Automated Large-Volume Injection for Direct Analysis of Alkylphenols and Ethoxylated Alkylphenols by Mixed-Mode Chromatography and Mass Spectrometry

Similar documents
Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Determination of urea in ultrapure water by IC-MS/MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

New Developments in Surfactant Analysis Using HPLC

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

columns Acclaim Mixed-Mode WCX-1 for Separating Basic Molecules

Monitoring Protein PEGylation with Ion Exchange Chromatography

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Determination of Additives in Carbonated Beverages

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

Enhancement of Linearity and Response in Charged Aerosol Detection

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Supporting Information

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

columns IonPac AS22 Anion-Exchange Column

columns IonPac AS17-C Anion Exchange Column

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

columns IonPac SCS 1 Silica Cation Separator

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

A S ENSIT IV E M E T HO D FO R T H E D E T E RM INAT IO N O F ENDO C RIN E- D IS RU P T ING COM P OUNDS IN RIV E R WAT E R BY L C / MS/MS

Advances in Sample Preparation for Accelerated Solvent Extraction

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Determination of Carbonyl Compounds in Workplace Air

Technical Note. Introduction

Supporting Information. Detection and Occurrence of Chlorinated By-products of Bisphenol A, Nonylphenol and

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Determination of Methylamine in Drug Products

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

LC-MS/MS Analysis of Phytocannabinoids and their

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

Transcription:

Automated Large-Volume Injection for Direct Analysis of Alkylphenols and Ethoxylated Alkylphenols by Mixed-Mode Chromatography and Mass Spectrometry Leo Wang, Bill Schnute, Stacy Henday, Dionex Corporation, Sunnyvale, CA, USA INTRODUCTION Alkylphenyl polyethoxylates (APEOs) are among the most commonly used nonionic surfactants, with over 6, tons produced annually. APEO degradation products alkylphenols (APs) and their short chain ethoxylates (mainly mono- and diethoxylates, EO and EO2) have been detected and reported in treated wastewater, sewage sludge, and soil. Studies reveal that these substances have varying estrogenic potencies. 2 Currently, these compounds are on the US E s priority testing list for evaluation of environmental and health effects. Because APEOs with different EO units exhibit different estrogenic potencies, it is important to develop an analytical method to profile and monitor their presence in the environment. The development of a chromatographic method for separation of APs and APEOs remains a challenge due to their structural complexity. Current methods, including normal-phase liquid chromatography (NPLC) and reversed-phase liquid chromatography (RPLC) separate these compounds by the difference of alkyl groups or EO units. 9 The preferred detection technique for APEO analysis is mass spectrometry, due to its sensitivity and selectivity. Here, a mixed-mode chromatographic method was developed for separation of these compounds. On-line concentration and desalting was fully automated using large-volume injection ( ml) to increase sample throughput and improve sensitivity. Three mixed-mode columns with different selectivities were evaluated and the results compared, as well as MS and MS/MS instrumentation modes as detection techniques. Parameters affecting chromatographic separation and MS detection were explored and discussed, with each method individually evaluated with respect to linearity, detection limits, precision, and accuracy. EXPERIMENTAL Instrumentation HPLC: P68 Dual Ternary Pump ASI- Autosampler or AS-HV autosampler TCC- Thermostatted Column Compartment UVDU detector Mass Spectrometer: MSQ Plus single quadrupole mass spectrometer with cone wash TSQ Quantum Access triple quadrupole mass spectrometer Software: Chromeleon 6.8 SR6 Xcalibur 2. DCMS Link for Xcalibur (version 2.)* for MS/MS analysis *DCMS Link is a Chromeleon-based software module providing the interface for controlling Dionex chromatography instruments from different mass spectrometer software platforms. Sample Loading Conditions for AS-HV* Concentrator: Acclaim Guard (. mm, µm) Injection Volume: ml Loading Speed: ml/min * For ASI- autosampler, use loop injection and µl injection volume. Chromatographic Conditions Analytical Column: Acclaim Surfactant (2. mm, µm) Column Temperature: C Mobile Phase: A: Methanol (CH OH); B: Buffer ( mm sodium acetate (NaOAc) for MSQ Plus, mm ammonium acetate (NH OAc) for Quantum TSQ); C: water. Time (min) A% B% C% 7 2 9 7 2 9 8 9 9 7 2 Flow Rate:.2 ml/min

