A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Similar documents
Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1

Application Note S/SL

Introduction to Capillary GC

Introduction to Capillary GC

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Sulfotepp impurities in Chlorpyrifos EC formulations

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Author. Abstract. Introduction

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

BUTYL ALCOHOL in urine by GC-FID Code GC05510

Pittcon P

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

U.S. EPA Method 8270 for multicomponent analyte determination

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

GAS CHROMATOGRAPHY (GC)

Application Note 032

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

STANDARD OPERATING PROCEDURES

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The Pulsed Flame Photometric Detector for analysis of phosphorus pesticides in wastewater and food commodities

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Understanding Gas Chromatography

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

GC/MS Application Note

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Elemental Analysis: CHNS/O characterization of polymers and plastics

Fast and Accurate Analysis of PBDEs in a Single Run, Including

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Highly Sensitive and Rugged GC/MS/MS Tool

Fast and Accurate Analysis of PBDEs in a Single Run, Including BDE-209

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

637. Thiamethoxam. HPLC method

Exercise in gas chromatography

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Multi-residue analysis of pesticides by GC-HRMS

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

GC Application Note. Dilution/standard addition: Efficient and traceable.

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

DEVELOPMENT AND VALIDATION OF GC-FID METHOD FOR THE DETERMINATION OF ETHANOL RESIDUE IN MARJORAM OINTMENT

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Elemental Analysis: NCS characterization of paper

Virtually Particle-Free Rt -Silica BOND Columns

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

NON-METHANE ORGANIC CARBON ANALYZER (NMOC Method 25)

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography

Organochlorine Pesticides by GCxGC-ECD

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Cyanide and sulfide analysis using amperometric detection and Metrosep A Supp /4.0

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Transcription:

Application Note 19010803 A Newly Approved ASTM Standard For Analysis of in Using a Pulsed Flame Photometric Detector (PFPD) Keywords ASTM Standard D4735-96 FPD Flame Photometric Detector Model 5380 PFPD Pulsed Flame Photometric Detector Introduction The ASTM Standard D4735-96, first approved in 1996 and reapproved in 2000, is a standard test method for the determination of trace levels of thiophene in benzene by gas chromatography with a flame photometric detector (FPD). The standard was recently reapproved in mid-2002 with modifications to include the use of the next generation of FPDs, the Pulsed Flame Photometric Detector (PFPD). Additionally, the method was upgraded to include the use of capillary columns. This application note presents data that were used to develop the new standard using the OI Analytical Model 5380 PFPD, including calibration range and linearity, precision and accuracy, intermediate repeatability, and ultimate sensitivity. All instrument configuration and operational parameters are included. The detector specified in the original ASTM standard was any flame photometric detector that would meet the performance criteria outlined in the method. The PFPD, shown in Figure 1, is often referred to as the newest generation of flame photometric detectors and brings with it a 10-fold increase in both sensitivity and selectivity over the older FPDs. The PFPD has the added advantage of providing an equimolar response to sulfur, which the FPDs were not capable, and allows quantitation of unknown sulfur compounds that may also be present in the matrix. The PFPD also has the ability to achieve significantly lower analytical detection limits than the FPD. The PFPD s linearized square root function output permits the easy use of either linear or quadratic calibration curves. These features, along with its improved stability, reduced gas usage, and minimal maintenance requirements, make the PFPD the ideal and logical choice of detectors for this method. Presented at the 2003 Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, Orlando, FL March 9 14, 2003 The objective of this project was to establish instrument operating conditions using the PFPD that would meet or exceed performance criteria outlined in Standard D4735-96. Performance objectives to be demonstrated include the following: Calibration linearity Precision and accuracy Intermediate repeatability Ultimate sensitivity Figure 1. The Model 5380 Pulsed Flame Photometric Detector (PFPD) is often referred to as the newest generation of flame photometric detectors.

