Validation of GC-MS/MS for Detection and Confirmation of Low-Level Dioxins

Similar documents
GC/MS/MS Dioxin and Dioxin-like PCBs in Feed and Food. Jessica Westland, LSAG Applications Chemist

Technical Report. Abstract: Pu Wang 1, Huizhong Sun 1, Qinghua Zhang 1, Feifei Tian 2, Lei Cao 2

ApplicationNOTE ULTRA TRACE ANALYSIS OF DIOXINS AND FURANS IN HUMAN ADIPOSE TISSUE USING SFE-LC EXTRACTION/CLEANUP AND THE WATERS AUTOSPEC ULTIMA NT

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

SPME-GC-MS/MS for Identification and Quantification of Migration Contaminants in Paperboard Food Packaging

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

T H E ANA LYSIS O F DIOX INS AND F U R A NS USING H RG C- H IG H R E SO LU T IO N MS W IT H T H E AU T OS P EC- ULTIMA NT

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

The Truth About Dioxin Analysis: Using High Resolution GC-TOFMS and GCxGC-TOFMS to. Analytical Approaches

1.0 INTRODUCTION AND BACKGROUND

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

ApplicationNOTE THE ANALYSIS OF POLYCHLORINATED BIPHENYLS (PCBS) BY GC-HIGH RESOLUTION MASS SPECTROMETRY USING THE MICROMASS AUTOSPEC ULTIMA NT

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Chapter 3: Development of a PTV-LVI-GC-MS/MS method

The following revisions are being implemented for the PCDD/F Supplemental QAPP:

Official Journal of the European Union REGULATIONS

ApplicationNOTE A STUDY OF THE ANALYSIS OF POLYBROMINATED DIPHENYL ETHER FLAME RETARDANTS BY GC/MS/MS. Introduction. Materials and Methods

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

L 364/32 Official Journal of the European Union

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

of mass spectrometry

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Impurity Profiling of Pharmaceutical Starting Materials Using Gas Chromatography Coupled with High- Resolution Accurate Mass Spectrometry

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

HILIC Method Development in a Few Simple Steps

Frank L. Dorman The Pennsylvania State University Department of Biochemistry and Molecular Biology

CORUS R, D and T, Swinden Technology Centre, Rotherham, UK S60 3AR. HALL ANALYTICAL LABORATORIES, Manchester, UK M23 9YJ

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

PosterReprint INTRODUCTION

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

Dioxins & PCBs concerns

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Detection of Mycotoxins in Corn Meal Extract Using Automated Online Sample Preparation with Liquid Chromatography-Tandem Mass Spectrometry

APPLICATION NOTE Authors Jane Cooper, 1 Jordi Parera, 2 Esteban Abad, 2 Richard Law,¹ and Cristian Cojocariu 1

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Multi-residue analysis of pesticides by GC-HRMS

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

ONE STEP CLOSER TO A COOK BOOK METHOD FOR DIOXIN ANALYSIS PART 2: ANALYSIS OF FOOD SAMPLES

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Estimation of LOQ for the Analysis of Persistent Organic Pollutants, in particular PCDD/Fs and PCBs

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Analysis of Alkylphenols Using GC-MS/MS and Automated SRM Development

MetWorks Metabolite Identification Software

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Measurements of Household Dust Concentrations of PCDDs, PCDFs, and PCBs from a Community in Michigan, USA. Lynn Zwica, MS

Methods of Analysis of Polychlorinated Biphenyl Congeners Using an Application-Specific Capillary GC Column

Fast, Effective XPS Point Analysis of Metal Components

Monitoring Protein PEGylation with Ion Exchange Chromatography

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

(Non-legislative acts) REGULATIONS

Automated and accurate component detection using reference mass spectra

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

Highly Sensitive and Rugged GC/MS/MS Tool

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Screening Method for 30 Pesticides in Green Tea Extract Using Automated Online Sample Preparation with LC-MS/MS

The Use of Tandem Quadrupole GC/MS/MS for the Determination of Polycyclic Aromatics Hydrocarbons (PAHs) in Food Products

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Introduction to Fourier Transform Infrared Spectroscopy