Mass Spectrometric Conditions MSQ Plus Interface: Electrospray Ionization (ESI) Detection Mode: Selected Ion Monitoring (SIM), see Table for scan details Nebulizer Gas: Nitrogen; 8 psi Probe Temperature: C Cone Wash: CH CN/H 2 O (/, v/v) at µl/min Quantum TSQ Interface: ESI Detection Mode: Selected Reaction Monitoring (SRM), see Table 2 for scan details Capillary Temperature: 27 C Needle Voltage: V Sheath Gas: units Auxiliary Gas: for the first segment, for the second segment Name Table. Scan Functions and Scan Events for MSQ Plus SIM Detection Adduct Mass m/z Span m/z SIM Group Time Range (min) Dwell Time (sec) Polarity Cone Voltage (V) [M+Na] + 27.2.. 8.. Pos. 6 [M+Na] + 7.2.. 8.. Pos. 6 SIM Group 2 [M+Na] + 287.2. 8. 2.. Pos. 6 NPEO2 [M+Na] +.2. 8. 2.. Pos. 6 SIM Group OP, tert-op [M+CH COO] 26.2.. 9.. Neg. 6 NP [M+CH COO] 279.2.. 9.. Neg. 6 Scan Segment ( ~ 2 min) 2 (2 ~ 2 min) Table 2. SRM Scan Functions for MS/MS Detection Analyte * Quantitative SRM ; ** Confirmative SRM. Polarity SRM Transitions Parent Ion (m/z) Product Ion (m/z) Collision Energy (V) (Q*) positive 268.2. (C**) positive 268.2 7. 7 (Q) positive 2.2 8. (C) positive 2.2 2. 22 (Q) positive 282.2 27. 7 (C) positive 282.2 8.2 NPEO2 (Q) positive 26.2 8. NPEO2 (C) positive 26.2 2. 22 Standard Preparation tert-op negative 2.. 2 -OP negative 2. 6. 2 -NP negative 29.2 6. 2 IGEL CA-2 and IGEL CO-2 were purchased from Sigma- Aldrich.,, and NPEO2 were chromatographically fractionated in our laboratory using an Acclaim WAX- column. Concentrations of collected fractions were calculated based on relative UV peak areas. N-octylphenol, tert-octylphenol, and n-nonylphenol were purchased from Sigma-Aldrich and dissolved in CH CN to µg/ml. Structures are shown in Figure. C 8 H 7 OH 2 CH 2 OH Octylphenol monoethoxylate m.w. = 2. C 9 H 9 OCH 2 CH 2 OH Nonylphenol monoethoxylate m.w. = 26. C 8 H 7 Octylphenol diethoxylate m.w. = 29. C 9 H 9 (OH 2 CH 2 ) 2 OH Nonylphenol diethoxylate m.w. = 8. (OCH 2 CH 2 ) 2 OH H C(H 2 C) 6 H 2 C OH H C(H 2 C) 7 H 2 C OH -OP -Octylphenol m.w. = 26. -OP -Nonylphenol m.w. = 22. OH t-bu tert-op tert-octylphenol m.w. = 26. 289 Figure. Structures of studied analytes. 2 Automated Large-Volume Injection for Direct Analysis of Alkylphenols and Ethoxylated Alkylphenols by Mixed-Mode Chromatography and Mass Spectrometry

