Optics Optical Testing and Testing Instrumentation Lab

Similar documents
Part 1 - Basic Interferometers for Optical Testing

n The visual examination of the image of a point source is one of the most basic and important tests that can be performed.

Metrology and Sensing

Optics.

Experiment O-2. The Michelson Interferometer

Revisiting Fizeau s Observations: Spectral study of Na source using Newton s rings. Abstract

Coherence and width of spectral lines with Michelson interferometer

Metrology and Sensing

Metrology and Sensing

8. The Michelson Interferometer

NORTHERN ILLINOIS UNIVERSITY PHYSICS DEPARTMENT. Physics 211 E&M and Quantum Physics Spring Lab #9: Diffraction Spectroscopy

Exam 3--PHYS 202--S10

Metrology and Sensing

Measurments with Michelson interferometers

Prac%ce Quiz 8. These are Q s from old quizzes. I do not guarantee that the Q s on this year s quiz will be the same, or even similar.

Phase Contrast. Zernike Phase Contrast (PC) Nomarski Differential Interference Contrast (DIC) J Mertz Boston University

Electricity & Optics

THE DIFFRACTION GRATING SPECTROMETER

Michelson Interferometer

The Diffraction Grating

Waves Part III Electromagnetic waves

THE ZEEMAN EFFECT PHYSICS 359E

PHY410 Optics Exam #3

Low Coherence Vibration Insensitive Fizeau Interferometer

Any first year text, sections on atomic structure, spectral lines and spectrometers

Chapter 20: Convergent-beam diffraction Selected-area diffraction: Influence of thickness Selected-area vs. convergent-beam diffraction

Einstein Classes, Unit No. 102, 103, Vardhman Ring Road Plaza, Vikas Puri Extn., Outer Ring Road New Delhi , Ph. : ,

Edward S. Rogers Sr. Department of Electrical and Computer Engineering. ECE318S Fundamentals of Optics. Final Exam. April 16, 2007.

Experiment 6: Interferometers

LC circuit: Energy stored. This lecture reviews some but not all of the material that will be on the final exam that covers in Chapters

Measurements in Optics for Civil Engineers

Engineering Physics 1 Prof. G.D. Vermaa Department of Physics Indian Institute of Technology-Roorkee

PRINCIPLES OF PHYSICAL OPTICS

Michelson Interferometer. crucial role in Einstein s development of the Special Theory of Relativity.

THE MICHELSON INTERFEROMETER Intermediate ( first part) to Advanced (latter parts)

High-Resolution Imagers

Interferometers. PART 1: Michelson Interferometer The Michelson interferometer is one of the most useful of all optical instru

Laser Speckle and Applications in Optics

Edward S. Rogers Sr. Department of Electrical and Computer Engineering. ECE426F Optical Engineering. Final Exam. Dec. 17, 2003.

Figure measurement of a large optical flat with a Fizeau interferometer and stitching technique

Optics Interference from Films Newton s Rings Michelson Interferometer

Introduction to Interferometer and Coronagraph Imaging

PS210 - Optical Techniques. Section VI

The Michelson Interferometer

The science of light. P. Ewart

PHYSICS 253 SAMPLE FINAL EXAM. Student Number. The last two pages of the exam have some equations and some physical constants.

General Physics II Summer Session 2013 Review Ch - 16, 17, 18

Interference- Michelson Interferometer. Interference lecture by Dr. T.Vishwam

PHYSICS 370 OPTICS. Instructor: Dr. Fred Otto Phone:

Chapter 10. Interference of Light

OPSE FINAL EXAM Fall 2015 YOU MUST SHOW YOUR WORK. ANSWERS THAT ARE NOT JUSTIFIED WILL BE GIVEN ZERO CREDIT.

Optical Instruments. Chapter 25. Simple Magnifier. Clicker 1. The Size of a Magnified Image. Angular Magnification 4/12/2011

A Question. Simple Magnifier. Magnification by a Lens 11/29/2011. The last lecture

2.71. Final examination. 3 hours (9am 12 noon) Total pages: 7 (seven) PLEASE DO NOT TURN OVER UNTIL EXAM STARTS PLEASE RETURN THIS BOOKLET

Prac%ce Quiz 8. These are Q s from old quizzes. I do not guarantee that the Q s on this year s quiz will be the same, or even similar.

