC

Inline sample preparation for the determination of anions in sodium hydroxide

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Thermo Scientific Dionex IonPac CS17 Cation-Exchange Column

Thermo Scientific Dionex eluent suppressors for ion chromatography

Expanded capabilities in ammonium and amine detection

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

The determination of trace anions in concentrated phosphoric acid

Determination of urea in ultrapure water by IC-MS/MS

Determination of trace anions in high-nitrate matrices by ion chromatography

columns IonPac SCS 1 Silica Cation Separator

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Utility of the Charge Detector in Ion Chromatography Applications

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

IonPac Trace Cation Concentrator

High-Performance Chromatography Superior isocratic separation of

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

Determination of Organic Acids in Beer Samples Using a High-Pressure Ion Chromatography System

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Determination of Trace Anions in Organic Solvents

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Overcoming Challenging Matrices in Ion Chromatography

suppressors Eluent Suppressors for Ion Chromatography

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

The Determination of Trace Anions in Concentrated Phosphoric Acid

The power to increase productivity. Thermo Scientific Dionex Reagent-Free Ion Chromatography (RFIC) System Capabilities

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Cation Exchange HPLC Columns

Determination of trace anions in concentrated glycolic acid

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

IC Columns Introduction

groups I and II cations, amines, divalent transition metals

suppressors SRS -ULTRA Self-Regenerating Suppressor

Determination of Trace Cations in Power Plant Waters Containing Morpholine

PRODUCT MANUAL. OMNIPAC PAX-100 GUARD COLUMN (4 x 50 mm, P/N ) OMNIPAC PAX-500 GUARD COLUMN (4 x 50 mm, P/N )

ProPac WCX-10 Columns

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Determination of Inorganic Anions in Wastewater by Ion Chromatography

Determination of Adsorbable Organic Halogen in Wastewater

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Determination of trace anions in organic solvents using matrix elimination and preconcentration

Determination of Common Inorganic Anions in Environmental Waters Using a Hydroxide Selective Column

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Determinations of Heat Stable Amine Salts and Amines in Amine Scrubbing Solutions Using Suppressed Conductivity and Charge Detection

IonPac SCS 1 IonPac SCG 1

Anion Exchange HPLC Columns

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

Sample Preparation for Ion Chromatography

Determination of Trace Anions in Concentrated Hydrofluoric Acid

4. Ion chromatography (IC) 4.1. Educational aims and objectives

INSTALLATION INSTRUCTIONS and TROUBLESHOOTING GUIDE. for the OnGuard- CARTRIDGE

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

IC5000 Ion Chromatography System

Product Manual for Dionex IonPac CS12A and CG12A Columns

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

Ion Chromatography: Green Chemistry for a Green Environment. Ken Kirkbride Somerset, New Jersey October 12, 2011

Determination of Trace Anions in Concentrated Glycolic Acid

Rapid Post-Blast Inorganic Explosive Analysis by Suppressed Ion Chromatography

PRODUCT MANUAL. for the. IONPAC CG14 GUARD COLUMN (2 x 50 mm, P/N ) (4 x 50 mm, P/N )

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Suppression in anion chromatography. More sensitive analysis of anions and organic acids

ION CHROMATOGRAPHY SYSTEM S 150

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Development of the Thermo Scientific Dionex IonPac AS26 Column for Haloacetic Acid Analysis

PRODUCT MANUAL OMNIPAC PCX-500 GUARD COLUMN (PCX-500 GUARD, P/N ) OMNIPAC PCX-500 ANALYTICAL COLUMN (PCX-500 ANALYTICAL, P/N )

Transcription:

CHROMATOGRAPHY Thermo Scientific Dionex IonPac AS-Fast-4µm Columns Product Specifications The Thermo Scientific Dionex IonPac AS-Fast-4µm column is a high-capacity, hydroxide-selective anionexchange column designed for the determination of trace-level concentrations of inorganic anions and low molecular weight organic acids, including fluoride, glycolate, acetate, formate, chloride, nitrite, carbonate, sulfate, oxalate, bromide, nitrate, and phosphate in highpurity water matrices encountered in the semiconductor and power generation industries. The Dionex IonPac AS-Fast-4µm column offers excellent retention of fluoride from the water dip and uses a simple hydroxide eluent easily supplied by the Eluent Generator (EG)..0 Gradient Separation of Trace Organic Acids and Inorganic Anions on a Dionex IonPac AS-Fast-4µm (4 0 mm) Column 4 0 Peaks: µg/l (ppb). Fluoride.0. Acetate.0. Glycolate.0 4. Formate.0. Chloride.. Nitrite.0. Carbonate.0. Sulfate.0. Oxalate.0 0. Bromide 0.0. Nitrate 0.0. Phosphate.0 The Dionex IonPac AS-Fast-4µm column is available in formats from 0.4 to 4 mm, allowing use of capillary to analytical flow rates, and supporting advanced IC IC applications. -0. 0 0 0 Superior Chromatographic Performance Universal column for trace-level concentrations of inorganic anions and low molecular weight organic acids in high-purity water matrices. The Dionex IonPac AS-Fast-4µm column offers improved peak shapes, efficiencies, and resolution compared to the Dionex IonPac AS column and is the recommended replacement for the Dionex IonPac AS column. Easily transfer applications from the Dionex IonPac AS column to the Dionex IonPac AS-Fast-4µm column due to the similar selectivity. The Dionex IonPac AS-Fast-4µm Capillary column (0.4 0 mm) offers reduced eluent consumption, reduced operating costs, and excellent reproducibility. Column selectivity optimized for a 0 C operating temperature to ensure reproducible retention times. Large-loop injections for ppb-level analysis. Superior retention and resolution of early-eluting anions, including fluoride, glycolate, acetate, and formate. Simplified operation is provided by the EG, which requires only a deionized water source to produce potassium hydroxide eluent. Hydroxide eluent suppresses to a very low background for trace-level determinations. Compatible with organic solvents to enhance analyte solubility, modify column selectivity, and facilitate effective column cleanup. Direct transfer of 4 mm applications to the mm and 0.4 mm column formats for more economical operation, reduced eluent consumption, and increased mass sensitivity.

High-Efficiency Particle Structure The Dionex IonPac AS-Fast-4µm packing is a unique structure composed of a highly crosslinked core and a MicroBead anion-exchange layer attached to the surface, as shown in Figure. The Dionex IonPac AS-Fast-4µm column substrate is composed of a supermacroporous 4 µm diameter resin bead consisting of ethylvinylbenzene crosslinked with % divinylbenzene. The anion-exchange layer is functionalized with alkanol quaternary ammonium groups and has a controlled thickness, which results in excellent mass transfer characteristics and consequently, highly efficient peaks. Versatile Capability The Dionex IonPac AS-Fast-4µm column is ideal for the isocratic separation of fluoride, chloride, carbonate, nitrite, sulfate, bromide, nitrate, and phosphate in high purity water samples using a simple hydroxide eluent and suppressed conductivity detection. Figure illustrates the excellent separation of chloride from carbonate, with highly retained analytes, such as phosphate, eluting in less than minutes. 4µm Capillary Format The Dionex IonPac AS-Fast-4µm Capillary column (0.4 0 mm) is packed with the same material as the equivalent standard bore version (producing the same performance as a 4 mm column), but requires only /00th the eluent flow rate. The capillary format offers the advantage of less eluent consumption providing reduced operating and disposal costs. Figure illustrates the effect of flow rate on run time using an isocratic eluent and the Dionex IonPac AS-Fast-4µm (0.4 0 mm) capillary column. By increasing the flow rate, common anions can be analyzed in under minutes with no change in selectivity. Excellent retention time reproducibility can be achieved with the capillary format. Figure. Structure of a Dionex IonPac AS-Fast-4µm packing particle. 0 4 4 Eluent: 0 mm KOH - 0 0 0 Eluent: mm KOH Figure. Effect of eluent concentration on the isocratic separation of common anions using the Dionex IonPac AS-Fast-4µm ( 0 mm) column. 0 4 4 4 Flow Rate: µl/min -0 0 0 0 Flow Rate: 0 µl/min Flow Rate: µl/min AS-Fast-4µm ( 0 mm) Eluent: See Chromatogram Eluent Source: Dionex EGC 00 KOH Flow Rate: 0. ml/min Inj. Volume:. µl Temperature: 0 C Dionex AERS 00 ( mm) Peaks: mg/l. Fluoride.0. Chloride.0. Carbonate 4. Nitrite 0.0. Sulfate 0.0. Bromide 0.0. Nitrate 0.0. Phosphate 0.0 AS-Fast-4µm (0.4 0 mm) Eluent: mm KOH Eluent Source: Dionex EGC 00 KOH Flow Rate: See Chromatogram Inj. Volume: 0.4 µl Temperature: 0 C Dionex ACES 00 Peaks: mg/l. Fluoride 0.. Chloride.. Carbonate 4. Nitrite.. Sulfate.. Bromide.0. Nitrate.0. Phosphate. Figure. Isocratic separation of common anions using the Dionex IonPac AS-Fast-4µm (0.4 0 mm) capillary column at various flow rates.

