Product data sheet Palas Nanoparticle measurement system DEMC 2000

Similar documents
Product data sheet Palas U-SMPS 1050 / 1100 / 1200

NANOPARTICLE MEASUREMENT SYSTEMS

ELECTROSTATIC CLASSIFIER MODEL 3082

Rapid Measurements of Aerosol Size Distributions Using a Fast Integrated Mobility Spectrometer (FIMS)

NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER

Mobility-Based Particle Size Classification and Detection down to 2nm Theory to Practice

ISO INTERNATIONAL STANDARD. Determination of particle size distribution Differential electrical mobility analysis for aerosol particles

The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch. Symposium Stockholm Mai

Brian Sargent, Assistant Director, Designee 3/21/18. See Thread for Approval

Jianfei Peng et al. Correspondence to: Jianfei Peng Min Hu and Renyi Zhang

A new versatile condensation particle counter for research and environmental monitoring

Differential Mobility Particle Sizer (Aerosol measurements)

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

Introduction and Initial Field Data of a Novel, Portable Nanoparticle Sizing Instrument

SOP SMPS BaltimorePM supersite Version /29/01 Page 1 of 13 SMPS

New Instruments from GRIMM

Comprehensive Measurement of Atmospheric Aerosols with a Wide Range Aerosol Spectrometer

Intercomparison of Mobility Particle Size Spectrometers

Measuring aerosol size distributions with the fast integrated mobility spectrometer

Causes of Concentration Differences Between a Scanning Mobility Particle Sizer and a Condensation Particle Counter

Particle counting efficiencies of new TSI condensation particle counters

CHAPTER 5 CONVECTIVE HEAT TRANSFER COEFFICIENT

Calibration checks of particle counter using primary and other techniques at the laboratory level

Model 2300XP PSL & Process-Particle Wafer Deposition System

Particle Size Distribution Measurements with the Novel 1 nm-smps

Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping

COUNTS SURFACE VOLUME MASS COLLECTION GENERATION

Defined calibration of the particle measuring system according to PMP

Intercomparison of Mobility Particle Size Spectrometers

Intercomparison of Mobility Particle Size Spectrometers

Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator.

Nanoparticle Trajectories In An Electrostatic Precipitator: Numerical Simulation And Experimental Validation

MEASURING NANOPARTICLE SIZE DISTRIBUTIONS IN REAL-TIME: KEY FACTORS FOR ACCURACY

Intercomparison of Mobility Particle Size Spectrometers

MEASURING NANOPARTICLE EXPOSURE

Multi-Instrument Manager Tool for Data Acquisition and Merging of Optical and Electrical Mobility Size Distributions

Experimental Investigation of the Collection Efficiencies of the AMS Aerodynamic Focusing Lens

A method of testing size resolution of a particle sizing instrument Toward traceable particle measurements

Intercomparison of Mobility Particle Size Spectrometers

Supplementary information

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

Clean Corona Ionization Technology. Peter Gefter Simco-Ion phone: (1)

Evaluation of the NanoAerosol Generator by Kanomax FMT Inc. in Aerosolization of Size Standard Nanoparticle and Protein

Measuring sub-50nm particle retention of UPW filters

CTC 155 / 350 / 660 Dry-block calibrator

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

Coating K = TFE wear resist, dry lubricant Kerkote X = Special coating, (Example: Kerkote with grease)

Generation and Evaluation of Monodisperse Sodium Chloride and Oleic Acid Nanoparticles

Nanoparticle Exposure in the Workplace

The Elzone II Particle Count and Size Analyzer

Improving the Nanoparticle Resolution of the ELPI

(manufactured) nano aerosols

Measuring Particle Mass and Other Properties with the Couette Centrifugal Particle Mass Analyzer

Fundamentals of Particle Counting

UFP instrument comparison at an urban background location in Antwerp

473-SHX Dew Point Mirror

Minimum (-31) every 10 s -35 (-31)

Helsinki, Finland. 1 Aerodyne Research Inc., Billerica, MA, USA. 2 Chemistry Department, Boston College, Chestnut Hill, MA, USA

The Scanning DMA Transfer Function

Site audit report Birkenes, Norway

Chapter 5: Nanoparticle Production from Cathode Sputtering. in High-Pressure Microhollow Cathode and Arc Discharges

Approval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Recent progress and future prospects for traceable submicron aerosol measurements

Aerosol size distribution estimation and associated uncertainty for measurement with a Scanning Mobility Particle Sizer (SMPS)

Bipolar Neutralization using Radioactive, X-ray, and AC Corona Methods

LABOTRON EXTRACTION & SYNTHESIS 2450 MHz

CONTINUOUS FINE DUST MEASURING SYSTEM

NANOTECHNOLOGY SAMPLING AND CONDITIONING

Performance of a New Commercial Electrical Mobility Spectrometer

Titration of Acids and Bases

JAPAN QUALITY. Automated Clinical Analyzer


High Temperature Condensation Particle Counter (HT- CPC)

DRS 61: Incremental encoders, number of lines and zero pulse width freely programmable DRS 60: Incremental Encoders with Zero-Pulse-Teach

SIFT-MS. technology overview

TSI PARTICLE TECHNOLOGY

Certificate: / 29 April 2014

Mobility particle size spectrometers: Calibration procedures and measurement uncertainties

BOXA-II ON-STREAM X-RAY FLUORESCENCE ANALYZER. Technical Description Version 2 BGRIMM AUTOMATION

473 Dew Point Hygrometer

Software BioScout-Calibrator June 2013

CHAPTER 7 NUMERICAL MODELLING OF A SPIRAL HEAT EXCHANGER USING CFD TECHNIQUE

Dewpoint Meter Data Sheet / Instructions

Differential Mobility Analysis of Aerosols: A Tutorial

Dimensional drawing face mount flange radial. R = bending radius min. 40 mm. Dimensional drawing face mount flange axial

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS.

