Response measurement of various neutron dose equivalent monitors in MeV neutron fields

Similar documents
Energy response for high-energy neutrons of multi-functional electronic personal dosemeter

Recent activities in neutron standardization at NMIJ/AIST

Estimation of Radioactivity and Residual Gamma-ray Dose around a Collimator at 3-GeV Proton Synchrotron Ring of J-PARC Facility

Benchmark Test of JENDL High Energy File with MCNP

ARTICLE. II. Method 1. Detector Figure 1 shows a schematic view of the spectrometer. The spectrometer consists of two 2-inch-diameter BF 3

Application and Validation of Event Generator in the PHITS Code for the Low-Energy Neutron-Induced Reactions

Neutron and Gamma-ray Emission Double Dierential Cross Sections. *5 Energy Conversion Engineering, Kyushu University, Kasuga-koen, Kasuga-shi 816.

Measurement of 40 MeV Deuteron Induced Reaction on Fe and Ta for Neutron Emission Spectrum and Activation Cross Section

Measurements and Calculations of Neutron Energy Spectra Behind Polyethylene Shields Bombarded by 40- and 65-MeV Quasi-Monoenergetic Neutron Sources

Integral Benchmark Experiments of the Japanese Evaluated Nuclear Data Library (JENDL)-3.3 for the Fusion Reactor Design

Progress in Nuclear Science and Technology, Volume 6,

ARTICLE. Progress in Nuclear Science and Technology Volume 4 (2014) pp Yoshiko Harima a*, Naohiro Kurosawa b and Yukio Sakamoto c

1.E Neutron Energy (MeV)

Fluence-to-Dose Conversion Coefficients for Muons and Pions Calculated Based on ICRP Publication 103 Using the PHITS Code

Analyses of shielding benchmark experiments using FENDL-3 cross-section data starter library for ITER and IFMIF applications

Altitude Variation of cosmic-ray neutron energy spectrum and ambient dose equivalent at Mt.Fuji in Japan

Author(s) Iwamoto, Y; Hagiwara, M; Nishiizumi Shibata, S. Citation Radiation protection dosimetry (201.

Recent Activities on Neutron Calibration Fields at FRS of JAERI

Fast-Neutron Production via Break-Up of Deuterons and Fast-Neutron Dosimetry

Neutronics Experiments for ITER at JAERI/FNS

Title of project: Influence of environmental parameters on secondary neutrons from cosmic rays at high altitudes in Alpine region

Sequential Measurements of Cosmic-Ray Neutron Energy Spectrum and Ambient Dose Equivalent on the Ground

Measurement of activation of helium gas by 238 U beam irradiation at about 11 A MeV

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup

Dosimetric Quantities and Neutron Spectra Outside the Shielding of Electron Accelerators

Characterization of the 3 MeV Neutron Field for the Monoenergetic Fast Neutron Fluence Standard at the National Metrology Institute of Japan

Status of the AGS Experiment for Mercury Spallation

Test of the rem-counter WENDI-II from Eberline in different energy-dispersed neutron fields

Analytical Functions to Predict Cosmic-Ray Neutron Spectra in the Atmosphere

Recent Developments of the PHITS code

RESPONSE MATRIX OF A BSS/ 6 LiI(Eu) Hector Rene Vega-Carrillo, Eduardo Manzanares-Acuña. Mexico. Eduardo Gallego, Alfredo Lorente.

External MC code : PHITS

Invited. Recent progress of nuclear data measurement in Japan. 1 Introduction. 2 Capture and fission cross sections

Comparison with simulations to experimental data for photoneutron reactions using SPring-8 Injector

Radiation Protection Dosimetry Advance Access published October 28, Radiation Protection Dosimetry (2010), pp. 1 8

BULK-I. Radiation Shielding Tool for Proton Accelerator Facilities

COMPARISON OF COMPUTER CODES APPLICABILITY IN SHIELDING DESIGN FOR HADRON THERAPY FACILITIES *

In our recent paper [1], applicability of the self-activation of an NaI scintillator has been studied for the photo-neutron monitoring around

Study on Radiation Shielding Performance of Reinforced Concrete Wall (2): Shielding Analysis

JRPR. Measurement of Neutron Production Doubledifferential Cross-sections on Carbon Bombarded with 430 MeV/Nucleon Carbon Ions.

