Method - Determination of aromatic amines in mainstream cigarette smoke

Similar documents
British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

CORESTA Recommended Method No. 83

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Sulfotepp impurities in Chlorpyrifos EC formulations

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Polymer Modifier in Asphalt

Uncontrolled Copy. SOP Ethylene Glycol Quantitation by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Chemistry Gas Chromatography: Separation of Volatile Organics

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

NORBUPRENORPHINE (Buprenorphine s Metabolite ) BUPRENORPHINE in urine by GC/MS Code GC Method of Confirmation by GC-MS

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

TSNAs in Tobacco Filler and Cigarette Smoke

California Environmental Protection Agency Air Resources Board

ENVIRONMENTAL analysis

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV*

CORESTA Recommended Method No. 86

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Method Formaldehyde, Isobutyraldehyde, and Furfural by Derivatization Followed by High Performance Liquid Chromatography

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

Technical Report. Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

METHANOLYSIS OF ACETAL

STANDARD OPERATING PROCEDURES

Smoke Analytes Sub-Group (SMA) Report 2018

Place a collection tube beneath cartridge. 3 ml methylene chloride/isopropyl alcohol/ hydrochloric acid (60/40/1) Collect the eluant

Validation of TNCO, TSNAs, and PAHs

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

EnAlgae SOP sheet: Silicate - SFA. 1. Purpose This procedure is to analyse seawater and freshwater for silicate, SiO2.

Determination of Volatile Substances Proof of Food Adulteration

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

The analysis of organic acid content of additives, premix, feed, and water.

ONDERSTEPOORT VETERINARY INSTITUTE METHOD: RTETH 042 DETERMINATION OF TETRACYCLINES IN HONEY 1. INTRODUCTION 2. EQUIPMENT

Toxicology Extraction Procedure for Base Drugs Using United Chemical Technologies Clean Screen Extraction Columns

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Determination of 2-Imidazolidinethione in Fruits and Vegetables by GC/MSD

NCASI METHOD CI/SG/PULP-94.02

STANDARD OPERATING PROCEDURES SOP: 1824 PAGE: 1 of 22 REV: 0.0 DATE: 04/21/95 ANALYSIS OF METHYL PARATHION IN SOIL SAMPLES BY GC/MS

METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

BUTYL ALCOHOL in urine by GC-FID Code GC05510

Method 548. July 1990 EPA EMSL-Ci. J.W. Hodgeson

JWH-018 and JWH-022 as Combustion Products of AM2201

TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010

TO-15 Checklist Determination of VOCs in Air by GC-MS

ANALYTICAL METHOD PROCEDURES

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

DETERMINATION OF PHENOLS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) Revision 1.

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Protocol 3: Protocol for the Analysis of Polybrominated Diphenyl Ethers (PBDE) in Human Milk, Air and Human Serum

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Determination of Phenoxyacid Herbicides in Water by Solid Phase Extraction and LC-MS/MS Detection

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

TEST METHOD. Ethoxyl Substitution in ETHOCEL TM Ethylcellulose Polymers. 1. Scope. 2. Principle. 3. Safety

637. Thiamethoxam. HPLC method

TKI CU01 V-1.0_EN Cu separation from irradiated targets 01/12/10

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

Acknowledgements. For more information about this method, contact:

The Nitrofurantoin Capsules Revision Bulletin supersedes the currently official monograph.

METHANOL REFERENCE SPECIFICATIONS

Report AFK0242/18 TABLE OF CONTENTS

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Determination and quantification by GC-MS of melamine (1 mg/kg), ammeline and cyanuric acid (20 mg/kg)

ACETONE IN URINE BY GC/MS Headspace Code GC10110

METHOD 3520B CONTINUOUS LIQUID-LIQUID EXTRACTION

Determination of urea in ultrapure water by IC-MS/MS

NICKEL-63/59 IN WATER

Certified Reference Material - Certificate of Analysis

Spectrophotometric Determination of Ferrocyanide in Effluents

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

IMPCA METHANOL REFERENCE SPECIFICATIONS

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Transcription:

British American Tobacco Group Research & Development Method - Determination of aromatic amines in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicaple to quantitative determination of the yields of 1-aminonaphthalene, 2- aminonaphthalene, 3-aminobiphenyl, and 4-aminobiphenyl (henceforth collectively termed Aromatic Amines ) in mainstream cigarette smoke, using gas chromatography with mass selective detection. 2 NORMATIVE REFERENCES ISO 3308:2000 Routine analytical cigarette smoking machine definitions and standard conditions ISO 3402:1999 Tobacco and tobacco products atmospheres for conditioning and testing ISO 4387:2000 Cigarettes - Determination of total and nicotine-free dry particulate matter using a routine analytical smoking machine ISO 8243:2006 Cigarettes - Sampling 3 PRINCIPLE Twenty conditioned cigarettes are smoked using a 20 port rotary Borgwaldt smoking machine. The mainstream smoke is collected onto a 92mm Cambridge filter pad. After smoking the Cambridge filter pad is extracted in a 5% HCl solution. Two SPE steps using different retention mechanisms are used to clean the HCl extract and to isolate the aromatic amines. The aromatic amines are then eluted with toluene, derivatised with heptafluorobutyric anhydride and analysed by GC/MS using selective ion monitoring (SIM). 4 HEALTH & SAFETY Read and understand the Material Safety Data Sheets for the chemicals used in this method. Read and understand the method risk assessment. Ensure that you understand the hazards and follow control measures relevant to the operation of this method. All preparation of standards and extraction of samples must be performed in a fume cupboard. 5 REAGENTS AND MATERIALS All reagents are Analytical Grade or equivalent unless otherwise stated. 1-Aminonaphthalene 10ng/µL in toluene 2-Aminonaphthalene 10ng/µL in toluene 3-Aminobiphenyl 10ng/µL in toluene 4-Aminobiphenyl 10ng/µL in toluene 2-Aminonaphthalene-d 7 0.1mg/mL in toluene 4-Aminobiphenyl-d 9 0.1mg/mL in toluene Hydrochloric acid (37%) Ammonium hydroxide (29%) Methanol HPLC grade Toluene Pyridine HPLC grade Heptafluorobutyric anhydride Page 1 of 9

Deionised water 6 APPARATUS Borgwaldt RM200 20 port rotary smoking machine Soap bubble manometer to measure puff volume Analytical balance capable of measuring to at least four decimal places 92mm Cambridge filter pads 50mL Volumetric flasks (class A) with stoppers 20mL Volumetric flasks (class A) with stoppers 250mL Schott bottles with lids 10mL Measuring cylinder 100mL Measuring cylinders 1L Measuring cylinder Calibrated variable volumetric multipipette capable of dispensing up to 2mL volume Air-displacement pipette capable of dispensing 4µL A range of calibrated positive-displacement pipettes capable of dispensing from 10µL to 1000µL 12mL Plastic (styrene) disposable test tubes Wrist action shaker 2mL amber GC vials with crimp top caps (PTFE coated) 2mL clear vials Vial crimper 50mL dispenser (accuracy + 5mL) 6mL 150mg MCX (cation exchange) SPE cartridges 6mL 200mg HLB (hydrophobic-lipophilic balance) SPE cartridges Heating block capable of holding 2 L GC vials and incubating to 80 C Manifold vacuum block capable of holding SPE cartridges 40mL SPE Polypropylene filter tubes with plastic connectors 5.5cm diameter filter paper SPE elution racks capable of holding 12mL tubes and 2mL glass vials ph 11 indicator sticks Agilent GC/MS with autosampler J&W DB5 MS Chromatography column (30m x 0.25mm i.d x 0.25µm) 1m Deactivated fused silica column 7 PRELIMINARY SAMPLE PREPARATION Cigarettes should be conditioned according to normal procedures (ISO 3402:1999). Unless specifically requested, cigarettes are not subject to any selection criteria other than the rejection of any obviously defective or damaged cigarettes. Butt marking is as specified in ISO 4387:2000 unless otherwise requested. 8 ANALYTICAL PROCEDURE SOLUTION PREPARATION 8.1 Extracting solutions Page 2 of 9

8.1.1 5% Hydrochloric acid solution 135mL (± 5mL) concentrated Hydrochloric acid (37%) is carefully added to 865mL (± 5mL) deionised water using a 1L measuring cylinder and stored in a container with dispenser attached. Prepare at least weekly or with each batch of samples. 8.1.2 1% Hydrochloric acid solution 20mL (± 1mL) of the 5% hydrochloric acid solution (8.1.1) is added to 80mL (± 1mL) de-ionised water using a 100mL measuring cylinder and stored in a Schott bottle. Prepare at least weekly or with each batch of samples. 8.1.3 5% Ammonium Hydroxide solution in methanol 17mL (± 1mL) ammonium hydroxide (29%) solution is added to 83mL (± 1mL) methanol using a 100mL measuring cylinder and stored in a Schott bottle. Prepare at least weekly and refrigerate when not in use. 8.1.4 ph11 Ammonium Hydroxide solution 75mL (± 5mL) ammonium hydroxide (29%) solution is added to 425mL (± 5mL) de-ionised water using a 500mL measuring cylinder. Check the alkalinity (ph 11 ± 1) using an appropriate ph stick and remake if necessary before storing in a Schott bottle. Prepare at least weekly. 8.1.5 30/70 Methanol/water solution 30mL (± 1mL) methanol is added to 70mL (± 1mL) de-ionised water using a 100mL measuring cylinder and stored in a Schott bottle. Prepare at least weekly. 8.2 Standard Preparation The following standards are stable for 12 months and should be refrigerated when not in use. 8.2.1 Internal standard Solution Pipette 2.5mL 2-aminonaphthalene-d 7 stock solution (0.1mg/mL) and 0.5mL 4-aminobiphenyl-d 9 Stock solution (0.1mg/mL) into a 50mL volumetric flask (class A) and make up to volume with toluene. Concentration of the Internal Standard Solution will be: 5ng/µL 2-Aminonaphthalene-d 7 1ng/µL 4-Aminobiphenyl-d 9 8.2.2 Calibration standards preparation Using positive displacement pipettes add the amounts of purchased aromatic amines stock solutions (10ng/µL) into volumetric flasks as set out in Table 1 and make up to volume with toluene. Table 1 Calibration Stock Solution Preparation Calibration Standard 1- Aminonaphthalene µl 2- Aminonaphthalene µl 3- Aminobiphenyl µl 4- Aminobiphenyl µl Volumetric Flask Size ml 1 50 50 10 10 50 2 250 250 50 50 50 3 500 500 100 100 50 4 400 400 80 80 20 5 600 600 120 120 20 6 800 800 160 160 20 7 1000 1000 200 200 20 Page 3 of 9

