Determination of urea in ultrapure water by IC-MS/MS

Similar documents
Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Dionex IonPac AS28-Fast-4µm column

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Determination of trace anions in concentrated hydrofluoric acid

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of Methylamine in Drug Products

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Inline sample preparation for the determination of anions in sodium hydroxide

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Expanded capabilities in ammonium and amine detection

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determination of Trace Cations in Concentrated Acids Using AutoNeutralization Pretreatment and Ion Chromatography

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Determination of trace anions in high-nitrate matrices by ion chromatography

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

The determination of trace anions in concentrated phosphoric acid

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

columns IonPac SCS 1 Silica Cation Separator

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

columns IonPac AS17-C Anion Exchange Column

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Determination of Trace Anions in Organic Solvents

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Determination of trace anions in concentrated glycolic acid

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

IonPac Trace Cation Concentrator

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Determination of trace anions in organic solvents using matrix elimination and preconcentration

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

The Determination of Trace Anions in Concentrated Phosphoric Acid

Determination of Trace Anions in Concentrated Bases Using AutoNeutralization Pretreatment and Ion Chromatography

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Determination of Adsorbable Organic Halogen in Wastewater

ProPac WCX-10 Columns

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Thermo Scientific. Anion-Exchange Column

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Conc n of A OAs present (um) Conc n of B OAs present (um)

Novel Electrodeionization Devices: Applications in Inorganic Analysis

Fast methods for the determination of ibuprofen in drug products

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Determination of adsorbable organic halogen in wastewater using a combustion ion chromatography system

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

IonPac SCS 1 IonPac SCG 1

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

anion-exchange column

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Determination of Trace Anions in Concentrated Hydrofluoric Acid

Utility of the Charge Detector in Ion Chromatography Applications

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Pelagia Research Library

Transcription:

APPLICATION NOTE 72482 Determination of urea in ultrapure water by IC-MS/MS Authors Soon Fatt Lee, 1 Fiona Teh Hui Boon, 1 Chris Cheah Hun Teong, 1 and Jeff Rohrer 2 ¹Thermo Fisher Scientific, Singapore CMD Application and Training Center 2 Thermo Fisher Scientific, Sunnyvale, CA, USA Keywords Trace analysis, Integrion, semiconductor, Dionex IonPac CS16-Fast-4µm column, HPIC Goal To develop an ion chromatography-mass spectrometry (IC-MS) method to determine low concentrations of urea in ultrapure water (UPW) Introduction Water used in the production of electronic devices, such as the components that make up computers and mobile phones, must be extremely pure. In particular, it must be free of ionic compounds. Typically, manufacturers aim to have the concentration of each ion below 1 µg/l. Another water contaminant that manufacturers seek to minimize is urea, which has been shown to cause problems with semiconductor production. 1 While anions and cations are routinely measured in ultrapure water (UPW) by ion chromatography (IC), the determination of urea is more challenging due to its poor retention by ion-exchange phases and the lack of a suitable detection technique for low concentrations.

This application note describes a method to determine urea that meets these challenges. Urea is retained on a high-performance cation-exchange column and detected by mass spectrometry (MS). Triple-quadrupole MS is used for greater detection specificity and sensitivity. Chromatography is performed using a Thermo Scientific Dionex Integrion HPIC system with RFIC capabilities. This is a compact IC system that automatically generates the eluent needed for chromatography. Positive electrospray ionization was used prior to the detection of urea using a Thermo Scientific TSQ Quantiva triple quadrupole mass spectrometer. This application note describes the development and execution of this method for determining urea in UPW. Experimental Equipment Dionex Integrion HPIC system Thermo Scientific Dionex AS-AP Autosampler TSQ Quantiva Triple Quadrupole Mass Spectrometer Auxiliary 6-port switching valve for Dionex Integrion HPIC system (P/N 22153-62027) for column outlet to divert to MS inlet or waste Thermo Scientific Chromeleon Chromatography Data System software 7.2 SR4 for IC control Thermo Scientific Xcalibur /Thermo Scientific TraceFinder 4.1 software for MS control and data processing Reagents and standards Deionized (DI) water, Type I reagent grade, 18 MΩ cm resistivity or better from Thermo Scientific Barnstead Easypure II system Urea (Sigma-Aldrich P/N U5128-1KG) Samples Seven samples were provided by a semiconductor manufacturer. In this application note, these samples are designated samples 1 through 7. LC Conditions Columns: Thermo Scientific Dionex IonPac CS16-Fast-4µm, 2 mm 150 mm (P/N 088601) Thermo Scientific Dionex IonPac CG16-Fast-4µm, 2 mm 30 mm (P/N 088602) Eluent: 1 mm Methanesulfonic acid (MSA) Eluent Source: Thermo Scientific Dionex EGC 500 MSA Eluent Generator Cartridge (P/N 075779) with a Thermo Scientific Dionex CR-CTC 600 Continuously Regenerated Cation Trap Column (P/N 088663) Flow Rate: 0.2 ml/min Inj. Loop Volume: 45 µl (full loop*) Temperature: 40 C Detection: MS (See below) Run Time: 6 min System Backpressure: ~2,250 psi *Loop prepared from 0.01 in. i.d. PEEK tubing MS Conditions Ionization Method: H-ESI (heated electrospray ionization) Polarity: Positive Spray Voltage: 3,500 V Vaporizer Temperature: 360 ºC Ion Transfer Tube Temperature: 350 ºC Sheath Gas Flow: 38 units Aux Gas Flow: 12 units Scan Mode: Selective Reaction Monitoring (SRM) Q1 & Q3 Resolution: 0.7 FWHM Dwell Time: 0.015 s CID Gas (Ar) Pressure: 1.5 mtorr Chrom Filter: 8 s 2