Thermo Scientific. Anion-Exchange Column

for XPS surface analysis

Official Journal of the European Union

Orbitrap GC-MS: An Opportunity to Help Address the Challenges of Chlorinated Paraffins Analysis

Analysis of alkylphenols using GC-MS/MS and automated SRM development

Dioxin, PCB analysis in fishmeal Comparison of two extraction techniques

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Enhancement of Linearity and Response in Charged Aerosol Detection

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Determination of urea in ultrapure water by IC-MS/MS

Transcription:

Validation of GC-MS/MS for Detection and Confirmation of Low-Level Dioxins Cristian Cojocariu, 1 Paul Silcock, 1 and Alexander Kotz 2 1 Thermo Fisher Scientific, Runcorn, U.K. 2 European Union Reference Laboratory for Dioxins and PCBs in Feed and Food, State Institute for Chemical and Veterinary Analysis of Food (CVUA), Freiburg, Germany Application Note 146 Key Words Food and Feed, Limit of Quantification, TargetQuan, TRACE 131 GC, TSQ 8 Evo GC-MS/MS Goal To validate the use of triple quadrupole GC-MS/MS technology for detection, quantification, and confirmation of low-level dioxins in feed and food samples Introduction Dioxins are highly toxic organic compounds often released into the environment unintentionally, mostly as by-products of human industrial activities (e.g. from the production, use, and combustion of chlorine containing organic compounds) or, to a lesser extent, from natural processes (such as forest fires or volcanic activity). Although the concentration of dioxins in the environment is usually low (parts per trillion level), these compounds are extremely toxic. Dioxins bioaccumulate and persist in the food chain. Humans are exposed to dioxins mainly through ingestion of food of animal origin. Due to their highly toxic potential, even low concentrations of dioxins can have negative effects on human health. They can cause reproductive and developmental problems, damage the immune system, interfere with hormones and cause cancer. 4 Hence, it is important to accurately detect, confirm, and quantify dioxins in food and feed samples even at low levels. However, detection and confirmation of low dioxin levels in such complex samples is a challenging task, requiring sensitive, selective, and robust analytical instrumentation. Newly introduced European Commission regulations permit the use of GC-MS/MS triple quadrupole technology for dioxin quantification and confirmation. 2, 3 In this study, the Thermo Scientific TSQ 8 Evo triple quadrupole GC-MS/MS is used to detect and confirm low-level dioxins in sample extracts. Moreover, a thorough validation of the TSQ 8 Evo GC-MS/MS for dioxin confirmation is described, including instrument sensitivity, linearity, limits of quantification (LOQs), and measurement precision. In addition, the data acquired using the GC-MS/MS was compared with the sector instrument data acquired for the same sample to assess measurement uncertainties. This work demonstrates that the TSQ 8 Evo triple quadrupole GC-MS/MS delivers consistent results and can be used with confidence to sensitively and selectively detect, quantify, and confirm dioxins in food and feed samples at low concentrations.