Status of standard data on shielding calculation in standard committee of the Atomic Energy Society of Japan

Estimate of Photonuclear Reaction in a Medical Linear Accelerator Using a Water-Equivalent Phantom

Calculations of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials*

Development of Neutron Detectors and Measurement of Cosmic-Ray Neutrons

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

Measurement of Neutron-Production Double-Differential Cross Sections for Nuclear Spallation Reaction Induced by 0.8, 1.5 and 3.

Bonner Sphere Spectrometer. Cruzate, J.A.; Carelli, J.L. and Gregori, B.N.

NEUTRON H*(10) INSIDE A PROTON THERAPY FACILITY: COMPARISON BETWEEN MONTE CARLO SIMULATIONS AND WENDI-2 MEASUREMENTS

Neutronic studies on a pulsed thermal neutron source based on the Be(p,n) reaction by using a compact proton accelerator

Validation of the UFS Bonner Sphere Spectrometer and Monte Carlo Methods at the CERN-EU high energy Reference Field (CERF)

Radiation damage calculation in PHITS

Some Comments to JSSTDL-300

Monte Carlo Simulation concerning Particle Therapy

Measurements of Neutron Capture Cross Sections for 237, 238 Np

Tadafumi Sano, Jun-ichi Hori, Yoshiyuki Takahashi, Hironobu Unesaki, and Ken Nakajima

arxiv: v1 [physics.ins-det] 9 Apr 2018

An introduction to Neutron Resonance Densitometry (Short Summary)

Absolute measurements of the response function of an NE213 organic liquid scintillator for the neutron energy range up to 206 MeV

Energy Response Characteristics of Several Neutron Measuring Devices Determined By Using the Scattered Neutron Calibration Fields of KAERI

Development of a quasi-monoenergetic neutron field using the Li(p, n) Be reaction in the MeV energy range at RIKEN

Measurement of induced radioactivity in air and water for medical accelerators

Neutron Energy and Time-of-flight Spectra Behind the Lateral Shield of a High Energy Electron Accelerator Beam Dump, Part I: Measurements

Monte Carlo Calculations Using MCNP4B for an Optimal Shielding Design. of a 14-MeV Neutron Source * James C. Liu and Tony T. Ng

Nuclear data evaluation of 206 Pb for proton- and neutron-induced reaction in energy region from 20 to 200 MeV

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Performance Test of the Electronic Personal Neutron Dosemeter in Neutron Fields Simulating Workplaces of MOX Fuel Fabrication Facilities

Calculation of Effective Dose Conversion Coefficients for Electrons

Development of displacement damage model in PHITS and comparison with other codes in a high-energy region

Beam Extraction by the Laser Charge Exchange Method Using the 3-MeV LINAC in J-PARC )

RADIOLOGICAL IMPACT OF THE TRIGAACCELERATOR-DRIVEN EXPERIMENT (TRADE)

A Practical Method for Evaluating the Neutron Dose

Light ion production in 175 MeV quasi mono-energetic neutron induced reactions on carbon, oxygen, and silicon

NEW ARRANGEMENT OF COLLIMATORS OF J-PARC MAIN RING

Status of J-PARC Transmutation Experimental Facility

ICRP Symposium on the International System of Radiological Protection

Nuclear Cross-Section Measurements at the Manuel Lujan Jr. Neutron Scattering Center

Fission Fragment characterization with FALSTAFF at NFS

Elastic Recoil Detection Method using DT Neutrons for Hydrogen Isotope Analysis in Fusion Materials. Abstract

TRADE Preliminary comparison of codes for the activation calculation of the water cooling the target

Radiation Protection At Synchrotron Radiation Facilities

Felix C. Difilippo. Oak Ridge National Laboratory P.O. Box 2008 Oak Ridge, TN 3783 l-6363 USA

Estimation of neutron and gamma radiation doses inside the concrete shield wall for 10 and 15 MV medical linear accelerators

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

Occupational Radiation Protection at Accelerator Facilities: Challenges

Recent Activities on Neutron Standardization at the Electrotechnical Laboratory

SINBAD Benchmark Database and FNS/JAEA Liquid Oxygen TOF Experiment Analysis. I. Kodeli Jožef Stefan Institute Ljubljana, Slovenia

Intercomparison between Monte Carlo codes and experiments for induced activities due to high energy electrons