To prepare the calibration standard vials, add 100µL of internal standard solution (8.2.1) to 900µL of each calibration stock solution standard in separate autosampler vials. This will achieve the amounts for each level of the working calibration standard as listed in Table 2. Table 2 Calibration Vials Summary Calibration Standard 1- Aminonaphthalene (ng) 2- Aminonaphthalene (ng) 3-Aminobiphenyl (ng) 1 9 9 1.8 1.8 2 45 45 9 9 3 90 90 18 18 4 180 180 36 36 5 270 270 54 54 6 360 360 72 72 7 450 450 90 90 4-Aminobiphenyl (ng) 8.2.3 Add 4µL of the derivatisation agent heptafluorobutyric anhydride, and 10µL of pyridine to each vial. Incubate the vials for 30 minutes at 80 C ± 5 C using a suitable heating block. 8.3 QC Preparation 8.3.1 Prepare the QC standard as described in Table 3. Table 3 QC Stock solution preparation Amount of QC Standard (10ng/µL) added (µl) Volume (ml) Concentration ng/ml 1-Aminonaphthalene 500 250 2-Aminonaphthalene 500 20 250 3-Aminobiphenyl 100 50 4-Aminobiphenyl 100 50 8.3.2 Add 4µL of the derivatisation agent Heptafluorobutyric anhydride, and 10µL of pyridine to each vial. Incubate the vials for 30 minutes at 80 C (± 5 C) using a suitable heating block. 9 ANALYTICAL PROCEDURE SAMPLE PREPARATION 9.1 Sample Collection Cigarettes are smoked on a Borgwaldt rotary 20 port smoking machine. Typically the RM200 is used, but if the cigarette style is not suitable, the RM20 CSR may be used. Warm-up the smoking machine for 20 minutes before smoking. Check the linear airflow is 200 mm/s (±30 mm/s), the system has no leaks and puff volume is 35mL ( ± 0.3mL) (for ISO smoking). 20 cigarettes are loaded and smoked. Record the number of lit puffs and the weight of Total Particulate Matter (TPM). Page 4 of 9

9.2 Sample Extraction After smoking, cut the pad into quarters and add 100mL (± 10mL) of 5% HCl to the pad in a 250mL Schott bottle using a 50mL dispenser. Add 100µL (± 2µL) of deuterated internal standard solution to the acidified sample using a positive displacement pipette. Place on a wrist-action shaker for 30 minutes at 180rpm (± 10rpm). 9.3 Sample Clean Up 9.3.1 Apply all of the HCl extract to an unconditioned Waters Oasis MCX cartridge. Use a slight vacuum to initiate the drip-through of the acidified sample. 9.3.2 Rinse the MCX cartridge with 2mL 1% HCl using the variable volume autopipette. Rinse with 2mL methanol. Discard the acidified waste. 9.3.3 Elute the compounds retained on the MCX with 2mL of 5% ammomium hydroxide in methanol into the plastic test tubes. Use the vacuum to remove any residual eluting solvent from the cartridge. 9.3.4 Using a 10 ml measuring cylinder dilute the eluant with 10mL (± 2mL) of ph 11 ammonium hydroxide solution to modify the sample for the reverse-phase HLB cartridge. 9.3.5 Condition the HLB cartridge with 3mL of methanol followed by 2mL of ph 11 ammonium hydroxide solution using an autopipette. 9.3.6 Apply the diluted eluant to the HLB cartridge to retain the aromatic amines. 9.3.7 Rinse the cartridge with 2mL of ph 11 ammonium hydroxide and then rinse with 2mL of 30/70 methanol/water to remove some of the more polar and less hydrophobic compounds. Dry the cartridge using the vacuum manifold for at least 15 minutes to ensure that no water remains in the final extract. Discard the waste solvent from the manifold. 9.3.8 Elute the Aromatic Amines from the HLB cartridge into the clear 2mL vials using 1.5mL of toluene. Transfer the toluene eluant to an autosampler vial using glass pasteur pipettes. Should the final extract contain traces of water ensure this is not transferred to the autosampler vial. Add 4µL of the derivatization reagent, Heptafluorobutyric Anhydride. Crimp the vial, shake, and place in a pre-heated heating block for 30 minutes at 80 C (± 5 C). The sample is ready for analysis on a GC/MS. Page 5 of 9

