The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane

Similar documents
environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Selection of a Capillary

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

Selection of a Capillary GC Column

Choosing the Correct GC Column Dimensions and Stationary Phase

The Suite for Environmental GC Analysis

Low Bleed Stationary Phases for Gas Chromatography

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

Rely on Rxi -PAH Columns

SUPELCO. Improved Performance for Cyanopropylphenyldimethylpolysiloxane (SPB-1701, SPB-1301) Capillary Columns

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Introduction and Principles of Gas Chromatography

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

Facility: Mansfield Revision 3. Title: PAHs by SPME Page 1 of 11

Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood Using GC/MS

Determination of 24 PAHs in Drinking Water

The Role of the Carbon Cartridge in the One-Pass System

Amec Foster Wheeler Environment & Infrastructure, Inc BIG SHANTY ROAD, NW, SUITE 100 KENNESAW, GEORGIA (770)

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Non-Potable Water A2LA Certificate

A Look at Column Choices. Ed Kim Application Engineer Agilent Technologies, Inc February 12, 2009

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

Application Note GCMS-09 Evaluation of Rapid Extraction and Analysis Techniques for Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood by GC/MS/MS

Rtx -CLPesticides Columns Reduce Analysis Time; Reduce System Maintenance

Application Note # CA

Appendix 1: Polycyclic Aromatic Compounds: Nomenclature and Analysis

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

STANDARD OPERATING PROCEDURES

Column Selection. there is more to life than a boiling point column. Jaap de Zeeuw Restek Corporation, The Netherlands. Copyrights: Restek Corporation

Table 1 Soil and Groundwater Sampling Summary Table

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

GC Instruments. GC Instruments - Sample Introduction

A Single-Method Approach for the Analysis of Volatile and Semivolatile Organic Compounds in Air Using Thermal Desorption Coupled with GC MS

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

EL608. Deodorants [EL /1/ ]

Methods of Analysis of Polychlorinated Biphenyl Congeners Using an Application-Specific Capillary GC Column

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

2. September 2015 Sub-Concrete Slab Soil Investigation

Slightly Polar Columns

Making Improvements to Today s Natural Gas Analysis Systems

Introduction to Capillary GC

Supporting Information

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission

Understanding Gas Chromatography

Volatile organic compounds (VOCs):

Results of a Sediment Survey in the Near Offshore Waters of the Proposed Quarry Site in the Vicinity of Whites Cove, Digby Neck, Nova Scotia

SUPELCO. SPME-Adsorption versus Absorption: Which Fiber is Best for Your Application? Robert E. Shirey and Raymond F. Mindrup

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Appendix 2 Data Distributions for Contaminants in Water and Sediment Samples in Response to the Deepwater Horizon Oil Spill, 2010

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

TETRA TECH, INC. December 8, Lynn Nakashima Berkeley Regional Office 700 Heinz Avenue, Suite 200C Berkeley, California 94710

ILLA SpA VIA GHISOLFI GUARESCHI, NOCETO (PR)

Dorothea Gilbert, Espen Eek, Naiara Berrojalbiz, Amy MP Oen and Hans Peter H. Arp. Norwegian Geotechnical Institute, Oslo

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Test Report. Report No. ECL01H Page 1 of 5

Analysis of Organic Pollutants by Micro Scale Liquid-Liquid Extraction. and On-column Large Volume Injection Gas Chromatography. Mark S.

What Does the EPA Method Round Robin Data Really Tell Us?

Andrei Medvedovici, Florina Micăle, Florentin Tache

Certificate of Accreditation

STANDARD OPERATING PROCEDURES SOP: 1821 PAGE: 1 of 33 REV: 0.0 DATE: 11/25/94

APPENDIX G. Data Management Rules. Dioxin Data Report Appendix G. Lower Duwamish Waterway Superfund Site: T-117 Early Action Area

Place a collection tube beneath cartridge. 3 ml methylene chloride/isopropyl alcohol/ hydrochloric acid (60/40/1) Collect the eluant

The direct effect of the harbour. micropollutant concentration. Workshop Characterization of atmospheric pollution in harbour areas 26 June 2013

by Dr. George M. Frame

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Reviewing Material Safety Data Sheets to Verify Significant Drinking Water Threats

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

APPENDIX D Laboratory Analytical Reports

Pursuant to Chapter 4 of the Code of Ethics, ETAD members may not manufacture or sell dyes referred to in Annexes A and B. Annex A

APPENDIX A TO PART 136 METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND INDUSTRIAL WASTEWATER METHOD 610 POLYNUCLEAR AROMATIC HYDROCARBONS

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Get Selective. By Jaap de Zeeuw

Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas

NEW Agilent J&W Ultra Inert Capillary GC Columns. Proven performance that raises the bar on. consistent. column inertness.

