University of Groningen. The HVEE C-14 system at Groningen Gottdang, Andreas; Mous, Dirk J.W.; van der Plicht, Johannes. Published in: Radiocarbon

Similar documents
AN ION SOURCE FOR THE HVEE 14C ISOTOPE RATIO MASS SPECTROMETER FOR BIOMEDICAL APPLICATIONS

A HIGH THROUGHPUT 14C ACCELERATOR MASS SPECTROMETER

C - AMS. Introduction Identification of 14 C. Remarks for users

University of Groningen. Laser Spectroscopy of Trapped Ra+ Ion Versolato, Oscar Oreste

Accelerator Mass Spectroscopy

ACCELERATOR MASS SPECTROMETRY AT THE LUND PELLETRON ACCELERATOR

University of Groningen. Hollow-atom probing of surfaces Limburg, Johannes

Citation for published version (APA): Hoekstra, S. (2005). Atom Trap Trace Analysis of Calcium Isotopes s.n.

FIRST RADIOCARBON MEASUREMENTS AT BINP AMS

University of Groningen. Extraction and transport of ion beams from an ECR ion source Saminathan, Suresh

3.5. Accelerator Mass Spectroscopy -AMS -

Theoretical simulation of nonlinear spectroscopy in the liquid phase La Cour Jansen, Thomas

Citation for published version (APA): Mollema, A. K. (2008). Laser cooling, trapping and spectroscopy of calcium isotopes s.n.

Optimization of 10 Be and 26 Al detection with low-energy accelerator mass spectrometry

MAURICE ARNOLD, EDOUARD BARD, PIERRE MAURICE HELENE VALLADAS and J C DUPLESSY

corresponding l Nuclear Physics Laboratory and Quaternary Isotope Laboratory University of Washington, Seattle, Washington INTRODUCTION

Citation for published version (APA): Shen, C. (2006). Wave Propagation through Photonic Crystal Slabs: Imaging and Localization. [S.l.]: s.n.

SUPPLEMENTARY INFORMATION

Citation for published version (APA): Kooistra, F. B. (2007). Fullerenes for organic electronics [Groningen]: s.n.

accelerating opportunities

SIMULATION STUDY FOR THE SEPARATION OF RARE ISOTOPES AT THE SEOUL NATIONAL UNIVERSITY AMS FACILITY

Can a Hexapole magnet of an ECR Ion Source be too strong? Drentje, A. G.; Barzangy, F.; Kremers, Herman; Meyer, D.; Mulder, J.; Sijbring, J.

CAPABILITIES OF THE NEW SUERC 5MV AMS FACILITY FOR 14 C DATING

THE NEW 14 C ANALYSIS LABORATORY IN JENA, GERMANY. A Steinhof 1,2 G Adamiec 3 G Gleixner 1 GJ van Klinken 4 T Wagner 1

Spin caloritronics in magnetic/non-magnetic nanostructures and graphene field effect devices Dejene, Fasil

R JUERG BEER, HANS OESCHGER, and MICHEL ANDREE Physikalisches Institut, Universitat Bern, CH-3012 BERN, Switzerland

J H SHEA*, T W CONLON, JAMES ASHER, and P M READ. in the dating of artifacts (Gove, 1978).

Enhancement of spin relaxation time in hydrogenated graphene spin-valve devices Wojtaszek, M.; Vera-Marun, I. J.; Maassen, T.

Citation for published version (APA): Mendoza, S. M. (2007). Exploiting molecular machines on surfaces s.n.

Superfluid helium and cryogenic noble gases as stopping media for ion catchers Purushothaman, Sivaji

RADIOCARBON ANALYSIS OF DIFFEREENT SAMPLES AT BINP AMS

Citation for published version (APA): Andogah, G. (2010). Geographically constrained information retrieval Groningen: s.n.

New Tandetron Accelerator Laboratory at MTA Atomki, Debrecen: Multi-Purpose and Multi-User Facility

Citation for published version (APA): Sok, R. M. (1994). Permeation of small molecules across a polymer membrane: a computer simulation study s.n.