Calibration standards were prepared in mobile phase the following concentration levels: 2,,, 2,,, 2,, and ng/ml. Each level contains seven analytes:,,, NPEO2, -OP, tert-op, and -NP. RESULT AND DISCUSSION Chromatography Among the three column candidates, the Acclaim Surfactant and Acclaim Mixed-Mode WAX- possessed the capability of separating APEO EO short oligomers. To achieve the goal of total resolution of APEO EO oligomers and APEO groups, i.e., OPEOs and NPEOs, these columns were compared with respect to their group separation efficiency. As shown in Figure 2, the Acclaim Surfactant column demonstrated the ability to differentiate APEOs by the EO units as well as alkyl group difference. mau Columns: Acclaim Surfactant and Acclaim WAX- (2. mm, µm) Mobile Phases: For Surfactant: 8% CH CN % mm NH OAc ph.2 9% H 2 O For WAX-: % CH CN % mm NH OAc ph.2 % H 2 O NPEOs on WAX- OPEOs on WAX- Column Temp.: C Flow Rate:. ml/min Inj. Volume: µl Detection: UV @ 27 nm Samples: IGEL CA-2 IGEL CO-2 NPEOs on Surfactant OPEOs on Surfactant 6 9 2 Minutes Figure 2. Resolution comparison of WAX- and Surfactant columns for APEO groups. 289 Column temperature was a key parameter for fine tuning resolution and adjusting run time. Seven temperatures from 2 C were examined to determine optimum conditions (acetonitrile was used in the mobile phase). Results are shown in Table. Three resolutions (Rs) were measured to evaluate overall method resolution: Rs represents the resolution for EO oligomers; Rs represents the resolution for oligomer groups; Rs measures the separation of APEOs and AP. Oligomers with the same alkyl chain but different EO units were better resolved using low temperatures. Temperature ( C) Table. Effects of Column Operating Temperature Chromatography Time* (min) Resolution Resolution Resolution 2 9. 2.7 2.82 2. 2 7.8 2.6 2.92 2.2 6. 2..2 2..7 2..6.9. 2.26.8.7 2. 2.2.9.. 2.6.9. * Minimum time required to elute last peak (-NP). Mobile phase: 2% CH CN, % mm NH OAc, % H 2 O; flow rate:. ml/min Temperature dependence for the separation of APs and APEOs followed the same trend, with lower temperature yielding better resolution. However, separation of oligomer groups (OPEOs and NPEOs) showed the opposite trend: OPEOs and NPEOs were better resolved with elevated temperature. Column temperature was examined further using methanol as an organic modifier for MS detection and C was selected as the optimum condition to achieve total separation. A mobile phase with higher organic content is favorable for MS sensitivity. The methanol-containing mobile phase showed significantly improved MS response for APEO. Therefore, methanol was used for HPLC-MS analysis. Method Performance with Single Quadrupole MS Instrumentation Mass Spectrometry with SIM Detection When coupling MS detection to chromatography, ESI is the preferred ionization interface for APEO analysis. ESI showed better sensitivity and specificity for a wider range of oligomers. To improve detection limits using MS SIM detection,. mm sodium acetate was added to the mobile phase to form sodium adducts. Because sodium acetate is nonvolatile, salt condensation was observed after operating for a short period of time. To ensure the long term stability of MS detection, the optional cone wash feature was activated with a wash stream of µl/min methanol/water (/, v/v). Significant improvement of system stability was observed when employing the cone wash (.% decrease in intensity with cone wash compared with 6% decrease without cone wash after 6 runs.)

