Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing

Similar documents
Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Application Note. Abstract. Authors. Introduction. Food Safety

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Highly Sensitive and Rugged GC/MS/MS Tool

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Anaylsis of Pesticide Residues in Rice Using Agilent Bond Elut QuEChERS AOAC Kit by LC-MS/MS Detection

Organophosphorus Pesticides in Apple Matrix by GC/MS/FPD using an Agilent J&W DB-35ms Ultra Inert GC Column

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

GC/Q-TOF MS Surveillance of Pesticides in Food

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Agilent Bond Elut QuEChERS Food Safety Applications Notebook: Volume 2 INNOVATIVE APPROACHES FOR TODAY S FOOD ANALYSIS CHALLENGES

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Application Note. Authors. Abstract. Xin Sun Shandong Agriculture University, Taian, , China

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Rugged GC/MS/MS Pesticide Residue Analysis Fulfilling the USDA Pesticide Data Program (PDP) Requirements

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

The Analysis of Organophosphate Pesticides by LC/MS Application

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

An Optimal Method for the Analysis of Pesticides in a Variety of Matrices

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Reduce Cost of Pesticide Residue Analysis

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Analysis of Pesticide Residues in Apple using Agilent Bond Elut QuEChERS European Standard EN Kits by LC/MS/MS Detection

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Analysis of Pesticides and Contaminants using HS-SPME-GC/MS and DRS Detection. Blair Surridge, Tyler Dyer, Rene Bilideau, Paloma Bertolin Presoti

Use of Analyte Protectants in GC-Analysis a way to improve peak shape and reduce decomposition of susceptible compounds

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Screening for Hundreds of Pesticide Residues Using a GC/Q-TOF with an Exact Mass Pesticide Database in Food

Validation of a Simple and Rapid Multiresidue Method (QuEChERS) and its Implementation in Routine Pesticide Analysis

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Analysis of Pesticide Residues in Apple Using Agilent Bond Elut QuEChERS AOAC Kits by GC/MS

Fully automated QuEChERS clean-up and LC/MS-QQQ analysis of pesticides in fruits and vegetables.

Understanding Gas Chromatography

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

ENVIRONMENTAL analysis

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

STANDARD OPERATING PROCEDURES

The Pulsed Flame Photometric Detector for analysis of phosphorus pesticides in wastewater and food commodities

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Sulfotepp impurities in Chlorpyrifos EC formulations

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Application Note. Abstract. Authors. Introduction

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Thermo Scientific HyperSep Dispersive SPE Products. Efficient sample preparation and clean-up using the QuEChERS Method

QuEChERS Pesticide Analysis for Fresh Produce*

Evaluation of common organic solvents for gas chromatographic analysis and stability of multiclass pesticide residues

Routine Multiresidue Pesticide Analysis using the Agilent 6470 Triple Quadrupole Mass Spectrometer

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Pure Chromatography. Food Safety Applications. Abstract. Introduction

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Author. Abstract. Introduction

Analysis of BTEX in Natural Water with SPME

Validation Report 18

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

Application Note. Abstract. Authors. Environmental

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Transcription:

Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing Agilent Ultimate Plus Deactivated Fused Silica Tubing for Guard Columns Application Note Food Safety Authors Yun Zou and Andy Zhai Agilent Technologies (Shanghai) Co. Ltd Abstract Agilent Ultimate Plus deactivated fused silica tubing was evaluated as a guard column and compared with an equivalent and popular guard column from another vendor to analyze tomato samples extracted with QuEChERS. Ultimate Plus deactivated fused silica tubing can provide better linearity, repeatability, and stability. The Agilent deactivated FS tubing was superior to another vendor s tubing for the analysis of active and difficult pesticides in tomatoes. Introduction Guard columns are widely used in GC and GC/MS applications to protect the analytical column from contamination. When the guard column is a short piece of uncoated, deactivated fused silica tubing, the use of a guard column is also an inexpensive technique to extend the lifetime of capillary columns. Since contamination is limited to the front of the column, trimming the guard column periodically to restore performance, instead of the capillary column, preserves the main column. Thus, chromatography, including retention time and resolution, is not affected.