The scanning microbeam PIXE analysis facility at NIRS

Citation for published version (APA): Martinus, G. H. (1998). Proton-proton bremsstrahlung in a relativistic covariant model s.n.

ION BEAM OPTIC SIMULATIONS AT THE 1 MV TANDETRON TM FROM IFIN-HH BUCHAREST

NEW AND IMPROVED AMS FACILITIES

University of Groningen. Study of compression modes in 56Ni using an active target Bagchi, Soumya

Citation for published version (APA): Kooistra, F. B. (2007). Fullerenes for organic electronics [Groningen]: s.n.

University of Groningen. Statistical inference via fiducial methods Salomé, Diemer

Citation for published version (APA): Shen, C. (2006). Wave Propagation through Photonic Crystal Slabs: Imaging and Localization. [S.l.]: s.n.

Beam Diagnostics Lecture 3. Measuring Complex Accelerator Parameters Uli Raich CERN AB-BI

It was first employed by Purser and others (1977) and Bennett

University of Groningen. Enantioselective liquid-liquid extraction in microreactors Susanti, Susanti

Ultratrace analysis of radionuclides by AMS

Citation for published version (APA): Fathi, K. (2004). Dynamics and morphology in the inner regions of spiral galaxies Groningen: s.n.

Published in: ELECTRONIC PROPERTIES OF NOVEL MATERIALS - PROGRESS IN MOLECULAR NANOSTRUCTURES

ARE COMPACT AMS FACILITIES A COMPETITIVE ALTERNATIVE TO LARGER TANDEM ACCELERATORS?

University of Groningen. Water in protoplanetary disks Antonellini, Stefano

University of Groningen

Defining quality standards for the analysis of solid samples

Sensitized solar cells with colloidal PbS-CdS core-shell quantum dots Lai, Lai-Hung; Protesescu, Loredana; Kovalenko, Maksym V.

of mass spectrometry

University of Groningen. Event-based simulation of quantum phenomena Zhao, Shuang

Secondary ion mass spectrometry (SIMS)

PROCEEDINGS OF THE 16 TH INTERNATIONAL 14 C CONFERENCE, EDITED BY W. G. MOOK

HIGH-PRECISION INTERCOMPARISON AT ISOTRACE

Peptide folding in non-aqueous environments investigated with molecular dynamics simulations Soto Becerra, Patricia

Carbon dioxide removal processes by alkanolamines in aqueous organic solvents Hamborg, Espen Steinseth

STATUS OF THE NEW AMS FACILITY IN MAGURELE: THE CENTER OF RADIOCARBON FOR ENVIRONMENT AND BIOSCIENCES *

SYSTEMATIC INVESTIGATIONS OF 14C MEASUREMENTS AT THE VIENNA ENVIRONMENTAL RESEARCH ACCELERATOR

Citation for published version (APA): Nouri-Nigjeh, E. (2011). Electrochemistry in the mimicry of oxidative drug metabolism Groningen: s.n.

University of Groningen

EBSS: Hands-on Activity #8. Operation of an electromagnetic isotope separator (EMIS) Dan Stracener Jon Batchelder

DEVELOPMENT OF LARGE SCALE OPTIMIZATION TOOLS FOR BEAM TRACKING CODES*

University of Groningen

System-theoretic properties of port-controlled Hamiltonian systems Maschke, B.M.; van der Schaft, Arjan

IRMS. perspective. Isotope Ratio Mass Spectrometry.

Information about the T9 beam line and experimental facilities

Substrate and Cation Binding Mechanism of Glutamate Transporter Homologs Jensen, Sonja

Citation for published version (APA): Boomsma, R. (2007). The disk-halo connection in NGC 6946 and NGC 253 s.n.

Progress of an accelerator mass spectrometry system on the Tsukuba 12UD Pelletron tandem accelerator

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

University of Groningen. Taking topological insulators for a spin de Vries, Eric Kornelis

University of Groningen. Morphological design of Discrete-Time Cellular Neural Networks Brugge, Mark Harm ter

Performance Studies of the Vibration Wire Monitor on the Test Stand with Low Energy Electron Beam

A Multi-beamlet Injector for Heavy Ion Fusion: Experiments and Modeling

University of Groningen. Levulinic acid from lignocellulosic biomass Girisuta, Buana

Citation for published version (APA): Wang, Y. (2018). Disc reflection in low-mass X-ray binaries. [Groningen]: Rijksuniversiteit Groningen.