Large Loop Injections for μg/l (ppb) Level Analysis The selectivity and high capacity of the Dionex IonPac AS-Fast-4µm column allows the determination of trace levels of inorganic anions and low molecular weight organic acids in high-purity water samples using a large-loop injection. Figure 4 illustrates the separation of inorganic anions and low molecular weight organic acids in a high-purity water sample using this technique. Separation was achieved using a potassium hydroxide gradient with suppressed conductivity detection at a temperature of 0 C. Low μg/l (ppb) levels of these analytes can easily be quantified using a ml injection loop with a 4 mm column, a 00 µl injection loop with a mm column, or a 0 µl injection loop with a 0.4 mm capillary column. The conductivity of the hydroxide eluent can be suppressed to a very low background, facilitating trace-level analysis. Extended Application Capabilities The unique selectivity and high capacity of the Dionex IonPac AS-Fast-4µm column make it an ideal column for methods development of specialized anion applications. For example, anionic additives in personal care products can easily be determined. Figure illustrates the excellent separation of fluoride, chloride, monofluorophosphate, sulfate, and phosphate found in dental care products. Trace Analysis is Easier with an Eluent Generator Combining the Dionex IonPac AS-Fast- 4µm column with an EG makes trace anion analysis even easier. An EG makes potassium hydroxide eluent electrolytically from water. The use of this ultrapure hydroxide as an eluent results in a low background and stable baseline.gradients can be programmed with ease on an EG, facilitating method optimization. Figure illustrates the optimization of a hydroxide gradient and flow rate to reduce analysis time of low molecular weight organic acids and inorganic anions commonly encountered in the chemical, semiconductor, and power generation industries...0 0-0 0 Eluent: mm KOH from 0 to 4 minutes, -0 mm KOH from 4 to minutes, 0- mm KOH from to0 minutes, mm KOH from 0 to 0 minutes -0. 0 0 0 4 4 4 0 Flow Rate: µl/min Eluent: mm KOH from 0 to minutes, Flow Rate: µl/min -0 mm KOH from to minutes, 0- mm KOH from to minutes, 0 mm KOH from to minutes 4 Eluent: -0 mm 0- mm KOH from to minutes, mm KOH from to minutes Standard Toothpaste Sample A Toothpaste Sample B -.0 0 0 0 0 0 0 AS-Fast-4µm ( 0 mm) Eluent: mm KOH from 0 to mins, -0 mm KOH from to min, 0- mm KOH from to min, mm KOH from to min Eluent Source: Dionex EGC 00 KOH Flow Rate: 0. ml/min Inj. Volume: 00 µl Temperature: 0 C Dionex AERS 00 ( mm) Suppressor, AutoSuppression, Peaks: µg/l (ppb). Fluoride.0. Carbonate.0. Acetate.0. Sulfate.0. Glycolate.0. Oxalate.0 4. Formate.0 0. Bromide 0.0. Chloride.. Nitrate 0.0. Nitrite.0. Phosphate.0 Figure 4. Determination of trace organic acids and inorganic anions using a hydroxide gradient, large loop injection, and the Dionex IonPac AS-Fast-4µm ( 0 mm) column. AS-Fast-4µm ( 0 mm) Eluent: mm KOH from 0 to mins, -0 mm KOH from to mins, 0- mm KOH from to mins, mm KOH from to mins Eluent Source: Dionex EGC 00 KOH Flow Rate: 0. ml/min Inj. Volume:. µl Temperature: 0 C Dionex AERS 00 ( mm) Peaks: mg/l (ppm). Fluoride.0. Acetate 0.. Glycolate 0. 4. Formate 0.. Chloride 0.. Carbonate. Monofluorophosphate.0. Sulfate.0. Phosphate.0 Figure. Determination of anionic additives in toothpaste using the Dionex IonPac AS-Fast-4µm ( x 0 mm) column. AS-Fast-4µm (0.4 0 mm) Eluent: See Chromatogram Eluent Source: Dionex EGC 00 KOH Flow Rate: See Chromatogram Inj. Volume: 0.4 µl Temperature: 0 C Dionex ACES 00 Suppressor, AutoSuppression, Peaks: mg/l (ppm). Fluoride 0.. Carbonate.. Acetate.. Sulfate.. Glycolate.. Oxalate. 4. Formate. 0. Bromide.0. Chloride 0.. Nitrate.0. Nitrite.. Phosphate. Figure. Optimization of hydroxide gradient and flow rate to reduce analysis time with the Dionex IonPac AS-Fast-4µm (0.4 0 mm) capillary column.