Due to complex matrices, multiresidue analysis of pesticides in fruits, vegetables, and other foods is always a challenge for sample preparation and detection. The QuEChERS method for pesticide analysis was introduced by USDA scientists in 2003 [1], and has been applied widely because of its Quick, Easy, Cheap, Effective, Rugged and Safe features. Agilent Bond Elut QuEChERS kits have demonstrated excellent recoveries for frequently used pesticides in tomato and other matrices [2,3]. However, food extracts processed by QuEChERS are still complicated, containing impurities such as high-boiling compounds that can cause contamination of the analytical column. Many pesticides are active analytes, and contamination of the analytical column leads to poor peak shape, loss of response, and shorter column lifetime. High inertness performance to minimize analyte degradation and peak tailing is required. To ensure accurate and reproducible results, using deactivated fused silica tubing as a guard column to protect the analytical column plays a key role in an inert flow path. In this application note, tomato was selected because it is a high-consumption fruit in many cultures, but also because it is purported to have many health benefits derived from its lycopene content, which helps to decrease oxidative stress. Agilent Ultimate Plus deactivated fused silica tubing has shown excellent performance as a GC restrictor in the analysis of pesticide checkout mixtures [4], and as a guard column in an endrin/ddt breakdown test [5]. A representative group of difficult pesticides, including organophosphates (OPs), organochlorines (OCs), carbamates, and pyrethroids, were spiked in tomato matrix blank samples and extracted with Bond Elut QuEChERS kits. The matrix-spiked standards were then analyzed by GC/MS with an Agilent J&W HP-5ms Ultra Inert Column connected with Agilent Ultimate Plus deactivated fused silica tubing as a guard column. In addition, tests were also performed on tubing from a different vendor for comparison under the same GC/MS conditions. Experimental Chemicals and reagents All reagents and solvents were HPLC or analytical grade. Acetonitrile (ACN) was from J&K Scientific (Beijing, China). Toluene was from ANPEL Scientific Instrument Co. Ltd (Shanghai, China). Water was from J. T. Baker. Pesticide standards and the internal standard (triphenyl phosphate, TPP) were purchased from Ultra Scientific (North Kingstown, RI, USA) and J&K Scientific. Matrix blank preparation Organic tomatoes were bought from a local food market. The tomatoes were frozen, chopped, and then homogenized thoroughly. A 10.0 g (± 0.1 g) amount of homogenized sample was placed into a 50 ml centrifuge tube and prepared with QuEChERS, as shown in Figure 1. Extraction and cleanup were achieved using an Agilent Bond Elut QuEChERS EN extraction kit (p/n 5982-5650CH) and a Bond Elut QuEChERS dispersive kit (p/n 5982-0029). Weigh 10 g comminuted sample into a 50 ml centrifuge tube. Add 10 ml ACN, and shake by hand. Add Agilent Bond Elut QuEChERS EN extraction salts. Add two homogenizers, and shake vigorously for 20 seconds. Centrifuge at 4,000 rpm for five minutes. Transfer 6 ml of the upper layer to a 15 ml dispersive SPE tube. Vortex for one minute, then centrifuge at 4,000 rpm for five minutes. Transfer the upper layer to a sample vial. Figure 1. QuEChERS procedure to extract pesticide residues in tomatoes. 2

