Initial Results on the Feasibility of Hybrid X-Ray Microscopy

Similar documents
A new method of detecting interferogram in differential phase-contrast imaging system based on special structured x-ray scintillator screen

PROBLEM OF X-RAY SYNCHROTRON BEAM COLLIMATION BY ZONE PLATE

Light Source I. Takashi TANAKA (RIKEN SPring-8 Center) Cheiron 2012: Light Source I

FACTORS AFFECTING IN-LINE PHASE CONTRAST IMAGING WITH A LABORATORY MICROFOCUS X-RAY SOURCE

Liquid refractometer based on immersion diffractometry

Michelson Interferometer

Observation of ferroelectric domains in bismuth ferrite using coherent diffraction techniques

Refinement of X-ray Fluorescence Holography for Determination of Local Atomic Environment

Chapter 10. Interference of Light

RADIATION SOURCES AT SIBERIA-2 STORAGE RING

Laser-Modulator System

AP5301/ Name the major parts of an optical microscope and state their functions.

Characterizing the Nova600 microfocus X-ray sourc for use with the Talbot interferometer for phase-contrast imaging

MIDTERM 3 REVIEW SESSION. Dr. Flera Rizatdinova

On the possibilities of x-ray phase contrast microimaging by coherent high-energy synchrotron radiation

The Chemical Control of Superconductivity in Bi 2 Sr 2 (Ca 1 x Y x )Cu 2 O 8+±

A 0.2 m s -1. B 10 m s -1. C 20 m s -1. D 40 m s -1

Brightness and Coherence of Synchrotron Radiation and Free Electron Lasers. Zhirong Huang SLAC, Stanford University May 13, 2013

Spatial coherence measurement of X-ray undulator radiation

(i.e. what you should be able to answer at end of lecture)


Problem Solving. radians. 180 radians Stars & Elementary Astrophysics: Introduction Press F1 for Help 41. f s. picture. equation.

Atomic and Nuclear Physics

LC circuit: Energy stored. This lecture reviews some but not all of the material that will be on the final exam that covers in Chapters

X-ray phase imaging with a grating interferometer

Electricity & Optics

High-Resolution. Transmission. Electron Microscopy

The Use of Synchrotron Radiation in Modern Research

Lab 3-4 : Confocal Microscope Imaging of Single-Emitter Fluorescence and Hanbury-Brown and Twiss Set Up, Photon Antibunching

Diffraction of Electrons

print first name print last name print student id grade

THE INDIAN COMMUNITY SCHOOL, KUWAIT SECOND SEMESTER EXAMINATION PHYSICS (Theory)

ON MICROMETRIC PATTERN FORMATION IN MAGNETIC FLUIDS THIN FILMS

Ultra-narrow-band tunable laserline notch filter

SSRL XAS Beam Lines Soft X-ray

UNIT-5 EM WAVES UNIT-6 RAY OPTICS

Quantum Interference and Duality

Single Emitter Detection with Fluorescence and Extinction Spectroscopy

Gratings in Electrooptic Polymer Devices

1. Waves and Particles 2. Interference of Waves 3. Wave Nature of Light

Measurements in Optics for Civil Engineers

TA/TI survey. Phy Phy

NEW CORRECTION PROCEDURE FOR X-RAY SPECTROSCOPIC FLUORESCENCE DATA: SIMULATIONS AND EXPERIMENT

Basics of Synchrotron Radiation Beamlines and Detectors. Basics of synchrotron radiation X-ray optics as they apply to EXAFS experiments Detectors

Beyond the Geometric toward the Wave Optical Approach in the Design of Curved Crystal and Multilayer Optics for EDXAS

Knot undulator to generate linearly polarized photons with low on-axis power density. Abstract

Lab 1 Entanglement and Bell s Inequalities

Lecture Outline Chapter 30. Physics, 4 th Edition James S. Walker. Copyright 2010 Pearson Education, Inc.

C. Incorrect! The velocity of electromagnetic waves in a vacuum is the same, 3.14 x 10 8 m/s.

Synchrotron Radiation Representation in Phase Space

arxiv: v3 [nucl-ex] 18 May 2018

Experiment #4 Nature of Light: Telescope and Microscope and Spectroscope

GEANT4 simulation about the laboratory astrophysics in Taiwan

WAVE PARTICLE DUALITY

PHYSICS 253 SAMPLE FINAL EXAM. Student Number. The last two pages of the exam have some equations and some physical constants.