Linearity and Method Detection Limits The linearity of this method was evaluated by three replicate injections of calibration standards. The method detection limit (MDL) for each analyte was calculated by the equation: MDL = t 99% S (n=7) Where t is Student s t at 99% confidence intervals (t 99%, n=7 =.) and S is the standard deviation. The standard deviation was obtained from seven replicate injections of a calibration standard at ppb for OPEOs and ppb for NPEO2 and APs (a calibration standard at ppb was used to obtain the standard deviation for due to its low response). The results for linearity and MDLs are summarized in Table. When using adducts (sodium adducts for APEOs and acetate adducts for phenols) as quantitative ions, linear responses were observed in the range from low ppb to 2 ppb with correlation coefficients greater than.99. The calculated MDLs vary from 7. () to 7.6 ppb (). Precision and Accuracy Method precision and accuracy were calculated from seven replicated injections of standards at different concentrations based on their MS response. Precision is shown by RSD from the replicate analyses, and accuracy was calculated by (observed amount)/(specified amount) %. Method precision for each analyte is less than % with the highest bias of 9.2% for (Table ). Accuracy measurements were within % bias for each analyte except for which was 8% higher than the specified amount. Table. Linearity, Precision, Accuracy, and Method Detection Limits Using MS SIM Analyte MS Adduct Linearity From To R 2 Precision (RSD, n = 7) Accuracy (%) MDL (pg) [M+Na] + 2.999 9.2 96.9 29 a [M+Na] +.999.8 9.2 82 a [M+Na] + 2 2.996 2. 8 6 b NPEO2 [M+Na] + 2.99 2.2 96.9 2 c tert-op [M+CH COO] 2.998. 9. 222 c -OP [M+CH COO] 2.99. 97.2 282 c -NP [M+CH COO] 2.997.6 22 c a Calculated from 7 replicate injections of a ppb standard. b Calculated from 7 replicate injections of a ppb standard. c Calculated from 7 replicate injections of a ppb standard. Method Performance with Triple Quadrupole MS/MS Instrumentation Mass Spectrometry with SRM Detection An ESI interface and methanol-containing mobile phase were also used for MS/MS SRM detection. NH OAc was used to buffer the mobile phase instead of NaOAc because analyte-sodium adducts did not show strong fragment ions with SRM conditions. Two SRM transitions were used for quantification and confirmation. The SRM scan details are shown in Table 2. Calibration Range and Method Detection Limits Coefficient of determination (R 2 ) measured greater than.99 for each analyte from low calibration levels (2 ppb) to 2 ppb. MDLs were calculated by the same means as for MS SIM detection, and details are shown in Table. Lower detection limits can be achieved using MS/MS SRM compared with MS SIM detection. The SRM chromatograms of APs and APEOs are shown in Figure with the injection amount of 2 ng of each analyte. LC Conditions Column: Acclaim Surfactant (2., µm) Column Temp.: C Flow Rate:.2 ml/min Injection: 2. ng in µl Mobile Phase: A: CH OH; B: NH OAc; C: H 2 O Relative Abundance Gradient: Time (min) A B C 7 7 2 7 2 9 7 2 9 8 9 2 7 2 RT: 6.7 S/N 77 RMS MS Conditions MS Detector: TSQ Quantum Access Ionization: Electrospray Ionization (ESI) Detection Mode: Polarity Switching SRM Needle Voltage: V Capillary Temp.: 27 C Sheath Gas: Arb. Auxiliary Gas: / Arb. (segment / segment 2) Ion Sweep Gas: See Table 2 for SRM Scan Segments and Events NL: 6.89E +SRM: 268 A RT:.6 S/N 77 RMS NL: 2.9E6 +SRM: 2 8 B NL: 2.8E RT: 7.97 S/N 9 RMS +SRM: 282 27 C NL:.2E6 RT: 7. S/N 79 RMS NPEO2 +SRM: 26 8 D NL:.8E2 RT: 8.77 S/N 27 RMS tert-octylphenol SRM: 2 E NL: 7.E RT:.77 S/N 2 RMS -Octylphenol SRM: 2 6 F NL: 8.87E RT: 6.67 S/N 298 RMS -Nonylphenol SRM: 29 6 G 2 Minutes Figure. SRM Chromatograms of alkylphenols and alkylphenol ethoxylates. Reconstructed SRM chromatograms A G show APs and APEOs with 2 ng injection. Automated Large-Volume Injection for Direct Analysis of Alkylphenols and Ethoxylated Alkylphenols by Mixed-Mode Chromatography and Mass Spectrometry

