IRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL

Similar documents
IRTS-P Precision Infrared Temperature Sensor

CR10XTCR THERMOCOUPLE REFERENCE

CR10TCR THERMOCOUPLE REFERENCE

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

3WHB10K 3-WIRE HALF BRIDGE TERMINAL INPUT MODULE INSTRUCTION MANUAL

CR10XTCR THERMOCOUPLE REFERENCE

MODEL HMP45C TEMPERATURE AND RELATIVE HUMIDITY PROBE INSTRUCTION MANUAL

SP-LITE SILICON PYRANOMETER INSTRUCTION MANUAL

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL

INSTRUCTION MANUAL. CR10XTCR Thermocouple Reference. Revision: 1/03. Copyright (c) Campbell Scientific, Inc.

3WHB10K 3-Wire Half Bridge Terminal Input Module

MET ONE RAIN GAGE MODELS 380, 385, 380M, 385M

CM3 PYRANOMETER INSTRUCTION MANUAL

INSTRUCTION MANUAL. SP Lite2 Pyranometer. August Copyright 2017 Campbell Scientific (Canada) Corp.

WARRANTY AND ASSISTANCE

Model ASPTC Aspirated Shield with Fine Wire Thermocouple

3WHB10K 3-Wire Half Bridge Terminal Input Module

CVD20 20:1 Voltage Divider Revision: 5/12

INSTRUCTION MANUAL. CSI Model TB4 Rain Gage. Revision: 8/04. Copyright (c) Campbell Scientific, Inc.

Model 108 Temperature Probe Revision: 4/07

INSTRUCTION MANUAL. CS500 Temperature and Relative Humidity Probe. Revision: 7/04. Copyright (c) Campbell Scientific, Inc.

SI-111 Precision Infrared Radiometer Revision: 7/13

INSTRUCTION MANUAL. LI200X-LC Pyranometer for MetData1 4/97. Copyright (c) Campbell Scientific, Inc.

WXT510 Weather Transmitter Revision: 4/07

INSTRUCTION MANUAL. CS547 Conductivity and Temperature Probe and A547 Interface. Revision: 9/00. Copyright (c) Campbell Scientific, Inc.

INSTRUCTION MANUAL. Model 207 Temperature and Relative Humidity Probe. Revision: 2/94. Copyright (c) Campbell Scientific, Inc.

INSTRUCTION MANUAL. CS100 Barometric Pressure Sensor 3/05. Copyright (c) Campbell Scientific, Inc.

Model HFP01 Soil Heat Flux Plate Revision: 7/12

INSTRUCTION MANUAL. Model HFP01SC Self-Calibrating Soil Heat Flux Plate Revision: 3/14. Copyright Campbell Scientific, Inc.

Eppley PIR Precision Infrared Radiometer

227 Delmhorst Cylindrical Soil Moisture Block Revision: 5/13

INSTRUCTION MANUAL. CS616 and CS625 Water Content Reflectometers Revision: 2/14. Copyright Campbell Scientific, Inc.

NOVALYNX CORPORATION MODEL 110-WS-16BP BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL

CS100 Barometric Pressure Sensor

LUDLUM MODEL 42-5 NEUTRON BALL CART. August 2015 Serial Number PR and Succeeding Serial Numbers

Amplified Pyranometer SP-212 & 215

4WFB120, 4WFB350, 4WFB1K 4 Wire Full Bridge Terminal Input Modules Revision: 5/07

User s Manual: MODEL 904-T CLINOMETER PAK

Using CSIM11 ph and ORP Probes with Campbell Scientific Dataloggers

WARRANTY AND ASSISTANCE

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

FIELD GUIDE FOR HORUS SCOPES

Instruction Manual SEFRAM Thermal Camera

INSTRUCTION MANUAL. 229 Heat Dissipation Matric Water Potential Sensor Revision: 5/09. Copyright Campbell Scientific, Inc.