SUPPLEMENTARY INFORMATION

Interaction X-rays - Matter

Homework 1. Nano Optics, Fall Semester 2018 Photonics Laboratory, ETH Zürich

Imaging & Microscopy

Two-dimensional mapping of the asymmetric lateral coherence of thermal light

Chapter 1 X-ray Absorption Fine Structure (EXAFS)

Praktikum zur. Materialanalytik

Einstein Classes, Unit No. 102, 103, Vardhman Ring Road Plaza, Vikas Puri Extn., Outer Ring Road New Delhi , Ph. : ,

Experiment Report Form

Chapter 5. Effects of Photonic Crystal Band Gap on Rotation and Deformation of Hollow Te Rods in Triangular Lattice

Stigmatic X-ray imaging using a single spherical Laue crystal

Magnetic measurements (Pt. IV) advanced probes

Structure analysis: Electron diffraction LEED TEM RHEED

Calibration of the IXPE Instrument

Characteristics and Properties of Synchrotron Radiation

X-Ray Diffraction as a key to the Structure of Materials Interpretation of scattering patterns in real and reciprocal space

Optical Systems Program of Studies Version 1.0 April 2012

Interference by Wavefront Division

Electron Microprobe Analysis and Scanning Electron Microscopy

Optics.

Polarization of Light and Birefringence of Materials

MEASUREMENT OF TEMPORAL RESOLUTION AND DETECTION EFFICIENCY OF X-RAY STREAK CAMERA BY SINGLE PHOTON IMAGES

Mock Examination Physics 2 nd Book Time 3 Hrs MM : 70

Optical Instruments. Chapter 25. Simple Magnifier. Clicker 1. The Size of a Magnified Image. Angular Magnification 4/12/2011

P5 Revision Questions

Roger Johnson Structure and Dynamics: X-ray Diffraction Lecture 6

Experiments on deflection of charged particles using silicon crystals at. Proton Synchrotron (KEK(

2D XRD Imaging by Projection-Type X-Ray Microscope

Development of 2-Dimentional Imaging XAFS System at BL-4

Unit 4 Parent Guide: Waves. What is a wave?

Transmission Electron Microscopy

Unbalanced lensless ghost imaging with thermal light

is the minimum stopping potential for which the current between the plates reduces to zero.

Two Dimensional SR-Interferometer at PETRA III. Artem Novokshonov DESY, Tomsk Polytechnic University

MODERN OPTICS. P47 Optics: Unit 9

Alignment of Liquid Crystals by Ion Etched Grooved Glass Surfaces. Yea-Feng Lin, Ming-Chao Tsou, and Ru-Pin Pan

Electron Diffraction

Chapter 10: Wave Properties of Particles

Resolution: maximum limit of diffraction (asymmetric)

Practical 1P4 Energy Levels and Band Gaps

Which of the following can be used to calculate the resistive force acting on the brick? D (Total for Question = 1 mark)

Interference- Michelson Interferometer. Interference lecture by Dr. T.Vishwam

Optics Optical Testing and Testing Instrumentation Lab


4. Other diffraction techniques

Transcription:

CHINESE JOURNAL OF PHYSICS VOL. 43, NO. 5 OCTOBER 2005 Initial Results on the Feasibility of Hybrid X-Ray Microscopy P. K. Tseng, 1 W. F. Pong, 1 C. L. Chang, 1 C. P. Hsu, 1 F. Y. Lin, 2 C. S. Hwang, 2 and H. S. Sheu 2 1 Physics Department, Tamkang University, Tamsui, Taiwan, ROC 2 National Synchrotron Research Center, Hsinchu, Taiwan, ROC (Received December 16, 2004) This study presents a hybrid X-ray microscope that magnifies an X-ray image using a group of asymmetrically cut crystals and an X-ray phase contrast imaging system. An asymmetrically cut crystal is inserted between the sample and a CCD-fluorescence assembly. The experimental results demonstrate that the phase contrast image is preserved. This system can be improved using a better diffraction crystal. PACS numbers: 07.05.Pj, 87.57.-s I. INTRODUCTION When a coherent X-ray passes through a sample, a phase contrast image that consists of fringes at every boundary of the various media is produced. Such a system can be used to observe a thin sample which would otherwise be difficult to observe, because the absorption differences among the various media are small. The image produced by such a system is called an X-ray phase contrast image (XPCI). XPCI experiments have been performed using sophisticated X-ray-optics to ensure coherency [1, 2]. However, if the X-ray source is very small, the X-rays become coherent [3]. The third-generation synchrotron radiation facility is an ideal X-ray source for an XPCI. It can be operated using a simple set up, as shown in Fig. 1 [4]. A synchrotron radiation beam is directly passed through the sample, and a single crystal fluorescence-screen at some distance from the sample transforms the X-rays into a visible light image. The visible pattern on the screen is observed through a microscope or a microscope objective connected to a CCD camera. The basic limitation of this method FIG. 1: Schematic diagram of a simple set up for an X-ray phase contrast image microscope. http://psroc.phys.ntu.edu.tw/cjp 979 c 2005 THE PHYSICAL SOCIETY OF THE REPUBLIC OF CHINA

