In-Tube Extraction Sample Preparation for Gas Chromatography

Similar documents
Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Maximizing Sample Throughput In Purge And Trap Analysis

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Methanol Extraction of high level soil samples by USEPA Method 8260C

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Optimizing. Abstract: is standardd. procedures. altered to

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

EPA TO-17 Volatile Organic Compounds

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Ed George and Anaïs Viven Varian, Inc.

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

CALA Directory of Laboratories

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Thermal Desorption Technical Support

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Optimizing Standard Preparation

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

Maximizing Production While Minimizing Costs

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Headspace Technology for GC and GC/MS: Features, benefits & applications

Determination of Volatile Organic Compounds in Air

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Analytical Trap Comparison for USEPA Method 8260C

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Study of Residual Solvents in Various Matrices by Static Headspace

2017 EPA Method Update Rule and EPA Method 624.1

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

COMBI. Gaschromatography. GC / GC-MS sample injector grows with your needs

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

Analysis of BTEX in Natural Water with SPME

Application. Gas Chromatography February Introduction

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Target Compound Results Summary

Volatile Organic Compounds in Water PBM

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

ChromTech options for PAL systems

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

STANDARD OPERATING PROCEDURES

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Associates. Project No.: October 25, 2017

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

AUTOMATED HEADSPACE. Headspace Principles. Graham Broadway. Headspace Gas Chromatography Background to the HS Technique

C146-E209. Headspace Samplers. HS-20 Series

Date Issued: April 01, 2013 Expiration Date: June 30, 2013

Meeting NJ Low Level TO-15 Air Testing Method Requirements

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

PAL SPME Arrow The Better SPME

SELECTED ION FLOW TUBE MASS SPECTROMETRY. Interscience Expert Center: VIP Meeting 15 Sept. 2009

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

GC Application Note. Determination of C2-C12 aldehydes in water by SPME on-fiber derivatization.

Transcription:

In-Tube Extraction Sample Preparation for Gas Chromatography CTC Analytics AG Industriestrasse 20 CH-4222 Zwingen Switzerland Tel: +41 61 7658100 Fax: +41 61 7658199 E-mail: support@ctc.ch Web: www.ctc.ch

In-Tube Extraction ITEX, a new Sample Preparation Technique for Gas Chromatography Bruno Baltensperger CTC Analytics AG, Industriestrasse 20, CH-4222 Zwingen, Switzerland Beat Schilling BGB Analytik AG, Lettenstrasse 97, CH-8134 Adliswil, Switzerland Range of possible applications Environmental Drinking Water Foods / Flavours Consumer Products Forensics / Toxicology Petrochemicals Polymers Pharmaceuticals Residual Solvents Rapid and efficient sample enrichment of volatile and semi-volatile compounds in solid, liquid and gaseous samples In-tube extraction and direct thermal desorption using proven industry standard adsorbents Syringe only concept for transparent sample handling, no sample loops, transfer lines, or switching valves No GC injector modifications, no cryo-focussing necessary Top mounted on GC's, saves valuable bench space Interfaces with any CombiPAL System controlled by all major GC/GC-MS Systems

ITEX Sample Extraction Procedure Sample is heated and / or agitated in a sealed sample vial until equilibrium is achieved The ITEX Trap needle pierces the sealed sample vial and the heated gas tight syringe pumps the headspace through the ITEX Trap The loaded ITEX Trap is flash heated up to 3 C and desorbed into the hot GC Injector After thermal desorption the hot ITEX Trap is cleaned with inert flush gas Sample conditioning Adsorption Desorption Trap cleaning

Enrichment Relationship between Sample Volume & Sensitivity 3 60 x 40 x 300 2 200 1 20 x 10 x 100 0 0 x Minutes 1 2 3 4 5 6 7 8 9 10 11 12 Enrichment of Methylesters on a Tenax TA ITEX trap. 1µl of a mixture of C4, C6, C8, C10 Methylesters in Methanol (100ng/µl) was injected into a 20ml HS-vial. After conditioning at 40 C for 10min. 1ml of the headspace using an ITEX trap without packing material was injected to determine a "static headspace value" (0x). Afterwards the needle was replaced by the TENAX TA ITEX trap. The enrichment of the solutes on the trap was studied using various numbers of pumping strokes. ITEX Parameter: Extraction Speed: 100µl/sec. Extraction Strokes: 0 / 10 / 20 / 40 / 60 Temperature Pumping Syringe / Sample Incubation: 40 C/10min. Desorption at 2 C, 15sec. splitless Chromatography: Injection: Splitless 15sec. at 2 C, Carrier gas: 0.2bar Hydrogen Column: BGB-1 15m x 0.32mm ID, 1.0µm film Temperature Program: 40 C - 1min. - 10 C/min to 200 C Detection: FID 2 C