Table 1. Optimized SRM transitions. Compound Name Retention Time (min) Time Window (min) Polarity Precursor (m/z) Product (m/z) Collision Energy (V) Dwell Time (ms) Urea * 2.34 1.5 Positive 61.1 44 30 15 Urea-dimer 2.34 1.5 Positive 121 61 25 15 *Transition used for quantitation of urea. Additional software used for MS analysis Thermo Scientific TSQ Quantiva Tune 2.0 software was used for system tuning and calibration. The compound SRM optimization wizard was also used. Thermo Scientific TraceFinder 4.1 software was used for instrument method setup, sample sequence setup and submission, automatic detection and integration of SRM peaks, plotting of calibration curve (linear and 1/x weighting), ion ratio confirmation, and experiment report generation. Eluent preparation The 1 mm MSA was prepared by the Dionex Integrion HPIC system eluent generator. The eluent generator also produced 50 mm MSA used to wash the system and column for 60 min after each day of analysis. It is possible to implement a 15 min wash at 50mM MSA into the gradient program when running samples containing higher cation concentrations. This facilitates washing of the column after each sample run. This wash solution is sent to waste and not to the electrospray interface. Preparation of standards Prepare the 1,000 ppm stock solution by weighing 1 g of urea into a 1,000 ml volumetric flask and filling the flask to the mark with DI water. Shake the flask to thoroughly mix the stock solution. To prepare 50 ml of 1 ppm intermediate standard solution, tare the balance with a 50 ml HDPP or HDPE bottle. Pipette 50 µl of the 1,000 ppm stock solution, add DI water to 50.0 g total, cap, and shake the bottle to thoroughly mix the solution. Prepare 50 ml solutions of working standards 1, 2, 3, 4, 5, 6, and 7 in a similar way, by pipetting 20, 25, 50, 100, 250, 500, and 1,000 µl of the intermediate standard into individual 50 ml labeled bottles. Dilute each one to 50.0 g total with DI water and thoroughly shake to mix the working standards. The final concentrations are shown in Table 2. Table 2 shows the low concentration urea standards prepared in DI water to generate the calibration curve for sample analysis. Standards should be prepared on the day of analysis. Sample preparation The samples were analyzed without preparation. After the initial analysis one sample was diluted 1:1 with DI water to have the measured amount of urea in the calibration range. System setup The system setup is shown in Figure 1. A diverter valve is installed after the column so that the column can be washed without sending the effluent to the MS and without having to disconnect the column. To reduce the delay volume between the column and the MS, a short piece of 0.003 in. PEEK tubing was used. Ion source cleaning is basic maintenance for MS systems. Routine ion source cleaning is recommended weekly to prevent any salt clogging or corrosion effects. Table 2. Concentrations of calibration standards. Concentration Std 1 Std 2 Std 3 Std 4 Std 5 Std 6 Std 7 Urea 0.4 0.5 1.0 2.0 5.0 10.0 20.0 3

