A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Similar documents
OA03 UNCERTAINTY OF MEASUREMENT IN CHEMICAL TESTING IN ACCORDANCE WITH THE STANDARD SIST EN ISO/IEC Table of contents

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

SWGDRUG GLOSSARY. Independent science-based organization that has the authority to grant

SAMM POLICY 5 (SP5) POLICY ON MEASUREMENT UNCERTAINTY REQUIREMENTS FOR SAMM TESTING LABORATORIES Issue 2, 28 February 2007 (Amd. 1, 11 August 2014)

Calibration Traceability Guidelines

Technical Procedure for General Laboratory Equipment

APPENDIX G ESTIMATION OF UNCERTAINTY OF MEASUREMENT

APPENDIX G EVALUATION OF MEASUREMENT UNCERTAINTY

Vocabulary of Metrology

Certificate of Accreditation to ISO/IEC 17025:2005

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

Traceability, validation and measurement uncertainty 3 pillars for quality of measurement results. David MILDE

What is measurement uncertainty?

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

A basic introduction to reference materials. POPs Strategy

Document No: TR 12 Issue No: 1

Comparability and Traceability Point of view from a metrological institute

CALIBRATION REPORT FOR THERMOHYDROMETER

MEASUREMENT UNCERTAINTY PREPARED FOR ENAO ASSESSOR CALIBRATION COURSE OCTOBER/NOVEMBER Prepared by MJ Mc Nerney for ENAO Assessor Calibration

The Fundamentals of Moisture Calibration

Measurement Uncertainty, March 2009, F. Cordeiro 1

ISO/IEC requirements. General matters of quality system, requirements for documentation and the list of the documents required for a

Precision estimated by series of analysis ISO and Approach Duplicate Approach

Technical Procedure for Measurement Assurance

R SANAS Page 1 of 7

NATIONAL ASSOCIATION OF TESTING AUTHORITIES (NATA) REQUIREMENTS FOR ACCREDITATION OF ICP-MS TECHNIQUES

New Thermometer Guidance Document from AASHTO Accreditation Program. Maria Knake Asphalt Binder ETG Meeting May 10, 2018

Uncertainty of Measurement A Concept

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

ANALYSIS CERTIFICATE OF

HOW TO ASSESS THE MEASUREMENT UNCERTAINTY

CALIBRATION CERTIFICATE # 503

MEASUREMENT TRACEABILITY AND ERRORS RELATED TO FORCE MEASUREMENT

Certificate of Analysis

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

Certificate of Analysis

Certificate of Analysis

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Analytical Measurement Uncertainty

SAMPLE. Accuracy Calibration Certificate. Instrument Type: EURAMET cg-18 v. 4.0 Sample ACC ISO17025 WI v.1.0

ISO/TS TECHNICAL SPECIFICATION. Water quality Guidance on analytical quality control for chemical and physicochemical water analysis

Mason Technology Ltd

1.11 Measurement Uncertainty

The Uncertainty of Reference Standards

Test Method Development and Validation as Pertaining to Microtrac Particle Size Measuring Instruments

Uncertainty of Measurement

Calibration and verification: Two procedures having comparable objectives and results

Certified Reference Material - Certificate of Analysis

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

Measurement Uncertainty Knowing the Unknown

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Draft EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement. Second Edition. Draft: June 1999

Part IVB Quality Assurance/Validation of Analytical Methods

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Measurement & Uncertainty - Basic concept and its application

Certificate of Accreditation

IS INTERNATIONAL STANDARD. Natural gas - Guidelines to traceability in analysis

Mason Technology Ltd

Certificate of analysis Certified reference material code: CRM-00-GTX1&4

Uncertainty Propagation for Force Calibration Systems

Essentials of expressing measurement uncertainty

An Introduction to the Differences Between the Two Most Recognized Force Standards

Practical Interpretation of Metrological Traceability on Gauge Block as to VIM 3

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

P103d Annex: Policy on Estimating Measurement Uncertainty for Construction Materials & Geotechnical Testing Labs Date of Issue 09/13/05