980 INITIAL RESULTS ON THE FEASIBILITY OF... VOL. 43 FIG. 2: Schematic diagram of the Bragg diffraction by a mis-cut crystal. is the same as that of an optical microscope. II. PROPOSED HYBRID X-RAY MICROSCOPE An asymmetrically cut crystal is used to widen the X-ray beam that has passed through the sample, thus overcoming the small photon beam size limitation. Accordingly, the original X-ray phase-contrast pattern is enlarged before it is converted to visible light by the fluorescent screen. Let the lattice constant of the crystal be d and the miss-cut angle with respect to the diffraction lattice plane be θ M. The Bragg angle is θ B. The width of the incoming X-ray beam is W in, and that of the outgoing beam is W out, as shown in Fig. 2. The relationship between W in and W out is W out = W in sin(θ B + θ M ) sin(θ B θ M ). (1) Therefore, the original X-ray phase contrast pattern is enlarged (in one direction) before it is converted into visible light by the fluorescenct screen. Another asymmetrically cut diffraction crystal pair is oriented with respect to the original amplification direction and the incoming X-ray beam as shown in Fig. 3, to amplify the X-ray phase contrasted pattern uniformly (in every direction). In Fig. 3, one asymmetrically cut crystal and one normal crystal (or another asymmetrically cut crystal) are used as a double crystal monochromater (DCM) arrangement to keep the X-ray beam parallel to the incoming synchrotron radiation. Fig. 3 presents a horizontal view and a schematic vertical diagram of this idea. The microscope is called a hybrid X-ray microscope (HXM) because it depends on two magnification principles. These amplification mechanisms and their limitations must be considered more carefully to demonstrate experimentally the feasibility of the hybrid X-ray microscope. Every set up should meet the following criteria.

VOL. 43 P. K. TSENG, W. F. PONG, et al. 981 FIG. 3: Hybrid X-ray microscope using X-ray phase contrast imaging and asymmetric refraction. 1. X-ray intensity. If the magnified factor is M, the flux of the incoming X-ray is L 0, equivalent to illuminating the magnified pattern with a flux of ( L 0 ) 2. M Therefore the luminosity of the synchrotron radiation source must exceed L Min, where L Min is the minimum luminosity required when the X-ray illuminates the object. The transmitted pattern can then be directly observed (by the same method). 2. Notably, the visibility of the phase contrast image is limited by the noise of the detector-ccd. 3. The incoming x-ray must be coherent to produce a phase-contrasted image. The synchrotron radiation becomes coherent if the electron beam diameter is small [5]. The light from the third-generation synchrotron radiation facility can become coherent if the source is of limited beam size, as indicated by the emittance of the storage ring. 4. As the emittance increases, the synchrotron becomes partially coherent. The phase contrast curve in the X-ray image gradually disappears. It is important to find the extent of the partial coherency of the small third generation ring, like NSRRC. 5. If the diffraction crystal is perfect, then the phase of the diffracted beam would be preserved. The hybrid microscope would work well in such a case. In the real world, it is very difficult to produce such a perfect crystal especially after asymmetric cutting. In general, the available single crystals have a more or less mosaic structure.