4 Trace Analysis of Inorganic Anions in High Ionic Strength Matrices High capacity columns allow large loop injections while maintaining the peak shapes of trace components, even in the presence of high concentrations of other anions (e.g., carbonate) during purity analysis of salts such as sodium bicarbonate and sodium carbonate. Figure shows the use of a large loop injection on the Dionex IonPac AS-Fast- 4µm ( 0 mm) column for trace analysis of chloride and sulfate in samples of sodium bicarbonate. Using an isocratic potassium hydroxide eluent, the bicarbonate peak is well resolved from chloride and sulfate and does not interfere with quantification of these trace anions. Figure shows the purity analysis of sodium carbonate using a large loop injection on the Dionex IonPac AS-Fast-4µm ( 0 mm) column. Determination of Inorganic Anions and Low Molecular Weight Organic Acids Low molecular weight organic acids and mono- and divalent inorganic anions commonly encountered in industrial process solutions and chemicals can be determined in a single run using the Dionex IonPac AS-Fast-4µm column. Figure illustrates the separation of weakly retained anions such as acetate, butyrate, and formate in a simulated industrial chemical sample using a hydroxide gradient at a controlled temperature of 0 C. Vendor B Vendor A AS-Fast-4µm ( 0 mm) Eluent: 0 mm KOH Eluent Source: Dionex EGC 00 KOH Flow Rate: 0. ml/min Inj. Volume: 00 µl Temperature: 0 C Dionex AERS 00 ( mm) Sample 000 ppm sodium bicarbonate Peaks:. Chloride. Bicarbonate. Sulfate -0 0 0 0 0 Figure. Purity analysis of sodium bicarbonate using a large loop injection on the Dionex IonPac AS-Fast-4µm ( 0 mm) column. AS-Fast-4µm ( 0 mm) Eluent: 0 mm KOH Eluent Source: Dionex EGC 00 KOH Flow Rate: 0. ml/min Inj. Volume: 00 µl Temperature: 0 C Dionex AERS 00 ( mm) Sample Peaks: ppm sodium carbonate. Chloride. Carbonate. Sulfate - 0 0 0 0 Figure. Purity analysis of sodium carbonate using a large loop injection on the Dionex IonPac AS-Fast-4µm ( 0 mm) column. 0 4 4-0 0 0 0 0 4 Dionex IonPac AS-Fast-4µm (4 0 mm) Eluent: mm KOH from 0 to minutes, -0 mm KOH from to min, 0- mm KOH from to min, mm KOH from to min Eluent Source: Dionex EGC 00 KOH Flow Rate:. ml/min Inj. Volume: 0 µl Temperature: 0 C Detection: Dionex AERS 00 (4 mm) Suppressor AutoSuppression, recycle mode Peaks: mg/l (ppm). Fluoride.0 4. Sulfate.0. Acetate.0. Oxalate.0. Iodate.0. Benzoate 0.0 4. Butyrate.0. Azide 0.0. Formate.0. Bromide 0.0. Valariate.0. Nitrate 0.0. Methylsulfonate.0 0. Chlorate 0.0. Chlorite.0. Phosphate 0.0. Bromate.0. Arsenate 0.0 0. Chloride.. Citrate 0.0. Nitrite.0 4. Chromate 0.0. Carbonate.0. Thiosulfate 0.0. Monofluorophosphate 0.0 Figure. Separation of various environmental anions using a hydroxide gradient and the Dionex IonPac AS-Fast-4µm (4 0 mm) column.