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS IN SOIL AT AL-NAHRAWAN BRICKS FACTORY Thamera K. M. Al-Rudaini 1 and Israa M.H.

Rapid Analysis of PAHs in Fly Ash Using Thermal Desorption and Fast GC TOF-MS

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Tri-n-propyltin Chloride Mixture

2401 Gas (liquid) Chromatography

Transcription:

The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane Jarl Snider, D. J. Hotnisky, Kristi Sellers, Dinesh V. Patwardhan Ph. D. Restek Corporation www.restekcorp.com

utline Background Definition of Terminology Unique Attributes of High Temperature Column Applications Conclusions

Polydimethylsiloxane H n H H 3 C n

Thermal Stability of PDMS Beginning of decomposition Adapted from loxane Polymers, ed. Clarson & Semlyen, 1993.

Bleed Chromatography of loose silicones from the column Bleed pa temperature

Typical Chemical Structure of Bleed Me Me Me Me Me Me D 3 Me Me Me Me Me Me Me Me D 4

Challenge nce the decomposition temperature for PDMS is 343 C, it is a challenge to make a column that is stable and has low bleed at 380 C.

rigin of Bleed Polymer Synthesis ligomers that are created in a column s lifetime

Polymer Synthesis rigin of Bleed... Adapted from Inorganic Polymers, Mark, Allcock, & West 1992.

Polymer Synthesis licone Polymers can be synthesized using well known catalysts (KH, HCl) under equilibration polymerization conditions. We use newer, more advanced catalysts, better synthetic techniques.

rigin of Bleed... Polymer Synthesis ligomers that are created in a column s lifetime

rigin of Bleed... ligomers that are created in a column s lifetime (FS tubing) [ ] (polymer coating) D 3 /D 4 bleed PDMS H H

Enhancement of Thermal Stability by Using Additives lphenylene CH3 CH3 CH3 x y z CH3 CH3 CH3

Additives... n m A column with true ne selectivity cannot be obtained by having silphenylene in the backbone.

Stx TM -1 HT The polymer or the phase is PDMS. n

rigin of Bleed... deactivation fused silica

rigin of Bleed... deactivation fused silica

Deactivation of the Fused lica Surface We are using an advanced deactivation called ltek. It is a deposition process, unlike silazane or silicone deactivation.

Aroclor 1242 8 16 24

Aroclor 1254 8 16 24

Aroclor 1260 8 16 24

Aroclor Run Conditions Column: Stx-HT1 Serial #: 215041 Description: 30m, 0.32mmID, 0.10µm Sample: Aroclor Solutions Part #32009, 32011, 32012 Concentration: Solvent: Sample ze: Instrument: Injector: 400ppb Hexane 1.0µL HP5890 Splitless/275 C Splitless Hold Time: 1.0min. SplitVentFlow: 40cc/min. Septa Purge: 5cc/min. Carrier Gas: Helium Head Pressure: 8.5psi Column Flow Rate: 1.9cc/min. Linear Velocity: 31cm/sec. Detector: ECD/310C Make up Gas Flow: 40cc/min. Temp. Program: 75 C(1.0min.) 15 C/min. to 150 C(0) 5 C/min. 300 C(10min.)