3.1.1 The method can detect, identify, and potentially measure the amount of (quantify) an analyte(s):

Chemistry 11. Unit 2: Introduction to Chemistry. Measurement tools Graphing Scientific notation Unit conversions Density Significant figures

IMPORTANT FEATURES OF ESTIMATION & INTERPRETATION OF MEASUREMENT UNCERTAINTY. B. Bhattacharya, K. Ramani, P.H. Bhave

INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT. VERIFYING THERMOMETERS ITM No T

ROLE AND SIGNIFICANCE OF UNCERTAINTY IN HV MEASUREMENT OF PORCELAIN INSULATORS A CASE STUDY

Implementation of ISO/IEC in a low level liquid s cintillation tritium laboratory

Chemistry 11. Measuring and Recording Scientific Data. Measurement tools Graphing Scientific notation Unit conversions Density Significant figures

Process Accuracy Ratio vs Process Uncertainty Ratio (Risk Mitigation: Calibration and the Customer s Process)

Measurement Uncertainty Principles and Implementation in QC

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Measurement & Uncertainty - Concept and its application

Scientific Literacy & the Scientific Method

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers

Reference Materials and Proficiency Testing. CropLife International Meeting October 2015 Gina M. Clapper

TITLE OF THE CD (English): New OIML Document

CERTIFICATE OF CALIBRATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ALLOMETRICS, INC Bayport Blvd. Seabrook, TX Terry Baldwin Phone: CALIBRATION

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

ESTA governance Quality Assurance System for Seed Treatment and Treated Seed

Calibration of Temperature Block Calibrators

Certificate of Accreditation

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

Examples of Method Validation Studies Conducted in Different Economies

Quantities and Measurement Units

A case study on how to maintain confidence of thermal properties test: Thermal conductivity of building insulation materials

Procedure for Measurement Assurance

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

LAB 15. Traceability: Volumetric Apparatus. Edition 3 February 2019 Draft for Consultation. LAB 15 / Issue: 3 Page 1 of 12

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Introduction to Uncertainty and Treatment of Data

Guidelines on the Calibration of Automatic Instruments for Weighing Road Vehicles in Motion and Measuring Axle Loads AWICal WIM Guide May 2018

Transcription:

Revision 1.2 October 15, 2012 PAGE II

TABLE OF CONTENTS TABLE OF CONTENTS... 1 GUIDANCE...2 1.0 Introduction... 2 2.0 Uncertainty and Traceability... 2 3.0 Need for Traceability... 3 3.1 What does traceability look like?... 5 4.0 Implementation of A61-01 CALA Traceability Policy...6 4.1 Selecting a Calibration Laboratory... 6 4.2 The Significance Test... 8 4.3 Balances...9 4.4 Thermometers and Thermocouples... 10 4.5 Volumetric Measurements...11 5.0 Recommended Calibration and Verification Intervals...12 6.0 Terms and Definitions...13 Rev 1.2 PAGE II

GUIDANCE 1.0 Introduction This document is provided for guidance on the implementation of A61-01 CALA Traceability Policy. It is not a requirements document. 2.0 Uncertainty and Traceability Traceability is related to the concept of Uncertainty of measurement. This diagram provides demonstration of their relation. This diagram shows a series of results as the small white dots. They are not the same as the actual value one would get if we lived in a perfect world. This perfect or true value is represented by the centre of the target. The black dot represents the average of all the results generated. It is not a result itself. It is also called the mean or average of the set of generated results. If a laboratory has produced a set of results from one large sample, then they may report this average as representative of the whole set. Rev 1.2 Page 2 of 13