Extrel Application Note

University of Groningen. Photophysics of nanomaterials for opto-electronic applications Kahmann, Simon

ICP-MS. High Resolution ICP-MS.

arxiv: v2 [physics.acc-ph] 21 Jun 2017

Huashun Zhang. Ion Sources. With 187 Figures and 26 Tables Э SCIENCE PRESS. Springer

Fig. 1 : Voltage vs hours graph for the 15 UD Pelletron

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur

University of Groningen

Accelerator Mass Spectrometry and Bayesian Data Analysis

Citation for published version (APA): Ruíz Duarte, E. An invitation to algebraic number theory and class field theory

University of Groningen. Statistical Auditing and the AOQL-method Talens, Erik

IAP NAN of Ukraine activity in. Possible areas of cooperation in. Roman Kholodov, PICS Meeting, Basivka, Ukraine,

Diagnostics Requirements for the ARIEL Low Energy Beam Transport

Mass Spectrometers. for Surface Analysis SURFACE ANALYSIS

Secondary Ion Mass Spectrometry (SIMS) Thomas Sky

has long-lived been used radioisotopes to detect by counting the accelerated ions. beams has been solved by combining

Dual photo- and redox- active molecular switches for smart surfaces Ivashenko, Oleksii

1 AT/P5-05. Institute of Applied Physics, National Academy of Sciences of Ukraine, Sumy, Ukraine

Refurbishment of the Electrostatic Accelerators of ithemba LABS

Alpha-Energies of different sources with Multi Channel Analyzer

Transcription:

University of Groningen The HVEE C-14 system at Groningen Gottdang, Andreas; Mous, Dirk J.W.; van der Plicht, Johannes Published in: Radiocarbon IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document version below. Document Version Publisher's PDF, also known as Version of record Publication date: 1995 Link to publication in University of Groningen/UMCG research database Citation for published version (APA): Gottdang, A., Mous, D. J. W., & Plicht, J. V. D. (1995). The HVEE C-14 system at Groningen. Radiocarbon, 37(2), 649-656. Copyright Other than for strictly personal use, it is not permitted to download or to forward/distribute the text or part of it without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license (like Creative Commons). Take-down policy If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim. Downloaded from the University of Groningen/UMCG research database (Pure): http://www.rug.nl/research/portal. For technical reasons the number of authors shown on this cover page is limited to 10 maximum. Download date: 18-09-2018

