Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Similar documents
Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

IDENTIFICATION OF STEROIDS IN COSMETIC PRODUCTS BY TLC AND HPLC 1 02/12/2005 ACM 007 A. THIN LAYER CHROMATOGRAPHY (TLC)

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Modernizing the USP Monograph for Acetaminophen

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

ANALYTICAL METHOD PROCEDURES

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Fast methods for the determination of ibuprofen in drug products

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

CIPAC. CIPAC Free relevant impurities methods:

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

with diode array detection

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

The Benefits of Gravimetric Sample Preparation

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Sulfotepp impurities in Chlorpyrifos EC formulations

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

TEMPLATE FOR AN EXAMPLE STANDARD TEST METHOD

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Quad-Z 215 with Disposable Tips: Implementing the Use of ZipTips for the Purification of Biological Samples

Determination of Alkaloids in Goldenseal Using Agilent Bond Elut Plexa Solid Phase Extraction Sorbent for Cleanup and HPLC-DAD Analysis

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Application. Gas Chromatography February Introduction

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Two-Stage Course-Embedded Determination of Caffeine and Related Compounds by HPLC in Caffeine Containing Food, Beverages and (or) Related Products

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

637. Thiamethoxam. HPLC method

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Chemistry Gas Chromatography: Separation of Volatile Organics

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Determination of Caffeine by HPLC

The Agilent 1290 Infinity II Multi Method Solution

ENVIRONMENTAL analysis

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analysis of Positional Isomers with Agilent Poroshell 120 PFP Columns

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Determination of Trace Cations in Power Plant Waters Containing Morpholine

High Performance Liquid Chromatography (HPLC) of Three Analgesics

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax:

Application Note. Abstract. Authors. Environmental

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

The Analysis of Organophosphate Pesticides by LC/MS Application

Agilent LC & LC-MS Cannabis Testing

Transcription:

Automated Sample Preparation in Quality Control of Eye-Drop Formulation Application Note Authors Bart Tienpont, Maria Rambla Alegre, Frank David, and Pat Sandra Research Institute for Chromatography Kennedypark 26 B-85 Kortrijk Belgium Abstract The Agilent 7696A Sample Prep WorkBench with WeighStation was used in a quality control method for eye-drop formulations. Automated sample preparation included the preparation of calibration standards (four levels) containing two active ingredients, and the preparation of eye-drop formulation samples by weighing and dilution. The two sequences were fully automated. The analyses show excellent repeatability and linearity.

Introduction The Agilent 7696A Sample Prep WorkBench with WeighStation was used for the determination of two active ingredients, dexamethasone and chloramphenicol, in eye-drop formulations. Dexamethasone is one of the most potent corticosteroids; it is 5 to 14 times more potent than prednisolone and 25 to 75 times more potent than cortisone and hydrocortisone. The addition of chloramphenicol, a broad-spectrum antibiotic, to dexamethasone leads to a combination that yields excellent results in inflammation of the anterior uvea (iritis, iridocyclitis). Eye-drop formulations against inflammation can contain one or both ingredients and are typically offered as a viscous aqueous solution. The typical analytical QC procedure consists of 1) preparation of a 4-level calibration series containing the active ingredients in water (aqueous mobile phase) and 2) dilution of an amount of the eye-drop formulation in water (aqueous mobile phase). As illustrated in Table 1A, sample preparation is normally done in volumetric flasks, starting from the preparation of two stock solutions, containing respectively dexamethasone and chloramphenicol. The resulting calibration solutions contain both ingredients in a concentration range between 16 and 64 µg/ml dexamethasone and between 8 and 32 µg/ml chloramphenicol. Samples are typically 25-fold diluted in aqueous mobile phase. After sample preparation, calibration standards and samples are analyzed by HPLC with UV detection. Since the standard solutions are not stable, they are typically prepared each time a series of samples has to be analyzed. In this application note, the automation of the preparation of the calibration standards and the dilution of the samples is described. The general method layout is shown in Table 1B. WeighStation was used to track the exact calibration standard concentrations and to measure sample amounts. Experimental Chemicals Dexamethasone and chloroamphenicol were obtained from Sigma-Aldrich (Beerse, Belgium). Water and acetonitrile (AcCN) were HPLC grade (BioSolve, The Netherlands). Phosphoric acid and sodium hydroxide were from Sigma-Aldrich. Configuration The Agilent 7696A Sample Prep WorkBench was equipped with two Agilent 7693A Automated Liquid Samplers. The front injector contained an enhanced syringe carriage with a 5-µL syringe (p/n G4513-6561). The back injector contained an enhanced syringe carriage with a 5-µL syringe (p/n 5183-314). WorkBench method Resource layout Figure 1 shows the resource layout for the WorkBench. Mobile phase A (.3% phosphoric acid in water, adjusted to ph = 3 with sodium hydroxide) and acetonitrile are placed as Chemical resource in positions 81-15 and 71-72 respectively. Solvent in vials A1 and B1 for front and back tower was water. All vials in which the standard dilutions are prepared are defined as Empty Container. WorkBench method program steps The 7696A Sample Prep WorkBench program for the preparation of the four calibration standards is shown in Figure 2. After weighing the empty vials, the system is placed on pause to allow the addition of solid standards. After this, the automatic process is resumed and solvent is added to give calibration standards at four levels. The lowest level is prepared in six-fold. 2