Waste Waste CS16-Fast-4µm Column with guard 6 5 1 4 3 2 45 µl Sample Loop Diverter Valve Autosampler TSQ Quantiva Triple Quadrupole MS Dionex Integrion HPIC system pump with RFIC EGC-MSA Figure 1. System configuration for IC-MS/MS urea analysis. Results and discussion Chromatography Urea was retained just beyond the void volume of the Dionex IonPac CS16 4-µm column. The 2 mm format of this small-particle-size column was chosen to enhance sensitivity as well as reduce eluent usage and reduce waste. Figure 2 shows the IC-MS/MS chromatograms of one injection of each of the urea calibration standards. The 66.1 to 44 m/z transition was used for detecting urea. Detecting urea by SRM reduces the possibility of interference from other early eluting compounds. The injection volume was optimized to produce the best peak shape and sensitivity. That study found an optimum injection volume of 45 µl. Acetonitrile is sometimes added after an IC separation to enhance MS sensitivity, but this was not required for this application. In fact, experiments using acetonitrile found that sensitivity was compromised due to formation of a urea-acetonitrile complex. This divided the urea between free urea and complexed urea. Calibration and minimum detection limit (MDL) determinations Because the goal was to determine low concentrations of urea in UPW, we used a small range of low concentrations (0.4 to 20 µg/l) with more points at the lower end of the concentration range. Three injections of each standard were made. The results of the calibration experiment showed a linear relationship between concentration and peak area response by least squares analysis (coefficient of determination (r 2 ) of 0.9979) and by visual inspection of the plotted data (Figure 3). The MDL was estimated by making seven injections a low concentration standard (0.4 µg/l). This study yielded a MDL of 0.14 µg/l using the formula: MDL = 3.142 average amount %RSD/100 (3.142 is the Student s t value for a 99% confidence level and a standard deviation estimate with n-1 degrees of freedom for seven replicates.) Figure 4 shows a chromatogram of one of the injections of the 0.4 µg/l standard. Interference study To determine if common cations could interfere with the determination of urea in UPW, 5 µg/l urea was spiked into a water sample containing 25, 100, 125, 250, 125, and 250 µg/l of lithium, sodium, ammonium, potassium, magnesium, and calcium, respectively. While these concentrations are not high, they are higher than would be expected in UPW, and because urea is weakly bound to the column, it is very easy to displace it. Three injections of this sample yielded an 82.7% recovery, suggesting the method is appropriate for UPW. 4

Blank 0.4 ppb 0.5 ppb 1 ppb 2 ppb 5 ppb 10 ppb 20 ppb Figure 2. Urea calibration standards chromatograms. Figure 3. Urea calibration plot. Concentration Figure 4. Chromatogram of a 0.4 µg/l urea standard. 5

Sample analysis Seven water samples from different parts of a semiconductor manufacturing process were provided by a customer (Figure 5). The results of our analysis are shown in Table 3. Sample #1 required dilution to have the urea concentration in the range of the calibration plot. Method stability during this analysis was checked by intermittently injecting 2 and 5 µg/l check standards. Each standard was injected three times during the sample analysis. The 2 µg/l standard had a 98.6% recovery and the 5 µg/l had a 101% recovery. This suggests good method stability during this analysis. To assess accuracy, samples #1, #2, and #6 were spiked with 5 and 10 µg/l urea and samples #2 and #6 were spiked with 2 µg/l urea. Each spiked sample was injected in triplicate. For these eight spiked sample preparations, the recovery ranged from 94.5% to 107%, suggesting that the determinations are accurate. #1* #2 #3 #4 #5 #6 #7 Figure 5. Sample chromatograms. 6

Table 3. Sample analysis. #1* #2 #3 #4 #5 #6 #7 Replicate 1 20.8 0.508 0.411 0.166 0.754 0.164 0.152 Replicate 2 19.5 0.481 0.456 0.172 0.815 0.150 0.162 Replicate 3 19.8 0.489 0.456 0.187 0.837 0.152 0.160 Average 20.1 0.493 0.441 0.175 0.802 0.155 0.158 RSD 3.28 2.81 5.89 6.18 5.36 4.88 3.35 *Required a 1:1 dilution with DI water Conclusion The IC-MS/MS method described in this application note can accurately determine low µg/l concentrations of urea in UPW, with a MDL of < 0.2 µg/l. Urea can be detrimental to the successful production of semiconductors. This method is easy to execute as the eluent is automatically prepared by the Dionex Integrion HPIC system with RFIC capability. The MS/MS detection provides both excellent sensitivity and confidence in peak identification. Reference 1. Godex, R. Measurement and Removal of Trace Levels of Urea in Semiconductor High-Purity Water. [Online] http://www.astsciencecorp.com/files/file/ge%20toc/ IntelGodecHenleyEdit32.pdf (accessed August 28, 2017). Find out more at thermofisher.com/icms 2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN72482-EN 1017S