Solutions and standards Standard and internal standard stock solutions (100 µg/ml) were made in acetonitrile and stored at 18 C. Due to lower response of the pesticides in Group 1, including fipronil, chlorfenapyr, iprodione, cyfluthrin, fenvalerate, difenoconazole, deltamethrin, and azoxystrobin, the concentration of these compounds in pesticide working solutions was twice as high as in Group 2. The high-qc solution, which was 20 µg/ml for Group 1 and 10 µg/ml for Group 2 pesticides, was prepared in toluene. This solution was used to prepare calibration curves in the matrix blank extract by appropriate dilution. Internal standard solution was added to give a final concentration of 100 ng/ml. Instrumentation Instruments and conditions are shown in Table 1. Two guard columns with different serial numbers were purchased from each supplier to demonstrate performance of different manufacturing lots. Two guard columns from each serial number were tested to confirm reproducibility. All guard columns were connected to analytical columns using Agilent Ultimate Unions, and analyzed in the same manner. Table 2 lists the flow path consumable supplies. Table 1. Instrumental conditions. Analytical column Agilent J&W HP-5ms UI, 30 m 0.25 mm, 0.25 µm (p/n 19091S-433UI) Agilent Ultimate Plus deactivated FS tubing, 5 m 0.25 mm (p/n CP802505) Guard column Deactivated FS tubing, 5 m 0.25 mm, from supplier R GC Agilent 7890B GC Autosampler Agilent 7683B Autosampler and sample tray, 5 µl syringe (p/n G4513-80213), 2 µl injection volume Carrier gas Helium, constant flow mode Split/splitless, 250 C, pulsed splitless, 25 psi pulse pressure for 0.75 min Inlet 50 ml/min purge flow at 0.75 min RT locking Chlorpyrifos-methyl locked to 13.443 min Oven 50 C (1 min), 25 C/min to 125 C (0 min), 10 C/min to 300 C (10 min) MSD Agilent 5977A MSD Solvent delay 4 min MS temperature 300 C (source), 150 C (quad) Transfer line 280 C MS EI, SIM Other parameters see Table 3 Table 2. Flow path supplies. Vials Amber, write-on spot, certified, 2 ml, screw top vial packs (p/n 5182-0554) Vial Inserts 150 µl glass with polymer feet (p/n 5183-2088) Septa Nonstick BTO septa (p/n 5183-4757) Column nut Self-tightening, inlet/detector (p/n 5190-6194) Ferrules 15% graphite: 85% Vespel, short, 0.4 mm id, for 0.1 to 0.25 mm columns (10/pk, p/n 5181-3323) UltiMetal Plus Flexible Metal, 0.4 mm id, for 0.1 to 0.25 mm fused silica tubing (10/pk, p/n G3188-27501) Union Agilent Inert Ultimate union (p/n G3182-60581) Liner Agilent Ultra Inert deactivated single taper splitless liner with wool (p/n 5190-2293) Inlet seal Ultra Inert, gold-plated, with washer (p/n 5190-6144) Internal nut CFT capillary fitting (p/n G2855-20530) 3

Table 3. Pesticides, CAS number, target ion, and retention time. No. Compound CAS no. Target ion RT No. Compound CAS no. Target ion RT ISTD Triphenyl phosphate 115-86-6 326 17.997 35 Parathion 56 38 2 291 14.401 1 Methamidophos 10265 92 6 94 6.412 36 Triadimefon 43121 43 3 57 14.439 2 DDV 62-73-7 109 6.619 37 Dicofol 115-32-2 139 14.441 3 Mevinphos 298-01-1 127 8.474 38 Isocarbophos 24353-61-5 136 14.519 4 Acephate 30560-19-1 136 8.481 39 Bromophos methyl 2104 96 3 331 14.738 5 Heptenophos 23560 59 0 124 10.119 40 Isofenphos-methyl 99675-03-3 199 14.889 6 Omethoate 1113 02 6 156 10.277 41 Pendimethalin 40487 42 1 252 15.008 7 Propoxur 114 26 1 110 10.477 42 Fipronil 120068-37-3 367 15.165 8 Ethoprophos 13194 48 4 158 10.729 43 Quinalphos 13593 03 8 146 15.232 9 Cadusafos 95465 99 9 159 11.300 44 Methidathion 950 37 8 145 15.511 10 Phorate 298 02 2 75 11.391 45 Tetrachlorvinphos 22248 79 9 329 15.685 11 alpha BHC 319 84 6 181 11.511 46 Fenamiphos 22224 92 6 303 15.869 12 Dimethoate 60 51 5 87 11.749 47 Profenofos 41198 08 7 208 16.095 13 Carbofuran 1563 66 2 164 11.872 48 Chlorfenapyr 122453-73-0 59 16.633 14 Atrazine 1912 24 9 200 11.930 49 4,4' DDD 72 54 8 235 16.951 15 beta BHC 319 85 7 219 12.041 50 Ethion 563 12 2 231 17.030 16 gamma BHC 58 89 9 181 12.178 51 Triazophos 24017 47 8 161 17.250 17 Quintozene 82-68-8 237 12.281 52 4,4' DDT 50 29 3 235 17.638 18 Fonofos 944 22 9 109 12.339 53 Iprodione 36734 19 7 187 18.326 19 Diazinon 333 41 5 179 12.480 54 Phosmet 732-11-6 160 18.518 20 Disulfoton 298 04 4 88 12.591 55 Phosalone 2310 17 0 182 19.212 21 Tefluthrin 79538 32 2 177 12.677 56 lambda Cyhalothrin 91465 08 6 181 19.527 22 Chlorothalonil 1897 45 6 266 12.744 57 Cyfluthrin I 68359 37 5 163 20.825 23 Pirimicarb 23103 98 2 166 12.999 58 Cyfluthrin II 68359 37 5 163 20.914 24 Phosphamidon 13171 21 6 127 13.243 59 Cyfluthrin III and IV 68359 37 5 163 21.003 25 Vinclozolin 50471-44-8 212 13.422 60 Cypermethrin 52315 07 8 181 21.145 26 Methyl parathion 298 00 0 263 13.435 61 Cypermethrin 52315 07 8 181 21.228 27 Chlorpyrifos methyl 5598 13 0 286 13.444 62 Cypermethrin 52315 07 8 181 21.331 28 Carbaryl 63 25 2 144 13.530 63 Fenvalerate 51630 58 1 167 22.070 29 Fenchlorphos 299 84 3 285 13.728 64 Fenvalerate 51630 58 1 167 22.236 30 Demeton S methyl sulfone 17040 19 6 169 13.799 65 Difenoconazole 119446 68 3 323 22.529 31 Fenitrothion 122 14 5 277 13.973 66 Difenoconazole 119446 68 3 323 22.614 32 Malathion 121 75 5 173 14.161 67 Indoxacarb 173584 44 6 150 22.814 33 Fenthion 55 38 9 278 14.339 68 Deltamethrin 52918 63 5 181 22.932 34 Chlorpyrifos 2921 88 2 197 14.385 69 Azoxystrobin 131860 33 8 344 23.256 4