HS515G-M. PARALOW Circle Dot Sight. User's Manual. Holosun Technologies Inc. Phone: (909) Fax: (909)

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

Using CSIM11 ph and ORP Probes with Campbell Scientific Dataloggers Revision: 4/10

TB-2012M BAROMETRIC PRESSURE SENSOR USER S MANUAL

Owner's Manual NSL100B BUILDERS LEVEL NSL500B TRANSIT LEVEL

SUGAR/BRIX REFRACTOMETER INSTRUCTION MANUAL

PAR Meter Instruction Manual. Product # Product # Year Warranty

Program Notes Use 2:1 voltage divider on Differential Input channel 1 to get full output range from C7

Melton Series. Owners Manual. Lab Centrifuge. Model # s A C, A C, B C, B C, C C, C C, D C, D C

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL

CAUTION: BEFORE YOU BEGIN, ALWAYS MAKE SURE THE WEAPON IS UNLOADED AND SAFE TO HANDLE.

CHROMA Colorimeter Model 252

Instruction Manual. Kite-22 Camera Crane Starter Package (KITE-22-STARTER) Exclusive of Weights

IF YOU NEED ANY FURTHER HELP WITH YOUR GALILEO TELESCOPE PLEASE CALL MON-FRI 9AM - 5PM E.S.COM

TDR Probes CS605, CS610, CS630, CS635, CS640, CS645 Revision: 9/13

INSTRUCTIONS FOR ATTACHING THE NORDIC PLOW TO THE ACTION TRACKCHAIR

INSTRUCTION MANUAL. BlackGlobe Temperature Sensor for Heat Stress Revision: 1/15. Copyright Campbell Scientific, Inc.

HMP60 Temperature and Relative Humidity Probe Revision: 7/12

PLOW MOUNT KIT FOR POLARIS RANGER P/N ASSEMBLY / OWNERS MANUAL

Driveline Components Catalog Center Bearings

MODEL H3022 MAGNETIC BASE / DIAL INDICATOR / CALIPER COMBO OWNER'S MANUAL

/2" Heavy Duty Swing-Out

Operating Instructions Validation Tool Kit, KT2 For use with USP Apparatus #1 & #2

SK Series BOX PLOW INSTALLATION & OWNER S MANUAL TABLE OF CONTENTS

Small Absolute Pressure Sensor

Field Pro Operation and Installation Manual

HFM-6 and HFM-7 OPERATOR S MANUAL

iweathar EDGE V Installation Guide Copyright 2012 iweathar. All rights reserved.

HEAVY DUTY SWING-OUT VALVE

Small Gage Pressure Sensor

Conversion Kit for: 7620 and 7820 Swing-Out Valves ft-lbs ft-lbs 8835/ ft-lbs

CBRD30CP3 & BOWRD30CP INSTRUCTION MANUAL

WS-7211U Wireless 433 MHz Weather Station. Instruction Manual

QUICK SET UP MANUAL WEATHER CENTER

MKIII -MB USER S GUIDE 09/18/18. Rev. 3

Owner's Manual. for AUTOMATIC LEVEL. NBL Series. FOR CUSTOMER SERVICE, PARTS & REPAIR, CALL Toll Free:

HEAVY DUTY SWING-OUT VALVE

Small, Gauge Pressure Sensor

IR/Thermistor Pen. Instruction Manual

HFM-6, HFM-7, & HFM-8 OPERATOR S MANUAL

Route and retain the clinometer cable so that it does not interfere with moving parts where the unit is mounted.

Instruction Manual HI Refractometer for Invert Sugar Measurements

OEM Silicon Pressure Die

CASSINI CQR-120 OPERATING INSTRUCTIONS

SM98A Harsh Media Backside Absolute Pressure Series

CAUTION: BEFORE INSTALLATION, CONFIRM THAT THE CROSSBOW IS UNLOADED AND IS IN SAFE OPERATING CONDITION.

One-Year Limited Warranty

INSTRUCTION MANUAL ANALOG COLORIMETER

1 YEAR LIMITED WARRANTY TELESCOPES

HR-Z SERIES Mild Radiation Resistance LVDT

DRIFTER PLOW-IN-A-BOX

INSTRUCTION MANUAL Fax: (503)

RP-100 TELESCOPE Instruction Manual

RP-200 TELESCOPE Instruction Manual

WHEN BRIGHTNESS COUNTS

566/568. Infrared Thermometers. Getting Started

Transcription:

IRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL 4/02 COPYRIGHT (c) 2002 CAMPBELL SCIENTIFIC, INC., APOGEE INSTRUMENTS, INC.