THE HVEE 14C SYSTEM AT GRONINGEN ANDREAS GOTTDANG, DIRKJ. W. MOUS High Voltage Engineering Europa B.V., Postbus 99, 3800 AB Amersfoort, The Netherlands and JOHANNES VAN DER PLICHT Centre for Isotope Research, University of Groningen, Nijenborgh 4, 9747 AG Groningen The Netherlands ABSTRACT. Since May 1994, a new-generation accelerator mass spectrometer (AMS) has been fully operational at the Centre for Isotope Research in Groningen, The Netherlands. The fully automated and high-throughput accelerator mass spectrometry (AMS) system, manufactured by High Voltage Engineering Europa (HVEE) is dedicated to radiocarbon analysis. The HVEE 413014C AMS is able to analyze up to 3000 samples per year. The system is characterized by simultaneous transport of all three isotopes (12C,'3C,14C) and 14C analysis with a precision below 0.5 pmc and a daily stability below 0.5 pmc. We present here a system description together with stability and performance measurements. INTRODUCTION Accelerator mass spectrometry (AMS) is an important dating tool in archaeology, geosciences, oceanography and other related research fields including biomedical applications. The ability to measure samples 1000 times smaller niques than together those used with in high liquid throughput scintillation broadens or gas counting the applications of tech- AMS for the future (Davis Mook 1984). When combined with precise radiocarbon dating, these 1994; advantages 14C02 allow, for example, tracing in seawater (Jones et a1.1990; von Reden et a1.1994). At the Centre for Isotope Research in Groningen, low background measurements (14C by decay counting and 3H) and mass spectrometry 15N ry (H, (2 180) ) have been used for many years. A new generation high-precision and high-throughput 14C AMS system has been in routine operation since May 1994. Because high-precision 14C dating with low detection thresholds requires improved control over fractionation and contamination effects of the samples (Beukens 1993; Bonani et a1.1990; Suter 1990), the sample preparation technique with the graphitization process, the procedure for filling the AMS sample carriers and the carrier itself have also been improved. These improvements are discussed elsewhere (van der Plicht et a1.1995). The HVEE 413014C AMS system (Mous, Gottdang and van der Plicht 1994; Purser et al. 1988, 1990; Purser 1992a), built by High Voltage Engineering Europa, is the second spectrometer of this type that measures all three carbon isotopes simultaneously. The high-throughput machine is capable of dating up to 3000 samples per year. Simultaneous transport of the three isotopes has clear advantages over serial injection (Wolfli 1987). With simultaneous injection, all individual isotopes are measured under the same conditions, which minimizes fractionation effects. Further, this "DC operation" provides an online diagnostic tool for the machine performance during measurements. After a brief description of the system, we show some stability measurements performed on various sections of the 14C AMS system. We present data that show the stability of the system together with performance measurements obtained from modern samples. Further results are presented in van der Plicht et al. (1995). Proceedings of the 15th International '4C Conference, edited by G. T. Coop D. D. Harkness, B. F. Miller and E. M. Scott. RADIOCARBON, Vol. 37, No. 2,1995, P. 649-656 649

650 A. Gottdang, D. J. W. Mous and J. van der Plicht SYSTEM DESCRIPTION The principles of operation of the HVEE 4130 AMS system were described previously (Purser et al. 1988,1990; Purser 1992a,b,c). Figure 1 shows the layout of the machine. The system can be divided into five main sections: 1) the ion source; 2) the recombinator; 3) the Tandetron tandem accelerator; 4) the high-energy mass spectrometer; and 5) the computer control system with data acquisition. 1. A high current cesium sputter-type ion source with a hemispherical ionizer and an improved cesium vapor delivery system produces a -36 kev negative ion beam from graphite samples. Up to 59 samples can be inserted and manipulated in this computer-controlled source, whereby the time for an unload and load cycle is <40 sec. To minimize cross-contamination among samples, the targets are stored in a revolving wheel 100 mm from the sample under analysis. To avoid serious sample cratering by the sputtering Cs+ beam, the loaded sample is moved during analysis perpendicular to the beam direction. Large cratering can introduce source fractionation and a varying beam emittance, which degrades overall system performance. The ion beam extracted from the source is focused with an einzel lens through a beam defining aperture. 2. The recombinator (Litherland and Kilius 1990, 1991) consists of four second-order corrected 45 magnets. The first pair of magnets analyzes the beam and separates the desired masses 12, 13 and 14 at the symmetry plane of the recombinator by ca. 20 mm. Moveable beam stops allow for every beam component to be injected and analyzed separately. A chopper wheel reduces the 12C beam load for the accelerator and the high-energy spectrometer by a factor of ca. 90 (Purser 1992a). The second pair of magnets recombines the individual beams back on one track for simultaneous injection in the tandem. The recombinator is achromatic, i.e., the beam positions at the image of the recombinator are not influenced by small changes in beam energy ("Brown Achromat") (Purser 1992a). 3. In the Tandetron accelerator, the singly charged negative ions are focused (Purser 1993) and accelerated to a terminal, which operates at +2.5 MV. After charge exchange to the mean charge state 3+, by collisions with argon atoms in a recirculating gas stripper canal, the ions are accelerated back to ground potential to reach an energy of 10 MeV. The 3+ charge state is selected to overcome background problems from metastable molecules such as 12CH2n+ or 13CH. Mol- TANDEM ACCELERATOR g STRIPPING CANAL HIGH ENERGY MASS SPECTROMETER 59 SAMPLE ION SOURCE Fig. 1. Schematic layout of the HVEE 14C AMS system