Results and Discussion The purpose of the tests was to evaluate the performance of Agilent Ultimate Plus deactivated FS tubing as a guard column, and compare it with a popular deactivated guard column from another supplier for the analysis of pesticides in a vegetable matrix by GC/MS. The system was inspected and carefully cleaned, if necessary, before each test. For consistency, new UI columns, UI gold seals, liners, and inert unions were used for each tubing test. Protecting the analytical column One potential issue with the use of GC/MS for analysis of QuEChERS samples is contamination and deterioration of the GC column. QuEChERS vegetable samples usually still contain some impurities that can accumulate on the head of the column, causing peak tailing, retention time shifting, and reduced response. Figure 2 shows chromatograms of a blank tomato extract and a 50 ng/ml spiked QuEChERS sample analyzed with deactivated FS tubing from different suppliers. As shown in Figure 2C, interference peaks are found in the blank chromatogram. These interference peaks are either completely separated with the target pesticide s peak, or with significantly lower intensity compared to the target pesticide s peak. Therefore, these interference peaks do not affect integration and quantitation of the pesticides of interest. However, these impurities can also cause some chromatographic problems in that peak shape and intensity may deteriorate faster as more complicated samples are injected. Using Agilent Ultimate Plus deactivated FS tubing as a guard column is an effective way to protect the analytical column by trimming the tubing periodically instead of the analytical column, to restore performance. Backflushing is recommended for complicated sample matrices to reduce analysis time, frequency of MSD source cleaning, and column head trimming [6]. Backflushing was not used in this study to reduce influences on the test results. However, we recommend the use of Ultimate Plus deactivated FS tubing as a guard column and backflushing in routine pesticide analysis. A Agilent Ultimate Plus deactivated FS tubing 200,000 17 20 33 *ISTD 16 18 34,35 21 23 150,000 2 15 25,26,27 36,37 7 12 19 42 49 14 100,000 3,4 5 8 10 11 29 38 13 22 30 32 39 40 50 52 56 59 43 48 58 60 4445 54 57 61 51 50,000 9 28 55 1 6 24 31 41 46 47 53 62 0 6 8 10 12 14 16 18 20 22 B Supplier R guard column 200,000 17 20 33 16 21 23 *ISTD 34,35 150,000 15 18 25,26,27 36,37 49 2 7 12 14 19 42 52 59 100,000 29 38 43 56 3,4 5 8 10 22 28 30 32 39 40 48 11 13 58 60 44 45 50 54 57 61 51 50,000 1 6 9 24 31 46 47 53 55 41 62 0 6 8 10 12 14 16 18 20 22 200,000 150,000 100,000 50,000 C Blank tomato extract 0 6 8 10 12 14 16 18 20 22 Figure 2. GC/MS chromatogram of tomato extract using deactivated FS tubing as guard columns. A) Agilent Ultimate Plus deactivated FS tubing, B) supplier R guard column, and C) tomato extract blank (peak identification in Table 3). *ISTD 63 64 63 64 65 66 67 68 69 65 66 67 68 69 5