Environmental Application EPA Method 2.2 ITEX 39 Headspace 1 2 3 4 56 7 8 12 13 10 9 11 14 15 16 17 18 19 20 21 22 23 24 25 28 29 27 30 31 32 33 26 37 38 34 3536 40 41 42 43 44 48 454647 49 52 51 53 54 Static Headspace Parameter 60 C / 10min / 1ml sample volume ITEX Parameter: Extraction Speed:100µl / sec. Total Pumping Strokes: Temperature Pumping Syringe / Sample Incubation: 60 C -10min. Desorption at 200 C, 15sec. splitless Comparison of ITEX analysis versus Static Headspace Sample: Purge and Trap calibration mix (Restek Cat.No. 30431 2.2 CAL2000 Mega-Mix) Chromatography: Injection: Splitless 15sec. at 2 C / Carrier gas: 0.2bar hydrogen Column: Rtx-2.2 60m x 0.32mm ID, 1.8µm film Temperature Program: 40 C - 1min. - 10 C - min to 220 C Detection: FID 2 C 1 1,1-Dichloroethylene 2 Methylene chloride (dichloromethane) 3 trans 1,2-Dichloroethylene 4 1,1-Dichloroethane 5 2,2-Dichloropropane 6 cis-1,2-dichloroethylene 7 Chloroform 8 Bromochloromethane 9 1,1,1-Trichloroethane 10 1,1-Dichloropropene 11 Carbon tetrachloride 12 1,2-Dichloroethane 13 Benzene 14 Trichloroethylene 15 1,2-Dichloropropane 16 Bromodichloromethane 17 Dibromomethane 18 cis-1,3-dichloropropylene 19 Toluene 20 trans-1,3-dichloropropylene 21 1,1,2-Trichloroethane 22 1,3-Dichloropropane 23 Tetrachloroethylene 24 Dibromochloromethane 25 1,2-Dibromoethane (EDB) 26 Chlorobenzene 27 1,1,1,2-Tetrachloroethane 28 Ethylbenzene 29 m-xylene 30 p-xylene 31 o-xylene 32 Styrene 33 Isopropylbenzene 34 Bromoform 35 1,1,2,2-Tetrachloroethane 36 1,2,3-Trichloropropane 37 n-propylbenzene 38 Bromobenzene 39 1,3,5-Trimethylbenzene 40 2-Chlorotoluene 41 4-Chlorotoluene 42 tert-butylbenzene 43 1,2,4-Trimethylbenzene 44 sec-butylbenzene 45 4-Isopropyloluene (p-cymene) 46 1,3-Dichlorobenzene 47 1,4-Dichlorobenzene 48 n-butylbenzene 49 1,2-Dichlorobenzene 1,2-Dibromo-3-chloropropane 51 1,2,3-Trichlorobenzene 52 Hexachloro-1,3-butadiene (Hexachlorobutadiene) 53 Naphthalene 54 1,2,3-Trichlorobenzene

Comparison Direct Injection vs ITEX Diesel Oil 40 Direct Injection ITEX 30 20 10 0 Minutes 2 4 6 8 10 12 14 16 18 20 22 1µl of a Diesel dissolved in Methanol (0ng / µl) was injected (splitless 15sec.) into the injector to determine a "100% value". 1µl of the same solution was added to 12ml water in a 20ml Headspace vial and then analysed with ITEX. ITEX Parameter: Extraction Speed: 120µl/sec. Extraction Strokes: 120 Temperature Pumping Syringe / Sample Incubation: C - 10min. Desorption at 2 C, 15sec. splitless Chromatography: Injection: Splitless 15sec. at 2 C / Carrier gas: 0.2bar Hydrogen Column: BGB-1 15m x 0.32mm ID, 1.0µm film Temperature Program: 40 C - 1min. - 10 C /min to 200 C Detection: FID 2 C

Environmental Application VOC, BTEX in Water ppt Level 100 90 80 SIM @ Mass=91 (ng/l) Ion Trap MS m-xylene p-xylene 70 60 Toluene Ethylbenzene o-xylene 40 30 20 10 0 Retention Time 11.40 12.00 12.20 12.40 13.00 13.20 13.40 14.00 14.20 14.40 15.00 Chromatogram shows BTEX Compounds at a concentration of ng/l using 20 Extraction strokes Monitoring of BTEX and VOC Compounds in Water (EPA Method 2.2) Key Words: VOC, BTEX, EPA Method 2.2, ITEX BTEX and VOC Compounds according to EPA Method 2.2 are analysed using ITEX sample preparation technique. Total sample preparation time of less than 15 minutes allows a high sample throughput. Sample Preparation: 10ml water is filled into 20ml Headspace sample vials. 3g Sodium chloride and 1µl of the internal standard VOC (ppb Fluorobenzene in Ethanol) is added. After sample conditioning at 60 C during 10 minutes 20 strokes of the headspace are pumped through the ITEXtrap with a velocity of 100µl/sec. The resulting sensitivity is sufficient to obtain the requested detection limit for drinking water of 0.05µg/l. ITEX Conditions: Sample Conditioning @ 60 C, 10 min. Extraction Strokes: 20 x 1ml Desorption @ 230 C with 1.3ml Headspace 20µl/sec. Chromatography: Column: Rtx-2.2, 60m x 0.32mm, 1.8µm film Carrier Gas: Helium 20psi Temperature Program: 40 C - 2 min.- to 240 C - 2 min. at 10 C / min. Precolumn: 1m x 0.32mm deactivated with DPTMDS Injector: Gerstel KAS3 with septa @ 1 C isothermal GC: Varian 3300 Detector: Varian Saturn 4D GC/MS/MS Courtesy of: Labor Veritas, Engimattstrasse11, CH-8027 Zürich, Natascha Meggiolaro