The HVEE 14C System at Groningen 651 ecule ions with a 3+ charge state have never been detected by AMS (Lee 1987). Specially designed inclined-field acceleration tubes (Purser 1992a) efficiently suppress ambiguities of ions with the same mass-energy product as the desired 10 MeV 1403+ particles, which result from charge exchange processes during the acceleration to ground potential. 4. The final analysis is done with an achromatic high-energy spectrometer, consisting of a 110 magnet, a 33 electrostatic g deflector and a 90 magnet. The first magnet selects the 3+ charge state and separates the three isotopes. The 12C and 13C beams are measured in individual Faraday cups, located in the image point of the double-focusing 110 magnet. ical currents obtained from p samples produced with the previously mentioned preparation technique are in the range of 300 to 800 na. The 14C particles pass two additional dispersive elements, a 33 electrostatic deflector, which blocks almost all residual ambiguities with the same mass energy product as 10 MeV 14C3, + and a 90 magnet for further background reduction, introduced by scattering effects at the 33 deflector plates or by two-step charge-exchange processes with the residual gas. The latter background is also minimized by maintaining vacuum conditions in the 14C spec- trometer below 10-8 mbar. 14C ions are detected in an isobutane- filled ' ionization chamber isolated from the vacuum system by a Mylar foil (.u.250,ug cm-2 thick). This detector provides a de/ dx and ainal signal. By gating g these signals with standard electronic devices all unwanted par- ticles are eliminated from the spectrum. For a highly stable terminal voltage, the 13C cup is provided with an integrated p slits Y stem. This setup p creates a slit error signal, which is related to the beam displacement in the cup. Application of low bias operational amplifiers (-1 pa) ensures that the measured 13 C current is not affected (Mows, Gottdang and van der Plicht 1994). 5. The complete low-level control of the system is done by fiber-optic links to a computer pro cessing unit. The resolution and the stability of these controls are better than 10-3. Every highlevel action can be performed with two personal computers, one for the source control and one for overall system control and data acquisition. This system configuration allows for unattended automatic operation around the clock. STABILITY OF INDIVIDUAL MACHINE SECTIONS Stability is the major key factor for reliable high-precision measurements. Thus, we put special emphasis on different machine sections to ensure that t he analysis is unaffected by drift or fractionation effects on a long-term basis. Recombinator The recombinator provides the simultaneous injection of the masses 12,13 and 14 into the accelerator and the high-energy spectrometer. To minimize fractionation effects resulting from possible misalignments of the individual beams with respect to the mechanical axis of the system, we optimized the recombinator components as described previously (Mows, Gottdang and van der Plicht 1994). Measurements showed that after this optimization, all three beams recombine within 0.3 mm and 0.4 mrad. The ion-beam radius at the image is ca. 3 mm and the half angle divergence 12 mrad. To confirm that no fractionation is created in the recombinator, we performed a scan with the recombinator magnets. Figure 2 shows the 13C/12C ratio, measured in the Faraday cups in the high-energy spectrometer. During the scan, the beams are shifted by ca. 8 mm in symmetry plane of the recombinator. The measured relative stability of the magnet current is -3.5 x 10-4 in 12 h. The range in which the system works during normal operation is indicated in the graph and corresponds to a variation of <0.2 mm. This measurement shows that fractionation effects in the recombinator are negligible.