Important components in the flow path Flow path inertness plays a critical role in the accuracy, precision, durability, and consistency of pesticide analysis in complicated sample matrices. Each component of the flow path has the potential to contribute to overall system activity. Therefore, the Agilent Inert Flow Path solution, which provides excellent surface inertness for the entire GC flow, includes Ultra Inert columns, Ultra Inert inlet liners and gold seals, with UltiMetal Plus inert inlet, capillary flow technology (CFT) devices, and other inertness-verified consumables. As shown in Table 4, the surface areas of liners and gold seals are about 4.4 and 1 cm 2, respectively; while the surface area of a 5 m 0.25 mm guard column is about 39.3 cm 2. If surface area is proportional to the potential for active sites, this means that the guard column has over eight times more active sites than the liner, and 39 times more active sites than the gold seal in the system. Chromatographically active compounds such as organophosphate pesticides can adsorb onto active sites, particularly at trace levels, compromising an analyte s response. High inertness performance of deactivated fused silica tubing is very important for the entire GC flow path. Table 4. GC Flow path surface areas Performance comparison Pesticides of interest in this study were from various pesticides groups, such as organophosphates, organochlorines, carbamates, and pyrethroids. These compounds also included many difficult active pesticides such as methamidophos, acephate, omethoate, dimethoate, carbaryl, chlorothalonil, DDT, phosmet, and iprodione. Evaluation and comparison focused on the performance of the guard column for these active compounds and some pesticides for routine inspection. Due to the inert flow path, most analytes showed sharp and symmetrical peak shapes with Ultimate Plus deactivated FS tubing and supplier R guard columns (see Figure 2A and Figure 2B). The more polar pesticides can be problematic, often yielding broad peak shapes or excessive tailing that makes reliable quantitation at low levels difficult. Figure 3 depicts a GC/MS/SIM chromatogram of more problematic polar pesticides on different supplier s tubing. Compared to supplier R s guard column, the high level of inertness of Ultimate Plus deactivated FS tubing results in better peak shape and decreased sample adsorption, allowing lower detection limits. Length (cm) Diameter (cm) Surface area (cm 2 ) Liner 7 0.2 4.4 Seal 0.4 0.8 1 Column 3,000 0.025 235.6 Guard column 500 0.025 39.3 Abundance 140,000 130,000 120,000 110,000 100,000 90,000 80,000 70,000 60,000 50,000 40,000 30,000 20,000 10,000 0 Methamidophos DDV Acephate ion 136 Mevinphos ion 127 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 Time Figure 3. Enlarged section of GC/MS chromatogram of more problematic polar pesticides. Agilent Ultimate Plus deactivated FS tubing (blue), and supplier R guard column (red). 6

Linearity Linearity was determined using calibration curves spiked into the tomato matrix. Calibration curves were constructed from 2 µl injections of selected standards at 10, 20, 50, 100, 250, and 500 ng/ml. Every standard solution contained 100 ng/ml of internal standards. Figure 4 compares calibration curve coefficients (R 2 ) for Ultimate Plus deactivated FS tubing and the supplier R guard column. The supplier R guard column is widely recognized for its inertness performance. All pesticides showed excellent linearity with calibration coefficients greater than 0.990 on both Ultimate Plus deactivated FS tubing and the guard column from supplier R. Overall, however, Agilent deactivated FS tubing exhibited better performance, particularly for very difficult pesticides, such as methamidophos, acephate, and omethoate. Calibration curve coefficients (R 2 ) (10 500 ng/ml) in matrix blank 1.002 1.000 0.998 0.996 0.994 0.992 0.990 0.988 0.986 0.984 Methamidophos Dichlorvos Agilent Ultimate Plus deactivated FS tubing Supplier R guard column Mevinphos Acephate Omethoate Phorate alpha-bhc Dimethoate Carbofuran Atrazine Lindane Quintozene Diazinon Disulfoton Chlorothalonil Pirimicarb Phosphamidon Vinclozolin Methyl parathion Chlorpyriphos-methyl Carbaryl Fenitrothion Malathion fenthion Chlorpyriphos Parathion Triadimefon Dicofol Isocarbophos Bromophos-methyl Isofenphos-methyl Pendimethalin Fipronil Quinalphos Methidathion Fenamiphos Profenophos Chlorfenapyr Ethion Triazophos 4,4'-DDT Iprodione Phosmet phosalone Figure 4. Comparison of calibration curve coefficients (R 2 ) for Agilent Ultimate Plus deactivated FS tubing and supplier R guard column. 7