2 Table 1. Sample extracts used for dioxin quantification. Mixed Animal Fat Fish Meal Egg Fat 1 Egg Fat 2 Weighted Sample 1. g fat 6.2 g sample (= 1. g fat).52 g fat.51 g fat Lipid Content [%] n/a 16.6 n/a n/a Moisture Content [%] n/a 5.3 n/a n/a Final Volume 4 µl 4 µl 1 µl 1 µl Instrument and Method Setup All of the experiments described below use a TSQ 8 Evo triple quadrupole GC-MS/MS instrument coupled with a Thermo Scientific TRACE 131 GC. Sample introduction is performed with a Thermo Scientific TriPlus RSH autosampler, and chromatographic separation is achieved using a Thermo Scientific TraceGOLD TG-5SilMS 6 m.25 mm I.D..25 µm film capillary column. Additional details of instrument parameters are displayed in table below. Instrument Conditions TRACE 131 GC Injection Volume (μl): 2 Liner: SSL single taper (P/N 453A2342) Inlet ( C): 26 Inlet Module and Mode: Splitless Carrier Gas (ml/min): He, 1.2 Oven Temperature Program: Temperature 1 ( C): 1 Hold Time (min): 2 Temperature 2 ( C): 25 Rate ( C/min) 25 Temperature 3 ( C): 285 Rate ( C/min) 2.5 Temperature 4 ( C): 33 Rate ( C/min) 1 Hold Time (min): 5 TSQ 8 Evo Mass Spectrometer Transfer Line ( C): 28 Ionization Type: EI Ion Source( C): 3 Electron Energy (ev): 4 Acquisition Mode: SRM Q2 Gas Pressure (Argon) (psi): 6 Q1 Peak Width (Da):.7 Q3 Peak Width (Da):.7 Operate the TSQ 8 Evo mass spectrometer in MS/MS mode using electron ionization (EI+). For each target compound, choose two selected reaction monitoring (SRM) transitions. 1 Acquire data using timed selection reaction monitoring (t-srm) with a minimum of 12 points/chromatographic peak. Sample Preparation Use both EPA1613 solvent standards (Wellington Laboratories, Inc.) and food and feed animal fat, fish meal (feed), and egg sample extracts (Table 1). Perform sample extraction and clean up in accordance with the existing analytical criteria for food and feed, 2,3 either automatically (PowerPrep Multi-column Sample Cleanup) or partly manual/partly automated (using GPC, multi-layer silica column, Florisil column, carbon column). To each of the samples, add 5 6 pg (mixed animal fat, fish meal) or 1 12 pg of 13 C-labeled internal standards (egg fat), and 5 pg (mixed animal fat, fish meal), or 1 pg (egg fat) of the 13 C-1234-TCDD recovery standard. Data Processing Carry out peak detection and compound quantification in the analyzed samples using Thermo Scientific TargetQuan software, which enables quantitation based upon relative response factors and includes toxic equivalence factors (TEFs) to automatically calculate toxic equivalence quotients (TEQs) and total TEQ of a sample.

Figure 1. Chromatography example for 2378-TCDD and its internal standard 13C-2378-TCDD present in the fish meal (a), mixed animal fat (b), and egg fat (c) samples. Absolute amounts of native 2378-TCDD on column (fg) are indicated. 3 Results and Discussion Chromatography Excellent chromatographic separation was achieved for all sample extracts, even at very low target levels. Example chromatograms are shown for 2378-TCDD in the feed, animal fat, and egg samples (Figure 1). Linearity of Response Dioxin linearity was assessed by determining the average RF % relative standard deviation (RSD) values from a six point calibration curve measured at the beginning and at the end of the sample batch. Excellent %RSD values were obtained. All measured compounds had %RSD values between.7 4.5%, well within the 15% limits established by the U.S. EPA 5 (Table 2). Table 2. Precision on the average response factor (%RSD) for each native compound calculated from a six-point calibration curve (CSL-CS4). Values represent duplicate measurements of each calibration point, measured at the beginning and end of a batch. Linearity/Calibration Compound Concentration Range (pg/µl) Average RF STDEV RF %RSD 2378-TCDF.1 4 1..4 3.7 2378-TCDD.1 4 1.4.2 1.4 12378-PeCDF.5 2 1..1.9 23478-PeCDF.5 2 1..2 1.8 12378-PeCDD.5 2 1.2.3 2.5 123478-HxCDF.5 2 1..1.8 123678-HxCDF.5 2 1..1 1.3 234678-HxCDF.5 2 1..1 1.3 123478-HxCDD.5 2.9.1.7 123678-HxCDD.5 2 1.1.1 1.3 123789-HxCDD.5 2 1..1.7 123789-HxCDF.5 2 1..5 4.5 1234678-HpCDF.5 2 1.1.1.9 1234678-HpCDD.5 2 1.1.1 1. 1234789-HpCDF.5 2 1.1.2 1.4 OCDD 1. 4 1.5.2 1.6 OCDF 1. 4 1.1.2 1.4