Experimental The PFPD was installed on an Agilent 6890N platform with a split/splitless injector. The PFPD was configured for selective sulfur detection and all injections were made with a liquid autosampler. Column selection is critical to ensure that the thiophene is sufficiently separated from the benzene matrix peak. The Carbowax -type columns have been shown to perform best for this type of separation. Complete instrument configuration and operating conditions are listed in Table 1. Table 1. Instrument Operating Conditions for ASTM Standard D4735-96 Parameter Injection GC Column Oven Program Detector Conditions Setting Injector temperature 200 C; 1-µL injection; 50:1 split ratio DB-WAX, 30 m x 0.25 mm I.D. x 0.5-µm film; He carrier gas; constant flow mode 50 C for 1 minute; 10 C/minute to 200 C; hold for 1 minute OI Analytical Model 5380 PFPD; 2-mm combustor; BG-12 broadband optical filter; 6 25 msec sulfur gate H 2 /air gas ratio tuned for sulfur; 250 C; data acquired in both linear and quadratic mode A stock solution of 100 ppm thiophene in benzene from Absolute Standards (part #92715) was diluted to prepare calibration and repeatability test standards. for preparation of standards was ACS Spectranalyzed grade from Fisher Scientific and was certified to be thiophene free. The benzene was demonstrated to be free of thiophene (i.e., below the detection limit with an injection volume five times the amount used for this method) prior to use by chromatographic analysis, as shown in Figure 2. Expected RT = 5.1 min Figure 2. Chromatogram of -free Used to Prepare Analytical Standards. Analysis was Performed Using a 1-µL Injection and a 10:1 Split Ratio (5x the Amount Used for the Method). No was Detected at the Expected Retention Time. Results and Discussion Calibration Six calibration standards spanning a range from 0.5 5.0 ppm thiophene were prepared and analyzed in triplicate using the conditions described above. The calibration curve was created using the average integrated peak areas at each concentration. Chromatograms of the 0.5 and 3.0 ppm standards are shown in Figure 3. Relative standard deviations over the 3 replicate injections for each concentration are shown in Table 2.

0.5 ppm 3.0 ppm RT Figure 3. Chromatograms of the 0.5 ppm and 3.0 ppm in Calibration Standards Acquired Using the PFPD. Data were Acquired Using the Conditions Listed in Table 1. Table 2. Repeatability of Triplicate Injections at Each Calibration Concentration Level Calibration Level %RSD (n = 3) 0.5 ppm 3.7 1.0 ppm 6.1 2.0 ppm 1.9 3.0 ppm 0.6 4.0 ppm 0.2 5.0 ppm 0.6 The flame chemistry that occurs within the PFPD generates an excited sulfur species, S 2 *, that produces a quadratic response. This response can then be digitally processed at the operator s discretion to produce a linear output signal. The PFPD is capable of outputting two simultaneous chromatographic signals, thus, generating two simultaneous chromatograms. For the purpose of this experiment one signal was configured in the quadratic mode and the second signal used the PFPD s automatic square root function to produce a linear response output. In this manner, linear calibration data from the PFPD could be directly compared to the quadratic and log/log plots specified in the original versions of the ASTM standard. These three different treatments of the same data are illustrated in Figure 4. All three charting techniques result in excellent and equivalent calibration curves. The use of linearized data simplifies the calibration process, and allows for quantitation of other unknown sulfur peaks that may be detected using the equimolar response feature of the PFPD.

Log/Log Calibration of in 0.5 ppm to 5.0 ppm Quadratic Calibration of in 0.5 ppm to 5.0 ppm 1.E+05 6.E+04 5.E+04 Log of PFPD Response (Area) 1.E+04 1.E+03 R 2 = 0.9995 PFPD Response (Area) 4.E+04 3.E+04 2.E+04 R 2 = 0.9997 1.E+04 1.E+02 0.1 1 10 Log of Concentration (ppm) 0.E+00 0 1 2 3 4 5 6 Concentration (ppm) Linear Calibration of in 0.5 ppm to 5.0 ppm 3.E+03 2.E+03 PFPD Response (Area) 2.E+03 1.E+03 R 2 = 0.9992 5.E+02 0.E+00 0 1 2 3 4 5 6 Concentration (ppm) Figure 4. Three Different Graphical Treatments of the Same PFPD Calibration Data. Log/log Plots were Required in the Original Version of ASTM Standard D4735-96 Because the Older FPDs were not Capable of Producing a Linear Response. The OI Analytical Model 5380 PFPD Automatically Linearizes the Signal Using Digital Processing, Thus Greatly Simplifying the Calibration Process. Precision and Accuracy Approximately eleven days after the initial calibration was completed, a fresh set of standards was made for the ASTM precision and accuracy studies. Seven replicate injections of four different concentrations across the calibration range were made over a period of about one week, and concentrations were calculated based on the original calibration curve. No additional tuning or optimization of the PFPD was necessary. Precision was measured as the percent relative standard deviation (%RSD) of the calculated concentrations from the seven injections. The %RSDs ranged from 1.4% for the 0.5 ppm standard to 0.6% for the 5.0 ppm standard. Complete data are reported in Table 3. Overlaid chromatograms of the replicate analyses are shown in Figure 5.