Repeatability and stability A 50 ng/ml spiked QuEChERS sample was analyzed over 15 injections to test repeatability and stability. The peak-area ratios of analytes/internal standards for most analytes were comparable between Ultimate Plus deactivated FS tubing and the supplier R guard column. However, for difficult active pesticides, better RSD values meant that the Agilent tubing provided more consistent responses of active pesticides, and thus supported more sample runs with acceptable results (Table 5). Conclusions Agilent Ultimate Plus deactivated fused silica tubing was evaluated for use as a guard column in the analysis of tomato QuEChERS samples, and compared with tubing from another supplier. Convincing proof for the surface inertness improvement of Ultimate Plus deactivated fused silica over another vendor s deactivated tubing includes better calibration curve linearity and longer durability, demonstrated by slower and reduced signal drop over multiple injections of critical active pesticides. This was visually apparent, and verified by RSD values. Ultimate Plus deactivated fused silica tubing can be used to improve GC flow path performance when used as guard columns for pesticide analysis in complicated matrices. Table 5. Injection repeatability and performance stability. Injection repeatability ( RSD%) n=15 No. Pesticide Supplier R guard column Agilent Ultimate Plus deactivated FS tubing 1 Methamidophos 6.69 5.86 2 Dichlorvos 2.25 2.14 3 Acephate 10.81 8.87 6 Omethoate 10.74 6.51 10 Phorate 1.36 1.15 12 Dimethoate 3.38 3.41 13 Carbofuran 3.73 3.15 14 Atrazine 1.18 1.26 16 Lindane 1.61 0.96 17 Quintozene 2.91 3.02 19 Diazinon 1.65 0.90 20 Disulfoton 1.59 1.66 22 Chlorothalonil 5.10 3.42 23 Pirimicarb 3.20 1.51 24 Phosphamidon 1.96 2.22 25 Vinclozolin 4.04 2.50 26 Methyl parathion 4.23 1.29 28 Carbaryl 7.42 6.17 31 Fenitrothion 2.28 1.67 32 Malathion 1.51 1.43 34 Chlorpyrifos 5.00 4.79 35 Parathion 5.04 2.78 36 Triadimefon 1.32 2.73 37 Dicofol 6.31 2.10 38 Isocarbophos 1.55 1.71 40 Isofenphos-methyl 1.37 1.34 41 Pendimethalin 3.48 1.52 42 Fipronil 3.48 1.89 43 Quinalphos 0.70 1.04 47 Profenofos 3.79 3.73 50 Ethion 2.83 1.24 51 Triazophos 2.28 1.21 52 4,4' DDT 1.25 0.65 53 Iprodione 2.88 2.46 54 Phosmet 3.37 1.98 55 Phosalone 3.98 2.36 8

References 1. Anastassiades, M.; Lehotay, S. J. Fast and Easy Multiresidue Method Employment Acetonitrile Extraction/Partitioning and Dispersive Solid-Phase Extraction for the Determination of Pesticide Residues in Produce. J. AOAC Int. 2003, 86, 412-431. 2. Zhao, Suli; Zhai, Andy. A Blind Study of Pesticides in vegetables by Agilent Bond Elut QuEChERS Extraction kits and Agilent 5975T LTM GC/MSD; Application note, Agilent Technologies, Inc. Publication number 5990-6323EN, 2012. 3. Zhao, Limian; Wylie, Philip L.; Stevens, Joan. Analysis of Pesticide Residues in Apple Using Agilent SampliQ QuEChERS EN Kits by GC/MS; Application note, Agilent Technologies, Inc. Publication number 5990-4073EN, 2012. 4. Dang, Ngoc A. Better Pesticide Analysis with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path; Application note, Agilent Technologies, Inc. Publication number 5991-5404EN, 2014. 5. Zou, Yun. Evaluation of Fused Silica Tubing for Active Compound Analysis in an Inert Flow Path; Application note, Agilent Technologies, Inc. Publication number 5991-5914EN, 2015. 6. Meng, C-K. Improving Productivity and Extending Column Life with Backflush; Application note, Agilent Technologies, Inc., Publication number 5989-6018EN, 2006. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 9

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2015 Printed in the USA June 9, 2015 5991-5974EN