This is a blank page.

Warranty and Assistance The IRTS-P PRECISION INFRARED TEMPERATURE SENSOR is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. Batteries have no warranty. CAMPBELL SCIENTIFIC, INC.'s obligation under this warranty is limited to repairing or replacing (at CAMPBELL SCIENTIFIC, INC.'s option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CAMPBELL SCIENTIFIC, INC. CAMPBELL SCIENTIFIC, INC. will return such products by surface carrier prepaid. This warranty shall not apply to any CAMPBELL SCIENTIFIC, INC. products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CAMPBELL SCIENTIFIC, INC. is not liable for special, indirect, incidental, or consequential damages. Products may not be returned without prior authorization. To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 753-2342. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. CAMPBELL SCIENTIFIC's shipping address is: CAMPBELL SCIENTIFIC, INC. RMA# 815 West 1800 North Logan, Utah 84321-1784 CAMPBELL SCIENTIFIC, INC. does not accept collect calls. Non-warranty products returned for repair should be accompanied by a purchase order to cover the repair. 815 W. 1800 N. Logan, UT 84321-1784 USA Phone (435) 753-2342 FAX (435) 750-9540 www.campbellsci.com Campbell Scientific Canada Corp. 11564-149th Street Edmonton, Alberta T5M 1W7 CANADA Phone (780) 454-2505 FAX (780) 454-2655 Campbell Scientific Ltd. Campbell Park 80 Hathern Road Shepshed, Loughborough LE12 9GX, U.K. Phone +44 (0) 1509 601141 FAX +44 (0) 1509 601091

This is a blank page.

IRTS-P Table of Contents 1. General Description...1 2. Specifications...1 3. Installation...2 4. Wiring...3 5. Example Programs...3 6. Maintenance...10 Figures 3-1. IRTS-P on UT018 Crossarm...2 3-2. Mounting IRTS-P to UT018 Crossarm...2 Tables 5-1. Wiring for Example Programs...4 i

This is a blank page.

IRTS-P Precision Infrared Temperature Sensor 1. General Description An infrared temperature sensor (IRTS) is a non-contact means of measuring the surface temperature of an object by sensing the infrared radiation given off. IRTS are widely used for measurements of leaf, canopy, and average surface temperature. With contact sensors it is difficult to avoid influencing the temperature, maintain thermal contact, and provide a spatial average. By mounting the infrared sensor at an appropriate distance from the target, it can be used to measure an individual leaf, a canopy, or any surface of interest. The IRTS-P is an infrared temperature sensor calibrated to output the signal for the target temperature with the same output voltage as if a Type K thermocouple were sensing the target temperature. A separate type K thermocouple is used to measure the temperature of the sensor body. The sensor body temperature is used to correct the target temperature for greater accuracy. 2. Specifications Power Requirements None: self-powered Accuracy ± 0.2 C from 15 to 35 C, ± 0.3 C from 5 to 45 C ( ± 0.1 C when sensor body and target are at the same temperature) Repeatability 0.05 C from 15 to 35 C Mass Less than 100 grams Dimensions 6.3 cm long by 2.3 cm diameter Response Time Less than 1 second to changes in target temperature Output Signal 2, type K, twisted, shielded pair thermocouple outputs (15 ft each), one for target temperature, one for sensor body temperature. The sensor body temperature is used to make a correction for target temperature. Optics Silicon lens Wavelength Range 6 to 14 micrometers Field of View 3:1 field of view (at 3 meters from sensor the Operating Environment FOV is a 1 meter diameter circle) Highly water resistant, designed for continuous outdoor use; temperature range: 0 to 50 C 1