982 INITIAL RESULTS ON THE FEASIBILITY OF... VOL. 43 FIG. 4: Schematic diagram of the demagnification experimental arrangement. III. EXPERIMENTAL RESULTS We carefully choose a reasonably good quality area in the asymmetric cut crystal, so as to diffract the X-ray wave without changing its phase. An experiment is performed to determine whether the image is diffracted with its phase contrast fringes. The photon flux in the proposed SR source, which is a 1.5 GeV storage ring (at a maximum electron current of 200 ma), is too weak to generate a magnified image by asymmetric diffraction, so a reverse experiment is designed to test the idea. An asymmetric crystal (Si(111)) (2000 ohm-cm) is oriented in a reverse way as shown in Fig. 4. A white synchrotron radiation beam is incident, and the cross section of the diffracted beam in only one direction becomes small. In the proposed experimental setup, the miss-cut angle is about 10 degree. The orientation of the crystal is fixed at approximately 13.3 degrees so the angle of the diffracted beam is 26.6 degrees (This angle is used because the intensity of the diffracted monochromatic X-ray from the source is maximal and tan(2θ) = 0.5). The diffracted monochromatic beam is extremely weak. Therefore, a cooled CCD detector with an exposure time of at least 10 minutes must be used. The best image is obtained by extending the exposure time from 10 minutes to as much as one hour. The characteristics of the phase-contrasted image are exploited to determine whether the phase of the diffracted beam can be retained. The phase-contrast curves become thicker as the distance from the sample to the fluorescent screen increases. IV. DISCUSSION Fig. 5(a) depicts the sample source used in this experiment, which is an insectentrapped in Baltic amber. After the X-ray microscope is de-magnified (or squeezed) by asymmetric diffraction, it becomes thin in the X-direction, as displayed in Fig. 5(b). Phasecontrast curves appear in every boundary as ordinary phase contrast images. The factor

VOL. 43 P. K. TSENG, W. F. PONG, et al. 983 FIG. 5: (a) Image of an insect in amber obtained by ordinary X-ray phase contrast microscopy. This mosquito-like insect, entrapped in Baltic amber, is used as the sample for asymmetrically cut diffraction imaging. (b) Image of the insect in Fig. 5(a) after the SR light passes through the insect and is diffracted by an asymmetrically cut crystal. (c) Simulated image of a mosquito-like insect after asymmetric refraction. Fig. 5(b) is demagnified only in the X-direction. of demagnification is about 0.3, according to Eq. (1). The simulated de-magnification of Fig. 5(a) yields the results in Fig. 5(c). The distance from the sample to the fluorescent crystal is changed from the minimum (30 cm) to 65 cm in steps of 5 cm to demonstrate that the dark curves came from the phase contrast image. Fig. 6 displays the results. The curves of the boundaries obtained by phase contrast imaging became broadened as the distance from the sample to the fluorescent crystal increased. This agrees qualitatively well with the figures. Only a small part of each image (the left half of the head of the insect) is enlarged. Fig. 7 depicts this image for comparison. This picture clarifies the change in the thickness of the boundary curves. However the comparison of the experimental results (Figs. 5 and 6) and the simulated asymmetric diffraction reveals that some fine phase contrast fringes are absent from the experimental results. The authors think that the diffraction surface was not sufficiently flat and the crystal used herein was not sufficiently perfect to yield the expected detailed structure. Perhaps the intensity of the diffracted beam was too weak and buried by noise. The surface of the diffraction crystal has been etched extensively so the original surface could not be observed before etching. We also believe that single crystal grains were exposed on the diffraction surface. We are trying now to improve the crystal and develop new surface processes.

984 INITIAL RESULTS ON THE FEASIBILITY OF... VOL. 43 FIG. 6: Series of X-ray phase contrast images obtained by changing the distance between the sample and the fluorescent crystal from 30 (left most picture) to 65 cm (in steps 5 cm). The phase contrast-boundary-curves of each figure become thicker in the right of pictures. FIG. 7: The left half of the head of the insect shown in Fig. 6 is enlarged and the images are shown together for comparison. Acknowledgments The author (PKT) would like to thank the National Science Council for financially supporting this research for many years. He would also like to thank the support team, including those who worked at the beam line and the operators of the storage ring. References [1] U. Bonse and M. Hart, Appl. Phys. Lett. 6, 155 (1965). [2] A. Momose, T. Takeda, and I. Itai, Rev. Sci. Instrum. 66, 1434 (1995). [3] T. J. Davis, D. Gao, T. E. Gureyev, A. W. Stevenson, and S. W. Wilkins, Nature 373, 595 (1995). [4] A. Snigirev, I. Snigireva, V. Kohn, S. Kuznetsov, and I. Schelokov, Rev. Sci. Instrum. 66, 5486

VOL. 43 P. K. TSENG, W. F. PONG, et al. 985 (1995). [5] When the dimensions of the light source are small, the emitted light will become (transversal) coherent. This fact was known when Young demonstrated his interference of light experiment. This is also true for synchrotron radiation.