Whether a laboratory reports a single result or the mean (average) of a set of results, the considerations for reporting uncertainty are the same in all cases: o Accuracy: Accuracy refers to closeness of agreement between a measured quantity value and a true quantity value of a measurand (VIM, 2.13). If one considers only the numbers, then one might examine the quantitative equivalent considerations. These are trueness and precision. o Trueness: Trueness is the closeness of agreement between the mean value obtained from a large series of test results and an accepted reference value; in the diagram above, trueness is represented by the black dot. o Precision: Precision is the closeness of agreement between measured quantity values obtained by replicate measurements under the same or similar specified conditions (VIM, 2.15) i.e., how dispersed the group of white dots are from each other and from the black dot (dispersion). Precision measures random contributions to uncertainty. Note: It is important to understand that trueness and precision are independent of each other. One has nothing to do with the other. Bias: Bias is the difference between the expectation of a measurement result and an accepted reference value. For example, if we determined from a number of measurements that the average trueness of several results had a certain relationship to a reference value (for example were 10% higher), then we might expect further measurements made in the same way to be similarly biased. While trueness is a property of a result or set of results, bias is a property of a measurement method. Bias is normally determined by measuring trueness. In the diagram, if we planned to do more measurements, we would expect most results to be high and to the right (i.e., we expect the result not to be true due to bias). 3.0 Need for Traceability Analytical data are often used for comparative purposes such as conformity to a standard or limit. To compare results from different laboratories or from the same laboratory at different times with known confidence, all labs must be using the same reference points. These common reference points are achieved by establishing traceability. That is, calibrations that can be related to primary national or international standards (ideally S.I. units of measurement) through an unbroken chain of comparisons all having stated uncertainties. Traceability and its Relation to Uncertainty Analytical data are the result of measurements. No measurement; no matter how carefully made, can be completely free of uncertainty. One can think of a test result as a best estimate of a property and the uncertainty as how sure we can be of that estimate. Uncertainty does not imply doubt about the validity of the measurement; it quantifies Rev 1.2 Page 3 of 13

confidence and therefore adds value to the test result. Part of the quantifiable true uncertainty is the uncertainty inherent in the measurement data, so the ability to evaluate and control these uncertainties is crucially important for appropriate data interpretation. Knowledge of measurement uncertainty allows the data user to draw valid conclusions with a known degree of confidence. Agreement between labs (or within labs over time) is limited by the uncertainties in the traceability chain. So traceability places measurements on a consistent and comparable measurement scale. Uncertainty characterizes the strength of the links in the chain and characterizes the agreement to be expected between labs making similar measurements. In the traceability chain, uncertainty of test results is the uncertainty in the reference together with the uncertainty of making the measurement relative to the reference. While in a real chain the strength is determined by the weakest link, the strength of the traceability chain is determined by the collective strengths of all the links. Traceability involves the competent propagation of uncertainties all along the chain of measurement to the test result produced by a testing laboratory. See the diagram on the following page. If uncertainties have not been propagated to the actual test result, the result is not traceable. If it is not traceable, it may appear precise but its trueness will always be suspect. At the same time, uncertainties established along the traceability chain are the basis for the establishment of traceability of the measurement. The only property of a reference item (e.g., instrument, mass, reference material, etc ) that counts, in considering traceability, is uncertainty. Test results, in order to be traceable, must be conducted using traceable reference items (e.g., instruments, masses, reference materials, etc ) only. Only those reference items that have been competently compared to others of known uncertainty (calibrated) can have their contribution to the overall uncertainty of the measurement objectively examined. When this condition is met, the measurement is considered traceable. If the reference items are not traceable, then the test result is not traceable. Rev 1.2 Page 4 of 13

3.1 What does traceability look like? NATIONAL METROLOGY INSTITUTE (e.g., NRC) Uncertainty Calibration Traceability CALIBRATION LABORATORY Calibration TESTING LABORATORY Calibration TESTING ACTIVITY Test Result to Client At the top of this diagram are all of the National Metrology Institutes (NMIs) that are responsible in each nation for quantifying specific parameters for use within their nation. Most NMI s have signed a multilateral recognition agreement based on the demonstration of competence (e.g., accreditation against ISO/IEC 17025). The first job of an NMI is to characterise a parameter, such as mass or temperature, to a specific level of uncertainty. They propagate this uncertainty in support of competent measurements. This affects legal measurements (butcher weigh scales and gas pumps) as well as other fields of science requiring competent measurement. Each NMI has the ability to conduct measurements with very small uncertainties. They are also able to calibrate the instruments from calibration laboratories seeking to establish traceability to the NMI. This is the start of the traceability chain for a testing laboratory. Rev 1.2 Page 5 of 13