Benchmark Experiments of Accelerator Driven Systems (ADS) in Kyoto University Critical Assembly (KUCA)

Neutron cross section measurement of MA

Radiation shielding for undulator beamline in Indus-2 synchrotron radiation source

Author(s) Tatsuzawa, Ryotaro; Takaki, Naoyuki. Citation Physics Procedia (2015), 64:

The First Results of K2K long-baseline Neutrino Oscillation Experiment

Secondary Radiation and Shielding Design for Particle Therapy Facilities

A. Žukauskaitė a, R. Plukienė a, A. Plukis a, and D. Ridikas b

Neutron Metrology Activities at CIAE (2009~2010)

Shielding Design Considerations for Proton Therapy Facilities

Measurement of the Westcott Conventionality Thermal Neutron Flux and Suchlike at Irradiation Facilities of the KUR

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

Calibration of the GNU and HSREM neutron survey instruments

Transcription:

DOI: 10.15669/pnst.4.704 Progress in Nuclear Science and Technology Volume 4 (2014) pp. 704-708 ARTICLE Response measurement of various neutron dose equivalent monitors in 134-387 MeV neutron fields Yoshihiro Nakane a*, Masayuki Hagiwara b, Yosuke Iwamoto a, Hiroshi Iwase b, Daiki Satoh a, Tatsuhiko Sato a, Hiroshi Yashima c, Tetsuro Matsumoto d, Akihiko Masuda d, Tomoya Nunomiya e, Yukio Sakamoto f, Hiroshi Nakashima a, Atsushi Tamii g, Kichiji Hatanaka g and Takashi Nakamura h a Japan Atomic Energy Agency, 2-4 Shirakata, Tokai-mura, Naka-gun, Ibaraki-ken, 319-1195, Japan; b High Energy Accelerator Research Organization (KEK), 1-1 Oho, Tsukuba-shi, Ibaraki-ken, 305-0801, Japan; c Research Reactor Institute, Kyoto University, 2 Asashiro-Nishi, Kumatori-cho, Sennan-gun, Osaka, 590-0494, Japan; d National Institute of Advanced Industrial Science and Technology, 1-1-1 Higashi, Tsukuba-shi, Ibaraki-ken, 305-8561, Japan; e Fuji Electric Co., Ltd., 1 Fuji-machi, Hino-shi, Tokyo, 191-8502, Japan; f ATOX Co., Ltd., 1201 Takada, Kashiwa-shi, Chiba, 277-0861, Japan; g Research center for Nuclear Physics, 10-1 Mihogaoka, Ibaraki-shi, Osaka University, Osaka, 567-0047, Japan; h Tohoku University, 6-6 Aoba, Aramaki, Aoba-ku, Sendai, 980-8578, Japan In order to verify the accuracy of dose monitors for the measurements of high-energy neutrons, energy responses of seven neutron dose equivalent monitors, four kinds of wide-energy range monitors applied to high-energy neutron fields above 20 MeV, and three kinds of conventional type monitors for neutrons up to 20 MeV, were measured for 134-, 197-, 244- and 387-MeV quasi-monoenergetic neutron fields. Measured dose rates were compared one another, and with those in the fields obtained from neutron energy spectra and flux-to-dose conversion coefficients considered on design of the monitors. Dose rates measured with the wide-energy range monitors were in agreement or overestimate with evaluated ones, though those measured with the conventional type monitors underestimate. It was found that the wide-energy range monitors used in the present work will give the dose rates with a safety side at several hundred MeV neutron fields, though the conventional type monitors underestimate the dose rate in the fields. Keywords: high-energy; quasi-monoenergetic neutron; neutron dose; dose equivalent monitors; wide-energy range monitors; energy response 1. Introduction 1 In recent years, a lot of experiments in the field of frontier science were performed with using high energy radiations generated by high intensity accelerators. In Japan, high intensity proton accelerator facility, J-PARC[1], has been constructed for science and physics research as the joint project between the Japan Atomic Energy Agency (JAEA) and the High Energy Accelerator Research Organization (KEK). For the radiation protection in high-energy neutron fields such as found in accelerator facilities, it is important to measure the dose rates caused by wide-energy range neutrons, accurately. Especially, the dose rates caused by neutrons from 20 to several hundred MeV are important for the radiation protection in the fields because these neutrons dominantly contribute to the dose equivalent in general. In low-energy neutron fields below 20 MeV, many kinds of commercial neutron dose equivalent monitors such as Andersson- Braun type [2] dose monitors have been used widely for the measurements of neutron dose rates. The monitors, however, generally underestimate the dose caused by high-energy neutrons above 20 MeV because these monitors are designed for the measurement up to 20 MeV. Therefore, wide-energy range monitors have been developed for the improvement of dose measurements at high-energy neutron fields [3-6]. In order to verify the accuracy of dose monitors for the measurements of high-energy neutrons, energy responses were measured in the present work by seven kinds of neutron dose equivalent monitors for 134-, 197-, 244- and 387-MeV quasi-monoenergetic neutrons generated from the 7 Li(p,n) reactions at the Research Center of Nuclear Physics (RCNP) ring cyclotron of Osaka University [7]. Dose rates measured by the monitors were compared with one another, and with those in the fields obtained from neutron energy spectra and flux-to-dose conversion coefficients. *Corresponding author. Email: nakane.yoshihiro@jaea.go.jp 2014 Atomic Energy Society of Japan. All rights reserved.