System Recommendations The Dionex IonPac AS-Fast-4µm column is recommended for use with Thermo Scientific Dionex ICS-4000, Thermo Scientific Dionex ICS-000 +, Thermo Scientific Dionex ICS-000, or Thermo Scientific Dionex Integrion HPIC systems equipped with an EG. The EG is used to automatically produce potassium hydroxide gradients from deionized water. For all systems, the use of Thermo Scientific Dionex IC PEEK Viper fittings is recommended to achieve consistent low dead volume connections and ensure optimum chromatographic performance. Suppressor Recommendations For optimum ease of use and economy, use the Dionex IonPac AS-Fast-4µm column with the Thermo Scientific Dionex ADRS 00 Anion Dynamically Regenerated Suppressor or Thermo Scientific Dionex ACES 00 Anion Capillary Electrolytic Suppressor. Operate the Dionex IonPac AS-Fast-4µm column at 0 C to ensure reproducible retention times. Anion Trap Columns When using the EG for eluent delivery, a Thermo Scientific Dionex CR-ATC 00 or CR-ATC 00 Continuously Regenerated Anion Trap Column column should be installed between the EG cartridge and the EG degas module. Alternately, 4 mm and mm systems can use the Thermo Scientific Dionex IonPac ATC-HC 00 Anion Trap Column, installed between the pump outlet and inlet of the Dionex EluGen cartridge in the Dionex EG module. When performing sodium hydroxide gradient applications on the Dionex IonPac AS- Fast-4µm column using manually prepared eluents, the ATC 00 should be installed between the gradient pump and the injection valve to remove anionic contaminants from the eluent. Concentrator Columns For concentrator work, when a single piston pump such as the AXP pump (pulse damper required) is used for sample delivery, use the Dionex IonPac AG-Fast-4µm Guard column, Thermo Scientific Dionex IonPac UTAC- Series Ultra Trace Anion Concentrator Columns (UTAC-LP, UTAC-ULP, or UTAC-XLP), or the Thermo Scientific Dionex IonPac TAC- Trace Anion Concentrator Columns. Use the IonPac UTAC LP, UTAC-LP or the TAC-LP when the sample is delivered with a syringe or with an autosampler. For concentrator work with a 0.4 mm capillary column, use the Dionex IonPac AG-Fast-4µm Capillary Guard column or the Thermo Scientific Dionex IonSwift MAC-00 Concentrator column. Specifications Dimensions Dionex IonPac AS-Fast-4µm Capillary Column 0.4 0 mm Dionex IonPac AG-Fast-4µm Capillary Guard Column 0.4 mm Dionex IonPac AS-Fast-4µm Analytical Columns 0 and 4 0 mm Dionex IonPac AG-Fast-4µm Guard Columns 0 and 4 0 mm Maximum Operating Pressure 000 psi (Standard, Microbore, and Capillary) Mobile Phase Compatibility ph 0 4; 0 00% acetonitrile, methanol, and isopropanol Substrate Characteristics Bead Diameter: 4 µm (Analytical/Capillary column)* µm (Guard/Capillary Guard column)** *Analytical/Capillary Column resin composition: supermacroporous ethylvinylbenzene polymer cross-linked with divinylbenzene **Guard/Capillary Guard Column resin composition: microporous ethylvinylbenzene polymer cross-linked with divinylbenzene Crosslinking (%DVB): % Ion-Exchange Group Surface-functionalized with alkanol quaternary ammonium ion Functional Group Characteristic Medium high hydrophobicity Capacity 0 µeq/column (4 0 mm). µeq/column ( 0 mm). µeq/column (0.4 0 mm) 0 µeq/column (4 0 mm) µeq/column ( 0 mm) 0. µeq/column (0.4 mm)