4 Determination of Limit of Quantification (LOQ) Data acquired in SRM often shows little background noise due to the high selectivity of this acquisition mode on a triple quadrupole GC-MS instrument. For this reason, the LOD/LOQ could not be reliably calculated from the signal-to-noise (S/N) values. Here, an empirical approach was used to determine the LOQ for all native dioxins and furans targeted. This approach took into account the lowest level of each of the congeners that can be quantitatively determined and confirmed with suitable precision (<15% RSD) and accuracy, in addition to meeting all other confirmation criteria stated in the European Commission regulations. 2,3 In the first set of experiments, an EPA1613 CSL standard was diluted serially. Each dilution level was injected repeatedly (n=1) and the LOQ for each native compound was calculated taking into account the Student s-t critical values for the corresponding degrees of freedom (99% confidence). The results of this experiment are shown below, with LOQ values for all 17 PCDD/Fs congeners ranging from.1.6 pg/µl (Table 3). The second set of experiments focused on assessing both the instrument sensitivity and the LOQ for the tetra chlorinated dioxins. A solvent standard containing six TCDD congeners with gradually increasing concentrations (ranging from 1 1 fg/µl) was used. Peak shape, the peak area reproducibility (n= 8 injections) and the ion ratio deviation (% from the theoretical value), were evaluated for each congener and concentration level. The results of this experiment confirmed that TCDD can be confidently detected and confirmed at 1 fg on a column with ion ratio values lower than the 15% maximum limit and with peak area reproducibility (no internal standard correction) better than 7% RSD (Figure 2). Table 3. Limits of quantitation (LOQ) for the 17 dioxin congeners calculated from CSLx5 and CSLx1 dilution and taking into account the ion ratio, response factors, and precision limits.. PCDD/Fs Limit of Quantification Compound pg/µl 2378-TCDF.1 2378-TCDD.1 12378-PeCDF.2 23478-PeCDF.2 12378-PeCDD.2 123478-HxCDF.4 123678-HxCDF.4 234678-HxCDF.3 123478-HxCDD.6 123678-HxCDD.3 123789-HxCDD.4 123789-HxCDF.4 1234678-HpCDF.2 1234678-HpCDD.5 1234789-HpCDF.2 OCDD.5 OCDF.3 1 9 8 7 6 Inj. No. Concentration 1 fg 25 fg 1 fg 25 fg 5 fg 1 fg 1 3129 711 2594 6219 87635 239216 2 345 629 25998 5918 85789 24162 3 347 755 27425 58262 87578 237274 4 2637 744 25734 59335 8779 233655 5 2678 6376 2785 58583 83897 23447 6 2964 6513 2797 5892 85373 238554 7 2828 6931 26347 59936 86844 24274 8 2699 6579 2533 55519 79471 23238 2378-TCDD 1 fg IR % = 4 5 4 3 Mean 2878 6764 26422 58733 85458 237263 STDEV 192 411 94 1452 273 3625 %RSD 6.7 6.1 3.6 2.5 3.2 1.5 1478-TCDD 25 fg IR % = 6 1234-TCDD 5 fg IR % = 11 2 1 1368-TCDD 1 fg IR % = 9 1379-TCDD 25 fg IR % = 6 1378-TCDD 1 fg IR % = 7 13.4 13.5 13.6 13.7 13.8 13.9 14. 14.1 14.2 14.3 14.4 14.5 14.6 14.7 14.8 14.9 Minutes Figure 2. Overlay SRM chromatograms (quan ion) of six TCDD congeners in a stepped standard (concentration ranging from 1 1 fg on column). Annotated are the peak area (AA), retention time (RT), percentage deviation of the ion ratio value from the theoretical value (IR %), corresponding concentration on column, and congener name. The included table indicates peak area reproducibility as %RSD for each congener and corresponding concentration level.