Progress in Nuclear Science and Technology, Volume 4, 2014 705 2. Experiments 2.1. Experimental setup Experiments were carried out at the neutron time-of-flight (TOF) beam course in the RCNP of Osaka University. Figure 1 shows a schematic view of the experimental arrangement. The experimental room is a tunnel of 100-m in length, 4-m in width and 3.5-m in height. Quasi-monoenergetic source neutrons, the peak energies were 134-, 197-, 244- and 387-MeV, were produced in a 1.0-cm thick Li target bombarded with 137-, 200-, 246- and 389-MeV protons, respectively. Source neutrons emitted in the forward direction reached an experimental room through a 10.0-cm-wide and 12.0-cm-height aperture, and 150-cm-long iron collimator embedded in a concrete wall. The protons penetrating through the target were bent towards a beam dump by the swinger magnet to measure the proton beam intensity with a Faraday cup inside the dump. The monitors were placed on the neutron beam at the position of 35.0 m from the Li target except for 387-MeV neutron, 30.0 m for 387-MeV neutron. At the 35.0 m position, the size of collimated beam is about 60 x 70 cm 2. Since the beam size is quite small to the room size, contributions of wall-scatter components were estimated to be quite small. The energy spectra of source neutrons were shown in Figure 2. The spectra have been measured by the TOF methods with three sizes of NE213 liquid organic scintillation detectors (25.4 x 25.4 cm 2 for neutrons above 100 MeV, 12.7 x 12.7 cm 2 for neutrons from 10 to 100 MeV, and 5.08 x 5.08 cm 2 for neutrons below 10 MeV in diameter and length) [8-9]. 2.2. Neutron monitors Table 1 shows the neutron dose equivalent monitors used in the present work. Four kinds of wide-energy range monitors applicable for high-energy neutron field (three kinds of modified Andersson-Braun type monitors with lead or tungsten breeder, and a liquid organic scintillator type monitor with the response functions applied for high-energy neutron dose) were used in the measurements. In addition, three kinds of conventional Figure 1. Schematic view of the experimental arrangement at RCNP. Neutron Spectrum (n MeV -1 sr -1 μc -1 ) 10 10 10 9 10 8 134MeV 197MeV 244MeV 387MeV 10 7 0 100 200 300 400 Neutron Energy (MeV) Figure 2. Measured energy spectra of source neutrons. type monitors (two Andersson-Braun type monitors and a proportional counter type monitor) were also used for comparisons. In the table, the JAEA wide range monitor was developed [6] for applying to neutron monitoring for high-energy proton accelerator facility, J-PARC [1], and products of the monitor have been stationed in the facility for monitoring neutron dose rates. To measure the neutron dose accurately for wide energy Table 1. Neutron dose equivalent monitors used in the present work. Monitor Reproduced Data of Andersson-Braun type Wideenergy dose conversion coefficients Detector type Breeder for range Counter high-energy Shape 20MeV > 20MeV neutron JAEA wide range yes H*(10) E Modified Andersson-Braun 3 He Lead Spherical HPI* 6060 yes H*(10) H*(10) Modified Andersson-Braun BF 3 Lead Cylindrical WENDI yes H*(10) H*(10) Modified Andersson-Braun 3 He Tungsten Cylindrical DARWIN yes H*(10) H*(10) Liquid organic scintillator --- --- --- Alnor(Studsvik) 2202D no H*(10) --- Andersson-Braun BF 3 none Cylindrical Fuji Electric NSN1 no H*(10) --- Andersson-Braun 3 He none Spherical Fuji Electric NSN3 no H*(10) --- Proportional counter --- --- --- * Health Physics instruments