Table. Calibration Range and Coefficient of Determination Using MS/MS SRM Analyte SRM Calibration Range NPEO2 From To R 2 Fitting Function Weighting 268 2.9972 Linear /X 268 7 2.996 Linear /X 2 8 2 2.9976 Quadratic /X 2 2 2 2.9976 Quadratic /X 282 27 2.996 Linear /X 282 8 2.996 Linear /X 26 8 2 2.9977 Quadratic /X 26 2 2 2.9976 Quadratic /X tert-op 2 2.997 Linear /X -OP 2 6 2.999 Linear /X -NP 29 6 2.998 Linear /X Precision and Accuracy Method precision and accuracy were measured at three levels:, 2, and 2 ppb. Calculations were performed by the same means as for MSQ SIM detection, and results are shown in Table 6. Accuracy varies from 86% to %, and RSDs range from.% to 9.%. Analyte Table 6. Precision and Accuracy Using MS/MS SRM Detection NPEO2 tert-op -OP -NP Level Mean Accuracy (%) RSD ND N/A N/A 2 2. 2 899 9..2 8.9 89. 2.7 2 96 98.. 2 76 87.8.9 ND N/A N/A 2 2 2 6.9 2 82 92.6.2 9.88 98.8 9. 2 87 9. 2. 2 82 9.6 2.2 ND N/A N/A 2 22 6 8.9 2 2.2.6 ND N/A N/A 2 72 86. 8.8 2 978 98.9. ND N/A N/A 2 2 2..9 2 26.8.6 MDL (pg) 27 2 8.6 6. 62 27 Automation of Large Volume Injection for Direct Analysis of Water Samples To improve the method detection limits and to reduce intensive sample preparationin the laboratory, an automated large-volume injection method was employed for direct analysis of these analytes using the AS-HV autosampler. The system schematic is shown in Figure. The water sample is loaded into a ml sample loop by the AS-HV then delivered by the second system pump unit through a concentrator column (Acclaim,.6 mm, µm) using DI water as the mobile phase. The analytes trapped on the concentrator column were back-eluted and separated on the analytical column (Acclaim Surfactant, 2. mm, µm), and detected by UV and MS detectors. Automation was achieved through Chromeleon software. Detection limits were significantly improved by using large-volume injections. Direct analysis of water samples for APs and APEOs can be performed using this method. Method specificity was evaluated by analyzing ml treated DI water (organic residues were removed by passing DI water through an analytical column (.6 mm, µm) as a system blank. No measurable peaks were observed at the specific retention times for target analytes except for NPEO2 (2% peak area of NPEO2 at ppt). Carryover was examined by analyzing a system blank sample after injection of a calibration standard at the highest concentration, i.e., 2 ppt. No carryover peaks were observed except for NPEO2 whose peak area was 6% of NPEO2 in a blank sample. Therefore, carryover was deemed to be negligible when using large-volume injection for direct analysis. The coefficient of determinations (R 2 ) achieved were greater than.99 for each analyte from low levels (: ppt; : ppt; : 2 ppt; NPEO2: ppt; t-op: ppt; -OP: 2 ppt; -NP: ppt) to 2 ppt. MDLs were measured by the statistical method mentioned in prior sections with n 6. MDLs for APEOs were measured at low ppt levels (:. ppt; : 6.79 ppt; : 8.2 ppt; NPEO2:. ppt), tert-op at 78 ppt and -OP at ppt. (MDL for -NP was not measured due to low response.)