1. 5 WHO pg TEQ/g fat.9.8.7.6.5.4.3.2 GC-HRMS GC-MS/MS egg-1 GC-MS/MS egg-2.1. 2378-TCDF 12378-PCDF 23478-PCDF 123478-HxCDF 123678-HxCDF 234678-HxCDF 123789-HxCDF 1234678-HpCDF 1234789-HpCDF OCDF 123678-HxCDD 123789-HxCDD 1234678-HpCDD OCDD Figure 3. Comparison of the individual dioxin concentrations (calculated as pg TEQ/g fat) using GC-HRMS and GC-MS/MS. Data shows the values for two individual extractions from the same egg sample (egg-1 extracted from.52 g of egg fat and egg-2 from.51 g of egg fat). 2378-TCDD 12378-PCDD 123478-HxCDD Determination of Dioxins in Sample Extracts As required by the EU regulations, determination of PCDD/Fs in a sample should be performed with a high degree of accuracy. 2,3,5 In order to evaluate the accuracy of measurements, the egg, animal fat, and fish meal samples were analyzed for their dioxin content. The calculated concentration of each individual dioxin congener (pg TEQ/g fat or ng/kg %dw) was compared with the values obtained from a GC-HRMS system. The data shows excellent agreement between the results obtained using the TSQ 8 Evo GC-MS/MS and that obtained using GC-HRMS (Figure 3). To increase the confidence in detection of the target compounds, the ion ratio (IR) abundance value of each of the native compounds was measured in the sample extracts and the values compared with both the theoretical ion ratio values and the values derived from the CS3 calibration check standard, as stated by the new dioxin regulations. 2,3 The results of this experiment show that all the IR abundance values for the compounds analyzed were within the 15% tolerance, increasing the confidence in the results and meeting the EU criteria for dioxin confirmation. 2 An example ion ratio abundance calculation is shown in Figure 4 for the egg samples. 1.2 1.8 Area Ratio.6.4.2 2378-TCDD 12378-PeCDD 123478-HxCDD 123678-HxCDD 123789-HxCDD 1234678-HpCDD OCDD 2378-TCDF 12378-PeCDF 23478-PeCDF 123478-HxCDF 123678-HxCDF 234678-HxCDF 123789-HxCDF 1234678-HpCDF 1234789-HpCDF IR CS3 IR theoretical Figure 4. Ion ratios of dioxins detected in egg fat samples. Comparison of individual IR average values (n=2) with the theoretical ion ratio values (red symbol) and with the values derived from the CS3 calibration standard (green symbol). OCDF Average

6 Precision of Measurement Obtaining a valid determination of dioxin concentration in sample extracts is critical. Dioxin concentrations are normally assessed by looking at the %RSD calculated from the results generated under reproducible conditions. 2,3 In this study, precision measurement of the total dioxin content (as WHO TEQ pg/g or ng/kg) was determined by running a sequence of n=13 repeat injections of the mixed animal fat and the fish meal samples. The results show excellent precision for both sample extracts with 6% RSD for the animal fat and 4% RSD for the fish meal sample (Figures 5 and 6). Where an analytical result for an individual dioxin congener was below the LOQ, the actual content could be anywhere between zero and the LOQ. For this reason, when a compound was not detected, the corresponding LOQ was used as the concentration value. This way, the upper bound TEQs for each sample extract were derived by summing the weighted concentrations for each of the 17 dioxin congeners analyzed and the results compared with the upper bound TEQs obtained from the GC-HRMS data. Total dioxin content (WHO-PCDD/ F-TEQ pg/g upper bound) of each sample in each of the 13 repeat injections obtained from the GC-MS/MS analysis was plotted against the sector instrument data acquired for the same samples. The calculated deviation did not exceed 16.6% for the animal fat sample (Figure 5) or 8% for the fish meal sample (Figure 6). Injection No WHO-PCDD/F-TEQ ub 1.91 2.76 3.88 4.9 5.94 6.83 7.84 8.85 9.88 1.87 11.78 12.82 13.84 Mean.85 STDEV.5 %RSD 5.97 % Difference form HRGC-MS 3 2 1 1 2 3 1 2 3 4 5 6 7 8 9 1 11 12 13 Injection No Figure 5. Mixed animal fat sample: deviation (%) of the total dioxin concentration (WHO-PCDD/F-TEQ pg/g upper bound) from the GC-HRMS results. Table shows the reproducibility of the total dioxin content (WHO-PCDD/F-TEQ pg/g upper bound) (n=13 injections) with the average value, standard deviation, and %RSD. Injection No WHO-PCDD/F-TEQ ub 1.26 2.26 3.24 4.25 5.26 6.26 7.26 8.27 9.24 1.26 11.24 12.24 13.27 Mean.25 STDEV.1 %RSD 4.3 % Difference form HRGC-MS 3 2 1 1 2 3 1 2 3 4 5 6 7 8 9 1 11 12 13 Injection No Figure 6. Fish meal sample: deviation (%) of the total dioxin concentration (WHO-PCDD/F-TEQ ng/kg 88% dw upper bound) from the GC-HRMS results. Table shows the reproducibility of the total dioxin content (WHO-PCDD/F-TEQ ng/kg 88% dw upper bound) (n=13 injections) with the average value, standard deviation, and %RSD.