706 Y. Nakane et al. range, a 1.5-cm-thick lead spherical shell was inserted to a conventional Andersson-Braun type rem monitor between inner and outer polyethylene spherical shells as a breeder for high-energy neutrons, which is similar concept for other modified Andersson-Braun type monitors. The thickness and the position of the lead shell inserted the monitor were determined [6] by neutron transport calculations using MCNP-4B code [10] with LA150 library [11] that the energy response reproduces the flux-to-dose conversion coefficients. For neutron energy below 20 MeV, the monitor was designed to reproduce the coefficients for ambient dose equivalent [12-14], H*(10)/φ, same as the other monitors. On the other hand, the responses reproduce the coefficients for effective dose [15], E/φ, for the energy above 20 MeV, though the other monitors were designed to reproduce those of ambient dose equivalent, H*(10)/φ, because the monitor responses do not underestimate dose rates at high-energy neutron fields. For verifying the calculations of energy response, the responses have been measured at neutron fields from thermal to 68-MeV quasi-monoenergetic neutron sources[6]. The calculation responses were in good agreement with measured ones within 15 %, and the measured dose rates were in good agreement with evaluated ones within 10 %. 2.3. Evaluations of dose rates in the fields 2.3.1 Evaluations of dose rates Neutron dose rates in the fields have been evaluated from the measured energy spectra of source neutrons and flux-to-dose conversion coefficients. According to the design concept of the monitors, two kinds of the coefficients, H*(10) and E, were applied for neutron energy above 20 MeV in the evaluation. 2.3.2 Dose conversion coefficients In general, monitors are designed that the energy responses reproduce to the coefficients of H*(10) as operational quantities. On the other hand, the coefficients for E were applied for the shielding calculations of the facility as protection quantities. Figure 3 shows the comparisons of the coefficients [12-15]. The coefficients for H*(10) are obviously smaller than those for E for the energy above 100 MeV, only 55-60% of those for E for 200-400 MeV, while those are in agreement for the low energy region on the whole. It means that the monitors designed for reproducing the coefficients of H*(10) have a possibility of underestimating effective dose rate in high-energy neutron fields above 100 MeV. The JAEA wide range Table 2. Calculated neutron dose rates at the position of 35.0 m from the Li target. Case of calculation Applied flux-to-dose conversion coefficients Dose Conversion Coefficients (psv cm 2 ) 10 4 10 3 10 2 10 1 Effective Dose (AESJ) Effective Dose (ICRP-74 (AP)) H*(10) (ICRP-74) H*(10) (Sato) H*(10) (FLUKA) 10 0 10-8 10-6 10-4 10-2 10 0 10 2 10 4 Energy (MeV) Figure 3. Flux-to-dose conversion coefficients for effective dose and ambient dose equivalent. monitor was designed to avoid this underestimation. For the comparisons of energy regions, both the coefficients above 1 MeV are several tens of times larger than those below 10 kev. This means that the contribution of low-energy neutrons is estimated to be quite smaller in dose evaluation than that of high-energy neutrons. 2.3. Evaluated results of dose rates Table 2 shows the evaluated dose rates per proton beam intensity(coulomb) at the monitor irradiation position, 35.0 m distance from Li target. The results show that dose rates applied for H*(10)+E are 44-65% larger than the those for H*(10)+H*(10) in the field because of the differences between H*(10) and E for the energy above 100 MeV shown in Figure 3. 3. Results and discussions 3.1. Comparison of measured results Measured results of monitor responses are listed in Table 3. The results for the 387-MeV source are normalized to those at the position of 35.0 m from the Li target. The measured results of all the conventional type monitors are smaller than those of all the wide-energy range monitors. Especially, the results of Alnor 2202D and NSN1 are about an order smaller than those of JAEA wide range. For the comparisons between wide-energy range Calculated dose rates per proton beam intensity(sv C -1 ) at the position of 35.0 m from the Li target En 20 MeV En > 20 MeV 134 MeV 197 MeV 244 MeV 387 MeV H*(10) + H*(10) H*(10) /φ H*(10) /φ 0.665 0.592 0.545 0.699 H*(10) + E H*(10) /φ E /φ 0.961 0.952 0.885 1.154