Table 3. Calculated Concentrations of in in Standards with Prepared Concentrations of 0.5 ppm, 1.0 ppm, 3.0 ppm, and 5.0 ppm. Precision is Reported as %RSD. Run No. Calculated Concentrations (ppm) Run 1 0.51 0.97 2.98 4.91 Run 2 0.49 0.96 2.92 4.96 Run 3 0.50 0.96 2.97 4.97 Run 4 0.49 0.98 2.98 5.01 Run 5 0.49 0.97 2.97 4.98 Run 6 0.49 0.97 2.96 4.99 Run 7 0.50 0.99 2.99 4.96 Average Conc. 0.50 0.97 2.97 4.97 %RSD 1.4% 1.3% 0.8% 0.6% 0.5 ppm 1.0 ppm 3.0 ppm 5.0 ppm Figure 5. Overlaid PFPD Chromatograms from Seven Replicate Analyses of in at Different Concentrations

Accuracy of the average calculated concentrations from the replicate injections is reported as percent recovery over the calibration range, and is tabulated in Table 4. Table 4. Accuracy of Calculated Concentrations of in Using the PFPD is Reported as Percent Recovery Expected Concentration (ppm) 0.50 1.0 3.0 5.0 Average Calculated Concentration for n = 7 (ppm) 0.50 0.97 2.97 4.97 % Recovery 99.1 97.2 98.9 99.3 The ASTM Standard D4735-02 specifies method criteria for precision in terms of intermediate precision for single laboratory results and reproducibility for results from two or more laboratories following the same method. These criteria are represented as average measured concentrations with standard deviations reported on a mg/kg (ppm) basis. Table 5 compares the single laboratory intermediate precision data from the PFPD in D4735-02 to the data originally reported for D4735-96. The single lab data demonstrate that the PFPD is capable of analytical precision 3 to 4 times better than the FPD used in the original study. Table 5. Single Laboratory Intermediate Precision Data for the OI Analytical Model 5380 PFPD Compared to the Original FPD Data Intermediate Precision Data From D4735-96 Concentration (mg/kg) Intermediate Precision (mg/kg) 0.80 0.040 1.80 0.078 Intermediate Precision Data From D4735-02 Concentration (mg/kg) Intermediate Precision (mg/kg) 0.50 0.010 1.00 0.010 3.00 0.023 5.00 0.031 Ultimate Sensitivity Since the analytical conditions used to validate the new standard included a split ratio of 50:1, there was significant capacity to improve the ultimate sensitivity of the method by simply reducing the split ratio. Although lower concentration ranges are beyond the scope of the official method, the ability of the PFPD to determine low-level ppb ranges was investigated. The lowest calibration test standard with concentration of 0.5 ppm was diluted by a factor of 10 to produce a sensitivity check standard at 0.05 ppm (50 ppb) thiophene. A 1-µL injection of this sensitivity check standard was made using a 10:1 split ratio and produced the chromatogram shown in Figure 6. The chromatographic peak in Figure 6 represents 50 ppb thiophene or approximately 1.9 pg sulfur on column, and it has a signal-to-noise ratio of about 3:1. Thus, the range of Standard D4735-02 can be easily expanded to include thiophene concentrations as low as 50 ppb or less simply by modifying the split ratio.

50 ppb Figure 6. Chromatogram of 50 ppb Acquired Using a 1-µL Injection and a 10:1 Split Ratio. Lower Split Ratios or Larger Injection Volumes Could Facilitate Analysis of in the 15 20 ppb Range. Conclusions The PFPD, the newest generation of flame photometric detectors, has been shown to meet all the performance criteria specified in ASTM Standard D4735-02, and the new modifications were officially approved in 2002. Calibration is simplified by digital processing of the PFPD signal to produce a linear output, and thus, a linear calibration curve, which is equivalent to the more complex log/log plots required by the original standard. Ultimate sensitivity of the method can be lowered by one order of magnitude or more simply by reducing the split ratio. The PFPD s superior stability allows data to be acquired over a period of days or even weeks with no additional detector tuning or optimization required. The PFPD s equimolar response provides a means to measure the concentration of any other unknown sulfur species that may also be present in the matrix.

Carbowax is a registered trademark of Union Carbide Co. Spectranalyzed is a registered trademark of Fisher Scientific Co. P.O. Box 9010 College Station, Texas 77842-9010 Tel: (979) 690-1711 FAX: (979) 690-0440 www.oico.com