Table 1A. Classical Sample Preparation (USP or EP Methods) Code Preparation stock solutions Target conc (µg/ml) Calibration standards DEX Stck weigh 1 mg dexamethasone in a 25-mL vial + add 25 ml water 4 CLO Stck weigh 1 mg chloramphenicol in a 25-mL vial + add 25 ml water 4 Preparation standard solutions LVL1 1 ml Dex Stck + 5 ml CLO Stck in 25 ml mobile phase 16/8 LVL2 2 ml Dex Stck + 1 ml CLO Stck in 25 ml mobile phase 32/16 LVL3 3 ml Dex Stck + 15 ml CLO Stck in 25 ml mobile phase 48/24 LVL4 4 ml Dex Stck + 2 ml CLO Stck in 25 ml mobile phase 64/32 Preparation eye-drop formulation samples Samples SAM 1 ml or 1 g eye-drop sample + 25 ml mobile phase Target conc (µg/ml) dexamethasone/chloramphenicol Table 1B. Agilent 7696A Sample Prep WorkBench Sample Preparation Code Preparation stock solutions Target conc (µg/ml) Calibration standards DEX Stck-1 weigh 4 mg dexamethasone in a 2-mL vial * + add 1 ml AcCN 4, DEX Stck 15 µl DEX Stck-1 + 1,35 µl mobile phase (1/1 dilution) 4 CLO Stck-1 weigh 4 mg chloramphenicol in a 2-mL vial * + add 1 ml AcCN 4, CLO Stck 15 µl CLO Stck-1 + 1,35 µl mobile phase (1/1 dilution) 4 Preparation standard solutions LVL1 2 µl Dex Stck + 1 µl CLO Stck + 38 µl mobile phase 16/8 LVL2 4 µl Dex Stck + 2 µl CLO Stck + 26 µl mobile phase 32/16 LVL3 6 µl Dex Stck + 3 µl CLO Stck + 14 µl mobile phase 48/24 LVL4 8 µl Dex Stck + 4 µl CLO Stck + 2 µl mobile phase 64/32 Preparation finished product samples Samples SAM 1 eye-drop sample* + 1 ml mobile phase (approximately 35 4 mg) Target conc (µg/ml) dexamethasone/chloramphenicol *solid powder or viscous liquid is added manually while Agilent 7696A Sample Prep WorkBench is on hold. This is followed by a weighing step. Sample vials Mobile phase (chemical resource) Figure 1. Acetonitrile (chemical resource) STD solution vials (empty vials) Agilent 7696A Sample Prep WorkBench resource layout for the serial dilution of dexamethasone and chloramphenicol. 3

Tare DEX stck-1 and CLO stck-1 vials Press PAUSE on the Agilent 7696A Sample Prep WorkBench and add standard (approximately 4 mg). Press START to resume. Weigh DEX stck-1 and CLO stck-1 vials (accurate standard weighing). Weigh exact amount of added solvent. Dilution Figure 2. Graphical flow chart used to program Agilent 7696A Sample Prep WorkBench method setup for calibration standards. 4

An additional sequence is prepared for the dilution of the samples: empty vials are tarred and the system is put on hold. One drop of viscous sample is added to each vial. This requires a manual action since the viscosities of the samples do not allow transferring the liquids using syringes. The workbench is then resumed. The vials are again weighed (Figure 3) and the exact sample amounts are calculated and reported in the sequence report. Finally, solvent is added for sample dilution (see method lay-out in Figure 4). Both programs for the preparation of the calibration standards and for the preparation of the samples can be programmed in one sequence. Figure 3. Tare Automatic Add 1 eye-drop Manual Weigh Automatic Illustration of taring, sample addition, and weighing of the sample amount. Tare the sample vials 1-minute hold Press PAUSE on the Agilent 7696A Sample Prep WorkBench and add sample (1 drop). Press START to resume. Weigh the samples Add 1 ml of mobile phase Weigh the total mixture 1. Begin a group. 2. Tare the sample. 3. End the group. 4. Wait 1 minute seconds. 5. Press PAUSE on the WorkBench. Press START after adding the sample. 6. Weight sample. 7. Flag the sample as RESULT. 8. Add 1, µl of the mobile phase to the sample at the front tower (washes, pumps). 9. Weigh the sample. 1. Flag the sample as RESULT. Figure 4. Agilent 7696A Sample Prep WorkBench method setup for samples. 5