652 A. Gottdang, D. J. W. Mous and J. van der Plicht 1.076 r N O X 1.075 U 1.074 N r r U M 1.073 1.072-58,6 58.7 58.8 58.9 59.0 59.1 59.2 59.3 59.4 59.5 Fig. 2.13C/12C ratio during a scan with the recombinator magnets Magnet Current (A) Slit Feedback Stabilization Figure 3 shows the measured current from the 13C cup and the corresponding slit error signal during a scan with the 110 magnet. During this measurement, the error signal was not taken as the reference for the terminal voltage. The linear slope of the slit error signal at ±1 mm around the zero position makes it possible to relate the error signal to the actual beam displacement in the image point after the 110 magnet. This information can be easily used to tune the high-energy spectrometer from day to day in exactly the same manner. The time dependence of the error signal and the related beam displacement are shown in Figure 4. The ion beam was intentionally shifted with the 110 magnet by 0.3 mm and the slit feedback was switched off. After 1.3 sec, the slit feedback was activated and the beam was forced back to the zero position. The graph shows that the small fluctuations of ca. 0.15 mm disappear when the slit feedback stabilization is switched on. We also note that by using slit feedback stabilization, the center of the beam in the high-energy mass spectrometer is inherently stable below 0.1 mm over the long term. MACHINE PERFORMANCE Figure 5 shows a typical 14C spectrum from a sucrose sample (150.61 pmc). Only the Efinal detector of the ionization chamber is used here; no gating with the de/dx detector signal was done. It took 22 min to measure the 201,258 counts in the 14C peak (counting statistics: 2.2%o). The insert shows that the 14C peak is clearly separated from the neighboring peaks and the 14C count rate can be easily extracted by using a software window. The exact origin of the other two peaks is still under investigation. We believe multiple charge exchanges and scattering effects are involved here.

The HVEE 14C System at Groningen 653 Q c 300 0 6 4 2 0-2 -4-6 -8-6 -4-2 0 2 4 Beam Displacement (mm) 6 8 Fig. 3.13C current with corresponding slit error signal during a scan with the 110 magnet over the 13C cup 2.0 1.5 feedback off 1.0 E 0.2 v E 0.5 0.1 w E 4, 0,0 t 0.0-4 a- (1) -0.5 0.1 o ::: 1-0.2 0.3-2.0,i,,f,,f --0.4 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 Time (sec) feedback on Fig. 4. Measured slit error signal and corresponding beam displacement with slit feedback stabilization switched off and on 0.4 0.3

654 A. Gottdang, D. J. W. Mous and J. van der Plicht 9000 5000 7000 E14C-201258 1 "C window J 6000 5000 340 380 420 460 500 4000 3000 2000 1000 100 200 300 400 500 600 700 800 900 Channels Fig. 5. Ungated 14C energy spectrum measured in.22 min from a sucrose (150.61 pmc) sample In Figure 6, the 13C/12C ratio is measured for >2 h. We use six sucrose samples in 12 runs; every sample is loaded twice. Each data point in the graph represents the mean of 79 current measurements carried out in 40 sec. No long-term drift is observed on a per mil level. Note that all data points are within 4%o. The small drift for each individual run is probably related to the warming of the sample by the cesium sputter beam. To avoid these drifts, an improved target cooling clamp has been installed and is being studied. Table 1 shows the mean and the standard deviation of all three isotope ratios, determined sequentially for two days. The 11 sucrose samples were produced in one batch using the preparation tech- TABLE 1. Results from measurements of two succeeding days. The 12C and 13C currents were in the range of 500 to 650 na on both days. 13C/12C 14C/13C 14C/l2C Sucrose 150.61 pmc Mean o (x10-2) any (x10-10) 1st day 6 samples 1.0719 2.2%o statistics 2nd day 5 samples 1.0696 1.7%o statistics Difference between means of both 2.1%o 0.99% 1.2%o (0.18 pmc) days

I I I I I I I I I I The HVEE 14C System at Groningen 655 sample #1 #2 #3 #4 #5 #6 #1 #2 #3 #4 #5 #6 0 1.073 I v 1 I I I f1 X I I I 1 I 1.072 I I I t IT i i,? 1.071 ' Lf I I I 1.069 F-! I 0 32 64 98 128 Time (min) Fig. 6. Stability measurement of the 13C/12C-ratio over 2 h for six different sucrose samples. Each sample was loaded twice in the ion source. nique developed by van der Plicht et at. (1995). The counting statistics on the two days were 2.2%o and 1.7%o, respectively. The ion source and the terminal voltage were switched off between days. We rejected no data block and used only a simple software window for the 14C counts from the detector. Table 1 shows that a precision <0.5 pmc is obtainable on a routine basis. We achieved a daily stability below 0.5 pmc with this AMS system. CONCLUSION The fully automated and high-throughput HVEE 14C AMS system at the University of Groningen is fully operational. Special emphasis has been placed on critical components of the AMS machine to reduce machine fractionation and to increase long-term stability. The recombinator provides simultaneous injection with negligible fractionation. The simultaneous analysis of all three carbon isotopes is an online diagnostic tool, which checks the performance of the AMS continuously. The slit feedback stabilization guarantees a stable beam transport in the high-energy spectrometer on a longterm basis. Precision measurements with modern samples have shown that a precision below 0.5 pmc is obtainable on a routine basis with a daily stability below 0.5 pmc. ACKNOWLEDGMENTS The Groningen 14C AMS was funded by the University of Groningen and the Ministry of Economic Affairs, The Hague, The Netherlands. We thank K. H. Purser and A. E. Litherland and his group for many fruitful discussions.