3. Installation The field of view for infrared sensors is calculated based on the geometry of the sensor and lens. However, optical and atmospheric scatter and unwanted reflections from outside the field of view may influence the measurement. Under typical conditions, 80 to 90 percent of the IR signal is from the field of view and 10 to 20 percent is from the area surrounding the field of view. If the target surface is small, for example a single leaf, try to mount the sensor close enough that the surface extends beyond the field of view. To obtain the desired view of the canopy or surface of interest, the IRTS-P is often mounted separately from a met station tower. A hole threaded for a standard tripod camera mount screw (1/4 inch diameter; 20 threads per inch) can be used to mount the sensor to a user-supplied support. The IRTS-P can be mounted to a Campbell Scientific tripod or tower with the UT018 Crossarm (Figure 3-1.) A mounting kit (Stainless steel bolt, 2 washers, and 2 nuts PN 14475) is included with the IRTS-P that can be used to attach the sensor to the crossarm (Figure 3-2.) FIGURE 3-1. IRTS-P on UT018 Crossarm Tighten against sensor to fix angle of sensor. Nut with two lock washers attaching bolt to the arm. FIGURE 3-2. Mounting IRTS-P to UT018 Crossarm 2

4. Wiring The IRTS-P has two thermocouple outputs in separate shielded cables. The cable with the black band on the jacket near the wires is the output from the detector. The cable without a black band is a thermocouple measuring the temperature of the sensor body. We recommend measuring the temperatures with differential voltage thermocouple measurements. 5. Example Programs The datalogger program to measure the IRTS-P measures the thermocouple outputs to obtain the IRTS-P sensor body temperature and the apparent (uncorrected) temperature of the target. The thermocouple temperature requires the temperature of the terminals to which the thermocouples are connected. The panel temperature is used as the reference for the CR23X and CR5000. The module temperature is not an accurate measurement of the CR10X panel temperature; a CR10XTCR is required to measure the reference temperature. After measuring the thermocouple outputs, the sensor body temperature is used to calculate correction coefficients that are then used to correct the target temperature. All three example programs measure the sensor once a second and output average values once an hour. The actual channels and outputs intervals need to be adjusted for the actual installation and application. The equations implemented in the program are (Bugbee et.al. 1996) Corrected Target Temperature, CTT = ATT SEC were ATT = Apparent Target Temperature, and Sensor Error Correction, 2 SEC = (0.25/ P )(( ATT H ) K ) ; SB Black Band Detector Temperature: Yellow TC + Differential High Red TC Differential Low Bare - Ground Sensor Body Temperature: Yellow TC + Differential High Red TC Differential Low Bare - Ground SB SB P SB = 26.168 + 2.8291 SB 0.03329 SB 2 3

H SB = 5.8075 0.08016 SB +.00849 SB 2 K SB = 85.943 + 11.740 SB + 0.08477 SB 2 and SB = Sensor Body Temperature The three example programs measure the sensor once a second and output average sensor and target temperatures once an hour. In an actual installation the measurement and output intervals, and channels used may be changed to accommodate the other measurements and outputs. TABLE 5-1. Wiring for Example Programs Sensor/Lead Description CR10X CR23X CR5000 CR10XTCR Reference Temp: Not Used Not Used Black Excitation E3 Red Signal SE1 Clear Ground AG IRTS-P Black Band Detector Temp Yellow TC + 2H 1H 1H Red TC - 2L 1L 1L Bare wire Shield G IRTS-P No Band Sensor Temp Yellow TC + 3H 2H 2H Red TC - 3L 2L 2L Bare wire Shield G CR10X Example Program ;{CR10X} ; *Table 1 Program 01: 1 Execution Interval (seconds) ;Measure CR10XTCR Reference Temperature 1: Temp (107) (P11) 2: 1 SE Channel 3: 3 Excite all reps w/e3 4: 1 Loc [ RefTemp ] 5: 1.0 Mult 6: 0.0 Offset ;Measure IRt/c apparent target temperature 4