Ordering Information In the U.S., call -00-4-0 or contact the Thermo Fisher Scientific Regional office nearest you. Outside the U.S., order through your local Thermo Fisher Scientific office or distributor. Refer to the following part numbers. Capillary, Analytical, and Guard Columns Part Number Dionex IonPac AS-Fast-4µm Capillary Column 0.4 0 mm 0 Dionex IonPac AG-Fast-4µm Capillary Guard Column 0.4 mm 0 Dionex IonPac AS-Fast-4µm Analytical Column 4 0 mm 04 Dionex IonPac AG-Fast-4µm Guard Column 4 0 mm 04 Dionex IonPac AS-Fast-4µm Analytical Column 0 mm 04 Dionex IonPac AG-Fast-4µm Guard Column 0 mm 00 Dionex IC PEEK Viper Fittings Kits Dionex IC PEEK Viper Fittings Kit for Dionex Integrion HPIC systems with conductivity detectors 0 Dionex IC PEEK Viper Fittings Kit for Dionex ICS-000 + /ICS-000 systems with conductivity detectors 00 Dionex IC PEEK Viper Fittings Kit for Dionex ICS-000 + /ICS-000 capillary systems with conductivity detectors 00 Dionex IC PEEK Viper Fittings Kit for Dionex ICS-4000 capillary systems with conductivity detectors 0 Anion Trap Columns Dionex CR-ATC 00 Continuously Regenerated Anion Trap Column, for use with Dionex Integrion HPIC systems equipped with an eluent generator 0 Dionex CR-ATC 00 Continuously Regenerated Anion Trap Column, for use with Dionex ICS-000 + systems equipped with an eluent generator 00 Dionex CR-ATC Continuously Regenerated Anion Trap Column (Capillary), for use with capillary HPIC systems 00 Dionex ATC-00 Anion Trap Column ( 4 mm), for use with manually prepared eluents and 4 mm columns) 0 Dionex ATC-00 Anion Trap Column (4 mm), for use with manually prepared eluents and mm columns) 00 Dionex ATC-HC 00 Anion Trap Column ( mm) 0 Trace Anion Concentrator Columns Dionex IonSwift MAC-00 Monolith Anion Concentrator Column (0. 0 mm) 040 Dionex IonPac TAC-LP Trace Anion Concentrator (4 mm) 040 Dionex IonPac UTAC-LP Ultra Trace Anion Concentrator Low Pressure (4 mm) 00 Dionex IonPac UTAC-LP Ultra Trace Anion Concentrator Low Pressure (4 mm) 0 Dionex IonPac UTAC-ULP Ultra Trace Anion Concentrator Ultra Low Pressure ( mm) 0 Dionex IonPac UTAC-XLP Ultra Trace Anion Concentrator Extremely Low Pressure ( mm) 0 Product Specifications Find out more at thermofisher.com/ic 0 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. PS04-EN 0S