Perfume Application Softener-Comparison Porapak Filter Technique vs. ITEX 100 17.66 RT Nr Name 8.79 1 Limonene 9.09 2 Eucalyptol 8.79 11.94 3 Methylhexylketone 90 13.2 4 cis-3-hexenylacetate ITEX 18.37 14.37 5 Rose oxide 3 gr sample, 30 minutes 17.66 6 Cyclal C 80 equilibration time, 40 C 17.85 7 Tetrahydrolinalool 22.56 26.94 18.37 8 Tetrahydromyrcenol 22.27 9 Linalool 70 22.56 10 Verdox 25.21/26.94 11 Vertenex 26.49 12 Citronellylacetate 60 26.59 13 Pelargol 27.75 14 Undecavertol 28.04 15 Gardenol Porapak 29.02 16 Benzylacetate Filter Technique 30.45 17 DMBC acetate 5 gr sample, 20 hours 31.33 18 Allylcyclohexylpropionate equilibration time, 25 C 32.09 19 Phenylethylacetate 40 13.20 25.21 29.02 33.04 20 alpha-methylionone 33.04 34.31 21 Verdylacetate 35.32 22 n-methyl-alpha-ionone 30 36.61 23 Verdylpropionate 38.56 24 Diphenyloxide 39.67 25 Lilial 11.94 39.91 26 Iso E super 49.15 27 alpha-hexylcinnamic aldehyde 20 31.33 22.27 23.91 36.61 39.91 10 14.37 9.09 35.32 19.47 39.67 5.51 7.78 40.48 45.30 46.21 10.41 14.92 49.15 0 Minutes 5 10 15 20 25 30 35 40 45 Chromatography (Thermo TraceGC): Column Stabilwax 30m x 0.25mmID x 0.25 mm Oven 35 C - 0.5min. 15 C / min. C - 0min. 5 C / min. 220 C / 1min SSL splitless with surge, surge pressure 20kPa/0.4min (0.5ml), split flow 100ml /0.3min Carrier He, 1ml constant flow with vacuum compensation MS conditions (Thermo Trace MS system): Ionisation mode EI+ Source temperature 230 C Interface 220 C Mass 20-3 amu Courtesy of: Givaudan Research Company, CH-8600 Dübendorf, Zürich. Switzerland, H. Koch ITEX conditions: Incubation temp 40 C Incubation time 30 min Syringe temp 45 C Extraction volume 1000 ml per stroke Extraction strokes 10 Extraction speed 100 ml/sec Desorption temp 200 C Desorption speed 100 ml/sec Flush time 5 min

Specifications ITEX adsorption step out of a sample vial 2.5ml with 1/4" 28 UNF fitting Extraction Speed: Selectable from 10µl / sec. up to 1000µl / sec. Extraction Strokes: Selectable from 1-999 Extraction Volume: Selectable from 2µl - 20µl / 1 stroke Desorption Temperature: +5 C above ambient - 3 C selectable in 1 C increments Desorption Time: 0-300 seconds Pumping Syringe and Trap Cleaning: Inert gas purging, 30sec. - 600sec. Heated Pumping Syringe: +5 C above ambient - 1 C selectable in 1 C increments Incubator Oven: 6 heated vial positions for 2ml / 10ml / 20ml vials +5 C above ambient - 200 C selectable in 1 C increments Agitation: Interval shaking 2rpm-7rpm, selectable in 1rpm increments Incubation Time: Up to 999 minutes selectable in 1 second increments Conclusion This new approach, ITEX or In-Tube Extraction, an automated sample adsorption and thermal desorption technique shows a great potential for volatile compounds in various matrices. The adsorption at the trap does continuously reduce the concentration of analytes in the headspace and disturbs the equilibrium in the vial. The result is a pumping effect of the analytes from the sample matrix to the headspace. This effect is similar as used by the SPME technique but the active control over the pumping strokes allows the user to utilize rather a dynamic headspace as known by the Purge&Trap technique. Using the flush gas and a user selectable trap heating temperature does allow to clean the trap very efficiently. Blank runs, even after several hundred injections, do not show a carry over effect. We are aware that this statement can not be generalized. It depends on the sample and matrices. The technique is still too young to be able to give a general statement The sensitivity level to be reached can be as far down as the ppt level.