656 A. Gottdang, D. J. W. Mous and J. van der Plicht REFERENCES Beukens, R. P. 1993 Radiocarbon accelerator mass spectroscopy: Background and contamination. Nuclear Instruments and Methods in Physics Research B79: 620-623. Bonani, G., Eberhardt, P., Hoffmann, H. J., Niklaus, Th. R., Suter, M., Synal, H. A. and Wolfli, W. 1990 Efficiency improvements with a new stripper design. Nuclear Instruments and Methods in Physics Research B52: 338-344. Davis, J. C. 1994 AMS beyond 2000. Nuclear Instruments and Methods in Physics Research B92:1-6. Jones, G. A., McNicol, A. P., von Reden, K. F. and Schneider, R. J. 1990 The National Ocean Sciences AMS Facility at Woods Hole Oceanographic Institution. Nuclear Instruments and Methods in Physics Research B52: 278-284. Lee, H. (ms.) 1987 Study of carbon-14 detection and dating using accelerator mass spectrometry. Ph. D. dissertation, University of Toronto: 152 p. Litherland, A. E. and Kilius, L. R. 1990 A recombinator for radiocarbon accelerator mass spectrometry. Nuclear Instruments and Methods in Physics Research B52: 375-377. 1991 Mass recombinator for accelerator mass spectrometry. US Patent No. 5.013.923. Mook, W. G. 1984 Archaeological and geological interest in applying 14C AMS to small samples. Nuclear Instruments and Methods in Physics Research B5: 297-302. Mous, D. J. W., Gottdang, A. and van der Plicht, J. 1994 Status of the first HVEE 14C AMS in Groningen. Nuclear Instruments and Methods in Physics Research B92:12-15. Purser, K. H. 1992a A high throughput 14C accelerator mass spectrometer. In Long, A. and Kra, R. S., eds., Proceedings of the 14th International 14C Conference. Radiocarbon 34(3): 458-467. 1992b Acceleration apparatus which reduced background of accelerator mass spectrometry measurements of 14C and other radio nuclides. U.S. Patent No. 5.120.956. 1992c Accuracy of AMS isotopic ratio measurements. U.S. Patent No. 5.118.936. 1993 Injection system for tandem accelerators. U.S. Patent No. 5.247.263. Purser, K. H., Smick, T., Litherland, A. E., Beukens, R. P., Kieser, W. E. and Kilius, L. R. 1988 A third generation 14C accelerator mass spectrometer. Nuclear Instruments and Methods in Physics Research B35: 284-291. Purser, K. H., Smick, T. H. and Purser, R. K. 1990 A precision 14C accelerator mass spectrometer. Nuclear Instruments and Methods in Physics Research B52: 263-268. Suter, M. 1990 Accelerator mass spectroscopy: State of the art in 1990. Nuclear Instruments and Methods in Physics Research B52: 211-223. van der Plicht, J., Aerts, A. T., Wijma, S. and Zondervan, A. 1995 First results from the Groningen AMS Facility. Radiocarbon, this issue. von Reden, K. F, Schneider, R. J., Cohen, G. J. and Jones, G. A. 1994 Performance characteristics of the 3 MV Tandetron AMS system at the National Ocean Sciences AMS facility. Nuclear Instruments and Methods in Physics Research B92: 7-11. Wolfli, W. 1987 Advances in accelerator mass spectrometry. Nuclear Instruments and Methods in Physics Research B29:1-13.