Progress in Nuclear Science and Technology, Volume 4, 2014 Table 3. Summary of measured responses and the ratios of evaluated dose rates. Monitor Measured dose rates per beam intensity(sv C -1 ) at the position of 35.0 m from the Li target 134 MeV 197 MeV 244 MeV 387 MeV * 134 MeV 197 MeV 244 MeV 387 MeV * JAEA wide range 1.11 1.17 1.32 1.74 (1.15) (1.23) (1.49) (1.51) HPI 6060 --- --- 0.621 0.855 --- --- 1.14 1.22 DARWIN --- --- 0.725 0.725 --- --- 1.33 1.04 WENDI --- --- 0.737 1.115 --- --- 1.35 1.60 Alnor 2202D --- --- 0.137 0.163 --- --- 0.252 0.233 Fuji NSN1 --- --- 0.116 --- --- --- 0.212 --- Fuji NSN3 0.513 0.414 0.364 --- 0.771 0.699 0.668 --- * Results of 387 MeV were normalized to those at the position of 35.0 m from the Li target. Ratio of measured and evaluated dose rates applied for H*(10) + H*(10) (for H*(10)+E ) 707 monitors, the results of HPI 6060, DARWIN and WENDI are in agreements one another. The results of JAEA wide range was 1.5-2.4 times larger than those of others. It was because the differences of reproducing the coefficients on the design. As shown in figure 3, the coefficients for E are about twice larger than those for H*(10) for the energy above 200 MeV. For the comparisons between conventional type monitors, measured result of Alnor 2202D is in agreement with that of NSN1 for the 244-MeV source. On the other hand, the result of NSN3 is about 3 times larger than those of Alnor 2202D and NSN1, and close to the results of wide-energy range monitors. This agreement and disagreement is ascribed to be the difference of detector types, as shown in Table 1. 3.2. Comparison of measured to calculated results Measured dose rates were compared with calculated ones obtained from the coefficients considered on design. The ratios of measured and evaluated dose rates are listed in Table 3, and shown in Figure 4. The results of all the monitors except JAEA wide range were compared with those for H*(10), though those of JAEA wide range were compared with those for H*(10)+E. The measured results with JAEA wide range were in agreement with evaluated ones within 23% for the 134- and the 197-MeV sources. Although the results of Ratio of measured/evaluated dose rates 2 1 JAEA wide range HPI 6060 DARWIN WENDI Alnor 2202D Fuji NSN1 Fuji NSN3 0 100 200 300 400 Neutron Energy (MeV) Figure 4. Ratios of measured and evaluated dose rates. NSN3 were in agreement with evaluated ones within 31%, it was a problem for dose monitor to underestimate dose rate at the fields. For the 244- and the 387-MeV sources, the results of JAEA wide range and WENDI tend to overestimate the evaluated ones, though the results of HPI and DARWIN were in agreement on the whole with the evaluated ones. On the other hand, the results of all the conventional type monitors underestimate the evaluated ones. Especially, the results of Alnor 2202D and NSN1 (conventional Andersson-Braun type monitors) are remarkably underestimate, only 21-25% of the evaluated ones. 4. Conclusion The energy responses for seven kinds of neutron dose equivalent monitors were measured for 134-387 MeV quasi-monoenergetic neutrons. Dose rates measured with four wide-energy range monitors are in agreement or overestimate with evaluated ones in the fields. On the other hand, all the conventional type monitors underestimate the evaluated ones, especially the results of conventional Andersson-Braun type monitors show only 21-25%. It is concluded that the wide-energy range monitors used in the present work will give the dose rates with a safety side at several hundred MeV neutron fields, though the conventional type monitors underestimate the dose rate in the fields. References [1] S. Tanaka High intensity proton accelerator project in Japan (J-PARC), Radiat. Protec. Dosim. 115 (2005), pp. 33-43. [2] I.O. Andersson and J. Braun, A neutron rem counter with uniform sensitivity from 0.025 ev to 10 MeV, Proc. of the IAEA Symposium on Neutron Detection, Dosimetry and Standardization, Harwell, England, Dec. 10-14, 1962, (IAEA:Vienna) Neutron Dosimetry Vol. II (1963), pp.87-95. [3] C. Birattari, A. Ferrari, C. Nuccetelli, M. Pelliccioni and M. Silari An extended range neutron rem counter, Nucl. Instrum. Methods Phys. Res. A297 (1990), pp. 250-257. [4] L. Jianping, T. Yueli, L. Shudong, S. Ban, T.