Calibration, Traceability and Uncertainty are all required at any point in the traceability chain. None of these are deemed to be present unless ALL are present. 4.0 Implementation of A61-01 CALA Traceability Policy The CALA Traceability Policy (A61-01) means three things to CALA laboratories. First, it requires CALA laboratories to make use of measurement instruments, whose measurement traceability goes all the way back to an international standard (the SI) through a national measurement laboratory (e.g., NIST or NRC). Second, it means that CALA laboratories must know how to spot the signs that an instrument is traceable or not. Third, it means that CALA laboratories must understand the following simple relationship. All three of these components must exist at every level in the traceability chain in order for the final test result to be traceable. NO CALIBRATION = NO UNCERTAINTY = NO TRACEABILITY 4.1 Selecting a Calibration Laboratory 4.1.1 General Traceability means that all of the comparisons back along the Calibration chain to the national measurement laboratory were done by competent organizations (calibration laboratories). The competence of a calibration laboratory is most easily established by ensuring that the calibration laboratory is accredited to ISO/IEC 17025 for the task. For example, the following table represents a sample Scope of Accreditation for an accredited calibration laboratory: Measurement Parameter Range of Measurement Related Uncertainty Temperature 0.01-140.01 degrees C +/- 0.003 degrees C Volume 0.05-200.00 Litres +/- 0.005 ml An accredited calibration laboratory that is recognized by the CALA will normally be found on one of the following websites: SCC/CLAS http://palcan.scc.ca/specssearch/calibrationsearchform.do NVLAP - http://ts.nist.gov/standards/scopes/calibration-index.htm A2LA http://www.a2la.org/dirsearchnew/newsearch.cfm IAS http://www.iasonline.org/calibration_laboratories/cl.html L-A-B http://www.l-a-b.com/content/directory-accredited-labs ACLASS - http://www.aclasscorp.com/search-accredited-companies.aspx Rev 1.2 Page 6 of 13

This list is not exhaustive as CALA recognizes other laboratory accreditation bodies, but these are the ones closest to you, the CALA laboratory. 4.1.2 Calibration Certificates The following information should appear on the calibration certificate: An indication that the calibration laboratory is accredited to ISO/IEC 17025 by a recognized and named accreditation body. Remember, a service provider that is only registered (or certified) to ISO 9000 (or other management system standard) is not considered a technically competent calibration service provider. The serial number of the measuring equipment (or reference standards) being used to calibrate your equipment and a statement that the measuring equipment (or reference standard) is traceable to SI units through an NMI; and, The measurement range for which your equipment was calibrated and the specific uncertainty measurements for that range. Before purchasing a traceable thermometer or calibration certificate, you may be able to obtain from your potential supplier an example of their calibration certificates. This will allow you to review the calibration certificate and if you have any concerns about whether you are meeting the traceability and calibration requirements, you can get answers from your supplier before you purchase the item or service. If newly purchased equipment is not provided with an appropriate calibration certificate, then a calibration must be performed prior to use. Rev 1.2 Page 7 of 13