AXP Sample Sample 2 Wash solvent CH OH/H 2 O (/) Figure. System schematic. /8 - /6 Adapter Eluent Trap 6 Pump Left - Off 2 2 6 Peristaltic Pump /8 - /6 Adapter Eluent Trap 6 Pump Left - DI H 2 O 2 /8 - /6 Adapter Eluent Trap 6 2 Peristaltic Pump Peristaltic Pump Pump Left - Off 2 6 2 6 Surfactant Surfactant Surfactant Step UV Pump Right Eluent In Step 2 UV Pump Right Eluent In Step UV Pump Right Eluent In MS MS MS 2897 REFERENCES. Guenther, K.; Heinke, V.; Thiele, B.; Kleist, E.; Prast, H.; Raecker, T. Environ. Sci. Technol. 22, 6, 676 68. 2. Loos, R.; Hanke, G.; Umlauf, G.; Eisenreich, S. J. Chemosphere 27, 66, 69 699.. Rice, C. P.; Schmitz-Afonso, I.; Loyo-Rosales, J. E.; Link, E.; Thoma, R.; Fay, L.; Altfater, D.; Camp, M. J. Environ. Sci. Technol. 2, 7, 77 7.. Nimrod, A. C.; Benson, W. H. Crit. Rev. Toxicol. 996, 26, 6.. Shang, D. Y.; Ikonomou, M. G.; Macdonald, R. W. J. Chromatogr. A 999, 89, 67 82. 6. Ferguson, P. L.; Iden, C. R.; Brownawell, B. J. Anal. Chem. 2, 72, 22. 7. Ferguson, P. L.; Iden, C. R.; Brownawell, B. J. J. Chromatogr. A 2, 98, 79 9. 8. Loyo-Rosales, J. E.; Schmitz-Afonso, I.; Rice, C. P.; Torrents, A. Anal. Chem. 2, 7, 8 87. 9. Andreu, V.; Ferrer, E.; Rubio, J. L.; Font, G.; Pico, Y. Sci. Total Environ. 27, 78, 2 29.. Wang, J.; Henday, S.; Schnute, W. Profiling Analysis of the Degradation Products of Alkylphenol Polyethoxylates by LC-MS Using an Acclaim Surfactant Column with Mass Spectrometric Detection. Presented at the 6th ASMS Conference, 28.. Petrovic, M.; Diaz, A.; Ventura, F.; Barcelo, D. Anal. Chem. 2, 7, 886 89. CONCLUSION This study produced methods for profiling degradation products of ethoxylated alkylphenols in water samples. Using a large-volume injection, automated direct analysis of APs and APEOs was demonstrated without labor-intensive sample preparation. Suppliers Sigma-Aldrich (St. Louis, MO, USA). Total separation of seven target analytes was achieved and selective, sensitive detection was achieved using mass spectrometric detection. Detection limits and specificity were improved over MS SIM using MS/MS SRM detection, and additional confirmation information can be obtained by introducing a second SRM transition. DCMS LINK is a trademark, and Acclaim and Chromeleon are registered trademarks of Dionex Corporation. MSQ Plus is a trademark and TSQ Quantum Access and Xcaliber are registered trademarks of Thermo Fisher Scientific, Inc. Passion. Power. Productivity. Dionex Corporation 228 Titan Way P.O. Box 6 Sunnyvale, CA North America U.S./Canada (87) 29-7 South America Europe Austria () 66 2 Benelux () 2 68 9768 (2) 29 Denmark () 6 6 9 9 France () 9 Germany (9) 626 99 Ireland () 6 6 Italy (9) 2 62 267 Sweden (6) 8 7 8 6988-6 Automated Large-Volume Brazil () 7 Injection Switzerland for () Direct 62 2 9966 Analysis United Kingdom of () Alkylphenols 276 69722 and (8) 77-7 Ethoxylated Alkylphenols by Mixed-Mode Chromatography and Mass Spectrometry www.dionex.com Asia Pacific Australia (6) 2 92 2 China (82) 228 282 India (9) 22 276 27 Japan (8) 6 688 2 Korea (82) 2 26 28 Singapore (6) 6289 9 Taiwan (886) 2 87 66 LPN 22- /9 29 Dionex Corporation