708 Y. Nakane et al. Suzuki, K. Iijima and H. Nakamura, Neutron energy response of a modified Andersson-Braun rem counter, Radiat. Protec. Dosim. 67 (1996), pp. 179-185. [5] R. H. Olsher, H.-H. Hsu, A. Beverding, J. H. Kleck, W. H. Casson, D. G. Vasilik and R. T. Devine, WENDI: An improved neutron rem meter, Health Physics 79(2) (2000), pp. 170-181. [6] Y. Nakane, Y. Harada, Y. Sakamoto, T. Oguri, M. Yoshizawa, F. Takahashi, T. Ishikura, T. Fujimoto, S. Tanaka and N. Sasamoto, Evaluation of Energy Response of Neutron Rem Monitor Applied to High-energy Accelerator Facilities, JAERI-Tech 2003-011, Japan Atomic Energy Research Institute, (2003). [in Japanese] [7] H. Sakai, H. Okamura, H. Otsu, T. Wakasa, S. Ishida, N. Sakamoto, T. Uesaka, Y. Satou, S. Fujita and K. Hatanaka, Facility for the (p,n) polarization transfer measurement, Nucl. Instrum. Methods Phys. Res. A369 (1996), pp. 120-134. [8] Y. Iwamoto, M. Hagiwara, D. Satoh, H. Iwase, H. Yashima, T. Itoga, T. Sato, Y. Nakane, H. Nakashima, Y. Sakamoto, T. Matsumoto, A. Masuda, J. Nishiyama, A. Tamii, K. Hatanaka, C. Theis, E. Feldbaumer, L. Jaegerhofer, C. Pioch, V. Mares and T. Nakamura, Quasi-monoenergetic neutron energy spectra for 246 and 389 MeV 7 Li(p,n) reactions at angles from 0 to 30 degree, Nucl. Instrum. Methods Phys. Res. A629 (2011), pp. 43-49. [9] Y. Iwamoto, M. Hagiwara, H. Iwase, H. Yashima, D. Satoh, T. Matsumoto, A. Masuda, C. Pioch, V. Mares, T. Shima, A. Tamii and K. Hatanaka, Characterization of quasi-monoenergetic neutron source using 137, 200, 246 and 389 MeV 7 Li(p,n) reactions, Proc. ICRS-12&RPSD-2012, Nara, Japan, Sept. 2-7 (2012) [to be published]. [10] J.F. Briesmeister (Editor), MCNP - A General Monte Carlo n-particle Transport Code, Version 4B, LA-12625-M, Los Alamos National Laboratory, (1997). [11] M.B. Chadwick, P.G. Young, S. Chiba, S.C. Frankle, G.M. Hale, H.G. Hughes, A.J. Koning, R.C. Little, R.E. MacFarlane, R.E. Prael and L.S. Waters, Cross section evaluations to 150 MeV for accelerator-driven systems and implementation in MCNPX, Nucl. Sci. Eng. 131(3) (1999), pp. 293-328. [12] A. Ferrari, M. Pelliccioni and M. Pillon, Fluence to effective dose conversion coefficients for neutrons up to 10 TeV, Radiat. Protec. Dosim. 71(3) (1997), pp. 165-173. [13] A. Ferrari and M. Pelliccioni Fluence to dose equivalent conversion data and effective quality factors for high energy neutrons, Radiat. Protec. Dosim. 76(4) (1998), pp. 215-224. [14] T. Sato, A. Endo, M. Zankl, N. Petoussi-Henss, H. Yasuda and K. Niita, Fluence-to-dose conversion coefficients for aircrew dosimetry based on the new ICRP Recommendations, Progress Nucl. Sci. Technol. 1 (2011), pp. 134-137. [15] Radiation Dose Conversion Coefficients for Radiation Shielding Calculations: 2004, AESJ-SC-R002:2004, Atomic Energy Society of Japan, (2004). [in Japanese]