4.2 The Significance Test Clause 5.6.1 of ISO/IEC 17025 states: All equipment used for tests and/or calibrations, including equipment for subsidiary measurements (e.g., for environmental conditions) having a significant effect on the accuracy or validity of the result of the test, calibration or sampling shall be calibrated before being put into service. The laboratory shall have an established programme and procedure for the calibration of its equipment. Significance can be established using an approach that compares results from a simple statistical expression. Any number, x, that is one third or less than another, y, will tend to have no significant effect on the outcome of the square root of the sum of their squares: If, x = 1.0 and y = 3.0, then z = 3.16 if x is included in the expression and 3.0 if it is not. The difference is approximately 5%, a relatively insignificant difference when considering uncertainties. Any number, x, that is one tenth or less than another, y, will tend to have negligible effect on the outcome of the square root of the sum of their squares. If, x = 1.0 and y = 10.0, then z = 10.05 if x is included in the expression and 10.0 if it is not. The difference is approximately 0.5%, a negligible difference when considering uncertainties. The following three examples may be used by laboratories to determine which of their instruments meet the requirement for significant effect in their calibration program. o o o Example 1: If the uncertainty of an instrument is smaller than one-tenth of the largest contributor in the estimation of uncertainty for all tests affected by that instrument, then the laboratory has evidence to increase the period of calibration from one to two years or more, decreasing its frequency of calibration. Example 2: If the uncertainty of an instrument is between one tenth and one-third of the largest contributor in the estimation of uncertainty for all tests affected by that instrument, then the laboratory must maintain the period of calibration at one year. Example 3: If the uncertainty of an instrument is larger than one-third of the largest contributor in the estimation of uncertainty for all tests affected by that instrument, then the laboratory has evidence to decrease the period of calibration to less than one year; for example every six months. Rev 1.2 Page 8 of 13

The choices to a laboratory are the following, and they depend on the significance of the contribution of the instrument to the overall uncertainty of the desired test result as described by the examples above: o When instrument performance is very much better than the requirement, the laboratory can reduce the calibration frequency (extend the cycle time). See example 1 above.; o When performance is at the same level as requirement, the calibration frequency must be increased. See example 3 above; or, o When the laboratory does not know, they must calibrate that equipment once per year. See example 2 above. In general, the only instruments in a typical analytical laboratory that require traceability through external calibration are those related to mass and temperature and subsidiary measurements (e.g., barometric pressure). Volume (e.g., pipettes) must also be traceable but their traceability is usually linked to mass. Chemicals and standards purchased for the purpose of instrument calibration must also be traceable. but their traceability is ensured through proper analytical methods and proper procurement procedures. A good description of chemical traceability can be found in Traceability in Chemical Measurements (Eurachem 2003). 4.3 Balances 4.3.1 Calibration Daily use of a balance often leads to subtle changes in the balance. This may be seen when reviewing the verification data and may also be indicated when the as found checks on the balance by the calibration laboratory are found to be out-of-specification compared to acceptance criteria and the as-left conditions. If these are all found to be in-specification, then perhaps the balance is stable enough to reduce the calibration frequency IF the measurement uncertainty of the balance at the lab is significantly less than the criteria required for the weighing (for example, if MU= 0.001 mg but the balance is only used to measure down to 0.1 mg). In these cases, the frequency might be decreased but not eliminated. 4.3.2 Verification Verification allows the laboratory to demonstrate stability of operation of the balance over the whole range normally used. The criteria for these checks should be fit-for-purpose. The documented procedure for the verification process should include: the frequency of verifications; Rev 1.2 Page 9 of 13

the use, storage, and handling of appropriate weights; acceptance criteria; and, actions to be taken when the acceptance criteria are not met (e.g., recall of results, etc ). 4.4 Thermometers and Thermocouples 4.4.1 Calibration The frequency of recalibration of thermometers will be dependent on the use of the thermometer. For example, incubator thermometers with a narrow temperature requirements should be calibrated more frequently than those used for sample storage. A laboratory may demonstrate traceability of their temperature measurements through external calibration using an accredited calibration laboratory, or performing internal calibrations. A laboratory may opt to send working thermometers to an accredited laboratory for calibration at an established frequency (e.g., annually). This is by far the easiest way to ensure traceability of all critical thermometers but can be costly and requires the purchase of more thermometers than are used on a daily basis in order to ensure the availability of calibrated thermometers when others are being calibrated. A laboratory may decide to maintain a limited number of thermometers that are sent to an accredited calibration laboratory on a scheduled basis. The laboratory may then use these reference thermometers to calibrate working thermometers. Thermometer and thermocouple calibrations and verifications against the laboratory s reference thermometer are typically conducted at a single temperature (e.g., temperature of use or ice-point). 4.4.2 Verification Thermometers exposed to wide fluctuations in temperature may require frequent verification. Similarly, any temperature device being moved frequently may also require a higher frequency of verification (e.g., a working thermometer being moved from incubator to incubator). Rev 1.2 Page 10 of 13