2: Thermocouple Temp (DIFF) (P14) 2: 21 2.5 mv 60 Hz Rejection Range 3: 2 DIFF Channel 4: 3 Type K (Chromel-Alumel) 5: 1 Ref Temp (Deg. C) Loc [ RefTemp ] 6: 2 Loc [ AppTargT ] 7: 1.0 Mult 8: 0.0 Offset ;Measure Sensor Body Temp 3: Thermocouple Temp (DIFF) (P14) 2: 21 2.5 mv 60 Hz Rejection Range 3: 3 DIFF Channel 4: 3 Type K (Chromel-Alumel) 5: 1 Ref Temp (Deg. C) Loc [ RefTemp ] 6: 3 Loc [ SenBodyT ] 7: 1.0 Mult 8: 0.0 Offset ;Calculate P, H, & K Coefficients 4: Polynomial (P55) 2: 3 X Loc [ SenBodyT ] 3: 4 F(X) Loc [ Psb ] 4: -18.35 C0 5: 7.0435 C1 6: -0.12 C2 7: 0.0 C3 8: 0.0 C4 9: 0.0 C5 5: Polynomial (P55) 2: 3 X Loc [ SenBodyT ] 3: 5 F(X) Loc [ Hsb ] 4: 21.15 C0 5: -1.12 C1 6: 0.0285 C2 7: 0.0 C3 8: 0.0 C4 9: 0.0 C5 6: Polynomial (P55) 2: 3 X Loc [ SenBodyT ] 3: 6 F(X) Loc [ Ksb ] 4: -499.19 C0 5: 39.97 C1 6: -.5057 C2 7: 0.0 C3 8: 0.0 C4 9: 0.0 C5 5

;Calculate correction factor (SEC) 7: Z=1/X (P42) ; {1/Psb} 1: 4 X Loc [ Psb ] 2: 4 Z Loc [ Psb ] 8: Z=X*F (P37) ; {.25/Psb} 1: 4 X Loc [ Psb ] 2: 0.25 F 3: 4 Z Loc [ Psb ] 9: Z=X-Y (P35) ; {ATT - Hsb} 1: 2 X Loc [ AppTargT ] 2: 5 Y Loc [ Hsb ] 3: 5 Z Loc [ Hsb ] 10: Z=X*Y (P36) ; {ATT - Hsb}^2 1: 5 X Loc [ Hsb ] 2: 5 Y Loc [ Hsb ] 3: 5 Z Loc [ Hsb ] 11: Z=X-Y (P35) ; {subtract Ksb} 1: 5 X Loc [ Hsb ] 2: 6 Y Loc [ Ksb ] 3: 6 Z Loc [ Ksb ] 12: Z=X*Y (P36) ; {calculate SEC} 1: 4 X Loc [ Psb ] 2: 6 Y Loc [ Ksb ] 3: 7 Z Loc [ SEC ] ;Calculate corrected target temperature (CTT) 13: Z=X-Y (P35) ; 1: 2 X Loc [ AppTargT ] 2: 7 Y Loc [ SEC ] 3: 8 Z Loc [ CTT ] 14: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 10 Set Output Flag High (Flag 0) 15: Real Time (P77) 1: 1220 Year,Day,Hour/Minute (midnight = 2400) 16: Average (P71) 1: 3 Reps 2: 1 Loc [ RefTemp ] 17: Average (P71) 2: 8 Loc [ CTT ] 6

*Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines End Program -Input Locations- 1 RefTemp 1 2 1 2 AppTargT 1 2 1 3 SenBodyT 1 3 1 4 Psb 1 3 3 5 Hsb 1 4 3 6 Ksb 1 2 2 7 SEC 1 1 1 8 CTT 1 0 1 CR23X Example Program ;{CR23X} ; *Table 1 Program 01: 1 Execution Interval (seconds) ;Measure Panel Reference Temperature 1: Panel Temperature (P17) 1: 1 Loc [ RefTemp ] ;Measure IRt/c apparent target temperature 2: Thermocouple Temp (DIFF) (P14) 2: 21 10 mv, 60 Hz Reject, Slow Range 3: 1 DIFF Channel 4: 3 Type K (Chromel-Alumel) 5: 1 Ref Temp (Deg. C) Loc [ RefTemp ] 6: 2 Loc [ AppTargT ] 7: 1.0 Mult 8: 0.0 Offset ;Measure Sensor Body Temp 3: Thermocouple Temp (DIFF) (P14) 2: 21 10 mv, 60 Hz Reject, Slow Range 3: 2 DIFF Channel 4: 3 Type K (Chromel-Alumel) 5: 1 Ref Temp (Deg. C) Loc [ RefTemp ] 6: 3 Loc [ SenBodyT ] 7: 1.0 Mult 8: 0.0 Offset 7