ALASTAIR MCLACHLAN Applications Cameron

TSI PARTICLE TECHNOLOGY

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

Corona Discharge Characteristics and Particle Losses in a Unipolar Corona-needle Charger Obtained through Numerical and Experimental Studies

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES

Calibration and Modeling of PMP compliant Condensation Particle Counters

Data Sheet / Instructions. Dewpoint MeterO

RS-C Flexible Reticle Stocker

UPDATED INVERSION MATRICES FOR ENGINE EXHAUST PARTICLE SIZER (EEPS ) SPECTROMETER MODEL 3090

Explosive Screening by Vapour Detection: Operational Aspects

CAL GAUSS CALIBRATION MAGNET

JAPAN QUALITY. Automated Clinical Analyzer

473 Dew Point Hygrometer

Transcription:

Product data sheet Palas Nanoparticle measurement system DEMC 2000 Applications Aerosol science Calibration of condensation particle counters (CPC) Monodisperse particle source System component of an SMPS

Benefits The user is able to select any size within the defined size range. The DEMC can be connected to many counters to form an SMPS. Continuous and fast-scanning principle of measurement Graphic display of measurement values Intuitive operation using 7" touchscreen and GUI Integrated data logger Low maintenance Reliable function Reduces your operating expenses

Description The Palas DEMC is available in two versions. The version with the long classifying column (2000 model) is able to classify particles in the size range of 8 to 1,400 nm. The DEMC size classifier (as defined in ISO 15900) selects aerosol particles based on their electrical mobility and directs them to the outlet. It is also often referred to as a DMA. The Palas DEMC universally connects to CPCs and aerosol electrometers from other manufacturers. Several counters are already supported (see Fig. 3). We will also integrate your counter into the software upon request. In combination with a polydisperse particle source*, a DEMC is used to obtain a very narrow (monodisperse) particle size distribution for nanoparticles of a certain size. Accurate size determination and reliable performance of the DEMC are extremely important especially for calibration set-ups. The size is adjusted by entering it directly on the touch screen (in nm) or by using the arrow buttons to increase or decrease the size. If the DEMC is used as a component of an SMPS system, it provides continuous and rapid scanning of the particle size distribution of the aerosol. Based on the user settings, a scan can be performed in as few as 30 seconds or in up to 64 size channels per decade. The user controls the DEMC using a graphical user interface, which provides linear and logarithmic display of measurement values and data management of the integrated data logger. The software provides sophisticated data evaluation (extensive statistics and averaging) and export capabilities. The DEMC is typically operated as a standalone device, but can also be connected to a computer or network using various interfaces (USB, LAN, WLAN, RS-232/485). * Additional information can be found in the product data sheets for the aerosol generators by Palas, e.g. the DNP 2000, RBG 1000 or AGF 2.0. Function Figure 1 presents the principle of operation of the DEMC. The aerosol is conditioned before it enters the DEMC column. A dryer (e.g. silica gel, Nafion) removes moisture from the particles. A bipolar neutralizer (e.g. Kr 85) is used to ensure a defined charge distribution of the aerosol. An impactor at the inlet of the DEMC is required in order to remove particles larger than the classifier size range. Fig. 1: Principle of operation of the DEMC The aerosol is then directed into the DEMC column via the inlet. The aerosol flow along the outer electrode is carefully combined here with a sheath air flow. It is important to avoid any turbulence here in order to ensure laminar flow. The surfaces of the electrodes must be of extremely high quality with respect to smoothness and tolerances. This sheath air is a dry, particle-free carrier gas (typically air) with a higher volume than the aerosol that is continuously circulated in a closed loop. The sheath air to sample air volume ratio defines the transfer function and thus the resolution capacity of the DEMC. A radially symmetric electric field is generated between the inner and outer electrode by applying voltage. The inner electrode is positively charged with a small slit at the end. By balancing the electrical force on each particle with its aerodynamic drag force in the

electrical field, negatively charged particles are diverted to the positive electrode. Particles with appropriate electrical mobility pass through the slit and exit the DEMC. These classified particles with the same electrical mobility are then available for use downstream. If the DEMC is used as a component of an SMPS system, then the voltage and thus the electrical field change continuously. Particles of varying mobility exit the DEMC and are counted consecutively by a nanoparticle counter and a condensation particle counter (e.g. Palas UF-CPC) or an aerosol electrometer (e.g. Palas Charme ). The tested and optimized Palas software combines the data (voltage, particle number, etc.) in order to obtain a particle size distribution, as shown in Figure 2. User interface and software Based on continuous customer feedback, the user interface and software have been designed for intuitive operation and real-time control and display of measurement data and parameters. In addition, the software provides data management with the integrated data logger, sophisticated export capabilities, and network support. The measured data are able to be displayed and evaluated with many available options. The DEMC software and firmware support the use of nanoparticle counters by other manufacturers. Figure 3 presents an example. Fig. 3: Screenshot of DEMC counter selection Fig. 2: Particle size distributions of an aerosol generated by the Palas DNP 3000 particle generator

Specifications Size channels Up to 256 (128/decade) User interface Touch screen, 800 480 pixels, 7" Data logger storage Software Classifying range (size) Adjustment range (voltage) Volume flow (sheath air) Impactor 4 GB PDAnalyze 8-1489 nm 1-10000 V 2,5-14 l/min Nozzle for 3 different cut-offs