4.5 Volumetric Measurements 4.5.1 Calibration Dispensing devices must be calibrated upon receipt and at an established frequency (e.g., annually) and verified. Dispensing devices may be sent to an accredited calibration laboratory for calibration. Alternately, the laboratory may perform an internal calibration following a generally accepted procedure. This procedure generally involves the repeated weighing of dispensed volumes of water, corrected for standard temperature and pressure. For adjustable dispensing devices, this procedure is performed at more than one volume. Acceptable performance (precision) is based on the precision required for the piece of dispensing equipment. Historical daily or asused verification data can be used to develop standard deviations for calibration provided all the requirements for calibration are met for each verification reading. This includes, but is not limited to, ensuring: o that the laboratory documents their verification procedure to include all of the same requirements for the performance of a calibration but with as few as one reading on each day of verification; o that a statistically valid number of readings are used in the determination of the standard deviations for the calibration; o that the person conducting the verification has been properly trained on the procedure used; o that they can competently propagate uncertainties from the reference standard (device) to the working instrument; and o that they use a calibrated balance that meets the requirements of ISO/IEC 17025 for this purpose. 4.5.2 Verification Daily or as-used verifications are performed by dispensing a measured volume to a tared balance and recording the weight (corrected for temperature and pressure). For adjustable dispensing devices, this procedure should be performed at more than one volume. While syringes generally come with certificates from an ISO 9000 registered company, this certificate in itself is not sufficient evidence of on-going integrity of the syringe. Syringes used for standard preparation and sample introduction for the GC etc... can be verified indirectly through the analysis of surrogates or internal standards. Rev 1.2 Page 11 of 13

5.0 Recommended Calibration and Verification Intervals The following table sets out recommended calibration and verification intervals. Equipment Calibration Interval Verification Frequency Balances Initial and every two (2) years Daily or as-used Masses Reference Working Initial and every five (5) years - Day that balance is calibrated by an external calibration provider. Day that balance is calibrated by an external calibration provider. Thermometers Reference Working Initial and every five (5) years Initial and every two (2) years Annually, using ice-point or boiling point (liquid-in-glass only). Annually, against the reference thermometer. Pipettors Initial and annually Daily or as-used Rev 1.2 Page 12 of 13

6.0 Terms and Definitions o Calibration: Calibration is a comparison of measurements between two standards or measurement devices. It involves the competent propagation of uncertainties from the instrument or standard whose measurement characteristics are known and traceable to the SI, to an instrument or standard whose measurement characteristics are to be quantified through this comparison. o Calibration Curve: This applies to analytical laboratory instrumentation. The term defines the relation between analyte concentration and instrument response. o Control Standard: A standard used as a basis for comparison with calibration standards, prepared independently from the calibration standards, and which undergoes sample processing identical to that carried out for the calibration standards. o Reference Material (RM): Material or substance one or more of whose property values are sufficiently homogeneous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to the materials. (ISO/IEC Guide 43-1) o SI (Système International d Unités): The name (International System of Units) adopted by the 11th General Conference on Weights and Measures (1960) for the recommended practical system of units of measurement. The base units are a choice of seven welldefined units: the metre, the kilogram, the second, the ampere, the Kelvin, the mole, and the candela. o Traceability: A property of the result of a measurement or the value of a standard whereby it can be related, with a stated uncertainty, to stated references, usually national or international standards, through an unbroken chain of comparisons. (ISO Guide 30). o Trueness: The closeness of agreement between the average value obtained from a large series of test results and an accepted reference value. (ISO 3534-1, 3.12). o Uncertainty of Measurement: Parameter that characterizes the dispersion of the quantity values that are being attributed to a measurand, based on the information used. (VIM) o Verification: Confirmation through examination of a given item and provision of objective evidence that it fulfils specified requirements. [modified from ISO 9000:2005, item 3.8.4] Rev 1.2 Page 13 of 13