;Calculate P, H, & K Coefficients 4: Polynomial (P55) 2: 3 X Loc [ SenBodyT ] 3: 4 F(X) Loc [ Psb ] 4: -18.35 C0 5: 7.0435 C1 6: -0.12 C2 7: 0.0 C3 8: 0.0 C4 9: 0.0 C5 5: Polynomial (P55) 2: 3 X Loc [ SenBodyT ] 3: 5 F(X) Loc [ Hsb ] 4: 21.15 C0 5: -1.12 C1 6: 0.0285 C2 7: 0.0 C3 8: 0.0 C4 9: 0.0 C5 6: Polynomial (P55) 2: 3 X Loc [ SenBodyT ] 3: 6 F(X) Loc [ Ksb ] 4: -499.19 C0 5: 39.97 C1 6: -.5057 C2 7: 0.0 C3 8: 0.0 C4 9: 0.0 C5 ;Calculate correction factor (SEC) 7: Z=1/X (P42) ; {1/Psb} 1: 4 X Loc [ Psb ] 2: 4 Z Loc [ Psb ] 8: Z=X*F (P37) ; {.25/Psb} 1: 4 X Loc [ Psb ] 2: 0.25 F 3: 4 Z Loc [ Psb ] 9: Z=X-Y (P35) ; {ATT - Hsb} 1: 2 X Loc [ AppTargT ] 2: 5 Y Loc [ Hsb ] 3: 5 Z Loc [ Hsb ] 10: Z=X*Y (P36) ; {ATT - Hsb}^2 1: 5 X Loc [ Hsb ] 2: 5 Y Loc [ Hsb ] 3: 5 Z Loc [ Hsb ] 8

11: Z=X-Y (P35) ; {subtract Ksb} 1: 5 X Loc [ Hsb ] 2: 6 Y Loc [ Ksb ] 3: 6 Z Loc [ Ksb ] 12: Z=X*Y (P36) ; {calculate SEC} 1: 4 X Loc [ Psb ] 2: 6 Y Loc [ Ksb ] 3: 7 Z Loc [ SEC ] ;Calculate corrected target temperature (CTT) 13: Z=X-Y (P35) ; 1: 2 X Loc [ AppTargT ] 2: 7 Y Loc [ SEC ] 3: 8 Z Loc [ CTT ] 14: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 10 Set Output Flag High (Flag 0) 15: Real Time (P77) 1: 1220 Year,Day,Hour/Minute (midnight = 2400) 16: Average (P71) 1: 3 Reps 2: 1 Loc [ RefTemp ] 17: Average (P71) 2: 8 Loc [ CTT ] *Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines End Program -Input Locations- 1 RefTemp 1 3 1 2 AppTargT 1 2 1 3 SenBodyT 1 3 1 4 Psb 1 3 3 5 Hsb 1 4 3 6 Ksb 1 2 2 7 SEC 1 1 1 8 CTT 1 1 1 9

CR5000 Example Program Public RefT, TargetCorrT, TC(2), SensorCorr, Psb, Hsb, Ksb Alias TC(1) = TargetApparentT Alias TC(2) = SensorBodyT DataTable (AllDat,1,-1) DataInterval (0,60,Min,10) Average (1,RefT,FP2,0) Average (2,TC(),IEEE4,0) Average (1,TargetCorrT,IEEE4,0) EndTable BeginProg Scan (1,Sec,3,0) PanelTemp (RefT,250) TCDiff (TC(),2,mV20C,1,TypeK,RefT,True,0,_60Hz,1.0,0) Psb = 26.168 + 2.8291 * SensorBodyT - 0.03329 * SensorBodyT^2 Hsb = 5.8075-0.08016 * SensorBodyT + 0.00849 * SensorBodyT^2 Ksb = -85.943 + 11.740 * SensorBodyT + 0.08477 * SensorBodyT^2 SensorCorr = (0.25/Psb) * ((TargetApparentT - Hsb)^2 - Ksb) TargetCorrT = TargetApparentT - SensorCorr CallTable Alldat NextScan EndProg 6. Maintenance As with any optical sensor, it is important to keep the lens and view clean. Otherwise the sensor will be measuring the temperature of the obstruction instead of the surface of interest. Clean the lens gently with a moistened cotton swab. Distilled water or alcohol works well for most dust/dirt. Salt deposits dissolve better in a weak acid solution (~0.1 molar). 10