Table 4. Comparison of the total dioxin content (WHO-PCDD/F-TEQ-upper bound) of the sample extracts analyzed by GC-HRMS and by GC-MS/MS. Average values of replicate injections (n) with corresponding standard deviations (in brackets) are given. Sample Type Additionally, calculation of the % deviation of the total dioxin content from the GC-HRMS data for all the samples analyzed shows a minimum of 2.2% for the fish meal (average for n=13 injections) and a maximum of 5.5% for the mixed animal fat (Table 4). Conclusion The TSQ 8 Evo triple quadrupole GC-MS/MS system is suitable for the quantification and confirmation of dioxins and furans in food and feed matrices at low levels. Excellent selectivity and sensitivity were achieved even at very low levels (e.g. 28 fg 2378-TCDD absolute amount on column) in matrix samples. Excellent precision measurements of the total dioxin content (WHO-PCDD/F-TEQ-ub) were obtained for two low level contaminated samples (fish meal and mixed animal fat) with %RSD values <6% (n=13). Deviation of the total dioxin content (WHO-PCDD/ F-TEQ-ub) from the GC-HRMS data was within ±2% for all the samples analyzed. WHO-PCDD/F-TEQ-ub GC-HRMS GC-MS/MS Deviation % Egg (n=2) 3.39 pg/g 3.27 (.3) 3.5 Fish Meal (n=13).26 ng/kg % dw.25 (.1) 2.1 Mixed Animal Fat (n=13).83 pg/g.85 (.4) 2.8 References 1. Thermo Scientific Application Note 138: Meeting the European Commission Performance Criteria for the Use of the Triple Quadrupole GC-MS/MS as a Confirmatory Method for PCDD/Fs in Food and Feed Samples. Runcorn, U.K. [Online] http://www. thermoscientific.com/content/dam/tfs/atg/cmd/ CMD%2Documents/Application%2&%2 Technical%2Notes/Mass%2Spectrometry/ GC%2MS/AN-138-TSQ-8-EVO-PCDD-FS- AN138-EN.pdf (accessed Sept. 26, 214). 2. Commission Regulation (EU) No 589/214 of 2 June 214 laying down methods of sampling and analysis for the control of levels of dioxins, dioxin-like PCBs and non-dioxin-like PCBs in certain foodstuffs and repealing Regulation (EU) No 252/212, Off. J. Eur. Union: Legis. 214. 164, p. 18 4 3. Commission Regulation (EU) No 79/214 of 2 June 214 amending Regulation (EC) No 152/29 as regards the determination of the levels of dioxins and polychlorinated biphenyls, Off. J. Eur. Union: Legis. 214. 188, p. 1 18. 4. World Health Organization, Fact sheet N 225: Dioxins and their effects on human health, Updated June 214. 5. US EPA Method 1613: Tetra-through octa-chlorinated dioxins and furans by isotope dilution HRGC/HRMS (Revision B), 1994. Application Note 146 www.thermoscientific.com/chromatography 214 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. PowerPrep is a trademark of FMS, Inc. Florisil is a registered trademark of U.S. Silica Co. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 91:28 Certified. Africa +43 1 333 5 34 Australia +61 3 9757 43 Austria +43 81 282 26 Belgium +32 53 73 42 41 Brazil +55 11 3731 514 Canada +1 8 53 8447 China 8 81 5118 (free call domestic) 4 65 5118 AN146-EN 114S Denmark +45 7 23 62 6 Europe-Other +43 1 333 5 34 Finland +358 9 3291 2 France +33 1 6 92 48 Germany +49 613 48 114 India +91 22 6742 9494 Italy +39 2 95 591 Japan +81 6 6885 1213 Korea +82 2 342 86 Latin America +1 561 688 87 Middle East +43 1 333 5 34 Netherlands +31 76 579 55 55 New Zealand +64 9 98 67 Norway +46 8 556 468 Russia/CIS +43 1 333 5 34 Singapore +65 6289 119 Sweden +46 8 556 468 Switzerland +41 61 716 77 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 8 532 4752