PCB Congeners Column: Stx-HT1 Serial #: 215041 Description: 30m, 0.32mmID, 0.10µm Sample: PCB Congener Standard #2 Part #32294 Concentration: 100ng Solvent: Iso-ctane Sample ze: 1.0µL 1 3 4 5 6 7 Elution rder: 1. BZ#28 2. BZ#52 3. BZ#101 4. BZ#118 5. BZ#153 6. BZ#138 7. BZ#180 Instrument: Injector: Splitless Hold Time: SplitVentFlow: Septa Purge: Carrier Gas: Head Pressure: Column Flow Rate: Linear Velocity: Detector: Make up Gas Flow: HP5890 Splitless/275 C 0.80min. 40cc/min. 5cc/min. Helium 8.5psi 1.9cc/min. 31cm/sec. ECD/310 C 40cc/min. 2 *PCBs numbered according to Ballschmiter Temp. Program: 100 C(0.80min.) 15 C/min. to 325 C(10min.) 4 8 12 16

10ppm Aliphatics 1 2 3 4 1. C24 2. C26 3. C28 4. C30 5. C36 5 30m, 3.32mmID, 0.01µm Stx -1HT (cat.# 11709) Solvent: Hexane ven temp.: 30 C (hold 0.5 min.) to 240 C at 15 C/min. to 380 C at 10 C/min. (hold10min.) Det.: FID at 340 C Inj.: splitless at 250 C Carrier: hydrogen Flow rate: 4.0cc/min. at 30 C Make-up gas flow: 40cc/min. 16 20 24 28

Bleed Comparison 15.61pA 39.02pA 11.27pA 6.07pA Rtx -1MS 6.94pA Stx -1HT 4.34pA 340 C 360 C 380 C

1 50ppm Hydrocarbons 2 1 3 2 4 3 4 5 5 6 7 6 Aliphatic hydrocarbons: 1. C9 2. C10 3. C12 4. C14 5. C16 7 8 8 9 9 6. C18 7. C19 8. C20 9. C22 10. C24 11. C26 12. C28 13. C30 10 11 12 Aromatic Hydrocarbons: 1. napthalene 2. 2-methylnaphthalene 3. acenaphthylene 4. acenaphthene 5. fluorene 6. phenanthrene 7. anthracene 8. fluoranthene 9. pyrene 10. benzo(a)anthracene 11. chrysene 12. benzo(b)fluoranthene 13. benzo(k)fluoranthene 14. benzo(a)pyrene 15. indeno(1,2,3-cd)pyrene 16. dibenzo(a.h)anthracene 17. benzo(g,h)perylene 11 10 12 13 14 15 13 16 17 4 8 12 16

Analysis Conditions 30m, 3.32mmID, 0.01µm Stx -1HT (cat.# 11709) ven temp.: 30 C (hold 0.5 min.) to 240 C at 15 C/min. to 380 C at 10 C/min. (hold10min.) Det.: FID at 340 C Inj.: splitless at 250 C Carrier: hydrogen Flow rate: 4.0cc/min. at 30 C Make-up gas flow: 40cc/min. 1.0µL injections of MA EPH Aliphatic Hydrocarbon standard (cat.# 31459, solvent: hexane) & MA EPH Aromatic Hydrocarbon standard (cat.# 31458, solvent: methylene chloride) @ 50ppm concentration.

4 5 ASTM D2887 1 2 3 6 7 8 9 30m, 3.32mmID, 0.01µm Stx -1HT (cat.# 11709) ven temp.: 30 C to 380 C at 10 C/min. Det.: FID at 380 C Inj.:coolon-column Carrier: helium Flow rate: 1.3cc/min. at 30 C Linear velocity: 22cm/sec. at 30 C Make-up gas flow: 45cc/min. Head pressure: 6.0psi 10 11 1. C7 2. C8 3. C9 4. C10 5. C11 6. C12 7. C14 8. C16 9. C18 10. C20 11. C24 12. C28 13. C32 14. C36 15. C40 16. C44 17. C46 18. C48 19. C50 20. C52 21. C54 22. C56 23. C58 12 13 14 15 16 17 18 19 20 21 22 23

XTI Mix on Rtx-1 & Stx-1HT 1 1 3 2 2 3 4 6 5 7 4 6 7 5 8 9 8 9 10 11 12 10 11 12 13 13 14 14 Stx -1HT Rtx -1 1. 1,2-hexanediol 2. nitro-di-n-propylamine 3. benzoic acid 4. C14 5. 2,4-dinitrophenol 6. nitrophenol 7. nitroaniline 8. pentachlorophenol 9. carbazole 10. C20 11. C21 12. C22 13. benzo(b)fluoranthene 14. benzo(k)fluoranthene

Summary Definitions The design of low bleed & stable column system was discussed Several applications were shown