Experimental conditions for HPLC analysis Analyses were performed on an Agilent 129 Infinity HPLC System. Separation was performed on a Poroshell 12 column (2.1 mm 4 mm, 2.4 µm dp (p/n 695775-92)). Mobile phase A was.3% phosphoric acid in water (adjusted to ph = 3 with sodium hydroxide) and mobile phase B was acetonitrile. A gradient from 2% B ( minutes) to 5% B (5 minutes) was used. Flow rate was.5 ml/min. Injection volume was 1 µl and detection was done by UV at 254 nm. Results and Discussion The chromatograms corresponding to the four calibration levels are overlaid in Figure 5. The calibration curves are given in Figure 6. The linearities for both dexamethasone and chloramphenicol are excellent (R² >.999). 4 35 3 25 2 15 1 5 1 2 3 4 5 min Figure 5. Chromatograms (overlay with X/Y-axis offset) of four calibration levels of chloramphenicol (2.35 minutes) and dexamethasone (3.55 minutes). Area Area 7 6 5 4 3 2 1 16 14 12 1 8 6 4 2 Figure 6. A Chloramphenicol Rel. res% (1): 4.134 Correlation:.99941 1 B Dexamethasone Rel. res% (1): 3.358 Correlation:.9996 1 2 1 2 Amount (ng/µl) 2 2 4 Amount (ng/µl) Calibration curves for chloramphenicol (2.35 minutes) and dexamethasone (3.55 minutes) obtained by automated calibration standard preparation using the Agilent 7696A Sample Prep WorkBench. 3 3 4 4 6

The repeatability of sample preparation was evaluated by preparing the lowest calibration sample in six-fold. The chromatograms are overlaid in Figure 7 and show excellent repeatability, with RSDs of 1.5% for chloramphenicol and 2.7% for dexamethasone, respectively. Next, three types of samples were analyzed. The chromatograms are shown in Figure 8. Sample A contains dexamethasone, sample B contains chloramphenicol and sample C contains a combination of both solutes. For all three samples, correct amounts of the active ingredients were detected (values all between 9% and 15% of labelled concentration). The analyses of different sample types that were prepared in the same sequence also revealed that no crosscontamination was observed (no chloramphenicol in sample only containing dexamethasone and vice versa). 8 Chloramphenicol 6 4 2 Dexamethasone.5 1. 1.5 2. Figure 7. 2.5 3. 3.5 4. 4.5 min Overlay of 6 chromatograms of six individually prepared calibration solutions of chloramphenicol (2.35 minutes) and dexamethasone (3.55 minutes) obtained by automated calibration standard preparation using the Agilent 7696A Sample Prep WorkBench. The percent relative standard deviation for chloramphenicol and dexamethasone are 1.5% and 2.7%, respectively. 4 A 2.5 1. 1.5 2. B 8 6 4 Dexamethasone 2.5 3. 3.5 4. 4.5 min Chloramphenicol 2.5 1. 1.5 2. 2.5 3. 3.5 4. 4.5 min C Dexamethasone 1 7.5 Chloramphenicol 2.5 5.5 1. 1.5 2. 2.5 3. 3.5 4. 4.5 min Figure 8. Chromatograms obtained for three eye-drop formulation types. 7

The repeatability of sample preparation is demonstrated in Table 2, showing the measured concentrations, the sample weights and the final concentrations of chloramphenicol in sample type B. This sample was prepared in 6-fold. The relative standard deviation of the measured amount is lower than 1%. In addition to these quantitative data, note that the total solvent consumption is drastically reduced, no volumetric glassware is used, and hardly any solutions are wasted. Conclusions The Agilent 7696A Sample Prep WorkBench with WeighStation was successfully applied in the automated preparation of calibration samples and sample dilutions. The viscous nature of the samples did not allow volumetric dilution with syringes, but using the weighing station, exact sample masses were determined. After dilution, accurate determination of the active ingredients was possible. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. Table 2. Repeatability of Sample Preparation for Eye-drop Formulation Type B Measured conc (µg/ml) Sample weight (mg) Corrected conc (µg/mg) SAM-rep1 124 27.72 4.59 SAM-rep2 121 26.95 4.61 SAM-rep3 114 25.41 4.61 SAM-rep4 13 23.7 4.57 SAM-rep5 143 32.58 4.53 SAM-rep6 138 31.27 4.53 RSD (%).74 www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 214 February 6, 214 5991-3591EN