10 ANALYTICAL PROCEDURE INSTRUMENTAL ANALYSIS 10.1 Instrument Set Up Parameters Analysis is performed on an Agilent 6890 Gas Chromatograph (GC) fitted with autosampler and 5973 Mass Selectvie Detector (MSD). Column Type J&W DB5 MS 30m x 0.25mm i.d x 0.25µm Injection type and temperature Splitless/ 280 C Column temperature programme Initial 80 C hold for 2 minutes Ramp 30 C/min to 150 C 16 C/min to 210 C hold for 2 minutes 30 C/min to 280 C hold for 2 minutes Total run time is 14.42 minutes Carrier gas Helium Transfer line temperature 280 C Injection Volume 1µL Column Flow 1mL/min (constant flow) Solvent Delay 5 minutes MS Source temperature 230 C MS Quadrupole temperature 150 C MS Mode SIM Ion Dwell time The following ions are used as target ions: Target 1-Aminonaphthalene 339 (100%) 2-Aminonaphthalene 339 (100%) 3-Aminobiphenyl 365 (100%) 4-Aminobiphenyl 365 (100%) 2-Aminonaphthalene-d 7 346 (100%) 4-Aminobiphenyl-d 9 374 (100%) Standards and samples are quantified on the target ions stated above. 10.2 System Suitability Criteria 10.2.1 MS Tuning Tune the MS weekly, or if the system has been vented. Check the following criteria are met on the tune report: Air and water peaks <10% EM Volts 1000 3000 Page 6 of 9

Ion ratios of m/z 69:219:502 are approximately 10:4:1 Peak width approximately 0.6 (+10%) 10.2.2 Peak check System suitability is based on 1-aminonaphthalene. If this analyte passes the system suitability criteria it is assumed that the other aromatic amines will also pass. For the QC: Peak height must be 500,000m or above USP tailing not grater than 2.0 and not less than 0.9 10.2.3 Calibration linearity The R 2 value of the calibration graph must be >0.995 10.3 Run Order Start run with two conditioning vials (the first of blank toluene and the second of a previously analysed smoked sample) Standards in ascending order (1-7) QC standard Reference cigarette sample 1 Smoked samples (1 x 5 replicates) Reference cigarette sample 2 Smoked samples (1 x 5 replicates) QC standard If further samples are on the sequence at least one QC standard should be run for every 12 sample vials. A QC standard should be run to complete the sequence. 11 CALCULATIONS Using the instrument software, plot a calibration graph of calibration standards concentration against peak area ratio, without forcing the line through zero. Peak area ratio = 2-aminonaphthalene peak area/2-aminonaphthalene-d 7 peak area Check the plots, coefficient of determination (R 2 ) and intercept before accepting the calibrations. Calculate the concentration of the aromatic amines in the sample solutions. The results obtained form the GCMS are in ng. To convert to ng/cigarette divide the result in ng by the number of cigarettes smoked for each sample (usually 20). 12 PRECISION AND REPORTING LIMITS Five replicate smokings and analyses are performed to determine the precision of the the analysis. Longer term precision is monitored through the maintenance of control charts. The lower reporting limit is defined by the concentration of the lowest calibration standard and equates to 0.45ng/cigarette for 1- and 2-aminonaphthalene and 0.09ng/cigarette for 3-and 4- aminobiphenyl. 13 QUALITY ASSURANCE AND CONTROL Control charts of the QC standard and the reference cigarette are maintained to allow inspection of the method performance. 14 SPECIAL CASES Under more intense smoking regimes, the number iof cigarettes per smoking run may need to be reduced in order to avoid smoke breakthrough on the Cambridge filter pad. Page 7 of 9

Date of Issue: 31 March 2008 Page 8 of 9

APPENDIX A SAMPLE CHROMATOGRAMS QC Standard : 225 ng/ml 1-ANP and 2-ANP, 45 ng/ml 3-ABP and 4-ABP Kentucky Reference Cigarette (2R4F) Page 9 of 9