Lab 10: Spectroscopy & the Hydrogen Atom Phy208 Fall 2008

Name Final Exam May 1, 2017

Emission Spectrum of Atomic Gases. Prelab Questions

Astronomical Techniques

= nm. = nm. = nm

Phys102 Lecture Diffraction of Light

4. Dispersion. The index of refraction of the prism at the input wavelength can be calculated using

Lab 5: Spectroscopy & the Hydrogen Atom Phy248 Spring 2009

UNIT-5 EM WAVES UNIT-6 RAY OPTICS

A beam of coherent monochromatic light from a distant galaxy is used in an optics experiment on Earth.

INTERFEROMETERS. There are 4 principal types of measurements that can be made with this type of interferometer.

Experiment 8 Michelson Interferometer

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 FRAUNHOFER DIFFRACTION

Laboratory #29: Spectrometer

CHARA Meeting 2017 Pasadena, California

Lecture 3 : Electrooptic effect, optical activity and basics of interference colors with wave plates

DISPERSION OF A GLASS PRISM

Michelson Interferometry and Measurement of the Sodium Doublet Splitting

EXPERIMENT NO. 8. OBJECT: To determine the wavelength of sodium light by method.

Confocal Microscopy Imaging of Single Emitter Fluorescence and Hanbury Brown and Twiss Photon Antibunching Setup

OPSE FINAL EXAM Fall 2016 YOU MUST SHOW YOUR WORK. ANSWERS THAT ARE NOT JUSTIFIED WILL BE GIVEN ZERO CREDIT.

Phys 531 Lecture 27 6 December 2005

Transmission Electron Microscopy

Constructive vs. destructive interference; Coherent vs. incoherent interference

Lecture PowerPoints. Chapter 24 Physics: Principles with Applications, 7 th edition Giancoli

Interference, Diffraction and Fourier Theory. ATI 2014 Lecture 02! Keller and Kenworthy

Week 7: Interference

A) n L < 1.0 B) n L > 1.1 C) n L > 1.3 D) n L < 1.1 E) n L < 1.3

Astronomy 203 practice final examination

AS 101: Day Lab #2 Summer Spectroscopy

Nature of Light Part 2

Astronomy. Optics and Telescopes

The Michelson Interferometer and the He-Ne Laser Physics 2150 Experiment No. 3 University of Colorado

LAB DEMONSTRATION OF INTERFEROMETRIC

Speed of Light in Air

Experiment 3 1. The Michelson Interferometer and the He- Ne Laser Physics 2150 Experiment No. 3 University of Colorado

Vågrörelselära och optik

Phys 100 Astronomy (Dr. Ilias Fernini) Review Questions for Chapter 5

Laboratory 3&4: Confocal Microscopy Imaging of Single-Emitter Fluorescence and Hanbury Brown and Twiss setup for Photon Antibunching

Diffraction I. Physics 2415 Lecture 37. Michael Fowler, UVa

Light as Wave Motion p. 1 Huygens' Ideas p. 2 Newton's Ideas p. 8 Complex Numbers p. 10 Simple Harmonic Motion p. 11 Polarized Waves in a Stretched

Imaging applications of Statistical Optics

FIRST YEAR PHYSICS. Unit 4: Light II

Introduction. Procedure and Data

Spectrum of Hydrogen. Physics 227 Lab

Transcription:

Optics 513 - Optical Testing and Testing Instrumentation Lab Lab #6 - Interference Microscopes The purpose of this lab is to observe the samples provided using two different interference microscopes -- the polarization differential microscope (Nomarski) and the two-beam Mirau. White light fringes should be observed using both instruments. By the time you complete this lab you should love white light fringes. Procedure: Nomarski Read the attached write-up on the Nomarski microscope. Place the ball bearing under the objective. Before finding focus adjust the sample so it is too close to the objective, and then find focus by moving the sample away from the objective. This procedure will minimize the chances of crashing the objective into the sample. Adjust the eyepieces so you are comfortable looking at the microscope image. After you think the sample is in focus, move the sample to make sure the image moves and you are really looking at the sample. Adjust the angle of polarization of the incident light and note the effect of changing the direction of polarization. Turn the ring on the objective to move the prism and vary the path difference between the two interfering beams. Note the direction of shear. Repeat the above procedure for looking at the magnetic hard disk. Note both the grooves on the hard disk and surface defects. Mirau Place the flat mirror under the objective. Before finding focus adjust the objective so it is too close to the sample, and then find focus by moving the objective away from the sample. This procedure will minimize the chances of crashing the objective into the sample. Adjust the eyepieces so you are comfortable looking at the microscope image. The easiest way to find focus is to first put in the red filter and then close down the field stop. A bright circular region of light will be seen from the light reflected from the reference surface. Next adjust the objective-sample distance until a faint halo is seen around the bright region of light from the reference. Further adjustment of focus will cause the halo to collapse into the bright region. At this point fringes should be seen. Open the field stop to see the entire field.

Adjust the two knobs on the objective to vary the tip-tilt of the reference surface. Adjust the focus to see the fringe contrast change. Take out the red filter and look at the white light fringes. Adjust the tip-tilt and focus. Fluff the fringes and vary focus. You should see some beautiful colors. By now you should love white light fringes. Repeat the above procedure for looking at the ball bearing, the Ronchi ruling, the curved diffraction grating, and any other sample you want to look at. Questions: 1) The two interferometers give you different data. Describe the difference. 2) The differential interferometer does not detect grooves in a surface if they have a particular orientation. Why? 3) Why do you see different colors with the differential microscope as the polarizer is rotated? 4) Why does translating the Wollaston prism cause different colors to appear? 5) How can you increase the contrast of what you see in the differential interferometer? 6) What condition must be satisfied in the Mirau interferometer to obtain white light fringes? 7) Is the central fringe dark or bright for the Mirau interferometer? Why? Does your answer depend upon the sample? 8) What is the coherence length of the white light source used in the Mirau interferometer? 9) How would the fringes change if the objective lens in the Mirau interferometer had spherical aberration? 10) How would the fringes change if the objective lens in the Mirau interferometer had chromatic aberration? 11) How would you determine if there is a bump or a hole on the test surface? 12) What is the effect of the aperture stop in the Mirau interferometer microscope? 13) Do both interferometers have to be placed on a vibration isolation table? Explain.

Nomarski Microscope The diagram below shows a drawing of the optical layout of a Nomarski microscope. The Nomarski microscope is sometimes called a differential interference contrast (DIC) microscope or a polarization interference contrast microscope. A polarizer after a white light source is used to set the angle of the polarized light incident upon a Wollaston prism. The Wollaston splits the light into two beams having orthogonal polarization, which are sheared with respect to one another. After reflection off the test surface the Wollaston recombines the two beams. A fixed analyzer placed after the Wollaston transmits like components of the two polarizations and generates an interference pattern. The resulting image shows the difference between two closely spaced points on the test surface. The point separation (shear at the test surface) is usually comparable to the optical resolution of the microscope objective and hence only one image is seen. The image shows slope changes and it appears as though the surface has been illuminated from one side. Like a shearing interferometer, only detail in the direction perpendicular to the shear is seen. In other words, if the shear is in the x direction, only features parallel to the y-axis will be seen. Detail parallel to the x direction will not be visible without rotating the test surface or the Wollaston prism. The path difference between the two beams can be adjusted by laterally translating the Wollaston prism. When the axes of the polarizer and analyzer are parallel and the prism is centered, the path lengths are equal and white light is seen for a perfect test surface with no tilt. When the polarizer and analyzer are crossed and the prism centered, no light gets through. When the prism is translated sideways, the two beams have unequal paths

and different colors are seen. The color for a specific feature on the test surface depends upon the path difference between the two beams for that point. The color changes indicate the surface slopes. When the polarizer before the prism is rotated, the relative intensities of the two orthogonal polarized beams change, and the colors change.

Optical Profiler for Measurement of Surface Microstructure CAMERA FIELD STOP APERTURE STOP BEAMSPLITTER EYEPIECE LAMP SPECTRAL FILTER MICROSCOPE OBJECTIVE PZT MIRAU INTERFEROMETER TEST SURFACE